Journal of Analytical Toxicology, Vol. 24, October 2000 SimultaneousDetection and Quantitation of Diethylene Glycol, EthyleneGlycol, and the ToxicAlcoholsin Serum using Capillary Column Gas Chromatography Robert H. Williams 1,*, Steve M. Shah 1, Jack A. Maggiore 1, and Timothy B. Erickson 2 IDepartmentof Pathology, Division of Clinical Pathology(M/C 750), Universityof Illinois at Chicago Medical Center, 840 South Wood Street, 201G CSB, Chicago, Illinois 60612 and 2Department of Emergency Medicine, Division of Toxicology (NVC724), Universityof Illinois at ChicagoMedical Center, 808 South Wood Street, 471 CME, Chicago, Illinois 60612 Abstract I Determination of toxic glycols and alcohols in an emergency setting requires a rapid yet accurate and reliable method. To simultaneously determine diethylene glycol (DEG) along with ethylene glycol, methanol, isopropanol,acetone, and ethanol, we modified a previously developed gas chromatographic (GC) method. The system used a Hewlett-Packard 6890 GC with EPC, a Gooseneck splitless liner, and an Rtx-200 capillary column (30 m x 0.53-ram i.d., 3 ram). After serum samples were deproteinized using ultrafiltration (MUlipore Ultrafree| 1 mL of the protein-free filtrate was manually injected into the GC. Internal standards for alcohols (and acetone) and glycolswere n-propanol and 1,3- butanediol, respectively. All compounds eluted within 3.5 rain (linear temperature gradient from 40 to 260~ total run time was 6.5 rain. Limit of detection and linear range for all compounds were 1 or 2.5 mg/dt and 0-500 mg/dL, respectively. In addition, there was no interference from propionic acid, propylene glycol, and 2,3- butanediol. The modifications in the equipment and temperature program allowed increased resolution and thus, detection and reliable quantitation of DEG and other common toxic glycols and alcohols of clinical interest. Introduction Acute intoxication by the common toxic alcohols, methanol, isopropanol (and its metabolite acetone), and ethylene glycol, is well known in toxicology. Methanol and ethylene glycol through the generation of toxic metabolites produce a profound metabolic acidosis (1). The metabolites of methanol can cause severe damage to the optic nerve, whereas the metabolites of ethylene glycol cause renal and pulmonary damage (1). Isopropanol is also toxic--it is very caustic to the gastrointestinal tract--and its metabolite, acetone, is a major CNS depressant (1). However, because no acidic products are produced during the metabolism ' Author to whom correspondence should be addressed. of isopropanol, metabolic acidosis is rarely observed unless the patient is very hypotensive (increased lactic acid production) (1). All three alcohols cause an increase in the osmolal gap. Diethylene glycol (DEG) is also very toxic; however, it is not routinely determined as part of a toxic alcohol panel. DEG is a derivative of ethylene glycol with both compounds causing acute renal failure (2,3). DEG, like ethylene glycol, can increase the osmolal gap. However, unlike ethylene glycol, DEG does not cause a metabolic acidosis (2,3). DEG has occurred as a contaminant or has been used a substi- tute for propylene glycol and glycerin in the pharmaceutical industry in some countries (2). In 1937, 105 patients died in the United States when 72% DEG was used as a diluent in the antibi- otic preparation called Elixir Sulfanilamide (2). Consequently, the 1938 the Federal Food, Drug, and Cosmetic Act was passed (2). Since that time, poisoning by diethylene glycol in North America has been rare. However, it has been associated with several inter- national epidemics having numerous fatalities over the past 60 years (3) (Table I). Because diethylene glycol, unlike ethylene glycol (or methanol), does not produce a profound metabolic aci- dosis, it is often missed in the differential diagnosis of a patient that presents to the emergency department with a mild acidosis and some signs of nephrotoxicity. In addition, the development of Table I. International Epidemics of Diethylene Glycol Poisoning Year Country Numberof deaths 1937 United States 105 1967 South Africa 7 1985 Spain 5 1986 India 14 1990 Nigeria 47 1990-1992 Bangladesh 51 * 1992 Argentina 26 * An additional 157 cases were also likely due to DEG poisoning, but not confirmed. Reproduction tphotocopying) of editorial content of this journal is prohibited without publisher's permission. 621
Transcript
Page 1: Simultaneous Detection and Quantitation of Diethylene ... · otic preparation called Elixir Sulfanilamide (2). ... Control materials for precision and accuracy. Quality-control

