+ All Categories
Home > Documents > Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime...

Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime...

Date post: 28-Apr-2018
Category:
Upload: vanhanh
View: 217 times
Download: 2 times
Share this document with a friend
23
Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design Kenneth Nordsieck University of Wisconsin Document Number SALT-3120AE0001 Revision 2.21 10 Mar, 2003
Transcript
Page 1: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

Southern African Large Telescope

Prime Focus Imaging Spectrograph

Optics Design

Kenneth Nordsieck

University of Wisconsin

Document Number SALT-3120AE0001

Revision 2.2110 Mar, 2003

Page 2: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 i

Change History

Rev Date Description

1.1 5 Oct, 2001 PDR

2.0 9 Aug, 2002 Critical Design, first iteration

2.1 27 Nov, 2002 Mods for Optomechanical design

2.2 4 Mar, 2003 CDR

2.21 10 Mar, 2003 Add longslit figure, and grating and filter complements

Table of Contents

1 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

2 Design Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.1 SALT Telescope model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.2 Design goals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22.3 Optical Indices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

3 Optics Subsystems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63.1 Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63.2 Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73.3 Collimated Beam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83.4 Polarimetric Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93.5 Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103.6 Slitmask / Slitviewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

4 Thermal Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

5 Imaging Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155.1 Image Quality Budget . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155.2 RMS Image quality vs configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

6 Throughput . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186.1 Coatings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186.2 Stray Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196.3 Efficiency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Page 3: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 1

1 Scope

This document presents the top level optical design for the Prime Focus Imaging Spectrograph. The following subsystems are described in more detail in the following PDR Subsystem Designor Trade Study Documents:

SALT-3120AA0002 Grating and Filter Trade Study

SALT-3120AA0003 Polarimetric Optics Design Study

SALT-3120AE0005 Camera/Collimator Optics Specification

SALT-3150AA0001 Slitmask Requirements and Fabrication Document

SALT-3180AA0001 Etalon and Filter Trade Study

SALT-3190AA0001 Detector Subsystem Design Study

A description of the alignment tolerance and thermal and opto-mechanical design may be foundin

SALT-3125AE0001 Camera and Collimator Mechanical Tolerancing

SALT-3125AE0002 CDR Camera and Collimator Conceptual Design

Detailed specifications for optical components may be found in the following documents

SALT-3120AS0006 Folding Flat Mirror Specification

SALT-3120AS0007 VPH Grating Substrate Specification

SALT-3120AS0008 Calcium Fluoride Blank Specification

SALT-3120AS0009 Fused Quartz/ Fused Silica Optical Blanks Specification

SALT-3120AS0010 Sodium Chloride Optical Blanks Specification

SALT-3120AS0011 Slit Viewer Optics Specification

SALT-3120AS0012 Spectrograph Lens Fabrication Specification

SALT-3120AS0013 Waveplate Specification

SALT-3120AS0014 Polarizing Beamsplitter Specification

SALT-3120AS0015 VPH Grating Specification

SALT-3120AS0016 Anti-reflection Coatings Specification

2 Design Procedure

The optical design was modeled using the ZEMAX optical CAD package, version 10 - EE.

2.1 SALT Telescope model

The input for the optical system started with a ZEMAX model of the telescope and Spherical

Page 4: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 2

Aberration Corrector (“SAC”) provided by the SALT project. The design described here uses theSAC specification for an 11m telescope pupil at F/4.2, and an eight arcmin field of view. Theprescription is given in the document "SALT Optical Model", SALT-3300AS0001, 13 June2002. The telescope focal plane is flat, and the entrance pupil is 586 mm from the focal plane. Since the spectrograph is designed to be able to use slits that are smaller than the seeing disk, itwas designed for best imaging of the slit, not the sky, so that the aberrations of the SAC weredeliberately ignored. To facilitate this, a “perfect” telescope/ SAC model was constructed usingZEMAX paraxial elements having the same effective focal length, F/ratio, and entrance pupilsize and position as the actual SAC, but with perfect imaging. It was also important to model theSAC vignetting of the marginal rays as a function of field angle, since this fortuitously reduceselement diameters and reduces the “waist” of the collimated beam. This was done usingZEMAX “vignetting factors”. They had to be adjusted by hand to match the actual SAC pupil,since the automatic ZEMAX vignetting factor algorithm fails for the complex SAC pupil.

2.2 Design goals

Overall scientific goals for PFIS are described in "PFIS Instrument Description", SALT-3170AE0001. The rationale behind the optical design goals listed therein is described below

• Coverage 320 - 900 nm. Maintain simultaneous IR beam (850 nm - 1.7:) upgrade possibility.

There was a strong desire within the consortium to have coverage down to the atmospheric limitat the Prime Focus. At the same time, there is a strong desire for a near IR instrument. Givenspace and weight constraints, these would have to have a common collimator. This seemsfeasible, if one uses very broadband (possibly Solgel) coatings in the common optics (seeCoatings below). The IR beam should be sub-thermal (wavelengths less than 1.7:) because thetelescope is not optimized to minimize thermal emissivity. The visible- IR break should be about850 nm since that is where the efficiency curves of CCD and HgCdTe IR detectors cross. Sincethere is no room for a third beam, the visible beam should cover all wavelengths below 850 nm. The original specification called for 320 - 850 nm for the visible beam, This was stretched to900 nm to cover the CaII triplet and to allow for some overlap with the future IR beam. TheUNC SOAR spectrograph covers 320 - 850 nm with conventional coatings, but requires NaClelements for good color correction. This spectrograph was used as a starting point for the visiblebeam optics.

• All- transmission optics for high efficiency and compactness;

The highest possible transmission is a general goal for this instrument. The gain in compactnesscomes from avoiding the wasted collimated beam space required with a reflective collimator, andfrom the use of transmission gratings. Also, a reflective camera would introduce vignetting afterthe polarizing beam-splitter, which greatly compromises polarimetric precision.

