2013/01 Stryker IFU P23276 A VSI IFU 13653 A (00413) Safe Handling and Hospital Transportation Shipping Handle the flexible tip with care. Never place items on the tip. Never push or hit the tip against hard surfaces. Always use the levers to articulate the tip—never bend the tip manually. Never place or stack other devices on the scope. The scope is delicate and can be damaged by other devices or from rough handling. To transport the scope within a hospital, carefully place it in Stryker sterilization tray 0233-032-113. Secure the scope in the silicone tabs. Ship the scope only in Stryker carrying case 0502-110-123. Ensure the vent cap is attached, or damage will result. Bends and Kinks Liquid Immersion Before immersing the scope in liquid for cleaning, disinfection, or sterilization, place the soaking cap over the video connector. Never attach the vent cap to the scope during liquid immersion, or damage will result. Carefully place the scope in the cleaning basin or sterilization tray. Do not sharply bend the insertion tube. Before sterilization in liquid, perform a leak test. Return any scope with leaks to Stryker for evaluation. Gas Sterilization Do Not Autoclave or Auto-Clean AUTOCLAVE AUTOMATED CLEANING If excessive force is required to pass the cleaning brush through a kink in the working channel, return the scope to Stryker for evaluation. Before gas sterilization, perform a leak test. Return any scope with leaks to Stryker for evaluation. Always attach the vent cap to the scope during gas sterilization. Failure to follow this instruction will damage the scope. Carefully place the scope in Stryker sterilization tray 0233-032-113. Secure the scope in the silicone tabs. Arrange the video cable along the perimeter of the tray for maximum sterilant exposure. Do not autoclave the scope. Do not use an automated cleaning method. The scope must be manually cleaned. After being secured in the sterilization tray, the insertion tube might retain a curved shape. This does not indicate damage; the scope should operate normally. Quick Reference Guide IDEAL EYES ® HD URT-7000S/7000Si Flexible Video Ureteroscope 0502110111, 0502110112. See Instructions For Use (PN #P18188) for complete reprocessing and handling instructions. 5900 Optical Court, San Jose, CA 95138 USA, +1.800.624.4422, +1.408.754.2000, www.stryker.com
Transcript
Page 1: Quick Reference Guide HD URT-7000S/7000Si Flexible Video ......IDEAL EYES® HD URT-7000S/7000Si Flexible Video Ureteroscope 0502110111, 0502110112. See Instructions For Use (PN #P18188)

2013/01 Stryker IFU P23276 A

VSI IFU 13653 A (00413)

Safe Handling and Hospital Transportation Shipping

Handle the flexible tip with care. Never place items on the tip. Never push or hit the tip against hard surfaces. Always use the levers to articulate the tip—never bend the tip manually.

Never place or stack other devices on the scope. The scope is delicate and can be damaged by other devices or from rough handling.

To transport the scope within a hospital, carefully place it in Stryker sterilization tray 0233-032-113. Secure the scope in the silicone tabs.

Ship the scope only in Stryker carrying case 0502-110-123. Ensure the vent cap is attached, or damage will result.

Bends and Kinks Liquid Immersion

Before immersing the scope in liquid for cleaning, disinfection, or sterilization, place the soaking cap over the video connector.

Never attach the vent cap to the scope during liquid immersion, or damage will result.

Carefully place the scope in the cleaning basin or sterilization tray. Do not sharply bend the insertion tube.

Before sterilization in liquid, perform a leak test. Return any scope with leaks to Stryker for evaluation.

Gas Sterilization Do Not Autoclave or Auto-Clean

AUTOCLAVE

AUTOMATED CLEANING

If excessive force is required to pass the cleaning brush through a kink in the working channel, return the scope to Stryker for evaluation.

Before gas sterilization, perform a leak test. Return any scope with leaks to Stryker for evaluation.

Always attach the vent cap to the scope during gas sterilization. Failure to follow this instruction will damage the scope.

Carefully place the scope in Stryker sterilization tray 0233-032-113. Secure the scope in the silicone tabs. Arrange the video cable along the perimeter of the tray for maximum sterilant exposure.

Do not autoclave the scope.

Do not use an automated cleaning method. The scope must be manually cleaned.

After being secured in the sterilization tray, the insertion tube might retain a curved shape. This does not indicate damage; the scope should operate normally.

Quick Reference Guide IDEAL EYES® HD URT-7000S/7000Si Flexible Video Ureteroscope 0502110111, 0502110112. See Instructions For Use (PN #P18188) for complete reprocessing and handling instructions. 5900 Optical Court, San Jose, CA 95138 USA,

+1.800.624.4422, +1.408.754.2000, www.stryker.com