Journal of Analytical Toxicology, Vol. 24, October 2000

Simultaneous Detection and Quantitation of Diethylene Glycol, Ethylene Glycol, and the Toxic Alcohols in Serum using Capillary Column Gas Chromatography Robert H. Williams 1,*, Steve M. Shah 1, Jack A. Maggiore 1, and Timothy B. Erickson 2 I Department of Pathology, Division of Clinical Pathology (M/C 750), University of Illinois at Chicago Medical Center, 840 South Wood Street, 201G CSB, Chicago, Illinois 60612 and 2Department of Emergency Medicine, Division of Toxicology (NVC 724), University of Illinois at Chicago Medical Center, 808 South Wood Street, 471 CME, Chicago, Illinois 60612

Abstract I

Determination of toxic glycols and alcohols in an emergency setting requires a rapid yet accurate and reliable method. To simultaneously determine diethylene glycol (DEG) along with ethylene glycol, methanol, isopropanol, acetone, and ethanol, we modified a previously developed gas chromatographic (GC) method. The system used a Hewlett-Packard 6890 GC with EPC, a Gooseneck splitless liner, and an Rtx-200 capillary column (30 m x 0.53-ram i.d., 3 ram). After serum samples were deproteinized using ultrafiltration (MUlipore Ultrafree| 1 mL of the protein-free filtrate was manually injected into the GC. Internal standards for alcohols (and acetone) and glycols were n-propanol and 1,3- butanediol, respectively. All compounds eluted within 3.5 rain (linear temperature gradient from 40 to 260~ total run time was 6.5 rain. Limit of detection and linear range for all compounds were 1 or 2.5 mg/dt and 0-500 mg/dL, respectively. In addition, there was no interference from propionic acid, propylene glycol, and 2,3- butanediol. The modifications in the equipment and temperature program allowed increased resolution and thus, detection and reliable quantitation of DEG and other common toxic glycols and alcohols of clinical interest.

Introduction

Acute intoxication by the common toxic alcohols, methanol, isopropanol (and its metabolite acetone), and ethylene glycol, is well known in toxicology. Methanol and ethylene glycol through the generation of toxic metabolites produce a profound metabolic acidosis (1). The metabolites of methanol can cause severe damage to the optic nerve, whereas the metabolites of ethylene glycol cause renal and pulmonary damage (1). Isopropanol is also toxic--it is very caustic to the gastrointestinal tract--and its metabolite, acetone, is a major CNS depressant (1). However, because no acidic products are produced during the metabolism

' Author to whom correspondence should be addressed.

of isopropanol, metabolic acidosis is rarely observed unless the patient is very hypotensive (increased lactic acid production) (1). All three alcohols cause an increase in the osmolal gap.

Diethylene glycol (DEG) is also very toxic; however, it is not routinely determined as part of a toxic alcohol panel. DEG is a derivative of ethylene glycol with both compounds causing acute renal failure (2,3). DEG, like ethylene glycol, can increase the osmolal gap. However, unlike ethylene glycol, DEG does not cause a metabolic acidosis (2,3).

