+ All Categories
Home > Documents > Stable, tunable calibration source for large-beam optical systems

Stable, tunable calibration source for large-beam optical systems

Date post: 03-Oct-2016
Category:
Upload: vicky
View: 214 times
Download: 0 times
Share this document with a friend
7

Click here to load reader

Transcript
Page 1: Stable, tunable calibration source for large-beam optical systems

Stable, tunable calibration source for large-beam opticalsystems

Todd G. Ulmer, Amy C. Raudenbush, and Vicky Loriaux

An optical calibration source for free-space systems with large beams is presented. It produces a 2 W,370 mm beam tunable from 1545–1570 nm with a wavefront error of ���15 peak to valley and �0.013�

rms at 1550 nm, power stability of ��0.02 dB, arbitrarily selected polarization, and polarization sta-bility of ��1°�h in orientation and ��0.02�h in ellipticity. © 2007 Optical Society of America

OCIS codes: 120.4640, 120.4820, 350.4800, 060.2310.

1. Introduction

Characterization of optical systems with beams largerthan a few inches in diameter is hampered by manyfactors, including accumulated wavefront error, lowpower densities, and power instability introduced byatmospheric turbulence in the laboratory environ-ment. Precise measurements require an optical refer-ence source with pristine wavefront quality that isstable in both power and polarization so that any ob-served deviations from the ideal can be accurately as-cribed to the system under test. Multiwatt powers aredesirable for sufficient illumination of diagnostic sen-sors and for convenience in aligning large infraredbeams. Furthermore, the generation of arbitrary po-larization states is needed for assessment of polariza-tion dependencies such as polarization-dependent loss.Finally, wavelength tunability is required for manyapplications.

In particular, large-beam systems incorporating awavefront sensor can benefit from a stable, multiwattoptical source. Instruments for characterizing thewavefront error of an optical system can be dividedinto two categories: double pass (e.g., an interferome-ter) and single pass (e.g., a wavefront sensor). Thegeometry of many systems precludes the use of double-pass configurations; additionally, interferometric tech-niques can become problematic in systems with lowpower densities. The preferred solution for these ap-plications is a single-pass configuration using a wave-

front sensor, which requires a stable optical source forthe generation of reference files. Optical power is at apremium in large-beam systems where the wavefrontsensor is a diagnostic tool tapped off the main signalpath, and thus a stable, high-power, high-quality op-tical source is invaluable.

Here we present an optical calibration source thatachieves the performance shown in Table 1. The cal-ibration source is installed in a testbed in an opticaltesting facility that is used to characterize large op-tical systems. Measurements made with the calibra-tion source require stability on the order of seconds tominutes; as will be shown in Section 3, this level ofperformance is readily achieved.

2. Method

A. Overview

The calibration source is shown in Fig. 1. To mini-mize the wavefront error, the approach taken is tolimit the number of free-space components and toutilize fiber-optic components to implement the de-sired functionality, including laser amplification,heat-source remoting, alignment transfer, and po-larization state generation. As shown in Fig. 1, lightis launched from an angle-polished single-mode op-tical fiber, reflected by a single fold flat, and thencollimated by an overfilled off-axis parabola (OAP)to create a quasi-Gaussian beam that is truncatedat the 9% intensity point. In addition to providing aconvenient optical point source, the fiber can berepositioned readily to allow access to a 633 nmFizeau interferometer used for alignment of the OAP.

B. Fiber-Optic Components

The laser source is a master oscillator–power am-plifier configuration. The master oscillator consistsof a pair of sampled-grating distributed Bragg re-

The authors are with MIT Lincoln Laboratory, 244 Wood Street,Lexington, Massachusetts 02420, USA. T. Ulmer’s e-mail addressis [email protected].

Received 31 July 2006; accepted 1 September 2006; posted 5October 2006 (Doc. ID 73600); published 25 January 2007.

