+ All Categories
Home > Documents > Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Date post: 12-Jan-2016
Category:
Upload: britton-porter
View: 227 times
Download: 0 times
Share this document with a friend
Popular Tags:
7
Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements
Transcript
Page 1: Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Section Four

Good Laboratory Practice:

Quality Assurance of Analytical Measurements

Page 2: Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

What is Good Laboratory Practice (GLP)?

• The goal of GLP is to certify that every step of the analysis is valid.

• Standard Operating Procedures (SOP)

• Quality Assurance Unit (QAU)

Page 3: Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Validation of Analytical MethodsDefintion Example

Technique Scientific principle gives compositional information

Spectrophotometry

Method Distinct adaptation of technique selected measurement purpose

Pararosaniline method to determine SO2

Procedure Written directions to use a method

ASTM D2914

Sulfur Dioxide

Protocol Set of definitive directions that must be followed, without exception

EPA Reference Method Determine SO2 Atmosphere

Page 4: Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Fig. 4.1. General process for evaluation/validation of methodology.

Validation involves determining:

•selectivity

•linearity

•accuracy

•precision

•sensitivity

•range

•limit of detection

•limit of quantitation

•ruggedness/robustness

Standard reference materials (SRMs) best for determining accuracy.

Validation involves determining:

•selectivity

•linearity

•accuracy

•precision

•sensitivity

•range

•limit of detection

•limit of quantitation

•ruggedness/robustness

Standard reference materials (SRMs) best for determining accuracy.

©Gary Christian, Analytical Chemistry, 6th Ed. (Wiley)

Page 5: Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Fig. 4.2. Response factor plot for Figure 3.8. (p. 104)

RF = (signal – y intercept)/(concentration).

A zero slope indicates linear response over the concentration range selected.

RF = (signal – y intercept)/(concentration).

A zero slope indicates linear response over the concentration range selected.

©Gary Christian, Analytical Chemistry, 6th Ed. (Wiley)

Page 6: Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Fig. 4.3. Dependence of relative standard deviation on concentration.

The precision becomes poorer at low concentrations.

(Also sometimes at high concentrations, as in spectrophotometric measurements –see spectrometric error, Fig. 16.27.)

The precision becomes poorer at low concentrations.

(Also sometimes at high concentrations, as in spectrophotometric measurements –see spectrometric error, Fig. 16.27.)

©Gary Christian, Analytical Chemistry, 6th Ed. (Wiley)

Page 7: Section Four Good Laboratory Practice: Quality Assurance of Analytical Measurements.

Fig. 4.4. Representative z-value distributions for proficiency tests with a series of laboratories.

Z = (Xi-X)/s = (mean conc. of lab – accepted conc.)/(S.D. accepted conc.)

A z-value of 2 means a lab is two S.D. away from the accepted value (95% chance of systematic error).

Z = (Xi-X)/s = (mean conc. of lab – accepted conc.)/(S.D. accepted conc.)

A z-value of 2 means a lab is two S.D. away from the accepted value (95% chance of systematic error).

©Gary Christian, Analytical Chemistry, 6th Ed. (Wiley)


Recommended