Date post: | 13-Apr-2018 |
Category: |
Documents |
Upload: | harish123456 |
View: | 218 times |
Download: | 0 times |
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 1/43
INSTRUCTION MANUAL
RADICAL
PRECISION CALIBRATOR
RADIX ELECTROYSTEMS PVT. LTDMUMBAI-400053
MAN#366R0/P1/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 2/43MAN#366R0/P2/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 3/43
TABLE OF CONTENTS
Title Page
Warranty Certificate.…………………………………………………………………. 4
1. Introduction……………………………………………………………………………. 5
2. Getting Acquainted with Radical……………………….....………………………… 62.1 Standard Equipment................................................................................... 62.2 Input and Output Terminal.......................................................................... 72.3 Keys……………………………………………………………………………… 82.4 Display…………………………………………………………………………… 102.5 Charging the Battery................................................................................... 102.6 Battery life................................................................................................... 10
3. Setting up Radical…………..………………………………………..………………. 11
4. MEASURE Mode………………..………………………….………………………... 124.1 Measure mV……………………….………........................……….......…….. 124.2 Measure mA………………………………………………………...………….. 134.3 Providing Loop power…………………………………………………...…….. 144.4 Measure Resistance…………………………………………………..…….…. 154.5 Testing Continuity………………………………………………..….….………. 164.6 Measuring Temperature………………………………………….…….….....… 17
4.6.1 Using Thermocouples…………………………………………….....….. 174.6.2 Using Resistance Temperature Detectors (RTDs)…………….......… 18
4.7 Measurements in percent of scale…………………………......….…….….… 194.8 Observing Min & Max measurement………………………….………….…… 20
5. SOURCE Mode………………………………………………......……………….….… 215.1 Sourcing Electrical Parameters…………………………..……………….…… 215.2 Source mV……………………………………………………….……….……… 225.3 Source V……………………………………………………………….……....… 235.4 Source mA……………………………………………………………….….…… 245.5 Simulating a 4 to 20 mA Transmitter…………………………………......…… 255.6 Source Resistance ..………………………………………………….....……… 26
5.6.1 Source Resistance using Ohms Module........................................... 275.7 Simulating Thermocouples……………………………………………......…… 295.8 Simulating RTDs……………………………………………….……….…..…… 30
5.8.1 Simulating RTDs using Ohms Module.............................................. 315.9 Stepping and Ramping the output value…………………………...…...……. 33
5.9.1 Stepping the output………………………………………………….…… 335.9.2 Ramping the output………………………………...……………….….… 33
6. Simultaneous MEASURE / SOURCE……………………….………......…………… 35
7. Maintenance.……………………………………………………..................….……… 367.1 Replacing the Battery Module ........................................................................367.2 Cleaning the calibrator ................................................................................ 377.3 Service Center Calibration or Repair............................................................. 377.4 Replacement Parts...................................................................................... 37
8. Specifications……………………………………………………..................….……… 38
MAN# /P3/42/300910/A366R0
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 4/43MAN#366R0/P4/42/300910/A
Warranty Certificate
Product : Precision Calibrator
Model : Radical
Serial No :
Date of Purchase :
Warranty valid upto :
Radix Electrosystems Pvt. Ltd. warrants this product to be free of defects in workmanship andmaterials for one year from the date of purchase.
1. Defects in workmanship and materials will be serviced free of charge during the warrantyperiod. Labour required for rectifying the defects is also covered by this warranty. Anyfurther claims are excluded.
2. Warranty servicing does not extend the period of warranty for the entire instrument or forthe parts that have been replaced.
3. From the warranty are excluded :a) all damage caused by improper handling, e.g. non-observance of the operatinginstructions, subjecting the product to excessive heat, moisture, corrosion, etc.b) damage caused by the intervention of the buyer or by unauthorised third partiesc) transport damage from the manufacturer to the dealer or user or on the way to servicingd) parts subject to normal wear, e.g. relays.
4. Any liability for direct or indirect consequential damages caused by the instrument isexcluded, even if the defect of the instrument is covered by the warranty.
Radix liability for any breach of this agreement shall not exceed the purchase price paid.
5. Any further claims, based on any title, are excluded.
6. In case repair service is required under this factory warranty, please send the instrument,indicating the exact address of the sender and defect, carefully packed, postage paid, to :Radix Microsystems, B-1, 2nd Floor, Ghanshyam Indl. Estate, Off Veera Desai Road, Andheri (West), Mumbai - 400 053.
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 5/43MAN#366R0/P5/42/300910/A
1. INTRODUCTION
RADICAL is a precision thermometer/calibrator for measurement of process signals and calibration of process instruments.
RADICAL is a microcontroller based design. It incorporates a highly stable reference, a high-resolution Analog - to - Digitalconverter, electronic switching circuitry & other specialized electronics.
Further, use of the micro-controller allows direct display in ºC/ºF for thermocouple and RTD inputs, as well as automatic andmanual cold-junction compensation for thermocouples.
Features
Measure mV, mA, RTDs, thermocouples, ohms
Source mV, volts, mA, thermocouples, RTDs, ohms
Continuity test (in MEASURE mode)
Simulate 4 to 20 transmitter (in SOURCE mode)
Loop 24V (in measure mA)
Simultaneous Measure and Source
Step, ramp, % scale, min-max valuesNon tactile membrane keypad
24 keys with 3 Softkeys
128 x 64 graphic LCD display with backlight
Auto shutoff and LO BAT
8~14 hours of use with Lithium-ion rechargeable battery
This manual provides complete operating instructions as well as terminal connection details, for different modes of operation.
TABLE 1 : SUMMARY OF SOURCE AND MEASURE FUNCTIONS
FUNCTION MEASURE SOURCE
DC Voltage -250 to +250mv 0 to 10V (5mA max) or -250 to 250mV
Resistance 0to 1000 0to 1000
DC current 0 to 25 mA 0 to 25 mA (with 600load)
Continuity Displays “SHORT” if measured
resistance is < 10 andDisplays “OPEN” if measured NA
resistance is >= 10
Thermocouple B, C, D, E, G, J, K, N, R, S, T B, C, D, E, G, J, K, N, R, S, T
100Platinum (385), 100Platinum (385),
RTD 200Platinum (385), Cu 53,Ni 120 200Platinum (385), Cu 53,Ni 1202,3 or 4 wire 2 wire
Loop Power 24V Nominal, 30 mA max
CAUTION :Do not block the Radical enclosure ventilation (e.g. by putting a file on Radical) as it can lead tohigh temperature inside the enclosure and a possible fire hazard.