0003-6935/07/050669-07$15.00/0© 2007 Optical Society of America

10 February 2007 � Vol. 46, No. 5 � APPLIED OPTICS 669

Page 2: Stable, tunable calibration source for large-beam optical systems

flector (SG-DBR) tunable diode lasers1 that togetherspan the desired wavelength range, and the poweramplifier is a dual-clad Er:Yb fiber amplifier. Al-though several individual tunable laser sources (e.g.,external-cavity lasers) are available that would spanthe entire wavelength range, two factors led to theselection of the pair of SG-DBR lasers. First, thespare laser can serve as a holding beam to preventdropouts in input power to the amplifier while theprimary laser is being tuned to a new wavelength;thus, the amplifier can remain operational at nomi-nal power during tuning to minimize thermal pertur-bations, enhancing power stability. Second, therelatively broad linewidth of the SG-DBR tunablediode lasers �5–10 MHz� helps minimize stimulatedBrillouin scattering (SBS) in the fiber following thefiber amplifier. The reduced SBS allows remoting ofthe fiber components as far as several meters fromthe main free-space optical system to minimize airturbulence from heat sources such as the power am-plifier. Traditional methods of SBS suppression suchas current dithering or optical modulation wereavoided here for reasons of stability and simplicity.Figure 2 shows an experimental comparison of theroom-temperature SBS threshold in slow-axis-alignedPANDA polarization-maintaining fiber at 1559.8 nmfor various commercial tunable laser sources; theSBS threshold for the SG-DBR laser is approximately

twice that of any of the four external-cavity laserstested.

An active polarization controller based on a lithiumniobate actuator2,3 is used to track out residual po-larization fluctuations4 in the fiber amplifier, which isa dual-axis polarization-maintaining design. An errorsignal is derived from the second port of a high-powermicro-optic polarization beam splitter (PBS). Theerror-signal power is scaled using two 1% fiber tapsas attenuators, and an electrical error signal to drivethe feedback circuit is generated using a 100 kHzphotoreceiver. The PBS establishes a linear polariza-tion with a polarization extinction ratio in excess of25 dB as the input to the remainder of the calibrationsource, and the active polarization controller ensuresthat the power passing through the PBS is nominallyas stable as the power generated directly by the fiberamplifier.

A high-power micro-optic circulator with polarization-maintaining pigtails on ports 1 and 2 transportsa well-defined linear state of polarization to the

Table 1. Calibration Source Performance

Parameter Performance

Beam diameter 370 mm 1�e2 diameter, truncatedat 406 mm

Wavelength range 1545–1570 nmWavefront error ���15 peak to valley, 0.013� rms

at 1550 nmMaximum power �2 WPower stability ��0.02 dBDynamic range �20 dBPolarization arbitraryPolarization stability ��1°�h in orientation

��0.02�h in ellipticity

Fig. 1. Calibration source layout. OAP, off-axis parabola; EDFA, erbium-doped fiber amplifier; PBS, polarization beam splitter; OSA,optical spectrum analyzer.

Fig. 2. SBS threshold at 1559.8 nm for various tunable lasersources in slow-axis-aligned PANDA fiber. ECL, external-cavitylaser; SG-DBR, sampled-grating distributed Bragg reflector laser.

670 APPLIED OPTICS � Vol. 46, No. 5 � 10 February 2007

Page 3: Stable, tunable calibration source for large-beam optical systems

non-polarization-maintaining input fiber of the sec-ond polarization controller, which functions as apolarization-state generator. The third port of thecirculator provides a backreflection monitor paththat is used for OAP alignment via a retroreflector;in addition, this path provides a means of monitor-ing for signs of SBS. The circulator also isolates thefiber amplifier from reflections.

The polarization-state generator is a fiber-squeezerdesign5 with minimal insertion loss, enabling high-power operation, and is used in conjunction with apolarimeter in the free-space optical testbed to adjustthe polarization to any desired state. It is also capableof high-speed operation for applications that requirepolarization scrambling. The single-mode fiber on ei-ther side of the polarization controller is kept to acombined length of �3 m and secured to minimizelong-term polarization drift. The final components inthe fiber-optic path are additional single-mode cou-plers to provide monitor points for outgoing powerand wavelength.

The fiber-optic components are mounted on a61 cm � 92 cm aluminum plate. Other than securingcomponents with standard screws and adhesive, nospecial precautions were necessary with respect tovibration or thermal isolation to achieve the stableperformance described in Section 3. With the excep-tion of the polarization controller fiber and the 1% tap

that follows, all components in the primary signalpath are polarization maintaining.

C. Free-Space Optics

The free-space optics consist of a high-quality foldflat and an OAP. The fold flat is a fused-silica sub-strate coated with protected silver to provide a sur-face figure of 0.048� peak to valley and 0.008� rms.The OAP is a Zerodur substrate coated with en-hanced gold. The surface figure of the OAP, over a419 mm clear aperture, is 0.050� peak to valley and0.006� rms. The reflectance variation between verti-cal and horizontal polarization states is less than�0.1% for each mirror, which is easily achieved be-cause the angle of incidence on both mirrors is lessthan 20°.

