7/29/2019 Site Planning for Medical centres
http://slidepdf.com/reader/full/site-planning-for-medical-centres 1/6
DOTmedbusiness news I january 2010 www.dotmed.com42
Site-Planning for Medical
Imaging Equipment
Informed Advance Planning Simplies
a Potentally Complex installaton
By Joel Kellogg, ETS-Lindgren and Dave Jordan, West Physics
7/29/2019 Site Planning for Medical centres
http://slidepdf.com/reader/full/site-planning-for-medical-centres 2/6
DOTmedbusiness news I january 2010 43
Designing space for medical imaging equipment
can be quite complex and involved as there are
many items that must be addressed in order to
successfully develop a site. Good site plan-
ning must thoroughly evaluate both the impact
of the imaging equipment on the surroundings (environmen-
tal concerns) and the impact of the space itself on the per-
formance of the imaging equipment and the personnel using
the equipment (performance concerns). Depending upon the
equipment, concerns may include radiation, magnetic and/or radio frequency (RF) shielding, electromagnetic interfer-
ence (EMI), vibration, and acoustic requirements. There may
also be concerns over co-siting medical imaging equipment
as one piece of equipment could have a negative impact on
another piece of equipment. As a result, it is critical to de-
velop a site plan and workow process well in advance that
is functional for users, patients, and the planned equipment.
Environmental ConcernsShielding (Magnetic, Radio Frequency and Radiation) and
Acoustics Magnetic Shielding
Shielding is critical to the proper development of a site for imaging equipment. Some imaging equipment will require ra-
diation shielding and other equipment, mainly Magnetic Reso-
nance Imaging (MRI) systems, will require Radio Frequency
(RF) and magnetic shielding. Careful planning that accounts for
workow and surrounding areas can also help reduce the level
of shielding required representing a cost savings to the owner
while providing a functional, efcient work environment.
When planning for MRI systems, there are some sim-
ple things that can be done to avoid excessive magnetic
shielding costs. While the majority of MRI systems re-
quire an RF shield, the magnetic shielding requirements
are driven by specic site selection. For example, in large
imaging suites with multiple MRI systems, placing MRI
systems side-by-side with limited spacing between MRI
suites will drive the requirements for magnetic shielding.
In some cases, this may also require expensive modica-
tions to the equipment itself. While many sites will only
be concerned with meeting FDA recommendations of
5-Gauss containment in public areas surrounding MRI
suites, placing MRI systems next to each other can cre-
ate a concern for crosstalk. Crosstalk results in MRI sys-
tems having an interactive impact on each other. Figure
1 shows crosstalk concerns which result in increased
magnetic shielding requirements. To avoid crosstalk in
a situation where MRI systems are placed in close prox-
imity, the magnetic shielding requirements may change.
For example, a magnet vendor may require that the 3
Gauss fringe elds do not intersect, which means that the
magnetic shielding will need to be designed to contain
3 Gauss rather than 5 Gauss. This will result in a heavier
magnetic shield and increased shielding costs. A simple solu-
tion may be to place the MRI equipment rooms, which house
the MRI’s electronic systems, between adjacent MRI systems.
This may provide enough spacing so that the 3 Gauss lines do
not intersect; however, even if the 3 Gauss lines intersect, the
amount of magnetic shielding required to separate the 3 Gauss
lines will cost much less than having the MRI systems located
side by side.
It is also important to consider the areas surrounding im-
aging equipment that will need radiation and MRI shielding.
There are many pieces of equipment that could be adversely
impacted by high static magnetic elds similar to those gener -
ated by an MRI system. Ultrasound equipment, computerized
tomography (CT), cathode ray tube (CRT) monitors, linear
accelerators, and electron microscopes are just a few exam-
ples of equipment that can be negatively impacted by the one
Gauss fringe eld. Magnetic shielding costs can be reduced by
Figure 1
Crosstalk concerns result in increased magnetic shielding
requirements due to MRI systems in close proximity.
Figure 2
Providing separation between adjacent MRI systems will reduce the
potential for crosstalk and decrease the amount of magnetic shielding
required.
