Date post: | 05-Apr-2018 |
Category: |
Documents |
Upload: | drsandeep-narula |
View: | 219 times |
Download: | 0 times |
of 90
7/31/2019 QS -Quality Control
1/90
1
Quality Control
Introduction
7/31/2019 QS -Quality Control
2/90
2
Purchasing
& Inventory
AssessmentOccurrence
Management
Information
Management
Process
ImprovementCustomer
Service
Facilities &
Safety
The Quality System
Organization Personnel Equipment
Documents
& Records
Process
Control
(QC & EQA) &
Specimen
Management
7/31/2019 QS -Quality Control
3/90
3
The Quality Assurance Cycle
Data and Lab
ManagementSafetyCustomerService
Patient/Client Prep
Sample Collection
Sample Receipt
and Accessioning
Sample TransportQuality Control
Record Keeping
ReportingPersonnel Competency
Test Evaluations
Testing
7/31/2019 QS -Quality Control
4/90
4
Quality Control
Definitions
Qualitative Quality Control
Quantitative QC How to implement
Selection and managing control materials
Analysis of QC data
Monitoring quality control data
7/31/2019 QS -Quality Control
5/90
5
What is Quality Control?
Process or system for monitoring the qualityof laboratory testing, and the accuracy and
precision of results Routinely collect and analyze data from every
test run or procedure
Allows for immediate corrective action
7/31/2019 QS -Quality Control
6/90
6
Designing a QC Program
Establish written policies and procedures
Corrective action procedures
Train all staff
Design forms
Assure complete documentation and review
7/31/2019 QS -Quality Control
7/90
7
Qualitative vs.Quantitative
Quantitative test
measures the amount of a substance
present Qualitative test
determines whether the substance being
tested for is present or absent
7/31/2019 QS -Quality Control
8/90
8
Qualitative QC
Quality control is performed for both, systemis somewhat different
Controls available Blood Bank/Serology/Micro
RPR/TPHA
Dipstick technology
Pregnancy
7/31/2019 QS -Quality Control
9/90
9
Stains, Reagents, Antisera
Label containers
contents
concentration date prepared
placed in service
expiration date/shelf life
preparer
7/31/2019 QS -Quality Control
10/90
10
Media Preparation
Record amount prepared
Source
Lot number Sterilization method
Preparation date
Preparer pH
Expiration date
7/31/2019 QS -Quality Control
11/90
11
Microbiology QC
Check: SterilityAbility to support growth
Selective or inhibitory characteristics of the medium Biochemical response
Frequency
Test QC organisms with each new batch or lot number
Check for growth of fastidious organisms on media ofchoice incubate at time and temp recommended
RECORD Results on Media QC form
7/31/2019 QS -Quality Control
12/90
12
Quality Control: Stains and Reagents
Gram stain QC
Use gram positive and gram negative
organisms to check stain daily Other :
Check as used positive and negative
reactions
7/31/2019 QS -Quality Control
13/90
13
Stock QC organisms
Organisms to be maintained must beadequate to check all media and test systems.
