Servo-u/n Cleaning MX-6989-02 MCV00052945 REVB
Transcript
Page 1: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Servo-u/n Cleaning

MX-6989-02 MCV00052945 REVB

Page 2: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Table of contents

MX-6989-02

General 1

Cleaning 2

Expiratory cassette 3

MCV00052945 REVB

Page 3: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Clean To free from dirt or pollution

Disinfect To reduce the number of harmful microorganisms

Sterilize To free from living microorganisms

General

Different levels of cleaning

MX-6989-02

Clean

Disinfect Sterilize

MCV00052945 REVB

Page 4: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator system.

General

Cautions:

Infected parts

MX-6989-02 MCV00052945 REVB

Page 5: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

The cleaning methods vary between different hospitals.

General

Cleaning intervals

MX-6989-02 MCV00052945 REVB

Page 6: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

User’s manual SERVO-U® / SERVO-n®

Cleaning and maintenance

General

Servo-u/n cleaning instructions

MX-6989-02 MCV00052945 REVB

Page 7: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Table of contents

MX-6989-02

General 1

Cleaning 2

Expiratory cassette 3

MCV00052945 REVB

Page 8: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Cleaning

Diagrams/Wipe off/discard

MX-6989-02

Cleaning when bacterial filter is used

Dismantle

Assemble

Wipe off/ discard

Disinfection when no bacterial filter is used

Dismantle

Rinse

Wipe off/ discard

Assemble

Steam autoclave

Dismantle

Clean/Rinse

Wipe off/ discard

Dry

Autoclave

Dry

Dry

Assemble

Washer- disinfector

Rinse

Disinfectant

MCV00052945 REVB

Page 9: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Wipe the ventilator system and all removable parts with cleaning wipes or a soft lint-free cloth moistened in soap & water or detergent-based disinfectant •  3M Respirator Cleaning Wipes •  Clorox Bleach Germicidal Wipes (may

cause discoloration) •  Super Sani-Cloth Germicidal Wipes

(may cause discoloration) •  Virkon S •  Accel TB (Virox 5). •  In case of more contaminated surfaces,

use ethyl alcohol (70%) or isopropyl alcohol (70%) together with lint free cleaning wipes.

Cleaning diagram

Wipe off/discard

MX-6989-02

Avoid contact with electrical connectors.

MCV00052945 REVB

Page 10: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Table of contents

MX-6989-02

General 1

Cleaning 2

Expiratory cassette 3

MCV00052945 REVB

Page 11: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

The need for cleaning of the expiratory cassette is heavily reduced by using an external bacterial filter. Disconnect and change the expiratory filter: •  Follow manufacturer’s instructions on

how long the filter can be used e.g. 24 or 48 hours.

•  For each new patient, or whenever needed

Expiratory cassette cleaning

When bacterial filter is used

MX-6989-02

Dismantle

Assemble

Wipe off / discard

MCV00052945 REVB

Page 12: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

When no bacterial filter is used, disinfection is necessary.

Expiratory cassette cleaning

Disinfection

MX-6989-02

Dismantle

Rinse

Wipe off/ discard

Dry

Assemble

Washer- disinfector

Rinse

Disinfectant

MCV00052945 REVB

Page 13: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Immediate rinsing in water (<35°C/95°F) can remove particles and reduces the risk of cross-contamination between patients.

Expiratory cassette cleaning

If rinsing before disinfection is not included in the washer disinfector program: •  Rinse the parts

thoroughly in water (<35°C/95°F).

•  Let the water flow through the parts.

Rinse

MX-6989-02

Dismantle

Rinse

Wipe off/ discard

Dry

Assemble

Washer- disinfector

Rinse

Disinfectant

MCV00052945 REVB

Page 14: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

•  Alcohol (ethyl or isopropyl alcohol 70%) •  Aniosyme DD1 •  Gigazyme Plus •  Salvanios pH7 (new instead of

Hexanios)

Expiratory cassette cleaning

Pre-disinfectant

MX-6989-02

Important: Follow the disinfectant manufacturer’s recommendations and instructions.

MCV00052945 REVB

Page 15: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

•  Rinse the parts thoroughly in water to remove all traces of disinfectant

•  Let the water flow through the parts

•  Carefully shake and tilt the cassette, turn upside down and repeat

Expiratory cassette cleaning

Rinse

MX-6989-02

Dismantle

Rinse

Wipe off/ discard

Dry

Assemble

Washer- disinfector

Rinse

Disinfectant

MCV00052945 REVB

Page 16: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

The expiratory cassette will last at least 200 cycles.