• A maximum of one asphere. The original PDR design called for all spherical surfaces to reducerisk and cost. This goal has since been revisited, resulting in a single asphere surface at theentrance to the camera. The cost/risk tradeoff is described below.

• UV Crystals and fused silica only.

Page 5: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 3

This is required for good UV throughput down to 320 nm. The current model uses only fusedsilica, fused quartz, CaF2, and NaCl.

• Beam size 150 mm, the maximum for practical Fabry-Perot etalons.

A major goal of the instrument is the highest possible first-order spectral resolution. For agrating spectrograph at Littrow, the resolution in first order is given by

R = (2 tan 2i / 2s) (d/D)

where 2i is the grating tilt, 2s is the slit width, d is the beam diameter, and D is the primarymirror diameter. With D fixed, the maximum grating tilt fixed by mechanical constraints, and theminimum slit width set by the seeing, the only free parameter is the beam diameter. (Similararguments apply to Fabry-Perot, where d/D is the parameter which fixes the angular size of the"bullseye", where the wavelength is constant to within the etalon resolution). The practicallimiting beam diameter is set by the maximum diameter available etalons of 150 mm. We havechosen the beam to be 150 mm, so that there is some vignetting at the edge of the field for theFabry-Perot. VPH gratings are available in larger sizes, so that in grating mode there will be novignetting.

• Images < 0.40 arcsec in the dispersion direction over the full wavelength range.

The specification is as follows:

< 0.4 arcsec RMS width in dispersion direction for field angles < 3 arcmin

< 0.5 arcsec RMS width in dispersion direction for field angles 3-4 arcmin

< 0.55 arcsec RMS diameter for 340 - 850 nm on axis

< 0.65 arcsec RMS diameter for 320 - 340 and field angles 3-4 arcmin

The tightest imaging requirements are in spectroscopic mode with a reduced slit, so the RMSwidth is a more appropriate specification. Since the field of view in spectroscopic mode is notsymmetric, some astigmatism can be tolerated perpendicular to the dispersion with this methodof optimization. For grating spectroscopy with a 0.65 arcsec slit, the slit image is degraded by nomore than 17% for field angles < 3 arcmin, and 26% for field angles 3-4 arcmin. (This slit, thenarrowest envisioned, transmits 50% of the telescope image in the best seeing, and 40% inmedian seeing). The RMS diameter specification is appropriate for imaging without a slit. Themedian SALT seeing has RMS diameter of 1.2 arcsec (at 37° zenith angle, with 0.6 arcsectelescope images). For imaging, the median image is degraded by no more than 10% (14% at320 nm). These specifications include the optics as designed (including thermal effects in therange -5 - 20 deg C), plus manufacturing errors and alignment errors. The specification ismonochromatic: the design will allows lateral color (about 1.6 arcsec from 320 to 900 nm), sinceit is assumed that broadband imaging will be performed by the SALT scientific grade acquisitioncamera.

The original goal called for these imaging specifications to be met with no refocus betweengrating configurations. This has been judged to be unnecessarily strict, since there will be a

Page 6: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 4

Figure 1. Thermal indexcoefficient

Material Fused Silica Fused Quartz CaF2 NaCl

VIS NIR

Formula Sellmeier Sellmeier Schott Schott

K1/A0 0.6961663 0.917032891 2.038847 2.330048

L1/A1 0.004679148 0.006098740 -3.232100×10-3 -9.408556×10-4

K2/A2 0.4079426 0.187250355 6.156896×10-3 1.806310×10-2

L2/A3 0.01351206 0.017132580 5.661271×10-5 3.505261×10-4

K3/A4 0.8974794 0.803245003 -4.095144×10-9 2.603943×10-6

L3/A5 0.97934 88.4236969 2.240656×10-8 3.533943×10-7

nd 1.458464 1.458562 1.433891 1.544369

vd 67.82 67.69 95.04 42.82

D0 2.4882×10-5 2.5435×10-5 -2.978×10-5 -2.521×10-5 -8.4428×10-5

E0 3.7588×10-7 5.4896×10-7 5.2587×10-7 -3.339×10-6 1.9989×10-6

8tk 0.13333 0.0 0.1105 0.0 0.14582

Table 1. Index of refraction coefficients

camera focus mechanism which will be required tocompensate for filter thickness differences anduncompensated thermal effects. The current design has a200: refocus range.

2.3 Optical Indices

The optical design is very sensitive to the assumed opticalindices and their variation with temperature. Fortunately, forthe chosen crystals, the indices do not vary appreciably withmanufacturer. Unfortunately, there is disagreement in theliterature as to what the indices are. The following indiceswere chosen based on consultation with Harland Epps andDarragh O'Donoghue. Table 1 lists the adopted coefficientsfor n(8) and dn(8)/dT. The formulae are as follows:

(Sellmeier) n2(8) - 1 = K1 82 / (82 - L1) + K2 8

2 / (82 - L2) +K3 8

2 / (82 - L3)

(Schott) n2(8) = A0 + A1 82 + A2 / 8

2 + A3 / 84 + A4 / 8

6

+ A5 / 88

dn(8)/dT = [ D0 + E0 / (82 - 8tk

2) ] (n2 - 1)/(2n)

The dn/dT representation is that used by ZEMAX. Thecoefficients were derived by fitting the quantity D(8) = 2n/(n2

- 1) dn/dT to data from the literature (Figure 1).

Page 7: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 5

Material FusedSil

CaF2 NaCl

)nd ×106 81 51 114

)vd 0.07 0.11 0.05

Table 2. Index of refraction sensitivity

A tolerance analysis was performed on the current design to determine the accuracy with whichthe indices must be known. Table 2 lists )nd and )vd, the error in the index and the Abbe'number at 586 nm that results in an rms image size degradation of 1%.