7/29/2019 Site Planning for Medical centres
http://slidepdf.com/reader/full/site-planning-for-medical-centres 3/6
DOTmedbusiness news I january 2010 www.dotmed.com44
placing equipment at distances that are
outside the maximum allowable static
magnetic eld as required by the OEM
specications. Figure 2 shows how
providing separation between adjacent
MRI systems will reduce the potential
for crosstalk and decrease the amount of
magnetic shielding required. It is pos-
sible to place equipment such as ultra-
sounds and CTs next to, above, or below
an MRI system, but it should be under-
stood that such a placement may change
magnetic shielding requirements and
increase shielding costs. However, there
may be workow reasons that make it
advantageous to place a piece of equip-
ment near an MRI.
Radio Frequency ShieldingUnlike magnetic shielding, Radio Fre-
quency (RF) shielding is required for
the majority of MRI applications and
consists of a highly conductive materialsuch as copper, aluminum, or galvanized
steel surrounding the MRI system. De-
termining RF shielding requirements is
fairly simple due to the fact that all RF
shields consist of a six sided structure
and the level of attenuation or shield-
ing effectiveness is determined by the
eld strength of the MRI system to be
installed. For example, MRI systems
with eld strengths of 1.5T or less re-
quire an RF shield that provides 100 dB
of attenuation at 100 MHz while 3.0T
MRI systems require a shielding system
that provides a 100 dB of attenuation at
150 MHz. As the eld strength of the
MRI systems increase, the performance
requirements of the RF shielding will
typically increase as well.
Special consideration must be giv-
en to all items that penetrate the exte-
rior of the RF shield into the MRI suite.Specialty shielded penetrations must be
provided to maintain the RF shielding
integrity. For example, HVAC must pass
through HVAC wave guides and plumb-
ing for re sprinklers, if required, must
also pass through a wave guide. Wave
guides are designed to specically limit
the frequencies that could pass through
the shielding. Wave guides designed for
a 100 MHz RF enclosure, for example,
will not allow frequencies below 100
MHz to pass through the shielding assome of those frequencies could inter-
fere with the MRI system. Additionally,
all electrical items including power for
outlets, lighting and specic sensors
required for an MRI installation, must
pass through electrical lters. Most
electrical lters are low pass lters that
only allow signals below a “cutoff” fre-
quency to pass through. For example, a
lter with a 1 MHz cutoff will only al-
low signals at frequencies below 1 MHz
to pass though the lter into the MRIsuite via the wiring of the lter.
Additionally, special attention
must be given to the ground isolation
of the room. This is a critical, but of-
ten overlooked, consideration when site
planning. The RF enclosure, when rst
constructed, should achieve a minimum
of 1000 ohms of ground isolation. This
is intended to prevent ground loops and
other issues, including less than optimal
MRI images, which will occur due to
poor grounding of an enclosure.In addition to specialty shielded
penetrations, other important aspects of
the shielding that should be considered
are the doors and windows. The doors
and windows are items that the MRI end
users will see and use every day. There-
fore, functionality should be considered
a top priority when designing and speci-
fying products for a site. Typical MRI
doors, as an example, use friction to cre-
ate the RF seal. This works well for cre-
ating the RF seal at the door, but can be
quite difcult to operate as these doors
require a reasonable amount of force to
open and close. This raises other con-
cerns and issues. These doors require
maintenance and repair of the door’s
RF ngers to maintain performance.
Without proper maintenance, the door
can become unreliable. There are doors
available that provide mechanical sealsto create the RF seal at the door with
simple “push button” entry and egress
operation. These doors have the look
and feel of a standard door adding to
user comfort while the ease of opera-
tion expedites patient throughput. It is
important to keep in mind windows and
doors will often have limited acoustic
performance. Careful attention should
be used when specifying doors and win-
dows if acoustics are a concern.
Another signicant issue to consid-er is the oor. All RF shielding systems
have a oor that is between 5/8” and
1-1/2” thick. In some applications, de-
pending upon the vibration and acoustic
requirements, the oors may be thicker
than 1-1/2”. This can create an Ameri-
cans with Disabilities Act (ADA) issue
requiring a ramp into the room; this may
be inconvenient and create logistical is-
sues for moving patients. Therefore,
when designing a new space, a oor de-
pression for the MRI suite should be in-cluded. This will allow the RF shielding
vendor to install the RF shielding sys-
tem with a at threshold or a threshold
that meets ADA requirements, eliminat-
ing the need for a ramp into the room.
For existing buildings, concrete should
be removed to depress the slab.
