+ All Categories
Home > Documents > QS -Quality Control

QS -Quality Control

Date post: 05-Apr-2018
Category:
Upload: drsandeep-narula
View: 219 times
Download: 0 times
Share this document with a friend

of 90

Transcript
  • 7/31/2019 QS -Quality Control

    1/90

    1

    Quality Control

    Introduction

  • 7/31/2019 QS -Quality Control

    2/90

    2

    Purchasing

    & Inventory

    AssessmentOccurrence

    Management

    Information

    Management

    Process

    ImprovementCustomer

    Service

    Facilities &

    Safety

    The Quality System

    Organization Personnel Equipment

    Documents

    & Records

    Process

    Control

    (QC & EQA) &

    Specimen

    Management

  • 7/31/2019 QS -Quality Control

    3/90

    3

    The Quality Assurance Cycle

    Data and Lab

    ManagementSafetyCustomerService

    Patient/Client Prep

    Sample Collection

    Sample Receipt

    and Accessioning

    Sample TransportQuality Control

    Record Keeping

    ReportingPersonnel Competency

    Test Evaluations

    Testing

  • 7/31/2019 QS -Quality Control

    4/90

    4

    Quality Control

    Definitions

    Qualitative Quality Control

    Quantitative QC How to implement

    Selection and managing control materials

    Analysis of QC data

    Monitoring quality control data

  • 7/31/2019 QS -Quality Control

    5/90

    5

    What is Quality Control?

    Process or system for monitoring the qualityof laboratory testing, and the accuracy and

    precision of results Routinely collect and analyze data from every

    test run or procedure

    Allows for immediate corrective action

  • 7/31/2019 QS -Quality Control

    6/90

    6

    Designing a QC Program

    Establish written policies and procedures

    Corrective action procedures

    Train all staff

    Design forms

    Assure complete documentation and review

  • 7/31/2019 QS -Quality Control

    7/90

    7

    Qualitative vs.Quantitative

    Quantitative test

    measures the amount of a substance

    present Qualitative test

    determines whether the substance being

    tested for is present or absent

  • 7/31/2019 QS -Quality Control

    8/90

    8

    Qualitative QC

    Quality control is performed for both, systemis somewhat different

    Controls available Blood Bank/Serology/Micro

    RPR/TPHA

    Dipstick technology

    Pregnancy

  • 7/31/2019 QS -Quality Control

    9/90

    9

    Stains, Reagents, Antisera

    Label containers

    contents

    concentration date prepared

    placed in service

    expiration date/shelf life

    preparer

  • 7/31/2019 QS -Quality Control

    10/90

    10

    Media Preparation

    Record amount prepared

    Source

    Lot number Sterilization method

    Preparation date

    Preparer pH

    Expiration date

  • 7/31/2019 QS -Quality Control

    11/90

    11

    Microbiology QC

    Check: SterilityAbility to support growth

    Selective or inhibitory characteristics of the medium Biochemical response

    Frequency

    Test QC organisms with each new batch or lot number

    Check for growth of fastidious organisms on media ofchoice incubate at time and temp recommended

    RECORD Results on Media QC form

  • 7/31/2019 QS -Quality Control

    12/90

    12

    Quality Control: Stains and Reagents

    Gram stain QC

    Use gram positive and gram negative

    organisms to check stain daily Other :

    Check as used positive and negative

    reactions

  • 7/31/2019 QS -Quality Control

    13/90

    13

    Stock QC organisms

    Organisms to be maintained must beadequate to check all media and test systems.

