+ All Categories
Home > Documents > Sources of uncertainty in a DVM-based measurement system ...

Sources of uncertainty in a DVM-based measurement system ...

Date post: 01-Mar-2022
Category:
Upload: others
View: 1 times
Download: 0 times
Share this document with a friend
14
Volume 99, Number 3. May June 1994 Journal of Research of the Nationai Institute of Standards and Technology [J. Res. Nail. InsL Stand. Technol. 99, 227 (1993)] Sources of Uncertainty in a DVM-Based Measurement System for a Quantized Hall Resistance Standard Volume 99 Number 3 May-June 1994 Kevin C. Lee, Marvin E. Cage, and Patrick S. Rowe National Institute of Standards and Technology, Gaithersburg, MD 20899-0001 TVansporUible lO IcH standard rcstslors h<[ve become fairly widespread in industrial, university, and government standards labo- ralories because of iheir \tr*/ tempcralurc coefficieni of resistance, ease of iransporta- Ijon. and convenient value. The values of these resistors, however, tend to drift with lime, requiring jieriodic recalibration against an inv-iu^iant standard such as the quantized Hali resistance. The availability of a simple, inexpensive measuremeni sys- tem for calibrating 10 kft resistors against such an invariant standard would be of great benefit to primary standards tab- oratories. This paper describes a simple au- tomated measuremeru system that uses a single, high accuracy, commercially avail- able digital voltmeter (DVM) to compare the voltages developed across a 10 kil stan- dard resistor and a quantized Hall resistor when the same current is passed through the two devices. From these measure- ments, the value of the 10 kQ standard re- sistor is determined. The sources of un- certainty in this system arc analyzed in detail and it is shown thai it is possible to perform calibraltons with lelative com- bined standard uncertainties less than lXl(r'(0.1 ppm). Key words: digital voltmeter. DVM method; electrical metrology: electrical ref- cicnce standards; quantized Hall resis- tance; quantum Hall effect; resistance cali- bration. Accepted: March 3. 1994 1. Introduction Resistors composed of coils of wire wound around suitable forms have been used as staixiards of resistance for many years [IJ. Such devices arc even tod^ widely used as worlcing standards of resistance by primary and secondary standards laboratories in industry, university, and government. Due to aging of the wire and other effects, however, the values of these resistors tend to drift with time, requiring periodic recalibration against a known standard. Because wire-wound resistors drift with time, many national standards laboratories have adopted a standard of resistance based on the quantum Hall effect [2]. When a sample containing a thin, two dimensional conducting layer known as a two-dimen- sional electron gas (2 DEG) is cooled to liquid helium temperatures in the presence of a very strong magnetic field, the resistance of the device becomes quantized, assuming well defined values given by R»{i) VH(I) h_ ie' i (I) where VH is the voltage across the Hall device. / is the current through the device, h is the Planck constant, e is ihe elementary charge, / is an integer, and R^, the von Klitzing constant, has been defined by international agreement to be exactly 25 812.807 Cl for the purposes of practical electrical metrology [3], The resistance is 227
Transcript
Sources of uncertainty in a DVM-based measurement system for a quantized Hall resistance standardVolume 99, Number 3. May June 1994
Journal of Research of the Nationai Institute of Standards and Technology
[J. Res. Nail. InsL Stand. Technol. 99, 227 (1993)]
Sources of Uncertainty in a DVM-Based Measurement System for a Quantized
Hall Resistance Standard
Kevin C. Lee, Marvin E. Cage, and Patrick S. Rowe
National Institute of Standards and Technology, Gaithersburg, MD 20899-0001
TVansporUible lO IcH standard rcstslors h<[ve become fairly widespread in industrial, university, and government standards labo- ralories because of iheir \tr*/ tempcralurc coefficieni of resistance, ease of iransporta- Ijon. and convenient value. The values of these resistors, however, tend to drift with lime, requiring jieriodic recalibration against an inv-iu^iant standard such as the quantized Hali resistance. The availability of a simple, inexpensive measuremeni sys- tem for calibrating 10 kft resistors against such an invariant standard would be of great benefit to primary standards tab- oratories. This paper describes a simple au- tomated measuremeru system that uses a single, high accuracy, commercially avail- able digital voltmeter (DVM) to compare the voltages developed across a 10 kil stan- dard resistor and a quantized Hall resistor
when the same current is passed through the two devices. From these measure- ments, the value of the 10 kQ standard re- sistor is determined. The sources of un- certainty in this system arc analyzed in detail and it is shown thai it is possible to perform calibraltons with lelative com- bined standard uncertainties less than lXl(r'(0.1 ppm).
Key words: digital voltmeter. DVM method; electrical metrology: electrical ref- cicnce standards; quantized Hall resis- tance; quantum Hall effect; resistance cali- bration.
Accepted: March 3. 1994
1. Introduction Resistors composed of coils of wire wound around
suitable forms have been used as staixiards of resistance for many years [IJ. Such devices arc even tod^ widely used as worlcing standards of resistance by primary and secondary standards laboratories in industry, university, and government. Due to aging of the wire and other effects, however, the values of these resistors tend to drift with time, requiring periodic recalibration against a known standard. Because wire-wound resistors drift with time, many national standards laboratories have adopted a standard of resistance based on the quantum Hall effect [2]. When a sample containing a thin, two dimensional conducting layer known as a two-dimen- sional electron gas (2 DEG) is cooled to liquid helium
temperatures in the presence of a very strong magnetic field, the resistance of the device becomes quantized, assuming well defined values given by
R»{i) VH(I) h_
ie' i (I)
where VH is the voltage across the Hall device. / is the current through the device, h is the Planck constant, e is ihe elementary charge, / is an integer, and R^, the von Klitzing constant, has been defined by international agreement to be exactly 25 812.807 Cl for the purposes of practical electrical metrology [3], The resistance is
227
Volume 99, Number 3, May-June 1994 Journal of Research of the National Institute of Standards and Technology
time invariant, and is, under appropriate conditions of measurement, independent of the measurement condi- tions, such as current, temperature, and magnetic field. Because of these properties, the quantized Hall resis- tance, by international agreement, has been used since January 1,1990 as a practical representation of the ohm.
The measurement systems in use at national standards laboratories are quite complex and are capable of achieving relative combined staixiard uncertainties [4] of I X10 * (or 0.01 ppm where 1 ppm - 1X10 *) or less [5]. Many government and industrial standards laborato- ries do not need such small uncertainties in their work. Indeed, the uncertainty required by many laboratories is sufficiently large that the drifts in the values of wire- wound artifact staiKlards are less than their measure- ment resolution and are therefore ignored. Such labiwa- tories are well served by wire-wound artifacts which can be sent periodically to national standards laboratories to be calibrated or to participants in NIST's Measurement Assurance Program, or MAP. Generally, the relative combined uncertainty achieved in the MAP is abtxit 0.1 ppm to 0.2 ppm.
