slanted viewscreen, you’ll appreciate how easy the SterilGARD ® III Advance° is to use. Your work is easy to reach. There’s more functional room inside. Eye-level controls are easy to see. There is less glare. As you work, your head and neck, arms and elbows, remain in the most natural position, less tiring than before. Along the streamlined viewscreen, concealed behind the attractive front panels, you’ll find another Baker exclusive—UniPressure TM Preflow, an elegantly simple yet effective air management system that puts more uniform air pressure on the supply filter for better vertical flow across the work surface. A simple idea, to be sure, but by moving the blower/motor to a horizontal orientation from vertical, we’ve created a more efficient, less demanding and quieter airflow system. And years later, if it comes time for maintenance, your certifier will find the cabinet easy to seal and easy to decon- taminate. Our new direct-to-plenum filter mounting technique puts the HEPA filter on a telescoping frame, within easy reach, easy to change. SterilGARD ® III Advance° is the most exciting development in biological safety cabinets in years. Consider the angle. Learn more about SterilGARD ® III Advance°. It’s what you would expect from The Baker Company. *Properly vented to the outdoors through a facility exhaust system, SterilGARD ® III Advance° meets or exceeds NSF 49 minimum standards for a Class II, Type A2 (formerly B3) cabinet. CSA Listed Meets NSF Standards Underwriter Laboratories Listed UniPressure Preflow Plenum SterilGARD ® III Advance° Class II, Type A2* Biological Safety Cabinet, Vertical Flow SterilGARD ® III Advance° Sets the Industry’s Highest Standard for Comfort, Performance and Safety. In the laboratory environment, safety is everything. Now, with the intro- duction of SterilGARD ® III Advance°, The Baker Company offers a new adap- tive ergonomic design, combined with a unique airflow management system and proven containment technology, to improve comfort, increase productivity and reduce the cost of ownership. SterilGARD ® III Advance° is a collective design, representing an esteemed reputation for performance, reliability and craftsmanship associated with The Baker Company worldwide for nearly 50 years. SterilGARD ® III Advance° starts with a simple idea: a better angle. From your first experience with the 10° Exclusive 10° slanted viewscreen– less glare, work easier to reach.
Transcript
Page 1: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

slanted viewscreen, you’ll appreciatehow easy the SterilGARD®IIIAdvance° is to use. Your work is easy toreach. There’s more functional roominside. Eye-level controls are easy tosee. There is less glare. As you work,your head and neck, arms and elbows,remain in the most natural position, lesstiring than before.

Along the streamlined viewscreen,concealed behind the attractive frontpanels, you’ll find another Baker exclusive—UniPressureTM Preflow, an elegantly simple yet effective air management system that puts more uniform air pressure on the supply filterfor better vertical flow across the worksurface. A simple idea, to be sure, but by moving the blower/motor to a horizontal orientation from vertical,we’ve created a more efficient, lessdemanding and quieter airflow system.

And years later, if it comes time formaintenance, your certifier will find thecabinet easy to seal and easy to decon-taminate. Our new direct-to-plenum filter mounting technique puts the HEPAfilter on a telescoping frame, within easyreach, easy to change.

SterilGARD®III Advance° is themost exciting development in biologicalsafety cabinets in years. Consider theangle. Learn more about SterilGARD®IIIAdvance°. It’s what you would expectfrom The Baker Company.

*Properly vented to the outdoors through a

facility exhaust system, SterilGARD® III Advance°

meets or exceeds NSF 49 minimum standards for

a Class II, Type A2 (formerly B3) cabinet.

CSAListed

MeetsNSF

Standards

UnderwriterLaboratories

Listed

UniPressure™

PreflowPlenum

SterilGARD®III Advance°Class II, Type A2* BiologicalSafety Cabinet, Vertical Flow

SterilGARD®III Advance° Sets the Industry’s HighestStandard for Comfort, Performance and Safety.

In the laboratory environment, safety is everything. Now, with the intro-duction of SterilGARD®III Advance°,The Baker Company offers a new adap-tive ergonomic design, combined with aunique airflow management system andproven containment technology, to improve comfort, increase productivityand reduce the cost of ownership.

SterilGARD®III Advance° is a collective design, representing anesteemed reputation for performance,reliability and craftsmanship associatedwith The Baker Company worldwide fornearly 50 years.

SterilGARD® III Advance° startswith a simple idea: a better angle. From your first experience with the 10°

Exclusive 10° slanted

viewscreen–less glare,

work easier to reach.

