+ All Categories
Home > Documents > Radiation Protection in Digital Radiology

Radiation Protection in Digital Radiology

Date post: 03-Jan-2016
Category:
Upload: cecilia-lancaster
View: 45 times
Download: 0 times
Share this document with a friend
Description:
Radiation Protection in Digital Radiology. Fundamentals of Digital Radiography L01. Educational Objectives. Explain how ordinary radiographic images can be captured in digital form Discuss the advantages and limitations of digital images - PowerPoint PPT Presentation
44
IAEA International Atomic Energy Agency Radiation Protection in Digital Radiology Fundamentals of Digital Radiography L01
Transcript
Page 1: Radiation Protection in Digital Radiology

IAEAInternational Atomic Energy Agency

Radiation Protection in Digital Radiology

Fundamentals of Digital Radiography L01

Page 2: Radiation Protection in Digital Radiology

IAEA 2

Educational Objectives

• Explain how ordinary radiographic images can be captured in digital form

• Discuss the advantages and limitations of digital images

• Explain how the dissociation of acquisition and display in DR can contribute to unnecessary radiation exposure to patients

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 3: Radiation Protection in Digital Radiology

IAEA 3

What is a digital image?

• An approximation of an analog image, with regard to• spatial information• contrast information

• A computer file composed of discrete picture elements, or pixels• location in file (array or matrix)

represents image position• numeric value represents signal

intensity

Etruscan Roman Mosaic circa 50BC

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 4: Radiation Protection in Digital Radiology

IAEA 4

Why would we want digital images?

• Availability • a digital image can be transmitted electronically

to distant locations and can exist simultaneously at multiple locations

• Flexibility• the appearance of a digital image can be

modified• Convenience

• a digital image can be stored electronically without occupying physical space

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 5: Radiation Protection in Digital Radiology

IAEA 5

Conventional screen-film radiography

• Radiation strikes intensification screen(s) producing fluorescence

• Fluorescent light exposes photographic film producing latent image

• Latent image is chemically developed to produce density in film

• Film density is viewed by trans-illumination

1895

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 6: Radiation Protection in Digital Radiology

IAEA 6

Developed film is effectively analogue

• Density is the result of many developed silver grains

• Grains in intensification screen are quite small

[1,0,0,2,3,4][1,0,0,2,3,4][1,0,0,2,3,4][0,0,0,0,0,0]

1 0 3 42

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 7: Radiation Protection in Digital Radiology

IAEA 7

Three approaches to digital radiography

I. Translate developed film into digital form.II. Capture the radiographic projection by

non-photographic method and digitize during development.

III. Capture the radiographic projection or its fluorescence directly in digital form.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 8: Radiation Protection in Digital Radiology

IAEA 8

Method I: Film Digitization

• Video of transilluminated radiographs• “Camera-on-a-stick”• Low cost, low quality

• LASER film digitisers• Best quality• Expensive and involves periodic maintenance

• CCD film digitisers• Less cost than LASER, less maintenance, better

quality than camera-on-a-stick• Old problems of drift, noise, non-uniform

illumination, and veiling glare – mostly rectified

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 9: Radiation Protection in Digital Radiology

IAEA 9

Process of film digitization

• Light is directed onto a film

• Light passing through the film is measured

• Amount of light attenuated is converted into a digital code value

Bushberg, Seibert, Leidholdt, BooneThe Essential Physics of

Medical Imaging 2nd EdRadiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 10: Radiation Protection in Digital Radiology

IAEA 10

Fundamental limitations of film digitization

• Prone to artefacts• Labour intensive – an extra step

• Best quality achievable is limited by original screen-film image

How many of us maintain capability to digitize film?Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 11: Radiation Protection in Digital Radiology

IAEA 11

Method II: Non-photographic capture with digital development

• Xeroradiography • Charged Selenium plate• Electrostatic latent image• Charge distribution transferred to paper using

toner• Selenium drum detector

• Selenium deposited on cylindrical Al drum• Selenium uniformly charged before exposure• X-rays partly neutralize the charge• Charge distribution measured by electrometer

array

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 12: Radiation Protection in Digital Radiology

