Slit Lamp Illumination Techniques
Author: Irina Jagiloviča
e-mail: [email protected]
www: www.optometristiem.lv
Page 2
Diffuse Illumination
Light is spread evenly over the entire observed surface
The beam is opened all the way.
Direct the light onto the eye at a 45 degree angle
The microscope is directed straight ahead.
Observe: eyelids, lashes, conjunctiva, sclera, pattern of redness, iris, pupil, gross pathology, and media opacities
Page 3
Diffuse Illumination
Page 4
Diffuse Illumination
Page 5
Direct Illumination Techniques
Beam The microscope is usually directed straight ahead but may also be moved to an
angle opposite the illuminator. The greater the angle between the illuminator and the microscope, the greater the
width of the illuminated section. A very narrow beam (optical section) directed onto the cornea can be used to
evaluate corneal shape, elevation, and thickness. Observe: cornea, iris, lens, vitreous
Page 6
Beam
Page 7
Beam
Page 8
Tangential Illumination
This technique is used to observe surface texture. Medium-wide beam of moderate height Swing the slit lamp arm to the side at an oblique angle Magnifications of 10X, 16X, or 25X are used Observe: anterior and posterior cornea, iris, anterior lens (especially useful for
viewing pseudoexfoliation)
Page 9
Tangential Illumination
Page 10
Specular Reflection
Specular reflection is used to visualize the integrity of the corneal and lens surfaces. If the surface is smooth, the reflection will be smooth and regular; if the surface is broken or rough
Position the illuminator about 30 degrees to one side and the microscope 30 degrees to the other side
To visualize the endothelium, start with lower magnification (10X to 16X). Direct a relatively narrow beam onto the cornea
Switch to the highest magnification available.
Endothelium is best viewed using only one ocular, so you may want to close one eye.
Observe: corneal epithelium and endothelium, endothelial mosaic, lens surfaces
Page 11
Specular Reflection
Page 12
Specular Reflection
Page 13
Indirect Illumination
Proximal
Use a short, fairly narrow slit beam.
Place the beam at the border of the structure or pathology
Observe: corneal opacities (infiltrates, vessels, foreign bodies)
Page 14
Proximal
Page 15
Sclerotic Scatter
A tall, wide beam is directed onto the limbal area.
When the light is properly aligned with regard to the eye, a ring of light will appear around the cornea.
The light is absorbed and scattered through the cornea
highlighting pathology.
Use 10X magnification, with the microscope directed straight ahead
Observe: general pattern of corneal opacities
Page 16
Sclerotic Scatter
Page 17
Sclerotic Scatter
Page 18
Sclerotic Scatter
Page 19
Retroillumination
Retroillumination is used to evaluate the optical qualities of a structure.
The light strikes the object of interest from a point behind the object and is then reflected back to the observer
Page 20
Retroillumination
Page 21
Indirect Retroillumination from the Iris
the beam is directed to an area of the iris bordering the portion of the iris behind the pathology
This provides a dark background, allowing corneal opacities to be viewed with more contrast
Page 22
Retroillumination from the Fundus (Red Reflex)
The slit beam at 2 to 4 degrees
Shorten the beam to the height of the pupil to avoid reflecting the bright light off of the iris.
Focus the microscope directly on the pathology using 10X to 16X magnification. Opacities will appear in silhouette.
This view is best accomplished if the pupil is dilated.
Page 23
Retroillumination From the Fundus (Red Reflex)
Page 24
Van Herrick Technique
Use to evaluate anterior chamber angle without gonioscopy Medium magnification Angle 60 degrees Narrow beam close to limbus
Depth of anterior chamber is evaluated it to the thickness of cornea:
4. grade – open anterior chamber angle 1:1 ratio
3. grade – open anterior chamber angle 1:2 ratio
2. grade – narrow anterior chamber angle1:4 ratio
1. grade – risky narrow anterior chamber angle less than 1:4 ratio
0. grade – closed anterior chamber , cornea “sits” on iris
Page 25
Van Herrick Technique
Page 26
Van Herrick Technique
1. grade – risky narrow anterior chamber angle less than 1:4 ratio
Page 27