Do not use any battery charger other than Model BC1 as it can lead to damage and a possible fire hazard.
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 6/43MAN#366R0/P6/42/300910/A
2. GETTING ACQUAINTED WITH RADICAL
2.1 Standard EquipmentThe item shown in table no. 2 below are included with your calibrator . If the calibrator is damaged or something is missing,contact the place of purchase immediately. To order replacement parts or spares, see the user-replaceable parts listed at the endof this manual
TABLE 2 : STANDARD EQUIPMENT
ITEM QUANTITY PHOTOGRAPH
Test leads (AEA-053) one set
Test clips ( AEA-051) one set
Test probes ( AEA-050) one set
Jumper for three-wire Resistance measurement connection 1(AEA-049)
TC mini jack for thermocouple mV measurement or Sourcing 1(JGX-001)
Rechargeable Lithium-ion battery pack (inside instrument) - 1Model LB918
Battery Charger - Model BC 1 1
Ohms Module - OM1000 1
Instruction Manual & Test & Calibration Report 1
Carrying Case(PKG-077) 1
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 7/43MAN#366R0/P7/42/300910/A
2.2 Input and Output Terminal
Figure 1 shows the calibrator input and output terminal. Table no. 3 explains their use.
Terminals Use Description
1 , 2 SOURCE mV, V Terminals for sourcing mV , V
1 , 3 SOURCE, RTDs Terminals for sourcing Resistance , RTD
4 , 6 MEAS mA, SOURCE mA Terminals for measuring mA or sourcing mA
5 , 6 MEAS(2 wire) Terminals for measuring Resistance , RTD (2 -wire)
5 , 6 , 7 , 8 MEAS(3 wire , 4 wire) Terminals for measuring Resistance , RTDs (3- wire , 4- wire )
7 , 8 MEAS mV Terminals for measuring mV
9 , 10 24V loop suply Terminals for 24V - loop supply
11 TC input / output jack Jack for measuring or simulating Thermocouple
3
6
9
+/
1 2
4 5
7 8
0
+OUT
mV/V
OUT
+
CE
MEAS
SOURCEmA
TC
RTDNULL
+ 24 V
IN
4-w
TX+
RADI CAL
mVIN
+
+
TC IN/OUT
mV
V
ENTER
IN mA OUT
+IN2-W
3-W
Figure 1. Terminal
1
2
4
6
8
7
5
3
10
9
11
TABLE 3 : INPUT AND OUTPUT TERMINAL
CAUTION : Read complete manual before use.
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 8/43MAN#366R0/P8/42/300910/A
2.3 Keys
Figure 2 shows the calibrator keys and table 4 explains their functions .The soft keys are the three unmarked keys just below thedisplay. Labels that appear above the soft key during operation define Soft key functions.
3
6
9
+/
1 2
4 5
7 8
0
+OUT
mV/V
OUT
+
CE
MEAS
SOURCEmA
TC
RTDNULL
+ 24 V
IN
4-w
TX+
RADI CAL
mVIN
+
+
TC IN/OUT
mV
V
ENTER
IN mA OUT
+IN2-W
3-W
Figure 2. Keys
1
2
3
4 5 6 7 8 9 10 11 12
13
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 9/43
TABLE 4 : KEY FUNCTIONS
ITEM FEATURE DESCRIPTION
1 Turns the power on and off.
2 Turns the backlight on and off.
Perform the function defined by the label above each key3
on the display.
Cycles the calibrator through MEASURE, SOURCE and4
MEASURE/SOURCE modes.
In MEASURE mode toggles between mA (current) and loop24V function.
5 In SOURCE mode toggles between mA and Simulatetransmitter function.
Selects the dc mV function in MEASURE mode or toggle6
between mV and Voltage function in SOURCE mode.
Toggles between resistance and continuity function in7 MEASURE mode or Selects the resistance function in
SOURCE mode.
Selects TC (thermocouple) or RTD (resistance temperature8
detector).
Used to nullify measured resistance in simultaneous9
MEASURE and SOURCE mode and in MEASURE mode.
Terminates a numeric entry when setting a source value.10
These keys are use to make selection when prompted.These keys also increase or decrease the output in Source11mode when using the step feature.
12 To clear entry in SOURCE mode.
13 Numerical keypad Use whenever a numeric entry is required.
CE
ENTER
TC
RTD
mV
V
mA
MEAS
SOURCE
NULL
MAN#366R0/P9/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 10/43MAN#366R0/P10/42/300910/A
2.4 Display
Figure 3. Elements of a typical display
Figure 3 shows the features of a typical display of MEASURE mA mode. The other parts of display are as follows:
Mode Indicator: shows whether the calibrator is in MEASURE or SOURCE mode. In split screen MEASURE /SOURCE mode, there is a Mode Indicator for each window.
Measured Value: shows the measured value in your choice of engineering units or percent of scale.
2.5 Charging the Battery
The Lithium-ion Battery Module Model LB918 is charged using the Battery charger Model BC1 supplied.Normally a charging period of 4 hours is enough for a full charge.
2.6 Battery Life
Radical performance is guaranteed to meet specifications until the low battery annunciator located at the top right side of
the display appears. When the annunciator appears, stop using the radical and charge the battery. If you have a fully
charged spare battery, exchange it with the low battery.
The battery is useful for up to 250 charge/discharge cycles if used optimally.
If you need to replace the battery, refer to “7.1 Replacing the Battery Module”.
For long battery life & best performance, wait for to appear before you recharge the battery.
- +
- +
LOOP 24V
10.390 mA
MEASURE mA
%Scale Min Max
Mode Indicator
Loop Power Annunciator
Measured Value
Soft Key Labels
- +
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 11/43MAN#366R0/P11/42/300910/A
3. SETTING UP RADICAL
In setup mode user can set Auto off time, Beeper condition, Backlight condition at power on & Power on mode.
Auto off An optional Auto off feature turns the Radical off after a selected time if no key operation is done in this period.
Beeper This feature allow user to set beeper sound ON or OFF for key press.
Backlight at power onThis feature allows user to set power on condition for backlight (ON or OFF).
( key is used to make backlight ON or OFF after power on.)