DEG has occurred as a contaminant or has been used a substi- tute for propylene glycol and glycerin in the pharmaceutical industry in some countries (2). In 1937, 105 patients died in the United States when 72% DEG was used as a diluent in the antibi- otic preparation called Elixir Sulfanilamide (2). Consequently, the 1938 the Federal Food, Drug, and Cosmetic Act was passed (2). Since that time, poisoning by diethylene glycol in North America has been rare. However, it has been associated with several inter- national epidemics having numerous fatalities over the past 60 years (3) (Table I). Because diethylene glycol, unlike ethylene glycol (or methanol), does not produce a profound metabolic aci- dosis, it is often missed in the differential diagnosis of a patient that presents to the emergency department with a mild acidosis and some signs of nephrotoxicity. In addition, the development of

Table I. International Epidemics of Diethylene Glycol Poisoning

Year Country Number of deaths

1937 United States 105 1967 South Africa 7 1985 Spain 5 1986 India 14 1990 Nigeria 47 1990-1992 Bangladesh 51 * 1992 Argentina 26

* An additional 157 cases were also likely due to DEG poisoning, but not confirmed.

Reproduction tphotocopying) of editorial content of this journal is prohibited without publisher's permission. 621

Page 2: Simultaneous Detection and Quantitation of Diethylene ... · otic preparation called Elixir Sulfanilamide (2). ... Control materials for precision and accuracy. Quality-control

severe metabolic acidosis from ingestion of ethylene glycol (or methanol) depends upon the time of ingestion and presentation to the emergency department because the formation of toxic "acidic" metabolites is time dependent and highly variable, espe- cially if high quantities of ethanol have been consumed (1). Thus, it is not uncommon for patients who present soon after ingestion of these compounds to have a normal acid-base status or be mildly acidotic.

The American Academy of Clinical Toxicology has developed a set of guidelines for laboratories that serve as regional toxicology treatment centers (4). To be clinically useful, results for toxic alcohol levels need to be available within 2 h (4). Thus, a rapid method that can simultaneously determine DEG along with the common toxic alcohols in an emergency setting would be clini- cally useful worldwide. However, most chromatographic methods used to measure alcohols and diols require separate conditions. They must have different columns and (or) chromatographic sys- tems because of lengthy analysis times involved. Thus, the sepa- rate analyses for these compounds are both labor intensive and time consuming (5-13).

Most alcohols of clinical interest are volatile, and thus can be readily determined by direct injection or headspace gas chro- matography (GC) using flame ionization detection (12,13). In the past, diols required derivatization to form boronic esters prior to GC analysis because underderivatized diols chromatographed poorly and yielded low sensitivity for flame ionization detectors (14,15). However, formation of boronic esters with diethylene glycol is not possible (16), so gas chromatography-mass spec- trometry (GC-MS) has been used to determine DEG and ethylene glycol (16). Methods have been developed for did analysis that do not require derivatization; however, they are not suitable for alcohol analysis (17,18). Enzymatic methods, although used for screening purposes, have had limited success and cannot be used to determine alcohols and diols simultaneously (12,19-21).

Recently, Livesey et al. (22) developed a GC method that has the capabilities to simultaneously determine many alcohols and diols of clinical interest without interference from compounds noted in the past, such as 2,3-butanediol, propylene glycol, proprionic acid, and valproic acid (19). However, the method was not opti- mized to detect diethylene glycol. Thus, the objective of this work was to modify the method of Livesey et al. (22) to detect DEG in addition to the common alcohols and diols encountered in the emergency setting.

Methods and Materials

Instrumentation Analysis was performed in the splitless mode with a Hewlett-

Packard (Palo Alto, CA) model 6890 GC equipped with electronic pneumatic control, a flame ionization detector, and a Restek Rtx-200 capillary column (Crossbond | trifluoropropylmethyl polysiloxane, 30 m x 0.53-ram i.d., 3-lain film thickness, Restek Corp., Bellefonte, PA). The injection port liner was a Gooseneck splitless liner (Restek Corp.). The temperature of the injector port and detector was set at 250~ The oven temperature program consisted of i rain at 40~ followed by a temperature gradient from 40~ to 260~ at a rate of 70~ The total run time was

622

Journal of Analytical Toxicology, Vol. 24, October 2000

6.5 min. Helium carrier gas flow rate was 80.1 mL/min; for the hydrogen-air flame, the hydrogen flow rate was 40 mL/min, and the airflow rate was 400 mL/min. Integration of peak area was obtained using an HP ChemStation/Integrator data system.