Radiation ShieldingImaging procedures that use ionizing
radiation pose a health risk to the clini-
cal staff and patients as well as to mem- bers of the public in spaces surrounding
the imaging suite. Unlike the patient,
who derives a medical benet from the
radiation used in the procedure, these
individuals must be carefully protected
from exposure. Medical radiation comes
from two types of sources: X-ray tubes,
such as in CT scanners, radiographic
rooms, and uoroscopy suites, and ra-
dioactive materials, which are used in
procedures such as nuclear medicine,
ETS-Lindgren’s Auto-Seal™ II Door offers me-
chanical seals and a push-button operation.
7/29/2019 Site Planning for Medical centres
http://slidepdf.com/reader/full/site-planning-for-medical-centres 4/6
DOTmedbusiness news I january 2010 45
single-photon emission computed to-
mography (SPECT), and positron emis-
sion tomography (PET). There are also
hybrid imaging systems such as PET/CT and SPECT/CT which utilize both a
radioactive source and an X-ray source.
All sources of ionizing radiation
are generally shielded the same way –
with layers of lead sheeting applied to
the existing structural barriers. Other
materials may be used, such as concrete,
steel, or gypsum wallboard. A radiation
physicist, such as a medical physicist
or health physicist, should be consulted
to determine these shielding require-
ments; in many states this consultation
is required by law. There are some im-
portant differences in the design ap-
proach to facilities using X-ray systemsand those using radioactive materials.
Given the additional cost and weight
associated with lead-shielded building
systems, facility designers should take
prudent steps to reduce the amount of
lead needed.
For X-ray imaging systems, the ef-
fective radiation source is the machine
itself. Figure 3 shows a typical layout
while Figure 4 shows a better, more ef-
fective layout. Lead shielding require-
ments can be mitigated by placing
equipment in larger rooms, effectively
increasing the distance between the
source of radiation and other occupied
spaces. Care should be taken when lo-
cating the wall or chest bucky (the de-
vice used to hold lm cassettes for chest
X-rays taken in a standing position) in
a radiographic room, since X-rays will
strike this wall directly. In addition toincreasing the distance, rooms adjacent
to X-ray imaging systems of all kinds
should be chosen for low-occupancy
uses – storage rooms, janitor’s closets,
restrooms, and outdoor areas are all
good choices to situate next to X-ray
and CT rooms. Areas that have high
occupancy, such as ofces, nurses’ sta-
tions, and so forth will require more
shielding if placed in close proximity to
an X-ray source.
In imaging operations using ra-dioactive material, the patient actually
becomes the radiation source once in-
jected with the radiopharmaceutical.
This requires a different approach, as
the source of radiation moves, occu-
pies, and exposes different areas of the
human body. For SPECT and general
nuclear medicine imaging, this is less
of a concern because the emitted radia-
tion is not very penetrating. However,
for PET and PET/CT suites, the site
planner should consult closely with theradiation physicist at the layout stage.
For these suites, the PET/CT scan room
is typically not the most critical shield-
ing design element. Rather, the “up-
take” or “quiet” rooms, where the pa-
tients wait for 30-60 minutes between
the radiopharmaceutical injection and
the scanning procedure, are the major
shielding concern. In a facility not laid
out with this in mind, the uptake room
walls can require several inches of lead
thickness to provide adequate shieldingto an adjacent ofce or waiting room.
Of course, for a PET/CT suite, the scan
room must be analyzed and shielded as
a CT scanner room in addition to the
shielding required due to the PET radio-
active material.
In either case, care should be taken
to congure the imaging suite so that
the imaging technologist will have full
control to prevent members of the pub-
lic from inadvertently coming into con-
Figure 3
Figure 4
7/29/2019 Site Planning for Medical centres
http://slidepdf.com/reader/full/site-planning-for-medical-centres 5/6
DOTmedbusiness news I january 2010 www.dotmed.com46
tact with sources of radiation. This can
mean simply conguring the direction
of a door swing so that a technologist
will have an unobstructed view of a
doorway when seated at his or her work-
station, or a more difcult approach in
providing for proper security of areas
where radioactive materials are stored.