Power on modeRadical allows user to set power on mode. e.g. if user selects Meas mA mode as power on mode then at every power onit will switch on in MEASURE mA mode.
Auto off 95 min
Beeper OFFBacklight OFFP.on mode Source mV
SETUP
Back Change Done
Auto off 95 min Beeper ON
Backlight OFFP.on mode Source mV
SETUP
Back Change Done
Auto off 95 minBeeper OFF Backlight ON
P.on mode Source mV
SETUP
Back Change Done
Auto off 95 minBeeper OFFBacklight ON P.on mode Source mA
SETUP
Back Change Done
0 mA
SOURCE mA
Step Ramp
To go into SETUP modePress any key, beforecount goes to zero(5 sec)
1Press any key to go
into setup … 5
Auto off 95 min Beeper OFF
Backlight OFFP.on mode Source mV
SETUP
Back Change Done
Setup mode is displayed
2 To select function to change Arrow position change.
3To turn ON / OFF auto offfeature.
Change Soft key
From ON to OFFOFF is displayed in front of Auto off.From OFF to ONPreviously set Auto off time outperiod is displayed in front of Auto off.New entry of Auto off time can bedone using Numerical key pad.(1 to 120 minutes)Change Soft key toggles betweenOFF & ON (Auto off time in minutes)
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
4 To turn ON/OFF Beeper Change Soft key
From ON to OFFOFF is displayed in front of Beeper.From OFF to ONON is displayed in front of Beeper.
5 To turn ON/OFF Backlight. Change Soft key
From ON to OFF
OFF is displayed in front ofBacklight.From OFF to ON
ON is displayed in front of Backlight.
6 To select power on mode. Change Soft key
Next mode is displayed.Keep pressing Change Soft key tillrequired mode is displayed.
With every key press of ChangeSoft key next value is displayed.
(Meas mV, Meas mA, Meas,Meas TC, Meas RTD, Source mV,
Source V, Source mA, Source,Loop 24V, Simu Tx, Meas+Srce)
7Exit from setup mode with outsaving new settings Back soft key
Switch to power on mode set insetup mode.
8Exit from setup mode & savenew settings made in Setup
mode.
Done soft keySwitch to power on mode set insetup mode.
0 mV ---
SOURCE mV
Step Ramp
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 12/43MAN#366R0/P12/42/300910/A
4. MEASURE MODE
Note1 : At power on, the 'Power ON Mode' selected (Sec SETUP mode) will appear (e.g. MEASURE mA)
Note2 : If MEASURE mode is selected by pressing , the instrument will go to the MEASURE mode for the selectedparameter(e.g. MEASURE mV)
4.1 Measure mV
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upperleft corner.Note: Press this key
till MEASURE isdisplayed at upper leftcorner of display.
0.005 mA
MEASURE mA
% Scale Min Max
2Enter into measure mVmode
Measure mV Screen is displayed.
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+ +_
Dc mV Source
Black
Red
Figure 4. Connections for measuring DC mV
5.021 mV ---
MEASURE mV
% Scale Min Max
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 13/43MAN#366R0/P13/42/300910/A
4.2 Measure mA
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upper leftcorner.
2 Enter into measure mA mode Measure mA Screen is displayed.
7.055 mA
MEASURE mA
% Scale Min Max
+OUT
mV /V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX
+
3-W
mVIN
+
Black
Red
Figure 5. Connections for measuring mA current
Note: Press this keytill MEASURE isdisplayed at upper leftcorner of display.
0.005 mV
MEASURE mV
% Scale Min Max
---
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 14/43MAN#366R0/P14/42/300910/A
4.3 Providing loop power
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upper leftcorner.
2 Enter into Loop 24 V modeLoop 24 V is displayed onmA Screen.
+OUT
mV /V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX
+
3-W
mVIN
+
4 to 20transmitter +
Red Black
Figure 6. Connections for supplying loop power
LOOP 24V
7.055 mA
MEASURE mA
% Scale Min Max
0.005 mV ---
MEASURE mV
% Scale Min Max
Note: Press this keytill MEASURE isdisplayed at upper leftcorner of display.
(Note : Press twice)
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 15/43MAN#366R0/P15/42/300910/A
4.4 Measure Resistance
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upper leftcorner.
2
3
4
5
Enter into measureresistance mode
To change wire configuration
Exit from change wire screen
To adjust zero ohm (null).
Screen to change. Wire isdisplayed.
Next wire is configuration isdisplayed.
MeasureScreen is displayed.
Zero setting is done.
Note: Press this keytill MEASURE isdisplayed at upper leftcorner of display.
3-wire
100.00
MEASURE
Wire Done
Wire soft key
4-wire
0.02
MEASURE
Wire Done
Done soft key
Null soft key
Figure 7. Connections for measuring resistance
2 wire 4 wire3 wire
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
+OUT
mV /V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX
+
3-W
mVIN
+
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
0.005 mV ---
MEASURE mV
% Scale Min Max
0.02
MEASURE
Null % Scale Min Max
4-wire
0.00
MEASURE
Null % Scale Min Max
4-wire
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 16/43MAN#366R0/P16/42/300910/A
4.5 Testing Continuity
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upper leftcorner.
2Enter in continuity testing.(Note: Press twice)
Measure Continuity screen isdisplayed.Word “SHORT” is displayed &beeper sounds if measure
resistance is < 10&Word “OPEN” is displayed if
measure resistance is >= 10.
SHORT0.120
MEASURE
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
Figure 8. Connections for testing continuity
Note: Press this keytill MEASURE isdisplayed at upper leftcorner of display.
0.005 mV
MEASURE mV
% Scale Min Max
---
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 17/43MAN#366R0/P17/42/300910/A
4.6 Measuring Temperature
4.6.1 Using Thermocouples The calibrator supports following standard thermocouples, each identified with an alpha character: B, C, D, E, G, J, K, N, R, S or T.
To measure temperature-using thermocouple, proceed as follows:
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upper leftcorner.
2
3
4
5
Enter into measurethermocouple mode
To change CJC mode(Auto mode or Manual mode)
To change unit.(°C or °F)
To change TC type
TC screen is displayed.
From Auto to Manual modeMan is displayed at the place of Auto.Previously entered CJC value isdisplayed in front of CJC.First digit of CJC will flash.(Use numerical keypad to enter CJCvalue.) After entering CJC value pressENTER key to exit from CJC entryscreen.From Manual to Auto mode Auto is displayed at the place of Man.CJC value read by sensor isdisplayed in front of CJC.
