Status and activity on LIF-technique development in NFI. I.Moskalenko , N.Molodtsov, D.Shcheglov. 1. Divertorplasm a diagnostic by LIF-m ethod in program ofITER. 2. The field ofLIF-technique activities. 3. He Ispectroscopic schem e forLIF m easurem entsin ITER divertor. 4. Alternative spectroscopic schem eon helium atom measurem ents. 5. Application ofLIF-technique to extrinsic im purity diagnostics.LIF- m easurem entson A rIIions. 6. A rIIm easurem entson PN X -U . 7. Com parison of ITER M easurem entRequirem ents and perform ance ofLIF-technique. 8. Schem atic diagram ofLaserM odule forLIF-Diagnosnic. 9. Spectroscopic schem esforLIF on N e I. 10. LIF-System basic conceptualdesign. 11. Conclusions.
Transcript
Page 1: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

Status and activity on LIF-technique development in NFI.

I.Moskalenko, N.Molodtsov, D.Shcheglov.

1. Divertor plasma diagnostic by LIF-method in program of ITER. 2. The field of LIF-technique activities. 3. He I spectroscopic scheme for LIF measurements in ITER divertor. 4. Alternative spectroscopic scheme on helium atom measurements. 5. Application of LIF-technique to extrinsic impurity diagnostics. LIF-

measurements on Ar II ions. 6. Ar II measurements on PNX-U. 7. Comparison of ITER Measurement Requirements and performance

of LIF-technique. 8. Schematic diagram of Laser Module for LIF-Diagnosnic. 9. Spectroscopic schemes for LIF on Ne I. 10. LIF-System basic conceptual design. 11. Conclusions.

Page 2: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

Divertor plasma diagnostics by Laser-Induced Fluorescence method in program of ITER

     LIF-technique is based on detection of fluorescence radiation of atom or ion excited by laser beam. The diagnostic gives opportunity to perform measurements with good spectral, temporal and spatial resolution.     Helium is “ash” of thermonuclear reactions, the measurement of helium density and temperature in divertor plasmas is a problem of importance.      Another inert gases (Ne, Ar, Kr) are also is targets for LIF-technique application because of idea to inject these extrinsic impurities in order to run experiments using RI-discharges.       A priory estimation of signal and signal-to-noise ratios were carried out by making use of real laser parameters; this laser considered to be a prototype of laser transmitter for measurements on ITER.   Parameters of laser system give a possibility to perform Doppler measurements of temperature by laser spectroscopy method (by scanning narrowband laser line over absorption spectral line profile). A large series of such measurements have been carried out on plasma machine in Kurchatov Institute (T-10, HELLA, PNX-U).   Integral part of activity on development of LIF-technique is work on interpretative (CRM). The activity includes development of methods to measure parameters of electron component with help of LIF-method.

Page 3: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

The field of LIF-technique activities

HeliumTemperature

T (He I)

Task of priority

Helium densityin divertor plasmas

CRM-modernizing(Non-stop activity)

-new databases for HeI and etc.

Electron density

ne(R,Z)=fI(1)/I(2)

New scheme

Extrinsic impuritiesCRM

n(Ar II)

Doppler measurements of f(vi) Ar II

Modelling experiments on PNX-U

Targets for Ti(R,Z)measurements

Supplementary functions

Supplementary functions

Additional opportunities

Supplementary function

Page 4: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

He I spectroscopic scheme for LIF measurements in ITER divertor

This scheme permit to avoid stray light problem. The application of laser spectroscopy gives possibility to measure three plasma parameters simultaneously. Local measurement of Doppler temperature by Laser Spectroscopy Method is based on scanning of pumped spectroscopic transition with radiation of narrow band tunable laser and detection of fluorescence light. After making the correction on laser line width, Zeeman splitting and etc. the atom temperature can be calculated using Doppler width.

Helium temperature at 587.6 nm has been also measured on T-10 tokamak. Temperature was (4-5) eV. Measurements of helium concentrationThe sum of fluorescence signals gives input data for He I density estimates via CRM.Estimates of local electron density by LIF-techniqueThe ratio of fluorescence signals at 388.9-nm and 706.5-nm lines depends on electron density. The modeling experiments have been performed on PNX-U and results demonstrated good agreement with electric probes measurements.