E. coli MacConkey, EMB, susceptibilitytests
Staphylococcus aureus Blood agar,
Mannitol Salt, susceptibility tests Neisseria gonorrhoeae chocolate, Martin-
Lewis
7/31/2019 QS -Quality Control
14/90
14
Detecting Errors
Many organisms have predictableantimicrobial test results
Staphylococcus spp. are usually
susceptible to vancomycin Streptococcus pyogenes are always
susceptible to penicillin
Klebsiella pneumoniae are resistant toampicillin
7/31/2019 QS -Quality Control
15/90
15
Sources of Error
If you encounter an unusual pattern
rule out error by checking identification oforganisms
repeat antimicrobial susceptibility test
Report if repeat testing yields same result, or referthe isolate to a reference laboratory for confirmation
7/31/2019 QS -Quality Control
16/90
16
Quality Control
Quantitative Tests
How to implement a laboratory
quality control program
7/31/2019 QS -Quality Control
17/90
17
Implementing a QC Program
Quantitative Tests Select high quality controls Collect at least20 control values over a period of
20-30
days for each level of control Perform statistical analysis Develop Levey-Jennings chart Monitor control values using the Levey-Jennings
chart and/or Westgard rules Take immediate corrective action, if needed
Record actions taken
7/31/2019 QS -Quality Control
18/90
18
Selecting Control Materials
Calibrators Has a known concentration of the substance(analyte) being measured
Used to adjust instrument, kit, test system inorder to standardize the assay
Sometimes called a standard, although
usually not a true standard This is nota control
7/31/2019 QS -Quality Control
19/90
19
Selecting Control Materials
Controls Known concentration of the analyte
Use 2 or three levels of controls
Include with patient samples whenperforming a test
Used to validate reliability of the test system
7/31/2019 QS -Quality Control
20/90
20
Control Materials
Important Characteristics Values cover medical decision points Similar to the test specimen (matrix)
Available in large quantity Stored in small aliquots
Ideally, should last for at least 1 year
Often use biological material, consider bio-hazardous
7/31/2019 QS -Quality Control
21/90
21
Managing Control Materials
Sufficient material from same lot number orserum pool for one years testing
May be frozen, freeze-dried, or chemicallypreserved
Requires very accurate reconstitution if this
step is necessary Always store as recommended bymanufacturer
7/31/2019 QS -Quality Control
22/90
22
Sources of QC Samples
Appropriate diagnostic sample
Obtained from:
Another laboratory
EQA provider
Commercial product
7/31/2019 QS -Quality Control
23/90
23
Types of Control Materials
Assayed mean calculated by the manufacturer must verify in the laboratory
Unassayed less expensive must perform data analysis
Homemade or In-house
pooled sera collected in the laboratory characterized preserved in small quantities for daily use
7/31/2019 QS -Quality Control
24/90
24
Preparing In-House Controls
7/31/2019 QS -Quality Control
25/90
25
Criteria for Developing
Quality Controls for HIV
Low positive
Between the cut off and positive control
At a level where variability can be followed
Generally ~2 times the cut off
7/31/2019 QS -Quality Control
26/90
26
Production of a QC Sample -
Production Protocol Materials
Calculation of Volume
stock sample
diluent
QC batch
Method Validation Acceptance Criteria
batch
stability
7/31/2019 QS -Quality Control
27/90
27
Process for Preparing
In-house Controls
Serial dilution of high positive stock sample
Select suitable dilution
Produce large batch Test stability
Test batch variation
Dispense, label, store
7/31/2019 QS -Quality Control
28/90
28
Making Suitable Dilutions
100 ul serumin tube 1
100ul diluent ineach tube
Mix and Transfer
Each tube is a 1:2 dilution
of the previous tube
Discard
7/31/2019 QS -Quality Control
29/90
29
Selecting a Suitable Sample Dilution
Serial Dilutions on Abbott AxSYM HIV-1/HIV-2 MEIA
Doubling Dilutions
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
32768
65536
131072
262144
524288
S/Co
Ratio
0
2
4
6
8
10
12
14
16
18
20
Pos Cont 3.3
Neg Cont 0.38Cut Off 1.0
7/31/2019 QS -Quality Control
30/90
30
Batch Production
Prepare positive sample
centrifuge
heat inactivate
Mix positive sample in diluent
magnetic stirrer Bottle batch in numbered lots of suitable
volume
7/31/2019 QS -Quality Control
31/90
31
Stability Testing
Assess the rate of deterioration
QC Sample
Storage
Day 7 Day 14 Day 21 Day 28
-20c
4c
16-25C
7/31/2019 QS -Quality Control
32/90
32
Batch Validation
Dispense aliquots
Test aliquots
Confirm desired titre level
compare against target value
Confirm minimal batch variation acceptable if CV
7/31/2019 QS -Quality Control
33/90
33
Storage of QC Samples
Validated batch aliquoted into smaller userfriendly volumes for storage
Establish a storage protocol:
store at -20oC in use vials stored at 4oC
use 0.5 ml vial maximum of one week
freeze-dried(requires accurate reconstitution)
chemically preserved
7/31/2019 QS -Quality Control
34/90
34
7/31/2019 QS -Quality Control
35/90
35
Quality Control -Quantitative
Analysis of QC Data
7/31/2019 QS -Quality Control
36/90
7/31/2019 QS -Quality Control
37/90
37
Analysis of Control Materials
Need data set of at least 20 points, obtainedover a 30 day period
Calculate mean, standard deviation,coefficient of variation; determine targetranges
Develop Levey-Jennings charts, plot results
7/31/2019 QS -Quality Control
38/90
38
Establishing Control Ranges
Select appropriate controls
Assay them repeatedly over time
at least 20 data points
Make sure any procedural variation is represented:
different operators
different times of day
Determine the degree of variability in the data to establishacceptable range
7/31/2019 QS -Quality Control
39/90
39
Measurement of Variability
A certain amount of variability will naturallyoccur when a control is tested repeatedly.