IAEA 12

Computed Radiography (CR) or Photostimulable Phosphor (PSP)

Radiography

• Latent image of trapped electrons is formed when x-rays hit the imaging plate

• Latent image is read out physically instead of chemical process

• As the latent image is read out…• Stimulated light emitted with the help of LASER is directed

to a Photomultiplier Tube (PMT)• The PMT signal is digitized using Analog-to-Digital

Converter (ADC)

• The digital image consists of an array of ADC Code Values• ADC Code values represent exposure information• Array locations represent spatial information

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 13: Radiation Protection in Digital Radiology

IAEA 13

Imaging plateImaging plate

Latent Image slow scanslow scan

fast scanfast scan

Light guideLight guide

Photomultiplier tubePhotomultiplier tube

Rotating polygon mirrorRotating polygon mirror

LaserLaser AmplifierAmplifier

Analog-to-Digital ConverterAnalog-to-Digital Converter

Photostimulable phosphor reader

??

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 14: Radiation Protection in Digital Radiology

IAEA 14

Characteristics of PSP systems

• Generally, but not exclusively, cassette-based systems

• Moderate initial capital investment• Simple retrofit to existing radiographic

equipment• Individual scanner can serve multiple

acquisition devices• Workflow comparable to daylight

loader film processing

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 15: Radiation Protection in Digital Radiology

IAEA 15

Method III: X-rays are converted immediately into digital signals without

latent image • Fluorescent screen with video camera (video-

fluoroscopy, image intensifiers)• Fluorescent screen with Charged-Coupled Devices (CCD)

or Complementary Metal Oxide Semiconductor (CMOS) array• Optical lens coupling

• Secondary quantum sink• Fiber optic coupling

• Small area

• Hydrogenated Amorphous Silicon (a-Si:H) with Thin Film Transistors (TFT)• alternative = switching diode• requires x-ray converter (Gadolinium oxysulphide or Caesium

Iodide)• Amorphous Selenium (a-Se) electronically coupled to TFT

“Fla

t pan

els

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 16: Radiation Protection in Digital Radiology

IAEA 16

Characteristics of “Direct” capture systems

+ Rapid acquisition and processing+ Typically integrated with x-ray

generator+ No mechanical scan mechanism—High initial capital investment—Challenging manufacturing

processes—Limited systems for bedside

radiography? Brief history of clinical operation

? Life cycle issues unknown (durability?)

? Image rendering unknown? Exposure factor issues

Courtesy JA Seibert, UC-Davis

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 17: Radiation Protection in Digital Radiology

IAEA 17

How good an approximation does the digital image make?

• Spatial information depends on …•Dimensions of the pixels (matrix size)•Blur

• Contrast information depends on …•Grayscales (Code values) per pixel (i.e.

quantization)•Characteristic function (Code values vs.

exposure)•Noise

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 18: Radiation Protection in Digital Radiology

IAEA 18

Effects of Pixel size and Bit-depth

1024 x 1024 64 x 64 32 x 32 16 x 16

For an image of M x N matrix, k bytes/pixel, the memory needed to store the image is k x M x N bytes

8 bits/pixel3 bits/pixel2 bits/pixel1 bit/pixel

Larger pixels

More bit-depth

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 19: Radiation Protection in Digital Radiology

IAEA 19

Practical resolution is less than the Nyquist frequency

• Factors besides sampling compromise sharpness; depends on• X-ray focal spot dimensions• Blur in Indirect DR and CR• Optical and mechanical imprecision

in IDR and CR• Afterglow in fast-scan dimension in

CR

• Limit of resolution is where Modulation Transfer Function (MTF) has decreased to 10%

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 20: Radiation Protection in Digital Radiology

IAEA 20

MTF of DR depends on more than just sampling

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 21: Radiation Protection in Digital Radiology

IAEA 21

Noise interferes with our ability to detect contrast

σ = √N

SNR = N/σ = √N

Air KERMA

(µGy)

Photons

/100mX100µm

Noise

(%)

9.0 1333 2.7

0.9 133 8.6

0.09 13 27.4

Bushberg, Seibert, Leidholdt, Boone The Essential Physics of Medical Imaging 2nd Ed