Some laboratories, however, require smaller uncer- tainties. At such levels, drifts of the values of wire- wound artifacts require that they be frequently recali- brated. For these laboratories, the availability of a simple and inexpensive invariant standard of resistance would be of great benefit. The fairly recent introductbn of high accuracy commercial digital voltmeters (DVMs) with 8 1/2 digit resolution has made it possible to con- ceive of a simple and fairly inexpensive measurement system [6] that would enable government and industrial standards laboratories to perform calibrations of their wire-wound resistors directly against a quantized Hall resistor with uncertainties of 0,1 ppm or less.
Such a calibration s>'stem has three distinct parts to it: a quantized Hall resistance device, a cryogenic system in which a superconducting solenoid and the Hall device are cooled to liquid helium temperatures, and a mea- surement system for comparing the standard resistor to this quantized Hall resistor. While the selection of the sample and cryogenic system are beyond the scope of this paper, a few words must be said about them, for they al'fcct the design of the measurement system [7]. The sample and the cryogenic system must be such that the conditions for accurate measurement of the Hall resistance are met [8]. The conditions pertinent to thus discussion are that the plateaus in the Hall voltage ex- tend over as broad a range of magnetic field as possible, and that the voltage drop along the sample in the direc- tion of the current flow, V^, be as small as possible utxler the conditions of measurement.
While in theory any Hall plateau [any value of i in Eq. (1)] can be used for resistance calibration, in practice, the plateaus corresponding to higher values of /' (i > 4) tend, except in unusual cases, to have values of Vj which are unacccptably large for precision resistance calibrations [91. In general, the plateaus corresponding to lower values of j (;=4,2,1) (X:cur at higher magnetic fields, are broader, and have values of V^ that are small enough to permit accurate resistance calibrations. The value of magnetic field at which any given plateau oc- curs LS a function of the electron concentration in the 2 DEC which in turn is a function of the design of the sample: samples can be designed to exhibit the j-4 plateau, for example, at very low fields of only a few tcsla, or very high fields of 10 T or more. In choosing a sample design, one must balance several factors: if the plateau occurs at lower field, it will be more accessible with an inexpensive magnet and cryogenic system, but it will be narrower, and possibly V^ may be too large; on the other hand, if the plateau occurs at a high magnetic field, the plateau will be broader, Kr vvill be smaller, but the cryogenic system and magnet will have to be much larger, and consequently more expensive. It is the opin- ion of the authors that the optimum sample exhibits the /=4 plateau [R»{4)-6 453.201 75 il] in a magnetic field range of 4.5 T to 6,5 T, and the (=2 plateau between 9.0 T and 13.0 T. TTiese fields are easily attainable with commercially available superconducting magnets. The discussion of the measurement system in this paper therefore assumes that the resistance of the Hall device, «H(0, will be 6 453.20175 O, 12 906,403 5 fi, or 25 812.807 fl, corresponding to the i-4,2, or 1 plateau.
2. Description of Measurement System
This measurement system is shown schematically in Fig. 1. The standard resistor to be calibrated (Rs) Ls placed in series with the Hall device (/?») A constant current is passed through both resistors, and the poten- tial difference across each resistor is measured with the DVM. All measurements of resistance are four-terminal measurements. The current source and DVM are con- nected to the resistors using switches, so that the direc- tion of current flow can be easily reversed, and the meter can be connected to either of the resistors. The measure- ments are performed in the order described by Marullo- Reedtz and Cage [5]. The potential drop VR across the standard resistor /?s is first measured with the current flowing in the "positive" direction. The current direc- tion is then reversed, and the potential drop across the standard resistor is measured again. These two
228
Volume 99. Number 3, May-June 1994
Journal of Research of the National Institute of StandanJs and Technology
1 A
Current Source
B ^fj,
Fig. I, DVM-basfd mcastircmcnl system for comparing wire-wound standard resistors (*,1 with a <)uantized Hall resistor (RH)- AL, and Ki; arc the net Icjikagc resistances to ground of the cahles. current source, resistors, and the resl of Ihe measurement system, The contacl resistances are denoted by n , r,^. The voltage across each resistor is mcasuned with ihc digital voltmeter (DVM). The thermal voltages generated at the connections between the DVM and the resistors are denoted by e«, and el,. Z is the internal impedance of the DVM and Yc is the Norton equivalent admittaiKc of the current source.
measurements are repeated in the reverse order, to give a set of four values:
VR(+;), v^-i). v^-i), VR(+/). (2)
The DVM is then connected to the Hall device, and this same sequence of four measurement.s is made on it twice. Finally, the standard resistor is measured again. The entire sequence of measurenienls is:
(1) VR(+/), V„(-/), VR(-/), V„(+/) -
(2) V„(+/), VH(-/), VH(-/), V„(+/)
(3) V„(+/), V„W). VH(-/), VH(+/)
(4) VR(+/), VR(-/), VR(-/), Vp(+/)
In all, a group of sixteen measurements is taken. The four individual values of each of these sets is averaged
<VR.> (3a)
{VH2) (3b)
<v„,) (3c)
<VR4) (3d)
to yield a group of four values shown at the right. It will be shown later that this eliminates the effects of thermal voltages that are either constant or vary linearly with time. Finally, the two voltage drops across the standard resistor, (VRI) and (VR4) are averaged, as are the two measurements of the voltage drop across the Hall resis- tor. <Vw> and (VH3>:
^^^.<M|0^_
(4a)
(4b)
The above measurement sequence is then repeated with the positions of the Hall resistor and the standard resistor interchanged: i.e., the Hall resistor is connected in the top position of the circuit in Fig. 1, and the standard
229
Volume 99, Number 3, May-June 1994 Journal of Research of the National Institute of Standards and Technology
resistor is placed in the bottom position. The average voltages in Eqs. (4a) and (4b) are then computed for this interchanged position, yielding {V^^ and (V»°^).
In theory, if the current through the two resistors is constant,
(vr> (5)
In practice, the situation is complicated greatly by a large number of factors, including thermally generated voltages, leakage resistances, contact resistances, insta- bilities in the current source, noise in the voltmeter, etc. It is shown in Sec. 3 that by application of the measure- ment procedure outlined above, the effects of several of these factors can be eliminated or minimized. Other factors cannot be eliminated by design of the measure- ment system, and their effects must be independently minimized. Some of these factors contribute to the er- ror of the determination of the value of the standard resistor. At this point it should be noted that the value of the standard resistor R is usually expressed as a deviation from its nominal value, denoted Rs^:
\ RT" J Rr"
This ratio can be expressed in terms of the voltage ratios in £q. (5) as:
" (v„) Rr '
As a result, the correction factors and uncertainties in the ratio of the voltages in Eq. (5) must be divided by the
ratio RH
in order to obtain the correctbn factors and
uncertainties in the deviation of the resLsttff from its nominal value.