Variability is affected by operator technique,environmental conditions, and theperformance characteristics of the assaymethod.
The goal is to differentiate betweenvariability due to chance from that due toerror.
7/31/2019 QS -Quality Control
40/90
40
Measures of Central Tendency
Data are frequently distributed about acentral value or a central location
There are several terms to describe thatcentral location, or the central tendencyof a set of data
7/31/2019 QS -Quality Control
41/90
41
Measures of Central Tendency
Median = the value at the center (midpoint)of the observations
Mode = the value which occurs with thegreatest frequency
Mean = the calculated average of the
values
7/31/2019 QS -Quality Control
42/90
42
Calculation of Mean
X = Mean
X1 = First result
X2
= Second result
Xn = Last result in series
n Total number of results
Calc lation of Mean O tliers
7/31/2019 QS -Quality Control
43/90
43
Calculation of Mean: Outliers
1. 192 mg/dL
2. 194 mg/dL
3. 196 mg/dL4. 196 mg/dL
5. 160 mg/dL
6. 196 mg/dL
7. 200 mg/dL
8. 200 mg/dL
9. 202 mg/dL10. 255 mg/dL
11. 204 mg/dL
12. 208 mg/dL
13. 212 mg/dL
Calculation of Mean
7/31/2019 QS -Quality Control
44/90
44
Calculation of Mean
1) 192 mg/dL
2) 194 mg/dL3) 196 mg/dL
4) 196 mg/dL
5) 196 mg/dL6) 200 mg/dL
7) 200 mg/dL
8) 202 mg/dL
9) 204 mg/dL
10) 208 mg/dL
11) 212 mg/dL
Sum = 2,200 mg/dL
Mean = the calculatedaverage of the values
The sum of the values (X1+ X2 + X3 X11) dividedby the number (n) ofobservations
The mean of these 11
observations is (2200 11) = 200 mg/dL
7/31/2019 QS -Quality Control
45/90
45
Calculation of Mean:
ELISA Tests
Collect optical density (OD) values for controlsfor each assay run
Collect cutoff (CO) value for each run Calculate ratio of OD to CO (OD/CO) for each
data point or observation
This ratio standardizes data Use these ratio values to calculate the mean
7/31/2019 QS -Quality Control
46/90
46
Normal Distribution
All values are symmetrically distributedaround the mean
Characteristic bell-shaped curve Assumed for all quality control statistics
7/31/2019 QS -Quality Control
47/90
47
Normal Distribution
Freque
ncy
4.7 4.8 4.9 Mean 5.1 5.2 5.3
X
Normal Distribution
7/31/2019 QS -Quality Control
48/90
48
Normal Distribution
0246
810121416
#ofOb
servations
192 194 196 198 200 202 204 206 208 210 212
Serum glucose (mg/dL)
Mean
7/31/2019 QS -Quality Control
49/90
49
Accuracy and Precision
The degree of fluctuation in the measurements isindicative of the precision of the assay.
The closeness of measurements to the true valueis indicative of the accuracy of the assay.
Quality Control is used to monitor both the precisionand the accuracy of the assay in order to provide
reliable results.