Reagents, supplies, and samples Preparation of standards. All reagents were analytical/high-

performance liquid chromatography (HPLC) grade. DEG, 1-n- propanol (n-propanol), 2-propanol (isopropanol), and acetone were purchased from Sigma (St. Louis, MO). Ethylene glycol, methanol, ethanol, 1,3-butanediol, 2,3-butanediol, and propylene glycol were purchased from Aldrich (Milwaukee, WI). Stock stan- dards for DEG, ethylene glycol, methanol, isopropanol, acetone, and ethanol were either prepared from pure chemicals (listed) at a concentration of 1000 mg/dL or purchased commercially as a custom alcohol/glycol mixture (0.1% or 1000 mg/dL) from Restek Corp. A working standard containing a mixture of alco- hols/diols at a concentration of 100 mg/dL was prepared by adding 1 mL of each compound to a lO-mL volumetric flask using distilled-deionized water as the diluent. Internal standards for alcohols and acetone (n-propanol) and for glycols (2,3-butane- diol) were prepared by making a 1:10 dilution of the stock stan- dard (800 mg/dL) to obtain a final concentration of 80 mg/dL.

Control materials for precision and accuracy. Quality-control materials used to assess precision were UTAK Volatiles Plus

Table il. Retention Times for Alcohols, Diols, and Internal Standards

Retention t ime Retention time window Compound* (min) (rain)

Methanol 0.84 0.80-0.88 Ethanol 1 .(/7 0.9& I. 15 Isopropano[ 1.28 1.24-1.32 n-PropanoP 1.54 1.50-1.58 Acetone 1.76 1.71-i .81 Ethylene glycol 2.59 2.54-2.64 Propylene glycol 2.69 2.61-2.77 2,3-Butane diol 2,73 2.65-2.81 1,3-Butane diol* 3,07 2.99 3.15 Diethylene glycol 3.44 3.35-3.53

* Target concentration = 100 rag/alL "~ Internal standard for alcohols. * Internal standard for diols.

Table III. Day-to-Day Precision of Commercial Controls (n = 10)

UTAK Volatiles Plus UTAK Volatiles Plus Compound 10 mg/dL - CV (%) 25 mg/dL - CV (%)

Methanol 3,7 7.3 Ethanol 3.4 &6 Isopropanol 5.3 7.5 Acetone 4.8 9.6 Ethylene glycol 9.8 4.0 Diethylene glycol 9.8 8.8

Page 3: Simultaneous Detection and Quantitation of Diethylene ... · otic preparation called Elixir Sulfanilamide (2). ... Control materials for precision and accuracy. Quality-control

Journal of Analytical Toxicology, Vol. 24, October 2000

serum-based controls (UTAK Corp., Valencia, CA) containing DEG, ethylene glycol, methanol, isopropanol, acetone, ethanol, propylene glycol, and 2,3-butanediol at final concentrations of 10 and 25 mg/dL. For accuracy studies, UTAK custom-made, serum- based calibrators containing diethylene glycol, ethylene glycol, methanol, isopropanol, acetone, ethanol, propylene glycol, and 2,3 butanediol at final concentrations of 10, 50, 100, 200, and 300 mg/dL were purchased. Additional calibrators at concentrations of 400 and 500 mg/dL were purchased for ethanol. The UTAK cal- ibrators were also used for calibration verification.

Ultrafiltration units. To filter the standards or obtain a protein- free filtrate from serum samples, Ultrafree-MC filter units having a 10,000-Da-cutoff were purchased from the Millipore Corp.