Page 2: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

Design and Performance FeaturesSterilGARD®III Advance° incorporatesUniPressureTM Preflow, the Bakerexclusive high performance airflow system which provides optimum protection from particulates, togetherwith extended filter life and reducedcost of ownership.• The unique UniPressureTM

Preflow Plenum designapportions and distributes airacross, then through theHEPA supply filter, improvingdownflow uniformity, extend-ing useful filter life and reducing noise.• The UniPressureTM Preflow filter housing creates negative pressure sur-rounding the positive pressure plenumto improve containment; any possiblegasket leaks are contained under negative pressure.• An innovative telescoping filter mountprovides a direct seal of the filter to the UniPressureTM Preflow Plenum, andsimplifies filter replacement.• Supply and exhaust HEPA filters arerated at 99.99% minimum efficiency incapturing 0.3 micrometer particulates.• The Baker exclusive momentum aircurtain contains particulates within thework area and inhibits migration ofroom air into the work area.• High velocity return air slots locatedalong the side walls of the work areaand along the top of the viewscreenmaximize the biological safety cabinetprotective capabilities by preventing theescape of particulates, ensuring that nounfiltered air enters the work area, andpreventing gases, vapors or particulatesfrom migrating up behind the view-screen and escaping into the operator’senvironment.• Aerodynamically designed airfoil atthe front opening facilitates the flow of room air into the front grille, ensur-ing that no room air flows into the work area.

• Innovative electronic controller includesLED displays and diagnostic functions.Optional PLC control system includestouchscreen display, digital pressuremonitor, RS232 data port; RS485 optionallows monitoring of up to 32 cabinets.• An audible/visual sash alarm systemwarns of unsafe viewscreen opening via

proximity switches.• A large, easy-to-read magnehelicgauge confirms cabinet operation.• Detachable side panels offer greaterflexibility for plumbing considerations;the reinforced panel design reduces thenumber of fasteners and simplifiesremoval and replacement.

• The enlarged viewing area, 92 sq.in. (403a), 140 sq.in. (603a), isslanted 10° to create a more natural head and elbow positionwhen working, minimizing arm extension required to reachthe work surface, and reducing glare.• New proximity switches replace mechanical limit switches forviewscreen height alarm and UV light interlock.• The eye-level control panel is positioned to face down towardthe user for greater visibility and easier access.• Reduced front grille depth and a slim profile lower plenumcombine to move the work surface closer to the front and yourlap for better arm position.• Petcocks are offset, not one on top of the other, for easieraccess and use.• A padded armrest (optional) across the cabinet frontimproves support and comfort.• An optional stand with telescoping legs allows the work surface height to be adjusted from 301⁄8 " to 385⁄8".• SterilGARD® III Advance° offers the largest unobstructed, useable workarea in the industry. Inside cabinet has been extended 2" deeper, creating 110 sq.in. (403a), 170 sq.in. (603a), additional work area.• Depth of the front grille is minimized to bring the work closerto the user. Rear grille is parallel to the back wall to increaseuseable work space.• A straight back wall maximizes available work area andaccommodates lab equipment and apparatus commonly used inthe biological safety cabinet.• A unitized, unobstructed drain pan with radius corners facilitates cleaning. New one piece work surface/air intakegrille with pull tabs replaces earlier two-piece design.• Streamlined design of the sliding viewscreen simplifies sealing prior to total cabinet decontamination.• The telescoping UniPressureTM Preflow Plenum simplifies filter changing, delivers quieter, more efficient performance.• All electrical control components are consolidated on a panel behind the hinged canopy for singular access outsidethe containment area.• The innovative horizontal blower/motor assembly is easilyremoved and replaced.

The SterilGARD®III Advance° design is based on extensive anthropometric data, and wasdeveloped by The Baker Company in cooperation with ergonomics researchers at a lead-ing university.

SterilGARD® III Advance°

Advantages Ergonomic Features and Benefits

Comfortable and Easy to Work In

More Space for Your Work, Your Equipmentand You

Easy to Clean and Maintain

P a g e 2 8 0 0 . 9 9 2 . 2 5 3 7 w w w . b a k e r c o . c o m

Page 3: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

S t e r i l G A R D ® I I I A d v a n c e °

P a g e 3

The enlarged viewing area, 92 sq.in.(403a), 140 sq.in. (603a), is slanted 10°to create a more natural head andelbow position when working, minimiz-ing arm extension required to reachthe work surface.

Low profile, unitized drain panbeneath the work surface allows bestarm position above the user’s lap. Newone piece work surface/air intakegrille with pull tabs replaces earliertwo-piece design.

Eye-level controls face downtoward the user, for greater visibilityand easier access.

Innovative electronic controllerincludes LED displays and diagnosticfunctions. Optional PLC control systemincludes touchscreen display, digitalpressure monitor, RS232 data port;RS485 option allows monitoring of upto 32 cabinets.

Optional stand with telescopinglegs allow the work surface to beadjusted from 301⁄8" to 38 5⁄8".

Audible/visual sash alarm withtimer and mute switch encouragessafe operation.

A large, easy-to-read magnehelicgauge confirms cabinet operation.

High velocity return air slots maximize protective capabilities.

HEPA supply and exhaust filters,99.99% minimum efficiency for 0.3micrometer particulates.

StediVOLT® voltage regulator automatically adjusts for low and highvoltage to maintain proper airflow.

Energy-saving fluorescent lightingwith solid-state ballasts reduces flicker,minimizes heat output, improves eyecomfort, and extends lamp life by asmuch as 50%.

1

2

Non-glare work surface improvescomfort.

Independent ground-fault interrupt(GFI) outlets with self-resetting circuitbreaker for user safety; trip lightwarns circuit has been disconnectedbecause of ground fault error.

Straight back wall with integratedrear grille creates the largest unobstructedwork surface. Inside cabinet is extended2" deeper, creating 110 sq.in. (403a), 170sq.in. (603a) additional work space.

