IAEAInternational Atomic Energy Agency
RADIATION PROTECTION INDIAGNOSTIC AND
INTERVENTIONAL RADIOLOGY
Part 18: Optimization of protection in CT scanner
Practical exercise - Quality Control
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 2
Contents
Quality control tests
• CT accuracy, uniformity, linearity. and noise,
• Low and high contrast resolution
• Z-axis sensitivity
• Alignment, table top travel accuracy
• Gantry tilt measurement
• Dosimetry
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 1: Quality Control
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 4
Physical parameters of CT image
Image quality• May be expressed in terms of physical parameters such
as uniformity, linearity, noise, spatial resolution, low contrast resolution
• Image quality depends on the technical characteristics of the CT scanner, the exposure factors used and image viewing conditions.
• Quality may be assessed by quantitative measurement using test phantoms, and by the appearance of artifacts.
• Measurements should be conducted regularly
IAEA 18: Optimization of protection in CT scanner 5
Scanner performance: technical parameters (I)
Test Phantoms:• Test phantoms of a standardized human shape
or test objects of a particular shape, size and structure, are used for the purposes of calibration and evaluation of the performances of CT scanners
• Phantoms should allow for evaluation of CT number; uniformity; noise; spatial resolution; slice thickness; dose; positioning of table top
IAEA 18: Optimization of protection in CT scanner 6
Scanner performance: technical parameters (II)CT Number Accuracy
• CT number depends on tube voltage, filtration, object thickness• CT number of water is by definition equal to 0 • Measured CT number should be < ± 4 HU in the central ROI
CT Number Linearity• The linear relationship between the calculated CT number and the
linear attenuation coefficient of each element of the object• Deviations from linearity should be < ± 5 HU
CT Number Uniformity• The CT number of each pixel in the image of an homogeneous object
should be the same over the image area• The difference in the CT number between a peripheral and a central
region of an homogeneous test object should be < 8 HU• Differences are largely due to beam hardening
IAEA 18: Optimization of protection in CT scanner 7
Noise• The local statistical fluctuation (standard deviation) of CT
numbers in a homogeneous Region Of Interest (ROI)• Noise strongly affects the low contrast resolution• Noise is dependent on the radiation dose
• Image noise should be measured over an area of about 10% of the cross-sectional area of the test object.
• Goal— obtain an image with an acceptable level of noise while keeping the patient dose as low as reasonably achievable
Scanner performance: technical parameters (III)
doseNoise 1
IAEA 18: Optimization of protection in CT scanner 8
Scanner performance: technical parameters (IV)
Spatial Resolution• The high contrast resolution determines the minimum
size of details visualized in the plane of the slice with a contrast >10%.
It is affected by:
• the reconstruction algorithm
•the detector width
• the effective slice thickness
•the object to detector distance
• the x-ray tube focal spot size
•the matrix size.
IAEA 18: Optimization of protection in CT scanner 9
Scanner performance: technical parameters (V)
Spatial Resolution• The low contrast resolution determines the size
of detail that can be visibly reproduced when there is only a small difference in density relative to the surrounding area• Low contrast resolution is limited by noise.
• The perception threshold in relation to contrast and detail size can be determined by creating contrast-detail curve.
IAEA 18: Optimization of protection in CT scanner 10
Scanner performance: technical parameters (VI)
Slice Thickness• The slice thickness is determined in
the center of the field of view as the distance between the two points on the sensitivity profile along the axis of rotation at which response has fallen to 50%.
• The use of post-patient collimation to reduce the width of the image slice leads to very significant increases in the patient dose
Positioning of couch• The accuracy of positioning of the
patient couch is evaluated by moving the loaded couch a defined distance and moving it back to the start position
IAEA 18: Optimization of protection in CT scanner 11
Minimum requirements: CT scanner (I)
• Image noiseThe Standard Deviation of CT numbers in the central 500 mm2 ROI for a water or tissue equivalent phantom should not deviate more than 20% from the baseline.
• CT number valuesThe deviation in the CT number values for water or tissue equivalent material and materials of different densities should <± 20 HU or 5%.
• CT number uniformityThe SD of the CT number over a 500 mm2 region of interest for water or tissue equivalent material at the center and at the periphery of phantom should vary by less than 1.5% of the baseline
IAEA 18: Optimization of protection in CT scanner 12
Minimum requirements: CT scanner (II)
• Computed tomography dose index (CTDI)
The CTDI for a single slice for each available beam shaping filter and for each available slice thickness should not deviate more than ± 20% from the baseline.
• Irradiated slice thickness
The FWHM of the dose profile should not differ more than ± 20% from baseline.
• High contrast resolution (spatial resolution)
The FWHM of the point spread function of a pin, or the edge response function of an edge should not differ more than ± 20% from baseline.
