+ All Categories
Home > Documents > QC in the Lab.pps

QC in the Lab.pps

Date post: 14-Apr-2018
Category:
Upload: saint5470
View: 217 times
Download: 1 times
Share this document with a friend

of 38

Transcript
  • 7/30/2019 QC in the Lab.pps

    1/38

    Quality Controlin the

    Pathology Laboratory

    John Santangelo

  • 7/30/2019 QC in the Lab.pps

    2/38

    Quality Assurance

    and

    Quality Control

  • 7/30/2019 QC in the Lab.pps

    3/38

    Quality assurance in pathology

    and laboratory medicine is the

    practice of assessing

    performance in all steps of thelaboratory testing cycle.

    including pre-analytic, analytic,and post-analytic phases.

  • 7/30/2019 QC in the Lab.pps

    4/38

    Quality control is an integral

    component of quality assurance andis the sum of processes and

    techniques to --

    detect, reduce, and correct

    deficiencies in an analytical process.

    e.g. designing new and bettermethods for analysing blood or

    serum.

  • 7/30/2019 QC in the Lab.pps

    5/38

    Quality improvement is the practice of continuously

    assessing and adjusting performance using statistically

    and scientifically accepted procedures.

    Preventive Maintenance of Equipment

    Continual monitoring of the temperature of water baths

    and refrigerators is important to the maintenance of

    reagent quality and test performance.

    Equipment such as microscopes, centrifuges, and

    spectrophotometers should be cleaned and checked

    for accuracy on a regular schedule.

    Failure to monitor equipment regularly can produce

    inaccurate test results and lead to expensive repairs.

  • 7/30/2019 QC in the Lab.pps

    6/38

    Factors in Quality AssuranceTo guarantee the highest quality patient care through laboratory testing, a

    variety ofpre-analytical, analytical and post-analytical factors in addition to

    analytical data must be considered.

    Pre-analytical, analytical and post-analytical factors are all part of Quality

    Control

    Non-analytical factors that support quality testing include the following:

    Qualified Personnel

    Established laboratory policies

    The laboratory procedure manual

    Proper procedures for specimen collection and storage

    Preventive maintenance of equipment

    Appropriate methodology

    Established quality assurance techniques

    Accuracy in reporting results

  • 7/30/2019 QC in the Lab.pps

    7/38

    Pre Analytical

    Proper Procedures for Specimen Collection and Storage

    Strict adherence to correct procedures for specimen collectionand storage is critical to the accuracy of any test.

    Pre-analytical errors are the most common

    source of laboratory error.

    The correct patient identification is critical.

    Date of Birth. (careful with overseas people as Month & daymight be reversed)

    Ask patient to spell their name.

  • 7/30/2019 QC in the Lab.pps

    8/38

    Pre Analytical (cont)

    Each blood specimen obtained should be labeled with the

    patients first and last name, hospital identification number,patient location, time, date and the phlebotomists initials.

    Correct storage of specimens is critical to obtaining accurateresults.

    Some specimens are stored at room temperature and somein the refrigerator.

    Specimen integrity is an important issue when blood iscollected at a site away from the testing facility.

    Blood must be centrifuged within 30 minutes ofcollection.

    Delayed centrifugation will result in wrong results.

  • 7/30/2019 QC in the Lab.pps

    9/38

    Pre Analytical (cont)

    Specimens are to be transported to the laboratory in the

    appropriate containers, e.g. coolers.

    Physiologic Factors Affecting Test Results

    Posture- Changing from a supine to a sitting or standingposition results in a shift of body water from inside the blood

    vessels to the interstitial spaces.

    Larger molecules cannot filter into the tissues and concentratein the blood. There will be significant increases in test values

    for lipids, enzymes and proteins.

  • 7/30/2019 QC in the Lab.pps

    10/38

    Pre Analytical (cont)

    Diurnal Rhythm- Diurnal pertains to daylight, and diurnal

    rhythm refers to the daily body fluctuations that occur.

    Certain hormone levels such as cortisol and adenocorticotropichormones, decrease in the afternoon.

    Other test values, such as iron and eosinophils, increase in the

    afternoon.

  • 7/30/2019 QC in the Lab.pps

    11/38

    Pre Analytical (cont)

    Exercise- muscle activity elevates creatinine, protein, creatine

    kinase, aspartate transaminase, and lactate dehydrogenase

    values.

    Research also suggests that exercise activates coagulation

    and fibrinolysis and increase platelet counts.

    Stress- Anxiety can cause temporary increase in white blood

    cells, catecholamines as well as an acid-base imbalance.

  • 7/30/2019 QC in the Lab.pps

    12/38

    Pre Analytical (cont)

    Dieting- Fasting means no food or beverages except

    water for 8-12 hours prior to blood collection.

    If a patient has recently eaten, there will be atemporary increase in glucose and lipid content in the

    blood.

