Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1
Transcript
Page 1: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Sensors irradiatedCalibration and Measurement at Genova

A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010

1

Page 2: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

test pre-Irradiation

Single chip tuned (@20V) to have• Threshold ~ 3200 e• TOT(Q=20ke) = 60 BC

Am Source scan to verify ToT tuning (0.5 Mhits, ~250 entries per pixel)

Noise scan vs HV

ID on Wafer ID FE-I3

Sensor Type V_bd

Irraggiamento

Test after

IRRAD

3 06051 2E 60 3 10e15

17 09051 2E - Genova -

18 05051 2E - 1 10e15 X

7 07051 3E 65 1 10e15 X – TB, X

9 07052 3E 60 3 10e15 X

13 08052 3E 60 5 10e15

12 08051 4E 25 1 10e15 X

14 09052 4E 60 3 10e15 X – TB, X

16 05052 4E 65 5 10e15

A.Rovani, C.Gemme, A. Micelli – 3D Irst test in Genova – June 25,20102

Page 3: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Test after irradiation • Test done @ T ~ -20° C

IV scan Re-tuning @ Vnominal (Higher than before irradiaton)

• Threshold ~ 3200 e• TOT(Q=20ke) = 60 BC

Standard calibration @ Vnominal: Xtalk, Threshold, Monleak, ToT calib, Am Source scan

Standard calibration repeated for a couple of other voltages settings

Noise scan vs HV

3

Page 4: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

General comments• The bias Voltage required is obviously changed• FE-I3 works fine also after the irradiation• re-tuning is required• The noisy pixels are so noisy that is necessary to disable

the HIT bus and the preamplifier to run the source SCAN

4

Page 5: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Preliminary results: few examples

5

18 (2E – 1*10e15) 7 (3E – 1*10e15)

IV SCAN before and after irradiation (T ~ -20°C)

Page 6: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

6

3 (2E – 3*10e15) 16 (4E – 5*10e15)

IV SCAN before and after irradiation (T ~ -20°C)

Preliminary results: few examples

Page 7: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Preliminary results: few examples

Noise increases after irradiation: Difference before and after irradiation

Pixel masked

7

Page 8: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Preliminary results: few examples

ToT calibration before and after irradiation:Calibration with Clow and Chigh differs a lot after irrad. 9 (3E – 3*10e15 - even if ToT re-tuning after irrad)

8

After irradPre irrad

Page 9: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Preliminary results: few examples

ToT spectrum before and after irradiation9 (3E – 3*10e15)

9

After irradPre irrad

Page 10: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Preliminary results: few examples

10

7 (3E – 1*10e15) 14 (4E – 3*10e15)

Source SCAN after irradiation (TB sensors)

After irradAfter irrad

Page 11: Sensors irradiated Calibration and Measurement at Genova A.Rovani, C.Gemme, A. Micelli 3D Irst test in Genova – June 30,2010 1.

Preliminary results: few examples

11

3 (2E – 3*10e15)

Noise vs HV after irradiation