RapidArc in Bergen Britt Nygaard, Harald Valen and Ellen Wasbø Haukeland University Hospital, Bergen, Norway
Transcript

RapidArc in Bergen

Britt Nygaard, Harald Valen and Ellen Wasbø

Haukeland University Hospital, Bergen, Norway

2

• 2007: – Trilogy with RapidArc option

• 2008: – Scandidos Delta4 QA tool– Aria upgrade: RapidArc on the Trilogy and 23iX

• Autumn 2009: – Course in Bellinzona and Zug– Stay-and-learn in Copenhagen– Eclipse AAA configuration– Machine QA and patient QA procedures

• 2010:– Decisions, decisions.. Which category of patients?– Learning RapidArc doseplanning in Eclipse– 1st patient on 14th of June – 2nd on 22nd of November

3

Quality control

• Commisioning tests as suggested by Memorial Sloan-Kettering CC and Varian– A picket fence test during RapidArc– 7 adjacent fields with varying Dose rate & Gantry

speed– 4 adjacent fields with varying MLC speed & Gantry

speed

– Possible to study combined effect of • dose rate and gantry speed• dynamic MLC and variable dose rate

C. C. Ling et. al: Commissioning and Quality Assurance of RapidArc Delivery System. Radiotherapy, Int. J. Radiation Oncology Biol. Phys., Vol. 72, No. 2, pp. 575–581, 2008.

4

MLC speed variation during RapidArc

Dose rate and Gantry speed variation during RapidArc

5

Analyse results

• MLC speed variation (”Test3”)

• Dose rate and Gantry speed variation (”Test2”)

Clinac 23EX (2004): T2 & T3

7

Trilogy (2007): T2 & T3

8

Clinac 23iX (2005): T2 & T3

9

TrueBeam (2011): T2 & T3

10

Analyse results• Dynalog files

– Log planned and actual leaf positions and leaf speed vs. time

– Log gantry speed vs. Time– How TrueBeam

• Tool: ”Analyse Dynalog”– In-house developed (EW)– Language: IDL

11

12

Patient QA

• Delta4– Daily dose correction– Run and measure Verification plan– Pass / Fail criteria

• Dose deviation– > 85% within ±3% deviation

• Distance to agreement– > 98% with DTA ≤ 3mm

• Gamma index 3%, 3mm– > 95% with index ≤ 1

13

1 arc, 135° to 225°, TrueBeam 6MV photons

14

15

Clinac 23EX (2004), RapidArc in 2011: Failed T2 & T3 commissioning tests

Patient QA Dose dev. within ±3%

DTA < 3mm γ < 1 (3%, 3mm)

PAB 90,7% 100% 100%

GB 83,7% 100% 100%

TER 95,8% 100% 99,4%

GDG 85,5% 100% 100%

EKGP 85,9% 100% 100%

MS 83,0% 100% 100%

16

More patient QA

• Independent dose calculation

• Point check of dose• Control of monitor units

17

Treatment planning, Autumn 2010:

• 5 years experience with IMRT– head and neck– prostate with and without lymph nodes (LN)– ani (and gyn) with LN – Sarcoma, lymphoma and other

• RA configuration and acceptance tests OK • RA installed on 2 Clinacs • Patient start up

18

Which patient groups?

• Increased efficiency for the department – Prostate with LN, 7 splitted fields

• Patients unable to keep the supine position for 10-15 min– Head and neck

• Less MU and less risk for secondary cancer• A category that is easy to create acceptable and

standardized plans for– Prostate intermediate risk

19

Which patient groups?

• Increased efficiency for the department – Prostate with LN, 7 splitted fields

• Patients unable to keep the supine position for 10-15 min– Head and neck

• Less MU and less risk for secondary cancer• A category that is easy to create acceptable and

standardized plans for– Prostate intermediate risk

20

Prostate intermediate risk, criteria:• Treatment of prostate and seminal vesicles

• Equal plan or better than IMRT (PTV and rectum)

• We made two plans, one IMRT (backup) and one RA, 1 arc 135-225° (avoid couch slides) for the 10 first patients

• PTV 95%-107%, median 100%,

• Rectum: max 10ml >60 Gy and less than 50 Gy to half the circumference

• Delta4 measurements OK; • Gamma index 3%, 3mm

– > 95% with index ≤ 1

• Dose deviation– > 85% within ±3% deviation

21

5 fields IMRT: 574 MU (2.15 Gy x 35) RA: 1 arc 135-225° 494 MU (2.15 Gy x 35)

22

5 fields IMRT: 574 MU (2.15 Gy x 35 = 75.25) RA: 1 arc 135-225° 494 MU (2.15 Gy x 35)

23

5 fields IMRT: RA: 1 arc 135-225°

24

IMRT

RA

25

RA today: (2.4 Gy sem.ves. and integrated boost 2.7 Gy prostate) x 25 = 67.5 Gy (EQD2= 81 Gy if α/β=1.5)

26

27

Gamma: 2mm 2%

Measured with Delta4

28

7 field-IMRT

1499 MU (2.7 Gy)

555 MU/Gy

(calibration factor 130MU/Gy)

2 full arc RA

611 MU (2.7 Gy)

Prostate high risk: 2 Gy to the lymph nodes, integrated boost; 2.4 Gy sem.ves. and 2.7 Gy prost, 25 fractions

29

IMRT RA

30

IMRT RA

31

Dose to rectum

32

IMRT

33

Future:

• We would like to treat our high risk protate with LN with two arcs– Prerequisite: RA plan equal or better than IMRT (PTV and

rectum)

• This autumn we have been focusing on commissioning TrueBeam..