SERVICE MANUAL
ECHO CAMERA
SSD-1000
Document NumberDocument Revision
: MN2-0218: 0
Copyright
English Edition
1 / 1
MN2-0218SSD-1000 SERVICE MANUAL
1/4
Contents of SSD-1000 SERVICE MANUALPAGE
Section1 How to use this service manual page 1-11-4 (4 pages)
1-1 Service Manual 1- 11-2 Contents of this Service Manual 1- 11-3 Construction of This Service Manual 1- 11-4 Contents of Each Section 1- 2
Section2 PRECAUTIONS (read without fail) page 2-12-4 (4 pages)
2-1 Precautions Against Electrical Hazards to Serviceman 2- 12-2 Precautions Against Mechanical Hazards to Serviceman 2- 12-3 Precautions Against Germ Hazards to Serviceman 2- 12-4 Precautions for Keeping Electrical Safety 2- 22-5 Precautions for Keeping Mechanical Safety 2- 22-6 Precautions for Keeping Chemicals Safety 2- 22-7 Preparation to be Made at Service Center 2- 22-8 Care to be Taken in the Field 2- 32-9 Precaution for Monitor repairing 2- 32-10 Handling of S. M. D. PCBs 2- 4
Section3 BEFORE REPAIRING page 3-13-14 (14 pages)
3-1 Repair work on the description of Service Manual 3- 13-2 Upgrade (Improvement) work on the description of Service Manual 3- 63-3 Massages 3- 12
Section4 PRINCIPLE OF SYTSTEM OPERATION page 4-14-86 (86 pages)
4-1 System Specification 4- 14-2 System Configuration 4- 54-3 System Block Diagram 4- 54-4 Principle of System Operation 4- 11
MN2-0218SSD-1000 SERVICE MANUAL
2/4
4-5 Principle of each PCBs Operation 4- 204-5-1 PROBE SELECTOR 4- 204-5-2 TX / PRE AMP 4- 244-5-3 RX FOCUS 4- 284-5-4 MAIN AMP 4- 324-5-5 GEU I.T.F. 4- 364-5-6 DIU 4- 404-5-7 CPU 4- 504-5-8 CINE MANAGER & CINE MEMORY 4- 544-5-9 MECHA. SERVO 4- 604-5-10MECHA. PROBE SELECTOR 4- 664-5-11PHYSIO. AMP. 4- 704-5-12PHI (PHYSIO. MEMORY) 4- 764-5-13DMS ITF / VOL / SERVO 4- 82
Section5 SCHEMATICS page 5-15-60 (60 pages)
5-1 PSC-129 CABLE CONNECTION 5- 15-2 CBL15 CBL16 5- 35-3 CBL314 CBL317 CBL318 5- 55-4 CBL600 CBL601 5- 85-5 CBL800 CBL801 CBL803 5- 105-6 CBL804 CBL806 CBL807 5- 135-7 CBL301 CBL550 CBL754 5- 165-8 CBL802 5- 195-9 FOOT SWITCH 5- 205-10 L-KEY-64* 5- 215-11 GEU MOTHER BOARD 5- 335-12 DIU MOTHER BOARD 5- 375-13 AC OUTLET JB-212 5- 415-14 AC OUTLET JB-215 5- 425-15 CABLE CONNECTION FOR POWER SUPPLY 5- 435-16 TRANSFORMER UNIT (EU-6024) 5- 445-17 POWER STABILIZING UNIT (EU-6025) 5- 455-18 8 VOLTAGE REGULATOR 5- 465-19 MONITOR IP-1233B* 5- 495-20 PHYSIO UNIT EU-5042 5- 51
MN2-0218SSD-1000 SERVICE MANUAL
3/4
5-21 CABLE752 CABLE750 5- 525-22 CBL756, CBL758 & PHYSIO. PANEL 1 5- 545-23 CBL753 & PHYSIO. PANEL 2 5- 555-24 CBL302 CBL312 CBL320 5- 565-25 CBL805 5- 595-26 POWER CABLE FOR MO DRIVE UNIT 5- 60
Section6 TROUBLE SHOOTING page 6-16-82 (82 pages)
6-1 Introduction 6- 16-2 Precautions 6- 16-3 Tools and Measuring Instruments Required 6- 46-4 Information 6- 5
6-4-1Automatically Setting after Powering On 6- 56-4-2Setting Switches on the PCB 6- 56-4-3Usage of the Functions to be Set by the User 6- 76-4-4Attention of connecting optional units 6- 86-4-5Resetting the Backup Memory (RAM) 6- 96-4-6How to reset Backup RAM 6- 106-4-7Location of PCBs 6- 116-4-8Location of ROMs 6- 126-4-9Correction for fail to start or locking up 6- 156-4-10Minimum Start-up 6- 176-4-11 Preset Data Sheet 6- 20
6-5 Check List Map 6- 356-5-1Flow chart for the usage of Check List Map 6- 356-5-2Phenomenon Code Table 6- 376-5-3MAP 6- 446-5-4PCB Check Procedure 6- 50
6-6 Waveform for Troubleshooting 6- 71
Section7 ADJUSTMENT PROCEDURE page 7-17-4 (4 pages)
7-1 Introduction 7- 17-2 Attention 7- 17-3 Tools and Measuring Instruments 7- 17-4 Adjustment Procedure 7- 2
MN2-0218SSD-1000 SERVICE MANUAL
4/4
Section8 PERFORMANCE CHECK page 8-18-12 (12 pages)
8-1 Introduction 8- 18-2 Precautions 8- 18-3 Making Entries in Repair Report 8- 18-4 Performance check 8- 2
8-4-1Function Check 8- 38-4-2Image Quality Check 8- 88-4-3Safety Check 8- 10
Section9 DISASSEMBLING PROCEDURE page 9-19-42 (42 pages)
9-1 SSD-1000 Disassembling Procedures 9- 19-2 EU-9075 Installation Procedures 9- 299-3 DMS-1000 Installation Procedures 9- 35
Section10PARTS LIST page 10-110-46 (46 pages)
10-1Contents of Parts List 10- 110-2Appliance of Parts List 10- 210-3Outline of Parts List 10- 210-4Explanation of Parts List 10- 310-5Attention 10- 410-6Parts List 10- 5
Section11SERVICE INFORMATION page 11-111-4 (4 pages)
11-1Introduction 11- 111-2Assignment of CTRL key and those usage 11- 111-3A pop-up menu display for probe registration 11- 3
Appendix SSD-1000 SERVICE MANUAL Appendix pageApndx-1Apndx-22 pages
Appendix-1Introduction....................................................................... Apndx- 1
Appendix-2Manual Change Information.............................................. Apndx- 1
MN2-0218SECTION 1 How to use this service manual
1 - 1
1-1 Service Manual
1) This service manual has been prepared for persons in charge of repair at the field.
2) This service manual is compiled according to the following basic principle.For service, pickout a faulty PCB and replace it with a new PCB .
3) Make the best use of this service manual, making also reference to available technical support
information such as Technical Bulletin.
1-2 Contents of this Service Manual
1) The equipment is repaired by PCB replacement. Therefore this service manual does not
include the circuit diagrams of the PCB unit. For the function of each PCBs whose circuit
diagram is not included, refer toSECTION 4 PRINCIPLE OF SYSTEM OPERATION.In SECTION 4, Specification of System, Principle of System, System Block Diagram, PCB
Block Diagram, the explanation of each PCB Block Diagram, and signal list are described.
However, Cable Connection Diagram, Circuit Diagram of PCB equipped with the panelswitches which are easily exchangeable at the field andCircuit Diagram composed ofgeneral circuit such as TV monitor and Power Supply unit are described inSECTION 5SCHEMATICS.
2) For changes and modifications of as well as additions to specifications, if any, prompt
information will be given to you by means ofAPPENDIX Manual Change Information.
IMPORTANT Always observe the manner specified for replacement, addition, ordeletion of Manual Change to prevent missing of necessaryinformation and keeping of erroneous information.
1-3 Construction of This Service Manual
The structure of Service Manual is as follows:
1) Service Instructions................................................................SECTION 13, 59, 112) Principle of Operation............................................................SECTION 4
3) Parts List................................................................................SECTION 10
4) Manual Change Information...................................................APPENDIX
MN2-0218SECTION 1 How to use this service manual
1 - 2
1-4 Contents of Each Section
SECTION 1 How to use this service manual
Describes the purpose of the Service Manual.
SECTION 2 PRECAUTIONS
Describes general precautions and preparations for maintenance service. Be sure to follow working
procedures if mentioned.
SECTION 3 BEFORE REPAIRING
Gives information peculiar to the equipment and care to be taken before starting repair work.
SECTION 4 PRINCIPLE OF SYSTEM OPERATION
Describes Specification of System, Principle of System, System Block Diagram, PCB Block
diagram, the explanation of each PCB Block Diagram, and Signal List. Gives the convenience of
grasping flow of major signals and mutual communication between units in the whole system.
SECTION 5 SCHEMATICS
Gives the cable connection diagram including all cables used, the circuit diagram of PCB equipped
with switches, and the circuit diagram of TV monitor and Power Supply unit.
SECTION 6 TROUBLESHOOTING
Describes precautions on actual repair work and shows the necessary tools and measuring
instruments. Also, includes many hints on primary diagnosis and measures to be taken in the field.
MN2-0218SECTION 1 How to use this service manual
1 - 3
SECTION 7 ADJUSTMENT PROCEDURE
Gives guides of adjustments of PCBs and units which some PCBs need when they are replaced .
SECTION 8 PERFORMANCE CHECK
Describes the procedure of checking for proper operation after repair and provides the forms of
check sheet.
SECTION 9 DISASSEMBLING PROCEDURE
Disassembling Procedure Illustrates the disassembly and assembly of main components. Be sure to
follow working procedures if specified.