    As a result, the serum or plasma may appear cloudyor turbid, which interferes with testing, especially with

    tests such as glucose, sodium, and complete blood

    counts.

  • 7/30/2019 QC in the Lab.pps

    13/38

    Pre Analytical (cont)

    Smoking- Patients who smoke prior to blood collection mayhave increased WBC counts and cortisol levels.

    Long-term smoking can lead to decreased pulmonary function

    and result in increased hemoglobin levels. (secondarypolycythaemia)

  • 7/30/2019 QC in the Lab.pps

    14/38

  • 7/30/2019 QC in the Lab.pps

    15/38

    Analytical

    Appropriate Methodology

    When new methods are introduced, it is important to

    check the procedure for accuracy and variability.

    Replicate analyses using control specimens are

    recommended to check for accuracy and to eliminate

    factors such as day-to-day variability, reagent

    variability and differences between technologists.

  • 7/30/2019 QC in the Lab.pps

    16/38

    Analytical (cont)

    Established Quality Assurance Techniques

    Each procedure should have an established protocol toassure the quality of the results.

    Usually, the normal and abnormal control samples are

    analyzed at the same time patient specimens are analyzed.

    Established limits of acceptable performance must bedetermined for each type of test.

    If control results are not within acceptable limits, patient

    results cannot be guaranteed to be accurate.

    In these cases, the source of error must be identified and the

    entire test repeated before a patients result can be reported.

  • 7/30/2019 QC in the Lab.pps

    17/38

    Analytical (cont)

    Accuracy in Reporting Results

    When extremely abnormal results or differences from

    previous test values are found, the laboratory

    protocol should establish the method of rechecking

    and reporting such results.

  • 7/30/2019 QC in the Lab.pps

    18/38

    Things that go wrong during analysis

    1. Equipment faults

    2. Reagents expired

    3. Reagents contaminated

    4. Reagents not stored under the right conditions

    5. Temperatures not checked

    6. Automatic pipettes not maintained

    7. Dirty glassware

    8. Bad operator technique

  • 7/30/2019 QC in the Lab.pps

    19/38

    Qualified Personnel

    The entry-level examination competencies of all certifiedpersons in hematology must be validated. Validationtakes in form of both external certification and newemployee orientation to the work environment.Continuing competency is equally important.

    Established Laboratory Policies Laboratory policies should be included in the laboratory

    reference manual that is available to all hospitalpersonnel.

    The Laboratory Procedure Manual

    Laboratory procedures should be included in the manual.

  • 7/30/2019 QC in the Lab.pps

    20/38

    Batch:

    A batch or lot is a collection of products all identical in size,

    type, conditions and time of production.

    All controls have a Batch Number and expiry date so that a

    fault can be traced easily.

    The number of tests between controls.

    Drift:

    The control results are steadily increasing or decreasing away

    from the true published mean value found in the packageinsert.

  • 7/30/2019 QC in the Lab.pps

    21/38

    Interference:

    Something outside or inside the autoanalyser causing

    unreliable control and test results.e.g. electrical interference, a light source losing brilliance, dirt or

    dust etc. Insects.

    Interpolation:

    Someone changing a test or control result because it doesnt

    look or feel right.

    Random error:

    An unexpected and unacceptable test or control result whichdisappears when repeated.

    Fibrin blocking the autoanalyser.

  • 7/30/2019 QC in the Lab.pps

    22/38

    Standard Deviation: (SD)Any control results greater than 2 standard deviation are

    flagged and the analyser usually stops.

    Calibrator: (standard):

    A know value supplied in the package insert to calculate the

    unknown value.

  • 7/30/2019 QC in the Lab.pps

    23/38

    Controls:

    There are Internal and External controls.

    Internal controls are reconstituted and assayed.The values are given in the insert package.

    External controls are sent from an outside national or

    international laboratory.

    The results are not supplied until after they receive your results.

    This way cheating is eliminated.

    Controls above and below normal reference ranges should be

    performed.

    Important when monitoring therapeutic Drugs.

  • 7/30/2019 QC in the Lab.pps

    24/38

    Bias:

    Bias is a problem and is widespread when doing internal controls.

    This is why external controls are done in parallel.

    Examples of bias:

    1. All controls should be done with the normal run, not separately and not in

    duplicate.

    Often the laboratory staff will do controls separately and several times until

    they get the answer they want. This is incorrect and cheating.

    Controls are run to detect problems with equipment and technique.

    2. Large batches should have a control about every 10 to 15 samples not at

    the beginning and end of a sample run.

    3. The external and internal controls should not be tested separately or in

    triplicate.

    4. Cheating.

  • 7/30/2019 QC in the Lab.pps

    25/38

    NATA

    National Association of Testing Authority

    RCPA

    Royal College of Pathologists of Australia

    QAP

    Quality Assurance Programs

    Organizations and programs that monitor

    Laboratories

  • 7/30/2019 QC in the Lab.pps

    26/38

    Book are kept to monitor refrigerator temperatures, water

    baths, autoclaves etc, and faults that occur and how they

    were rectified.