New proximity switches replacemechanical limit switches for view-screen height alarm, and UV light interlock.

Innovative horizontal blower/motorwith multi-voltage speed controllermaintains air velocity, provides moreuniform airflow to the filter.

New internal balancing damperreduces exhaust velocity by more than 75percent, minimizing room air turbulence.

Versatile canopy exhaust connectionprotects cabinet performance.

Unipressure™ Preflow Plenum deliversquieter, more efficient performance,simplifies filter changing.

Streamlined design simplifies sealingfor pressure leak testing or cabinet deconta-mination.

4

5

6

7

8

9

10

Momentum Air Curtain

3

4

5

6

7

8

9

10

12

12

11

11

13

13

14

14

3

1

2

SterilGARD®III Advance° Design Features • The front panels are lighter andeasier to remove for filter changing.

Class II, Type A2 ApplicationsThe SterilGARD®III Advance°biological safety cabinet isdesigned for sterile product prepa-ration and biological experimenta-tion involving agents of low andmoderate risk.• When exhausted to the room,SterilGARD®III Advance° is classi-fied a Class II, Type A2 cabinet.• Properly vented to the outdoorsthrough a facility exhaust system,SterilGARD®III Advance° exceedsminimum standards for a Class II,Type A2 cabinet (formerly B3).

Momentum AirCurtain IncreasesProtection WithoutRestricting AccessSterilGARD®III Advance° employs aunique momentum air curtain whichoffers an added measure of contain-ment and protection exclusive to theBaker design.• This unique airflow design creates amore impenetrable air barrier ormomentum air curtain at the front ofthe cabinet, increasing the cabinet’sprotective capabilities.• The strategic position of thestainless steel diffuser just belowthe supply filter creates a faster air-flow of a nominal 100 FPM at thefront of the work zone.

• Airflow over the center of the worksurface is relatively slower at a nominal55 FPM.• The resulting air curtain, high velocityreturn air slots, aerodynamically

contoured front opening surfaces, and optimum air intake velocity, combine to minimize turbulence and prevent migration of airborne contaminants intoor out of the work area.

15

16

17

18

19

20

15

16

1718

19

20

100 FPM

55 FPM

SterilGARD®III Advance°

Page 4: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

Balanced Airflow and Exhaust Assure UniformityThe SterilGARD®III Advance°features superb airflow characteristicswhich deliver unidirectional, downflowair over the work area for maximumcontainment and protection.• Filtered air descends from top to bottom of the work area in a unidirec-tional flow. Near the center of the work surface, the air current divides; a portion of the downflow air is pulledthrough the back wall grille and theremainder is pulled through the frontperforated grille.

• Simultaneously, room air is pulledthrough the front access opening and into the front perforated grille. It does not enter the work area. The calculated average intake air setpointvelocity at the front access opening is 105 FPM.

• All air is combined under the worksurface. From this point, contaminatedair is pulled under negative pressurethrough a plenum up the rear and sidesof the cabinet into the blower/motor,which blows it into a positive pressureplenum.• From the positive pressure plenum,approximately 30% of the air exits the system through the exhaust filter.The remaining 70% passes through thesupply filter and re-enters the work areaas particulate-free air.

High Velocity Return Air Slots CaptureUnfiltered Air Containment and cleanliness areaccomplished with precise control of airflow volumes and velocities. Aunique feature in the Baker cabinetdesign, the high velocity return airslots, have been proven to maximizethe biological safety cabinet protectivecapabilities.

• High velocity return air slots are located along the side walls of the work area and along the top of theviewscreen.• Air is drawn into these slots at approximately 1,000 FPM, preventingthe escape of particulates and ensuringthat no unfiltered air enters the workarea.• High velocity return air slots locatedalong the top of the sliding viewscreenprevent gases, vapors or particulatesfrom migrating up behind theviewscreen and escaping into the operator’s environment, and preventroom air from migrating down behindthe viewscreen and contaminating thework area.

SterilGARD®III Advance° Airflow

When installed as a Type A2 cabinet with direct exhaust to the facility exhaust system bya canopy or hard connection, the SterilGARD®III Advance° provides particulate protection,plus additional protection of personnel from minute amounts of vapors and gases. Ifexhausted to a facility exhaust system specially treated with activated charcoal, scrub-bing or catalytic conversion, protection from exhausted vapors and gases is extended tothe environment.

ProtectionFromParticulates

ProtectionFrom Gasesand Vapors

Exhausted to untreated facilityexhaust systemProtects personnel,product and the environment

Protects personnel

Exhausted to treated facilityexhaust systemProtects personnel,product and the environment

Protects personnel and theenvironment

Protects personnel, product and theenvironment

SterilGARD®III Advance° Containment and Protection

Exhaust to room

P a g e 4 8 0 0 . 9 9 2 . 2 5 3 7 w w w . b a k e r c o . c o m

Type A2 Type A2 Type A2

Page 5: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

S t e r i l G A R D ® I I I A d v a n c e °

• The slanted viewscreen minimizesglare.• The vertical lift viewscreen is constructed of 1⁄4 " safety plate glass.• The sash has an 8" (or optional 10")working opening with an average calculated air intake velocity of 100-110FPM.• A maximum viewscreen sash openingof 193⁄4" simplifies equipment and instrument loading and unloading.• An integrated viewscreen sash alarmincludes an audible and visual indicatorto warn of improper sash opening abovethe 8" (or optional 10") opening. Thealarm is disabled when the sash is fullyclosed.• The viewscreen sash alarm timer and audible alarm mute button overridethe alarm for 5 minutes when the sash is raised for cleaning, loading orunloading.