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 22: Radiation Protection in Digital Radiology

IAEA 22

Combination of quantum noise and anatomic noise limits low contrast

detection

DR Image CT Image

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 23: Radiation Protection in Digital Radiology

IAEA 23

Detective Quantum Efficiency (DQE) of DRCombines SNR and Resolution

(200 um)

(143 um)

(200 um)

Ideal detector

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 24: Radiation Protection in Digital Radiology

IAEA 24

0

0.5

1

1.5

2

2.5

3

0.1 1 10 100 1000

Air kerma (μGy)

Den

sity

(O

D)

1

10

100

1000

10000

Inte

nsit

y (r

el)

Film/screenPSL

1023

0 Raw Plate Exposure0.1 µGy 1000 µGy

High kVL=2.2, S=50Over-Exposed

Low kV, L=1.8, S=750Under-Exposed

ED

R S

igna

l

DR has wide dynamic range (latitude)

Histogram re-scaling

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 25: Radiation Protection in Digital Radiology

IAEA 25

Rescaling of DR images is a double-edged sword

• Variations in exposure factor selection are automatically compensated by rescaling for consistent appearance

• Inappropriately high or low exposures are not immediately apparent from the appearance of the image until the range-of-adjustment is exceeded• Underexposure makes noisy images• Overexposure makes crisp, noiseless images,

that are preferred by radiologists

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 26: Radiation Protection in Digital Radiology

IAEA 26

There is a documented tendency to overexpose in CR and DR

• Oversight of exposure factor selection is impossible without an exposure indicator

Barry Burns, UNC

Freedman et al. SPIE 1897 (1993),472-479.Gur D et al. Proc 18th European Congress of Radiology. Vienna Sep 12-17.(1993)154.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 27: Radiation Protection in Digital Radiology

IAEA 27

What would be the appropriate exposure?

• Seibert, et al Acad Radiol (1996) 4: 313-318• QA based on exposure indicator reduces doses

• Willis Ped Radiol (2002) 32: 745-750• 33% dose reduction if exposure indicator target followed

• AAPM Task Group #116 is effort to standardize indicators

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 28: Radiation Protection in Digital Radiology

IAEA 28

Five ways of making a radiographic projection with lower dose to the

patient

• Each method has consequences with respect to the five aspects of image quality

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 29: Radiation Protection in Digital Radiology

IAEA 29

Important information about DR acquisition and processing is in

metadata

• CR vs. DX object• Mandatory vs.

optional vs. private tags

• Automatic vs. manual entry of data

• PACS interpretation of metadata

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 30: Radiation Protection in Digital Radiology

IAEA 30

New artefacts from the discrete nature

of DR

• Interference pattern between fixed grid lines and down-sampling rate for display

• Disappeared on zoom

• Bad choices• Display default

magnification factor

• Line rate of grid

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 31: Radiation Protection in Digital Radiology

IAEA 31

Photon Energy (keV)

0 20 40 60 80 100 120 140

0.01

0.1

1.0

Qu

a ntu

m E

f fic

ien

c y

BaFBr 100 mg/cm² (CR)

Gd2O2S:Tb 120 mg/cm2 (Lanex)

CsI:Tl 45 mg/cm2 (a-Si/CsI)

Martin Yaffe/Tony Seibert

A-Selenium 25 mg/cm2

X-ray energy sensitivity differs among detectors

Would you expect to use the same kVp?Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 32: Radiation Protection in Digital Radiology

IAEA

Emerging digital technologiesThe EOS Imaging system

32

Two simultaneous digital planar radiographs (PA and LAT) in the standing position by linear scanning of a fan-shaped collimated X ray beam from 5 cm to 180 cm (whole body).

Charpak Detector: Detector amplification by photon gaz cascade

High gain signal, sensitivity maximized

X ray tubes

Detectors

EOS allows for a dose reduction up to 10 times compared to CR

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 33: Radiation Protection in Digital Radiology

IAEA 33

Advantages of digital radiology

• Digital imaging has practical technical advantages compared with film techniques:•wide contrast dynamic range, •post-processing functionality, •multiple image viewing options, •electronic transfer, •electronic archiving possibilities.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 34: Radiation Protection in Digital Radiology

IAEA 34

How to move from film-screen to digital?