Expiratory cassette cleaning

Washer-disinfector

MX-6989-02

Dismantle

Rinse

Wipe off/ discard

Dry

Assemble

Washer- disinfector

Rinse

Disinfectant

MCV00052945 REVB

Page 17: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

•  Wash the parts with water only in a

washer-disinfector at a temperature of 85-95°C (185-203°F)

•  Cleaning agent that can be used: -  neodisher® MediClean forte -  Getinge MIS Detergent

Expiratory cassette cleaning

Washer-disinfector

MX-6989-02

Dismantle

Rinse

Wipe off/ discard

Dry

Assemble

Washer- disinfector

Rinse

Disinfectant

MCV00052945 REVB

Page 18: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

• Alcohol (ethyl or isopropyl alcohol 70%) • Anioxyde 1000 • Cidex OPA

Expiratory cassette cleaning

MX-6989-02

Disinfectant

Important: Follow the disinfectant manufacturer’s recommendations and instructions.

MCV00052945 REVB

Page 19: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Expiratory cassette cleaning

Drying cabinet Room air

Drying

MX-6989-02

Dismantle

Rinse

Wipe off/ discard

Dry

Assemble

Washer- disinfector

Rinse

Disinfectant

Important: Never dry the cassette by applying high pressure air. Internal tubing may be damaged.

MCV00052945 REVB

Page 20: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

•  Steam autoclaving will reduce the lifetime of the expiratory cassette and is not recommend as a cleaning method

•  Before placing the expiratory cassette in a steam autoclave, make sure that no water remains inside the cassette

Expiratory cassette cleaning

Steam autoclave

MX-6989-02

Assemble

Dismantle

Clean/Rinse

Wipe off/ discard

Dry

Autoclave

Dry

MCV00052945 REVB

Page 21: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

•  Instrument parts should be steam autoclaved in a validated process, at a temperature of 134°C (273°F) for 3 minutes.

•  Rubber parts should be steam autoclaved in a validated process, at a temperature of 121°C (250°F) for 15-20 minutes.

Expiratory cassette cleaning

Steam autoclave

MX-6989-02

Assemble

Dismantle

Clean/Rinse

Wipe off/ discard

Dry

Autoclave

Dry

MCV00052945 REVB

Page 22: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

The expiratory cassette will last at least: •  100 autoclaving cycles with 3 minutes

sterilization time at 134ºC (273ºF)

•  30 autoclaving cycles with 18 minutes sterilization time at 134ºC (273ºF)

Cleaning the expiratory cassette

Steam autoclave

MX-6989-02 MCV00052945 REVB

Page 23: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

A pre-use check must be performed after all types of cleaning. If the pre-use check fails: •  Carefully shake/tilt the cassette (5-7 times)

•  Run the cassette in a ventilator system with a test lung for 10 minutes

•  Dry the cassette in room air (if no drying cabinet is available) 12-24 hours, depending on surrounding conditions.

•  Re-do the pre-use check

Cleaning the expiratory cassette

Pre-use check

MX-6989-02 MCV00052945 REVB

Page 24: Servo-u/n - Getinge...filter: • Follow manufacturer’s instructions on how long the filter can be used e.g. 24 or 48 hours. • For each new patient, or whenever needed Expiratory

Getinge is a leading global provider of innovative solutions for operating rooms, intensive-care units, hospital wards, sterilization departments, elderly care and for life science companies and institutions. With a genuine passion for life we build quality and safety into every system. Our unique value proposition mirrors the continuum of care, enhancing efficiency throughout the clinical pathway. Based on our first-hand experience and close partnerships, we are able to exceed expectations from customers – improving the every-day life for people, today and tomorrow.

www.getinge.com

Getinge is a registered trademark of Getinge AB, its subsidiaries, or affiliates in the United States or other countries • Copyright 2020 Getinge AB or its subsidiaries or affiliates • Servo-u, Automode, NAVA, Safety Scale, and Servo Compass are trademarked by Maquet Critical Care AB . All rights reserved • CAUTION: Federal (US) law restricts this device to sale by or on the order of a physician. • Refer to Instructions for Use for current indications, warnings, contraindications, and precautions. Sales Office: Getinge · 45 Barbour Pond Drive · Wayne, NJ 07470 · USA Manufacturer: Maquet Critical Care AB · 171 54 Solna, Sweden · Phone: +46 (0)10-335 00 00 · [email protected]

MX-6989-02 MCV00052945 REVB