2.3.1 Fused Silica and Fused Quartz

The Sellmeier fit for fused silica was taken fromMalitson (1965, JOSA 55,10). It is valid for 0.21 -3.71 : at 20°C. Comparing the adopted indiceswith those used by Epps (private communication)for the SOAR Goodman spectrograph, we find()nd×106, )vd) = (1, 0.04). Malitson (1965) alsocompared four samples of Corning 7940, and found(10 - 30, 0.04). The data used for the dn/dT fit isfrom the manufacturers data sheet for Corning 7980 fused silica. Comparing the resulting indexat 0°C with the Epps 0°C indices, we find (5, 0.08). The uncertainty in index and dispersion thusseems acceptable.

The Sellmeier fit for fused quartz is a ZEMAX fit of index data for Heraeus Infrasil taken fromthe manufacturer's specification. It is valid for 274nm - 1.7: at 20°C. The data for the dn/dT fitwas taken from the same manufacturers specification, for 238 - 643 nm, extrapolated to 1.7 : Both fused silica and fused quartz have a rather small thermal variation of the index of refraction,so the optical design is insensitive to this data.

2.3.2 Calcium Fluoride

For CaF2 we have adopted the Schott fit used by Epps for SOAR Goodman at 20°C. The dataused for the dn/dT fit is from Malitson (1963, Appl Opt 2, 1103). It was fit in two pieces, VIS:0.297 - 0.89 : and NIR: 0.89 - 1.7:, because the ZEMAX formula cannot represent data with anextremum. The PFIS model uses the VIS formula for the Visible beam and the NIR formula forthe NIR beam. Malitson also gives an index fit for 24 °C; if this is corrected with the VIS dn/dTto 20 °C, we find ()n×106, )vd) = (0, 0.01). Similarly comparing with the Epps 0°C indices, wefind (4, 0.01). The CaF2 indices appear to be well in hand.

2.3.3 Sodium Chloride

NaCl is the most problematic material. We have again chosen the Schott fit used by Epps forSOAR Goodman at 20°C. This is based on a re-analysis of data in Li (1976, J. Phys. Chem. Ref.Data, 5, 329). These indices have been used successfully in previous Epps designs using NaCl. Li (1976) gives a fit which differs from the Epps indices by ()n×106, )vd) = (-155, 0.04), anunacceptable difference. The data in Li is very heterogeneous in quality and temperature, so acareful choice of data and temperature correction is important, especially considering the largedn/dT for NaCl. The dn/dT adopted is based on data in Feldman (1978, NBS Technical Note#993, p50) for 0.46 - 3.39:, and Epps (private communication) for 0.26 - 0.4:. This gives arather larger value for dn/dT than data in Li (D0 ~ 84 vs 76), which may account for the

Page 8: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 6

Figure 2. PFIS Optical Layout. () = removable.

discrepancy. Applying our dn/dT fit to arrive at indices for 0°C, and comparing with the Epps0°C indices gives an error of ()n×106, )vd) = (64, 0.01), which is acceptable.

3 Optics Subsystems

The current PFIS optical design isshown in Figure 2. Here we describethe design proceeding from thedetector backwards along the beam.

3.1 Detector

The detector geometry was chosenbased on issues of sampling and CCDavailability. The fastest affordablerefractive UV camera was judged to beF/2.2. For an 11m telescope, the 8arcmin SALT field is then 56 mmacross, which almost fills the longdimension (61 mm) of the mostcommon modern 2048 x 4096 , 13.5 -15 micron pixel CCDs. A mosaic oftwo of these chips is often chosen, butanother factor to consider is thenumber of spectral resolution elements. For a slit matching the median SALTimages of 1.2 arcsec, there are 400spatial resolution elements, and asquare array would provide only 435resolution elements for gratingspectroscopy. This is well short of thenumber of spectral resolution elementson large telescope slit spectrographs,which are in the range 600 -1200. Since simultaneous resolution elementsdefine the multiplex advantage of aspectrograph, it was felt that thisnumber should be competitive, so thata 3 mosaic, with the long dimension inthe dispersion direction, was baselined.

The baseline CCD chosen (see Detector Subsystem Design Study) is a mosaic of three Marconi/EEV 42-82 (2048×4096 15: pixels) chips, for a total of 6144×4096 pixels (95 ×61 mm). For anF/2.2 camera, the pixels are 0.13 arcsec, so that the 1.2 arcsec seeing disk is critically sampled for2×2 binning, and a 0.5 arcsec slit is critically sampled for unbinned readout. The number of 1.2arcsec spectral resolution elements is 655.

Page 9: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 7

Figure 3. PFIS camera optics

3.2 Camera

For VPH observations, the camera is articulated about the grating axis as the grating is tilted toallow tuning of the grating blaze while the wavelength range is shifted. Since the VPH blazepeak is at Littrow, the camera articulating angle is twice the grating tilt. A maximum grating tiltof 50° has been chosen (from mechanical and grating efficiency considerations), leading to amaximum camera articulation of 100°. The camera aperture has been sized to accept this rangeof configurations with no vignetting for any wavelength on axis, and < 5% vignetting for theextreme wavelengths at the extreme articulation angles 4 arcmin off axis.