SECTION 10 PARTS LIST
Lists the mechanical parts and electrical part which replacement possibility are considered .
SECTION 11 SERVICE INFORMATION
Provides available information about maintenance service.
APPENDIX SERVICE MANUAL Appendix
Manual change information, the revision list of this manual, is filed in this section.
MN2-0218SECTION 1 How to use this service manual
1 - 4
( Blank page )
MN2-0218SECTION 2 PRECAUTIONS
2 - 1
2-1 Precautions Against Electrical Hazards to Serviceman
When disassembling the equipment after checking it for a trouble symptom, give care to the
following:
1) Be sure to unplug the equipment before disassembly.
2) Be sure to turn off the main switch on the equipment when removing electrical parts such as
PCBs, probe, and cable.
3) Safety alert symbols
The indication used on this equipment and in this service manual have the following
meaning.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury.
A caution message is inserted here.
2-2 Precautions Against Mechanical Hazards to Serviceman
When disassembling the equipment, give care to the following to protect serviceman from
hazards :
1) Keep the working environment neat.
Wear working gloves to protect your hands from getting injured by burrs on the unit and
casing.
2) Use only proper tools suited to work being made.
3) Be sure to observe the specified disassembly procedure shown in SECTION 9.
4) Take sufficient care not to damage component with undue load.
2-3 Precautions Against Germ Hazards to Serviceman
1) When it is necessary to touch the equipment, options and/or other peripheral devices at a
customer who uses intracorporeal (transesophageal, transurethral, transvaginal, transrectal)
probes that need sterilization, take special care to protect your hands against germs,
irrespective of the usage of the equipment: whether it is used in the operation room or not.
2) Service tools are subject to germ pollution in hospitals and, therefore, need periodical
sterilization.
3) Be careful not to directly touch anything assumable to have germ pollution. If necessary, ask
the customer for effective protection against germs.
MN2-0218SECTION 2 PRECAUTIONS
2 - 2
2-4 Precautions for Keeping Electrical Safety
1) Be sure to ground the equipment securely.
2) Perfectness in grounding, screw tightening, and cover installation is essential. Negligence of it
could cause a possibility of leakage current from outer fitting which may lead to serious
damage to a patient being diagnosed.
2-5 Precautions for Keeping Mechanical Safety
Take care to the following to prevent the equipment from being damaged or broken during
disassembly and reassembly work.
1) Be sure to observe the specified disassembly procedure.
2) Take care not to damage component parts by undue load.
3) When reassembling the equipment, carefully check every part for loosening, distortion and
creak.
4) Use only the specified screws and nuts. Using any other screws and/or nuts would affect not
only mechanical performance, but also electrical performance of the equipment.
2-6 Precautions for Keeping Chemicals Safety
Whenever grease, oil or other chemicals is used for maintenance service, options and/or
peripheral devices, be sure to clean the equipment and/or devices after service work.
2-7 Preparation to be Made at Service Center
1) When called by a customer on the telephone, note the followings:
l Name of equipment
l Serial number of equipment
l Name of hospital
l Telephone number
l Name of person in charge
l Detail of trouble symptom as far as possible
l State of connection to optional devices
2) Go over the Technical Bulletin to see whether the complained trouble can be mended bymeans of regular repairing method.
MN2-0218SECTION 2 PRECAUTIONS
2 - 3
2-8 Care to be Taken in the Field
1) Check for trouble symptoms.
2) Check for connection to optional devices and other peripheral devices.
3) Record the contents of the battery backup memory.
4) After working, restore the equipment according to the above mentioned contents of memory if
necessary.
5) After completion of work, put back the peripheral devices to the original condition.
2-9 Precaution for Monitor repairing
1) Subjecting the unit to strong shocks may result in damage to the CRT or malfunction, therefore
care must be taken when transporting or installing the unit.
2) CRT with the deflecting yoke is already adjusted to the best condition. Do not touch the
deflecting yoke and the magnet of the neck part.
3) Be sure to detach the metallic goods such as a wrist watch from your body before doing the
repair work.
To prevent the secondary damage and the electrical shock, the matters above should be taken
into careful consideration.
DANGER High voltages are present inside the display chassis. Only experiencedtechnicians should touch internal parts.
DANGER The electric charge has remained in CRT after the power switch is turnedoff. Because the high voltage is usually used for CRT. So make the
electric charge escape with a grounding stick which is connected to the
ground of the chassis and through the resistance for high voltage (Approx.
1M) before removing the anode cap.Some electric charge remain in CRT after escaping with a grounding
stick. Do not touch the metallic part of the anode cap with bare hands,
when detaching the anode cap directly.
MN2-0218SECTION 2 PRECAUTIONS
2 - 4
2-10 Handling of S.M.D. PCBs
It is an Aloka's policy that neither repair nor upgrade (improvement) of PCBs used for S.M.D. is
made in the field as a rule because of the following reasons:
[REMARKS] PCB does not need repairing or modifying in the field as a rule.
When handling a PCB, do not touch the IC unless it is necessary.
IC soiled with workers hands may cause corrosion. Additionally, foreign particles such as fine
solder dust could be the cause of short-circuited IC lead wires whose pitch is smaller than that of
the traditional ones.
Do not give excessively large shocks to the PCB.
When replacing the ROM (Read Only Memory) on the PCB, attempting to force the ROM into
its socket would cause the PCB to be subjected to an undue force, and the following faults may :
1) Damage to PCB intermediate-layer patterns,
2) Peeling of chip devices (resistor, capacitor, diode, etc.)
3) Damage to a junction between electrode and internal element of chip devices,
4) Peeling of patterns (especially those for mounting the parts) together with chip devices since
those patterns are rather fragile compared with PCBs used before now, and
5) Damage to parts on the reverse side in the case of PCBs of both-side mounting type.
Also, a PCB mounted improperly or a warped PCB mounted as it is may cause the chip devices to
come off and the fine patterns to be cut.
Additionally, reuse of chip devices (including resistors, capacitors, diodes, etc.) is strictly
inhibited because of the following reason: Since the chip devices are lacking in lead wires, such
as those found in the traditional component parts, heat given to the PCB will be directly conducted
to the inside of chip devices. As a result, a thermal stress will occur due to a difference in thermal
expansion coefficient between each chip device and PCB, giving rise of the possibility of cracks
inside of or on the surface of chip devices or the possibility of thermal breaking (internal burning).
Very thin wiring patterns require extreme care in handling of the PCB
Be sure to observe the precautions mentioned above also to prevent the secondary accidents.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 1
3-1 Repair work on the description of Service Manual
The typical processes for the repair work are shown as the Flow Chart on the next page. Do the
repair work according to this procedure. In the case of upgrade (improvement) of the Technical
Bulletin or Upgrade Kit, see the next item 3-2.
Each procedures of flow chart are numbered to refer its detail shown from page 3-3. Furthermore,
the Flow Chart and its explanation show the time when each section of service manual are required
on repair work. This is a guide for the usage of service manual.
The service manual is very important for the repair work, especially readjustment and
performance check after completion of repair work. This is to keep the safety and quality of
equipment. If you make them, you have to describe that the treatment has been done according to
the applied section of service manual, on the repair report or the like.
The circled numbers shown in the Flow Chart on next page, are corresponded to the procedure
number shown from page 3-3.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 2
at customer side
at your site
Yes
No
4
3
2
1
No
Yes
Reception andInvestigation
Selection and order ofrequired part(s)
Preparation
Confirmation
Repair andadjustment
Operation check
Work as normal?
Check by customerFill repair report
Approve by customerRepair again?
Present repair reportDemand to repair the
defective part(s)
Show comments of theprohibition to use on the
equipment
Report to customer
Section 46, 11
Section 6, 10
Technical Bulletin
History
Section 3
Section 2, 6, 9
Section 8
5
6
7
8
9 11
10
12
Demand of repair(START)
Completion(END)
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 3
Procedure 1Reception of repair and investigation
Accept the repair request from the customer or distributor. At this time, the following points have
to be confirmed and checked,
l Model name/number, and serial number
l Name of customer, address, phone number, and name of person in charge
l Configuration of the connection of peripheral devices
l Software version or the like shown on the start up display(if possible)
l Detail of phenomenon appeared on the function of equipment
Make an examination what circuit may be defective as the function of equipment based on the
above information. If you need to know about the basic operation and special information for the
maintenance, refer to the following sections, or ask to the Technical Support,
u Section 4 PRINCIPLE OF SYSTEM OPERATION
u Section 5 SCHEMATICS
u Section 6 TROUBLESHOOTING
u Section 11 SERVICE INFORMATION
The reported phenomenon may be the original problem on the equipment. Because, refer to the
Technical Bulletin separately issued to check it whether defectiveness or not. If it has been reported
as the original problem, make a work according to the Technical Bulletin.
Procedure 2Selection of required parts and order
If you find the doubtful circuit, order the necessary parts. Then check the delivery date and decide
the date to visit on the consultation with the customer.
For the selection and order of parts, refer to the following sections,
u Section 6 TROUBLESHOOTING
u Section 10 PARTS LIST
For the electrical parts such as UNIT, check the history information on the HISTORY of this
equipment separately issued.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 4
Procedure 3Preparation of visiting the customer
Check the required tools, measuring devices and parts to be replaced before the visiting the
customer. Then check the special information for the equipment reference with the following
section,
u Section 3 BEFORE REPAIRING
Procedure 4Confirmation of phenomenon
Confirm the appeared phenomenon and condition to happen it with the customer. If you dont
know about the operation of equipment, refer to the Operation Manual attached to the equipment.
Procedure 5Repair and readjustment
Repair the defective circuit with the brought parts. For the repair work, read the following section
carefully,
u Section 2 PRECAUTIONS
And, examine the trouble reason depending on the situation with following section,
u Section 6 TROUBLESHOOTING
The electrical or mechanical readjustment may be requested depending on the replaced parts.