    E. coli MacConkey, EMB, susceptibilitytests

    Staphylococcus aureus Blood agar,

    Mannitol Salt, susceptibility tests Neisseria gonorrhoeae chocolate, Martin-

    Lewis

  • 7/31/2019 QS -Quality Control

    14/90

    14

    Detecting Errors

    Many organisms have predictableantimicrobial test results

    Staphylococcus spp. are usually

    susceptible to vancomycin Streptococcus pyogenes are always

    susceptible to penicillin

    Klebsiella pneumoniae are resistant toampicillin

  • 7/31/2019 QS -Quality Control

    15/90

    15

    Sources of Error

    If you encounter an unusual pattern

    rule out error by checking identification oforganisms

    repeat antimicrobial susceptibility test

    Report if repeat testing yields same result, or referthe isolate to a reference laboratory for confirmation

  • 7/31/2019 QS -Quality Control

    16/90

    16

    Quality Control

    Quantitative Tests

    How to implement a laboratory

    quality control program

  • 7/31/2019 QS -Quality Control

    17/90

    17

    Implementing a QC Program

    Quantitative Tests Select high quality controls Collect at least20 control values over a period of

    20-30

    days for each level of control Perform statistical analysis Develop Levey-Jennings chart Monitor control values using the Levey-Jennings

    chart and/or Westgard rules Take immediate corrective action, if needed

    Record actions taken

  • 7/31/2019 QS -Quality Control

    18/90

    18

    Selecting Control Materials

    Calibrators Has a known concentration of the substance(analyte) being measured

    Used to adjust instrument, kit, test system inorder to standardize the assay

    Sometimes called a standard, although

    usually not a true standard This is nota control

  • 7/31/2019 QS -Quality Control

    19/90

    19

    Selecting Control Materials

    Controls Known concentration of the analyte

    Use 2 or three levels of controls

    Include with patient samples whenperforming a test

    Used to validate reliability of the test system

  • 7/31/2019 QS -Quality Control

    20/90

    20

    Control Materials

    Important Characteristics Values cover medical decision points Similar to the test specimen (matrix)

    Available in large quantity Stored in small aliquots

    Ideally, should last for at least 1 year

    Often use biological material, consider bio-hazardous

  • 7/31/2019 QS -Quality Control

    21/90

    21

    Managing Control Materials

    Sufficient material from same lot number orserum pool for one years testing

    May be frozen, freeze-dried, or chemicallypreserved

    Requires very accurate reconstitution if this

    step is necessary Always store as recommended bymanufacturer

  • 7/31/2019 QS -Quality Control

    22/90

    22

    Sources of QC Samples

    Appropriate diagnostic sample

    Obtained from:

    Another laboratory

    EQA provider

    Commercial product

  • 7/31/2019 QS -Quality Control

    23/90

    23

    Types of Control Materials

    Assayed mean calculated by the manufacturer must verify in the laboratory

    Unassayed less expensive must perform data analysis

    Homemade or In-house

    pooled sera collected in the laboratory characterized preserved in small quantities for daily use