In this paper, relationships between the magnitudes of the vaiious effects and the resulting error in the value of the standard resistor are determined. Ft is shown in Sec. 3 that if certain limits are placed on the magnitudes of the various systematic effects, the magnitudes of the factors required to correct for their effect on the value of the standard resistor are significantly less than the uncer- tainty in the value of the standard resistor due to random effects, and the corrections can be neglected while still maintaining a relative combined standard uncertainty of
0.1 ppm. In Sec. 4, the uncertainties due to random effects, such as nonlinear drifts in the output of the current source and in the gain and offset of the DVM, etc., arc estimated. Limits arc derived for the maximum values of these factors to ensure that the relative com- bined standard uncertainty in the value of the standard resistor is of the order of O.i ppm. It is shown that an uncertainty of 0.1 ppm or less can be obtained when calibrating resistors with Rs as much as a factor of 4 different from the quantized Hall resistance, /?K(/) [this holds true whether the i = 1, 2, or 4 plateau is used for R„ii), i.e., whether the value of/?„(') is 25 812.807 H, 12 406.403 5 n, or 6 453.201 75 ill
3. Corrections Arising From Systematic Effects
Systematic effects that can contribute significantly to the error of the determination of the value of the stan- dard resistor arc associated with four main parts of the measurement system. These are:
i) the wires, cables, and switches used to connect the Hall resistor and the standard resistor.
ii) the Hall device and standard resistor;
iii) the current source used to supply the current thnxigh the resistors; and
iv) the DVM used to measure the voltages across the two resistors.
The magnitudes of the errors arising from each of these parts of the mea.surement system are estimated in this section.
3.1. Measurement System
In the measurement system shown in Fig. 1, the Hall resistor and the standard resistor to be calibrated are connected in scries, and are connected, by means of cables and switches, lo a current source. In principle, the same current flows through each resistor, so the ratio of the voltage measured by the DVM across the standard resistor and the Hall resistor should equal the ratio of the values of the resistors. In practice, however, there are thermally generated voltages in the wires, switch contacts, and variiws connections in the circuit which cause the measured voltages to differ from the actual voltage drops across the resistors. It is shown in Sec, 3.1.1 that the averaging technique described in Eq. (3) above eliminates the effect of thermal voltages that are either constant or vary linearly with time. Variable
230
Volume 99, Number 3. May-June 1994 Journal of Research of the National Institute of Standards and Technology
contact resistances in the switches connecting the resis- tors to the current source can cause variations in the current supplied by the current source. In Sec. 3.1.2, an upper limit is derived for the magnitude of the permissi- ble variations in the contact resistances. Leakage cur- rents in the cables, switches, and other components of the system can also result in errors in the measurements of the voltages. The errors due to the leakage resistances can be corrected for, as described in Sec. 3.1,3.
3.1.1 Thermal Electromotive Forces Thermally generated voltages in the switch contacts, the wires, and the connections to the DVM result in erroneous deter- minations of the voltage drops across the resistors in the circuit. These thermal voltages are denoted by e,i, and e^ in Fig, I, and arise when there are temperature varia- tions between various parts of the circuit and when there are junctions between dissimilar materials, such as in switches or at solder coniKCtions. Thermal voltages are therefcB^ generated at the connections between the cur- rent source and each resistor, and at the connections between the DVM and each resistor. TTic thermal voltages generated in the contacts to the current source add to the current produced by the current scxirce. As is shown in the next section, if the equivalent admittance of the current source (K in Fig. 1) is bw enough, the current source will adjust its output to maintain a con- stant current through the circuit, and these thermal voltages have no effect on the measurement. Thermally generated voltages in the contacts to the DVM, on the other hand, are significant, and must be corrected for. Since the thermal voltages are a function of the temper- ature differences in the circuit, they do not change sign when the current is reversed. Thus, if one averages the voltage across one of the resistors measured with the current in one direction and the voltage across the same resistor obtained with the current in the opposite direc- tion, the thermal voltage docs not contribute to the aver age. Specifically, if the Hall resistor is in the BOTTOM position of the circuit of Fig.), the voltage drop across it with the current flowing in one direction (which will be denoted as the "positive" direction) is
vTw) -/ GL, + G„
+ e*. (7)
^ (+/) - +/ Gu + GH
+ e* , (6)
where GL," I //?LJ is the leakage conductance in parallel with the Hall resistor (described in more detail in Sec. 3.13) and GH =1/^H is the Hall conductance. With the current flowing in the opposite direction (denoted as the "negative" directkin) one obtains
The average value of the voltage across the Hall resistor is then independent of the thermal voltage:
(vT') = I [vr (+/) -vW' (-/)] - ^^^ (8)
Tt has been assumed that the magnitudes of the ther- mal voltages are the same and independent of the current direction. Practically, the current is reversed using switches, but the connections between the resistor and the DVM are not broken when the direction of the current is changed. Thus, the thermal voltages in the DVM contacts should not change when the current is reversed, and the assumption that the thermal voltages remain constant or vary linearly with time during the measurements should be a gotxi one. It is also important to note that it was assumed that the current dtx;s not drift with time. This is discussed in more detail in Sec. 3,2.
3.1.2 Contact Resistances Contact resistances occur at all of the junctions in the system, including solder connections, switches, and other connectors in the circuit. In addition, the ohmic contacts to the 2 DEG in the quantized Hall device also exhibit contact resis- tance. Voltages develop across only those contacts that have current flowing thr<iugh them. The effects of the resistances of the contacts between the resistors and the DVM arc therefore minimized by making 4-terminal measurements of the voltages across each resistor, i.e., the voltage across a resistor is measured between two terminals that are separate from the terminals that carry the current.
The contact resistances TJ,, ..., r^,, between the resis- tors and the current source shown in Fig. I change the total resistance of the circuit, but if they are constant and reproducible, as would be the resistances at solder con- nections, then it is apparent that they have no effect on the measurement. The resistances of the contacts in the switches connecting the resistors to the current source may change, however, for the switch contacts are con- stantly being opened and closed. If these contact resis- tances are reproducible, then they also have no effect on the total measurement. To see this, consider the case where the total contact resistance for one current direc- tion always differs by an amount hZ from that for the opposite current direction. If the equivalent admittance of the current source is zero, then obviously this has no effect. In practice, however, the admittance of the
231
Volume 99, Number 3, May-June 1993
Journal of Research of the National Institute of Standards and Technology
cuiTcnt source is nonzero, and this difference in contact resistance changes the total load impedance, and slightly affects the current through the resistors by an amount S/L. In this case, the average in Eq. (8) is multiplied by the factor
(-i^)- (9)
where /L is the current through the resistors. If the differ- encc hZ in the contact resistances is reproducible, then the averages of both the voltages across the Hall resistor and the standard resistor are multiplied by this same factor, and it cancels when ratios of the voltages are taken (as described in Sec. 2, and in more detail in Sec. 3.1.3).