Because, refer to the following section after completion of repair,
u Section 9 ADJUSTMENT
Procedure 6Operation check
Check the system behavior to keep its condition as same as before in trouble, reference with the
following section. Be sure to do according to the description because check items are depending on
the portion to be treated.
u Section 8 PERFORMANCE CHECK
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 5
Procedure 7Judgment of the operation quality
If the result of Procedure 6 is passed to the all standards, do the next Procedure 8. On the
other side, if not, make a judgment of Procedure 10.
Procedure 8Confirm by customer, make repair report and approve
Reconfirm the solution of trouble phenomenon with the customer. Then make a repair report and
obtain approval of customer.
The repair report shows not only the treatment but also the method of readjustment and operation
check. If they have been done according to the service manual, the followings have to be shown,
Readjusted according to the Section 9 of service manual.
Checked according to the Section 8 of service manual, and passed.
Procedure 9Presentation of report and order to repair parts
Fill the repair report with necessary item, and present it according to the certain procedure.
If the defective parts that trouble cause included is available to use again by repair, make an order
to do. If you cannot judge whether the part can be used again or not, ask to the Technical Support.
Procedure 10Judgment of possibility to repair again
As the result of judgment on Procedure 7, if the trouble is not solved, judge the possibility to
make the repair work again.
If available, return to Procedure 5 and continue to work.
If unavailable, go to Procedure 11.
Procedure 11Indication of the prohibition to use
As the result of judgment on Procedure 10, if you judge that it is impossible to continue the
repair work at this time, indicate that the equipment is still out of order, and also show the
prohibition to use, on the equipment.
Procedure 12Report to the customer
Report the reason why the trouble cannot be solved to the customer. Then consult about the plan
of next repair work.
And do the same way from Procedure 2.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 6
3-2 Upgrade (Improvement) work on the description of Service Manual
The typical processes for the upgrade (improvement) work are shown as the Flow Chart on the
next page. Do the upgrade (improvement) work according to this procedure. In the case of repair
work, see the previous item 3-1.
Each procedures of flow chart are numbered to refer its detail shown from page 3-8. Furthermore,
the Flow Chart and its explanation show the time when each section of service manual are required
on upgrade (improvement) work. This is a guide for the usage of service manual.
The service manual is very important for the upgrade (improvement) work, especially
readjustment and performance check after completion of upgrade (improvement) work. This is to
keep the safety and quality of equipment.
The circled numbers shown in the Flow Chart on next page, are corresponded to the procedure
number shown from page 3-8.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 7
at customer side
at your site
Yes
No
Yes
4
3
2
1
No
No
Yes
Selection and order ofrequires parts/kits
Preparation
Upgrade (improvement)
Can recover?
5
6
7
8
9
11
12
Operation check
Work as normal?
Do the repair work,according to item 3-1
Operation check
Work as normal?
Check by customer
Consultation withTechnical Support
10
13
14
Demand to upgrade (improvement)(START)
Section 3
Section 8
Upgrade
Upgrade(improvement)
ProcedureSection 8
Tech.BulletinInstllationProcedure
Show comments of theprohibition to use on the
equipment
Report to customer
Return unnecessaryparts, and report of
upgrade (improvement)
Completion(END)
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 8
Procedure 1Selection of required parts / kits and order
Accept the upgrade ( improvement) request from the customer, distributor or person in charge of
sales. At this time, the following points have to be confirmed and checked to decide the parts and
kits,
l Document name that announced the upgrade ( improvement) or kit requested
l Model name/number, and serial number
l Name of customer, address, phone number, and name of person in charge
l Configuration of the connection of peripheral devices
l Software version or the like shown on the start up display
Make an examination what parts or kits are required based on the above information. For the
selection, refer to the following document separately issued, or ask to the Technical Support,
u Technical Bulletin
To confirm the detail of upgrade (improvement), see the Installation Procedure attached with
applied Technical Bulletin.
Depending on the upgrade (improvement), hardware, or software, the other upgrade
(improvement) may be required. Check it with the Technical Bulletin.
Then, confirm the delivery date of required parts or kits, and decide the date to visit on the
consultation with the customer.
Procedure 2Preparation of visiting the customer
Check the required tools, measuring devices and parts or kits to be used before the visiting the
customer. Then check the special information for the equipment reference with the following
section and document,
u Section 3 BEFORE REPAIRING
u Technical Bulletin and/or Installation Procedure
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 9
Procedure 3Operation check before upgrade (improvement)
On the basis of work, the upgrade (improvement) to the defective equipment is prohibited.
Because, before upgrade (improvement) work, check the behavior of equipment whether normal or
not according to following section and document,
u Section 8 PERFORMANCE CHECK
u Operation Manual
Procedure 4Judgment of the operation quality
If the result of Procedure 3 is passed to the all standards, do the next Procedure 5. On the
other side, if not, go to Procedure 10.
Procedure 5Upgrade (Improvement) work
Do the upgrade (improvement) work according to the following document,
u Installation Procedure attached with kit or Technical Bulletin
Procedure 6Operation check after upgrade (improvement)
Check the system behavior to keep its condition as same as before the upgrade (improvement),
reference with the following section. Be sure to do according to the description because check
items are depending on the portion to be treated.
u Section 8 PERFORMANCE CHECK
u Installation Procedure
Procedure 7Judgment of the operation quality
If the result of Procedure 6 is passed to the all standards, do the next Procedure 8. On the
other side, if not, make a judgment of Procedure 11.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 10
Procedure 8Confirmation by customer
Reconfirm any functions of equipment with the customer. Then, if need, introduce and explain
about the new functions and specification added by this upgrade (improvement).
Furthermore, if need, make a report to be approved by the customer. The report shows not only
the treatment but also the method of operation check. If it has been done according to the service
manual, the following has to be shown,
Checked according to the Section 8 of service manual, and passed.
Procedure 9Return of unnecessary parts and report of completion
According to the Technical Bulletin, return the unnecessary replaced or unused parts as soon as
possible if suggested.
And, if the report of upgrade (improvement) is suggested on the same document, report it with the
information required.
Procedure 10Work for the abnormal behavior of equipment
On the result of judgment in Procedure 4, if the equipment does not work normal, solve the
problem according to item 3-1 Repair work on the description of service manual shown in this
section.
When the problem is solved, return to Procedure 5 of this item and continue to do the upgrade
(improvement) work.
Procedure 11Judgment of possibility to recover
As the result of judgment on Procedure 7, if the problem has been made by this upgrade
(improvement), judge the possibility to recover it.
If available, return to Procedure 5 and continue to work.
If unavailable, go to Procedure 12.
Procedure 12Indication of the prohibition to use
As the result of judgment on Procedure 11, if you judge that it is impossible to recover at this
time, indicate that the equipment is the out of order, and also show the prohibition to use, on the
equipment.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 11
Procedure 13Report to the customer
Report to the customer that the upgrade (improvement) has not been completed because of the
problem on the upgrade (improvement) work. Then make a schedule to fix and complete it.
Procedure 14Asking to the Technical Support
Report to the Technical Support about the happening of problem on the upgrade (improvement)
work, make an examination to solve and order the additional parts. Before the asking, check the
following points,
l Name of kit, or the issue number of Technical Bulletin showing the upgrade (improvement)
l Model name/number, and serial number
l Configuration of the connection of peripheral devices
l Software version or the like shown on the start up display
l Indication of equipment such as Upgrade (improvement) or History Label
l Detail of phenomenon appeared on the function of equipment
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 12
3-3 Messages
In this equipment, messages are displayed warning that the equipment is malfunctioning or
advising the correct operation method.
There are two types of message, differing according to their content.
1) WARNING
This appears at the center of the screen and an alarm tone is also emitted.
(You can switch over the warning display position at the bottom of screen by selecting
WARNING DISPLAY in Common Preset of PRESET, then selecting BOTTOM.)
2) Assistance Messages
When you use the keyboard to operate a function in a menu, assistance messages appear in
the message area at the bottom of the screen .
Warning Area
Message Area
Fig. Message Areas
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 13
1) WARNING
No. Message Cause Treatment
1 Invalid probe connected.A probe was connected not
intended for use with the
equipment.
Connect a suitable probe.
2 Store memory clear
Data in the memory will be deleted.
Press to proceed.
:Quit
The NEW PATIENT switch
was pressed or MEMORY
CLEAR was selected from the
menu.
If it is permissible to erase image
data, press the SET switch. Ifnot, press the M AR K REFswitch.
3 MEMORY FULL
Press key again to clear
old images.
:Quit
You attempted to store a larger
amount of image data than the
remaining capacity of the store
memory.
Delete unwanted images from
the data already stored in the
memory.
2) Assistance Messages
No. Message Cause Treatment
1 Enter data from keyboard.
Then press key.
When the name of the
hospital and the data was
input.
Enter characters from the
keyboard, then press the SETkey.
2 INPUT ERROR
Invalid entry. Enter correct
value.
When incorrect values were
input during input of the data
and time.
Input the appropriate values.
3 the image.
Then try again.
When the PRINT switch onthe operation panel, STOREand REVIEW switches in theMENU were pressed during
freeze or when HISTmeasurement was carried out.
Freeze the image then press the
switch.
4 the image
Then continue measurement.
You attempted to perform an
auto trace without freezing the
image.
Press FREEZE . Then try again.
5 To change frequency or angle ,
press key.
After having changed to a
probe including two vibrator
elements.
Use the ELEMENT SELECTswitch to change the frequency.
6 System AUTO-FREEZE. Press
key to resume.
When panel operation was
not carried out for 20
minutes.
Press the FREEZE switch.
7 USE to
activate zoom.
You pressed the ZOOMswitch.
Turn rotary encoder to perform
a zoom operation.
MN2-0218SECTION 3 BEFORE REPAIRING
3 - 14
No. Message Cause Treatment
8 RANGE LIMIT
Selection is not available.