    This is required by NATA.

    NATA inspections are done about every three years.

    Any problems have to be fixed before NATA gives approval to

    continue to operate as a laboratory.

    Many laboratories have been closed down for not complying

    with the standards required by NATA.

  • 7/30/2019 QC in the Lab.pps

    27/38

    Conclusion (Quality Assurance QA)

    QA is the sum of all those activities in which the

    laboratory is engaged to ensure that informationgenerated by laboratory is correct.

    QA includes all aspects of laboratory activities that

    affects the results produced, from the choice of

    methods, to the education of personnel, to the

    handling of specimens and reporting results.

    The real purpose of QA activities is to determine how

    correct or incorrect the results coming from the labare, and to allow those managing the lab to determine

    whether or not the lab is fulfilling its functions

    satisfactorily.

  • 7/30/2019 QC in the Lab.pps

    28/38

    3 major activities of QA :

    1) Preventive those activities that are done prior to the

    examination of the specimen or sample and that are intendedto establish systems conducive to accuracy testing ( eg :

    preventive maintenance and calibration of instruments, testing

    of media, orientation and training of personnel )

    2) Assessment those activities that are done during testingto determine whether the test systems are performing correctly

    ( eg : the use of standard and controls, maintenance of control

    charts )

    3) Corrective those activities that are done, when error isdetected, to correct the system ( eg : equipment

    troubleshooting, recalibration of instruments )

  • 7/30/2019 QC in the Lab.pps

    29/38

    QA in Haematology Laboratory

    QA in haematology lab is intended to ensure the reliability ofthe lab tests.

    The objective is to achieve precision and accuracy

    4 components of QA programme :

    1) Internal Quality Control ( IQC )

    2) External Quality Control ( EQC )

  • 7/30/2019 QC in the Lab.pps

    30/38

    Accuracy

    - the closeness of the estimated value to the true value.

    - can be checked by the use of reference materials which have

    been assayed by independent methods of known precision

    Precision

    - reproducibility of a results, whether accurate or inaccurate

    within a define frame time ( eg: within the same day, from week

    to week etc )

    - can be controlled by replicate tests, check tests on previously

    measured specimens and statistical evaluation of results

  • 7/30/2019 QC in the Lab.pps

    31/38

    Internal Quality Control ( IQC )

    Based on monitoring the Haematology tests procedures that are performed

    in the lab.

    includes measurements on CONTROLS & is intended to ensure that thereis continual evaluation of the reliability of the work of the lab and that

    control is exercised over the release of the results.

    External Quality Control ( EQC )

    The objective evaluation by an outside agency of the performance by a

    number of laboratories on material which is supplied specially for the

    purpose.

    is usually organized on a national or regional basis.

    analysis of performance is retrospective.

    the objective is to achieve comparability with results of other labs.

  • 7/30/2019 QC in the Lab.pps

    32/38

    Control materials

    Specially preparedIt may be anticoagulated whole blood, preserved pooled red

    cells, plasma or serum.

    It can be used to check for accuracy if the value has been

    reliably determined ( eg : reference centre )

    Should have controls of high, normal and low values

    At least 1 control specimen should be used for every batch.

    If large specimens, use 1 control for every 20 specimens.

  • 7/30/2019 QC in the Lab.pps

    33/38

    Analysis of data

    Standard deviation of control specimensDispersion of results around the mean will indicate the error of

    reproducibility.

    95% of results on the same specimen should be within 2 SD

    and 99.7% within 3 SD.

    by chance, 1 in 20 of measurement might expected to fall

    outside 2 SD and only 1 in 333 outside 3 SD.

    If measurement more widely dispersed, this indicates an errorin the test.

  • 7/30/2019 QC in the Lab.pps

    34/38

    Control Charts

    Originally described by Shewhart, 1st applied in clinical chemistry by Levey

    and Jennings.

    Samples of the control specimen are included in every batch of patients

    specimens and the results checked on a control chart.

    To check precision, it is not necessary to know the exact value of the

    control specimen.

    Value has been determined reliably by a reference method, the same

    material can be used to check accuracy or to calibrate an instrument.

    If possible, controls with high, low and normal values should be used.

    Advisable to use at least one control sample per batch even if the batch is

    very small.

    The results obtained with the control samples can be plotted on a chart.

  • 7/30/2019 QC in the Lab.pps

    35/38

  • 7/30/2019 QC in the Lab.pps

    36/38

  • 7/30/2019 QC in the Lab.pps

    37/38

    A minimum of 100 patients are selected to establish a

  • 7/30/2019 QC in the Lab.pps

    38/38

    Total = 95.4%

    A minimum of 100 patients are selected to establish a

    Normal Reference Range

    Within 2 Standard Deviations


Recommended