An unavoidable resultof the length of thedetector is a ratherlarge camera field ofview, 16°. Typicalspectrograph camerasare in the range 10 -16°, so this is a designdriver. The largewavelength range isanother design driver. The starting designtook elements from theSOAR Goodmanspectrograph ofHarland Epps; this camera has a similar wavelength range, but only a 10.4° FOV. The use ofNaCl triplets is notable in this design. Another source was the Epps camera for the Keck LRIS-B, which has a smaller wavelength range but an FOV of 14.6°. The PFIS design has 9 elementsin 4 groups (Figure 3). The first group is a large fused silica /CaF2 quadruplet. The first surfaceis an asphere, on fused silica. The original camera design was all spherical, having 12 elements in5 groups, including 2 NaCl triplets, and starting with a CaF2/ silica doublet Putting an asphereon the initial CaF2 surface eliminated 4 elements, including one NaCl triplet. However, it wasjudged to be too risky to put an asphere on CaF2, and an extra silica element was added to thegroup to take the asphere. The resulting asphere is well within the experience of astronomicalspectrographs: it has a Maximum Aspheric Deviation ("MAD") of 276 : over the clear aperture,compared to a MAD of >1 mm in DEIMOS. An order of magnitude estimate from Hilyardshows that the extra cost of the asphere is offset by cost of the elements saved. The reduction ofair-glass interfaces and the reduced risk resulting from removal of a NaCl triplet weighs in favorof the asphere design. The quadruplet is followed by a CaF2 singlet, which provides much of thepower, and a Silica/ NaCl/ silica triplet, which provides additional power with color correction. The field flattener is fused silica. The flattener is also the detector cryostat window.

A filter magazine for Fabry-Perot interference filters and order blockers is located just before thedetector. The choice for filter location is between this position and the collimated beam. Thelatter would result in very large, expensive, heavy filters which would require very good optical

Page 10: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 8

Filter 81 > 82

Fused Silica (Atmos) 310 < 620

Hoya UV-32 330 620 - 640

Hoya UV-34 350 640 - 680

Hoya UV-38 390 680 - 760

Schott GG455 465 760 - 900

Table 3. Order Blocking Filters

l/mm 8 (nm) D (:) dn

300 400 - 900 ruled

900 320 - 900 3.0 0.1

1300 400 - 900 3.0 0.1

1800 480 - 900 3.5 0.1

2300 450 - 700 3.0 0.1

3000 320 - 520 2.0 0.1

Table 4. Grating Complement

quality, so this was ruled out. The disadvantageof locating the filters close to the detector is thepossibility of out-of focus ghosts fromreflections off the CCD surface, and the effectof the fast beam on the interference filters. Thefilter magazine will have a capacity of 20 filters;since a minimum of 5 order blockers for gratingspectroscopy and 30 interference filters forFabry-Perot will be required ultimately, thefilter complement will have to be adjusted basedon the current observation al programs. Table 3 gives a list of order separators, based on thegrating complement described below.

The camera was first optimized using the perfect SALT telescope model described above, plus aperfect collimator yielding the 150mm beam and the desired pupil placement. The field flattenerback focal distance was constrained to be greater than 9 mm, and the space between the lastcamera element and the field flattener was constrained to be big enough for the filters. Theoverall length was constrained to be < 625 mm, required by the overall envelope of theinstrument. The merit function consisted of image rms in the dispersion direction for field anglesof 0, 2, 3, and 4 arcmin, plus image diameter at the same field angles, but with a relative weightof 1/3. The dominant aberration is longitudinal color, so the system was optimized with thecamera in three configurations of the lowest-dispersion VPH grating (900 l/mm), with the cameraat Littrow and the grating tilted at 12.3°, 14.5° and 19.95°: this gives wavelength coverage,respectively, of 320 - 640 nm, 400 - 700 nm, and 610 - 900 nm. These are the most demandingconfigurations of the spectrograph; the higher dispersion configurations have better imagingbecause of the smaller range of wavelengths. The goal for the camera design images was adispersion -direction RMS width < 0.2 arcsec (23 microns), 2/3 the goal of the total system rmsof 0.3 arcsec (35 microns). This leaves 25% margin for fabrication and alignment errors.Roughly one hundred lens configurations were evaluated; no configuration without NaCl wasfound to be acceptable.

3.3 Collimated Beam

The dispersors are located in the 150mm diameter, 350 mm long collimated beam. Thecollimated beam will accommodate either two Fabry-Perot etalons (each 250 mm in diameter and155 mm thick), or one rotatable VPH grating. The etalon positions straddle the VPH positionnear the pupil.

The optical properties of the Fabry-Perot etalons and the VPHgratings is described in detail in their Trade Study documents.

As a summary, the VPH grating complement is listed in table 4and the efficiency vs resolution and wavelength plot is shownin figure 4. The VPH hologram thickness D and indexmodulation dn are listed in the table. In the figure, resolutionis for a 1.25 arcsec slit. The detector coverage is shown as a

Page 11: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 9

Figure 4. VPH Grating Efficiency vs resolution and wavelength

heavy tilted line atgrating tilts of 20, 30,40 and 50°. Thecontours showefficiency 50, 70 and90%. Commonastrophysical lines areshown as dashedlines. The gratingroles are as follows:

• The 300 l/mmgrating is aconventional ruled300 l/mmtransmission gratingfrom RichardsonGrating Lab. The freespectral range of thisgrating fills only 1/3of the detector, so it isintended for very low-resolution (R ~ 400) multi-object surveys using the entire field of view.

• The 900 l/mm VPH grating covers 320 - 620 nm in the UV (R ~ 900) and 610 - 900 nm in thered (R ~ 1600), with the free spectral range that covers the chip. This is a low resolution single-object or moderate field multi-object grating.

• The 1300 l/mm VPH has highest efficiency from 450 - 700 nm (R ~ 2000), and covers both H"and H$ for redshifts up to 0.15.

• The 1800, 2300, and 300 l/mm gratings provide their highest efficiency from 500- 700, 450 -600, and 320 - 500 nm, respectively (R ~ 3000 - 4000) and have their highest resolution atgrating tilt 50°, covering the CaII triplet, H", and H$/ [OIII]/ MgI lines, respectively.