3 3P

706.5

3 3S 3 3D

2 3P

2 3S

388.9

587.6

Page 5: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

Alternative spectroscopic scheme of helium

atom measurements

706.5

3 3S3 3P

3 3D

2 3P

2 3S

388.9

587.6

Ne<pse>Ne<pde>

This scheme also provides avoiding the stray light problem. Detuninig FLU - L is large for both

fluorescence lines. From point of view of application to divertor plasma diagnostic, this scheme is preferable in case when transmission of optics in the near UV became low.

Page 6: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

4p 2Fo

7/2

461.0 611.5

4p 2Fo7/2

4s 2D5/2

Fluorescence Laser

0

1

2

3

4

5

6

7

-14 -10 -6 -2 2 6 10 14

Inte

nsity,

arb

. u

nits

.

- pm

Ar II ion. Spectral profile of 611.5-nm line.

The next step was to apply LIF technique to diagnosis of “extrinsic” impurities. Doppler profiles of Ar II have been measured on plasma neutralizer PNX-U. Narrow band (L 3.6 pm) dye laser was scanning

across 611.5 nm absorption line. Ti range was (4-25 eV).

It correspond for helium temperatures (588 nm absorption line) to range values of 0.5 – 2.7 eV because width of Doppler- broadened intensity distribution D 0(Ti/Mi)

1/2 , M(Ar) = 40 and M(He) = 4.

Helium Doppler temperature was measured to be (1.5-2) eV in PNX-U plasma.

Page 7: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

Ar II measurements on PNX-U

Page 8: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

Comparison of ITER Measurement Requirements and performance of LIF-technique.

 

 

  Parameter

Parameter range

SpatialResolution

Accuracy

ITER(Level 1)

Ti,a 0.3 – 200 eV5 сm "along legs "

3 mm "across legs "20%

PNX--U Ti,(Ar1+) 4 – 25 eV(experiments)

Δl|| = 4 cm

Δl┴ = 3 mm

10 - 20% (depending

on discharge)

Page 9: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

Schematic diagram of Laser Module for LIF-Diagnostic

Pump laser

Wavelength, nm 308

Repetition rate, Hz 100

Output energy, mJ 200 – 240

Pulse duration, ns 28 – 32

Dye laser

Output energy, mJ 15 – 20

Pulse duration, ns 18 – 20

Tuning range, nm 340 –800

Bandwidth, pm 3.5

Output beam size, mm 2 x 2

Page 10: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

“3-level schemes” for Ne I LIF-technique probing2p2

1s2

1s3

1s5

588.

2

659.

8

616.

4

603.

0

1s3 , 1s5 – metastable

levels (Paschen’s notation)

Possible scheme for electron density measurements on Ne I

1sj

LCIF

2pi

e-

λL

1sm

1sl

LIF

2pk

)()(

)(e

FLU

FLU nfLIFI

LCIFI

Spectroscopic schemes for LIF on Ne I

Page 11: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

LIF system in zone A “right divertor leg” is shown. The probing laser beam is transmitted trough labyrinth mirrors and windows in closure plates and then directed with the help of large plane mirror M1 to cylindrical laser mirror ML. Changes of angle beam direction can be provided observe all zone A of investigation. Fluorescence light from part of laser beam equal ~ 40 mm and plasma light in zone A is collected by scanning mirror MS, transmitted by large mirror M1 to Cassegrain telescope and is directed to spectrometer. Using this configuration, it is possible to achieve the spatial resolution, equal to ~ 40 mm along “leg” and ~30 mm across “leg”.

Page 12: Status and activity on LIF-technique development in NFI. I.Moskalenko, N.Molodtsov, D.Shcheglov.

Conclusions

• Making use of LIF-technique in experiments on plasma devices has demonstrated the ability of the system to measure density, temperature (atoms, ions) and electron density in real physical experiments.

• The laser diagnostic system is occurred to be reliable from technical point of view and can be considered as to be the prototype of LIF transmitter for ITER.

• The work on development of Collisional Radiative Models is integral part of LIF program activity

• It is necessary to develop diagnostics by LIF-technique based on new “targets” (trace elements) for measurements.