  • 7/31/2019 QS -Quality Control

    24/90

    24

    Preparing In-House Controls

  • 7/31/2019 QS -Quality Control

    25/90

    25

    Criteria for Developing

    Quality Controls for HIV

    Low positive

    Between the cut off and positive control

    At a level where variability can be followed

    Generally ~2 times the cut off

  • 7/31/2019 QS -Quality Control

    26/90

    26

    Production of a QC Sample -

    Production Protocol Materials

    Calculation of Volume

    stock sample

    diluent

    QC batch

    Method Validation Acceptance Criteria

    batch

    stability

  • 7/31/2019 QS -Quality Control

    27/90

    27

    Process for Preparing

    In-house Controls

    Serial dilution of high positive stock sample

    Select suitable dilution

    Produce large batch Test stability

    Test batch variation

    Dispense, label, store

  • 7/31/2019 QS -Quality Control

    28/90

    28

    Making Suitable Dilutions

    100 ul serumin tube 1

    100ul diluent ineach tube

    Mix and Transfer

    Each tube is a 1:2 dilution

    of the previous tube

    Discard

  • 7/31/2019 QS -Quality Control

    29/90

    29

    Selecting a Suitable Sample Dilution

    Serial Dilutions on Abbott AxSYM HIV-1/HIV-2 MEIA

    Doubling Dilutions

    2

    4

    8

    16

    32

    64

    128

    256

    512

    1024

    2048

    4096

    8192

    16384

    32768

    65536

    131072

    262144

    524288

    S/Co

    Ratio

    0

    2

    4

    6

    8

    10

    12

    14

    16

    18

    20

    Pos Cont 3.3

    Neg Cont 0.38Cut Off 1.0

  • 7/31/2019 QS -Quality Control

    30/90

    30

    Batch Production

    Prepare positive sample

    centrifuge

    heat inactivate

    Mix positive sample in diluent

    magnetic stirrer Bottle batch in numbered lots of suitable

    volume

  • 7/31/2019 QS -Quality Control

    31/90

    31

    Stability Testing

    Assess the rate of deterioration

    QC Sample

    Storage

    Day 7 Day 14 Day 21 Day 28

    -20c

    4c

    16-25C

  • 7/31/2019 QS -Quality Control

    32/90

    32

    Batch Validation

    Dispense aliquots

    Test aliquots

    Confirm desired titre level

    compare against target value

    Confirm minimal batch variation acceptable if CV

  • 7/31/2019 QS -Quality Control

    33/90

    33

    Storage of QC Samples

    Validated batch aliquoted into smaller userfriendly volumes for storage

    Establish a storage protocol:

    store at -20oC in use vials stored at 4oC

    use 0.5 ml vial maximum of one week

    freeze-dried(requires accurate reconstitution)

    chemically preserved

  • 7/31/2019 QS -Quality Control

    34/90

    34

  • 7/31/2019 QS -Quality Control

    35/90

    35

    Quality Control -Quantitative

    Analysis of QC Data

  • 7/31/2019 QS -Quality Control

    36/90

  • 7/31/2019 QS -Quality Control

    37/90

    37

    Analysis of Control Materials

    Need data set of at least 20 points, obtainedover a 30 day period

    Calculate mean, standard deviation,coefficient of variation; determine targetranges

    Develop Levey-Jennings charts, plot results

  • 7/31/2019 QS -Quality Control

    38/90

    38

    Establishing Control Ranges

    Select appropriate controls

    Assay them repeatedly over time

    at least 20 data points

    Make sure any procedural variation is represented:

    different operators

    different times of day

    Determine the degree of variability in the data to establishacceptable range

  • 7/31/2019 QS -Quality Control

    39/90

    39

    Measurement of Variability

    A certain amount of variability will naturallyoccur when a control is tested repeatedly.

    Variability is affected by operator technique,environmental conditions, and theperformance characteristics of the assaymethod.

    The goal is to differentiate betweenvariability due to chance from that due toerror.

  • 7/31/2019 QS -Quality Control

    40/90

    40

    Measures of Central Tendency

    Data are frequently distributed about acentral value or a central location

    There are several terms to describe thatcentral location, or the central tendencyof a set of data

  • 7/31/2019 QS -Quality Control

    41/90

    41

    Measures of Central Tendency

    Median = the value at the center (midpoint)of the observations

    Mode = the value which occurs with thegreatest frequency

    Mean = the calculated average of the

    values

  • 7/31/2019 QS -Quality Control

    42/90

    42

    Calculation of Mean

    X = Mean

    X1 = First result

    X2

    = Second result

    Xn = Last result in series

    n Total number of results

    Calc lation of Mean O tliers

  • 7/31/2019 QS -Quality Control

    43/90

    43

    Calculation of Mean: Outliers

    1. 192 mg/dL

    2. 194 mg/dL

    3. 196 mg/dL4. 196 mg/dL

    5. 160 mg/dL

    6. 196 mg/dL

    7. 200 mg/dL

    8. 200 mg/dL

    9. 202 mg/dL10. 255 mg/dL

    11. 204 mg/dL

    12. 208 mg/dL

    13. 212 mg/dL

    Calculation of Mean

  • 7/31/2019 QS -Quality Control

    44/90

    44

    Calculation of Mean

    1) 192 mg/dL

    2) 194 mg/dL3) 196 mg/dL

    4) 196 mg/dL

    5) 196 mg/dL6) 200 mg/dL

    7) 200 mg/dL

    8) 202 mg/dL

    9) 204 mg/dL

    10) 208 mg/dL

    11) 212 mg/dL

    Sum = 2,200 mg/dL

    Mean = the calculatedaverage of the values

    The sum of the values (X1+ X2 + X3 X11) dividedby the number (n) ofobservations

    The mean of these 11

    observations is (2200 11) = 200 mg/dL

  • 7/31/2019 QS -Quality Control

    45/90

    45

    Calculation of Mean:

