of 38
7/30/2019 QC in the Lab.pps
1/38
Quality Controlin the
Pathology Laboratory
John Santangelo
7/30/2019 QC in the Lab.pps
2/38
Quality Assurance
and
Quality Control
7/30/2019 QC in the Lab.pps
3/38
Quality assurance in pathology
and laboratory medicine is the
practice of assessing
performance in all steps of thelaboratory testing cycle.
including pre-analytic, analytic,and post-analytic phases.
7/30/2019 QC in the Lab.pps
4/38
Quality control is an integral
component of quality assurance andis the sum of processes and
techniques to --
detect, reduce, and correct
deficiencies in an analytical process.
e.g. designing new and bettermethods for analysing blood or
serum.
7/30/2019 QC in the Lab.pps
5/38
Quality improvement is the practice of continuously
assessing and adjusting performance using statistically
and scientifically accepted procedures.
Preventive Maintenance of Equipment
Continual monitoring of the temperature of water baths
and refrigerators is important to the maintenance of
reagent quality and test performance.
Equipment such as microscopes, centrifuges, and
spectrophotometers should be cleaned and checked
for accuracy on a regular schedule.
Failure to monitor equipment regularly can produce
inaccurate test results and lead to expensive repairs.
7/30/2019 QC in the Lab.pps
6/38
Factors in Quality AssuranceTo guarantee the highest quality patient care through laboratory testing, a
variety ofpre-analytical, analytical and post-analytical factors in addition to
analytical data must be considered.
Pre-analytical, analytical and post-analytical factors are all part of Quality
Control
Non-analytical factors that support quality testing include the following:
Qualified Personnel
Established laboratory policies
The laboratory procedure manual
Proper procedures for specimen collection and storage
Preventive maintenance of equipment
Appropriate methodology
Established quality assurance techniques
Accuracy in reporting results
7/30/2019 QC in the Lab.pps
7/38
Pre Analytical
Proper Procedures for Specimen Collection and Storage
Strict adherence to correct procedures for specimen collectionand storage is critical to the accuracy of any test.
Pre-analytical errors are the most common
source of laboratory error.
The correct patient identification is critical.
Date of Birth. (careful with overseas people as Month & daymight be reversed)
Ask patient to spell their name.
7/30/2019 QC in the Lab.pps
8/38
Pre Analytical (cont)
Each blood specimen obtained should be labeled with the
patients first and last name, hospital identification number,patient location, time, date and the phlebotomists initials.
Correct storage of specimens is critical to obtaining accurateresults.
Some specimens are stored at room temperature and somein the refrigerator.
Specimen integrity is an important issue when blood iscollected at a site away from the testing facility.
Blood must be centrifuged within 30 minutes ofcollection.
Delayed centrifugation will result in wrong results.
7/30/2019 QC in the Lab.pps
9/38
Pre Analytical (cont)
Specimens are to be transported to the laboratory in the
appropriate containers, e.g. coolers.
Physiologic Factors Affecting Test Results
Posture- Changing from a supine to a sitting or standingposition results in a shift of body water from inside the blood
vessels to the interstitial spaces.
Larger molecules cannot filter into the tissues and concentratein the blood. There will be significant increases in test values
for lipids, enzymes and proteins.
7/30/2019 QC in the Lab.pps
10/38
Pre Analytical (cont)
Diurnal Rhythm- Diurnal pertains to daylight, and diurnal
rhythm refers to the daily body fluctuations that occur.
Certain hormone levels such as cortisol and adenocorticotropichormones, decrease in the afternoon.
Other test values, such as iron and eosinophils, increase in the
afternoon.
7/30/2019 QC in the Lab.pps
11/38
Pre Analytical (cont)
Exercise- muscle activity elevates creatinine, protein, creatine
kinase, aspartate transaminase, and lactate dehydrogenase
values.
Research also suggests that exercise activates coagulation
and fibrinolysis and increase platelet counts.
Stress- Anxiety can cause temporary increase in white blood
cells, catecholamines as well as an acid-base imbalance.
7/30/2019 QC in the Lab.pps
12/38
Pre Analytical (cont)
Dieting- Fasting means no food or beverages except
water for 8-12 hours prior to blood collection.
If a patient has recently eaten, there will be atemporary increase in glucose and lipid content in the
blood.
As a result, the serum or plasma may appear cloudyor turbid, which interferes with testing, especially with
tests such as glucose, sodium, and complete blood
counts.