<=

[ eO00~

] J~ I

O~ . . . . . . .

q I

0.11 1 1.5 2 2.5 3 3.6

Time (rain)

"i I i I

gO00~

I B

-t [ I ! I I !

Time (mln)

_ [ ~ ] o

i P t

i ~ N NN i cJ~

05 1 1.S 2 2.5 3 3,$

Time (min)

Figure I. Chromatogram of the high alcohol/did standard, concentration of I00 mg/dt of each alcohol and diol (A). Chromatogram of the low alcohol/diol standard, concentration of 10 mg/dl of each alcohol/diol (B). Chromatogram of a patient positive for isopropanol ingestion (C). Chromatogram of patient positive for ethylene glycol and diethylene glycol poisoning (D). Chromatogram of patient pos- itive for methanol poisoning (E).

(Bedford, MA). The ultrafiltration membranes were preserved with triethylene glycol (not glycerol) to avoid potential interfer- ence.

Patient samples. The Institutional Review Board at the University of Illinois at Chicago Medical Center approved the use of this method to analyze serum samples obtained from patients who presented to our emergency department suspected of a toxic alcohol or diol ingestion.

Procedure Calibration. To prepare standards for calibration, 25 ~L of

internal standard (containing 80 mg/dL of n-propanol/1,3- butanediol) was added to an Uttrafree-MC filter unit followed by

the addition of 100 p~L of the working standard A (100 mg/dL alcohols/diols). Samples were vortex

mixed briefly then centrifuged for 5 min. One microliter of the centrifuged standard was manu-

l ally injected directly onto the column using a Hamilton 1-~L syringe. Calibration is set to 100 mg/dL for alcohols/diols.

t Sample analysis. In this method 25 [aL of internal standard (containing 80 mg/dL of n- propanol/1,3-butanediol) was added to labeled Ultrafree-MC filter units followed by the addition of 100 ~tL of patient sample, working standard alcohol/diol mixture (duplicate), and UTAK con- trols to the appropriate filter units. Samples were vortex mixed for 15 s followed by centrifugation for 5 min. After centrifugation, 1 ~tL of the protein- free ultrafiltrate containing the internal standards was manually injected directly onto the column using a Hamilton syringe.

Results

Table II shows the average retention time and retention time window in minutes for some of the common alcohols and diols encountered in an emergency department setting. Several chro- matograms using this direct injection method with a protein-free ultrafiltrate are depicted in Figure 1. A serum-based calibrator spiked with a target value of 100 mg/dL is shown Figure 1A, and a serum-based calibrator with a target value of 10 mg/dL is illustrated in Figure lB. Regardless of concentration the chromatographic peaks are dis- tinct and symmetrical and baseline separation was achieved. Figure I also illustrates three chro- matograms of samples obtained from patients who presented to our emergency department with isopropanol ingestion (Figure 1C), ethylene glycol/diethylene glycol poisoning (Figure 1D), and methanol poisoning (Figure 1E). It should be noted that the ethanol peak due to ethanol admin- istration as part of the therapeutic regimen is also depicted in both cases (Figure 1D and 1E). With

623

Page 4: Simultaneous Detection and Quantitation of Diethylene ... · otic preparation called Elixir Sulfanilamide (2). ... Control materials for precision and accuracy. Quality-control

Journal of Analytical Toxicology, Vol. 24, October 2000

the isopropanol case, acetone is separated from the other common alcohols/diols; in the case of ethylene glycol poisoning, DEG is also clearly separated and distinguishable on the chro- matogram. Alcohols and diols of clinical interest evaluated by this method eluted within 3.5 rain. No interfering peaks appeared to be present in the chromatograms with this method modification.

Precision, accuracy, and carryover The day-to-day precision expressed as the percent coefficient of

variation (%CV) for the two levels of the UTAK commercial con- trol (10 mg/dL, 25 mg/dL) over a period of 10 days is given in

i

J

:]

Table III. All are within acceptable limits (less than 10%) and comparable to values reported in the literature with the use of GC. The CV for within-run precision was less than 5% for all com- pounds.

