QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy Marion, MA
Transcript
Page 1: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

QUIKLOOK II SoftwareQUIKLOOK Users' Group

2011 Annual Meeting

August 18-19,2011Tabor Academy

Marion, MA

Page 2: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

QUIKLOOK II Software

● New Features 2010

● New Features 2011

● Software Error Notices

● User discussion, questions, and suggestions

Page 3: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Version 2010.227– Released first week September.

Page 4: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Acquisition– For AOV after Start is click the control

signal is set to the initial value then a second prompt to click to continue when valve is in position

– For AOV tests instead of showing text boxes above show channels values similar to monitor screen and also include test time, elapsed time & time remaining in bigger font to read from a distance.

Page 6: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Configuration– Custom Wave Form Library (AOV)

Page 7: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Configuration– Calc String Pot Sensitivity for rotary valves

Page 8: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Configuration– Add System Data Form with M&TE field

Page 10: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Replay– Overlay tests from different directories

Page 11: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Replay– Ability to change lines to dots or triangles– Ability to change line weights and color on

trace by trace basis.

Page 13: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Replay– Multiple Monitors – Ensure that popup

windows occur on same monitor– AOV - Ability to trend Friction and other

calculated channels

Page 14: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● Analysis– Export To Excel AOV– Export to Excel with Office 2007– AOV - Variable Moment Arm– AOV - Variable Actuator Area

Page 15: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● QUIKLOOK Reports– All Reports

• Added the ability to display a company logo along side Teledyne logo

Page 16: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● QUIKLOOK Reports– QLReportAOV

• Added Transducer Report • Added Overall Report • Added Positioner Report • Added Mechanical Properties Report • Added Drop Test Report• Added Calibration Report• Added Configuration Report

Page 17: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2010

● QUIKLOOK Reports– QLReportTrend

• New Report• Added to QUIKLOOK ability to extract multiple

test data from tests selected for trending from Test Listing or Replay

Page 21: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2011

● Acquisition – Monitor Screen– Zero all pressure channels option.– Prompt to save config if changed.– Warning if pressure channels are not

zeroed.

Page 22: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2011

● Test Menu– Combine test

• Combine two tests end to end• Split test into two tests• Crop a test saving only data between two

markers

• In each case a new test file will be created leaving the original tests intact.

Page 23: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2011

● Configuration– Load Sensor – Remember filter

● Replay– Plot between markers (xy plots)

Page 24: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

New Features 2011

● Analysis– Automark MOV traces

Page 25: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Version 2010.227– Error Notice 2010.227-1

• Pretension graph does not work properly .• Pretension value units show mV/V but the actual

number reported is in V/V .• Sensitivities for sensors using mV ranges are

shown as volts but should be mV

Note: This error only occurs on acquisition systems which utilize the new Validyne board, model 160026V. This error does not apply to systems using the Keithley board, model 160026, or to desktop systems

Page 26: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Error Notice 2010.227-1– Workaround:

• When pretensioning a C-Clamp use the numbers above the graph. When they are >=0.0027 and <= 0.003 then the clamp is properly pretensioned

• Ignore mV/V label since voltage is V/V• Ignore Sensitivity units. Use configuration

screen to check units.

Page 27: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Version 2010.227– Error Notice 2010.227-2

• When testing a Double Acting AOV if the Top & Bottom Cylinder areas are set to the same value then an error will occur. This will not cause any errors in the results.

– Workaround: • Since Top & Bottom Cylinder areas should not be

the same, the bottom is smaller by the area of the piston rod, set the areas to different values. If actual areas are not available then set bottom area equal to top area minus 0.001 sq in.

Page 28: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Version 2010.227– Error Notice 2010.227-3

• A filter on a trace acquired at an acquisition rates of 2,000 Hz, 5,000 Hz or 10,000 Hz trace will not produce correct results .

– Workaround: • There is no workaround. Do not use the filter

function on traces acquire above a 1,000 Hz acquisition rate .

Page 29: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Error Notice 2010.227-3– This error applies to all versions of

QUIKLOOK– The error involved varies depending on

acquisition rate, noise frequency and filter frequency. Each filtered trace should be compared to the original trace to access the inaccuracies involved.

Page 30: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Error Notice 2010.227-3– QLReportFindTests

Page 32: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Version 2010.227– Error Notice 2010.227-4

• When entering the acquisition screen you may get a message of “Low Excitation Voltage”.

• Background: When entering the acquisition screen QUIKLOOK first resets the acquisition boards, uploads the configuration then checks the excitation. The reset of the acquisition boards causes the excitation to change and if no load is attached to the system, the voltage does not settle down for approximately two seconds. Occasionally the excitation voltage check will happen before the voltage settles down and causes the low excitation message

Page 33: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

Software Error Notices

● Error Notice 2010.227-4

– Workaround:• Workaround: Make sure you have your

sensors attached before going into the acquire screen. If this message appears go to the Monitor Screen. Check the excitation message in the lower right corner. If it states : "Excitation Voltage OK" then the excitation voltage is OK.

• Note: This error only applies to systems containing the Validyne boards.

Page 34: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

● When printing trace, the cursor still shows. Can you provide an option to NOT SHOW the cursor when printing traces?

User Suggestion

Page 35: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 36: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 37: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 38: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

● Can running loads and stroke times be added as part of the CTG configuration file, such that they are already preselected for each valve? I give the config files to my field testers before they test, and I have to spend the time adding all the points into each of the traces after they take them. Being able to do it in advance (once) and having it consistent would be a big plus.

User Suggestion

Page 39: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

● Another benefit here is that the field testers have the ability to go to the Analysis....Running Loads and Stroke Times and HAVE THESE VALUES ALREADY THERE. This would be a huge benefit to field testing.

User Suggestion

Page 40: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 41: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 42: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 43: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 44: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

● Contained in Quiklook.mdb file located in Quiklook directory

● Copy Quiklook.mdb from computer with Running Loads & Stroke Time customized

● Load into each acquisition system

User Suggestion

Page 45: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

● All Times will be added to each test at time of acquisition

● All Running Loads, where the channel name matches a channel name acquired, will be added to each test at time of acquisition

User Suggestion

Page 46: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 47: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

User Suggestion

Page 48: QUIKLOOK II Software QUIKLOOK Users' Group 2011 Annual Meeting August 18-19,2011 Tabor Academy

● User discussion, questions, and suggestions