7/30/2019 QC in the Lab.pps
13/38
Pre Analytical (cont)
Smoking- Patients who smoke prior to blood collection mayhave increased WBC counts and cortisol levels.
Long-term smoking can lead to decreased pulmonary function
and result in increased hemoglobin levels. (secondarypolycythaemia)
7/30/2019 QC in the Lab.pps
14/38
7/30/2019 QC in the Lab.pps
15/38
Analytical
Appropriate Methodology
When new methods are introduced, it is important to
check the procedure for accuracy and variability.
Replicate analyses using control specimens are
recommended to check for accuracy and to eliminate
factors such as day-to-day variability, reagent
variability and differences between technologists.
7/30/2019 QC in the Lab.pps
16/38
Analytical (cont)
Established Quality Assurance Techniques
Each procedure should have an established protocol toassure the quality of the results.
Usually, the normal and abnormal control samples are
analyzed at the same time patient specimens are analyzed.
Established limits of acceptable performance must bedetermined for each type of test.
If control results are not within acceptable limits, patient
results cannot be guaranteed to be accurate.
In these cases, the source of error must be identified and the
entire test repeated before a patients result can be reported.
7/30/2019 QC in the Lab.pps
17/38
Analytical (cont)
Accuracy in Reporting Results
When extremely abnormal results or differences from
previous test values are found, the laboratory
protocol should establish the method of rechecking
and reporting such results.
7/30/2019 QC in the Lab.pps
18/38
Things that go wrong during analysis
1. Equipment faults
2. Reagents expired
3. Reagents contaminated
4. Reagents not stored under the right conditions
5. Temperatures not checked
6. Automatic pipettes not maintained
7. Dirty glassware
8. Bad operator technique
7/30/2019 QC in the Lab.pps
19/38
Qualified Personnel
The entry-level examination competencies of all certifiedpersons in hematology must be validated. Validationtakes in form of both external certification and newemployee orientation to the work environment.Continuing competency is equally important.
Established Laboratory Policies Laboratory policies should be included in the laboratory
reference manual that is available to all hospitalpersonnel.
The Laboratory Procedure Manual
Laboratory procedures should be included in the manual.
7/30/2019 QC in the Lab.pps
20/38
Batch:
A batch or lot is a collection of products all identical in size,
type, conditions and time of production.
All controls have a Batch Number and expiry date so that a
fault can be traced easily.
The number of tests between controls.
Drift:
The control results are steadily increasing or decreasing away
from the true published mean value found in the packageinsert.
7/30/2019 QC in the Lab.pps
21/38
Interference:
Something outside or inside the autoanalyser causing
unreliable control and test results.e.g. electrical interference, a light source losing brilliance, dirt or
dust etc. Insects.
Interpolation:
Someone changing a test or control result because it doesnt
look or feel right.
Random error:
An unexpected and unacceptable test or control result whichdisappears when repeated.
Fibrin blocking the autoanalyser.
7/30/2019 QC in the Lab.pps
22/38
Standard Deviation: (SD)Any control results greater than 2 standard deviation are
flagged and the analyser usually stops.
Calibrator: (standard):
A know value supplied in the package insert to calculate the
unknown value.
7/30/2019 QC in the Lab.pps
23/38
Controls:
There are Internal and External controls.
Internal controls are reconstituted and assayed.The values are given in the insert package.
External controls are sent from an outside national or
international laboratory.
The results are not supplied until after they receive your results.
This way cheating is eliminated.
Controls above and below normal reference ranges should be
performed.
Important when monitoring therapeutic Drugs.
7/30/2019 QC in the Lab.pps
24/38
Bias:
Bias is a problem and is widespread when doing internal controls.
This is why external controls are done in parallel.
Examples of bias:
1. All controls should be done with the normal run, not separately and not in
duplicate.
Often the laboratory staff will do controls separately and several times until
they get the answer they want. This is incorrect and cheating.
Controls are run to detect problems with equipment and technique.
2. Large batches should have a control about every 10 to 15 samples not at
the beginning and end of a sample run.
3. The external and internal controls should not be tested separately or in
triplicate.
4. Cheating.
7/30/2019 QC in the Lab.pps
25/38
NATA
National Association of Testing Authority
RCPA
Royal College of Pathologists of Australia
QAP
Quality Assurance Programs
Organizations and programs that monitor
Laboratories
7/30/2019 QC in the Lab.pps
26/38
Book are kept to monitor refrigerator temperatures, water
baths, autoclaves etc, and faults that occur and how they
were rectified.