7/31/2019 QS -Quality Control
50/90
50
Precise and inaccurate Precise and accurate
Precision and Accuracy
7/31/2019 QS -Quality Control
51/90
51
Imprecise and inaccurate
7/31/2019 QS -Quality Control
52/90
52
Measures of Dispersion
or Variability There are several terms that describe thedispersion or variability of the data aroundthe mean:
Range
Variance
Standard DeviationCoefficient of Variation
7/31/2019 QS -Quality Control
53/90
53
Range
Range refers to the difference or spreadbetween the highest and lowest observations.
It is the simplest measure of dispersion. It makes no assumption about the shape of
the distribution or the central tendency of the
data.
7/31/2019 QS -Quality Control
54/90
54
Calculation of Variance (S2)
222
1N
)X(X2/dlmgS
1
7/31/2019 QS -Quality Control
55/90
55
Calculation of Variance
Variance is a measure of variability about themean.
It is calculated as the average squareddeviation from the mean.
the sum of the deviations from the mean,squared, divided by the number ofobservations (corrected for degrees offreedom)
7/31/2019 QS -Quality Control
56/90
56
Degrees of Freedom
Represents the number of independentdata points that are contained in a data set.
The mean is calculated first, so thevariance calculation has lost one degree of
freedom (n-1)
7/31/2019 QS -Quality Control
57/90
57
Calculation of Standard Deviation
mg/dlS1N
)x(x 21
variance
7/31/2019 QS -Quality Control
58/90
7/31/2019 QS -Quality Control
59/90
59
Standard Deviation and Probability
For a set of data with anormal distribution, a valuewill fall within a range of:
+/- 1 SD 68.2% of the
time +/- 2 SD 95.5% of the
time
+/- 3 SD 99.7% of thetime
68.2%
95.5%
99.7%
Freq
uency
-3s- 2s -1s Mean +1s +2s +3s
X
7/31/2019 QS -Quality Control
60/90
60
Standard Deviation and Probability
In general, laboratories use the +/- 2 SD criteria forthe limits of the acceptable range for a test
When the QC measurement falls within that range,there is 95.5% confidence that the measurement iscorrect
Only 4.5% of the time will a value fall outside of
that range due to chance; more likely it will be dueto error
C l l ti f
7/31/2019 QS -Quality Control
61/90
61
Calculation of
Coefficient of Variation
The coefficient ofvariation (CV) is the
standard deviation (SD)expressed as apercentage of themean
Ideally should be lessthan 5%
100xmean
SDCV
7/31/2019 QS -Quality Control
62/90
62
Monitoring QC Data
7/31/2019 QS -Quality Control
63/90
7/31/2019 QS -Quality Control
64/90
64
Levey-Jennings Chart
A graphical method for displaying controlresults and evaluating whether a procedure
is in-control or out-of-control Control values are plotted versus time
Lines are drawn from point to point to accentany trends, shifts, or random excursions
7/31/2019 QS -Quality Control
65/90
65
Levey-Jennings Chart
+3SD
+2SD
+1SD
Mean
-1SD
-2SD
-3SD
7/31/2019 QS -Quality Control
66/90
66
Levey-Jennings Chart -
Record Time on X-Axis and the Control Values on Y-Axis
80
85
90
95100
105
110
115
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24Co
ntrolValues
(e.g.mg/dL)
Time (e.g. day, date, run number)
7/31/2019 QS -Quality Control
67/90
67
Levey-Jennings Chart -Plot Control Values for Each Run
80
85
90
95100
105
110
115
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24Co
ntrolValues
(e.g.mg/dL)
Time (e.g. day, date, run number)
Levey Jennings Chart
7/31/2019 QS -Quality Control
68/90
68
Levey-Jennings ChartCalculate the Mean and Standard Deviation;
Record the Mean and +/- 1,2 and 3 SD Control Limits
80
85
90
95100
105
110
115
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
L J i Ch t
7/31/2019 QS -Quality Control
69/90
69
Levey-Jennings Chart -Record and Evaluate the Control Values
80
85
90
95
100
105
110
115
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
7/31/2019 QS -Quality Control
70/90
70
Findings Over Time
Ideally should have control values clustered aboutthe mean (+/-2 SD) with little variation in the upwardor downward direction
Imprecision = large amount of scatter about themean. Usually caused by errors in technique
Inaccuracy = may see as a trend or a shift, usuallycaused by change in the testing process
