+ All Categories
Home > Documents > SLAC TN 84-3 6920382

SLAC TN 84-3 6920382

Date post: 21-Feb-2018
Category:
Upload: mel
View: 220 times
Download: 0 times
Share this document with a friend

of 4

Transcript
  • 7/24/2019 SLAC TN 84-3 6920382

    1/4

    m-

    j SLA C-TH --84 -3 U ~

    ^

    *

    0 ^ DE84 011176

    L

    ' SLAC-TN-84-3

    / Jonuary , 1084

    C A B L E F A B M C A T I O N P R O C E D U R E S F O R T H E B PM S Y S T E M *

    W . B . S M IT H

    Stanford Linear Accelerator Center

    Stanford Univcnity, Stanford, California UfS S

    The B eam Position Monitor (BPM) System is

    t

    critical component or beam steering

    and control. It consists of the Beam Position Monitors (in the bcam linc), electronic

    signal processors (U> ed rermitely) and the ca bles tha t interconnect the two.

    Each BPM produces four signals. These four signals arc then compared electroni

    cally with each other. Differences in amplitude among these signals is converted into

    information about the position of the beam as it passed that particular BPM.

    In orer to provide the nee ded accuracy Tor this system, all four signals from

    each BPM must ar.ivc at the processing electronics simultaneously. This means the

    interconnecting cables must be matched very precisely. Th e following procedures,

    when followed precisely: will assure th e matching of BPM cables.

    TOLER ANCE STANDARDS : We use RG223/U cosxial cable for the BPM System.

    The m anufac turer specifies its propagation velocity is 1.54 nS/ft. For the BPM elec

    tronics to operate accurately, each of the four signals from a single BPM must arrive

    at the electronics within 0.5 uS. And since cable runs range from 100 to 300 feet, this

    requires an accuracy of 3.0 inches or 0.3% (or better)

    THE PROBLEM : Conventional cable measurement/cutt ing techniques produce typ

    ical accuracies on the order of 1 , far too coarse for the BFM System. Hence, the

    more sophisticated technique that follows.

    THE TIME DOM,UN REFLECTOMETER (TDK): This

    Ls

    a marvclously sensitive

    instrument. When properly calibrated, it is also extremely accurate. Phase I describes

    '.he procedure to be followed to calib rate the instrum ent. All subsequent work is highly

    dependent upon the "arc you give this calibration.

    * Work supported by the Department of Energy, contract DE-ACOM6SFO0515.

    PORT IONS OF THIS nEP OR T ARE ILLEGIBLE. I t

    h i s b een r ep ro d u ced f r o m t h * .b t r t v i l r t f c

    co p y t o p e rm i t th e b r o i d r t p * * i rb H W i l l -

    abi l i ty .

    P H A S E I , T D R C A L I B R A T I O N

    s t e

    P 1

    :

    Interconnect the various instrument* as shown on the attach ed Test System

    Interconnection Diagram.

    s t r

    P 8= Turn on all electronic equipment. Allow at least Jhom for warm-up, longer

    if at

    all

    possible.

    Step 3 : Measure PRECISELY and cut one RGS23/U cable. Make its length equal

    to or greater than the maximum length of the sets of cables to be made.

    Actual cable length should be selected in orJer to facilitate fabrication a nd

    operator preference. Terminate one end or the cable with SMA connector,

    Leave the other end open.

    S

    "> P * Repeat Step 3, only this time make the cable length EXACTLY one-hail

    the length of the Grst cable. PRECISION is the key to Strjs t, 3 and 4.

    Any errors here will lead to greater inaccuracies in the BPM cable* to b e

    fabricated. Th e cables made in Step 2 and Step 4, along with a ' lero-

    length" (i.e. no cable) comprise a ibrce-point calibration system for t i e

    TDR

    Step 5 : Set the SPDT coaxial switch "A' tc T D R Mode". Set the TDR " V/D IV

    dial to 0.05. Set the DP-IT uax iei sw itch *B* to position 1 . Set "Ri rc-

    t ion" to "VAR" niode. Set "SCAN" to Normal". Set 'PERSIST ENCE "

    to -NORMA L" mode. S "DISPLAY" to 'INTE RNA L". Set "MAGNI

    FIER" to "XI" . And set 'MARKE R P0ST1ON (Div) to TIM ES 100"

    and expand to 1 ft /div.