D. Alignment Procedure

The process used to align the calibration beam isshown in Fig. 3, and the associated alignment budgetsfor each step are shown in Table 2. Alignment of anOAP to an interferometer is a well-known procedure;the main challenge for creating a large collimatedbeam with the approach used here is transferring thisalignment to the fiber source. This alignment transferis accomplished using a spherical reference mirror tocoalign the interferometer and the fiber. First, the633 nm interferometer is aligned to the OAP so that

Fig. 3. Alignment procedure.

10 February 2007 � Vol. 46, No. 5 � APPLIED OPTICS 671

Page 4: Stable, tunable calibration source for large-beam optical systems

the focus of the interferometer is located at the realfocus of the OAP. Without moving the interferometer,the spherical reference mirror is inserted and alignedto it; i.e., the center of curvature of the sphericalmirror is positioned at the focus of the interferometer,and thus the focus of the OAP. The optical fiber isthen inserted in place of the interferometer andaligned to the transfer mirror using a five-axis posi-tioning stage; fiber alignment is optimized via thepower reflected back into the fiber, as monitored viaport 3 of the circulator. When the power is maxi-mized, the transfer mirror is removed and the fiberalignment to the OAP is complete. Note from Table 2that the wavefront error due to the initial alignmentof the OAP to the interferometer is 0.021� rms at633 nm, corresponding to 0.0086� rms at 1550 nm.The final alignment error of the fiber to the OAPis calculated to be 0.010� rms. This difference of0.0014� is within the resolution limit of the wave-front sensor used to characterize the system. Thecontribution of the fiber alignment to the wavefronterror is negligible; within the limits of measurement,the interferometer alignment is transferred to thefiber.

The alignment budget in Table 2 shows how eachalignment contributes to the final wavefront error.Each alignment step was modeled using the Zemax

software package to determine the amount of mis-alignment required to achieve either a specific wave-front error or a reflected power measurement. Thefinal alignment was bounded based on the stack-up ofthe individual misalignments. A large beam splitterin the testbed optics (described in Section 3) is used asa return mirror for the interferometer alignment andis included in the wavefront error budget for Step 1 inTable 2; in subsequent steps, the wavefront errorfrom this beam splitter was included in the budgetfor the testbed optics and does not contribute to thewavefront error of the calibration source.

To predict the wavefront error of the final calibrationbeam, several tolerance analyses were also performedusing Zemax. Two systems were analyzed interfero-metrically: the OAP alignment and the transfer mirroralignment. The OAP alignment was modeled using apoint source with an F�5 cone angle �5.7°�. Perturba-tions were made in the x and z directions as indicatedin Fig. 4. Similarly, the transfer mirror alignment tothe fiber source was analyzed using the fiber-couplingfeature in Zemax. This analysis provided a predictionof the returned power as a function of alignment andindicated that the critical alignment was the trans-verse direction �x�. The theoretical predictions wereconfirmed experimentally using a test laser, a spher-ical mirror, and a circulator. Figure 5 shows the re-

Table 2. Calibration Source Alignment Budget

Alignment 633 nm Mini-Fizeau Interferometer

Typical Wavefront Error (SinglePass) Optics and Alignment

(� rms)

Wavefront Error due to Alignment(Root-Sum-Square Out Optics)

(� rms)

Misalignment (ZemaxTolerancing)

x (�m) y (�m) z (�m)

� � 633 nmCalibration optics: Step 1 0.032 0.021 8.4 8.4 8.4Reference sphere: Step 2 0.012 0.011 �0.1 �0.1 �0.1

Alignment withCirculator: Step 3

Power Returned(Submicrometer Resolution Ensures �90%)

Micrometer Resolution(�m)

Fiber feedback 91.8% 0.5 0.5 0.5Stack-up (rounding up) 10 10 10

Fiber Cal Beam: Step 4

Single-Pass Wavefront Errordue to Optics (Excludes Return

Flat)

Wavefront Error due to Alignment(Zemax Prediction Based on

Stack-Up)

Total Wavefront ErrorPrediction (Root-Sum-

Square Value)

� � 1550 nm 0.0082� rms 0.010� rms 0.013� rms

Fig. 4. Configuration for angular sensitivity analysis measurement.