Measured value & CJC value isconverted into selected unit. (°C or °F)
Next TC type is displayed.Keep pressing Type soft key tillrequired Type is displayed.
Cj Mode Soft key
Unit Soft key
Type Soft key
TC K
100.0 °CMan
MEASURE TC
Cj Mode Unit Type
3.014 mV
TC E
100.0 °C Auto
MEASURE TC
Cj Mode Unit Type
4.709 mV
TC K
212.0 °F Auto
MEASURE TC
Cj Mode Unit Type
3.014 mV
TC K
3.014 mV
100.0 °C
MEASURE TC
Cj Mode Unit Type
1 2
4 5
7 8
0
MEAS SOURCE mA TC
RTD NULL mV V
ENTER
3
6
9
+/
+ OUT
mV/V OUT
+
CE
+ 24 V
IN 4 -w
TX +
mV IN +
TC IN/OUT
IN mA OUT + IN 2-W 3-W
Colour depends ontype of TC
Colour depends ontype of TC
Figure 9. Connections for measuring temperature using thermocouple
Note: Press this keytill MEASURE isdisplayed at upper leftcorner of display.
0.005 mV
MEASURE mV
% Scale Min Max
---
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 18/43MAN#366R0/P18/42/300910/A
4. 6.2 Using Resistance-Temperature Detectors (RTDs)
The calibrator supports following standard RTD Pt 100,Pt 200,Cu 53 ,Ni 120
The calibrator accepts RTD types. RTD are characterized by their resistance at 0°C which is called the “ice point”. The most
common R is 100and mostly come in a three-terminal configuration.
The calibrator accepts RTD measurement inputs in two, three or four wire connection. A four-wire configuration provides thehighest measurement precision and two wire provide the lowest measurement precision.
To measure temperature using an RTD input, proceed as follows:
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upper leftcorner.
2
3
4
5
To enter in to Measure RTDmode
To change wire configuration(2 wire, 3 wire, 4 wire)
To change Unit(°C or °F)
To change RTD type
RTD screen is displayed.
Next wire configuration isdisplayed.Keep pressing Wire soft key tillrequired Wire configuration isdisplayed.
Measured value is converted intoselected unit.(°C or °F)
Next RTD type is displayed.Keep pressing Type soft key tillrequired Type is displayed.
Wire Soft key
Unit Soft key
Type Soft key
(Note: Press twice)
3 Wire Pt 100
100.0 °C
138.50
MEASURE RTD
Wire Unit Type
4 Wire Pt 100
100.0 °C
138.50
MEASURE RTD
Wire Unit Type
3 Wire Pt 100
212.0 °F
138.50
MEASURE RTD
Wire Unit Type
3 Wire Pt 200
100.0°C
277.00
MEASURE RTD
Wire Unit Type
+OUT
mV /V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
3 wire2 wire 4 wire
Figure 10. Connections for measuring temperature using RTD
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
Note: Press this keytill MEASURE isdisplayed at upper leftcorner of display.
0.005 mV
MEASURE mV
% Scale Min Max
---
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 19/43MAN#366R0/P19/42/300910/A
4.7 Measurements in percent of scale
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure modeDisplay MEASURE at upper leftcorner.
2
3
4
5
6
7
8
Enter into measure mV or
mA or.
Go into% Scale mode
Return to main screen
To change low scale & highscale value
To select High scale or Lowscale
Return to main screen,with out saving changes
Go in to %Scale screen& save the changes madein high Scale & Low Scalevalue
Measure Screen is displayed forselected parameter.
(e.g. mA, mV,)
% Scale screen is displayed
Measure mV Screen is displayed.
Screen shows old low Scale &high Scale value.
Arrow position is changed.New entry can be done usingnumerical keypad.
Measure mV Screen is displayed.
% Scale screen is displayed
%Scale soft key
Back Soft key
Change Soft key
Back Soft key
Done Soft Key
Low Scale 0 mVHigh Scale 10 mV
MEASURE mV
Back Done
Low Scale 0 mVHigh Scale 10 mV
MEASURE mV
Back Done
0.005 mA
MEASURE mA
% Scale Min Max
or
50.000 %
0 10
MEASURE mV
Back change
50.000 %
0 10
MEASURE mV
Back change
Note: Press this keytill MEASURE isdisplayed at upper left
corner of display.
5.000 mV
MEASURE mV
% Scale Min Max
---
5.000 mV
MEASURE mV
% Scale Min Max
---
5.000 mV
MEASURE mV
% Scale Min Max
---
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 20/43MAN#366R0/P20/42/300910/A
4.8 Observing Min and Max Measurements
You can set the display to show the maximum and minimum readings.To observe minimum & maximum reading proceed as follows:
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select measure mode Display MEASURE at upper leftcorner.
2
3
4
5
Enter into measure mV ormA or .
Go into Min Max mode
Reset min & max value
Return to main Screen
Measure Screen is displayed forselected parameter.
(e.g. mA, mV,)
Min & Max value is displayedbelow the measured value.
Current value is loaded in min& max register
Measure mV Screen is displayed.
Min Max Soft key
Reset Soft key
Back Soft Key
0.005 mA
MEASURE mA
% Scale Min Max
Note: Press this keytill MEASURE isdisplayed at upper leftcorner of display.
5.000 mV
MEASURE mV
% Scale Min Max
---
12.708 mVmin max
5.013 13.813
MEASURE mV
Back Reset
---
10.584 mV
min max
10.584 10.584
MEASURE mV
Back Reset
---
5.000 mV
MEASURE mV
% Scale Min Max
---
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 21/43MAN#366R0/P21/42/300910/A
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
Select parameter to be
source. mV, mA,, TC,RTD,
Entering desired output value
To terminate numerical entry
Source screen for selectedparameter is displayed
Entry is displayed.
Sourced value is displayed
+ / -
Numerical keypad,Decimal point,
5. SOURCE MODE
Note1: At power on, the 'Power ON Mode' selected (Sec SETUP mode) will appear(e.g. SOURCE mA)
Note2: if SOURCE mode is selected by pressing , the instrument will go to the SOURCE mode for the selectedparameter (e.g. SOURCE mV)
5.1 Sourcing Electrical parameters
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
0 mA
SOURCE mA
% Scale Min Max
0 mV
SOURCE mV
Step Ramp
---
– 8.459 mV
SOURCE mV
Step Ramp
---
– 8.459 mV
SOURCE mV
Step Ramp
---ENTER
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 22/43MAN#366R0/P22/42/300910/A
5.2 Source mV
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
Enter into Source mV mode.
