QUALITY ASSURANCE OF RADON MEASUREMENTS IN PORTUGAL
FILIPA DOMINGOS (M.Sc)
LABORATORY OF NATURAL RADIOACTIVITY
QA SYSTEM FOR RN MEASUREMENTS INDOORS AND IN WATER IN LNR (UC)
• LABORATORY OF NATURAL RADIOACTIVITY
• ESTABLISHED IN 1999 IN THE EARTH SCIENCES DEPARTMENT, UNIVERSITY OF COIMBRA
• ACCREDITED AS A TESTING LABORATORY (COMPLIANT
WITH ISO/IEC 17025) SINCE 2019:
Assay ISO Method Product
Radon 13164-4:2015 Liquid scintillation Water
Gross alpha and beta 11704:2018 Liquid scintillation Water
Ra-226 13165-1:2013 Liquid scintillation Water
U-238 and U-234 13169:2018 Liquid scintillation Water
Po-210 13161:2011 Alpha spectrometry Water
Sample collection - Internal method Water
Radon 11665-4:2012 Passive detectors (CR-39) Indoor
Ra-226, Th-232 and K-40 18589-3:2015 Gamma spectrometry Solid
QA SYSTEM FOR RN MEASUREMENTS INDOORS AND IN WATER IN LNR (UC)
• ELEMENTS OF THE QA SYSTEM
• INTERNAL QUALITY CONTROL:
• BLANK TESTS (INDOOR: UNEXPOSED DETECTORS; WATER: SCINTILLATION COCKTAIL)
• CERTIFIED REFERENCE MATERIALS (WATER)
• DUPLICATES (WATER)
• STANDARDS WITH KNOW ACTIVITY (INDOOR: DEMOSLIDE; WATER: BLANK, C-14 AND H-3)
• EXTERNAL QUALITY CONTROL:• PROFICIENCY TESTING:
• INTERCOMPARISON EXERCISES (BILATERAL ASSAYS)
• STATISTICAL ANALYSIS (INTERNAL/EXTERNAL DATA):• SHEWHART CHARTS (BLANKS, STANDARDS)
• WARNING AND ACTION LEVELS
• ESTIMATION OF THE ANNUAL PRECISION CRITERIA
Periodicity: for each batch of samples
QC in Sampling (contamination during sampling & transport)
and Sample preparation
QA SYSTEM FOR RN MEASUREMENTS INDOORS AND IN WATER IN LNR (UC)
• ELEMENTS OF THE QA SYSTEM
• INTERNAL QUALITY CONTROL:
• BLANK TESTS (INDOOR: UNEXPOSED DETECTORS; WATER: SCINTILLATION COCKTAIL)
• CERTIFIED REFERENCE MATERIALS (WATER)
• DUPLICATES (WATER)
• STANDARDS WITH KNOW ACTIVITY (INDOOR: DEMOSLIDE; WATER: BLANK, C-14 AND H-3)
• EXTERNAL QUALITY CONTROL:• PROFICIENCY TESTING:
• INTERCOMPARISON EXERCISES (BILATERAL ASSAYS)
• STATISTICAL ANALYSIS (INTERNAL/EXTERNAL DATA):• SHEWHART CHARTS (BLANKS, STANDARDS)
• WARNING AND ACTION LEVELS
• ESTIMATION OF THE ANNUAL PRECISION CRITERIA
Periodicity: for each batch of samples; 5 % of
all samples collected
QC in Sampling (contamination during transport) and Sample preparation; Estimation
of precision
QA SYSTEM FOR RN MEASUREMENTS INDOORS AND IN WATER IN LNR (UC)
• ELEMENTS OF THE QA SYSTEM
• INTERNAL QUALITY CONTROL:
• BLANK TESTS (INDOOR: UNEXPOSED DETECTORS; WATER: SCINTILLATION COCKTAIL)
• CERTIFIED REFERENCE MATERIALS (WATER)
• DUPLICATES (WATER)
• STANDARDS WITH KNOW ACTIVITY (INDOOR: DEMOSLIDE; WATER: BLANK, C-14 AND H-3)
• EXTERNAL QUALITY CONTROL:• PROFICIENCY TESTING:
• INTERCOMPARISON EXERCISES (BILATERAL ASSAYS)