A standard Prontor 150mm shutter is just before the dispersors in the collimated beam. Theshutter cannot be placed in the collimator because of the desire for future simultaneous visible/IR observing. It could have gone near the detector, allowing a smaller shutter. However, theposition in the collimated beam is advantageous because very short exposures such as would benecessary for flux calibration would have no field angle exposure correction. It is true that theshutter is not at the pupil, and requires an aperture of 161mm, which is larger than anycommercially available shutter. However, we have determined that a standard Prontor 150 mmshutter may be modified to apertures of 160 - 165 mm without affecting reliability.

3.4 Polarimetric Optics

The polarimetric optics utilizes a "wide-field" design. As described in the Polarimetric Optics

Page 12: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 10

Design Document, such a system has a polarizing beamsplitter in the collimated beam whichtakes the central half of the field and splits it into two orthogonally polarized fields, the"ordinary" and "extraordinary" beams. A polarization modulator preceding the beamsplittermodulates the polarization state with time, and the difference between the intensities of the O andE images as a function of time yields the polarization.

The beamsplitter, an array of calcite Wollaston prisms, may be inserted just before the camera inthe collimated beam. This ensures that there is no vignetting of the split beam by the dispersors,especially the Fabry-Perot etalons, which would compromise the polarimetric precision. Also, byplacing the beamsplitter after the etalons, both the E and O fields have the same wavelengthgradient in Fabry-Perot mode, enabling direct differencing of the two fields. Since it has nopower and is in the collimated beam, a perfect beamsplitter would have no effect on the imagequality of the optical system. However, the prism surfaces must be flat and precisely aligned toavoid excessive degradation of the imaging. Also, there is a concern for ghost images arising indouble reflections within the prism and between the prism and the detector.

The modulator consists of two rotating superachromatic waveplates near the beam waist in thecollimator. The modulator should be ahead of any optical elements with polarization sensitivity,like the fold mirror and the dispersors. The collimator beam waist was chosen because itminimizes the waveplate size, a serious cost driver. This puts the modulator in a diverging beam,which must be considered in the modulator design (see Design Document). Also, when themodulator (a plane-parallel element) is removed, it must be replaced by a plane-parallel elementof the same material (fused quartz) and thickness (16 mm) as the waveplate to match thesubstantial focus change and spherical aberration of the waveplate. The collimator must bedesigned to include this plate. The first waveplate is a 100mm diameter halfwave plate (thelargest waveplate available). This is followed either by a fused quartz blank, for linearspectropolarimetry covering a 4×7.2 arcmin field of view, or by a single 60mm quarterwaveplate, providing a 3.0 arcmin diameter unvignetted field for circular or all-Stokes polarimetry. When neither waveplate is inserted, a single double thickness blank provides focuscompensation. Thus the only compromise on non-polarimetric modes imposed by thepolarization capability is the introduction of two air-glass interfaces, which may be anti-reflection coated. Because the waveplates (and compensator) are located so close to thetelescope focal plane, the flatness requirements on the elements are quite loose (5 8).

All elements between the waveplate and the polarizing beam-splitter must have low stress-birefringence to avoid depolarizing the E and O polarization destined for the beamsplitter. Atotal effective birefringence budget in the E-O axis of < 20 nm has been established for thisoptical train. This limits the depolarization to <1, 2, and 4% at 586, 430, and 320 nm,respectively. Fused Silica, fused quartz, and CaF2 with stress birefringence of < 2 nm/cm andNaCl < 10 nm/cm will meet this requirement, if the stress is not aligned in the elements. Annealing will be required on all these elements to meet this requirement.

3.5 Collimator

Compared to refractive collimators on other large telescopes, the requirements for the PFIScollimator are unusual in three ways:

Page 13: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 11

Figure 5. PFIS collimator optics

Figure 6. Future NIR beam model

• High speed (F/4.2)

• Closeness of the entrance pupil (586 mm)

• Very large wavelength range; 320 nm - 1.7 : to accommodate simultaneous visible - near IRobservations

The wavelength rangeand the speed againappear to require use ofNaCl. Fused quartz isused in place of fusedsilica in the commonNIR/ visible optics, toavoid the 1.39:absorption band in fusedsilica. The design (Figure5: this shows the visiblebeam only, without fold/dichroic) starts with adoublet field lens, fusedquartz/ CaF2, placed close to the focal plane. This rather strong field lens is due to the closenessof the entrance pupil; the requirement of good imaging down to 320 nm drives the use of adoublet rather than a singlet. This is followed by a space for the waveplates (a plane parallelfused quartz element) and a negative fused quartz element, which corrects the field curvatureintroduced by the field lens. This arrangement is required in imaging grating spectrographs withlarge field of view perpendicular to the dispersion, because any residual field curvature due to thecollimator is seen perpendicular to the dispersion but not along the dispersion, so it cannot becorrected by a spherical element in the camera. The negative element is followed by a singletfused quartz element,and the main triplet,CaF2 / NaCl/ CaF2. Thefinal element in thevisible beam collimatoris a CaF2/ fused silicadoublet.

The collimator wasdesigned to permit thefuture addition of thenear-IR beam. Theoptical prescriptionmust be optimized toallow good collimationover the entire 320 nm -

Page 14: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 12

Figure 7. Slitviewing optics

1.7: wavelength range, and the pupil must be placed such that there is adequate room in thecollimated NIR beam. Figure 6 illustrates the NIR beam model that was used. A fold mirrorbefore the final doublet will become the visible/ IR dichroic when the IR beam is added. Aseparate NIR collimator doublet (CaF2/ Schott K7) follows the dichroic. A fold follows this tobring the NIR beam within the instrument envelope. The visible camera is shown replicated onthe NIR side; it was used to define the NIR support structure, since the NIR camera is as yetundefined.