Microprocessor ControlsThe SterilGARD®III Advance° micro-processor control system incorporates aunique membrane control panel which iseasily cleaned, moisture-proof and excep-tionally reliable with no relays or switchesto corrode or fail. A variety of importantsafety features have been designed intothe control system.• Innovative electronic controller includesLED displays and diagnostic functions.Optional PLC control system includestouchscreen display, digital pressuremonitor, RS232 data port; RS485 optionallows monitoring of up tp 32 cabinets.• The control panel is positioned toface down towards the user for greatervisibility and easier access.• The sash alarm timer and audible alarmmute button override the alarm for 5minutes when the sash is raised for cleaning,loading or unloading.• The membrane panel is just .003 inchthick, with a smooth surface for cleanability.• Large buttons with icons are easy toidentify and operate.

Negative Pressure Double-Wall PlenumsEnhance SafetyThe all-metal, double wall design inSterilGARD®III Advance° creates base,side and rear wall plenums which cap-ture and contain contaminated air undernegative pressure. This unique Bakerdesign feature prevents contaminated airfrom escaping into the laboratory envi-ronment in the event of a puncture orleak in the cabinet.

Any plenum which contains bio-logically contaminated air under positivepressure is completely surrounded bynegative pressure areas.

High VisibilityViewscreen With SashAlarm SterilGARD®III Advance° has a stream-lined 10° slanted sliding viewscreenwhich makes the cabinet easier to useand more comfortable to work in.• The viewing area, enlarged by 92 sq.in.(403a), 140 sq.in. (603a), is slanted 10ºfor a more comfortable head and elbowposition when working; minimizes armextension required to reach the worksurface.• New one-piece work surface/air intakegrille with pull tabs replaces earlier two-piece design.• The operator’s head, neck, arms and elbows remain in the most naturalposition to improve comfort and reduce fatigue.• All work is within reach; the angledviewscreen design provides a safe, highly visible and easily accessible workarea for a range of procedures.

The microprocessor-based membrane control panel simplifies operation.

The SterilGARD®III Advance°

has a unique negative

pressure enclosure which

provides additional contain-

ment capability. The cabinet

work area and plenums

containing contaminated air

under positive pressure are

surrounded by negative

pressure areas.

P a g e 5

Page 6: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

UniPressureTM PreflowAdvanced Filter Plenum DesignThe SterilGARD®III Advance°incorporates UniPressureTM Preflow, a revolutionary air delivery system whichoffers significant benefits throughunique integration of supply and exhaustHEPA filter assemblies, a telescoping filter mount, an all-steel plenum withadjustable internal damperassembly, and a new horizontal blower/motor configuration.• Located between the HEPA filters, the supply air plenum is config-ured to provide more uniform airflow to the supply filter, while reducingnoise and increasing reserve blower/motor capacity.• The all-steel telescoping plenumclamps directly to the HEPA filters.Force is applied to the full perimeter ofthe filters rather than point force.• A closed-cell neoprene gasket forms an airtight seal around the filter periphery. • The horizontal blower assembly creates more uniform air direction to the supply HEPA filter, thereby reducingnoise, increasing motor reserve andextending filter life.• A new internal damper simplifies airflow balance and cabinet sealing fordecontamination.

Supply and Exhaust Filter PerformanceBecause filters remove microorganismsand airborne particulates (generallycalled aerosols) from the air, the quality, performance and useful life of supply and exhaust filters are criticalconsiderations in the biosafety environment.

• Each filter in each Baker cabinet isscan tested for leaks and tested for overall efficiency by the manufacturer,then individually scan tested by Bakerafter installation to assure leak-free performance.• Minimum filter efficiency is 99.99%for particulates 0.3 micrometers,increased efficiency for particulatesgreater and smaller than 0.3 micro-meters.• Both exhaust and supply HEPA filters can be inserted and removed from the front of the cabinet. This time-saving feature is useful if the cabinet isconnected to an exhaust duct.• A closed-cell neoprene gasket provides an airtight seal between the filter assembly and the metal plenum.

Blower/Motor System Helps ExtendFilter LifeBaker has selected an optimumblower/motor combination to assureperformance and to extend filter life toan average of 7 to 10 years, the longestlife in the industry. The Bakerblower/motor provides a constant

volume of air despite increases in resis-tance due to filter loading, resulting inextremely long filter life as verified bysimulated filter loading tests.