If the contact resistances of the switches vary in a raralom, uncoirelated, and irreproduciblc manner every time they are opened and closed, each voltage ha.s a different correction factor of the form given by Eq. (9). These correction factors do not cancel when voltage ratios are taken. As a result, it is necessary to keep 8/iy/i. of the order of 10"" or less to assure uncertainties of 1 part in 10' in ihe determination of the value of the standard resistor. This requirement places a limit on the variation in contact resistance SZ which depends on the admittance of the current source: the smaller the admit- tance, the larger the permissible variations in SZ. Specifically, if the admittance of the current source is K^. it can be shown that (10]'
YcSZ [lHR»+!is)YA
(10)
If the admittance of the current source is about 50 jiS, /f„(j-4) - 6 453.201 75 fi, and ;?s-IO kO,
afL = 2.7 X lOr* SZ.
and SZ must be less than 0.36 mfl for dljli. to be less than 10^. Ff a low admittance current source is used, SZ can be larger, but in general, the contact resistances in the switches and contacts of the circuit .shcxild be kept very small, and should be reproducible to within a few tenths of a milliohm, in order for them not to affect the accuracy of the mea.surcment. This should not be diffi- cult if care is taken to ensure that all of the contact surfaces are very clean and not covered with a thin oxide or a layer of organic contaminant.
' TTie Current source is represented by ils Norttw equivalenl. i.c.. an ideal curreni source In in parallel with an equivulenl iniemal admii- taivx Y;. See Rcf. [10] for more mfonnation.
The problem posed by the resistances of the contacts to the quantized Hall device Ls a vastly more subtle one. As described above, the difference between the resis- tances of the current carrying contacts to the 2 DEC with forward and reverse directions of current must be fess than a few tenths of a millk^m. It is very difficult to produce contacts to the 2 DEG with such small con- tact resistances [11], but fortunately, the actual contact resistances need not be this low: the important point is that the contact resistances must be independent of the direction (and magnitude) of the current. Nevertheless, it is neces.sary that the current carrying contacts have con- tact resistances of less than 10 mO, or they generate substantial amounts of noLse which prevents accurate resistance comparisons. One would think that the resis- tance of the contacts used to measure the Hall potential are unimportant, for no current flows through them. These contact resistances, however, must also be in the range of a few milliohms, for reasons which arc beyond the scope of this paper [7, 121.
3.1^ Leakage Resistances Leakage resistances arise from the noninfinite resistance of the electrical insulations used in constructing the system. As a result, the leakage resistance is distributed throughout the sys- tem: (here are contributions from the current source, the cables, the DVM, and even the standard resistor and the wires leading to the quantized Hall resistor. The leakage resistances from the high terminal of the current source, the current reversal switch, the cables, the standard resis- tor, and the DVM have been combined in the idealized "leakage resistance" R^ shown in Fig, 1. The leakage resistances from the low terminal of the current source, cables, etc., and the quantum Hall resistor have been combined in the idealized "leakage resistance" Ri_.. The circuit is grounded between the Hall resistor and the standard resistor to minimize the effects of the leakage resistance between the low terminal of the DVM and ground: the low terminal of the DVM is always con- nected to this point throughout the entire measurement sequence. The high terminal is alternately connected to point A in Fig. 1 to measure the standard resistor, or point B to measure the Hall resistor. If it can be assumed that the leakage resistances /?L, and /JL, remain constant throughout the measurement cycle, and that the Hall resistor arxl the standard resistor have nominally the same values, it can be shown that the error due to fcak- age resistance is eliminated by averaging the ratio of the voltage across the standard resistor to the voltage across the Hall resistor with the resistors in the standard config- uratk>n shown in Fig. 1 and the same ratio obtained when the positions of the resistors in the measurement circuit are interchanged. With the resistors connected as shown in Fig. 1 (the standard position), the average
232
Volume 99, Number 3, May June 1994 Journal of Research of the National Institute of Standards and Technology
voltages across the resistors [as determined from Eq. (8)1 are:
<vr>- GL, + Gs
(11a)
01b)
where G —\/R for each of the resistors in Fig. 1. The ratio of these two voltages is then
<vr> " GL, + Cs
-^(1+-^ ^). (12) Gs \ GH GS /
if it is assumed that the leakage resistances are inde- pendent of the positions of the resistors, the positions of the Hall resistor and the standard resistor can be usefully interchanged. In this case, the Hall resistor is in parallel with Rt, and the standard resistor is in parallel with R^^. The ratio of Eq. (12) is then
tonstnicted current sources will have leakage resis- tances greater than about 10'' H (GL < 10"'^ S). If/?s - 10 kfl, and /fH - 6 453.201 75 O or 12 906.403 5 fl. the correction to the value of the standard resistor will be of the order of 0.{X)3 ppm, which is more than an order of magnitude less than the uncertainty due to random ef- fects, and can therefore be neglected. For resistance ratios of 4 or more, as would result from the comparison of a 6 453.201 75 H and a 25 812.807 11 resistor, how- ever, the correction to the value of the standard resistor can be as large as 0.02 ppm, which is comparable to the uncertainly due to random effects and cannot be ne- glected. In this case, a correction factor can be estimated by measuring the leakage resistance between the point C in Fig. 1 and earth which will be approximately equal to I/(GL,+GL;) (in Fig. 1 point C is connected to earth, but for this leakage resistance measurement this connection must be removed).
The uncertainty associated with the assessment of this leakage resistance, however, will be quite large, possibly as large as the conection factor itself, so one can treat this correction factor as a component in the assessment of the combined uncertainty. If lower uncer- tainties are required, however, it will be necessary to increase the leakage resistance of the system another order of magnimde, something which is quite difficult to do.
(vD GL,+GH a <Vff*> GL, + GS G ;('-|;-t)- '">
Averaging Eqs (12) and (13) yields:
2 l<vr)"- {v^-})
GLI + G|.,
2Cs (14)
If the standard resistor has exactly the same value as the Hall resistor, then the last two terms on the right side of Eq. (14) cancel, and the leakage resistances have no effect. If the resistors are not equal in value, then there Is a small correction factor. If the leakage resistances change when the current is reversed or the resistors arc interchanged, then the above analysis does not hold. Even in this ca.sc, however, the error in the determina- tion of the value of the standard resistor caused by the leakage resistances will be of the order of the ratio of the leakage conductance CL=GI ,+Gi.j to the conductance of the larger of the two resistances K,i or Rs- Typically, cables insulated with PTFE Teflon, and carefully
3.2. Current Source
Several factors determine the optimum current for these measurements. The higher the current, the larger the voltages across the resistors, and the smaller the averaging time required to obtain voltage measurements with a given experimental .standard deviation of the mean. Too large a current, on the other hand, can cause self-heating of the standard resistor, which will change its value and, more importantly, can cause breakdown of the dissipationlcss current flow in the Hall device, rcn dcring it unusable for resistance calibrations [ 13J. Typi- cally, currents between 10 jjiA and 50 p.A satisfy these conditions. It should be noted that when performing calibratk>ns of 10 kXl resi.slors against a quantized Hall resistor, it is the maximum current that the quantized Hall device can sustain withcHit breaking down that lim- its the maximum current at which mca.surements can be performed. Generally, this critical current is far below the current at which even a typical 10 kil resistor would start to show self-heating effects.