The collimator was optimized using the perfect SALT model and a perfect 330mm focal lengthcamera for each beam. The merit function minimizes the rms image diameter simultaneously forwavelength 320, 400, 700, and 900 nm in the visible beam, and 0.9, 1.2, 1.4, and 1.7 : in theNIR beam. The exit pupil was constrained to be 360 mm from the last element of the collimator. This is farther than is required by the visible beam, but allows for the extra 200mm required inthe NIR beam by the NIR fold flat. Final constraints set the minimum separation of the field lensfrom the focal plane to be 10 mm (to allow for the slitmasks), provided minimum clearance forthe waveplates, and constrained the overall length to be less than 825 mm.

The goal for the collimator images was an rms diameter less than 0.15 arcsec (8 microns), 1/2 thegoal of the total system rms of 30 :. Again, a large number of lens configurations wereevaluated; no configuration without NaCl was found to be acceptable.

3.6 Slitmask / Slitviewer

As described in the Slitmask Trade Study Document,multi-object spectroscopy is facilitated throughcustom slitmasks loaded into a slitmask magazine,selected and inserted at the focal plane. Two kinds ofslitmasks are envisioned, thin flat carbon fiber maskswhich are placed coincident with the focal plane, sothat they are useful over the full field of view, andtilted metallic slits for work on-axis in the directionof dispersion. The former, while more flexible infield, present well-known difficulties with acquisitionsince the slit cannot be viewed directly. Theacquisition scenario for the carbon fiber slitmasksinvolves centering with the SALTICAM acquisitioncamera, followed by imaging the field with the PFIScamera unarticulated, comparing this with acalibration lamp image taken through the mask,commanding a telescope offset to align these images,finally followed by science observations with thecamera articulated, guided by off-axis probes. Incontrast, the tilted metallic masks are provided toallow direct viewing of the slit using relay optics tothe acquisition camera. We estimate that at least 50% of PFIS programs will be single object or

Page 15: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 13

Figure 8. Long slit and holder

longslit observations which are compatible with conventional slitviewing. This will allow veryaccurate acquisition/ centering of difficult objects, and even guiding off lost light on the slit. Itis felt that these programs should not be unnecessarily subject to the lost time and risk of theblind pointing scenario. The slit viewing optics, based on the Offner relay design, is shown inFigure 7. The optics is fixed: whenever the acquisition camera fold mirror is out, a view of theslit is presented to the acquisition camera if PFIS isusing a tilted slit. The slitviewer imaging is betterthan 0.1 arcsec, and the field of view is 2 x 8 arcmin,unvignetted in the center, and 50% vignetted at theedges. These optics are in the crowded volume infront of the focal plane, and so must become part ofthe telescope payload.

The multislit slitmasks will be 100: thick carbon-epoxy (see slitmask mechanism specification), andthe longslits will be either nickel-plated carbon-epoxy, nickel-plated aluminum, or nickel (Figure 8). All are to be cut by a commercial laser micromachining facility, delivered with the instrument.

7.5 ± 12.5 C Visible NIR

focus(±:)

scale(±:)

rms(+:)

focus(±:)

scale(±:)

rms(+:)

Collimator 227 -30 3 113 -22 1.5

Compensated 50 -4 0.2 -55 5 0.3

Camera 211 0 2

Compensated -50 -4 1.4

System -Comp 0 -8 1.7

Table 5. Thermal Effects on Optical Performance

4 Thermal Design

Thermal effects on optical performance are very important for SALT/ PFIS because of the largetemperature range experienced at the prime focus, and because the crystals used in the design forgood UV performance have large thermo-mechanical and thermo-optical coefficients. Normaloperating conditions are -5 - 20°C with a maximum rate of change of 1.5°C / hour; marginaloperating conditions (20% reduced imaging performance) are -10 - 25°C (2°C/ hr), and survivalconditions are -20 - 45°C. An exploration of the optical performance over the operatingconditions shows that 90% of the effects on the imaging are due to the thermal coefficient of theindex of refraction of CaF2 and (especially) NaCl. These affect the focus, the focal plane scale,and the image size. These are all serious enough to demand a passive thermal compensationsystem to adjust the element positions with temperature.

Table 5 lists the uncompensated (white) and compensated (grey) values for the shift of the focalposition, the focal plane scale (listed as the position of an image 4 arcmin off axis), and the

Page 16: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 14

Figure 9 Thermal Compensation System

image degradation, over the nominal operatingtemperature range of -5 - 20 °C. The total focal planeshift is so large (34: /deg) that at the maximumtemperature rate of change the spot size would bedegraded by 25: over the longest expected exposure ofone hour. This means that adjusting the focus betweenobservations is not sufficient, and passive focuscorrection is required. Also, the focal plane scalechange and image degradation for the camera over theoperating range are large and require compensation,particularly since it is desired that flat-field calibrationsbe performed during the day, when the temperature istypically 7°C warmer. Separate passive compensationschemes consisting of shifts of internal groups wereevaluated for the camera and collimator.

For the collimator the focal plane and the final doubletwere assumed to be fixed (they are attached to the maininvar structure), and motions of the two singlets andtriplet were considered. The focus shifts for the visibleand NIR beams were quite different, and could not besimultaneously compensated, so a compromise wasadopted that split the uncompensated focus motionbetween the two beams. This reduces the collimatorfocal shift by a factor of 5 in the visible beam and afactor of 2 in the NIR beam. Motion of no single groupnor combination of two groups was sufficient to simultaneously compensate the remaining focusshift, focal scale change, and image degradation. However, a scheme in which the first singletand triplet were treated as a one group and the singlet between them as a second group gave goodresults. A schematic implementation of this scheme is shown in figure 9. Delrin, with a largethermal coefficient of expansion of 125×10-6/°C, is used to shift the groups passively. One delrinspacer moves the first singlet and triplet relative to the rest of the collimator, and a second movesthe second singlet relative to this group.