Extended filter life minimizes filterreplacement and decontamination costs,and diminishes the hazard to personnelby reducing the frequency of openingthe contaminated areas of the cabinet.The more filter resistance the blower/motor system can overcome, the longerbetween filter changes.• As particulates fill the filter media,resistance to airflow increases.• The Baker blower/motor is a directdrive permanent split capacitor motor,designed to deliver the maximumamount of filter life for each cabinet.• The blower/motor automatically compensates for filter loading up to60% above the initial pressure dropacross the filter. Beyond that point, the blower/motor has a manual speed control which can be turned up toachieve even longer filter life.• The filter need not be changed untilthe blower/motor system cannot deliveradequate air volume to maintain thenominal setpoint, ±5 FPM.

The Baker exclusive UniPressureTM Preflow Plenum provides quieter, more efficient operation.

P a g e 6 8 0 0 . 9 9 2 . 2 5 3 7 w w w . b a k e r c o . c o m

Page 7: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

Easier On-Site Certification and TestingSeveral SterilGARD®III Advance° designfeatures combine to simplify certificationand maintenance for qualified servicepersonnel, thus reducing downtime andimproving life cycle costs.• A telescoping plenum assembly putsthe supply and exhaust HEPA filterswithin easy reach, and allows the filters to be clamped directly to theplenum against the closed-cell neoprene gasket.• Both exhaust and supply filters can be easily inserted and removed from thefront of the cabinet. This time-saving feature is useful if the cabinet is connected to an exhaust duct or if theroom has a low ceiling.• An internal adjustable damper allowsthe certifier to compensate for changingresistance of supply and exhaust filters;the damper regulates the balancebetween intake and downflow so that theproper air circulation ratios are main-tained.• A manual speed controller allows thecertifier to increase blower/motor speedto compensate for filter loading.

StediVOLT® Motor Speed ControllerMaintains OptimumPerformanceThe Baker exclusive StediVOLT®motorspeed controller reduces the risk of performance degradation and possibleproduct loss due to line voltage fluctuations.• StediVOLT®minimizes changes incabinet performances due to routinevoltage fluctuations, as well as voltagechanges caused by high-demand,brownout conditions.

S t e r i l G A R D ® I I I A d v a n c e °

P a g e 7

Motor Loading Capacity

Model Pressure Maximum Increase With Drop Increase Manual Blower Speed Control

SG403a, 4-foot 60% 180%SG603a, 6-foot 60% 180%

SterilGARD®III Advance° is capable of automatically handling a 60% increase in pressuredrop across the filter without reducing total air delivery more than 10%. With the use of themanual speed controller, the SterilGARD®III Advance° can handle up to a 180% increase in pressure drop across the filter.

• StediVOLT® reads incoming voltage, automatically compensates forlow and high voltage, and maintains proper blower speed to prevent periodsof unbalanced airflow which may contribute to loss of containment orproduct protection.

Comfortable LightingImproves Visibility,Reduces HeatErgonomic benefits of theSterilGARD®III Advance° include newflicker-free fluorescent lamps and electronic ballasts which provide bettervisibility, less heat at the face (front) ofthe cabinet, and improved user comfort.• Lamps provide an average of 80 foot-candles of illumination at the work surface.• Electronic ballasts eliminate fluorescent lamp flicker, minimizing eye strain and improving productivity.• The warm, natural illuminationexhibits better color fidelity.

Optional Ultraviolet Lamp Includes SafetyInterlockAn optional ultraviolet germicidal lampassists in contamination control.• The UV light interlock assures thatUV illumination occurs only when thesash is fully closed.• For added safety, the UV germicidallight switch and cabinet lighting cannotbe turned on simultaneously.

Cabinet ConstructionEnhances Performance,Minimizes DowntimeBaker cabinet designs represent many years of experience in stainlesssteel fabrication and craftsmanship.Design considerations such as wideradius corners, aerodynamically shapedsurfaces and non-glare satin finish interiors combine to improve comfort,simplify cleaning and maintain proper containment.• SterilGARD®III Advance° includes a corrosion-resistant, 16-gauge, Type 304 stainless steel work surface with smooth radiused corners to permit easy cleaning. The recessed work area retains spills.• The satin finish work surface diminishes harsh light reflection.

StediVOLT®

Page 8: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

• A protective grille, located under thework surface at the bottom of the rearand side return air plenums, preventswipes and other paper materials frombeing inadvertently drawn into the blower system. This eliminates costlydecontamination and down time toextract materials from the blower.• Work area side walls and rear wall are one-piece, 16-gauge stainless steelconstruction with 7⁄16" radiused corners for cleaning convenience.• The position of the stainless steel airdiffuser/filter protector located belowthe supply filter in the work area createsthe momentum air curtain at the frontopening and provides filter protection.• The 16-gauge cold-rolled steel exterior cabinet is protected with a whitebaked enamel finish.• The entire cabinet is airtight. Eachcomponent is welded, gasketed orassembled with hermetically sealedjoints. Each cabinet is bubble testedunder pressure to ensure the integrity ofthese seals.• To facilitate cleaning the drain pan,the work surface and supports are easily removed and the drain pan haslarge 7⁄16" radii on all sides.• Outlets are provided with drip-proofcovers and are protected by one self-resetting circuit breaker.