This is required by NATA.
NATA inspections are done about every three years.
Any problems have to be fixed before NATA gives approval to
continue to operate as a laboratory.
Many laboratories have been closed down for not complying
with the standards required by NATA.
7/30/2019 QC in the Lab.pps
27/38
Conclusion (Quality Assurance QA)
QA is the sum of all those activities in which the
laboratory is engaged to ensure that informationgenerated by laboratory is correct.
QA includes all aspects of laboratory activities that
affects the results produced, from the choice of
methods, to the education of personnel, to the
handling of specimens and reporting results.
The real purpose of QA activities is to determine how
correct or incorrect the results coming from the labare, and to allow those managing the lab to determine
whether or not the lab is fulfilling its functions
satisfactorily.
7/30/2019 QC in the Lab.pps
28/38
3 major activities of QA :
1) Preventive those activities that are done prior to the
examination of the specimen or sample and that are intendedto establish systems conducive to accuracy testing ( eg :
preventive maintenance and calibration of instruments, testing
of media, orientation and training of personnel )
2) Assessment those activities that are done during testingto determine whether the test systems are performing correctly
( eg : the use of standard and controls, maintenance of control
charts )
3) Corrective those activities that are done, when error isdetected, to correct the system ( eg : equipment
troubleshooting, recalibration of instruments )
7/30/2019 QC in the Lab.pps
29/38
QA in Haematology Laboratory
QA in haematology lab is intended to ensure the reliability ofthe lab tests.
The objective is to achieve precision and accuracy
4 components of QA programme :
1) Internal Quality Control ( IQC )
2) External Quality Control ( EQC )
7/30/2019 QC in the Lab.pps
30/38
Accuracy
- the closeness of the estimated value to the true value.
- can be checked by the use of reference materials which have
been assayed by independent methods of known precision
Precision
- reproducibility of a results, whether accurate or inaccurate
within a define frame time ( eg: within the same day, from week
to week etc )
- can be controlled by replicate tests, check tests on previously
measured specimens and statistical evaluation of results
7/30/2019 QC in the Lab.pps
31/38
Internal Quality Control ( IQC )
Based on monitoring the Haematology tests procedures that are performed
in the lab.
includes measurements on CONTROLS & is intended to ensure that thereis continual evaluation of the reliability of the work of the lab and that
control is exercised over the release of the results.
External Quality Control ( EQC )
The objective evaluation by an outside agency of the performance by a
number of laboratories on material which is supplied specially for the
purpose.
is usually organized on a national or regional basis.
analysis of performance is retrospective.
the objective is to achieve comparability with results of other labs.
7/30/2019 QC in the Lab.pps
32/38
Control materials
Specially preparedIt may be anticoagulated whole blood, preserved pooled red
cells, plasma or serum.
It can be used to check for accuracy if the value has been
reliably determined ( eg : reference centre )
Should have controls of high, normal and low values
At least 1 control specimen should be used for every batch.
If large specimens, use 1 control for every 20 specimens.
7/30/2019 QC in the Lab.pps
33/38
Analysis of data
Standard deviation of control specimensDispersion of results around the mean will indicate the error of
reproducibility.
95% of results on the same specimen should be within 2 SD
and 99.7% within 3 SD.
by chance, 1 in 20 of measurement might expected to fall
outside 2 SD and only 1 in 333 outside 3 SD.
If measurement more widely dispersed, this indicates an errorin the test.
7/30/2019 QC in the Lab.pps
34/38
Control Charts
Originally described by Shewhart, 1st applied in clinical chemistry by Levey
and Jennings.
Samples of the control specimen are included in every batch of patients
specimens and the results checked on a control chart.
To check precision, it is not necessary to know the exact value of the
control specimen.
Value has been determined reliably by a reference method, the same
material can be used to check accuracy or to calibrate an instrument.
If possible, controls with high, low and normal values should be used.
Advisable to use at least one control sample per batch even if the batch is
very small.
The results obtained with the control samples can be plotted on a chart.
7/30/2019 QC in the Lab.pps
35/38
7/30/2019 QC in the Lab.pps
36/38
7/30/2019 QC in the Lab.pps
37/38
A minimum of 100 patients are selected to establish a
7/30/2019 QC in the Lab.pps
38/38
Total = 95.4%
A minimum of 100 patients are selected to establish a
Normal Reference Range
Within 2 Standard Deviations