    ELISA Tests

    Collect optical density (OD) values for controlsfor each assay run

    Collect cutoff (CO) value for each run Calculate ratio of OD to CO (OD/CO) for each

    data point or observation

    This ratio standardizes data Use these ratio values to calculate the mean

  • 7/31/2019 QS -Quality Control

    46/90

    46

    Normal Distribution

    All values are symmetrically distributedaround the mean

    Characteristic bell-shaped curve Assumed for all quality control statistics

  • 7/31/2019 QS -Quality Control

    47/90

    47

    Normal Distribution

    Freque

    ncy

    4.7 4.8 4.9 Mean 5.1 5.2 5.3

    X

    Normal Distribution

  • 7/31/2019 QS -Quality Control

    48/90

    48

    Normal Distribution

    0246

    810121416

    #ofOb

    servations

    192 194 196 198 200 202 204 206 208 210 212

    Serum glucose (mg/dL)

    Mean

  • 7/31/2019 QS -Quality Control

    49/90

    49

    Accuracy and Precision

    The degree of fluctuation in the measurements isindicative of the precision of the assay.

    The closeness of measurements to the true valueis indicative of the accuracy of the assay.

    Quality Control is used to monitor both the precisionand the accuracy of the assay in order to provide

    reliable results.

  • 7/31/2019 QS -Quality Control

    50/90

    50

    Precise and inaccurate Precise and accurate

    Precision and Accuracy

  • 7/31/2019 QS -Quality Control

    51/90

    51

    Imprecise and inaccurate

  • 7/31/2019 QS -Quality Control

    52/90

    52

    Measures of Dispersion

    or Variability There are several terms that describe thedispersion or variability of the data aroundthe mean:

    Range

    Variance

    Standard DeviationCoefficient of Variation

  • 7/31/2019 QS -Quality Control

    53/90

    53

    Range

    Range refers to the difference or spreadbetween the highest and lowest observations.

    It is the simplest measure of dispersion. It makes no assumption about the shape of

    the distribution or the central tendency of the

    data.

  • 7/31/2019 QS -Quality Control

    54/90

    54

    Calculation of Variance (S2)

    222

    1N

    )X(X2/dlmgS

    1

  • 7/31/2019 QS -Quality Control

    55/90

    55

    Calculation of Variance

    Variance is a measure of variability about themean.

    It is calculated as the average squareddeviation from the mean.

    the sum of the deviations from the mean,squared, divided by the number ofobservations (corrected for degrees offreedom)

  • 7/31/2019 QS -Quality Control

    56/90

    56

    Degrees of Freedom

    Represents the number of independentdata points that are contained in a data set.

    The mean is calculated first, so thevariance calculation has lost one degree of

    freedom (n-1)

  • 7/31/2019 QS -Quality Control

    57/90

    57

    Calculation of Standard Deviation

    mg/dlS1N

    )x(x 21

    variance

  • 7/31/2019 QS -Quality Control

    58/90

  • 7/31/2019 QS -Quality Control

    59/90

    59

    Standard Deviation and Probability

    For a set of data with anormal distribution, a valuewill fall within a range of:

    +/- 1 SD 68.2% of the

    time +/- 2 SD 95.5% of the

    time

    +/- 3 SD 99.7% of thetime

    68.2%

    95.5%

    99.7%

    Freq

    uency

    -3s- 2s -1s Mean +1s +2s +3s

    X

  • 7/31/2019 QS -Quality Control

    60/90

    60

    Standard Deviation and Probability

    In general, laboratories use the +/- 2 SD criteria forthe limits of the acceptable range for a test