It should be recognized, however, that a reduction in averaging time due to a higher current is only realized if the primary factor limiting the accuracy of the voltage measurement is the signal-to-nolsc ratio and not the resolution of the meter. If the primary limiting factor is
233
Volume 99, Number 3, May-June 1994 Journal of Research of the National Institute of Standards and Technology
the resolutbn of the meter, then the benefits of increas- ing the current are somewhat limited. For example, if the resolution of the meter is 10 nV on the 100 mV range, then one benefits from choosing a current that produces a voltage near the top of this range, as in this case the voltage measurement has the minimum relative uncertainty. Increasing the current so that the voltage is so large that the meter must use the next range, e.g., the 1 V range, however, may not result in any benefits if the resolution of the 1 V range is 100 nV. In this case, the relative uncertainty of the voltage measurement will be the same or worse than at the lower current. In practice, some benefit may be obtained by using the 1 V range in this example, tor the internal DVM noise is often less on the higher voltage ranges than on the 100 mV range, so lower uncertainties in the measurement of the voltage can be achieved with shorter measurement times, even though the resolution of the meter is poorer on the 1 V range.
The method for determining the ratio of the value of the standard resistor to the Hall resistance described in Sec. 2 assumes that the current through the resistors is constant during the time that the voltage measurements are made. Any variations in the current, such as drifts or noise, will appear as drifts or noise in the voltages across the resistors and will affect the accuracy of the calibration of the standard resistor. In order to obtain a reladve combined standard uncertainty in the calibration of 10"\ it is necessary to keep the noise in the current source aixl nonlinear deviations in the current below 0.01 ppm; for currents between 10 fj,A and 50 jiA, this dictates that the current variations be less than 0.1 pA to 0.5 pA. Such low noise levels are rather difficult to achieve with active current sources containing opera- tional amplifiers, transistors, or other solid state compo ncnts which usually have noise levels of the order of a few ppm, and are therefore generally unacceptable for use with this method. Current sources made using mer- cury batteries and current-limiting wire-wound resistors are capable of meeting these stringent requirements, even though the output of these current sources tends to decrease with time in a predictable, linear manner.
Fortunately, .such stable linear drifts in current do not affect the measurements if the measurement sequence described in Sec. 2 is used. If the current decreases at a constant rate B, and the current at the beginning of the first measurement is /, then the current at a time r after the first measurement was begun is
/(0 = I-S^ (15)
each voltage measurement in the group is given by the expression in colunm 2 of Table 1. As described above in Sec. 2, the first four voltage measurements (of the standard resistor with positive and negative current) are averaged to eliminate thermal voltages, yielding the value shown in the third column of Table 1. The average voltages across the standard resistor are then averaged, as are the average voltages across the Hall device, as described in Sec. 2, Eq. (4a), to give
(vp, . (iiiliiiW . «. (^-)
<vr> = <VH2) + <Vm) «„[,,-(f«i,)]
If each individual voltage measurement described in Sec. 2 takes a time Ar, then the current at the start of
(16b)
The effective current is the .same in both of these equations, and is eliminated when the ratio of VH to VR
is taken. Note that the effective current in both of these equations is the current at the exact mid-point of the measurement cycle.
33 Digital Vtdtmeter (DVM)
The quality of the digital voltmeter used to measure the voltages across the resistors is the ultimate factor limiting the accuracy of this technique. In order to ob- tain resistance calibrations with relative combined stan- dard uncertainties of 10"\ the DVM must be capable of measuring voltages with uncertainties about a factor of 5^10 less than this. If a 20 JJLA current is used, the voltages across the resistors will be of the order of 200 mV, and the DVM must be capable of resolving voltages of the order of 0.2 X10 "^ V, or 20 nV [14]. Commer- cial DVMs arc now available from several manufactur- ers that have such a high resolution. In addition to the high resolution, however, the DVM must have very high accuracy, a high degree of linearity, a high input impedance, high stability, and very low noise. Commercial 8 1/2 digit multimeters from several manu- facturers arc on the market that meet these specifica- tk>ns. In this section, various systematic effects
234
Volume 99. Number 3, May-June 1994
Journal of Research of the National Institute of Standards and Technology
IfaMe I. Assuming a constani rate of change in the currcnl pro(tu<K<l by the current sounre and thai each ^^lage mcasurcincnt lakes a time A/, the current is calculated at the beginning of each voltage measure- ment on the standard resistor (R^) and and the Hall resistor (AH)- Each set of four measurements on a resistor is averaged to eliminate thermal voltages and other constant offsets, resulting in the average voltages shown in column 3.
Resistor Mean voltage
/fH(+) +(/(r-llBAf)
«s[/^(f)BA<]
/fK[/n-(-YjBA/j
>?H[/t--(y)sArj
associated with the DVM that contribute to the total uncertainty arc analyzed. In Sec. 3.3.1, a correction tenn accounting for offsets and nonlinearities in the response of the DVM to applied voltages is deriwd. In Sec. 3.3.2, the effect of the small current source between the input terminals of the DVM is con.sidered. The ef- fect of noninfinite input impedance is considered in Sec. 3.3.3.
While this measurement system is quite similar to potentiometric measurement systems achieving smaller ultimate uncertainties [5], the accuracy, range, resolu- tion, and linearity requirements on the DVM used with this system are greater. In the potentiometric measure- ment systems, a potentiometer is used to cancel the Hall voltage, so the detector is only used to measure very small differences between the Hall voltage and the voltage drop across the standard resistor being cali- brated. Therefore, the detector need not have a very great range, but must have very low noise, high resolu- tion, and high accuracy. Furthermore, since the detector is only measuring small deviations from zero, the linear ity of the detector over large ranges is not crucial. In the measurement system described in this paper, the DVM is used to measure voltages that differ widely from zero, and that are both positive and negative. This requires the DVM to have a very high degree of linearity.
Offsets and nonlinear responses of the DVM can be determined by calibrating the DVM against a Josephson airay. The Josephson array produces a time-invariant voltage that is related to fundamental constants, and, by international agreement, provides a practical metroJogi- cal represcntaiit>n of the volt. If the Josephson array produces a defined voltage V, the voltage indicated by the DVM will be:
VDVM = A + gr + A^(T), (17)
where A is the offset, g is the gain of the DVM, and N(y) is a nonlinear correction. For most modern high quality meters,
g - 1 + 5. (18)
where 5 is a small number. The values of A, ^, and N should be determined by measuring VDVM with applied array voltages in the neighborhood of the values ex- pected to be encountered in the resistance measure- ments. A least-squares fitting procedure should be used to determine the gain, offset, and nonlinear corrections for both positive and negative voltages.
If the offset voltage A is the same for both positive and negative voltages, then the same averaging proce- dure that eliminates the thermal voltages will eliminate the offset voltage: the offset will cancel when V'(-(-/) and V{-/) are subtracted, as in Eq. (8). In practice, however, neither A nor g need be the same for positive and nega- tive voltages.