• Low contrast resolution
Polystyrene pins of 0.35 cm diameter inserted in a uniform body water phantom should be visible in the image.
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 2: Noise
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 14
Imaging performance (Noise)
• Noise is assessed using cylindrical phantoms, which are either filled with water or made of a tissue equivalent material
• Once an axial image of the phantom has
been acquired, noise is measures as the standard deviation in CT number in a region of interest (ROI) placed in the center of the image
IAEA 18: Optimization of protection in CT scanner 15
Region of Region of interest interest (ROI)(ROI)
Imaging performance (Noise)
IAEA 18: Optimization of protection in CT scanner 16
Imaging performance (Noise)
• Noise values are given in manufacturers’ specifications are quoted for a specific phantom (e.g., manufacturer’s QA phantom) and for specified scan techniques
• These conditions must be matched exactly for the purposes of the acceptance test
• Manufacturers quote noise at a particular surface dose
• If this is the case, dose for axial scans can be measured by taping an ion chamber to the surface of the phantom
IAEA 18: Optimization of protection in CT scanner 17
Imaging performance (Noise)
• Baseline noise values should be obtained for several scan protocols that will be used clinically, using the routine QC noise phantom
• To ensure that noise figures are both accurate and representative, it is essential to find the mean value from several scans (10 scans.)
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 3: CT uniformity and linearity
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 19
CT number uniformity
• CT number uniformity can be assessed at the same time as measuring noise, by placing four additional ROI (N, E, S and W) at positions near the edge of the image of a uniform phantom
• The mean CT number is then noted for these four regions, as well as the center
• The deviation from the central value should be calculated
• The CT number uniformity should be measured for large fields of view
IAEA 18: Optimization of protection in CT scanner 20
Axial image of Axial image of a homogenous a homogenous phantomphantom
CT number uniformity
IAEA 18: Optimization of protection in CT scanner 21
CT number uniformity
• CT number linearity is assessed using a phantom containing inserts of a number of different materials (materials should cover a wide range of CT numbers)
• One example of a suitable phantom to use at acceptance is the Catphan (The Phantom Laboratory, Salem, NY), which contains four inserts with CT numbers ranging from approximately -1000HU to +1000HU
IAEA 18: Optimization of protection in CT scanner 22
CT number uniformity
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 4: Low and high contrast resolution
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 24
Low contrast resolution
• Low contrast resolution (or low contrast detectability) is often quoted in specification documentation as the smallest visible object at a given contrast for a given dose
• Since this measurement relates directly to imaging performance, it is an important value to verify at acceptance testing
• At least 20 images of the low contrast insert (LCR) should be acquired and then viewed by at least 3 observers under optimal viewing conditions to obtain an average
IAEA 18: Optimization of protection in CT scanner 25
Typical Typical image of the image of the Catphan LCR Catphan LCR insertinsert
Low contrast resolution
IAEA 18: Optimization of protection in CT scanner 26
Spatial resolution (high contrast)
• There exist two broad categories of measuring techniques : • analysis of the point spread function, usually by
calculation of the modulation transfer function (MTF) • either objective analysis or visual assessment of images
of a resolution bar phantom.
• The resolution is quoted as the spatial frequency (in line pairs/cm) at which the modulation falls to 50%, 10% or 2% MTF.
• These figures are often given for more than one reconstruction algorithm, e.g., for standard and high-resolution scans.
IAEA 18: Optimization of protection in CT scanner 27
Spatial resolution (high contrast)
• The number of line pairs per cm just visible in the image is approximately equivalent to the 2% value of the MTF
• This result can then be compared with the 2% MTF, as quoted in the manufacturer’s specifications
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 5: Z-axis sensitivity
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 29
Z-Sensitivity (Imaged slice width)
Axial measurements • Phantoms used for axial measurements may contain
thin metal plates, wires or arrays of air holes, inclined at an angle to the image plane
• Manufacturers should be able to supply an appropriate phantom or, alternatively, the Catphan contains an insert suitable for this test
Note: to obtain meaningful measurements, the thickness of the plates, wires or holes cannot be greater than the nominal slice width concerned. This may create problems for the sub-millimetre slice widths offered on multi-slice scanners.
IAEA 18: Optimization of protection in CT scanner 30
Z-Sensitivity (Imaged slice width)
Axial measurements • Phantoms manufacturers may quote the tolerance
for each nominal slice width setting in their specification documentation
• Z-sensitivity measurements in axial mode can be used to check that imaged slice widths are within the tolerances given
• This can also be used in conjunction with irradiated slice width measurements to assess the accuracy of post patient collimation and to calculate the geometric efficiency for the scanner
IAEA 18: Optimization of protection in CT scanner 31
Plan view of a test object used to measure imaged slice widths for axial scans
Z-Sensitivity (Imaged slice width)
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 6: Alignment
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 33
Alignment of indicating lights with scan, coronal and sagittal planes
• Several methods can be used to perform these tests
• The techniques described here are straightforward to implement and require minimal test equipment.