Electrical System Offers Safety and ConvenienceThe SterilGARD®III Advance°electrical system is designed for safetyand convenience. GFI protected duplexoutlets, one on each work area side wall, are provided to accommodate most commonly used laboratory instrumentsand equipment.• Each outlet is provided with a drip-proof cover, and one self-resetting circuit breaker, located on left side wall.• The internal work area outlets are on a separate circuit from the cabinetlights and blower/motor so that an over-load caused by research equipment willnot affect cabinet function.• Hermetically sealed bulkhead connectors are used to provide reliablecontainment protection for electricalconnectors which penetrate the cabinetcontaminated areas.• For safety reasons, the SterilGARD®III Advance° cabinet utilizes a single power cord and plugarrangement. This assures that a secondpower source is not left on if mainten-ance functions are necessary.• The SterilGARD®III Advance° is ULand CSA Listed for electrical safety andintegrity.

Functional Utilities WithPetcocks, Valves andPlumbing ConnectionsPlumbing and drainage connectionsare strategically placed for both convenience and proper air managementwithin the airflow plenums.• One petcock and one plugged pene-tration are provided as standard on theright side wall. Left side wall is pre-punched for optional/additional plumbingconnections.• Petcocks and penetrations are offset,not one on top of the other, for easieraccess and use.• External plumbing connections aremade to the bottom (or optionally topor back) of the cabinet rather thanthrough its sides. This arrangementallows installation next to building wallsor furniture to the right or left of thecabinet, saving valuable lab space.• A stainless steel ball valve providessafe and effective drainage of the drain pan.

SterilGARD®III Advance° features an optional stand with telescoping legs to permit multiple work surface heights. This unique stand allows the cabinet to be adjusted tostandard work surface heights ranging from 30 1⁄8" to 38 5⁄8".

Knee Space Height Work Surface Height Overall Height

Minimum Setting 28 1⁄8" 301⁄8" 917⁄8"Maximum Setting 36 5⁄8" 38 5⁄8" 100 3⁄8"

Telescoping Stand (Optional)

P a g e 8 8 0 0 . 9 9 2 . 2 5 3 7 w w w . b a k e r c o . c o m

Page 9: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

Options and AccessoriesFor convenience, most options, acces-sories and modifications are factoryinstalled and should be specified whenordering. Commonly requested optionsare listed below. For detailed informationon accessories and modifications, contactThe Baker Company.• 10" viewscreen sash opening• UV germicidal lamp• Stainless steel IV bar• Additional petcock (specify label

and location)• Plumb to top• Plumb to rear• Plastic storage bins• Mass Airflow Monitor• Padded armrest• Ergonomic adjustable footrest• Type A2 exhaust kit, canopy

connection• Type A2 exhaust kit, hard

connection• Reinforced work surface• Seismic restraints• Optional stand with telescoping legs• Optional stand with casters• Remote controlled petcocks

To OrderFor ordering information, terms and conditions of sale, contact TheBaker Company or visit the Baker Web site at www.bakerco.com for thename of your authorized BakerCompany representative.

Model No. Size Opening Electrical Amps/ BTU Interior Exterior Net Ship MAX. Breaker Dimensions1 Dimensions1 Weight* Weight

SG403a 4' 8"(19 3⁄4") 115V, AC 12.4/20 2906 46"W x 201⁄8"F-B*** 535⁄8"W x 30 9⁄16"F-B 525 lbs 650 lbsx 271⁄2"H x 63 1⁄8"H

SG603a 6' 8"(19 3⁄4") 115V, AC 14.7/20 4358 70"W x 201⁄8"F-B*** 775⁄8"W x 309⁄16"F-B 700 lbs 890 lbsx 271⁄2"H x 63 1⁄8"H

*Net weight SG403a with the optional stand is 625 lbs., with a total shipping weight of 750 lbs.*Net weight SG603a with the optional stand is 810 lbs., with a total shipping weight of 1000 lbs.1Inside cabinet is extended 2" deeper, creating 110 sq.in. (403a), 170 sq.in. (603a) additional work area.

S t e r i l G A R D ® I I I A d v a n c e °

Caution: A Class II, Type A2 biological safety cabinet is suitable for work with agents in the absence of volatile toxic chemicals and volatile radionuclides. With proper ventilation to the outside, a Class II, Type A2 biological safetycabinet is suitable for work with agents assigned to biosafety levels 1, 2 or 3,treated with minute quantities of volatile toxic chemicals and trace amountsof radionuclides required as an adjunct to microbiological studies, that willnot interfere with the work when recirculated in the downflow air (as stated in National Sanitation Foundation International Standard #49).

Note: The adequacy of this containment cabinet for the user’s personal safety, as with any containment cabinet, should be determined by an indus-trial hygienist or safety officer. Site preparation information, architecturaldrawings, detailed dimensions and purchase specifications are available.

P a g e 9

Summary Specification

SterilGARD®III Advance° Class II, Type A2 Biological Safety Cabinet, Vertical Flow

The SterilGARD® IIIAdvance° is also availablein a 3' Model SG303.** Forliterature and information

contact Baker.

**SG303 is not NSF or UL listedSG303 is ETL listed.

***Useable work space dimensions.

SG303SG403aSG603a

Page 10: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

Optional Stand

SterilGARD®III Advance°Model Number: SG403a, 4'SG603a, 6'

NSF Classification: Class II, Type A2Cabinet Type: Bench (Stand Optional)

Site PreparationElectrical System• 115V-1 Phase, 60 Hz.• One 16' power cord with 20-amp plug,NEMA 5-20P.

