+ All Categories
Home > Documents > SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer...

SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer...

Date post: 29-Jun-2020
Category:
Upload: others
View: 4 times
Download: 1 times
Share this document with a friend
8
Overview The SPCM-AQR is a self-contained module that detects single photons of light over the 400nm to 1060nm wavelength range - a range and sensitivity that often outperforms a photo-multiplier tube. The SPCM-AQR-IX uses a unique silicon avalanche photodiode with a circular active area that achieves a peak photon detection efficiency of more than 65% at 650nm over a 180μm diameter. The photodiode is both thermoelectrically cooled and temperature controlled, ensuring stabilized performance despite ambient temperature changes. Count speeds exceeding 10 million counts per second (Mc/s) are achieved by the SPCM-AQR-1X module. There is a “dead time” of 50ns between pulses. As each photon is detected, a TTL pulse of 2.5 Volts (minimum) high in a 50 load and 35ns wide is output at the rear BNC connector. To avoid a degradation of the module linearity and stability, a case temperature between 5˚C and 40˚C should be maintained. Saturation The photon count decreases at higher incoming light levels. The count at which the output rate starts to decrease is called the saturation point. As an extreme example, if the module is exposed to intense light the count rate will fall to zero. Consequently, in certain applications, tests should be performed by the operator to ensure that a low count rate is not caused by detector saturation. Precautions should be taken to avoid any excessive light level that will damage the SPCM module. Applications LIDAR Photon correlation spectroscopy Astronomical observation Optical range finding Adaptive optics Ultra sensitive fluorescence Particle sizing Key Features and Benefits Peak photon detection efficiency at 650nm: 65% typical Active area: SPCM-AQR-IX: 175 m User friendly Gated input Single +5V supply www.optoelectronics.perkinelmer.com D A T A S H E E T BIOMEDICAL SOLUTIONS SPCM-AQR Single Photon Counting Module
Transcript
Page 1: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

Overview

The SPCM-AQR is a self-containedmodule that detects single photonsof light over the 400nm to 1060nmwavelength range - a range andsensitivity that often outperforms a photo-multiplier tube.

The SPCM-AQR-IX uses a uniquesilicon avalanche photodiode witha circular active area that achievesa peak photon detection efficiencyof more than 65% at 650nm over a 180µm diameter. The photodiodeis both thermoelectrically cooledand temperature controlled,ensuring stabilized performancedespite ambient temperaturechanges.

Count speeds exceeding 10 millioncounts per second (Mc/s) areachieved by the SPCM-AQR-1Xmodule. There is a “dead time” of50ns between pulses.

As each photon is detected, a TTLpulse of 2.5 Volts (minimum) highin a 50� load and 35ns wide isoutput at the rear BNC connector.To avoid a degradation of themodule linearity and stability, a case temperature between 5˚Cand 40˚C should be maintained.

Saturation

The photon count decreases athigher incoming light levels. The count at which the outputrate starts to decrease is calledthe saturation point. As an extremeexample, if the module is exposedto intense light the count rate will fall to zero.

Consequently, in certainapplications, tests should beperformed by the operator toensure that a low count rate is not caused by detector saturation.

Precautions should be taken toavoid any excessive light level thatwill damage the SPCM module.

Applications

LIDAR

Photon correlationspectroscopy

Astronomical observation

Optical range finding

Adaptive optics

Ultra sensitive fluorescence

Particle sizing

Key Features and Benefits

Peak photon detection efficiencyat 650nm: 65% typical

Active area: SPCM-AQR-IX:175 �m

User friendly

Gated input

Single +5V supply

www.optoelectronics.perkinelmer.com

DA

TA

SH

EE

T

BIO

ME

DIC

AL

SO

LU

TIO

NS

SPCM-AQRSingle Photon Counting Module

Page 2: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

SPCM-AQR

Fiber connection option

The SPCM-AQR-WX-FC has an“FC” fiber-optic receptacle pre-aligned to the optical detector.Optical fibers with an FC connectoron one end are available separately,(see Ordering Guide 2). Due to thewavelength dependence of thegraded index coupling lens, theoperating wavelength range must bespecified (see Ordering Guide 2).The photon detection efficiency ofconnectorized modules is about95% of that quoted for standardmodules.