As described in Sec. 2, the ratio of the resistor values is determined from the arithmetic mean of the ratios of the averaged voltages across each resistor in the standard and interchanged posilbn, as in Eq. (14). Each voltage in Eq. (14) must be corrected for the offset, nonunity gain, and nonlinearities in the DVM before the ratios are taken. A factor taking all of these corrections into ac- count can be derived for Eq. (14) as follows: the voltage indicated by the DVM when it is connected to the Hall resistor in the bottom position in the circuit of Fig. 1, with positive current, is given by (neglecting thermal voltages)
V^(+/) = A^ + S*Tr + N(rr), (19)
where "VH^ is the' 'true" voltage across the Hall resi.stor in the bottom position in the circuit of Fig. 1. Likewise, with the current reversed.
vr(-/)=A- - g- rr+N{-.rry (20)
235
Vblume 99, Number 3, Mj«y-June (994 Journal of Research of the National Institute of Standards and Technology
Averaging as in Eq. (8) gives:
A*-A where
SN^~-^ ^-^ ^. (2ld)
If leakage resistances and other systematic effects are ignored, the ratio of the "true" voltages across the resistors is equal to the ratio of the resistors, i.e..
Inverting Eq. (21a) above gives
rr- (VD -a -SNfT
(22)
(23)
Repeating the calculation for each of the voltages and substituting into the left hand side of Eq. (22) yields
(W y-a-SNT
(24)
Generally, the corrections a and 5JV are very much smaller than the values of the voltages (V), so that the ratio of a+8N to (V) is of such small magnitude that terms of second order in this quantity can be neglected. Equation (24) can Chen be simplified to
^"2 l(vr) (vn^
({vf^\ ( « + &vr . a + mT VvT) )(- <vf^ <vr) )] (25)
The first term on the right side of Eq. (25) is the ratio of the voltages across each resistor [averaged as in Eq. (3)], while the second term on the right side of Eq. (25) is the correction factor that must be applied to correct for nonzero offsets and nonlinearities in the DVM. This expression can be considerably simplified by noting that both a and dN are very much smaller than the voltages (V), so the denominators can be replaced by their nom- inal values. ]n addition, the ^ are the differences between the nonlinearity corrections for positive and negative voltages, and since the magnitudes of the voltages across a given resistor are alt essentially the same, little error will be incurred by making the approx- imation that SN is a constant for each resistor. Thus, with the approximations
<vr') - <vn - {Vu)
^(^(«-HM^„)-(a.6Ar.)). (27)
It is important to remember that a is die difference between the offsets for positive and negative voltages, as given by Eq. (21) above. Usually, the offsets are negli- gibly small and are the same for positive and negative voltages, so that a in die above equation can be ne- glected. In this case, Eq. (27) can be written
The quantity 8N in the above equation is the difference between the nonlinearity correction for positive voltage and for negative voltage. This need not be zero and must be determined by calibrating the meter against a Joseph- son array. It should be noted that if both resistors have nominally the same value, then even though bN^ and ?>N„ may not be zero, Ihcy will have essentially the same value, and the correction term in Eq. (28) will vanish.
236
Volume 98, Number 3. May-June 1994
Journal of Research of the National Institute of Standards and Technology
Corrections for the non linearity of the meter therefore need only be made in the case that the resistors being compared have nominally different values. In the case that the ratio of the resistances is 4, this correction term can be quite significant. Using the DVM calibration data in Fig. 2 of Cage et al. [6], and assuming Rs-25 812.807 il and /?H-6 453.201 75 «, the correc- tion terin of Eq. (28) would be 0.40 ppm ± 0.28 ppm. It is important to note that the correction to the value of the resistor /?s, expressed as a relative deviation from its nominal value, (/?s-A?s°"y/^, due to nonlinearitics in the DVM is obtained by dividing Eq. (28) by the ratio i?s™//?H (sec Sec. 2), which in the case of this example is 4. The correction to the value of (Rs-Rs^}/Rr" due to the DVM nonlinearily is therefore only O.I ppm ± 0.07 ppm in this example.
Since the nonlinearity correction term above may change with time, the DVM should be calibrated before each resistaiKe calibration is performed. It should be noted that if either the offsets or the gain vary randomly with time, they will not cancel in the above equations, and will give rise to an uncertainty in the final determi- nation of the value of the standard resistor as discussed in Sec. 4,
3.3^ Internal Current Source Modern digital voltmeters inject a small current into the circuit to which their input terminals arc attached. This current is often very small: for a Hewlett Packard 3458A DVM^ it was measured to be 10^"* A (with no bias applied to the meter input terminals) by connecting a Keithley 602 electrometer directly across the input terminals. If this current does not change sign or magnitude with varying applied bias, it should have no effect on the average voltage measurements, as any offset produced by it wouki cancel when the voltages measured with positive and negative current are averaged. If it changes sign and magnitude with changing applied bias, but generally retains a magnitude in the range of lOr'* A, and if the measurement current is in the range of 10 JLA to 50 n,A, this small current would result in an error in the ratio of the resistors of only 0.0002 ppm to O.OOlO ppm.
3.3J Input Impedance Whenever a DVM with finite input impedance is connected across one of the resistors, some current will be shunted through the DVM. If the input impedance of the DVM is ZDVM=
l/GuvM, and if the admittance of the current source is negligibly small (that is, the current source is assumed to
be ideal), then the correction to the resistance ratio with the resistors in the standard pc)sition will be:
' Certain commercial cquipmem, instmmems. or materials arc idemi- fied in ihis paper in order to adequately specify the experimental piDCfdure. Such idenlificaiion does not imply reeominendaiion or endorsemcni by the National Institute of Standards and Technology, nor dors it imply thai (he materials or equipment identified are neces- sarily ihc best available for the purpose.
<vr> GH + GDVM
^H \ 2DVM / (29)
In the event that the two resistors being compared are of nominally equal value, the correction factor in Eq. (29) will vanish. If however, the resistors have different nominal values, the factor must be evaluated, and can be quite appreciable, particularly if the input impedance of the DVM is less than 10" 11. In this case, if /?s-/rH((=l)=25 812.807 H and «H('-4)«6 453.201 75 n (a 4-to-l resistance ratio) the correction to the value of Rs will be as much as 0.019 ppm. In practice, the input impedaiKes of modern DVMs are somewhat higher than 10" fl. The input impedances of two Hewlett Packard 3458A DVMs were measured using a method described by Cage et al. [6]. The impedances, measured at 22 °C with a relative humidity of 43%, were 5.9X10'" D. for one DVM and 3.6xlO'^n for the other DVM. Such input impedances would lead to only a 0.001 ppm correction to the resbtance ratio when calibrating a 10 kCt resistor against /f„(i_4)-6 453.201 75 ft, and a 0.005 ppm collection lo the resistance ratio when calibrating a 10 kft resisted against R»{i-\)- 25 812.807 ft.