Entering desired output value
To terminate numerical entry
Source mV screen is displayed
Entry is displayed.
Sourced value is displayed
+ / -
Numerical keypad,Decimal point,
0 mA
SOURCE mA
% Scale Min Max
+OUT
mV /V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
Black Red
Figure 11. Connections for sourcing mV
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
0 mV
SOURCE mV
Step Ramp
---
– 8.459 mV
SOURCE mV
Step Ramp
---
– 8.459 mV
SOURCE mV
Step Ramp
---ENTER
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 23/43MAN#366R0/P23/42/300910/A
5.3 Source V
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
Enter into Source V mode.
Entering desired output value
To terminate numerical entry
Source V screen is displayed
Entry is displayed.
Sourced value is displayed
Numerical keypad,Decimal point.
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
Black Red
Figure 12. Connections for sourcing V
0 mA
SOURCE mA
% Scale Min Max
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
(Note: Press twice)
0 V
SOURCE V
Step Ramp
---
1.8746 V
SOURCE V
Step Ramp
---
1.8746 V
SOURCE V
Step Ramp
---ENTER
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 24/43MAN#366R0/P24/42/300910/A
5.4 Source mA
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
Enter into Source mA mode.
Entering desired output value
To terminate numerical entry
Source mA screen is displayed
Entry is displayed.
Sourced value is displayed
Numerical keypad,Decimal point
0 mA
SOURCE mA
Step Ramp
4.000 mA
SOURCE mA
Step Ramp
4.000 mA
SOURCE mA
Step Ramp
+OUT
mV /V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX
+
3-W
mVIN
+
Black
Red
Figure 13. Connections for sourcing mA
0 mV ---
SOURCE mV
% Scale Min Max
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
ENTER
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 25/43MAN#366R0/P25/42/300910/A
5.5 Simulating a 4 to 20 mA Transmitter
You can configure the calibrator as a load on a current loop through the SOURCE mA function. The difference in operation is that,when you Source mA, the calibrator is sourcing current and when you simulate Transmitter, the calibrator is sourcing a variableresistance to regulate current to the specified value.
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
Enter into Simulating 4 to 20transmitter mode.
Entering desired output value
To terminate numerical entry
Screen for Simulate Transmitteris displayed.
Entry is displayed.
Sourced value is displayed
Numerical keypad,Decimal point
Simulate Transmitter
0 mA
SOURCE mA
Step Ramp
Simulate Transmitter
6.548 mA
SOURCE mA
Step Ramp
Simulate Transmitter
6.548 mA
SOURCE mA
Step Ramp
Figure 14 Connection for Simulating a 4 to 20 mA Transmitter
+OUT
mV/V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
Black
Red
+24V
UUT
LoopPower Supply
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
(Note: Press twice)
0 mV
SOURCE mV
% Scale Min Max
---
ENTER
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 26/43MAN#366R0/P26/42/300910/A
5.6 Source resistance
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
Enter into source resistancemode (2 wire)
Entering desired output value
To terminate numerical entry
Source resistance screen isdisplayed
Entry is displayed.
Sourced value is displayed
Numerical keypad,Decimal point
+OUT
mV /V
OUT
IN mA OUT
+
+
IN2-W
+ 24 V
IN
4-w
TX+
3-W
mVIN
+
Black Red
UUT
Figure 15. Connection for sourcing resistance
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
0 mV
SOURCE mV
% Scale Min Max
---
ENTER
0
SOURCE
2-wire
54
SOURCE
2-wire
54.00
SOURCE
2-wire
NOTES1) The unit under calibration must provide a continuous (unswitched) excitation current to Radical.2) The excitation current should be between 0.1 mA and 2.0 mA.3) If unit under calibration does not satisfy above requirements, then use Ohms Module OM1000 (Ref. Sec. 5.6.1) .
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 27/43MAN#366R0/P27/42/300910/A
SR.NO. ACTIONDESCRIPTION INFORMATION
1Select Measure Resistancemode in 4-wire configuration
2
Insert Ohms Module interminal provided for 4-wireresistance measurement(Terminals 5, 6, 7 & 8) Sec 2.2
Sec 4.1
See Figure 17
5.6.1 Source Resistance using Ohms Module OM1000
Switch SW1
COARSE potentiometer
FINE Potentiometer
Radical SideTerminals
Output Terminals
Figure 16 - Elements of Ohms Module OM1000
Cont.....
Figure 17 - Connection f Ohms Module OM1000 with RADICALo
NOTE
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 28/43MAN#366R0/P28/42/300910/A
4 Adjust required resistanceusing Coarse & Finepotentiometers
Radical will display resistanceadjusted using Coarse & Finepotentiometers.
See Figure 16
See Figure 16
SR.NO. INFORMATION ACTIONDESCRIPTION
5
To connect adjustedresistance on outputterminals of Ohms Module,toggle switch SW1 toSIMULATE position
Adjusted resistance is availableat output terminals of OhmsModule for connection to external
device.
Cont..... Source Resistance using Ohms Module OM1000
Figure 18 - Connection for adjusting resistance
3
Set switch SW1 to SETposition to measureresistance on Radical.
See Figure 18Radical will measured and displayresistance set on Ohms Module
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 29/43
5.7 Simulating Thermocouples
The calibrator supports following standard thermocouples, each identified with an alpha character : B, C, D, E, G, J, K, N, R, S, TTo generate temperature using a thermocouple, proceed as follows:
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
5
6
7
Enter into sourcethermocouple mode
To change cjc mode
(Auto mode or Manual mode)
To change unit.(°C or °F)
To change TC type
Entering desired output value
To terminate numerical entry
Source TC Screen is displayed.
From Auto to Manual mode.Man is displayed at the place of Auto.Previously entered CJC value isdisplayed in front of CJC.First digit of CJC will flash.(Use numerical keypad to enterCJC value.) After entering CJC value pressENTER key to exit from CJC entry.From Manual to Auto mode Auto is displayed at the place of Man.CJC value read by sensor isdisplayed in front of CJC.