When a setting was made
exceeding the change limit in
case of changing a setting
value with the rotary encoder.
9 USE to move the
mark. Then press key.
You attempt to shift a body
mark.
Shift the body mark with the
trackball.
10 Press key to rotate
Fetus mark.
To display the fetus body
mark.
The fetal rotation function is
activated by the M AR K REFswitch.
11 DETECTION ERROR
R-wave of ECG is not detected
When it is not possible to
detect the R wave of the ECGfor 5 seconds or more.
12 IMAGE stored in memory
frame ##.
When operations were finished
after carrying out store.
13 DATA/image deleted. When there was erasing of 1 or
all images in the store memory.
14 No images have been stored in
memory.
If you attempt to erase an
image when there are no
images in the store memory.
15 Press in the MENU to
display the image to be cleared
from the store memory. Then press
key.
If you attempt to erase an
image in the store memory
when a review image is not
displayed.
Press the REVIEW in theMENU, select the image to be
erased, then erase it.
16 PROCESSING When the histogram is being
calculate.
17 Measurement memory area is full.
To proceed clear measurement,
press key to clear.
You attempt to perform
measurement when the
measurement display area was
empty.
MN2-0218SECTION 4 Principle of System Operation
4 - 1
4-1. System Specifications
Scanning method Electronic linear scanningElectronic convex scanning
Mechanical sector scanning(Requires SCU-1000)
Mechanical radial scanning(Requires SCU-1000)
Simultaneously connectable probes: Electronic scanning probes: 2
Mechanical scanner probes: 2(Requires SCU-1000)
Display modes B, 2B, B/M, M
Line mode (M mode)
Display method ScrollSweep speed 1, 1.5, 2, 3, 4, 6, 8 sec/screenEcho erase Available
Request function Available
Image display magnification (dependent on connected probe)....
Electronic linear probe 4 levels from 3, 4, 5, 6, 8, 10, 12, 15, 17, 19, 21, 24 cmElectronic convex probe 8 levels from 3, 4, 5, 6, 8, 10, 12, 15, 17, 19, 21, 24 cmMechanical scanner 5 through 8 levels from 3, 4, 5, 6, 8, 10, 12, 15, 17, 19,
21, 24cm
Screen display direction Reversible vertically and horizontally (in B modeonly)
Zoom function B scrolling method with write zoomMove position (B mode) AvailableScan area Variable (movable)Positive/negative reverse Available in M mode only
MN2-0218SECTION 4 Principle of System Operation
4 - 2
Ultrasound focusing method
Electronic scanning probeTransmit focusingB mode..........4 steps max.
B/M mode......1 of 4 steps
M mode.........1 of 4 steps
Acoustic lens
Receive focusingMax. 16 levels, continuously variable
Acoustic lens
Mechanical scanning probe Concave focusing
Acoustic output control 64-level selective
Image adjustment and processing functions
Gain B mode: Continuously variableM mode: Operates as B-mode compensation value
STC 8 sliding volumesContrast B mode8-level switching
M mode8-level switching
AGC B mode8-level switchingM mode8-level switching
FTC ON/OFF(M mode only)Relief B mode4-level switching
M mode4-level switching
ABC
Automatic B-Gain Control 64-level switchingIPSImage Process Select 8-level switchingVIEW gamma compensation 3 typesPHOTO gamma compensation 3 typesFrame correlation 4-level switching(also, Auto-function for circulatory
system)
Post-processing 5 typesLine density Electronic probe..............2-level switching
Mechanical sector............2-level switching
Mechanical radial............2-level switching
Image select 3 typesPresets 15 types
MN2-0218SECTION 4 Principle of System Operation
4 - 3
Measurement functions Basic measurementObstetric measurement (with OB report function)Cardiac function measurement
Annotation function Available
Characters and automatic display ID ..........................14 characters x 1 lineAge........................3 characters x1 line
Heart rate...............5 characters x1 line
Name.....................23 characters x 1 line
Sex.........................1 character x1 line
Hospital name. .......21 characters x 2 lines
Comments.............57 characters x31 lines (NTSC)
...............................57 characters x36 lines (PAL)
Date and time
Probe number
Probe frequency
Image select
Line density
DVA (Drive Voltage Amplitude (%) )
BSA, GA
Plane number for B mode
Frame rate
Depth range
Gain
Contrast
AGC
Graphic display Gray-scale bar...............(64 gradations)Scale mark
Index mark
Focus mark
Scan area mark
Cine/Search mark
M cursor
M time marker
Puncture guide line
Caliper mark
Body mark.............ABDOM
...............................OB/GYN
MN2-0218SECTION 4 Principle of System Operation
4 - 4
...............................CARDIO
...............................PV
...............................SMALL PART
...............................Four limbs
Cine memory (optional)
Search function Yes (Max. 32 frames in B mode)Scroll function YesStore function YesReview function YesLoop playback function 4-level switching of playback speed
Physiological signals (optional)
ECG display 1 channel Sensitivity Continuously variable Position B mode.....................Fixed
M mode....................Continuously variable
ECG synchronization Yes (1 channel) R-DELAY 0 to 2.55 sec in 10 msec stepsPCG display 1 channel (only with M mode display) Sensitivity Continuously variable Position Continuously variable
Image display system Black and white composite video signal (NTSC, PAL)
External I/O signal connectors Black and white composite video signal input: 1 lineBlack and white composite video signal output 2 lines
Power supply input voltage 100 to 120 VAC 10%, 200 to 240 VAC10%,50/60 Hz
Power consumption 600VA or less
External dimensions Overall width.......420mmOverall height......1200 - 1335mm
Depth...................630mm
Unit weight Approx. 75 kg
MN2-0218SECTION 4 Principle of System Operation
4 - 5
Operating ambient conditions Temperature......10 to 40 Humidity...........30 to 85 %
Barometric pressure..700 to 1060 hPa
MN2-0218SECTION 4 Principle of System Operation
4 - 6
4-2. System Configuration
The SSD-1000 ultrasound diagnostic device comprises the following units:
Main unit USI-145
TX/RX partDIU part
Operation panel unit L-KEY-64*
Trackball L-TB-8
Observation TV monitor IP-1233*(NTSC) / IP-1233*V(PAL)
Power supply unit PSU-S1000
Options
Mechanical scanner unit SCU-1000
Physiological signal display unit PEU-1000
Volume mode unit EU-9075
Data management Subsystem DMS-1000
Foot-switch MP-2345
3-way foot-switch MP-2614
Recording device (Black and white printer)
VCR
4-3. System Block Diagrams
Lists of all PCBs mounted in the device (except power supply unit and external options) and
system block diagrams are shown on the following pages.
See Section 4-5 "Principle of PCB Operation" for the PCB block diagrams.
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
4 - 7
SSD-1000 PCB LIST(1/2)SSD-1000 Unit PCB model
nameName Description
EP420500 CONNECTOR PANEL Connectors for B/W printer, TV signal for VCR, printer control signalsMai unit PSC-129
EP-3349-4 Connector for foot switch
Operation panel unitL-KEY-64 Operation Panel
Image storage unit USM-23 EP401600 GEU MOTHERBOARD
EP401500 PROBE SELECTOR Two 260-pin connectors, PROBE 1/2 selection, HVS, HVS control, probe codeoutput
EP404400 TX/PRE AMP Transmission trigger signal generation, transmission pulse generation, PRE AMPcircuit, PRE STC signal generation, transmission power supply voltage monitoringcircuit
EP400600 RX FOCUS Delay & addition, electronic/mechanical scanner probe received signal switching,V.B.P.F. circuit, LOG AMP circuit, received signal output
EP401700/EP430900
DIU MOTHER BOARD
EP400700 MAIN AMP Receive dynamic focus selection, gain, AGC, contrast, relief circuit and control signalgeneration
EP395601 GEU I.T.F A/D conversion, frame rate and heart rate calculation, mode control, ultrasoundaddress generation, transmission voltage control, timing signal generation
EP396604/05 DIU D/A conversion, pre-processing, main memory, post-processing, GEU interface,parameter memory, display control, TV timing signal generation
EP399000/EP403502
CPU Overall system control, character and graphic generation
Cine memory unit EU-9056 EP399600 CINE MANAGER Cine memory read/write control
EP399700 CINE MEMORY B/W cine memory
4 - 8
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
SSD-1000 PCB LIST(2/2)SSD-1000 Unit PCB model
nameName Description
EP400800 MECH. SERVO Transducer position detection, motor drive signal generation, shaking compensationcircuit, transmission timing signal generation, Tx circuit for mechanical sector andannular array probes, transmission power supply voltage monitoring circuit
Mechanical scanner unitSCU-1000
EP400300 MECH. PROBESELECTOR
Two 60-pin connectors, PROBE 1/2 selection, PRE AMP circuit for mechanical andannular array probes, probe code output
EP401200 PHYSIO. AMP ECG AMP, PCG AMP, R-wave detection, PCG filter circuit, hum filter circuit
EP399500 PHYSIO. MEMORY A/D conversion, interpolation circuit, physiological signal memory, memory control,physiological signal synchronization mark generation
EP420700 PHYSIO. PANEL 1 Adjustment trimmers for ECG and PCG sensitivity and display position
Physiological signalunit
PEU-1000
(EU-5042
EP420600 PHYSIO. PANEL 2 ON/OFF switches for ECG, PCG, ECG SYNC and SOUND
Volume mode unit EU-9075 EP429300 DMS ITF/VOL/SERVODMS interface, Video signal input/output circuit, Volume mode calculating block,motor servo circuit
Data management DMS-1000 DMU-200 Data management unitCPU, VCM board, SCSI card, Ethernet card, Floppy disk drive, Hard disk drive
subsystem EP429301 DMS ITF DMS interface, Video signal input/output circuit
Power supply unit PSU-S1000 +5.1V5V15V+15VTV monitor+24VHVAHVBIP-1233* Black and White TV monitor (NTSC)Observation TV
monitor IP-1233*V Black and White TV monitor (PAL)
MN2-0218SECTION 4 Principle of System Operation
4 - 9
5421 3 15421 35421 35421 35421 35421 35421 35421 35421 35421 35421 35421 35421 35421 35421 3K L M N P Q R S T U V W X Y ZO
J12345
I12345
H12345
G12345
F12345
E12345
D12345
C12345
B12345
A5
43
R E
V I
S I
O N
S
SCALE
UNITS mm
DRAWING NO.