4. Uncertainties Arising from Random Effects
The discussions in the previous sections have con- cerned only errors and sources of uncertainty in the determination of the value of the standard resistor due to systematic effects. As can be seen from the summary in Table 2, the errors associated with most of these system- atic effects can be reduced to the level of 0.0] ppm by appropriate design of the measurement system. Ran- dom variations in the various parts of the measurement system, such as nonlinear drifts in the current, thermal voltages, contact resistances, and even the offset and gain of the DVM can also contribute to the uncertainty in the determination of the value of the standard resistor. Many of these have been discussed in Sec. 3 aixl their effects are summarized in Table 3. Appropriate design and construction of the measurement system, as noted in Sec. 3, generally results in contributions to the com- bined standard uncertainty from these effects of a few parts in 10*. External noise picked up by the cables in
237
Volume 99, Number 3. May-June 1994
Journal of Research of the National Institute of Standards and Technology
the measurement system, and circulating currents aris- ing from multiple ground connections in the circuit (commonly referred to as "ground loops") can also contribute to the combined standard uncertainty. Fortu- nately, the influence of these effects on the measure- ment can be greatly reduced by adequately shielding the measurement cables and equipment, preferably with two layers of shielding, and checking the circuit to ensure that there is only a single connection to ground (sec Ref. [5J). Often this latter task is complicated considerably by the fact that some instruments have various internal connections to ground that are not obvious, so that while
TbUe 2. Systematic effects thai give rise to coircciions lo the value of the standard resistor dctemuned using this method. If the value of each effect is kept within the limits shcFwn in the second cdumn, the value of the correction lo the deviaiicm of the value of Rs from its nominal value, given by the quantity (R^-Rf"VK%^. will be shown in the thiid ct^umn. The fiiS value was calculated assuming that lfs-10 kn and Rt,(i-4}-6 453.201 75 fl; the second value was calcu- lated assuming As-10 kO and Rn-25 812.807 n. Since the values of these corrections are generally very small in comparison with the uncertainty due to random effects, these correction factors are not applied, but for simplicity arc included in the uncertainty of the final result.
Effect
Afttuc to result calculated: (in ppm)
leakage resistance >10''n 0.004 to 0.016 3.1.3
DVM input impedance >10"n <0.003 to 0.015 1.3 J
DVM nonlincarity <0-05 n-V Oio<0.10*0.07 3.3.1 ±0.01 |iV
IWble 3. Randomly varying effects that com ributc to the uncertainty in the determination of the value of the standard resistor. In the first row, the first number in the third column was calculated assuming Ks-IO kfl and RH-6 453.201 75 fl; the second number was calculated assuming flH-25 812.807 il.
Effect Value Standard Section in uncertainty which
(in ppm) calculated
DVM current source <10r"A 0.000 to O.OOI 3.3.2
Noise in DVM 5nV to 15 nV 0.015 lo 0.046 4 for 30 s for single group of
measurements measuremenis
it may appear that the meter is isolated from ground, it in fact is cither connected internally to ground through its power cable, or through the chassis, which may be screwed to a rack, which itself may be groutxled.
Even if the effects of external noLse, random drifts in the system, etc., can be eliminated, the fundamental fac- tor limiting the accuracy of this technique is the internal noise and resolution of the DVM. Modern commercial 8 1/2 digit multimeters are available that have resolu- tions as fine as 10 nV out of 1 V, as mentioned in Sec. 3.3, so the DVM resolution does not really limit the accuracy of this technique. The internal noise in the DVM, however, can have rms values as high as 0,04 ppm to 0.1 ppm. TTiLs, coupled with random, long-term drifts in the gain, offset, and nonlinearity, are primarily responsible for limiting the accuracy of this technique.
In order to minimize the effect of such noise on the measurement, we must first examine the nature of the noise and how it affects the measurement. What we measure are the voltages across the resistors. These may be regarded as fixed voltages, upon which are super-im- posed noise voltages, which can have different frequen- CKS and amplitudes. Very high frequency noise, with a mean period less than the time required for a single measurement, will not greatly affect the measurement, for it will average to zero during the measurement time. Thus, the uncertainties in the voltage measurements in Eq. (3) can be greatly reduced by averaging each voltage measurement for a bng time. This can be done by making repeated measurements of each voltage over an extended period of time and then averaging these mea- surements to produce a mean voltage. The uncertainty in this mean will be given by [see Ref. 14]
5 = N{N - 1)
(30)
It would appear from Eq. (30) that the more measure- ments of each voltage are made, the smaller the uncer- tainty in the mean. This, however, is not the case, for in addition to the obvious high-frequency noise that gives rise to scatter in the individual measurements (each of which are assumed to be made over a short time interval of a second or less), there are long term drifts in the current, thermal voltages, and components of the mea- surement system. The.se drifts can be considered to be "noise" with very low frequency. In addition, there may be long-term, predictable drifts in the system: an example of this would be the stow, linear decrease in current produced by a mercury battery-powered current .source as the batteries get depleted. Averaging a voltage measurement for longer periods of time will decrease
238
Volume 99. Number 3. May-June 1994 Journal of Research of the National Institute of Standards and Technology
the effects of higher frequency noise wiih periods less than the measurement time, but if these long-term linear drifts in current (for example) are present, there is some measurement time beyond which the effects of short term noise become negligible, and the primary coniribu tion to the uncertainty given by Eq. (30) will be the linear drift in measurement current. For a drift of atxxit 0.2 ppm in 10 min, this time can be as short a.s 10 h or as long as several days, depending on the magnitude of the noise voltage.
For optimum results, the individual voltage measure- ments indicated in Eq, (3) should be averaged for short enough times that long term drifts do not contribute to the uncertainty of the measurement at ail. Then, as was shown in Sec. 3.2, appropriate averaging of sets of such measurements causes the effect of the long-term drifts to vanish. With modern 8 1/2 digit DVMs, averaging each of the voltage measurements in Eq. (3) for a period of about a minute results in uncertainties in the determi- nation of the mean value of each voltage of between 5 nV and 15 nV, depending on the internal noise of the DVM. This results in a relative standard uncertainty in the determination of the resistance ratio from a single group of measurements of between 0.015 ppm and 0.046 ppm.
It is important to understand that this estimate of the uncertainty in the resistance ratio oidy considers the contributions from the noise, and does nor include other random effects, such as irrcproducible contact resis- tances, and various random drifts in the system. There- fore, while the uncertainty due to noise m^ be quite small, the actual value of the ratio may be considerably different from the value determined from a single group of measurements- In order to reduce the combined un- certainty due to all raitdom effects in the determination of the resistance ratio, it is therefore necessary to per form numerous groups of measurements, and then aver- age the resistance ratios r determined from each group to obtain a final average < n >:
1 "' (31)
Because long-term drifts and other random effects tend to cause fairly large fluctuations in the values of the r^ determined from different groups (usually larger than the uncertainty in each r, due to noise), the uncertainty in the final resistance ratio determined from Eq. (31) is determined from the set of group means < r^ >
N{N-i) ^{r.-<r>f (32)
The uncertainty calculated from Eq, (32) includes the effects of noise, and of random variations in the system, so there is no need to include these effects explicitly. The magnitude of the uncertainty will depend on the magnitudes of these various random effects. When the internal noise in die DVM is sufficiently low that each voltage measurement has an uncertainty on the order of 5 nV to 7 nV. noise in the DVM will actually be only a minor factor contributing to the final uncertainty: more significant contributions will come from the irrepro- ducibility of the contact resistances and other factors shown in Table 3. If these random effects are kept within the limits shown in Table 3, however, it is usually possi- ble to obtain relative uncertainties (due to random ef- fects) in the determination of the mean resistance ratio [Eq. (31)1 of less dian 0.01 ppm after 20 groups of measurements.
