XM Canine PROCEDURETEST
N°1: PREPARATION OF BLOOD SAMPLES (MAJOR = Donor RBC + Recipient serum)
Recipient Donor
Serum or plasma pRBCs
(packed Red Blood Cells)
Whole blood• EDTA• ACD
Do not use Heparin
• CPD• Plain tube
Centrifuge blood tube in order to collect serum or plasma.
Centrifuge blood tube.Discard the serum or plasma to collect pRBCs.
OR
Collect blood from blood bag segment.
4
pRBCs
Mix gently & incubate 10’ at room temperature
N°2: INCUBATION AND WASHING PROCEDURE
1
4
In the clean test tube
10 min
1 Wash Buffer
1 white top buffer 1
1 Blood collector strip
2 large pipettes
1 small pipette
Material provided:
1 green top buffer 2
1 Clean test tube (1,5ml)
2Bu�er
1Bu�er
R D
XM
1
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
1 XM Quick Test
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
2
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5With blood collector strip
Donor’s tubeOR
Donor’s blood bag segment
EDTA
D
3
Recipient’s tube
With the small pipette
3Serum or plasma
R
in the test tube
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
1 Empty well
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
7s
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
MAJOR
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
XM
5
A
Suspension washing procedure
XM
CB
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
1,5ML
With a large pipette
2min
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
Don’t forget the balance tube(not provided)
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
Discard the supernatant
Add
Centrifuge
1000 g or2300 rpm
STEP 5 MUST BE DONE 3 TIMES
N°3: XM TEST PROCEDURE
1 2
3
3
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
4 5
lot: 154684769.25.63
Pa
t :
Da
te :
6
lot: 154684769.25.63
Pa
t :
Da
te :
2 to 5min
WAIT UNTIL COMPLETE MIGRATION
7
lot: 154684769.25.63Exp: 2009.06
Pa
t :
Da
te :
!
+
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :-
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
C = control
: www.alvedia.com - [email protected]
Contact us to ask for the
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
Result Interpretation
The XM test line will often be weaker than the control line
Collect entire suspensionwith a large pipette.
Fill the well.
Read immediately after testing.
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
lot: 154684769.25.63Exp: 2009.06 Pat :
Date :
And mix gently
2
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
1,2
ML
0,5
Mix gently
Incompatibility Compatibility