MODELTITLE
3RD ANGLE PROJECTION DESINEDDRAWN CHECKED APPD
11
Detector
LPF
EP400700 MAIN AMP
A/DGEU Timing
GEU BUS Window
LineBuffer
FrameMemory
US Interface
GEU Interface
AddressGenerator
DIUMemory
CINEManager
ABC Control
R-DelayControl
R-R
DIU Control Signals
R-Delay GATE
HV CONT
PHYSIOPanel
PHYSIO AMP
EP401200PHYSIO AMP
HUM FilterPCG IN
R-TRIG
DIU Timing
ROMTable
MPU
GDC
EP3990**/EP4305**
CPU
PostProcess
D/A
EP399600CINE MGR
EU-9056(Option)
EP399500
PEU-1000 (Option)PHI
EP3966**DIU
12inchMonitor
VCR(Option)
Printer(Option)
Foot SW (Option)
VIDEO OUT
VIDEO IN
VIDEO OUT
PRINTERCONTROL
VCR
FOOT SWFOOT SW
PR
INT
ACOUTLET
To Mother Board
HV CONT
EP3956** GEU ITF
SSD-1000
Main Panel Assembly
L-KEY-64* EP420600 & EP420700PHYSIO PANEL1 & 2
ECG IN
DIU Control
PSU-S1000
Power Supply Unit
SYSTEM BLOCK DIAGRAM
PEU-1000(Option)PSC-129
USM-23
USI-145
EP400800MECHASERVO
SCU-1000 (Option)
MECHATX
MECHASERVO
PRE STCGenerator
PR
OB
E S
ELE
CT
OR
EP401500
Pro
be
Con
nect
or
VBPFCONTROL
US
BLK
etc
.
EXCHANGEV
Contrast
STC
128-32Selector
EP400300
ME
CH
A R
X
MECHAPROBE SELECTOR
SCU-1000(Option)
ME
CH
APro
be
Con
nect
or
MECHA TX
AnnularPREAMP
MECHAPREAMP
Annular RX
Delay& Adder
AnnularDelay&Adder
Delay& Adder
AnnularDelay&Adder
VBPF LOGAMP
VBPF LOGAMP
PRE AMP
Transmitter
EVEN
ODD
TX Trig.Generator
EP400600RX FOCUS
EP400600RX FOCUS
EP404400
TX/PRE AMP
Fo CONT
Fo CONT
MA4-0026
EP399700
EU-9056(Option)DIU MEM
VideoSelector
IP-1233*/IP-1233*V
DMU-200orDMU-201
DMS-1000(Option)
DMS-1000,EU-9075(Option)
DMS ITF/VOL SERVO
EP4293**
PHI
(Gain,STC ,AGC,AAFL)
MechanicalSectorProbe
Linear orConvexProbe
MotorControl
VOLScanner
ServoCPU
A/DVOLGamma
OpacityTable
VOLCalc.
ImageMemory
FrameCorrelation
ImageCont.
Signal Processor
BrightTable
Register
CPU Data Address
(Ver.2.0 onward
MN2-0218SECTION 4 Principle of System Operation
4 - 10
(Blank page)
MN2-0218SECTION 4 Principle of System Operation
4 - 11
4-4. Principles of System Operation
4-4-1. System Control Unit
Fig. 4-4-1
The system control unit includes the HOST CPU (MC68340) to provide overall system control.The CPU controls all parts of the device based on the software stored in ROM, according to
information from the operation panel.
The local CPU (V50) is mounted in the cine memory unit to provide cine memory control.
CPU
OPERATIONPANEL
GEUBUS WINDOW
GEUI.T.F.
MAINAMP
RXFOCUS
TXPRE AMP
MECH.SERVO
(OPTION)
PROBESELECTOR
MECH. PROBESELECTOR(OPTION)
CPU ADRS. DATA
GEU CPU ADRS. DATA
PANEL CPUADRS. DATA
ARTOP(HD64410)
PHI(OPTION)
CINE MANAGERV50(CPU)(OPTION)
CINEMEMORY(OPTION)
DIU
CPU ADRS. DATA
V50 ADRS. DATA
CPU
(MC68340)
DMS ITF(OPTION)
Data managementsubsystem(OPTION)
DITF ADRS. DATA
MN2-0218SECTION 4 Principle of System Operation
4 - 12
4-4-2. TX/RX Unit
The Tx/Rx unit provides functions to generate transmission pulses, conduct analog
processing of received echoes, and transfer ultrasound image data to the DIU.
The transmission pulses are converted to high-voltage in the driver circuit that drives the
oscillator and are then supplied to the selected probe.
The transmission voltage is detected by the transmission power supply voltage monitoring
circuit on the TX / PRE AMP board and this data is sent to the system control unit. The
system control unit immediately stops transmission if it receives abnormal transmission
power supply data.
The electronic scanning probe has a maximum of 128 Tx/Rx lines, allowing it to
simultaneously drive up to 32 transducers. Up to 32 of the 128 Tx/Rx lines can be selected
by the selector circuit, which comprises 16 HVSs (8-channel High Voltage Analog
Switches).
The received echoes are received by the transducers that transmitted and are amplified in
the pre-amp circuit. The received echoes are analog processed after received dynamic
focusing, then A/D converted and transferred to the DIU.
With an electronic probe, phase compensation is conducted in the matrix switch and delay
line because the phases of the echoes received by the multiple transducers are displaced
from each other.
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
4 - 13
TX/RX PART BLOCK DIAGRAM 1/1
PRE STCCONT
GEU BUS
MECH.SERVO
MECH. PROBECONNECTOR
A, B, C, Z,MOTOR+
GEU BUSWINDOW
USBLK
USBLK GAINCONTROLLER
OPERATIONPANEL
GEU TIMINGGENE.
GEU BUS
STC
ABC CONT. GAIN,From DSC
CPU BUSMECH.
TX TRIG.
MECH.TX
LOGAMP
LOGAMP
F0 CONT
VBPF
VBPF
F0 CONT
TX TRIG.GENE.
TX
32ch
32ch
PROBECONNECTOR
128-32chSELECTOR
128ch
PRE AMP
32ch
MECH.PRE AMP
DELAY&ADDER
DELAY&ADDER
Annulardelay &adder
Annulardelay &adder
A/DAAF
RELIEFCONT-RAST
AGCGAINCONT.
EXCHANGV
GAIN CONT CONTRAST
3
RELIEF
3
AAF CONT
3
DSC VIDEOAnnularPRE AMP
6
AGC
3
MN2-0218SECTION 4 Principle of System Operation
4 - 14
4-4-3. DIU Unit
The DIU uses a high-speed X-Y system.
The ultrasound image data received from the Tx/Rx unit as a digital signal is sent
for pre-processing. Pre-processing involves input selection of the ultrasound image data
sent from the Tx/Rx unit, frame correlation, and storing in frame memory.
Post-processing involves the interpolation and correlation of the B-mode image in the TV
horizontal scan direction, addition of gray-scale bar, characters and graphics, and output to
the D/A conversion circuit. In addition, the DIU unit supplies the ultrasound image data
for ABC (Auto B-Gain Control).
After D/A conversion, TV synchronization and blank signals are added and the signal is
output as a composite video signal.
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
4 - 15
DSC PART BLOCK DIAGRAM 1/1
DSC VIDEOFrom TX/RX part
FRAMEMEMORY
D/A
PHI
PREPROCESS
TO / FROMCONNECTORPANEL
CINEMEMORY
CINEMANAGER
TO TV MONITOR/
ABCCONTROL
ABC CONT.
PHYSIO. SIGNALFrom PHYSIO. AMP
CPU BUS
V50 BUS
CPU BUS
M PHYSIO.
B PHYSIO. SIG.
CPU
MEMORYCONTROLLER
USINTERFACE
DIUCONTROL
CMP_BLKCMP_SYN
GRAY
TIMINGGENERATOR
RASCASWEOE
VIDEOSELECTOR
GDC
POSTPROCESS
TO GEU
DMSITF
DITF BUSTO / FROM DMS
CPU BUS
CPU BUS
MN2-0218SECTION 4 Principle of System Operation
4 - 16
4-4-4. ABC (Auto B-Gain Control) Function
This function evaluates the brightness from the B-mode ultrasound image data stored in
frame memory and automatically maintains the optimal B-mode gain for MAG etc. ,
regardless of the probe type, frequency, and individual differences between test subjects.
The ABC function can be turned ON or OFF and the ABC level can be set at the operation
panel.
The B-mode ultrasound image data stored in frame memory is sent to the ABC control unit,
which evaluates the brightness and outputs an ABC gain level analog signal according to
the ABC setting on the operation panel. Also, the gain volume setting is output by the
ABC controller as the ABC gain. The ABC gain is added to STC and output to the
MAIN AMP.
Fig. 4-4-2
4-4-5. Cine Memory Unit (Option)
The cine memory unit is made up of two PCBs: the controller unit (CINE MANAGER) and
memory unit (CINE MEMORY). Cine memory is a display memory that can store 32 pre-
frozen screens of ultrasound images (in 1B mode).
The cine memory unit offers search, scroll, store, review, and loop playback functions under
the control of the V50 processor on the CINE MANAGER board.
