Il loop recorder esterno Roberto Maggi
Transcript
Page 1: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Il loop recorder esterno

Roberto Maggi

Page 2: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ECG monitoring and syncope

• In-hospital monitoring• Holter Monitoring • Event recorder• External loop recorder• Remote (at home) telemetry• Implantable loop recorder

Page 3: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

External loop recorder (ELR)

Retrospective (loop) memory which continuously record and delete patient’s ECG:- patient-activation (symptomatic events) - auto-activation (asymptomatic events)

Retrospective memory differentiates ELR from event recorders

Page 4: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.
Page 5: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Event recorder

Page 6: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ELR: available devices

• MCOT, Cardionet• 3300 BT, Vitaphone• King of the Heart, Instromedics• SpiderFlash, Sorin• Cardiocall, Reynolds‐Esaote• Cardio PAL, Medicomp• etc….

Page 7: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ELR remote data trasmission:Intermittent patient-activated (symptoms)

Cardiology Department,Arrhythmologic Centre

Data stored in the device are sent on demand trough an analogic telephone transmission to physician. Local software for analysis

Page 8: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.
Page 9: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.
Page 10: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.
Page 11: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.
Page 12: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

CALL CENTER

Cardiology Department,Arrhythmologic Centre

Automatic ECG transmission of predefined events via Bluetooth wireless link. Physician accesses data via Internet with a secure log-in

ELR remote data trasmission:Automatic intermittent trasmission

(+ patient activated events)

Page 13: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

V-Patch

Page 14: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.
Page 15: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ELR remote data trasmission:continuous automatic (remote telemetry)

CALL CENTER

Cardiology Department,Arrhythmologic Centre

Continuous or 24 hour loop memory,wireless (real-time) to personal data manager and then trans-telephonic to service center

Page 16: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.
Page 17: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ELR remote data trasmission:continuous automatic (remote telemetry)

Page 18: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ELR remote data trasmission:continuous automatic (remote telemetry)

Page 19: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ELR models

At present, no system is proven to be superior to the other

Page 20: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Gula L et al. Am Heart J 2004; 147:644-648

Predictors of ELR diagnosis

Page 21: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ECG monitoring and syncope

• In-hospital monitoring• Holter Monitoring • External loop recorder• Remote (at home) telemetry• Implantable loop recorder

Same

positivity

Criteria

Page 22: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

• Correlation between syncope and an ECG abnormality (brady- or tachyarrhythmia)

• (In the absence of such a correlation):- ventricular pause >3 sec during waking state- periods of Mobitz II 2nd or 3rd degree AV block

during waking state- rapid paroxysmal atrial/ventricular tachycardia

• Correlation between syncope and sinus rhythmexcludes arrhythmic syncope

Positivity criteriaECG monitoring and syncope

Page 23: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

ECG monitoring and syncope

Diagnostic power

In-hospital monitoring 1-7

Holter Monitoring 1-7

External loop recorder 30

Implantable loop recorder 500

Page 24: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

External loop recorder

Linzer M et al. Am J Cardiol 1990; 66: 214-9Sivakumaran et al. AJM 2003; 115:1-5Gula L et al. Am Heart J 2004; 147:644-648

Linzer et alAJC 1996

Sivakumaran et alAJM 2003

Gula et alAm Heart J 2004

Symptom frequency

Median 10/year(>1/month in 70%) 5-10/year 7/month

End-point Pre-syncope or syncope

Pre-syncope orSyncope

Pre-syncope orSyncope

Monitoring duration 4 weeks 4 weeks 4 weeks

Diagnostic value 25% 56% 59%

Page 25: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

External loop recorder

Linzer M et al. Am J Cardiol 1990; 66: 214-9Schuchert A et al . PACE 2003; 26: 1837-1840

Linzer et alAJC 1996

Schuchert et alPACE 2003

Symptom frequency

Median 10/year(>1/month in 70%)

<1/month(3±4 last 6 mos)

End-point Pre-syncope or syncope

Syncope

Monitoring duration

4 weeks 6 weeks

Diagnostic value

25% 4%

Page 26: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Mobile cardiac outpatient telemetry (MCOT)

Rothman SA et al. JCE 2007; 18: 241-7Olson JA et al. JCE 2007; March

Rothman SA et al JCE 2007

Olson JA et alJCE 2007

End-point Syncope or presyncope

Syncope or presyncope

Monitoring duration Up to 30 days Up to 30 days

Diagnostic value 89% 59%

Page 27: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Mobile cardiac outpatient telemetry (MCOT) A prospective multi-center randomized study comparing mobile

cardiac outpatient telemetry versus standard loop event monitoring

Rothman SA et al. JCE 2007; 18: 241-7

MCOT External Loop p

End-point Syncope or presyncope

Syncope or presyncope

Monitoring duration 4 weeks 4 weeks

Diagnostic value 89% 69% 0.008

Arrhythmias 41% 15% 0.001

Page 28: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Mobile cardiac outpatient telemetry (MCOT)

Olson JA et al.J Cardiovasc Electrophysiol 2007

Patients 17 18 21

End-point Syncope or presyncope Palpitations Drug titration

Monitoring duration 4 weeks 4 weeks 4 weeks

Diagnostic value 59% 73% 33%

Page 29: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

External loop recorder: palpitations

Fogel et al. Am J Card 1997; 79:207-208Schmidt et al. Am J Med Sci 1988;296:299–302Bhandari et al. Am Heart J 1992;124:381–386Martinez et al. Int J Cardiol 2004; 97:495-498

End-point Diagnostic value

Fogel et al Symptom/ECGcorrelation 66%

Schmidt et al Symptom/ECGcorrelation 71 %

Bhandari et al Symptom/ECGcorrelation 63% - 69%

Martinez et al Symptom/ECGcorrelation 75%

Page 30: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

External loop recorder: comparative resultsSyncope Unit, Lavagna, 1/2005 – 7/2007

125 pts

55 (44%)

80 (64%)

25 (20%)

125 pts

Inclusion: symptom frequency >1/month

ECG-symptomcorrelation

Arrhythmias

FU 30±16 days

No arrhythmias

VPB/SVPB #18AF/AFLU #18SVT #12VT #4Pauses #3

Palpitations

SyncopePre-syncope

The e-Journal of Cardiology Practice, Volume 7, n° 5, 2008 www.escardio.org

Page 31: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

External Loop Recorder and syncope:Indications

Class I

• None

Class II

• May be useful in patients with recurrent (pre)syncopes who have the clinical or ECG features suggesting an arrhythmic syncope and inter-symptominterval ≤ 4 weeks

ESC Guidelines on Management of Syncope – Update 2004

Page 32: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Holter Event Recorders

LRE LRI

Syncope 0 0 + +++

Palpitations + ++ +++ +++

ECG trace quality

++ + ++ +++

Patient compliance

++ ++ ++ +

Automaticarrhythmiasdetection

++ 0 ++ ++

Costs +++ +++ ++ +

ECG monitoring and syncope

Page 33: Roberto Maggi - GIMSI · Roberto Maggi . ECG monitoring and ... sent on demand trough an analogic telephone transmission to physician. Local software for analysis. CALL CENTER.

Key points•ELRs are much more useful for palpitations than for syncope evaluation

•ILRs consequently is less frequently indicated•Event records are also useful when symptoms last enough to allow the patient to activate the recorder

•The diagnostic value of loop recorders is higher than Holter  


Recommended