Accuracy was performed on the UTAK commercially spiked, serum-based samples with target values ranging from 10 to 500 mg/dL depending on the compound. Table IV shows the relative accuracy for DEG, ethylene glycol, methanol, isopropanol, ace- tone, and ethanol with their corresponding CVs over a period of four days. At 10 mg/dL and 50 mg/dL, all CVs were less than 15%; at all other levels, the CVs were less than 10%.

Carryover for each alcohol or diol was assessed by analyzing a high-level serum-based calibrator (300 mg/dL) followed by three consecutive serum-blank sam-

D pies. No detectable carryover was noted with any of the compounds.

tinearity and limits of detection (LOD) and quantitation (LOQ)

Linearity was assessed using blank serum and serum spiked with each alcohol or diol over the range of 10 to 500 mg/dL. All calibration curves were linear over the entire concentration range; the calibration curves along with the linear

' . . . . . . . . . . . . . r . . . . . . . . . . . . - . . . . . . . . . . . . . . . . , . . . . i . . . . 0,5 1 1 . 5 2 2.15 3 3.5

Time (min)

7 0 0 0 -:

4OOO Z

�9 'o5' ' �9 q- ,!5 ' 2 2ts ~ '~ie

E

, J l !

Time (min)

Figure 1. (continued) Chromatogram of the high alcohol/diol standard, concentration of 100 mg/dL of each alcohol and diol (A). Chromatogram of the low alcohol/diol standard, concentration of 10 mg/dL of each alcohol/diol (B). Chromatogram of a patient positive for isopropanol ingestion (C). Chromatogram of patient positive for ethylene glycol and diethylene glycol poisoning (D). Chromatogram of patient positive for methanol poisoning (E).

regression lines are illustrated in Figure 2. Limits of detection were determined by analyzing, in trip- licate, specimens that contained 0 to 10 mg/dL of each compound. The LOD and LOQ for the alco- hols and acetone were determined to be 1.0 mg/dL and be 2.5 mg/dL, respectively. For DEG and ethy- lene glycol, the LOD was 2.5 mg/dL and the LOQ was 5 mg/dL.

Discussion

Concomitant analysis of alcohols and diols poses several problems since most methods are not amenable to both classes of compounds. Most alcohols of common clinical interest are volatile, and thus these compounds are readily analyzed without pretreatment by direct injection using headspace GC (12,13). Compounds such as ethyl- ene glycol and DEG generally require manipula- tion of the sample by derivatization prior to

Table IV. Interday Accuracy Data for Alcohols and Diols (n = 4)

Target Concentration (% target • % CV) Compound 10 mg/dL 50 mg/dL 100 mg/dL 200 mg/dL 300 mg/dL 400 mg/dL 500 mg/dL

Ethanol 105.0 + 6.7 103.0 + 8.3 102.8 • 5.1 106.9 • 10.1 108.5 + 5.6 Methanol 108.0 • 13.2 95.0 • 15.6 99.4 _+ 4.5 100.8 • 6.9 99.8 • 5.7 lsopropanoI 114.0 _+ 19.3 94.0 • t 5.6 97.0 • 6.t t02.2 + 5.2 99.5 • 4.2 Acetone 123.0• 97.0+13.4 94.3• 102.9_+6.8 99.3+8.9 Ethylene glycol 119.0_+4.7 90.2_+5.2 95.9• 101.5• 100.0_+4.9 Diethylene glycol 137.0 _+ 5.6 95.4 + 6.8 97.0 _+ 1.2 101.9 _+ 4.0 99.5 • 5.7

107.9+8.6 109.8+8.9

624

Page 5: Simultaneous Detection and Quantitation of Diethylene ... · otic preparation called Elixir Sulfanilamide (2). ... Control materials for precision and accuracy. Quality-control

A

40

8oo] y=13.969x+2.1145

t 1 1 0 10 20 30

r ~ I I I

y = 21.561x + 8.3428 E

600 .J

E E 500 g

i 400

8 u 300

'~ 200

100 o

'< 0

D y = 15 .536x+ 1.5916

Journal of Analytical Toxicology, Vol. 24, October 2000

10 20 30

Acetone - average peak area

800 B 600.

500 _ 500.