5. Conclusion Starvlards laboratories requiring resistance calibra-
tions with relative combined standard uncertainties of less than 0.1 ppm could benefit from a simple, inexpen- sive, intrinsic resistance standard. While the quantum Hall effect provides such a standard of resistance, the measurement systems used at most national standards laboratories are far too expensive, complex, and time consuming to construct and use in government and in- dustrial standards laboratories. This paper has analyzed the sources of uncertainty arising from both systematic and random effects in a simple quantized Hall resistance measurement system that uses a modern commercially available 8 1/2 digit digital voltmeter to compare the voltages developed across a standard resistor placed in series with a quantized Hall resistor. A measurement sequence has been presented which minimizes the ef- fects of thermal voltages and linear drifts in the current on the final determination of the unknown resistance. Criteria have also been presented for minimizing the effects of the contact resistances and leakage resis- tances.
Table 2 summarizes the systematic effects that cause errors in the determination of the resistance using this technique. Most of these errors are so small as to be negligible in comparison with the uncertainty in the final resistance ratio due to random effects, aixi these corrections arc therefore neglected and for simplicity are included in the uncertainty of the final result. In the case of the nonlinearity of the DVM, however, the error can be fairly large, and a correction factor [derived in Eq. (25)J must be applied to the final result. Table 3 summarizes the pcrmi,ssible magnitudes of random vari- ations and drifts in the various components of the mea- surement system.
239
Volume 99, Number 3, May-June 1994
Journal of Research of the National histitute of Standards and Technology
The combined standard uncertainty in the value of die standard resistor is the square root of the sum of the squares of the standard uncertainties arising from both random and systematic effects in the voltage measure- ments. If the uncertainty from random effects is less than 0.06 ppni (easily achievable with just a single group of measurements), the relative combined standard un- certainty in the determination of the resistance can be as low as 0.06 ppm if the DVM nonlinearity correction in Table 2 is negligible (including for simplicity the correc- tion factors in Table 2 as sources of uncertainty). If the DVM nonlinearity correctbn is as high as that shown in Table 2, the combined uncertainty will be as high as 0.09 ppm.
If the uncertainly due to random effects is less than 0.01 ppm, which can be achieved by averaging up to 20 groups of data as described in Sec. 4, the relative com- bined standard uncertainty in the value of the standard resistor can be less than 0.03 ppm (again assuming the DVM nonlinearity correction to be negligible, as waukl be the case if the resistors being compared had the same nominal value). This DVM-based measurement system can, therefore, be used to compare wire-wound standard reference resistors with quantum Hall resistors with a relative combined uncertainty of less than 0.1 ppm, and in the most favorable cases, with uncertainties less than 0.06 ppm. Since the quantized Hall resistance does not drift with time, this measurement system can be used to calibrate wire-wound resistors, the values of which tend to drift with time.
6. Acknowledgements
Tlie authors gratefully acknowledge the support of the Calibration Coordination Group of the Etepartment of Defense for this work. They also wish to thank Drs. B. F. Field, B. N. Taylor, R. F. Dziuba. and A. F. Clark of NIST and Prof. C. A. Lee of Cornell University for helpful discussions.
guantizeri Hail Resistance, J. Res. Natl. Bur. Stand. (U.S.) 92, 303^310 (1987).
[6] M. E. Cage, D. Y. Yu. B. M. Jeckelmann, R. l. Steiner, and R. V. Duncan, Investigating (he Use of Multimeters to Measure Quantized Hall Resistance Standards. IEEE TYans, Instnim. Meas. 40. 262-266(1991).
[7] For more information on the selecticm of samples, see W, van dcr V/e\. R G. Haajiappe). J. E. Mooij. C. J. P. M. Hannans, J. P. Andrf. G. Weimann. K, Ploog, C.T. Foxon, and JJ. Harris, Selection Criteria for AlGaAs-GaAs Hetcrostmciures in View of Their Use as a (Juanlum Hall Resistance Standard, J. Appl, Phys. (SS. 3487-3497 (1989).
[8] F, Dclah^e, Ttchnical Guidelines for Reliable Measurements of the Quantized Hall Resistance, Metrolc^ia 26, 63-68 (1989).
[9] See, for example Fig. 6 in K. Jaeger, P. D. Levjnc, and Cj\. Zack. Industrial Experience with a Quantized Halt Effect Sys- tem. IEEE Trans. Inslrum, Meas. 40. 256-261 (1991).
[101 P. Horowiu and W. Hill, The Art of Electronics (Cambridge Uriv. Press, Cambridge, England, 1980), p. 8.
[Ill D. Jucknischke,H. J. Buhlmann. R. Houdi^. M. Itcgcms, M. A. Py, B. Jeckelmann, and W. Schwiiz, Properties of Alkiyed Au- GeNi Contacts on GaAs/GaAIAs HetetDstnsclures. IEEE Trans. Instrum. Meas. 40, 228-230 (1990),
(12} H. Hirai and S. Komiyama, A Comacl Limited decision of the Quantized Hall Resistance, J. Appl. Phys. 69, 655-662 (1990).
(13) M. E. Cage. R. F. Dziuba. B. F. Field, E. R. WHUams. S. M. Girvin. A. C. Gossard. D. C.Tsui, and R, J. Wagner, Dissipation and Dynamic Nonlinear Behavior in the QusUilum Hail Regime, Phys. Rev. Lett. SI, 1374-1377 (1983).
114] The uTKertainly of the DVM measttfement is 1/2 V3 of its resolution. The derivation of this result is given in Aruiex F2 of the Guide to the Expression of Uncertainty in MeasurettKiU (Intemaiiona) Organization for Standardizalion. Geneva, Switzerland, 1993). p. 53.
About the authors: Kevin C. Lee and Marvin E. Cage are physicists in the Electricity Division at NIST. Patrick S. Rawe, a student from the Massachusetts Institute of Technology, was a Guest Researcher in the Electricity Division when this work was done. The National Insti- tute of Standards and Technology is an agency of the Technology Administration, US Department of Com- merce.
7. References
|5|
See, for example. J, L, Thomas, Precision Resistors and their Measurement, in Precision Mcasurcmcni and Calibration: Elec- tricity and Elecirtmics, Natl. Bur. Stand. (U.S.) Handbook 77, Vd. 1(1961), pp. 111-142. K. v. Ktiizing, G. Dorda. and M. Ftpper. New Method for High-Accuracy Deicrraination of the Fine-Structure Ctwisiam Based on Quantized Hall Resistance. Phys. Rev. L^ii. 45, 494- 497(1980). BIPM Proc,-Vert). Com, Im. Poids ei Mcsurcs 56. 20 (1988). B. N. Taylor and C. E. Kuyatt, Guidelines for Evaluating and Expressing the Uncertainty of NIST Mcasurctnent Results. Natl. Inst. Stand. Technol. Itch. Note 1297 (January 1993). See. for example, G, Manillo-Recdtz and M, E. Cage. An Auto mated Poteniiomciric System for Precision Measurement of the
240

Recommended