Publication No. J39897 2018-10-26 Supersedes J39897 2018-10-11 Site Specifications for the VITROS ® 5,1 FS Chemistry System and VITROS ® 4600 Chemistry System
Transcript

Publication No. J398972018-10-26

Supersedes J398972018-10-11

Site Specificationsfor the

VITROS® 5,1 FS Chemistry Systemand VITROS® 4600 Chemistry System

PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the subject matter gained by Ortho Clinical Diagnostics (Ortho) prior to publication.No patent license is granted by this information.Ortho reserves the right to change this information without notice, and makes no warranty, expressed or implied, with respect to this information. Ortho shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by the negligence or other fault of Ortho.

This equipment includes parts and assemblies sensitive to damage from electrostaticdischarge. Use caution to prevent damage during all service procedures.

Description Page

Table of Contents

Electrical Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Electrical Specifications for the 5,1 and 4600 Systems . . . . . . . . . . . . . . . . . . . . . . . . . 4Electrical Specifications for the Supplied PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . . 5Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6RECEPTACLE Location Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Networking Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Requirements at Customer Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Shipping Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Specifications for the System and the PRINTER. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Specifications for the 5,1 and 4600 HEAT REJECTION PLENUM . . . . . . . . . . . . . . 13

2 2018-10-26 – J39897

Electrical Power Requirements

Section 1: Electrical Power Requirements

AC Power QualityThe VITROS® 5,1 FS Chemistry System (hereafter referred to as the 5,1) and the VITROS® 4600 Chemistry System (hereafter referred to as the 4600) are designed to perform satisfactorily with reasonable disturbances and voltage tolerance to the ac power. The quality and reliability of the power source is often dependent on the geographic location of the site. Some sites, where power quality is exceptionally poor, may require special power conditioning equipment to ensure that the tolerances are maintained within acceptable limits. Some models of UNINTERRUPTIBLE POWER SUPPLIES (UPS) are recommended as optional equipment for use with VITROS® Systems, and may be purchased from the Distribution Center. For additional information, see the UPS User Guide located at: http://apps.orthoclinical.com/notes/contentpages/techdoc.aspx?view=USOTHER&culture=en-us

Static ElectricityThe System should be adequately grounded to prevent the effects of static charges generated. Static problems can also be reduced by ensuring that the relative humidity of the room is maintained above 30 percent. Special anti-static floor pads can be purchased and positioned where the operator is required to make contact with the instrument.

Electromagnetic InterferenceSeveral types of electrical interference may be indigenous to the site location and may require special filtering to prevent malfunctions in the 5,1 or 4600 systems. If interference is suspected to be a problem, shielding may be required, or filtering of the ac line power to the site may be necessary.

Cellular phones, two-way pagers, and other RF transmitting devices should not be used within 1 m of the 5,1 or 4600 Systems.

2018-10-26 – J39897 3

Section 2: Electrical Specifications for the 5,1 and 4600 Systems

* When measuring supply voltage: • Use a MULTIMETER. • Check a number of times and use the average of the values.• Wait a minimum of 5 minutes between checks. • The difference between the values should not be more than 6%.

ImportantIf an optional UPS is to be installed, make certain that the correct receptacle is installed in the lab.

** or an equivalent to meet local electrical codes

NoteCEE7 “Schuko” plug may be supplied, either straight or right angle.

Input Voltage* 200 to 240 VInput Frequency Limits 50 - 60 Hz 2 HzPower Operating (Normal) 1225 WattsMaximum Current 15APOWER CORD 4 m (13 ft) PLUG Type North America IEC 320 to NEMA L6-20P Continental Europe IEC 320 to CEE7 “Schuko” Other As required by local electrical codesCIRCUIT Type Dedicated with Isolated GroundRECEPTACLE Type North America NEMA L6-20R Continental Europe CEE7 “Schuko” Other As required by local electrical codes

UPS Rating Amps Receptacle Type3 KVA 20 A L6-20R**5 KVA 30 A L6-30R**

208 V Check between Voltage 60 Hz L1 - L2 190 - 220 V AC

L1 - SAFETY GROUND 113 - 127 V ACL2 - SAFETY GROUND 113 - 127 V AC

240 V Check between Voltage 60 Hz L1 - L2 226 - 254 V AC

L1 - SAFETY GROUND 113 - 127 V AC L2 - SAFETY GROUND 113 - 137 V AC

4 2018-10-26 – J39897

Electrical Specifications for the Supplied PRINTER

Section 3: Electrical Specifications for the Supplied PRINTER

ImportantThe following information is only applicable to PRINTERS supplied by Ortho. If the PRINTER is being supplied by the customer, refer to the manufacturer’s PRINTER manual for the electrical specifications.

Detach 1

1 RECEPTACLE required

3-wire, 120 V AC, polarized, SAFETY GROUND, maximum distance of 182.7 cm (72 in.) from each side of the system.