400 400.

6 . 9 300 ~ .B 300. O ~

~ 200 ~ ~ r ~ 200.

loo N loo.

o 0 0 5 10 15 20 25 30

Methanol - average peak area

5 10 15 20

Ethylene glycol - average peak area

25

600 600 y = 12.772x + 2.0829 C

500 ~;~'" 400 , Z ' "

30O

2OO

100

0 0 10 20 30 40

I sopropano l - average peak area

Ethanol - average peak area

I

, 0

Q . O 8o

'~ 500 . 1

| ,oo ;.B 300 i2~176 ,~ loo

o

y = 23.37x- 0.2291

0 5 10 15 20

Diethylene glycol - average peak area

25

Figure 2. Linearity plots using regression analysis for ethanol (A), methanol (B), isopropanol (C), acetone (D), ethylene glycol (E), and diethylene glycol (F). Theoretical concentrations were O, 10, 50, 100, 200, 300, 400, and 500 mgldL.

625

Page 6: Simultaneous Detection and Quantitation of Diethylene ... · otic preparation called Elixir Sulfanilamide (2). ... Control materials for precision and accuracy. Quality-control

Journal of Analytical Toxicology, Vol. 24, October 2000

analysis with GC (as with ethylene glycol) or GC-MS (as with DEG) (14-16).

Methods have been developed to measure ethylene glycol and/or DEG that use direct injection onto the column in a GC (17,18); however, they cannot be adapted to simultaneously measure the common alcohols. One of the problems encountered with GC of diols, especially with direct injection onto the column, is incom- plete elution from the column, causing contamination at the GC inlet/liner interface, carryover, and peak tailing. Unlike other methods in use, the present method did show a problem with car- ryover or peak tailing after 100 consecutive manual injections. Use of protein-free ultrafiltrates rather than whole serum permitted direct column injection for the analysis of DEG and the other common toxic diols and alcohols. By using the Hewtett-Packard 6890 GC in the splitless mode along with modification of the liner and temperature program, an increase in resolution was noted compared to the original method of Livesey et al. (22). Simultaneous detection of DEG along with some of the other common diols and alcohols of clinical interest was also achieved. The increase in inlet/detector temperature and higher tempera- ture gradient provided complete elution and excellent baseline separation of DEG and ethylene glycol along with the common toxic alcohols. This method is rapid (calibration, controls, and patient samples analyzed within an hour) and thus meets the cri- teria established by the American Academy of Clinical Toxicology for turnaround time for toxic alcohol and diols (4). This method has been in clinical use for over a year for patients seen in our emergency department and has had a positive impact on the treat- ment of our patients because of the reduction in turnaround time.

References

1. K.E. Burkhart and K.W. Kulig. The other alcohols--methanol, ethy- lene glycol, and isopropanol. Emerg. Med. Clin. N. Am. 8:913-928 (1990).

2. P.M. Wax. Elixirs, d iluents, and the passage of the 1938 federal food, drug, and cosmetic act. Ann. Int. Med. 122:456-461 (1995).

3. K.L. O'Brien, J.D. Selanikio, C. Hecdivert, M.F. Placide, M. Louis, D.B. Barr, J.R. Barr, C.J. Hospedales, M.J. Lewis, B. Schwartz, R.M. Philen, S. St. Victor, J. Espindola, Li . Needham, and K. Denerville. Epidemic of pediatric deaths from acute renal failure caused by diethylene glycol. J. Am. Med. Assoc. 279:1175-1180 (1998).