IAEA 18: Optimization of protection in CT scanner 34
Agreement between internal and external scan plane lights
• Use an envelope-wrapped film (therapy verificaiton film) for the measurement. However, a piece of paper or card can also be used
• The film is placed flat on the table and illuminated by the external scan plane light
• The position of the light is marked on the film envelope and the table is moved automatically to the scan plane
• If the distance between the internal and external lights is correct, the internal light should now coincide with the mark on the film envelope.
IAEA 18: Optimization of protection in CT scanner 35
Co-incidence of internal scan plane lights and scan plane
• Pin pricks are made in a piece of therapy verification film (or similar) along the line of the internal scan plane light
• The film is wrapped around a phantom and scanned with a narrow axial and developed
• Coincidence between the pin pricks and the x-ray beam exposure indicates alignment between the alignment lights and the scan plane
IAEA 18: Optimization of protection in CT scanner 36
Co-incidence of internal scan plane lights and scan plane
• For multi-slice scanners, pin pricks (and, thus, the internal alignment light) are usually found to coincide with the center of the four slices
• To plan a scan so that the x-ray beam is centered over the internal scan plane lights at zero, it will be necessary to center the first and last slices symmetrically around zero (e.g. four slices from –7.5 mm to +7.5 mm on a 4 x 5 mm scan)
IAEA 18: Optimization of protection in CT scanner 37
Z
X
Pin pricks made in film at position of scan plane light
X-ray beam exposure
Co-incidence of internal scan plane lights and scan plane
IAEA 18: Optimization of protection in CT scanner 38
Coronal and Sagital plane lights
• A long, thin object, with a high CT number relative to air, such as the ‘lead’ in a pencil or a straightened paper clip, can be used as a marker to perform this test
• The marker is supported above the patient table and aligned, using the indicating lights, so that it is positioned at the isocentre, parallel to the z-axis and perpendicular to the scan plane
• If indicating lights are accurately aligned to the coronal and sagital planes, the marker should appear as a dot at exactly x = 0, y = 0 on the axial image.
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 7: Table travel accuracy
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 40
Table travel accuracy
• A ruler or tape measure placed alongside the table, can be used to check that the amount of table movement indicated on the gantry agrees with the actual distance moved.
• A load of approximately 70- 80 kg should be placed on the table in order to simulate the weight of a patient.
• The test should be performed twice: by driving the table top both away from and towards the gantry.
IAEA 18: Optimization of protection in CT scanner 41
Assessment of distance indicator accuracy
Table travel accuracy
IAEA 18: Optimization of protection in CT scanner 42
Axial scan incrementation accuracy
• Verification of incrementation accuracy between successive axial slices can be achieved by placing therapy verification film on the couch (in the isocentre plane) and exposing it to an axial scan sequence
• Narrow slices separated by a couch increment greater than 1 slice width can be used, and the distance between the lines on the film measured
IAEA 18: Optimization of protection in CT scanner 43
Table travel accuracy for helical scans
• In helical scanning, it is not sufficient to use a simple mechanical test because the distance imaged depends on couch speed and scanner software
• One method of assessing distance accuracy is to use an acrylic test object containing two small radio-opaque markers, separated by a fixed distance (e.g., 20 cm)
• The test object is scanned in Scan Projection Radiography (SPR) mode and a helical scan is planned to start at the first marker and to end at a distance x from the first marker
• If couch travel is accurate during the helical scan, the markers should be clearly seen on the first and final images of the series.
IAEA 18: Optimization of protection in CT scanner 44
Couch travel accuracy for helical scans
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 8: Gantry tilt measurement
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 46
Gantry Tilt
• The accuracy of displayed gantry tilt can be assessed by supporting therapy verification film at the gantry end of the patient table
• The film must be held vertically (e.g., by taping to an acrylic block), so that it is parallel to the sagital plane and intersects scan and coronal planes at right angles
IAEA 18: Optimization of protection in CT scanner 47
Gantry Tilt
• Three axial exposures are made using the same film:• one for the maximum superior gantry tilt,
• one for the maximum inferior gantry tilt
• one at 0º gantry tilt
• The three scan planes should then be visible on the developed film
• The angles + and - between scan planes at maximum tilt relative to that at 0º tilt should equal tilt angles displayed on the gantry.