Fiber shielding

When used with optical fibers, both the fiber itself and theconnector shrouds must becompletely opaque. Otherwise,stray light will increase the countrate. The SPCM-QCX pigtailsconform to this requirement (seeOrdering Guide 2).

Gating function

A gating function is provided witheach module – useful for viewing asignal that occurs only in a smalltimeframe window. Also, in someapplications the background lightflux is higher than the signal. Inthis case. the gating option could beused to improve the S/N ratio byopening a window only when thelight signal is present. The output ofthe module is disabled when a TTLlow level is applied to the modulegate input.

Light emission duringphoton detection

One peculiarity of silicon avalanchephotodiodes is that as an incomingphoton is detected, a small amountof light is emitted from theavalanche region. The emitted lighthas a broad spectral distribution. Inmost cases, this is not a problem.However, it can cause some confusion if another detector is

monitoring light, or if the opticalsystem is such that light emittedfrom the SPCM-AQR is reflectedback on itself. If these photonsreturn 35 ns after the initial event,they will be detected.

Safety

The SPCM-AQR contains a highvoltage power supply. All internalsettings are pre-set; there are nouser adjustments. Units thatappear defective or have sufferedmechanical damage should not beused because of possible electricalshorting of the high voltage powersupply.

Warranty

A standard 12-month warrantyfollowing shipment applies. Anywarranty is null and void if themodule case has been opened.

EDS warning

Modules should only be handled atan ESD-safe work station.

2

Page 3: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

3www.optoelectronics.perkinelmer.com

Table 1. Specifications SPCM-AQR-WX at 22˚C, all models, unless otherwise indicated

Parameter Minimum Typical Maximum Units

Supply current 0.5 1.9 Amps

Supply voltage: (1) 4.75 5.0 5.25 V

PerkinElmer power cable total resistance 0.2 �

Case operating temperature (1.3) 5 40 ˚C

Active area (diameter) at minimum Pd 170 175 �m

Photon detection efficiency (Pd) at 400 nm 2 5 %650 nm 50 65 %830 nm 35 45 %

1060 nm 1 2 %Pd variation at constant case temperature ± 1 ± 3 %(2hrs at 25˚C)

Pd variation 5˚C to 40˚C ± 4 ± 10 %

Dark count (4,5,6) = SPCM-AQR-12 — 500 Counts/Sec.SPCM-AQR-13 — 250 Counts/Sec.SPCM-AQR-14 — 100 Counts/Sec.SPCM-AQR-15 — 50 Counts/Sec.SPCM-AQR-16 — 25 Counts/Sec.

Average dark count variation at constant casetemperature (6hrs at 25˚C) for (4,5,6,);

SPCM-AQR-12 & 13 ± 10 %SPCM-AQR-14 & 15 & 16 ± 1 �

Average dark count variation at 5˚C to 40˚Ccase temperature for (4,5,6,);

SPCM-AQR-12 & 13 ± 20 %SPCM-AQR-14 & 15 & 16 ± 2 �

Single photon timing resolution Contact factoryfor availability

Dead time (Count rates below 5 Mc/s) 50 60 ns

Output count rate before saturation 10 15 Mc/s

Linearity correction factor: (7) at 200 kc/s 1.01at 1 Mc/s 1.08 1.15at 5 Mc/s 1.40 1.67

Afterpulsing probability 0.5 %

Settling time following power up (1% stability)at 1 meg counts/sec and 25˚C 15 30 s

Threshold setting required on counter fordigital output pulse (terminate in 50�) 0.75 1.0 2.0 V

Output pulse width 35 ns

Gating turn on/off: (50� output)Disable = TTL Low 2 4 nsEnable = TTL High 45 55 ns

Gating threshold voltage: (at V supply = 5 V)Low level (sink current >90mA) 0 0.4 VHigh level (sink current >30mA) 3.5 5.25 V

Table 2. Absolute maximum ratings

Supply voltage (1) 5.5 V

Mean count rate 5 Mc/s (Above this point, dead time will increase due to diode self-heating)

Peak light Intensity 104 photons per pulse and pulse width less than 1 ns.