    S tep 6 : F indtheleadingrdgeor tbereBectcdpi i l seoatheTDR'fCRTd> i t> lay . Tki t

    wil l appear as in abrupt transition ia the trace from a hotiro ttaltu t to oe

    tha t rises steeply. We will use the "knee" where this IrUM lk* takes place

    aa the standard measuring point for all subsequent measurements with the

    TDR . The waveform will M t l ike this:

    M ' Leoaing edgt 5

    / (ejected pjite

    > K n

    2

  • 7/24/2019 SLAC TN 84-3 6920382

    2/4

    Step 7 : Having found the reflected pulse, set the "knee" on the X-AxU0 graticule,

    using 4 combination of the marker dial tod the coarse and fine Horizontal

    Position adjustments. When properly done, the "knee" will ait squarely on

    the 0% graticule, the Marker Dial will read "000" and Switch "B" will be

    in position "1".

    Step 8 : Set the Marker Dial to the length of the longest cable (or to its nanosecond

    equivalent, if in the time m ode). Set Switch

    fc

    B" to position "2". Use a

    jeweler's screwdriver to adjust *VAR" so that the new "knee" falls exactly

    on the0 graticuje.

    Step 9 The adjustm ents you made in steps 7 and 8 affect each other. So repeat

    steps 7 and 8 until you can step from one to the other without having to

    adjust either one.

    Step 1 0: Set Switch "B" to position "3" and adjust the Marker Dial for the length

    of the shorter cable (or its nanosecond equivalent if in the time mode). The

    new "knee

    9

    should be very close to the 0% graticule. If not, combine Steps

    C and 10 and "rock" the calibration to the best compromise you can find,

    ^ou will find that the accuracy of your cable cutting will have a major

    impact on the ease with which you are able to accomplish this calibration

    procedure.

    The calibration procedure for the TDR b now complete.

    ** WARNING ** WARN ING ** WARNING * WARN ING _**

    IT IS VERY EASY TO BUMP OTHER ADJUSTMENTS WHILE MAKING AD

    JUSTMENTS TO THE TDR OR WHILE WORKING WITH THE CABLES. EVEN

    THE SLIGHTEST BUMP WILL DESTROY T HE ACCURACY. IT IS fMP_ERATJrVE

    THAT YOU REPEAT THIS CALIBRATION PRODECURE ANYTIME YOU SUS-

    PECT TH E TDR'S SETTINGS HAVE BEEN DISTURBED IN ANY WAY

    WARNING " WARNING " WARNING " WARNING * "

    THIS CALIBRATION PROCEDURE SHALL BE PERFORMED AT THE BEGIN

    NING OF EACH DAYS WORK AND BEFORE BEGINNING WORK ON EACH

    NEW SET OF FOUR BPM CABLES TO BE TRIMMED.

    DISCLAIMER

    mfMmmm

    4

  • 7/24/2019 SLAC TN 84-3 6920382

    3/4

    PHASE n, CABLE TRIMMING

    Step 1: Cut four cables. Make each two to four feet longer than the length pre-

    stribed. Terminated one end of each cable with

    AH

    SMA connector (see

    attached trim code charts for details).

    Step 2

    As in the calibration procedure (Phase I), Switch

    "B"

    should have no cable

    attached to position "1", the shorter of the tw o calibration cables ( the

    length of the longer calibration cable) to position "3". Attach one of the

    cables to be trimmed to position "4".

    Step 3 : Confirm TDR calibration by switching Switch "B" through the Grst three

    positions.

    Step

    4'.

    Set the vernier scale to the desired length (length mode) or delay (time

    mode). Turn Switch "B" through the Grst three positions.

    Step 5 : Carefully "prune" the unterminated end of cable # 4 until the reflected

    pulse's "knee" falls within ONE ( ) MINOR DIVISION of the 0 graticule

    of the TDR's display.

    Step 6: Disconnect the two calibration cables. Connect the three remaining cnbles

    of the set of four positions. * T , "2" and "3".