The camera has no focal plane scale shift, appreciable image degradation, and a large focus shiftwith temperature. The passive thermal compensation system (figure 9) will shift the combinationof the CaF2 singlet and the NaCl triplet relative to the first camera group, and the camera housingwill be aluminum, which aids in the compensation. This removes all of the camera focus shift,the remaining collimator focus shift, and improves imaging. The entire system has been re-optimized for the temperature range -5 - 20 °C.

In addition to the passive focus compensation, one must allow for an active focus adjustment. This must allow for an imperfect passive adjustment, for the different focus between thespectrograph configurations, and for residual differences in filter optical paths. Allowing 10% ofthe thermal travel (±65:), a configuration range of ±100:, and a filter thickness error of ±50:,

Page 17: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 15

Figure 10. Focus position vs configuration andwavelength

we require ±235: for the active focus. Thiscannot be done in the collimator, since thevisible and NIR beams need to be separatelyfocused. The only collimator elements whichare unique to a beam are the final doublets, andtheir focus sensitivity is much too low. Thisleaves the camera. It was found that focusingthe CaF2 singlet/ NaCl triplet group (the samegroup as the passive compensator) is best. Focusing the detector dewar would involve aheavy stage to cantilever the dewar withrequired stiffness. The singlet/triplet group hasgood focus sensitivity, and the end of the delrinspacer used for passive thermal compensationis conveniently available for a focus actuator. About ±250: motion of this group will suffice. The focus as a function of central wavelength forthe five VPH gratings and for the Fabry-Perot is shown in figure 10.

5 Imaging Performance

5.1 Image Quality Budget

The image quality budget was based on a merit function similar to that used for optimizing thecamera: RMS-Y (dispersion direction) + 1/3 RMS-X, at four wavelengths, fields 0, 2, 3, and 4arcmin, and three configurations of the 900 l/mm grating.

State Design(7.5 C)

Thermal(20 C)

Flexure Fabrication Alignment Pol Total

50% 90% 50% 90% 50% 90%

MF 1312 1327 1334 1397 1384 1447 1357 1505 1631

% incr 1.1 3 1.7 6.5 5.5 10.3 3.4 14.7 24.3

% budget 1.5 3.5 6 10 4 25

Table 6. Image Quality Budget

Table 6 shows the merit function value for the design at optimum temperature (7.5°C), at 20°C,and for various degradations of the 20°C design by flexure, fabrication errors, alignment errors,and insertion of the polarizing beamsplitter. The computed percent degradation of the meritfunction is compared to the budgeted degradation. A total 25% degradation of the image isallowed due to all such effects.

The flexure image degradation value is based on the root-sum-square of the maximum flexureduring a track in the dispersion direction (0.1 arcsec) and the specification for the on-axis RMSwidth in the dispersion direction (0.4 arcsec). The fabrication value is based on a Monte Carlotolerance analysis for manufacturing errors specified in table 7:

Page 18: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 16

Specification Value

Surface Radii ±0.1% Deviations 1/2 8 at 630 nmCenter thickness ±0.075 mm

Wedge< 30 : edge thicknessdifference

Table 7. Lens Fabrication Specifications

Values are given for the median and the 90% point in the merit function distribution. Thealignment value is based on a Monte Carlo analysis described in the optomechanical designdocument SALT-3125AE0001 "CDR Camera and Collimator Conceptual Design", again for the50 and 90% merit function distribution points. The polarimetric mode image degradation isbased on a 148 wavefront distortion specification for the alignment of the polarizingbeamsplitter. The %degradation values were added (more conservative than root-squaredsummed) to give the total values at the right. We see that the 90% value is just within budget,and the 50% values are well within budget. We feel that the 50% values are appropriate for bothfabrication and alignment: for fabrication most of the degradation is due to radius errors, whichcan be compensated using the post-fabrication pickup surfaces and adjustment of spacings; andthe experience of the optomechanical subcontractor (Alan Schier of Pilot Group) is that thealignment process almost always comes out near or below the 50% point, never at the 90% point.

5.2 RMS Image quality vs configuration

The merit function gives an overall sense of the imaging of the most difficult configurations. Toget a feel for the imaging in all configurations and at a large range of wavelengths and fieldangles, we have computed (Figure 11) the RMS dispersion-direction width and the RMSdiameter as function of wavelength for monochromatic imaging (ie Fabry-Perot imaging) and forthe full range of grating tilts for all 5 VPH gratings, on axis and 4 arcmin off-axis.

The temperature is 20 °C, at the warm limit of the nominal operating range. The panels on theleft give the RMS image width at the detector of an infinitely thin slit oriented parallel to thespectrograph slit. The panels on the right give the RMS diameter of the spots for the sameconfigurations. The bottom right panel shows the monochromatic imaging performance,appropriate to the Fabry-Perot mode of the instrument. The specification in the FPRD is shownas a thick grey line. Note that the imaging is progressively better as the bandpass of theobservation decreases, consistent with the dominant aberration of longitudinal color.

Figure 12 shows the RMS-Y and RMS diameter image degradation due to thermal, fabricationerrors and alignment errors, for 320 and 630 nm, on axis and at 4 arcmin. The 320 nm, on-axis,RMS diameter is the worst case approach to the specification.

Page 19: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 17

Figure 11. Design Imaging Performance

Page 20: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 18

Figure 12. RMS image degradation (dashed: off-axis; thin: polarimetric). Grey: FPRD specification

Figure 13. PFIS Anti-reflection coatings

6 Throughput

6.1 Coatings

For the fold flat, we willuse the same LLNLmultilayer coating that isto be used by the SALTSAC. This has roughly 95% reflectivity 320 - 900nm. It is an extremelydurable coating.