672 APPLIED OPTICS � Vol. 46, No. 5 � 10 February 2007

Page 5: Stable, tunable calibration source for large-beam optical systems

sults, which are in excellent agreement with theprediction. Stable coupling was achieved using com-mercial translation stages based on standard differ-ential micrometers with 0.07 �m resolution.

3. Results

A. Wavefront Error

To determine the wavefront quality, the calibrationbeam was imaged through the optical testbed to aShack–Hartman wavefront sensor, and wavefrontdata were generated using vendor-supplied referencefiles. The testbed consists of several high-quality op-tics, including a large turning mirror, a telescope thatreduces the beam size, a lens relay, and several beamsplitters. The testbed optics were all staked down withwrenches, pins, and�or epoxy to minimize alignmentdrift. In contrast, the calibration source is intended tobe optimized on demand, and thus its optics are notpermanently fixed. Based on test results for the indi-vidual testbed components, the wavefront error dueto the testbed was estimated to be 0.0298� rms at1550 nm. Combined with the 0.013� rms predictedwavefront error for the calibration source (see Table2), the predicted performance for the combination ofthe calibration source and the testbed optics was0.033� rms. Upon initial alignment and optimiza-tion, performance of the combined system is often asgood as 0.030� rms, which is consistent with the pre-diction of 0.013� rms for the calibration source. Thiscorresponds to 0.065� peak to valley or ���15.

The wavefront error of the combined calibrationsource and optical testbed was measured at 1545 nmover a 66 h period. Figure 6 shows the results, whichexclude tilt but include all other aberrations. Thedata exhibits a clear temperature dependence of ap-proximately �0.15��°C; the facility is temperaturecontrolled via an air handling system that operateswith a 25–30 min cycle, resulting in the periodicitythat is apparent in the data. The average value of the

wavefront error begins at �0.04� rms and graduallyincreases to �0.05� rms over a period of 30 h as theoptical mounts settle. Beyond 30 h, the average valueremains constant at �0.05� rms. Both the long-termdrift and the periodic variation are believed to bedominated by the mounts for the free-space optics inthe calibration source.

B. Power Spectra and Dynamic Range

Power spectra for the calibration source are shown inFig. 7 for three wavelengths that span the wavelengthrange. The three rows show power spectra at 20 mW,200 mW, and 2.0 W, as confirmed by a National In-stitute of Standards and Technology traceable, free-space powermeter. The spectra were collected at theoptical spectrum analyzer (OSA) tap point shown inFig. 1. The length of fiber carrying the high-powersignal beyond the tap point is less than 3 m, andnonlinearity-induced spectral distortion is negligible;thus the tap provides an accurate representation ofthe power spectrum launched into free space. As ex-pected for a fiber-amplifier-based system, the perfor-mance at the middle wavelengths is excellent; at1557 nm, the optical signal-to-noise ratio (OSNR) ex-ceeds 44 dB in a 0.1 nm bandwidth at 2.0 W. Evenat the edge wavelengths, where amplifier gain tiltcauses preferential amplification of the suppressedmodes of the master oscillator, the OSNR exceeds 31dB. The optical power dynamic range is at least 20 dBat all wavelengths; maximum output power is limitedto slightly more than 2 W by the rated damagethreshold of the circulator.

C. Power Stability

Data from a long-term power stability measure-ment is shown in Fig. 8. After an initial warm-upperiod of �30 min, the amplifier was operated ina constant current mode to produce �1.45 W oflaunched power at 1553 nm. As shown in Fig. 8, thepower was stable to within 0.03 dB, or ��0.4%, over24 h. Also shown in Fig. 8 are data from a tempera-ture sensor located near the calibration source in thelaboratory as well as a temperature sensor internal to

Fig. 5. Theoretical and measured fiber–retrosphere alignmentsensitivity.

Fig. 6. rms wavefront error at 1545 nm and facility tempera-ture.

10 February 2007 � Vol. 46, No. 5 � APPLIED OPTICS 673

Page 6: Stable, tunable calibration source for large-beam optical systems

the optical amplifier. The periodicity of the residualvariation in the calibration source output power re-veals that the limiting factor for power stability isagain the air-handling system of the facility: varia-tion of the ambient temperature in the laboratorycauses temperature fluctuations with the same peri-odicity inside the amplifier, and because the amplifierefficiency varies with temperature, the output powervaries for a constant applied current.