    When the QC measurement falls within that range,there is 95.5% confidence that the measurement iscorrect

    Only 4.5% of the time will a value fall outside of

    that range due to chance; more likely it will be dueto error

    C l l ti f

  • 7/31/2019 QS -Quality Control

    61/90

    61

    Calculation of

    Coefficient of Variation

    The coefficient ofvariation (CV) is the

    standard deviation (SD)expressed as apercentage of themean

    Ideally should be lessthan 5%

    100xmean

    SDCV

  • 7/31/2019 QS -Quality Control

    62/90

    62

    Monitoring QC Data

  • 7/31/2019 QS -Quality Control

    63/90

  • 7/31/2019 QS -Quality Control

    64/90

    64

    Levey-Jennings Chart

    A graphical method for displaying controlresults and evaluating whether a procedure

    is in-control or out-of-control Control values are plotted versus time

    Lines are drawn from point to point to accentany trends, shifts, or random excursions

  • 7/31/2019 QS -Quality Control

    65/90

    65

    Levey-Jennings Chart

    +3SD

    +2SD

    +1SD

    Mean

    -1SD

    -2SD

    -3SD

  • 7/31/2019 QS -Quality Control

    66/90

    66

    Levey-Jennings Chart -

    Record Time on X-Axis and the Control Values on Y-Axis

    80

    85

    90

    95100

    105

    110

    115

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24Co

    ntrolValues

    (e.g.mg/dL)

    Time (e.g. day, date, run number)

  • 7/31/2019 QS -Quality Control

    67/90

    67

    Levey-Jennings Chart -Plot Control Values for Each Run

    80

    85

    90

    95100

    105

    110

    115

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24Co

    ntrolValues

    (e.g.mg/dL)

    Time (e.g. day, date, run number)

    Levey Jennings Chart

  • 7/31/2019 QS -Quality Control

    68/90

    68

    Levey-Jennings ChartCalculate the Mean and Standard Deviation;

    Record the Mean and +/- 1,2 and 3 SD Control Limits

    80

    85

    90

    95100

    105

    110

    115

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

    Mean

    Day

    +1SD

    +2SD

    +3SD

    -1SD

    -2SD

    -3SD

    L J i Ch t

  • 7/31/2019 QS -Quality Control

    69/90

    69

    Levey-Jennings Chart -Record and Evaluate the Control Values

    80

    85

    90

    95

    100

    105

    110

    115

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

    Mean

    Day

    +1SD

    +2SD

    +3SD

    -1SD

    -2SD

    -3SD

  • 7/31/2019 QS -Quality Control

    70/90

    70

    Findings Over Time

    Ideally should have control values clustered aboutthe mean (+/-2 SD) with little variation in the upwardor downward direction

    Imprecision = large amount of scatter about themean. Usually caused by errors in technique

    Inaccuracy = may see as a trend or a shift, usuallycaused by change in the testing process

    Random error = no pattern. Usually poor technique,malfunctioning equipment

  • 7/31/2019 QS -Quality Control

    71/90

    71

    Statistical Quality Control Exercise Hypothetical control values (2 levels of

    control) Calculation of mean

    Calculation of standard deviation

    Creation of a Levey-Jennings chart

    When does the Control Value

  • 7/31/2019 QS -Quality Control

    72/90

    72

    When does the Control Value

    Indicate a Problem? Consider using Westgard Control Rules

    Uses premise that 95.5% of control values

    should fall within 2SD Commonly applied when two levels of control

    are used

    Use in a sequential fashion

  • 7/31/2019 QS -Quality Control

    73/90

    73

    Westgard Rules

    Multirule Quality Control

    Uses a combination of decision criteria or

    control rules Allows determination of whether an analytical

    run is in-control or out-of-control

  • 7/31/2019 QS -Quality Control

    74/90

    74

    Westgard Rules

    (Generally used where 2 levels of controlmaterial are analyzed per run)