8
FROMPROBE TO
TV MONITOR
ABCCONTROL
OPERATIONPANEL
GAINCONT.
ABC GAIN
STC
GAIN CNT
8
CPU DATA
MAINAMP
A/D DIU D/A
MAIN AMP
DIU
OPERATIONPANEL
GEU ITF
TO CPU
MN2-0218SECTION 4 Principle of System Operation
4 - 17
4-4-6. Mechanical Scanner Unit (Option)
The mechanical scanner unit comprises a controller unit (MECH. SERVO) and selector
unit (MECH. PROBE SELECTOR). The transmission circuit is mounted on the MAIN
AMP board.
The transmission pulses generated by MAIN AMP are converted to high-voltage in the
driver circuit that drives the oscillator and are then supplied to the selected probe.The transmission voltage is detected by the transmission power supply voltage monitoringcircuit on the MAIN AMP board and this data is sent to the system control unit. Thesystem control unit immediately stops transmission if it receives abnormal transmissionvoltage data.The received echoes are amplified by the PRE AMP circuit on MECH. PROBE
SELECTOR and supplied to the RX FOCUS board.
Fig. 4-4-3
MECH.TX
GENE.
SERVOCONT.
RX
FOCUS
PROBE
CONNECTOR
MECHA. SERVO MECHA. PROBE SELECTOR RX FOCUS
TO MAIN AMP
A, B, C, Z,MOTOR+
TX TRIG
GEU BUS PRE
AMP.
ANNULARPREAMP.
MN2-0218SECTION 4 Principle of System Operation
4 - 18
4-4-7. Physiological Signal Unit (option)
The physiological signal unit is made up of the physiological amplifier unit (PHYSIO
AMP), physio. panel, and physiological signal memory unit (PHI).
ECG (electrocardiogram) and PCG (phonocardiogram) are input to PHYSIO AMP, where
they are amplified by amplifier circuits with isolated power supplies, and are output by the
photo-couplers as the isolated signals B_ECG, M_ECG, and M_PCG.
For noise elimination, the ECG circuit contains the HUM FILTER (50 Hz/60 Hz) and the
PCG circuit has a three-level PCG FILTER.
In addition, the ECG R-wave detection circuit (R-DETECT) is mounted on PHYSIO AMP.
The PHI board A/D converts the physiological signals.
The B_ECG signal is interpolated and stored in the physiological signal memory. The
stored signal is synchronized with the TV signal, read, added in the DIU post-processing
unit, and displayed on the plane mode (B mode) image.
The M_ECG and M_PCG signals are interpolated, written to the physiological signal
storage area of the DIU display memory, and displayed on the sweep mode (M mode)
image.
Fig. 4-4-4
PHIOPERATION
PANEL
PHYSIO. AMP.
ECGM_ECG
B_ECG
R-GATE TOGEU ITF
TOPHI
LIMITER
R-DETECT.
PHOTO-COUPLER
HUMFILTER
ECGAMP
M_PCGPCG TOPHI
LIMITERPCGFILTER
PHOTO-COUPLER
PCGAMP
HUM FL
PCG FLOPERATION
PANEL
M_ECG
B_ECG
M_PCG
A/D
ECGMEMORY
INTER-POLATION
INTER-POLATION
FRAMEMEMORY
POSTPROCESS
M_PCGM_ECG
B_ECG
MN2-0218SECTION 4 Principle of System Operation
4 - 19
4-4-8. Power Supply Unit
The power supply unit comprises a series power supply and switching power supply.
It supplies the power required by the main unit. It also has an isolated power supply for
the optional recording device.The transmission voltage for the electronic scanning probe (HVA), and mechanical scannerprobe (HVB) is controlled by the control signals (HVS DATA0-5) from the main unit.
4-4-9. Volume mode unit (option)
This unit is designed to display Volume mode image on a monitor.
4-4-10. Data Management Subsystem (option)
This unit is for handling DICOM communications and DICOM specification image filing.
MN2-0218SECTION 4 Principle of System Operation
4 - 20
4-5. Principles of PCB Operation
4-5-1. PROBE SELECTOR
This board comprises probe connectors, probe select relays, HVS, and the relevant control
circuits.
LIN/CONV CONNECTOR 1
260-pin connector for linear/convex probe.
LIN/CONV CONNECTOR 2
260-pin connector for linear/convex probe.
RELAY
Comprises 64 relays to switch between PROBE 1 and PROBE 2.
HVS
Comprises 16 HVSs (8-channel High Voltage Analog Switches) to select and
switch 32 of 128 channels
CONTROL & PROBE CODE INTERFACE
Comprises ROM (Read-Only Memory), logic circuits, relays, and PLDs
(Programmable Logic Devices). It controls the probe switching relays and HVSs
and outputs probe codes to the data bus.
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
4 - 21
TITLEPROBE SELECTOR 1/1
MODEL
EP4015**
RELAY
LIN/CONVCONNECTOR 1
PROBEON1PROBEON2M_E_SELRAMEN
GEU DATA(0-7)
CONTROL &PROBE CODEINTERFACE
TD1(1-128)
GEU ADRS (0-3)EN1EN2EN3READHVDTCLKHVADRS(0-2)
GEU BUS
LIN/CONVCONNECTOR 2
HVS
TD1(1-128)
TD1(128-1)TXRX (1-32)
PROBE CODE1 (0-9)
PROBE CODE2 (0-9)
HV DATA
RSSR
PROBE CODE2 (0-9)
MN2-0218SECTION 4 Principle of System Operation
4 - 22
SIGNAL LIST ( PROBE SELECTOR )
J100-1 J100-2
MN2-0218SECTION 4 Principle of System Operation
4 - 23
(Blank page)
MN2-0218SECTION 4 Principle of System Operation
4 - 24
4-5-2. TX/PRE AMP
This board comprises 32 Tx/Rx circuits and transmission trigger signal generation circuits to
carry out ultrasound Tx/Rx.
TX POWER SUPPLY VOLTAGE DETECTOR
A/D converts the transmission power supply voltage and outputs it to the data bus.
TX
Carries out electronic probe transmission.
(Max. simultaneously driven channels: 32. Max. transmission voltage: 90 V)
TX TRIGGER
Generates the transmission triggers for electronic probes (max. 32 channels).
TX FOCUS CONTROL
Generates the control signals to control electronic transmission trigger generation
for electronic probes, and the PROBE SELECTOR control signals.
PRE STC DRIVER
Converts the STC CNT and GND CNT signals to positive/negative signals and
outputs them to the PRE AMP.
PRE AMP
Conducts first-stage amplification of the received signals from the electronic probes
according to the STC_P and STC_N signals generated by the PRE STC DRIVER.
MN2-021
8SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
4 - 25
1/1MODEL
EP4044**TITLE
TX / PRE AMP
TX TRIGGENE. TX
TX FOCUSCONT.
TXVDETECTOR
TXCLKTXLD_ EN1USADRS EN2GEUDATA READGEUADRS TXFADRSUSBLK BSADRS
STCCNT
USBLKFrom GEU I.T.F.
TP1
32
AXENA1ENA2DSCLKPSTXCLRLDTXCLK1
PRESTCDRIV.
TRIG
32
TXRX
STC_PSTC_N
HHV
GE8GEUDATA
HVLEHVCLKHVDTCLKHVSWCLRHVADRSRESET
From MAIN AMP
From GEU I.T.F.
To GEU I.T.F
To PROBE SELECTOR
PRE AMP32
PRE OUT To RX FOCUS
From MAIN AMP
To PROBE SELECTOR
GEU BUS
GEU BUS
MN2-0218SECTION 4 Principle of System Operation
4 - 26
SIGNAL LIST ( TX/PRE AMP )
J101-1 J101-2
J101-3
MN2-0218SECTION 4 Principle of System Operation
4 - 27
(Blank page)
MN2-0218SECTION 4 Principle of System Operation
4 - 28
4-5-3. RX FOCUS
Two of these boards are used to alternately conduct receive focussing and receive dynamic
focussing.
V/I CONVERTER
Converts the PRE AMP voltage output to a current.
CROSS POINT SWITCH
The CROSS POINT SWITCH connects the received signals (max. 32 channels) to
the optimal delay tap.
DELAY & ADDER
Conducts delay and adding (receive focus) on the received signals (max. 32
channels) and outputs the received signals as one line.
LIN/CONV H.P.F. (High pass Filter)
Allows only the high-frequency component of the received signals to pass when an
electronic high-frequency probe is used.
MECH. H.P.F. (High pass Filter)
Allows only the high-frequency component of the received signals to pass when a
mechanical scanner high-frequency probe is used.
ANALOG SWITCH
Selects the received signals from the electronic or mechanical scanner probe, turns
ON/OFF the Doppler output and H.P.F.
V.B.P.F. (Variable Band Pass Filter)
Allows the optimum frequency signals for the probe frequency and received signal
depth to pass.
LOG. AMP
Comprises two +30 dB amplifiers and LOG ICs to carry out logarithmic
amplification of the received signals.
CROSS POINT SWITCH CONTROL
Generates cross point switch control signals and analog switch control signals
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
4 - 29
TITLERX FOCUS 1/1
MODELEP4006**
From TX/PRE AMP
V/I32
PRE OUT DELAY &ADDER
CROSSPOINT SW
CROSSPOINT SW
CONT
F0 CNT
MECH. RX
From MECH.PROBECONNECTOR
H.P.F.
ANALOGSWITCH
H.P.F.
LCOFFHPF ON
To MAIN AMP.