D. Polarization Stability

Figure 9 shows the polarization in terms of orienta-tion and ellipticity over the same 24 h period as Fig.8. An arbitrary polarization state was selected at the

beginning of the test after the warm-up period byapplying distinct voltages to each of the polarization-state generator’s four actuators. The observed varia-tion is ��1°�h in orientation and ��0.02�h inellipticity, limited by drift in the first few hours of themeasurement. Because this initial drift is monotonic,we believe it to be caused by mechanical relaxation ofthe short length of single-mode fiber following thepolarization state generator, which extends �70 cmabove the height of the optical table to the fiberlaunch positioner. The residual temperature varia-tion in the facility has a minimal effect on polariza-tion stability, manifesting itself as a slight periodicripple in the data.

Fig. 7. Calibration source power spectra. PSD, power spectral density.

Fig. 8. Long-term power stability. Fig. 9. Long-term polarization stability.

674 APPLIED OPTICS � Vol. 46, No. 5 � 10 February 2007

Page 7: Stable, tunable calibration source for large-beam optical systems

4. Summary

A 370 mm optical calibration source that is tunablefrom 1545 to 1570 nm and is stable in wavefront,power, and polarization has been demonstrated. Bylimiting the number of free-space optical componentsand utilizing the near-Gaussian mode of a single-mode optical fiber as the source, a wavefront betterthan 0.013� rms has been achieved. The fiber-opticcomponents allow the remoting of heat sources suchas the optical amplifier to minimize thermal pertur-bations in the optical testing environment. Wave-front, power, and polarization stability are more thanadequate to make accurate measurements on timescales from minutes to hours. Long-term wavefrontand power stability are limited by residual thermalvariation, while long-term polarization stability islimited by residual motion in the short length ofsingle-mode fiber. Stability could be further en-hanced using active thermal control or by using apower-stabilization control loop and an active polar-ization control loop via a tap in the free-space path.Likewise, the OSNR could be further enhanced byincorporating a tunable filter between the master os-cillator and the fiber amplifier. The current imple-mentation has only been exercised to �2 W becauseof the power rating of the micro-optic circulator; how-ever, similar components have been demonstrated tohandle �10 W with �1 dB of insertion loss, and thusthe output of the calibration source should readilyscale to �8 W.

The authors are grateful to several of their col-leagues at MIT Lincoln Laboratory: Alan DeCew,Phil Chapnik, Kurt Abdelmaseh, Evan Stryjewski,Sarah Klein, and Tim Williams for optical support;John Peters, Bill Wilcox, and Cliff Williams for soft-ware support; and Colm Cryan and Peter Schulz fordirection and useful comments. This work was spon-sored by the Department of the Air Force under AirForce contract FA8721-05-C-0002. Opinions, inter-pretations, conclusions, and recommendations arethose of the authors and are not necessarily endorsedby the United States Government.

References1. M. C. Larson, Y. A. Akulova, C. Coldren, T. Lijeberg, G. A. Fish,

S. Nakagawa, A. Dahl, P. Kozodoy, D. Bingo, M. Bai, N. Ram-das, S. Penniman, T. Wipiejewski, and L. A. Coldren, “Highperformance widely-tunable SG-DBR lasers,” in Novel In-PlaneSemiconductor Lasers II, C. F. Gmachl and D. P. Bour, eds.,Proc. SPIE 4995, 66–80 (2003).

2. S. Thaniyavarn, “Wavelength independent, optical damageimmune Z-propagation LiNbO3 waveguide polarization con-verter,” Appl. Physics Lett. 47, 674–676 (1985).

3. F. Heismann, “Integrated-optic polarization transformer forreset-free endless polarization control,” IEEE J. QuantumElectron. 25, 1989–1906 (1989).

4. T. H. Woods, “Highly reliable high power optical amplifiersat 1.55 �m,” in Optoelectronic Integrated Circuits VIII, L. A.Eldada and E. Lee, eds., Proc. SPIE 6124, 61240Q1–9 (2006).

5. R. Noé, H. Heidrich, and D. Hoffmann, “Endless polarizationcontrol systems for coherent optics,” J. Lightwave Technol. 6,1199–1207 (1988).

10 February 2007 � Vol. 46, No. 5 � APPLIED OPTICS 675


Recommended