• Training should be planned with anticipation.• Selection of equipment, connectivity and

quality control requires good advice (not only from the manufacturers) and visits to other installations.

• Patient dose and image quality should be audited carefully during the transition. The risk of increasing patient doses exists.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 35: Radiation Protection in Digital Radiology

IAEA 35

How to move from film-screen to digital?

• Typically the initial setting for the automatic exposure control using CR, should be the same or similar to the existing with film-screen. Optimisation (including possible kV changes) and potential dose reduction should be initiated later once radiologists become familiar with digital.

• Audit DICOM header. It contains a lot of useful information.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 36: Radiation Protection in Digital Radiology

IAEA 36

Checklist of practical advice (ICRP 93)

• When introducing a new digital system into clinical practice, the system should be set up to achieve the best balance between image quality and patient dose.

• Avoid deleting non diagnostic images at the workstation and carry out a statistical rejection rate analysis periodically.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 37: Radiation Protection in Digital Radiology

IAEA 37

Checklist of practical advice (ICRP 93)

• Be familiar with your workstation capabilities (post-processing capabilities, options in the monitor to visualise the images, etc).

• Identify correctly all the images to avoid their loss in the PACS.

• Ask for a calibration of the automatic exposure control that is appropriate for the sensitivity range of the digital system and the selected post processing.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 38: Radiation Protection in Digital Radiology

IAEA 38

Checklist of practical advice (ICRP 93)

• Avoid manual exposures if the automatic exposure control is usable. However, ensure it is used correctly, as only some possible errors are correctable by post processing.

• Control the number of images per examination to maintain it at a number similar (or lower) than that for conventional radiology.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 39: Radiation Protection in Digital Radiology

IAEA 39

Checklist of practical advice (ICRP 93)

• Make available a workstation for radiographers to post-process images in order to avoid retakes.

• Pay attention to the dose indication on the panel of the x-ray system or in the in-room monitors and utilise the information to manage patient doses.

• Establish easy access to the PACS to review previous images, in order to avoid retakes.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 40: Radiation Protection in Digital Radiology

IAEA 40

Checklist of practical advice (ICRP 93)

• Pay attention to proper collimation for the desired anatomical area. Once the image is acquired, numerical methods (i.e. software) may automatically crop part of the image, and when the image is received for reading the radiologist will not be aware that a larger anatomical area than necessary was irradiated.

• Select the correct pre-programmed technique e.g. using an abdominal technique (70-80 kV) for chest imaging (120-130 kV) will result in a higher entrance surface dose.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 41: Radiation Protection in Digital Radiology

IAEA 41

Summary• A digital radiographic image is a matrix of numbers with discrete

physical pixel dimensions and gray-levels.• The discrete nature of the DR image is the source of its

advantages and limitations• Digital radiographs can be produced by three methods:

• Digitization of screen-film radiographs• Non-photographic capture and digitization• Direct capture with or without conversion to light

• DR technologies have advantages of availability, flexibility, and convenience over conventional screen-film

• The utility of the DR image is enhanced by demographic, exam, and processing information in the DICOM header

• Except for digitized radiography, DR has the potential for unnecessary patient radiation exposure

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 42: Radiation Protection in Digital Radiology

IAEA 42

Answer True or False

• A digital image contains numeric value representing signal intensity.

• In computed radiography (CR), the electronic latent image is developed by chemical process

• Amorphous selenium is used in direct capture flat panel detectors

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 43: Radiation Protection in Digital Radiology

IAEA 43

Answer True or False

• True. A digital image is a computer file composed of discrete elements or pixels representing signal intensity.

• False. The electronic latent image in the CR plate is read out using LASER

• True. Amorphous selenium (aSe) is electronically coupled to the thin film transistor (TFT) in a direct capture flat panel detector.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography

Page 44: Radiation Protection in Digital Radiology

IAEA 44

References

• Managing patient dose in Digital Radiology ICRP Publication 93 Ann ICRP 2004 Elsevier.

Radiation Protection in Digital Radiology L01 Fundamentals of Digital Radiography


Recommended