Case temperature (3) 50˚C storage, 40˚C operating

Page 4: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

4

Operating instructions

1. Connection to incorrect voltageor reverse voltage may destroythe module. The warranty isinvalid where such damageoccurs.

2. These modules are not qualifiedfor shock or vibration otherthan normal instrumentationenvironments.

3. The module dissipates a meanpower of 2.5W and a maximumpower of 6.5W at high countrate and 40°C. Adequate heatsinking must be provided byclamping the module to asuitable heat sink via the holesin the module base. For thespecification performance, themodule case temperature mustnot exceed 40°C.

4. Bi-stability of the dark count:On a small percentage ofdelivered modules, bi-stabilityof the dark count has beenobserved. Research indicatesthis bi-stability is probably due to transitions at a singleimpurity site between a lowenergy and a high energy state.The phenomenon is seen as anabrupt change in the dark countrate, e.g., 350 to 390 c/s., andthe dark count switchesbetween the two states at a ratedependent upon the detectortemperature. Multilevelswitching has also beenobserved, where more than oneimpurity site is switching.

5. Long-term bi-stability is relatedto fundamental semiconductorphysics and is beyond

PerkinElmer’s control. Warrantyclaims will not be consideredagainst bi-stability alone.Warranty claims will only beconsidered if the high level ofthe dark count exceeds themaximum level in thespecification.

6. In the dark, the modulegenerates random counts thatfollow a Poisson distribution. Ina Poissonian process, thestandard deviation is equal tothe square root of the averagecount. In this specification the“dark count variation” refers tothe stability of the averagecount of the module.

7. The actual photon rate could becalculated using the followingequation, as indicated below:

ACTUALCOUNTRATE Photons = (OUTPUT ModuleCountRate x CORRECTIONFACTOR @ the Module CountRate) - DARK COUNT ModulePHOTON DETECTION EFFICIENCY Module

The theoretical value, at low count rate, of the Correction Factor follows this equation:

Correction Factor = 1 Where: td = Module Dead Time

1- (td X CR) CR= Output Count Rate

The deviation from an ideal linear system is another way of looking at the saturation effect. The following equations show how to calculate this departure from the linearity:

OUTPUT ModuleCountRateLINEARITY = -1

(PHOTONS Actual Count Rate x PHOTON DETECTION EFFICIENCY Module) + DARK COUNT Module

1

= -1Correction Factor

Page 5: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

5www.optoelectronics.perkinelmer.com

Figure 1. Block diagram of module Figure 2. Electrical connections

Figure 3. Detector scan Figure 4. Photon detection efficiency (pd) vs. wavelength

Figure 5. Typical afterpulse probability Figure 6. Typical correction factor

Page 6: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

6

Figure 7. Dimensional 0utline

Figure 8. Optical power vs. number of photons

Page 7: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

North AmericanCustomer Support CenterPerkinElmer Optoelectronics22001 Dumberry Road Asia HeadquartersVaudreuil, QC J7V 8P7 European Headquarters PerkinElmer OptoelectronicsCanada PerkinElmer Optoelectronics 47 Ayer Rajah Crescent #06-12Telephone: (+1) 450-424-3300 Wenzel-Jaksch-Str.31 Singapore 139947Toll Free Telephone: (+1) 866-574-6786 65199 Wiesbaden, Germany Telephone: (+65) 6775-2022Fax: (+1) 450-424-3345 Telephone (+49) 611-492-430 Telephone: (+65) 6770-4366Toll Free Fax: (+1) 877-574-1329 Fax: (+49) 611-492-165 Fax: (+65) [email protected] [email protected] [email protected]

For a complete listing of our global offices, visit www.optoelectronics.perkinelmer.com

©2005 PerkinElmer, Inc. All rights reserved. The PerkinElmer logo and design are registered trademarks of PerkinElmer, Inc. Slik™ is a registered trademark of PerkinElmer, Inc. or itssubsidiaries, in the United States and other countries. All other trademarks not owned by PerkinElmer, Inc. or its subsidiaries that are depicted herein are the property of their respectiveowners. PerkinElmer reserves the right to change this document at any time without notice and disclaims liability for editorial, pictorial or typographical errors.