Measured value & CJC value isconverted into selected unit.(°C or °F)
For every key press next TC typeis displayed.
Keep pressing Type soft key tillrequired Type is displayed.
Entry is displayed & previouslysourced mV is displayed
Sourced value is displayed withnew source mV
Cj Mode Soft key
Unit Soft key
Type Soft key
Numerical keypad,Decimal point,
+ / -
TC K
Auto
SOURCE TC
Cj Mode Unit Type
3.014 mV
100.0 °C
1 2 4 5
7 8
0
MEAS SOURCE mA TC
RTD NULL mV V
ENTER
3 6
9
+/
+ OUT mV/V OUT
+
CE
+ 24 V
IN 4 -w
TX +
mV IN +
TC IN/OUT
IN mA OUT + IN 2-W 3-W
Colour depends ontype of TC
Colour depends ontype of TC
Figure 19. Connections for simulating thermocouple
Note: Press this key
till SOURCE isdisplayed at upper leftcorner of display.
0 mV
SOURCE mV
% Scale Min Max
---
ENTER
TC K
0°C Auto
SOURCE TC
Cj Mode Unit Type
0.000 mV
TC K
212.0 °F Auto
SOURCE TC
Cj Mode Unit Type
3.014 mV
TC E
0 °C
Auto
SOURCE TC
Cj Mode Unit Type
3.014 mV
TC E
524.0 °C Auto
SOURCE TC
Cj Mode Unit Type
3.014 mV
TC E
524.0 °C Auto
SOURCE TC
Cj Mode Unit Type
37.324 mV
MAN#366R0/P29/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 30/43MAN#366R0/P30/42/300910/A
5.8 Simulating RTDs
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
5
6
To enter in to Source RTDmode
To change Unit(°C or °F)
To change RTD type
Entering desired output value
To terminate numerical entry
Source RTD screen is displayedwith previously set RTD type.
Measured value is converted intoselected unit (°C or °F)
Next RTD type is displayed.Keep pressing Type soft key tillrequired Type is displayed.
Entry is displayed.
Sourced value is displayed
Unit Soft key
Type Soft key
Numerical keypad,Decimal point,
+ / -
Pt100
0°C
100.00
SOURCE RTD
Unit Type
Pt100
0°F
93.34
SOURCE RTD
Unit Type
Pt200
0°C
200.00
SOURCE RTD
Unit Type
Pt100
100 °C
138.50
SOURCE RTD
Unit Type
Pt100
100.00 °C
138.50
SOURCE RTD
Unit Type
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
(Note: Press twice)
0 mV
SOURCE mV
% Scale Min Max
---
ENTER
+OUT
mV / V
OUT
IN mA OUT
+
+
IN
2-W
+ 24 V
IN
4-w
TX
+
3-W
mV
IN+
Black
Red
UUT
Figure 20. Connections for Simulating RTDs
NOTES1) The unit under calibration must provide a continuous (unswitched) excitation current to Radical.2) The excitation current should be between 0.1 mA and 2.0 mA.
3) If unit under calibration does not satisfy above requirement, then use Ohm’s Module OM1000 (Ref. Sec. 5.8.1
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 31/43
MAN#366R0/P31/42/300910/A
SR.NO. ACTIONDESCRIPTION INFORMATION
1
Select Measure RTD mode in4-wire configuration. Alsoselect unit & RTD type
2
Insert Ohms Module interminal provided for 4-wireresistance measurement(Terminals 5, 6, 7 & 8)(Ref. Sec. 2.2)
Sec 4.4
See Figure 22
5.8.1 Simulating RTDs using Ohms Module
Switch SW1
COARSE potentiometer
FINE Potentiometer
Radical SideTerminals
Output Terminals
Figure 21 - Elements of Ohms Module OM1000
Cont.....
Figure 22 - Connection f Ohms Module OM1000 with RADICALo
NOTE)
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 32/43MAN#366R0/P32/42/300910/A
4 Adjust required temperatureusing Coarse & Finepotentiometers
Radical will display resistance &corresponding temperatureadjusted using Coarse & Finepotentiometers.
See Figure 21
See Figure 21
SR.NO. INFORMATION ACTIONDESCRIPTION
5
To connect adjustedresistance on outputterminals of Ohms Module,toggle switch SW1 to
SIMULATE position
Adjusted resistance is availableat output terminals of OhmsModule for connection to external
device.
Cont..... Simulating RTDs using Ohms Module
Figure 23 - Connection for adjusting resistance
3
Set switch SW1 to SETposition to measureresistance on Radical.
See Figure 23Radical will measured and displayresistance set on Ohms Module& corresponding temperature
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 33/43
5.9 Stepping and Ramping the output value
5.9.1 Stepping
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2
3
4
5
Enter into source mV or Vor mA
Enter into Step mode
To change step size
Save changes & return instep mode
Source Screen is displayed forselected parameter.(e.g. mA, mV ,V)
Previously set Step size is displayedabove source value.
Enter step size is displayed aboveSource value.To Enter new Step size usenumerical key pad.
New step size is saved &displayed above sourced value.
Step Soft key
Change Soft key
Done Soft key
6
7
To source value with stepsize
Return to main screen Back Soft key Main screen is displayed.
With key value
incremented by step size.
With key valuedecremented by step size.
or
Note: Press this keytill SOURCE is
displayed at upper leftcorner of display.
0 mV
SOURCE mV
% Scale Min Max
---
0 mV
SOURCE mV
Step Ramp
---
step size 25 mV
15.000 mV
SOURCE mV
Back Change
---
Enter step size 25 mV
15.000 mV
SOURCE mV
Back Done
---
step size 10 mV
15.000 mV
SOURCE mV
Back Change
---
step size 10 mV
25.000 mV
SOURCE mV
Back Change
---
15.000 mV
SOURCE mV
Step Ramp
---
5.9.2 Ramping
Ramping is sweeping the source up or down in value. Ramping can be used to smoothly increase or decrease output parameter.
This feature is available in engineering units mV, V, mA.
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE modeDisplay SOURCE at upper leftcorner.
2Enter into source mV or Vor mA
Source Screen is displayed forselected parameter.(e.g. mA, mV ,V)
Note: Press this keytill SOURCE isdisplayed at upper leftcorner of display.