CIRCUIT Type Dedicated120 V Check between Voltage 60 Hz N - L1 102 - 132 V AC 15 amperes L1 - SAFETY GROUND 102 - 132 V AC

N - SAFETY GROUND Less than 1 V AC

E002_1331ACE002_1331ACA

RECEPTACLEpolarized power

GROUNDSAFETY

L1

N

NEMA 5-15R

2018-10-26 – J39897 5

Section 4: Electrical Requirements

Tension 200 to 240 VoltsIntensity 15.0 Amps MaximumFrequency 50 / 60 HzConsumption (nominal) 3.0 kVA Max Power 3000 WNumber of socket-outlet 1, plus 1 for printerProtection against micro cuts (Momentary power interruption)

Yes

Maximum time Tolerant of missing one full cycle of 50 Hz (20 milliseconds)

System tolerance to electrical variation

180 to 264 VAC 48-62 Hz extreme

6 2018-10-26 – J39897

RECEPTACLE Location Specifications

Section 5: RECEPTACLE Location Specifications

Specifications for PRINTER POWER and NETWORK CONNECTION

Specifications for the System only

Detach 2

Reference Description Q (height

from floor) 30.5 cm (12 in.) minimum 61.0 cm (24 in.) maximum

S (printer power)

2.9 m (8 ft. 10 in.) maximum

T (Network)

5.3 m (17 ft. 6 in.) maximum

Reference Description Q (height

from floor) 30.5 cm (12 in.) minimum 61.0 cm (24 in.) maximum

R (system power)

1.98 m (6 ft. 6 in.) maximum

2018-10-26 – J39897 7

Section 6: Environmental Specifications

Detach 3

Maximum altitude: 2438 m (8000 ft.)Laboratories operating a VITROS® System at altitudes greater than the Ortho published site specification of 8000 ft (2438 m) are responsible for validating those systems in their respective laboratories. In addition, Ortho recommends following good laboratory practice and adhering to all required quality control and maintenance procedures to ensure result integrity.This product has been evaluated and is suitable only for Pollution Degree II and Insulation Category II.Additional Requirements for the VITROS® 5,1 and VITROS® 4600 HEAT REJECTION PLENUMIf the customer chooses to purchase and use the optional Heat Rejection Plenum, the EXHAUST SYSTEM and DUCT must be installed by the customer.

Acoustic Noise Level 61 db Thermal Characteristics On 4187 Btu/hr

SpecificationsAir Capacity7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute259 - 366 metres (850 - 1200 ft.) per minuteDUCT20.5 cm (8 in.) flexible AIR DUCT and attaching hardwareAir Flow - Ventilation System7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute at 5.1 mm (0.2 in.) of water vacuum

8 2018-10-26 – J39897

Networking Specifications

Section 7: Networking Specifications

The requirements for e-Connectivity network connection are located in the e-Connectivity® Network Connection Specifications and Network Form located on-line at: https://techdocs.orthoclinicaldiagnostics.com/notes/contentpages/techdoc.aspx?view=USOTHER&culture=en-usThe customer must complete the form and return it to Ortho-Clinical Diagnostics.

ImportantAll requirements for lab network connections must be completed before installation of the analyzer(s).

2018-10-26 – J39897 9

Section 8: Requirements at Customer Site

TELEPHONE A TELEPHONE located so that the user can move around the System must be available.

Air Flow No air flow within 3.0 m (10 ft) of the System. Entrance to the room where the System will be installed

Minimum width of 81.3 cm (32 in.)

LAB COMPUTER Interface See the Interface Specification section of the Operator’s Manual.

Network Connection Refer to the Networking Specifications on page 9.Floor Loading 322 kg/m2 (65 lb./ft2)

10 2018-10-26 – J39897

Shipping Specifications

Section 9: Shipping Specifications

* Minimum overhead clearance required for unpacking.Detach 4

Shipping SpecificationsReference Description

A 274.3 cm (108 in.)B 110.5 cm (43.5 in.)C 161.3 cm (63.5 in.)D 215.9 cm (85 in.) *

Weight 811.9 kg (1790 lb)

2018-10-26 – J39897 11

Section 10: Specifications for the System and the PRINTER

System Specifications Supplied PRINTER Specifications ****Reference Description Reference Description

A 233.7 cm (92 in.) * J 40.0 cm (15.7 in.)B 83.8 cm (33 in.) K 51.3 cm (20.2 in.)C 132.1 cm (52 in.) L 34 cm (13.4 in.)D 205.7 cm (81 in.) M 61 cm (24 in.)

Weight 612.4 kg (1350 lb) Weight 19.7 kg (43.5 lb.)* 264.7 cm (104.2 in.) if AT Accessory is installed. Port

Printer LanguageUSB Version (TBD)Post Script

Installation Specifications for the System only

Installation Specifications for the Supplied PRINTER ****

Reference Description Reference DescriptionE 76.2 cm (2 ft 6 in.) N 17.8 cm (7 in.)F 76.2 cm (2 ft 6 in.) O 45.7 cm (18 in.)G 30.5 cm (12 in.) ** P 61.0 cm (24 in.)H 4.3 m (14 ft. ) *** **** If the PRINTER is being supplied by the

customer, refer to the manufacturer’s PRINTER manual for any specifications.

Q 76.2 cm (30 in.)** 91 cm (36 in.) if SEISMIC ANCHORAGES, AT Accessory Kit if installed from the rear or otherwise 30.5 cm (12 in.) if installed from the front, or HEAT REJECTION PLENUM are to be installed.*** 4.6 m (15 ft.) if AT Accessory is installed.

12 2018-10-26 – J39897

Specifications for the 5,1 and 4600 HEAT REJECTION PLENUM

Section 11: Specifications for the 5,1 and 4600 HEAT REJECTION PLENUM

Detach 5

HEAT REJECTION PLENUMSpecifications

Reference DescriptionA 107.1 cm (42.2 in.)B 77.5 cm (30.5 in.)C 38.7 cm (15.2 in.) D 17.7 cm (7.0 in.)

2018-10-26 – J39897 13

Ortho-Clinical Diagnostics1500 Boulevard Sébastien BrantB.P. 3033567411 IllkirchCEDEX, France

Ortho-Clinical Diagnostics, Inc.100 Indigo Creek DriveRochester, NY 14626USA