600092_01 DTS1005

“Your Partner of Choice”

With a broad customer base in all majormarkets, built on ninety years of solidtrust and cooperation with our cus-tomers, PerkinElmer is recognized as areliable partner that delivers high quan-tity, customized, and superior "one-stop" solutions. Our products - fromsingle photocells to complex x-rayinspection systems - meet the highestquality and environmental standards.

Our worldwide Centers of Excellence,along with our Customer and TechnicalSupport teams, always work with youto find the best solutions for yourspecific needs.

PerkinElmer Optoelectronics

PerkinElmer Optoelectronics is a globaltechnology leader providing market-driven, integrated solutions for a widerange of applications, which leverageour lighting, sensors, and imaging

expertise. Our technologies, servicesand support are fueling the medical,genomic and digital revolutions byenhancing our customers' productivity,optimizing performance, and accelerat-ing time to market.

So contact us and put PerkinElmer'sexpertise to work in your demandingapplications. We will show how ourinnovations will help you deliver theperfect product.

Fiber type ordering guide

Part Number Fiber Type Connector Type Diameter Numerical

Core Cladding Outer Aperature

SPCM-QC4 Multimode FC / Bare 62.5 �m 125 �m 2.5 mm 0.27

SPCM-QC6 Multimode FC / Bare 100 �m 140 �m 2.5 mm 0.29

SPCM-QC8 As SPCM-QC6 but 905 SMA on free end, 100 microns core fiber

SPCM-QC9 As SPCM-QC6 but FC connector on free end, 100 microns core fiber

Please contact PerkinElmer Optoelectronics for other configurations.

SPCM ordering guide

Page 8: SPCM-AQR Single Photon Counting Module Datasheetphys191r/Bench... · North American Customer Support Center PerkinElmer Optoelectronics 22001 Dumberry Road Asia Headquarters Vaudreuil,

Appendix I

Declaration of Conformity

Declaration stating that PerkinElmer has conformed to CSA (Canadian Standards Association) which provides theright to bear the CSA mark on this product.

This product is eligible to bear the CSA mark with adjacent indicator ‘C’ and ‘US’.

Products:CLASS 8721 84 ELECTRICAL EQUIPMENT FOR LABORATORY USE – Certified to US standardsCLASS 8721 04 LABORATORY EQUIPMENT – Electrical.

Single photon counting module SPCM-AQR model SPCM-AQR-WX (where W = 1 and X = 2, 3, 4, 5, 6), rated 5 Vdc,1.9 A, Continuous operation, installation category l, pollution degree 2

APPLICABLE REQUIREMENTS:

CAN/CSA-C22.2 No. 1010.1-92 (R1999) – Safety requirements for electrical equipment for measurement, control, andlaboratory use, part 1: General requirementsCAN/CSA-C22.2 No. 1010.1B-92 – Amendment 2:1997 to CAN/CSA 22.2 No. 101.1-92, Safety requirements forelectrical equipment for measurement, control and laboratory use, Part 1: General requirementsUL std no. 61010A-1 – Electrical equipment for laboratory use ; part 1:General requirements

----------------------------------------------------------------------------------

This product is eligible to bear the CE mark in accordance with:

EN 61326:1997 Electrical equipment for measurement, control and laboratory useEN 61010-1:2001 Safety requirements for electrical equipment for measurement, control and laboratory useThis product has been tested as per the following standards:

• Emission CISPR 11• IEC 61000-4-2 ESD• IEC 61000-4-3 Radiated susceptibility• IEC 61000-4-4 Burst• IEC 61000-4-5 Surge• IEC 61000-4-6 Conducted susceptibility• IEC 61000-4-11 Voltage dips and interruptions

www.optoelectronics.perkinelmer.com


Recommended