0 mV
SOURCE mV
Step Ramp
---
0 mV
SOURCE mV
% Scale Min Max
---
MAN#366R0/P33/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 34/43
cont. ...5.9.2 Ramping
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
3 Enter into ramp mode Ramp mode screen is displayed.Ramp Soft key
4
5
6
7
To change start value, Finalvalue & ramp time.
To select start value, Finalvalue & Ramp time tochange.
To save the changes made& go into Ramp mode
Return to main screenwithout saving changesmade.
Screen with previously set startvalue, Final value & Ramp timeis displayed.
Arrow position is changed.New entry can be done usingnumerical keypad.
Ramp mode screen is displayed.
Main screen is displayed.
Done Soft key
Back soft key
Change Soft key.
8
9
Start ramping from startingvalue
To Pause value Pause Soft key Ramping is Paused.
Ramping is start from starting
value.Start Soft key
Start val 5 mVFinal val 100 mVRamp time 20 sec
SOURCE mV
Back Done
Start val 5 mV Final val 100 mV
Ramp time 20 sec
SOURCE mV
Back Done
10
11
To start Ramping from
paused value
Return to main Screen Back soft key Main screen is displayed.
Pause Soft key
Ramping is start from paused
value.
or
15.000 mV
SOURCE mV
Back Start Change
---
0 mV
SOURCE mV
Back Start Change
---
0 mV
SOURCE mV
Step Ramp
---
5.000 mV
SOURCE mV
Back Start Pause
---
12.916 mV
SOURCE mV
Back Start Pause
---
12.916 mV
SOURCE mV
Back Start Pause
---
0 mV
SOURCE mV
Step Ramp
---
MAN#366R0/P34/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 35/43
6. SIMULTANEOUS MEASURE/SOURCE
SR.NO. DISPLAY ACTIONDESCRIPTION KEY TO BE PRESSED
1 Select SOURCE mode
Screen for SimultaneousMEASURE & SOURCE modeis displayed.
2
3
To change measureparameter
To change source parameter
0.010 mV
0 mV
MEASURE mV
SOURCE mV
0 mV
50.000 mV
MEASURE mV
SOURCE mV
180.010 mV
0 mV
MEASURE mV
SOURCE mV
Measured reading is cleared andMeasured unit start flashing.Flashing of unit stop after pressing
or or or
TABLE 5 : SIMULTANEOUS MEASURE/SOURCE FUNCTION
MEASURE FUNCTION SOURCE FUNCTION
DC mV DC V mA T
DC mV
mA
Loop 24V
(2 wire)
Continuity
C RTD
Note: Press this key tillboth MEASURE &SOURCE is displayedsimultaneously.
&
or
or
MAN#366R0/P35/42/300910/A
Sourced reading is cleared and
Source unit start flashing.
Flashing of unit stop after pressing
or or or
or
&
or
or
or
NOTESFor using source TC and Source RTD in simultaneous mode. TC/RTD type, unit CJC mode must be selected in Source TC/RTD mode before selecting simultaneous mode.
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 36/43
MAN#366R0/P36/42/300910/A
7. MAINTENANCE
7.1. Replacing the Battery Module
Replace the Battery Module when it no longer holds charge for an adequate duration by following the procedure given below.See 7.4 for ordering information.
1 Remove fitting nuts by turning nuts anti-clockwise.
2 Gently pullout the Battery Module.
3 Remove connector J1 from the Battery Module.
4 Connect new Battery Module to connector J1.
5 Insert Battery Module back into Radical.
6 Fix both nuts.
Battery Module
Model 9V18 Connector J1
Battery Module Model 9V18
Figure 18. Replacing Battery Module
Fitting Nuts
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 37/43
MAN#366R0/P37/42/300910/A
ITEM Model Number Radix Part no.
1 Radical Probe Set ( Test leads, Test clips, Test probes,Jumper for Three-wire Resistance measurement connection, - AEA-055TC mini jack )
2 Rechargeable Lithium-ion Battery Module 9V18 -(battery compartment + LB918 rechargeable battery )
3 Battery Charger BC1 -
4 Carrying case - PKG - 077
7.2. Cleaning the Calibrator
Clean calibrator with a soft cloth dampened with water or water & mild soap. Do not spill droplets.CAUTION : To avoid damaging the plastic lens & case, do not use solvents or abrasive cleansers.
7.3. Service Center Calibration or Repair
Calibration , repairs , or servicing not covered in this manual should be performed only at our service center. If the calibrator fails ,check the Lithium-ion battery pack first, and replace it if needed. Verify that the calibrator is being operated in accordance with theinstructions in this manual. If the calibrator is faulty, send a description of the failure with the calibrator. Be sure to pack the
calibrator securely, using the original container if it is available. Send the equipment postage paid to Service Center.
7.4. Replacement Parts
The table no.6 lists the user-replaceable parts for Radical. If the item has Model Number, order it by that number. Otherwise, useRadix part no.
TABLE 6 : USER REPLACEABLE PARTS
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 38/43
8. SPECIFICATIONS
GENERAL
Display : 128 x 64 monochrome LCD with backlight
Modes : 1. Millivolts Measure2. Millivolts Source3. Thermocouple Measure in ºC / ºF4. Thermocouple Source in ºC / ºF5. Milliamps Measure
6. Milliamps Source7. Resistance Measure8. Resistance Source9. Volts Source
10. RTD Measure in 2, 3, and 4 wire 11. RTD Source in 2 wire
12. Simultaneous Measure & Source mode13. Simulating 4 to 20 transmitter
ADC conversion rate : 12 readings/second
Maximum burden : 600(mA source mode)
Error messages : See Table 7.6
Other messages : See Table 7.7
Weight : 1.75 kg
Size (in mm) : 241(W) x 153(H) x 197(D)
Operating temperature : 0 to 50ºC
Battery : 9 V, 1800 mAh, Lithium-ion rechargeable battery
Battery life : 8 to 14 hours per charge, depending on mode, backlight use etc
Battery charger : Included
Range : See Tables 7.1 ~7.5
Accuracy : See Tables 7.1 ~ 7.5
Resolution : See Tables 7.1 ~ 7.5
SPECIAL FEATURES & FUNCTIONS
24 V loop supply : To power 2-wire transmitter without using external supply.
Bar-Graph display : Gives a graphic indication of the measured parameters on a scale of 0~100% in stepsof 2.5%.
Min-Max function : Shows 'peak' & 'trough' values over a measurement interval.