TP5
LOG0* PLOG0*N
LOG AMPV.B.P.F
TP10 TP12/13
AXAYMRSTBCS
From MAIN AMP
FCSE_0 RESETEXCHNG USBLKBSADRS USADRSEN1 GEUDATAEN2 GEUADRSREAD
GEU BUS
MECH. RXANNULAR ANALOG
SWITCHDELAY &ADDERV/I
ANNON
MN2-0218SECTION 4 Principle of System Operation
4 - 30
SIGNAL LIST ( RX FOCUS 1 )
J102-1 J102-2
J102-3
MN2-0218SECTION 4 Principle of System Operation
4 - 31
SIGNAL LIST ( RX FOCUS2 )
J103-1 J103-2
J103-3
MN2-0218SECTION 4 Principle of System Operation
4 - 32
4-5-4. MAIN AMP
This board switches the receive focus, conducts ultrasound analog processing, and generates
the mechanical scanner transmission pulses.
RX EXCHANGE
Receives the received signals from the two RX FOCUS boards while switching the
depth in approximately 1 cm increments. The received signal is switched by
analog ICs on both the rising edge and falling edge of the EXCHANGV signal.
DETECTOR
Detects the received signal and adds the gain signal.
AGC (Automatic Gain Control)
Comprises delay circuits. Subtracts the delayed received signal from the original
received signal. Eight levels can be selected by the AGC 0 to 2 control signals.
CONTRAST
Comprises resistive potential divider circuits to output a potential-divided signal.
Eight levels can be selected.
RELIEF
Comprises differentiation circuits. Adds the differentiated received signal to the
original signal. Eight levels can be selected.
AAF (Anti-Aliasing filter)
Eliminates the aliasing that occurs when monochrome image data is sampled.
GAIN CONTROLER
Generates the gain, F0 (6 types), and PRE STC (6 types) signals.
4 -
33
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
ystem
Opera
tion
TITLEMAIN AMP 1/1
MODELEP4007**
USBLK2MCLKREADEN1-3
M
SELECTOR
STC (0-7)
B GAIN
GAINCONTROLLER
GAIN STCDATAGENE.
SELECTOR
A/DA/DA/D
B GAIN. STC 0. DLV.
DVD
TP8
GBS
SOS
DVS
CONFDOP
M SEL
To RX
From DIU From OPERATION PANEL
From GEU I.T.F.
D/A
FO/PRE STCGENE.
D/A
D/APRE STC CONT.
F0 CONT.
TP9 To TX / PRE AMP
From GEU I.T.F.
DETECTOR
AGC
CONTRAST
RELIEF
AAFSELECTOR
3 AGC 3 AAF3 RELF3 CONT
TP1 TP13
TP7
TP3TP5TP4
LOG 01
LOG 02
AAF OUT
From RX FOCUS 1, 2To GEU I.T.F.
GEU DATA (07)
GEU ADRS(03)
GAIN CONT.
MN2-0218SECTION 4 Principle of System Operation
4 - 34
SIGANAL LIST ( MAIN AMP )
J411
1 VIDEO_OUT
2 GND
J104-1 J104-3
MN2-0218SECTION 4 Principle of System Operation
4 - 35
(Blank page)
MN2-0218SECTION 4 Principle of System Operation
4 - 36
4-5-5. GEU I.T.F.
This board generates the control signals for the GEU unit, generates the memory write
control signals for the DIU unit, and opens the bus between the CPU BUS and the GEU
BUS and PANEL BUS.
CPU BUS WINDOW
Controls opening between the CPU BUS and the GEU BUS and PANEL BUS.
PROBE CONNECTOR DETECTOR
Detects whether a probe is connected and outputs this data to the CPU BUS. The 5-
bit data has 2 bits for linear/convex probes, 2 bits for mechanical scanner probes,
and 1 bit for the independent probe. The corresponding data bit becomes LOW
when a probe is connected.
FRAME RATE CALCULATOR
The frame interval is counted from the 100sec clock signal by a 16-bit counterand the result is output to the CPU BUS. When the count is complete, the IRQ/
signal is output to apply an interrupt to the CPU. The CPU calculates the frame
rate from the counted data.
HEART RATE CALUCULATOR
Each time the optional physiological signal unit inputs the R/ signal, the IRQ/ signal
is output to apply an interrupt to the CPU. The CPU calculates the heart rate from
the data.
MODE CONTROL
Generates PROCOD and SCNMOD signals corresponding to the mode when mode
data is sent from the CPU.
PROCOD SCNMODB MODE 0 1M MODE 0 0D MODE 1 0 PROCOD : Process code(Not used) 1 1 SCNMOD : Scan Mode
MN2-0218SECTION 4 Principle of System Operation
4 - 37
BEAM ADDRESS
Outputs addresses in the B mode by counting a clock signal synchronized with an
ultrasound line using a 9-bit counter in which the CPU start and end values can be
set.
HV CONTROL
Generates the HVS DATA 0 to 7 signals from the high-voltage (transmission) data
sent from the CPU. This block also contains a CPU runaway detector circuit,
which sets all the HVS DATA 0 to 7 signals to HIGH level to stop transmission if a
CPU lock-up occurs.
TIMING GENERATOR
Generates all timing signals from the 12 MHz, 1 MHz, and 0.5 MHz clock signals
produced by dividing the 60 MHz/48 MHz clock frequency from the crystal
oscillator, according to the data set by the CPU. Operates at 48 MHz with a 2 MHz
probe but operates at 60 MHz with all other frequency probes.
A/D converter
US video (VIDEO_OUT) signal, which is sent from MAIN AMP board, is
converted into 6 bits of digital signal, and sent to DIU board.
This circuit is also used for A/D convert of VCR playback signal.
MN2-02
18
SEC
TIO
N 4
Prin
ciple
of S
yste
m O
pera
tion
4 - 38
TX CLK/
US BLK/
DF AREA
OFCENT/
96 PRF
D SMPL
HEARTRATE
CALCULATOR
PROBECONNECTION
DETECTOR
BEAMADDRESS
FRAMERATE
CALCULATOR
MODECONTROL
Line Buffer(FIFO)
A/Dconverter
HVCONTROL
TIMINGGENERATOR
CPU BUS WINDOW
IRQ/R/
0.5M
12M
1M
PROBE ON/
FRMEND/
IRQ/
PROCOD/
SCNMOD
USADRS(0-7)
DSCADRS(0-7)
HVS DATA
GEU BUS
PNL BUS
CPU BUS
TITLEGEU I.T.F. 1/1
MODELEP3956**
US video (VIDEO_OUT)6
ADDAT(0-5)VCR_IN
6
VCR_IN
MN2-0218SECTION 4 Principle of System Operation
4 - 39
SIGNAL LIST ( GEU ITF )
J207-1 J207-2
J207-3
J314
1 VIDEO_IN
2 GND
J315
1 VCR_IN
2 VCR_IN_R
3 NC
4 NC
J316
1 VCR_OUT
2 VCR_OUT_R
3 NC
MN2-0218SECTION 4 Principle of System Operation
4 - 40
4-5-6. DIU
The B/W DIU (Digital Imaging Unit) fetches ultrasound wave digital data coming fromthe GEU in synchronized with the DIU clock, writing ultrasound wave data to the displaymemory in XY writing method, then reading out the ultrasound wave data in synchronizedwith the TV signal. After this, gradation conversion of display data, addition of gray scalebar and other post processing are executed by this block.
In the writing system, ultrasound wave scanning data are thought of as scanning linevectors, and from the starting point indicated by the vector (X0, Y0) and the time it takes topass the boundary (X), (Y), one scanning line can freely be written to main memory.
Also, through the dynamic focus area signal (DFAREA), multi-level dynamic focusedultrasound wave data can be written to memory.
The boundary passing signal (INCX, INCY) is detected by the XY Boundary Calculatorfrom the boundary passing time (X), (Y) in 1 pixel units. If the boundary passing signal isdetected, data are written to the line buffer the following system is used to determine if thesignal is INCX or INCY by the scanning address and register setting.
(a) XY Sampling Writing System (X and Y System)In this system, writing to the line buffer is done each time either INCX or INCY is
generated. This system is used for all scanning lines.
(b) X Sampling, Y Sampling Writing System (X or Y System)In this write system, the line buffer write area (X system) is set when INCX is
generated and the line buffer write area (Y system) is set when INCY is generated.Data are written to the area specified in the X system each time INCX is generated,
while writing is not performed when INCY is generated. INCY information is held untilINCX is generated, then writing of both is done simultaneously. For this reason, thenumber of times INCY is generated must be fewer than the times INCX is generated.Accordingly, the area which can be specified in the X system is limited to 45 or more leftand right. In the same way, the area which can be specified in the Y system is limited to45 or less left and right.
Also, the area setting is stored in X0 and Y0 (16-bit) in the parameter table.The boundary pass signal (INCX, INCY) is written to the line buffer simultaneously
with the ultrasound wave data, and is used to increment the main memory write address.
ABC (Auto B-gain Control)This block generates B_GAIN signal, which controls the gain for B-mode, referring
ultrasound brightness data in the image memory and the Gain Level setting by operationpanel. Then B_GAIN signal is sent to MAIN AMP board, and controls the gain ofreceiving signal. In case ABC off, B_GAIN becomes the signal referring B GAIN knobsetting.
PRE PROCESS
This block is a circuit that executes various types of processing of ultrasound imagedata sent and received at the GEU for writing to MEMORY.
Test Pattern GenerationGenerates the various types of test pattern.
MN2-0218SECTION 4 Principle of System Operation
4 - 41
Echo EraseIn order to make physiological signals easy to see, the bottom portion of ultrasoundM image data is erased. The erased portion of ultrasound image data is written toimage memory as "0" level. This differs from USERASE and ERASE in the pointthat it is executed when ultrasound image data are stored in image memory.
Memory EraseIn order not to display unnecessary data when changing modes, etc., either a portionor all of the image memory is erased. The erased portion of ultrasound image data iswritten to image memory as "0" level.Except in the point that Erase is executed in screen units, which image memory isconfigured into, this function is the same as Echo Erase.
Black and White Frame CorrelationCorrelation between two frames is executed in order to reduce noise.