    12S

    rule

    13S rule

    22S rule

    R4S rule

    41S rule

    10X rule

  • 7/31/2019 QS -Quality Control

    75/90

    75

    Westgard 12S Rule

    warning rule

    One of two control results falls outside 2SD

    Alerts tech to possible problems Not cause for rejecting a run

    Must then evaluate the 13S rule

    1 R l A i i f l i i

  • 7/31/2019 QS -Quality Control

    76/90

    76

    12S Rule = A warning to trigger careful inspectionof the control data

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

    Mean

    Day

    +1SD

    +2SD

    +3SD

    -1SD

    -2SD

    -3SD

    12S rule

    violation

  • 7/31/2019 QS -Quality Control

    77/90

    77

    Westgard 13S Rule

    If either of the two control results falls outsideof 3SD, rule is violated

    Run must be rejected If 13S not violated, check 22S

  • 7/31/2019 QS -Quality Control

    78/90

  • 7/31/2019 QS -Quality Control

    79/90

    79

    Westgard 22S Rule

    2 consecutive control values for the samelevel fall outside of 2SD in the samedirection, or

    Both controls in the same run exceed 2SD

    Patient results cannot be reported

    Requires corrective action

    22S Rule = Reject the run when 2 consecutive control

  • 7/31/2019 QS -Quality Control

    80/90

    80

    22S Rule = Reject the run when 2 consecutive controlmeasurements exceed the same

    +2SD or -2SD control limit

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

    Mean

    Day

    +1SD

    +2SD

    +3SD

    -1SD

    -2SD

    -3SD

    22S ruleviolation

  • 7/31/2019 QS -Quality Control

    81/90

    81

    Westgard R4S Rule

    One control exceeds the mean by2SD, andthe other control exceeds the mean by +2SD

    The range between the two results willtherefore exceed 4 SD

    Random error has occurred, test run must be

    rejected

    R4S Rule = Reject the run when 1 control

  • 7/31/2019 QS -Quality Control

    82/90

    82

    R4S Rule = Reject the run when 1 controlmeasurement exceed the +2SD and the other

    exceeds the -2SD control limit

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

    Mean

    Day

    +1SD

    +2SD

    +3SD

    -1SD

    -2SD

    -3SD

    R4S rule

    violation

  • 7/31/2019 QS -Quality Control

    83/90

    83

    Westgard 41S Rule

    Requires control data from previous runs

    Four consecutive QC results for one level of

    control are outside 1SD, or Both levels of control have consecutive results

    that are outside 1SD

  • 7/31/2019 QS -Quality Control

    84/90

    84

    Westgard 10X Rule

    Requires control data from previous runs

    Ten consecutive QC results for one level of

    control are on one side of the mean, or Both levels of control have five consecutive

    results that are on the same side of the mean

    10 Rule = Reject the run when 10 consecutive control

  • 7/31/2019 QS -Quality Control

    85/90

    85

    10x Rule = Reject the run when 10 consecutive controlmeasurements fall on one side of the mean

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

    Mean

    Day

    +1SD

    +2SD

    +3SD

    -1SD

    -2SD

    -3SD

    10x rule

    violation

    Westgard Multirule QC

  • 7/31/2019 QS -Quality Control

    86/90

    86

    Westgard Multirule QC

  • 7/31/2019 QS -Quality Control

    87/90

  • 7/31/2019 QS -Quality Control

    88/90

    88

    Solving out-of-control problems

    Policies and procedures for remedial action

    Troubleshooting

    Alternatives to run rejection

  • 7/31/2019 QS -Quality Control

    89/90

    89

    Summary

    Why QC program?

    Validates test accuracy and reliability

    Summary:

  • 7/31/2019 QS -Quality Control

    90/90

    Summary:

    How to implement a QC program? Establish written policies and procedures Assign responsibility for monitoring and reviewing Train staff Obtain control materials Collect data Set target values (mean, SD) Establish Levey-Jennings charts Routinely plot control data Establish and implement troubleshooting and corrective

    action protocols Establish and maintain system for documentation


Recommended