Step function in Source mode : Allows a sourced parameter to be increased or decreased in steps of 1 mV, 1 mA or 1.
Ramp function in Source mode : Allows a sourced parameter to be smoothly increased or decreased at a specified rate.
Auto shutoff : Switches the instruments off if no key is pressed for the specified time.
Battery low warnings : Gives visual alerts when the battery is low and needs recharging & also when it is about to shut down.
TABLE 7.1 : MEASURE MODES : RANGE, RESOLUTION, ACCURACY
Mode Function Resolution Range Accuracy
Low High Unit
Measure DC mV 1V -250.000 250.000 mV 0.025% of reading + 0.015% of full scale
Measure DC mA 1 A 0.000 25.000 mA 0.025% of reading + 0.025% of full scale
Measure Ohms 10m 0.000 110.00 0.05% of reading + 0.05Measure Ohms 10m110.00 999.99 0.05% of reading + 0.5
MAN#366R0/P38/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 39/43
TABLE 7.2 : SOURCE MODES : RANGE, RESOLUTION, ACCURACY
Mode Function Resolution Range Accuracy
Low High Unit
Source DC mV 1V -250.000 250.000 mV 0.01% of reading + 0.005% of full scale
Source DC V 100V 0.0000 10.0000 V 0.01% of reading +0.005% of full scale.
Source DC mA 1 A 0.000 25.000 mA 0.025% of reading + 0.025% of full scale
Source Ohm 10m0.00 100.00 0.05% of reading + 0.05Source Ohm 10m100.00 1000.00 0.05% of reading + 0.2
TABLE 7.3 :
Range (°C) Resolution (°C) Error (°C)
Measure Source
TC Type Typical Max. Typical Max.
400 to 800
TC B 800 to 1000 0.1 0.8 2.3 1.2 2.3
1000 to 1820 0.7 1.8 1.0 1.8
40 to 800 0.5 1.1 0.7 1.1
800 to 1200 0.5 1.1 0.5 1.1TC C 0.1
1200 to 1800 0.7 1.3 0.9 1.3
1800 to 2316 1.0 1.8 1.0 1.8TC D 30 to 2310 0.1 1.0 1.2 0.8 1.2
-270 to 200 0.5 1.4 0.5 1.4
-200 to -100 0.3 0.4 0.3 0.4TC E 0.1
-100 to 600 0.2 0.3 0.2 0.3
600 to 1000 0.2 0.3 0.2 0.3
TC G 10 to 2310 0.1 1.0 1.2 0.8 1.2
-210 to 100 0.4 0.6 0.4 0.6TC J 0.1
-100 to 760 0.3 0.4 0.3 0.4
-270 to 100 1.0 1.2 0.8 1.2
-100 to 400 0.3 0.5 0.3 0.5TC K 0.1
400 to 1200 0.3 0.5 0.3 0.5
1200 to 1372 0.5 0.6 0.5 0.6
-270 to -100 0.5 1.0 0.6 0.6
TC N -100 to 900 0.1 0.1 0.5 0.5 0.5
900 to 1300 0.3 0.6 0.3 0.3
0 to 100 0.9 3.1 0.9 3.1TC R 0.1
100 to 1768 0.7 1.6 0.7 1.6
0 to 200 0.9 2.8 0.9 2.8
TC S 200 to 1400 0.1 0.7 1.8 0.7 1.8
1400 to 1768 0.7 1.7 0.7 1.7
-270 to -200 1.0 2.1 0.5 2.1
TC T -200 to 0 0.1 0.5 0.6 0.3 0.6
0 to 400 0.3 0.4 0.2 0.4
MEASURE & SOURCE MODES FOR THERMOCOUPLES : RANGE, RESOLUTION, ACCURACY
1.0 3.4 1.8 3.4
TABLE 7.4 : MEASURE & SOURCE MODES FOR RTDS : RANGE, RESOLUTION, ACCURACY
Pt100 (385) -200 to 0 0.1 0.3 0.3 0.5
0 to 400 0.01 0.1 0.3 0.3 0.5400 to 850 0.3 0.6 0.5 0.5
Pt200 (385) -200 to 0 0.1 0.3 0.3 0.5
0 to 400 0.01 0.1 0.3 0.3 0.5
400 to 850 0.3 0.6 0.3 0.5
Ni120 (Ni) -70 to 300 0.01 0.1 0.5 0.1 0.5
Cu 53 0 to 180 0.01 0.3 0.5 0.3 0.5
* Specifications are for 3-wire RTD in measure mode. Add 0.4°C for 2 & 4-wire RTD.
TABLE 7.5 : CONTINUITY
Continuity testing
TONE RESISTANCE
CONTINUOUS TONE < 10
NO TONE >= 10
MAN#366R0/P39/42/300910/A
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 40/43
MAN#366R0/P40/42/300910/A
DISPLAY CAUSE
To indicate over range in measure mode.
Invalid entry
Connection error This message is displayed in Source Resistance & in Source RTD mode if connection toRadical or UUT is not made.
If entry is out of limit
DISPLAY CAUSE
OPEN Displayed in testing continuity mode if measure resistance is >=10
SHORT
Wait Displayed in measure screen of simultaneous mode while adjusting resistance in SourceResistance or in Source RTD mode.
Displayed in testing continuity mode if measure resistance is <10
TABLE 7.6 : ERROR MESSAGES
TABLE 7.7 : OTHER MESSAGES
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 41/43
MAN#366R0/P41/42/300910/A
THIS PAGE LEFT INTENTIONALLY BLANK
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 42/43
MAN#366R0/P42/42/300910/A
Head Office
Radix B-17, 1st Floor, Ghanshyam Indl EstateOff Veera Desai Road, Andheri (West)Mumbai - 400 053, IndiaTel : + 91 22 26730101
Fax: + 91 22 26731891
Email : [email protected] www.radix.co.in
Radix Electrosystems Pvt Ltd, B-14, 2nd Floor, Ghanshyam Indl. Estate, Veera Desai Road, Andheri (W), Mumbai - 400 053Tel : 26730101 • Fax : + 91 22 66925021 • Email : [email protected] • www.radix.co.in
AN ISO 9001:2008 COMPANY
7/27/2019 Radical MAN366
http://slidepdf.com/reader/full/radical-man366 43/43
RADICALSOFTWARE & MANUAL
IS UPDATED.
PLEASE OBSOLETEEARLIER MANUAL
(MAN#195RX)