CPU AccessWhen making a histogram, etc. of ultrasound image data, the CPU can access imagememory. Access by the CPU is limited to times when the ultrasound image isfrozen.
MEMORY
This block is capable of storing enough ultrasound image data for two screens afterthey have been processed by the pre process block.
Scanning conversion in the DIU is executed during writing to or reading from thisblock.
Main memory is used as M image display memory during real operation and as framememory for frame correlation. During B image display and when the image is frozen,CINE memory is used as display memory.
Memory Configuration
2 x 2 pixels is the smallest configuration element and 2 screens of 512 x 512 pixelx 8 bit ultrasound image data are stored.
Black and White Main Memory
bits 0-5 : Black and White Image Data
bit 6 : Organism signal for M
bit 7 : Sync. Mark for M organisms.
Write/Read SpeedWriting to or reading from RAM is carried out by RMW at 320 nsec/1 pixel.
However, by shifting the access cycle one half period, apparently, RMW is done at160 nsec/1 pixel. On the other hand, reading from the SAM block is by serial Readat 160 nsec/1 pixel. By doing 2 pixels simultaneously, apparently, serial read can bedone at 80 nsec/1 pixel.
EraseUsing the Dual-port RAM flash write function, the image memory can be erased.
MN2-0218SECTION 4 Principle of System Operation
4 - 42
Erasing is executed by flash write at 160 nsec/1 Row, but when 4 chips are workingsimultaneously, erasing can be done at high speed, with one 512 x 512 pixel screenerased in approx. 45 sec. Furthermore, since it is not dependent on ultrasoundimage data write operations, it differs from USERASE, ECHO ERASE and MemoryErase.
POST PROCESS
This block carries out various types of processing of ultrasound image data read fromimage memory.
H InterpolationSpaces between lines of ultrasound B image data are interpolated in the TV H
direction. The area which can be interpolated is from 1-11 pixels, and selection ofthe number of pixels to be interpolated is set by the register.
Ordinarily, the linear interpolation method is used. For black and white anddispersion, linear interpolation is used.
Also, since the display area of ultrasound B image data is dependent on thedisplay format, it is controlled by TV synchronized signals.
Furthermore, ultrasound M image data do not pass through this processingsystem.
H CorrelationIn order to show connection of ultrasound image data in the TV H direction
smoothly, the arithmetic sum is taken between pixels in the TV H direction. Thesystem is equipped with independent correlation circuits for ultrasound B image dataand for M image data, with selection of whether these circuits are switched ON orOFF being set by the register. Also, since the display area of ultrasound image datais dependent on the display format, it is controlled by TV synchronization signals.
Gray Scale Bar GenerationA gray scale bar with 64 gradations ranging from the white level to the black level is
generated in the TV V direction. The display width of each gradation is fixed at 16 pixelsx 2 lines. Also, the display position is dependent on the display format, so it is controlledby TV synchronization signals.
View Gamma correction is carried out in accordance with the imaging device's
brightness characteristics. The following three types of gamma curve can beselected.
(1) Standard (Linear)(2) For ABDOM(3) For CARDIAC
There are two types (mainly for ultrasound B image data and for ultrasound Mimage data) which can be held simultaneously. This function shares the Enhance
MN2-0218SECTION 4 Principle of System Operation
4 - 43
RAM, so selection is made by rewriting the Enhance RAM. Furthermore, forultrasound M image data, Standard (Linear) is selected automatically.
This function is implemented for black and white images only.
Overlay displayThis is for displaying character and graphic. Two input circuits of OVL0 and
OVL1 are prepared for character and graphic signals allowing two kinds ofluminance level displaying. ON/OFF of each displaying and luminance level are setby the register. The following shows priority among overlay signals to bedisplayed.
(1) OVL0 High(2) OVL1 |(3) Gray scale bar |(4) Ultrasound image data Low
In case there is no overlapping between OVL0 and OVL1, shadow appears on theright side.
GEU I.T.F.This block generates and controls parameter information for fetching ultrasound
data to the DIU. It also carries out movement control and freeze control.
Parameter ROMContains ultrasound scanning line vector information (parameters) for each probe.
This ROM has a capacity of 512 Kbytes, enough for 42 probes.
Reading and Conversion of Parameters(a) Reading of ROM Data
Ultrasound scanning line vector information (parameters) X0, Y0, WXF,WYF, X and Y, are sent to the control block each time a scanning linewrite start signal (Line Start) arrives.
Data corresponding to DSCADRS (8:0), which are sent from the GEU, areread according to the type of probe and display magnification.
It also performs conversion of parameters for vertical reversal, horizontalreversal, display mode and probe code, in accordance with the read out data.Parameter conversion is accomplished by hardware according to the displaymode and control register in all cases.
(b) Parameter ConversionSince each unit of parameter data is 16 bits in length, the high byte and low
byte are read independently from ROM, then latched and sent to each block.Vertical reversal and horizontal reversal display modes and conversion ofparameters (X0, Y0) in accordance with probe codes is done at this time.
Parameter ROM for SoftwareThe system includes parameter ROM for reference by software. It is
accessible by the CPU at any time.
Scanning Area Display
MN2-0218SECTION 4 Principle of System Operation
4 - 44
This function makes the B image display area narrow and increases theframe rate. This is accomplished by varying BOF/, EOF/ and DSCADRS (8:0)in the GEU. Data from BOF/ to EOF/ are written as DIU.
Freeze FunctionThis function stops the writing of an image to display memory.Freeze is executed by setting the register. In the case of freezing of a real
display simultaneous with B/M, it is necessary to set the Freeze Register inadvance.
Forced Freeze, where writing to memory is forcibly stopped, andsynchronous freeze mode, which is synchronized with the ultrasound writeimage, are also provided.
B image synchronous freeze stops motion after one frame of B image datahas been written. Ordinarily, the sweep image in the B/M sweep motion modeis frozen as a B image. Also, synchronous freeze which is independent of Band sweep can be set. Sweep image synchronous freeze stops motion afterwriting of one line of sweep image data is written. When in the B Mode, theimage is frozen after writing of the B image is completed.
When in the Sweep Mode, the image is frozen when writing of a sweepimage line is completed.
When in the B/M Sweep Mode, B is frozen after writing of one frame, andthe M sweep image is frozen after freezing of B is completed.
WRITE CONTROL
This block is configured from a block which generates write control signals andaddresses when image data read from the line buffer is written to display memory andwhich controls erasing of display memory.
Image data (8-bit) read from the line buffer contain pixel border passing signals (INCX,INCY) in the X and Y directions. Write addresses are generated by operating the addresscounter for each direction by means of this signal. Setting of this address counter's initialvalue and controlling the up-down count makes it possible to control the image direction.
Line Buffer Interface FunctionThis function returns a Write End signal for handshaking in response to
memory write request signals from the line buffer.
Write Control Signal Generator FunctionThis function generates a write memory cycle (Read Modify Write) which
writes ultrasound data sent from the line buffer to memory. There are twosystems which generate this cycle, making it possible to repeat writing of thenext requested pixels in a 160 nsec. cycle. This is accomplished byoverlapping the 320 nsec. cycle of the two Read Modify Write systems by half.
Off Screen FunctionThe write address in both the X and Y directions is 12 bits, and a virtual
MN2-0218SECTION 4 Principle of System Operation
4 - 45
image can be developed in a 4096 x 4096 area, but in reality, the memory areawhere ultrasound data can be written is only 512 x 512 in size. Therefore, thevirtual write area is regarded as off screen area and the writing operation is notperformed. However, since the virtual write area also operates the addresscounter, the XY Boundary Calculator's INCX and INCY are used as is in the offscreen area. After entering the memory area (on screen area), fetching fromline buffer memory begins and writing to actual memory is done, and INCX andINCY, which are added to the memory write data, operate the write addresscounter.
MEMORY CYCLE CONTROL
In this block, control of writing to and reading from memory is executed in accordancewith memory control signals generated by WRITE CONTROL and DISPLAY CONTROL.Also, this block includes the function for reading and writing of display memory data fromthe CPU during a freeze operation, and generates memory cycles and addresses foraccessing direct image memory by the CPU. It sets the X direction address (9 bit) and Ydirection address (9 bit) of the data to be accessed, specifies the surface and setsincrement/decrement. After setting is completed, the CPU can read and write to theaccess data register.
If the CPU is accessing continuous pixels in the vertical direction, the auto incrementfunction, which makes resetting of the address (Y direction) unnecessary after each access,operates. After the initial access address is set, the CPU can read the data for 1 line in thevertical direction continuously.
DISPLAY CONTROLThis block reads image data written in main memory in synch with the TV signal and
carries out TV display.Since main memory uses dual port memory, writing to RAM and reading from SAM is
done asynchronously. Therefore, TV display of image data can be done according to thepredetermined display format, reading of the contents of the SAM block can be executedand can be displayed by Serial Reading.
Display Mode (One Screen Display Function, Complex Display Function)
The following display controls can be performed.
1B2BWide 2B1MLeft-Right B/MWide Left-Right B/M
In the case of one screen display, in accordance with the display address determinedby the display mode, transfer to the SAM block of memory is executed before TV
MN2-0218SECTION 4 Principle of System Operation
4 - 46
display begins, then in accordance with the display area signal, reading of display datafrom the SAM block is carried out.
In the case of complex display, in order to display Left-Right 2 screen mainmemory according to the display format, while TV display of screen 1 is being carriedout, data for display of screen 2 are being transferred to the SAM block (Real TimeRead Transfer) to carry out complex display.
Scroll Display FunctionThis function causes the latest write data of the sweep image to be constantly
displayed at the far right of the TV display screen. The scroll display transferaddress is generated by deleting the value of the TV display width from the sweepimage data's write address, and fetching the data in the resulting address in synchwith the TV field signal (VD).
Request Display FunctionIn the co