    Step 7 j Repeat Step 5 for cables 1, 2 and 3.

    Step 8 : Visually compare the four cables with the TDR by switching "B" through

    the four positions. NO TW O CABLES SHOULD DIFFER BY MORE

    THAN 1.5 MINOR DIVISIONS ON THE TDR GRATICULE. ALWAYS

    TRTM CABLES # 1 , # 2 or # 3 as needed to make this match. Cable # i a

    your standard now, and should NOT BE TRIMMED.

    Step 0

    ;

    Now that the cables have been carefully trimm ed, terminate the untcrmt*

    natcd end with the appropriate connector (SMA or SHV) according to the

    attached trim code charts.

    NO TE. : If difficulty is experienced IT the termination procedure and a cable must be

    ret rimmed in order to be properly terminated, repeat Step 8. If the cable

    still mcetr the standard in Step 8, it is acceptable. If not, then a new cable

    must be fabricated.

    S

    PHASE m, ATTENUATION MATCHING

    Mismatches in attenuation ca n not be remedied. Thus any cable of a four-cable

    aet that can not meet attenuation tolcraaces must be discarded from the set, and a

    DCW cable fabricated to take its place. However, a longer cable discard might prove

    satisfactory (once properly trimmed to the new length)

    at>

    a substitute for a mismatched

    cable of a ahorter set. Hence it is recommended that the longest sets of cable* be t t* 4

    for attenuation matching first.

    ATTENUATION STANDA RD: Each cable of a four-cable set must be within

    1

    of

    each of the other three.

    Step 1: Set Switch "B" to position "I". Connect cable #1 to Kaput #1 . Now

    connect the free end of cable #1 to ouput # 1 . Observe the TDK display.

    The reflected pulse shouM disappear, confirming proper connection.

    Step 2 : Set Switch "B" to position " 2 \ Connect cable #2 to Input # 2 and output

    # 2 , in much the same fashion aa in Step 1. Repeat the procedure for cable*

    # 3 and #4 . Be sure to use the TDR display to confirm i roper cooaectton

    in each case.

    Step 31 Set SPDT coaxial Switch *A" to *Rw" pofition.

    Step 4: Using the oscilloscope, adjust the DELAY GATE GENERATOR (DGG)

    that its output overlaps the leading aegatire pub* of li e PULSE

    STRETCHER.

    Step 5 : Maximize this overlap by maximizing the DIGITAL VOLTMETER (DVM,

    reading.

    Step 0: Step through the four positions of Switch

    v

    tl*. citing the DVM value* far

    each position.

    Step 7: Compute the deviation between the maxima and minima of the cable aet.

    Reject any cable which eihibiti a deviation greater than A 0 535 of the

    mean.

    EX AM PLE : The DVM readings for a wt of cables arc: 610, 611, 600 u d M

    (

    respectively. The mean would be: (610 -- 611 4* 600 + 6 M) /4 or 3432/4 *

    0.5%of 608 3.04.

    6

  • 7/24/2019 SLAC TN 84-3 6920382

    4/4

    Hence allowable attenuation for this yet would have to be within the span from

    604.06 to 611.04. O bviously, cables # 1 , #2 , and #3 meet this standard. But at a

    value of 602, cable # 4 must be rejected.

    Carrying this exam ple further, the replacement for cable # 4 produce s a DVM

    reading of 614, A new m ean must be calculated (611) and the acceptable 0 .5 % span

    calculated (3.055 ). Tbis new span is 607.045 to 014.055. Now all four cables arc

    acceptable, although Cable #4 just barely makes it.

    TRIM CODE CHART

    S MA C O N N E C T O R :

    S HV C O N N E C T O R :

    A - t f

    ii

    A

    *>'

    R ' "

    n _ 22"

    O 3 2

    ^ 32

    Use crimp die

    L

    Use coDDector body

    KTN 2001

    ii

    '

    i

    1

    King s # 170S-2 with

    the sleeve from a

    Kings # 1705-1.

    Crimp first with a

    111

    \

    B

    KTN 2002 die and

    again with a

    T

    c

    \

    K TN 1 81 die.


Recommended