• Unit is UL Listed as certified for electrical, fire and personal safety.• Two ground-fault interrupt protectedinterior duplex receptacles at 5.0 (SG403a),4.0 (SG603a) total amps, controlled bya self-resetting circuit breaker.

Exhaust Requirement*SG403a• For an 8" opening: 269 CFM at .02"to .04" water column suction directlyabove the exhaust filter before anyreductions, elbows or restrictions.

• For a 10" opening: 335 CFM at .02" to.04" water column suction directlyabove the exhaust filter before any reductions, elbows or restrictions.

SG603a• For an 8" opening: 408 CFM at .02"to .04" water column suction directlyabove the exhaust filter before anyreductions, elbows or restrictions.• For a 10" opening: 510 CFM at .02"to .04" water column suction directlyabove the exhaust filter before anyreductions, elbows or restrictions.

SG403a 1625 RPM, 6.6 amps 0.8 amps 5.0 amps 12.4 ampsSG603a 1625 RPM, 9.9 amps 0.8 amps 4.0 amps 14.7 amps

SG403a, SG603a

P a g e 1 0 8 0 0 . 9 9 2 . 2 5 3 7 w w w . b a k e r c o . c o m

Amperage ChartModel Blower Motor Lighting Ballast Duplex Receptacle Total

365/8" max.281/8" min. Stand Height(includes 21/2" of leg leveler adjustment).Add 2" for Work SurfaceHeight

*The above exhausting requirements are basic, should the user wish to build their own connections. HEC and CEC are asnoted in the chart on page 11.

Page 11: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

Exhaust OptionsThe SterilGARD® III Advance° is designedfor Type A2 installations. For added convenience, the cabinet design includesseveral innovative features which offerflexibility.

Both supply and exhaust HEPA filters can be inserted and removed from thefront of the cabinet, allowing filterreplacement without removing exhaustconnections.

Canopy Exhaust Connection (CEC)Laboratories which share commonexhaust systems typically experiencechanges in airflow which can compro-mise cabinet operation. Such problemsinclude (a) too much room air beingdrawn through the cabinet, causingpotential product contamination, or (b)too little air drawn through the cabinet,compromising personnel protection.Located between the cabinet exhaustHEPA filter and the building exhaust system, the canopy exhaust air gap maintains Type A2 cabinet performanceover a range of exhaust flow rates (see chart). Type A2 cabinet performanceis also maintained in the event of totalbuilding exhaust failure. TheSterilGARD® III Advance° canopyexhaust connection (CEC) incorporatesthe following features:• Designed for quick, efficient installa-tion to protect cabinet from performancedegradation caused by in-house exhaustsystem fluctuation.• Unique combination design consists ofan Exhaust Transition (ET) which createsthe desired air gap in the overall connection.• Exhaust connection collar is 10" diameter on the CEC403a (for the 4' cabinet) and 10" diameter on theCEC603a (for the 6' cabinet).

Exhaust Transition (ET) for Hard ConnectionThe exhaust transition (ET) provides ahard connection, or airtight mating ofthe cabinet to the exhaust duct, in whichflow rates for cabinet intake, cabinetexhaust and building exhaust are allequal. The ET403a and ET603a aredesigned for the 4' and 6' cabinets,respectively. The exhaust transitionincorporates the following features: • Rectangular base is easily installed ontop of the cabinet over the exhaust filter,requiring only 4 or 5 bolts.• Design enables filter leak checkingthrough a convenient access panel. • Exhaust connection collar is 10" diameter.

Filtration System• Exhaust and supply filters zero-probed HEPA, both 99.99% efficienton all particles 0.3 micrometers byDOP test.

1

2

3

S t e r i l G A R D ® I I I A d v a n c e °

Exhaust Requirements

Canopy Exhaust Connection

SG403a CEC403a8" 322-672 0.02-0.20"wc

10"10" 400-750 0.03-0.22"wc

SG603a CEC603a8" 490-840 0.04-0.31"wc10" 612-965 0.06-0.33"wc

P a g e 1 1

Cabinet Exhaust Sash Exhaust Flow Pressure Loss DuctModels Connection Height Rate (CFM) Range Diameter

SterilGARD®III Advance°

CEC Exhaust transitionTop of cabinetExhaust HEPA filter

1

2

3

Exhaust Transition (Hard Connection)*

SG403a ET403a8" 269 0.1"wc

10"10" 335 0.1"wc

SG603a ET603a8" 408 0.2"wc10" 510 0.3"wc

*Cabinets previously classified as Class II, Type A/B3 are now classified as Class II, Type A2. These cabinets may beexhausted to the room (previous Type A classification) or connected to a facility exhaust system (previous Type B3classification) via a "canopy" connection (previously referred to as "thimble"). NSF 49 (2002) specifies the "canopy"connection only because vented exhaust through a "hard" connection, though technically possible, may not meet NSFcriteria. Consult The Baker Company or your laboratory safety officer for details.

