December 9, 2016 1
Solutions pour l’analyse de routine des petites molécules
Fabrice Moriaud, Ph.D. - Applications Development - Fällanden
30ème Réunion d’Utilisateurs RMN Bruker
At university …
… Academia benefits from
• Free TopSpin 3.5 processing SW!
• Download from website
At the spectrometer …
… In next topspin 3.5 release, an acquisition license allows to run:
• Free CMC-assist 2.5
• Free Fastlane, SmartDriveNMR, qNMR
Brian runs an NMR lab
Brian
• … Works in a hypothetical pharma company / university
• … Runs the NMR service lab
• … Is the “resident NMR expert” in the organization
• … Brian has a solid NMR background
• … He needs help to make the right choices for numerous experiments
Brian needs to select for Kevin, the optimal experiments for QC or Structure verification of diverse samples
Numerous and diverse sample: concentrations, structural complexity, spectra complexity, water suppression in some DMSO samples
Brian thought about handling these numerous sample at night or week-end, without coming to a smart solution:
• Running all samples with a complete set of experiments to ensure optimal data (enough scans, HSQC on everything, maybe even auto water suppression…). That would take the spectrometer time for the other tasks
• Run the 1D’s and let Kevin figure it out on Monday (knowing that several spectra will need to be re-run )
• … Or check back on the progress every few hours, look at the results and setup follow-up experiments where needed. Brian is not at the lab at night and at week-end!
A week in the life of Brian
SmartDriveNMR: – makes life easier for measuring and analysing
Questions asked to the spectrometer are delegated to SmartDriveNMR : • Measure a routine 1H spectrum with optimized parameter sets: number of scans, and
suppressions.
• Measure a set of experiments with optimized information and acquisition time
SmartDriveNMR decision-making algorithm
• Step by step decision for further experiments based on both structure and already acquired spectra: 1H, 1H suppressed, more scans, edited HSQC, 13C, HMBC.
• Saves spectrometer time by automatically running only these experiments that add the most value to solve a given problem – using problem-tailored parameters.
Automatic spectra interpretation done by CMC-assist algorithm
SmartDriveNMR: decides on-the-fly from Proton spectrum to HSQC
• SmartDriveNMR analyses both the structure and the acquired NMR Data. Yields result with
confidence level:
With confidence level:
Auto-result
SmartDriveNMR requests HSQC
Higher confidence result !
Automatic decision: auto-result II
SmartDriveNMR: decision-making analysis
• SmartDriveNMR decisions are based on an analysis of both structure and spectra:
• This data analysis is the same as for CMC-q.
• CMC quantification is tested/proven on Expert analysed Data and integrated in
industry workflows.
Concentration
Deviation < 20% < 30% < 40% < 50%
Consistent (85%) 91% 96% 97% 99%
Inconsistent (15%) 60% 68% 77% 83%
Total 86% 92% 95% 96%
Comparison between CMC-q and expert interpretation (~1’700 samples!)
High reliability of the software’s results ! For concentration AND for “consistent”
Example: Actelion’s analysis of the CMC-q auto results
Courtesy of Julien Grimont, Actelion, Allschwil
SmartDriveNMR: consistency check
In addition to the experiment selection, SmartDriveNMR offers a consistency check with CMC-assist analysis, when Consistent, then a full assignment is given. It is done on top of the CMC-q analysis.
• At least “someone” looks at it and suggests an interpretation which is stored along with the data
• Having the right structure is crucial for further work in pharma industry: medicinal chemistry, bioassays. So FP should be as low as possible.
CMC-assist consistency check is validated on Expert analysed Data
CMC-assist – Industry workflow with Structure Verification
I want to improve on selecting sample with the correct structure within a synthesis library
AND look at as little as possible spectra:
what do I achieve with an auto-analysis algorithm if I only look at the red dots?
Idea and reasoning: John Hollerton @ GSK UK
10000 compounds: 5% expected wrong
CMC-assist – Industry workflow with Structure Verification
Idea and reasoning: John Hollerton @ GSK UK
8235 compounds: 0.65% expected wrong
1765 compounds 75% expected correct
False positives
13%
False negatives
14%
(0.65% = 10000 * 5% * 13%)
These I look at or simply mark
These I take for further testing & synthesis
From 5% to 0.65% -> 7 fold better With HSQC, False Positives=3% -> 30 fold better
Predicted Inconsistent with CMC-assist
Predicted Consistent with CMC-assist
Examples for typical confidence values
13 December 9, 2016
Expert: “Small and simple molecule with a nice and disperse spectrum – assignment straight forward”
No follow-up experiment
SmartDriveNMR:
1D 1H Consistent, Confidence 3.5 Stars
No follow-up experiment
Final: Consistent, Confidence: 3.5 Stars
Examples for typical confidence values
14 December 9, 2016
Expert: “Small and simple molecule with an ok spectrum – assignment possible”
maybe more Scans?