We propose to use threetypes of anti-reflectioncoatings for the refractiveoptics, depending onplacement and exposure. Solgel is preferred. SolGel is a chemicalcoating producing a stack of 200D pure silicaspheres, having an effective index ofrefraction of 1.22. Used over a single layer ofMgF2 (n = 1.38), the single surfacereflection can be less than 1% from 320 -850 nm, degrading to no worse than 2% at1.7:. However, there are durability concernsfor SolGel coatings, particularly for thoseover MgF2. They are especially sensitive tooil drop contamination, which can quickly"wick" to damage a large portion of thesurface. We have therefore chosen to useSolGel only on surfaces which are internal tolens assemblies, four in the collimator andfour in the camera. A simple MgF2 coatingwill be used on exposed surfaces common tothe visible - NIR beams (6 surfaces), and aconventional multilayer optimized for 320 -900 nm will be used on exposed surfaces inthe collimator doublet and the camera (6surfaces). Figure 13 compares SolGel/ MgF2

with MgF2 and a conventional multilayer(designed by Spectrum Thin Films). Thebottom panel shows the predictedtransmission for the PFIS system using

Page 21: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 19

Solgel on internal surfaces.

6.2 Stray Light

The PFIS optical system is closed to external stray light- ie, the only light getting into the systemcomes through the focal plane slitmask. The chief stray light concern is therefore from ghostsoriginating from stray reflections off the optical surfaces. We have performed a ZEMAX "ghostfocus generator" analysis using double bounces off the refractive elements from the field lensthrough the detector. (The Feb 4, 2003 version of ZEMAX, which supports aspheric surfaces,was used). A ghost relative brightness b was computed b = R1 R2 (1.2 arcsec/d)2, where R1

and R2 are the reflectivity of the two surfaces, and d is the RMS diameter of the ghost image atthe detector. Table 8 lists the reflectivities used for the various surfaces. Table 9 lists thebrightest ghosts in order of decreasing brightness.

Surface Reflectivity

SolGel 0.005

Multilayer 0.01

MgF2 0.02

CaF2-Silica 0.000072

CaF2-NaCl 0.001360

Silica-NaCl 0.000806

CCD 0.1

Table 8. Reflectivities for Ghost Analysis

Surf1 Surf2 R1 R2 d/1.2" b/10-5 FPRD

CCD Pupil 0.1 0.01 1.0 97.12 <100

CCD Pupil 0.1 0.01 1.0 97.12 <100

Pupil Pupil 0.01 0.01 1.0 10.00 <100

CCD Fld Lens 0.1 0.02 6.6 4.60 <10

1/2 Wave 1/2 Wave 0.02 0.02 5.2 1.48 <10

1/4 Wave 1/4 Wave 0.02 0.02 5.3 1.44 <10

Pupil Fld Lens 0.01 0.02 6.8 0.43 <10

Pupil Fld Lens 0.01 0.02 6.8 0.43 <10

CCD Flattener 0.1 0.01 22.3 0.20 <10

CCD Fld Lens 0.1 0.02 36.2 0.15 <10

Flattener Flattener 0.01 0.01 9.2 0.12 <10

Cam Trip 1/4 Wave 0.005 0.02 9.3 0.12 <10

CCD 1/2 Wave 0.1 0.02 44.0 0.10 <10

Table 9. The brightest PFIS ghosts

This is a conservative estimate of ghost brightness, since it does not take into account vignettingof the ghost beam nor light losses in the lenses and dispersors as it takes an extra double passthrough the system. The first three ghosts are in-focus ghosts caused by plane surfaces at thepupil: the two surfaces of the Fabry-Perot etalons, the gratings, and the polarizing beamsplitter.

Page 22: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 20

Figure 14. SALT PFIS efficiency

Figure 15. On-sky sensitivity comparison - grating spectroscopy

these are unavoidable, and must be minimized by good coatings. The FPRD puts a limit onfocused ghosts of 10-3, and 10-4 for unfocused ghosts. The brightest predicted out-of-focus ghostresults from reflections off the CCD propagating all the way back to the first surface in thesystem, the field lens, and returning; this should be suppressed by the double pass light loss. Ithas been suggested that a similar ghost might result from reflections off the back of the slitmask. The fact that this has never been observed in slitmask multi-object spectrographs suggests thatcomplete double pass ghosts are not important.

6.3 Efficiency

The estimated spectrograph/ detector efficiencyis shown in Figure 14. This estimate uses themixed Sol-Gel/ MgF2 / Multilayer coatingsdescribed above, LLNL reflection efficiencyfor the fold mirror, the EEV/ Marconiquantum efficiency curve for the "astronomybroadband" coating and deep depletion silicon,and a peak VPH grating efficiency of 90%. The detector/ spectrograph efficiency is 60% atpeak and 28% at 3200 D.

Figure 15 shows a comparison of the predictedsensitivity on the sky for PFIS VPH gratingspectroscopy for low (R ~ 1200) and medium(R = 3000) resolution, compared to thepredicted sensitivities of slit spectrographs onother large telescopes. For all of thesepredictions, the resolution is for a slit width comparable to the mean seeing at the telescope, andslit losses are not included. For SALT, we assume 1.25 airmasses of extinction (ZD = 37°), an11m pupil with 17% obscuration, a mean 85% illumination of the primary for a ±6 degree track,and one aluminum and four LLNL reflections. We see that PFIS will be comparable to or better

Page 23: Southern African Large Telescope Prime Focus Imaging ... · Southern African Large Telescope Prime Focus Imaging Spectrograph Optics Design ... 3.4 Polarimetric Optics ... or Trade

PFIS Optics Design V2.21 Mar 10, 2003 21

than other instruments over the entire 3200 - 9000 D band, with best performance in the blue. Predictions for one 3000 sec observation in the V-band are S/N = 46 at R = 1200 and S/N = 28 atR = 3000 for a V = 22 magnitude object


Recommended