• STATISTICAL ANALYSIS (INTERNAL/EXTERNAL DATA):• SHEWHART CHARTS (BLANKS, STANDARDS)
• WARNING AND ACTION LEVELS
• ESTIMATION OF THE ANNUAL PRECISION CRITERIA
Periodicity: according to the verification plan (ex. monthly); exceptionally
Check the equipment (drift and fitness to use);QC in data acquisition
(measurement)
QA SYSTEM FOR RN MEASUREMENTS INDOORS AND IN WATER IN LNR (UC)
• ELEMENTS OF THE QA SYSTEM
• INTERNAL QUALITY CONTROL:• BLANK TESTS (INDOOR: UNEXPOSED DETECTORS;
WATER: SCINTILLATION COCKTAIL)
• CERTIFIED REFERENCE MATERIALS (WATER)
• DUPLICATES (WATER)
• STANDARDS WITH KNOW ACTIVITY (INDOOR: DEMOSLIDE; WATER: BLANK, C-14 AND H-3)
• EXTERNAL QUALITY CONTROL:
• PROFICIENCY TESTING:
• INTERCOMPARISON EXERCISES (BILATERAL ASSAYS)
• STATISTICAL ANALYSIS (INTERNAL/EXTERNAL DATA):• SHEWHART CHARTS (BLANKS, STANDARDS)
• WARNING AND ACTION LEVELS
• ESTIMATION OF THE ANNUAL PRECISION CRITERIA
Periodicity: variable; depends on available
exercises/entities willing to do bilateral assays
QA SYSTEM FOR RN MEASUREMENTS INDOORS AND IN WATER IN LNR (UC)
Decision rules:• 1 value surpasses the
action level (AL);• 2 out of 3 consecutive
values surpass the warning level (WL);
• 6 consecutive values increase/decrease;
• 9 consecutive values in the same side of the mean;
• ELEMENTS OF THE QA SYSTEM
• INTERNAL QUALITY CONTROL:• BLANK TESTS (INDOOR: UNEXPOSED DETECTORS;
WATER: SCINTILLATION COCKTAIL)
• CERTIFIED REFERENCE MATERIALS (WATER)
• DUPLICATES (WATER)
• STANDARDS WITH KNOW ACTIVITY (INDOOR: DEMOSLIDE; WATER: BLANK, C-14 AND H-3)
• EXTERNAL QUALITY CONTROL:• PROFICIENCY TESTING:
• INTERCOMPARISON EXERCISES (BILATERAL ASSAYS)
• STATISTICAL ANALYSIS (INTERNAL/EXTERNAL DATA):
• SHEWHART CHARTS (BLANKS, STANDARDS, CRM)
• WARNING AND ACTION LEVELS
• ESTIMATION OF THE ANNUAL PRECISION CRITERIA
Periodicity: calculated with 20 results; updated
upon new data;
CALIBRATION OF RN MEASUREMENT INSTRUMENTS IN LNR (UC)
• CALIBRATION SYSTEM
• ALL EQUIPMENT WITH DIRECT INFLUENCE ON THE TEST RESULTS IS SUBJECT TO PERIODIC CALIBRATION/VERIFICATION/MAINTENANCE, ACCORDING TO PRE-DEFINED PLANS;
• THE PERIODICITY IS ESTABLISHED BY:
• MANUFACTURER'S RECOMMENDATIONS;
• TREND ANALYSIS OF QC DATA (EX: INSTRUMENTAL SHEWHART CHARTS, INTERLABORATORY COMPARISONS);
• PREVIOUS EXPERIENCE;
• FREQUENCY, TYPE AND CONDITIONS OF USE OF THE EQUIPMENT;
• ENVIRONMENTAL CONDITIONS (TEMPERATURE, HUMIDITY, VIBRATION, ETC.);
• DESIRED DEGREE OF ACCURACY/THE MAXIMUM ALLOWABLE ERROR;
• ACCEPTANCE/EVALUATION OF CALIBRATION CERTIFICATES;
• ORGANIZATIONS RESPONSIBLE
• CALIBRATION CERTIFICATES;
• CERTIFIED REFERENCE MATERIALS (EX. DEMOSLIDE, H-3, C-14 & BLANK)
EXPERIENCE OF INTERCOMPARISON OF PASSIVE OR ACTIVE MEASUREMENT INSTRUMENTS