IAEA 18: Optimization of protection in CT scanner 48
Assessment of accuracy of gantry tilt
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CT scanner
Topic 9: Dosimetry
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 18: Optimization of protection in CT scanner 50
Dosimetry - CTDI in air
• The Computed Tomography Dose Index (CTDI) in air can be measured using a 10 cm pencil ionization chamber, located in the scan plane at the isocentre
• The ion chamber can be supported using a laboratory test stand and clamp
IAEA 18: Optimization of protection in CT scanner 51
Ionization chamber
Table
Dosimetry - CTDI in air
IAEA 18: Optimization of protection in CT scanner 52
Dosimetry - CTDI in air
• It is important to measure the CTDI in air under the following conditions When commissioning a CT scanner:• For all beam shaping filters
• For all nominal slice widths
• For all clinical kV settings
• For a range of scan times
• For a range of mA settings
IAEA 18: Optimization of protection in CT scanner 53
Dosimetry - CTDI in air
IAEA 18: Optimization of protection in CT scanner 54
Dosimetry - CTDI in air
IAEA 18: Optimization of protection in CT scanner 55
Axial slice positionsAxial slice positions
Helical scan (pitch 1)Helical scan (pitch 1)
Dosimetry - CTDI in air (helical)
IAEA 18: Optimization of protection in CT scanner 56
Dosimetry - CTDI in Acrylic Phantoms
• The body phantom is placed on the patient table and the head phantom is supported in the head rest
• Phantoms are aligned and centred at the scan isocentre
• The ion chamber is inserted into either the central or one of the peripheral holes of the phantom (all other holes being filled with acrylic rods)
IAEA 18: Optimization of protection in CT scanner 57
Dosimetry - CTDI in Acrylic Phantoms
• Dose measurements at the center are used to calculate the central CTDI
• Peripheral CTDI is measured in at least four positions around the phantom, so as to achieve a true average
Note: If gantry rotation is initiated from different angular positions for successive scans it may be necessary to take a number of measurements at each position in order to get a representative mean dose
IAEA 18: Optimization of protection in CT scanner 58
Head phantomHead phantomBody phantom Body phantom (or annulus (or annulus
to fit over head phantom)to fit over head phantom)
Insert to plug holesInsert to plug holes
Dosimetry - CTDI in Perspex Phantoms
IAEA 18: Optimization of protection in CT scanner 59
• Central and peripheral CTDI’s are used to calculate weighted CTDI, CTDIw:
• CTDIws can be compared against diagnostic reference levels for standard scan examinations
Dosimetry - CTDI in Perspex Phantoms
)( CTDI3
2 + CTDI
3
1
C
1 = CTDI p100,c100,wn
IAEA 18: Optimization of protection in CT scanner 60
Example of CTDI Values for some CT units
MODEL KV FAD S nCDTI (mGy/mAs) Air weighted
Head Body Siemens HIQ
133 700 10 0.195 0.161 (0.83) 0.093 (0.48)
Siemens Plus S
120 700 10 0.128 0.110 (0.86) 0.062 (0.48)
Siemens Plus S
137 700 8 0.161 0.082 (0.51)
GE Pace 120 525 10 0.344 0.200 (0.58) 0.094 (0.27)
Philips LX
120 606 10 0.200 0.160 (0.80) 0.081 (0.41)
FAD(mm): Focus to Axis distanceS (mm): Slice thickness
IAEA 18: Optimization of protection in CT scanner 61
Dose Profiles (Irradiated slice width)
• Measurement of irradiated slice widths, for all nominal slice width settings, provides a direct test of pre-patient beam collimation and allows geometric efficiencies to be calculated for the scanner
• Geometric efficiency (GE) is defined as:GE = width slice irradiated x 100 %
width slice imaged• It is recommended that GE value is displayed on
the console if it is less than 70%
IAEA 18: Optimization of protection in CT scanner 62
Dose Profiles (Irradiated slice width)
• First method : expose therapy verification film, supported in air at the isocentre, at each of the slice width settings
• Once developed, optical density profiles may be plotted using a digitizer or the width measured using a ruler
• To provide an accurate dose profile, a calibration curve can be applied to convert optical density profiles into dose profiles, from which irradiated slice widths (FWHM of dose profiles) may be derived
• Alternative technique: measure dose profiles using TLD’s.
IAEA 18: Optimization of protection in CT scanner 63
Measurement of Measurement of irradiated slice irradiated slice
widths for a range widths for a range of nominal slice of nominal slice width settingswidth settings
Dose Profiles (Irradiated slice width)
IAEA 18: Optimization of protection in CT scanner 64
Quality control program
Example of a quality control program for a CT unit:
Baseline ± 20%H1 yBHigh contrast resolution
> ± 20% or 1 mmM1 yBImage slice thickness
> ± 20% or 1 mmH1 yBIrradiated slice thickness
Baseline ± 20%H1 yACTDI
> ± 1.5% H1 yBCT number uniformity
Water: ± 10 HU
Other materials ± 20 HU
*more scans/parameters
H
H
D/W
1 Y*
A
B
CT number values
Baseline ± 20%
*more scans/parameters
H
H
D/W
1 y*
A
B
Image noise
Remedial levelPriorityFrequencyLevel of expertise
Physical parameter