4. American Academy of Clinical Toxicology. Facility assessment guidelines for regional toxicology treatment centers. J. Toxicol. Clin. Toxicol. 31: 211-217 (1993).

5. M. Balikova and J. KoNicek. Rapid determination of ethylene glycol

at toxic levels in serum and urine. J. Chromatogr. 434:469-474 (1988).

6. S. Shahangian, V.U Robinson, and T.A. Jennison. Formate concen- trations in a case of methanol ingestion. Clin. Chem. 30:1413-1414 (1984).

7. J.A. Jonsson, A. Eklund, and L. Molin. Determination of ethylene glycol in postmortem blood by capillary gas chromatography. J. Anal. Toxicol. 13:25-26 (1989).

8. S-T. Cheung and W-N. Lin. Simultaneous determination of methanol, ethanol, acetone, isopropanol and ethylene glycol in plasma by gas chromatography. J. Chromatogr. 414:248-250 (1987).

9. N.B. Smith. Determination of serum ethylene glycol by capillary gas chromatography. Clin. Chim. Acta 144:269-272 (1984).

10. R.J. Flanagan, S. Dawling, and B.M. Buckley. Measurement of ethy- lene glycol (ethane-l,2-diol) in biological specimens using derivati- zation and gas-liquid chromatography with flame ionization detection. Ann. Clin. Biochem. 24:80-84 (1987).

11. D.J. Doedens. Methods for determination of ethylene glycol. Vet. Hum. ToxicoL 25:96-101 (1983).

12. R.H. Williams. Clinical laboratory assessment of ethanol and the toxic alcohols. In Lexicomp's Poisoning & Toxicology Handbook, 3rd ed., J.B. Leikin and F.P. Paloucek, Eds. Lexicornp, Hudson, OH, 1998, pp 894-905.

13. W.H. Porter. Clinical toxicology. In Tietz Textbook of Clinical Chemistry, 3rd ed., C.A. Burtis and E.R. Ashwood, Eds. W.B. Saunders, Philadelphia, PA, 1999, pp 906-981.

14. R.L. Peterson and D.O. Rodgerson. Gas chromatographic determi- nation of ethylene glycol in serum. Clin. Chem. 20:820--824 (1974).

15. D.W. Robinson and D.S. Rewe. A gas chromatographic procedure for quantitation of ethylene glycol in postmortem blood. J. Anal Toxicol. 5:69-72 (1981 ).

16. R.R. Freeman, Ed. High-resolution gas chromatography, 2nd ed. Hewlett Packard, Palo Alto, CA, 1981.

17. K. Aarstad, O. Dale, O., Aakervik, O. Steiner, and K. Zahlsen. A rapid gas chromatographic method for the determination of ethylene glycol in serum and urine. J. Anal. Toxicol. 17:218--221 (1993).

18. L.E. Edinboro, C.R, Nanco, D.M. Soghioan, and A. Poklis. Determination of ethylene glycol in serum utilizing direct injection on a wide-bore capillary column. Ther. Drug Monit. 15:220-223 (1993).

19. K.W. Ryder, M.R. Glick, and S.A. Jackson. Emergency screening for ethylene glycol in serum. Clin. Chem. 32:1574-1577 (1986).

20. J. Standefer and W. Blackwell. Enzymatic method for measuring ethylene glycol with a centrifugal analyzer. Ctin. Chem. 37: 1734-1736 (1991 ).

21. B. Vinet. An enzymatic assay for the specific determination of methanol in serum. Clin. Chem. 33:2204-2208 (1987).

22. J. F. Livesey, S.L. Perkins, N.E. Tokessy, and M.J. Maddock. Simultaneous determination of alcohols and ethylene glycol in serum by packed- or capillary-column gas chromatography. Clin. Chem. 41: 300-305 (1995).

Manuscript received March 31, 2000; revision received June 9, 2000.

626