• WHEN PARTICIPATED IN WHICH INTERCOMPARISON
External quality control Year Assay Product Type Organiser
Proficiency testing 2017 Radon Indoor Passive PHE, England
Interlaboratory comparison 2018 Radon Indoor Passive MetroRadon, Spain
Interlaboratory comparison 2018 Radon Indoor Active NRPI, Czech Republic (IAEA)
Proficiency testing 2019 Radon Water - IPROMA, Spain
Proficiency testing 2019 Radon Indoor Passive BfS, Germany
Interlaboratory comparison 2020 Radon Water - LARUC
Interlaboratory comparison 2020 Radon Indoor Passive LARUC
Proficiency testing 2020 Radon Water (Sampling) Relacre, Portugal
EXPERIENCE OF INTERCOMPARISON OF PASSIVE OR ACTIVE MEASUREMENT INSTRUMENTS
• HOW DO YOU USE THE RESULTS OF INTERCOMPARISON :
• ASSESS ACCURACY;
• MONITOR/DETECT SYSTEMATIC ERRORS;
• IDENTIFY THE NEED FOR EQUIPMENT MAINTENANCE/VERIFICATION;
• MONITOR HUMAN ERRORS (EX. SAMPLE PREPARATION/HANDLING);
• EVALUATE/COMPARE THE LABORATORY PERFORMANCE;
• INTERNAL QUALITY CONTROL;
• ASSESS THE PERFORMANCE OF LABORATORY TECHNICIANS;
• QUALIFICATION OF LABORATORY TECHNICIANS (ANNUALLY);
EXPERIENCE OF INTERCOMPARISON OF PASSIVE OR ACTIVE MEASUREMENT INSTRUMENTS
• PROFICIENCY TESTING SCHEME AS AN ORGANISER:
• ASSAY: RADON IN WATER SAMPLES;
• OBJECTIVE: QUALITY CONTROL IN SAMPLING;
• STRATEGY: SAMPLE COLLECTION OF SPRING WATER AT THE SAME LOCATION BY ALL PARTICIPANTS
• ORGANISER: RELACRE (ASSOCIATION OF PORTUGUESE ACCREDITED LABORATORIES)
• RESPONSIBILITIES OF LNR AS A PROFICIENCY TESTING PROVIDER:
• STABILITY AND HOMOGENEITY STUDIES FOR THE SELECTION OF THE SAMPLING LOCATION;
• STABILITY AND HOMOGENEITY STUDIES ON THE DAY OF SAMPLE COLLECTION;
• DETERMINATION OF RADON IN ALL WATER SAMPLES COLLECTED AND DELIVERED BY THE PARTICIPATING LABORATORIES;
• PROTOCOL: SAMPLING ACCORDING TO THE RECOMMENDATION 1/2017 (ERSAR), DECREE-LAW 152/2017 (7TH DECEMBER), AND ISO 5667-5:2006;
• PARTICIPATION OF 18 LABORATORIES (54 WATER SAMPLES COLLECTED);
CHALLENGES FOR GOOD LABORATORY PRACTICES AND QUALITY ASSURANCE OF RN MEASUREMENT INSTRUMENTS IN LNR (UC)
• METHODS FOR SAMPLING FOR RADON IN WATER SAMPLES (SOURCE OF VARIABILITY/ UNCERTAINTY)
• SAMPLING OF INDOOR RADON IN REAL CONDITIONS (INTERCOMPARISON EXERCISES ARE HELD IN “IDEAL” CONDITIONS)
• EXECUTION OF INTERCOMPARISON EXERCISES DEPENDS ON AVAILABILITY (LOW PERIODICITY AND LONG WAIT FOR RESULTS)
• CONTAMINATION DURING TRANSPORT IN INTERCOMPARISON EXERCISES:• RESPONSIBILITY IDENTIFY
• HOW TO INTERPRET/ANALYSE THE DATA?
• DECISION RULES IN SHEWHART CHARTS/ACCEPTABILITY CRITERIA ARE SUBJECTIVE;
• FUTURE CHALLENGE FOR THE LNR: COMPLY WITH THE REQUIREMENTS OF ISO/IEC 17043 FOR THE COMPETENCE OF PROVIDERS OF PROFICIENCY TESTING SCHEMES, THEIR DEVELOPMENT AND OPERATION