AcousticsAcoustics is an important issue that
should be investigated when designing a
site for medical imaging systems. Unlike
EMI and vibration concerns, acoustic
issues are often created by the imaging
equipment itself. In particular, MRI sys-
tems can be quite disruptive to the sur-
rounding environment. MRI systems can
create airborne noise, which is the propa-
gation of acoustic noise through the air,
and structure-borne noise, which is the
propagation of acoustic noise through the building structure. As a result, acoustic
solutions typically need to address both
airborne and structure-borne noise in or-
der to be effective. Typical airborne
acoustic solutions involve detailed
wall, ceiling and oor construction
to meet predetermined acoustic cri-
teria. The solution should also de-
tail how penetrations, HVAC ducts,
and gaps in construction around the
imaging suite should be treated.
Recommended wall and ceilingconstruction usually involves some
combination of gypsum board, stud
placement, sound batt insulation,
air gaps, and isolation clips to ad-
dress both the transmission and re-
ection of acoustic noise.
Structure-borne acoustic
solutions can be more complicated.
Typically, a structure-borne solution in-
volves some combination of weight and
isolation material. For example, with
proper site planning, a vibration slab
that is isolated from the surrounding
structure and is placed on spring isola-
tors could also be a cost-effective solu-
tion for structure-borne acoustic noise.
Performance ConcernsElectromagnetic Interference (EMI)
Medical imaging systems can be sen-
sitive to electromagnetic interference
(EMI). EMI may be caused by electri-
cal equipment, motors, moving metal
objects such as cars, trucks and eleva-
tors, and by transportation systems that
run on electrical power such as subways
and trains. Proper site layout based upon
an awareness of the impact this existing
equipment may have on the medical im-
aging system can avoid many of these
issues. For example, avoiding the place-
ment of EMI sensitive imaging equip-
ment near electrical rooms, large trans-formers and motors, parking garages,
roadways, and elevators can aid in the
prevention of potential EMI issues (as
well as the signicant vibrations associ-
ated with many of these elements).
Unfortunately, EMI issues cannot
always be avoided especially in large
urban areas where an owner may have
limited placement options. This does not
mean that imaging equipment cannot be
installed in an EMI environment that
exceeds manufacturer specications. Infact, there are solutions to EMI issues.
EMI shielding consultants are available
that can survey existing buildings to
quantify the EMI environment to deter-
mine if a facility meets the equipment
EMI criteria. If the building is under
construction, a consultant can approxi-
mate the source(s) of EMI based upon
the electrical layouts and the proximityof EMI sensitive equipment to items
such as moving metal found in elevators
or subways or electrical sources. Ideally,
addressing performance considerations
prior to or during the construction pe-
riod results in the most cost effective
solution for optimal performance of the
medical imaging equipment.
EMI issues also may be endemic to
the facility itself in the case of a retro-
t of an MRI suite. MRI scanners can
permanently magnetize steel and other
ferromagnetic components within the
building structure, and the resulting
magnetic eld can cause EMI within the
room after the MRI system is removed.
Steel shielding from the old MRI sys-
tem, steel beams, and corrugated metal
deck and rebar inside concrete oors
and ceilings are common culprits when
this type of interference exists.In the event that a site does not or
will not meet the manufacturer’s re-
quirements for EMI, a shielding con-
sultant could propose shielding options
that will reduce the EMI in the area
around EMI sensitive equipment. These
solutions typically come in two forms;
passive shielding and active shielding.
Passive shielding involves using
materials with magnetic shielding prop-
erties. Common materials that are used
for passive shielding are aluminum, sili-con steel, and low carbon steel. Passive
shielding can be very effective in resolv-
ing 60 Hz and higher frequency EMI
issues, but has limited effectiveness
in resolving lower frequency distur-
bances that can be created by mov-
ing metal objects such as subways
and trains that operate on DC elec-
tric power. Passive shielding also
requires no maintenance after instal-
lation and can often be the most cost
effective solution for higher frequen-cy (frequencies of 50 Hz and greater)
EMI issues. However, this solution is
dependent upon the level of the EMI
and the amount of material required
to resolve the issue.
Active shielding, on the other
hand, utilizes electronics and coil sys-
tems to create a cancelling magnetic
eld over a predetermined volume. Ac-
tive shielding, unlike passive shielding,
can be very effective at lower frequen-
cies (frequencies below 100 Hz). Thereare several benets associated with ac-
tive shielding solutions. First, an active
system has the ability to respond to a
changing environment. Therefore, if
the EMI environment becomes worse,
a well designed active system will be
able to respond to those changes and
maintain an EMI environment that
meets manufacturer specications for
EMI sensitive equipment. Second, an
active system requires considerably
Transportation systems such as subway trains can
cause electromagnetic interference with medical
imaging systems.