SmartDriveNMR:
1D 1H Consistent, Confidence 2.8 Stars
More Scans
Final: Consistent, Confidence: 3.2 Stars
Examples for typical confidence values
15 December 9, 2016
Expert: “Spectrum dominated by Water (and Solvent DMSO)”
Suppression
SmartDriveNMR:
1D 1H Consistent, Confidence 0.3 Stars
Suppression
Final: Consistent, Confidence: 3.1 Stars
SmartDriveNMR: spectrum and structure
Expert: “Structure is complex and 1H spectrum is not easy to assign”
HSQC
SmartDriveNMR:
1D 1H Consistent, Confidence 2.1 Stars
HSQC
Final: Consistent, Confidence: 3.9 Stars
Expert: “Even though structure looks complex and worth an HSQC, the 1H spectrum is fairly easy to assign”
No HSQC
SmartDriveNMR:
1D 1H Consistent, Confidence 3.5 Stars
No HSQC
Final: Consistent, Confidence: 3.5 Stars
M = 422 g mol-1
M = 520 g mol-1
Examples for typical confidence values
17 December 9, 2016
Expert: “Structure is complex and 1H spectrum is not easy to assign”
HSQC
SmartDriveNMR:
1D 1H Inconsistent, Confidence 2.9 Stars
HSQC
Final: Consistent, Confidence: 3.9 Stars
Courtesy of Valentin Poirier, Structuralis, Romainville
SmartDriveNMR: HSQC statistics
When are HSQCs suggested:
• In 32% of > 1000 1D test cases (Ø 300 – 500 g/mol)
• More likely for high MW compounds
0%
20%
40%
60%
80%
100%
Ratio of suggested HSQCs on top of 1D by MW
Summary SmartDriveNMR
19 December 9, 2016
• SmartDriveNMR carries out experiments automatically in an optimized way based on the previous experiment
• Optimization of spectrometer’s time while keeping best quality in results - saving user time “in front of the spectrometer”, e.g. day, night, week ends
• Both NMR experts and non expert benefit
• Interactive / Manual Structure Verification using CMC-assist 2.5
• SmartDriveNMR is new and available in TopSpin 3.5 fully integrated in IconNMR - just a few clicks to try it !
Meet - Rachel
Rachel
• … Works in drug development, Quality Control, Galenic Formulation
• … In charge of determining the amount of active drug in a formulation
• … Needs a rapid and cost effective solution
• … She needs qNMR in automation
qNMR Internal Standard
IS
IS
IS
IS
Mw
Mw
Wt
Wt
I
IPP
a
a
aa
Accuracy Eliminates errors introduced by inherent sample differences
Pa, PIs = potency of analyte/standard Ia, IIs = integral area of analyte/standard from NMR spectrum normalized by number of nuclei Mwa, MwIs = molecular weight of analyte/standard Wta, WtIs = weight of analyte/standard.
e.g. Potency = 90 % means that out of 100mg, only 90 mg are from analyte. Remaining 10% can be anything, moisture, solvent, impurities.
Quantification en solution avec reference externe et interne (qNMR)
Meet - Kim
Kim
• … Works in a hypothetical analytical lab: pharma, forensic
• … In charge of providing consistency check using HRAM MS and NMR
• … Has expertise in NMR or HRAM MS
• … Needs to deliver consistency checks with a high confidence
• … She needs Fusion-SV
Kim’s full verification…
•She wants to be sure, not to file the wrong structure
•Orthogonal data is needed to minimize the chance for false positives
•MS can do things NMR can not (Cl/Br, S/O: totally different mass and small change in chemical shift …)
•Nominal mass leaves 1’000s of potential false positives
•HRAM-MS verifies the molecular formula
NMR can do things MS can not
•100’s of possible structures per molecular formula potential false positives
•1D1H
•HSQC
•HMBC
•13C
Further reduction of
potential wrong passes
Bruker Fusion-SV 1.1
HRAM-MS and NMR data
Details
info in a nutshell
one unified result
shared projects
accurate mass and
isotopic pattern information
automated analysis and
assignment of NMR data
Bruker Fusion-SV 1.1
Automatic analysis of combined 1D1H, HSQC, HMBC and 1D13C data for
Automatic processing of NUS data (HSQC, HMBC)
Spectra interpretation easily understandable
13C assignment HMBC correlations
New: Bruker Fusion-SV 1.1
• License
• TopSpin 3.5pl7: acquisition license allows to use SmartDriveNMR and qNMR (No need of CMC-assist license anymore at the spectrometer)
• SmartDriveNMR
• automatically provides an optimal set of optimized experiments
• Delivers pre-interpreted spectra by CMC-assist
• Qnmr
• Rapid and cost effective solution to determine the amount of active drug in a formulation
• Integrated in IconNMR and CMC-assist, side by side with external reference quantification
• Fusion-SV
• High confidence Structure Verification combining High Resolution Accurate Mass Spectrometry and NMR
December 9, 2016 28