Random error = no pattern. Usually poor technique,malfunctioning equipment
7/31/2019 QS -Quality Control
71/90
71
Statistical Quality Control Exercise Hypothetical control values (2 levels of
control) Calculation of mean
Calculation of standard deviation
Creation of a Levey-Jennings chart
When does the Control Value
7/31/2019 QS -Quality Control
72/90
72
When does the Control Value
Indicate a Problem? Consider using Westgard Control Rules
Uses premise that 95.5% of control values
should fall within 2SD Commonly applied when two levels of control
are used
Use in a sequential fashion
7/31/2019 QS -Quality Control
73/90
73
Westgard Rules
Multirule Quality Control
Uses a combination of decision criteria or
control rules Allows determination of whether an analytical
run is in-control or out-of-control
7/31/2019 QS -Quality Control
74/90
74
Westgard Rules
(Generally used where 2 levels of controlmaterial are analyzed per run)
12S
rule
13S rule
22S rule
R4S rule
41S rule
10X rule
7/31/2019 QS -Quality Control
75/90
75
Westgard 12S Rule
warning rule
One of two control results falls outside 2SD
Alerts tech to possible problems Not cause for rejecting a run
Must then evaluate the 13S rule
1 R l A i i f l i i
7/31/2019 QS -Quality Control
76/90
76
12S Rule = A warning to trigger careful inspectionof the control data
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
12S rule
violation
7/31/2019 QS -Quality Control
77/90
77
Westgard 13S Rule
If either of the two control results falls outsideof 3SD, rule is violated
Run must be rejected If 13S not violated, check 22S
7/31/2019 QS -Quality Control
78/90
7/31/2019 QS -Quality Control
79/90
79
Westgard 22S Rule
2 consecutive control values for the samelevel fall outside of 2SD in the samedirection, or
Both controls in the same run exceed 2SD
Patient results cannot be reported
Requires corrective action
22S Rule = Reject the run when 2 consecutive control
7/31/2019 QS -Quality Control
80/90
80
22S Rule = Reject the run when 2 consecutive controlmeasurements exceed the same
+2SD or -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
22S ruleviolation
7/31/2019 QS -Quality Control
81/90
81
Westgard R4S Rule
One control exceeds the mean by2SD, andthe other control exceeds the mean by +2SD
The range between the two results willtherefore exceed 4 SD
Random error has occurred, test run must be
rejected
R4S Rule = Reject the run when 1 control
7/31/2019 QS -Quality Control
82/90
82
R4S Rule = Reject the run when 1 controlmeasurement exceed the +2SD and the other
exceeds the -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
R4S rule
violation
7/31/2019 QS -Quality Control
83/90
83
Westgard 41S Rule
Requires control data from previous runs
Four consecutive QC results for one level of
control are outside 1SD, or Both levels of control have consecutive results
that are outside 1SD
7/31/2019 QS -Quality Control
84/90
84
Westgard 10X Rule
Requires control data from previous runs
Ten consecutive QC results for one level of
control are on one side of the mean, or Both levels of control have five consecutive
results that are on the same side of the mean
10 Rule = Reject the run when 10 consecutive control
7/31/2019 QS -Quality Control
85/90
85
10x Rule = Reject the run when 10 consecutive controlmeasurements fall on one side of the mean
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
10x rule
violation
Westgard Multirule QC
7/31/2019 QS -Quality Control
86/90
86
Westgard Multirule QC
7/31/2019 QS -Quality Control
87/90
7/31/2019 QS -Quality Control
88/90
88
Solving out-of-control problems
Policies and procedures for remedial action
Troubleshooting
Alternatives to run rejection
7/31/2019 QS -Quality Control
89/90
89
Summary
Why QC program?
Validates test accuracy and reliability
Summary:
7/31/2019 QS -Quality Control
90/90
Summary:
How to implement a QC program? Establish written policies and procedures Assign responsibility for monitoring and reviewing Train staff Obtain control materials Collect data Set target values (mean, SD) Establish Levey-Jennings charts Routinely plot control data Establish and implement troubleshooting and corrective
action protocols Establish and maintain system for documentation