7/29/2019 Site Planning for Medical centres
http://slidepdf.com/reader/full/site-planning-for-medical-centres 6/6
DOTmedbusiness news I january 2010 47
less construction. An active system
typically requires the installation of
coils unlike passive shielding, which
often requires the installation of a six
sided structure that requires interior
nishes. Additionally, since passive
shielding is a much less effective solu-
tion for lower frequency applications,
active shielding is often a more cost
effective solution to EMI issues gener-
ated by subways and moving metal. In
the event that a particular site does or will experience EMI issues, solutions
should be evaluated based upon the ef-
fectiveness of the solution and the cost
to implement the solution.
VibrationImaging equipment, particularly CT
and MRI scanners, can be impacted by
building vibrations. Fortunately, most
vibration concerns can be addressed
through careful design and construc-
tion of the area to house sensitive piecesof equipment. In existing buildings, it
is critical to perform a vibration sur-
vey. A survey will quickly determine
whether the existing structure meets
the vibration requirements of the equip-
ment being sited. If the equipment does
not meet the vibration criteria, a good
survey will analyze the cause of the
vibration and provide general recom-
mendations to resolve vibration issues.
Many vibration issues can be resolved
through inexpensive solutions. For ex-
ample, a nearby mechanical room may
not include isolation pads or isolators
for mechanical equipment that induce
vibrations into the structure. This may
simply require placing vibration isola-
tors under such equipment. It may also
be possible to use vibration isolators on
some imaging equipment such as MRIs,
which would allow for the attenuation
of vibrations and decoupling of the
equipment from the structure due to the benets of the isolators. However, in
some situations, it may be necessary to
stiffen the structure under the imaging
equipment or build an isolated slab to
meet the equipment’s vibration criteria.
The unfortunate aspect is that isolated
slabs and modications to the structure
can be quite expensive. However, such
a building retrot is much less costly to
perform during the process of outtting
a space and installing imaging equip-
ment. Performing a retrot after imag-ing equipment has been installed and
found to be adversely affected by vibra-
tions can be extremely expensive.
In the case of new construction,
a vibration consultant can aid in the
development of the site by reviewing
equipment specications and site plans
to determine the vibration response of
the structure due to vibration sources
such as mechanical, other imaging
equipment, and trafc in the building.
Further, a vibration consultant can aid
in the development of isolated slabs as
necessary, which is much less expensive
to implement at the time of construc-
tion and can have the added benet of
providing structure-borne sound attenu-
ation due to high noise levels that are
produced by some pieces of imaging
equipment, in particular MRI systems.
SummaryInformed site planning when designing
space for medical imaging equipment,
especially prior to construction, results in
a cost effective and successful site – one
that is completed on time and on budget.
An experienced shielding design team
with an experienced acoustic and/or vi-
bration consultant will aid the user in
this often complex process. The upfront
investment in an experienced consultant
for site planning will result in construc-tion and operational savings that more
than return the initial investment. To pro-
tect your investment, ask any consultant
or rm you may hire for customer ref -
erences and a list of recently completed
projects. Follow up on the references and
visit the project if prossible to assess the
capability of the consultant and value-
added services provided to the customer.
AcknowledgementThe authors would like to acknowledgeand thank Rick Ottolino, AIA, of Ottoli-
no Winters Huebner for his invaluable
contributions to and review of this article.
About the AuthorsJoel Kellogg is Manager of Technical
Engineering and Consulting with ETS-
Lindgren in Glendale Height, IL. His ex-
perience includes design of MRI suites to
meet various acoustic, vibration, EMI, and
radio frequency shielding requirements.
He may be reached at 630-307-7200 or [email protected]
Dave Jordan is Senior Medical
Physicist with West Physics Consulting
in Atlanta, GA. His experience includes
acceptance, accreditation, and safety
evaluations of medical imaging systems
and design of radiation shielding. He
may be reached at (770) 435-9186 or
• Online: dotmed.com/dm11080
ETS-Lindgren uses isola-
tors under the magnet to
perform vibration surveys