Page 12: SterilGARD III Advance° Sets the Industry’s Highest Standard for … · 2017-12-15 · sealing prior to total cabinet decontamination. • The telescoping UniPressureTM Preflow

1. Manufacturer shall provide a certified copy of the personnel, product and cross-contamina-tion (biological) tests, equivalent to or moredemanding than as specified in NSF InternationalStandard #49, performed on the unit selectedfrom the corresponding statistical sample. Testsmay be witnessed by a representative of the purchaser.2. Cabinet shall have momentum air curtain

downflow velocity profile, i.e., a higher velocity ofdownflow behind the viewscreen relative todownflow velocity over the work surface.3. High velocity return air slots shall be located ateach end of the front access opening. These slotshelp to prevent contaminated air from beingdrawn into the work area along the edges of theside wall and from escaping the work area to theambient environment.4. The sliding viewscreen shall be slanted at anangle of 10° from vertical, capable of moving to afully closed position during shutdown periods.5. Viewcreen shall be constructed of 1⁄4" safetyplate glass, with a maximum opening of 173⁄4" forequipment loading.6. All biologically contaminated ducts, plenumsand work area side walls shall be permanentmetal construction and maintained under negativepressure or enclosed within a negative pressurezone.7. Cabinet shall be capable of automatically handling a 60% minimum increase in filter loading without reducing total air delivery bymore than 10%. With use of the speed controller,a 180% increase in airflow shall be attainable.Test data to verify these capabilities shall beavailable upon request. 8. Interior work area shall be 27 1⁄2" high.9. Cabinet shall have The Baker Company’sexclusive UniPressure Preflow Plenum, designedto provide more uniform airflow to the supply filter.10. Supply and exhaust filters shall be front-loading.11. A telescoping plenum assembly shall be provided to allow the filters to be directlyclamped to the plenum against a closed-cell neoprene gasket. Plenum applies force to fullperimeter of filters, rather than point force.

12. Complete unit shall be listed as certified byUnderwriters Laboratory (UL) for electrical, fireand personal safety.13. Sliding sash high velocity air return slots shallbe provided along the entire top of the work areato prevent migration of contaminated air behindthe glass viewscreen.14. Cabinet shall have a microprocessor-basedcontrol system with a membrane control panelmounted on the front of the cabinet facing downtowards the user when sitting at the unit.15. Unit shall have an audible alarm and a flash-ing LED to indicate when the sliding viewscreenis in an unsafe position. An alarm mute switchshall be provided on the front-mounted cabinetcontrol panel to allow the operator to mute thealarm tone for brief adjustments. The alarm shallautomatically reset after five minutes if theviewscreen remains in an unsafe position.16. Intake velocity through the 8" front accessopening shall be minimum of 100 FPM.17. Cabinet exterior construction:

a. seal panels of 16-gauge cold-rolled steel b. dress panels of 16-gauge cold-rolled steel,painted PermaWhite™.

18. Cabinet interior (work area) construction: one-piece 16-gauge, Type 304 stainless steel,with a smooth, 7⁄16" radius between rear and sidewalls, and easily cleanable, radiused corners onthe work surface tray.19. Work area side walls and rear wall to be one-piece construction. A straight back wall shall beprovided to maximize work area and easilyaccommodate laboratory equipment.20. Work area shall be provided with two GFIprotected duplex outlets with drip-proof coversand circuit breaker.21. Lower front work area airfoil shall be provided to improve access opening containmentcapability.22. Cabinet shall have a unitized drain pan with7⁄16" radius on all sides and a fully-removable worksurface and work surface supports to facilitatecleaning.23. Stainless steel air diffuser and filter protectorprovided in work area.24. Unit must be listed by NSF International asmeeting Standard #49.25. Each unit, before shipping, shall have a complete physical test to assure cabinet meets Class II requirements. A copy of this test will beprovided with the unit.

26. Speed controller shall automatically compensate for voltage change to maintain constant voltage to motor while allowing for manual adjustments during filter loading.27. Adjustable internal damper provided to compensate for changing resistance of exhaustand supply filters.28. A single power cord and plug shall be provided for electrical power source.29. One petcock and one plugged penetration areprovided as standard on the right side wall. Leftside wall is prepunched for optional/additionalplumbing connections.30. All external plumbing connections to the petcocks shall be made through the bottom of the cabinet and not the sides, allowing zeroclearance between the unit and the building wallsor equipment to its right and left.31. The unit shall have standard HEPA filters for aprotection effectiveness of 99.99% when filteringparticles of 0.3 micron size.32. The unit shall have (optional) ultraviolet lightwith a shutoff safety feature when the viewscreenis raised.33. The unit shall have electronic ballasts for UVand fluorescent lighting to provide longer life andlower heat output.34. The unit is available with an optional stand,which includes telescoping legs that allow thework surface height to be set from 30 1⁄8" to 38 5⁄8".

WarrantyThe Baker Company, Inc., expressly represents and warrants all goods (a) to be as specified (and described) in The BakerCompany catalogs and literature, and (b) to be free under normal use, service and testing (all as described in The BakerCompany catalogs and literature) from defects in material andworkmanship for a period of thirty-six months from the invoice date.

The exclusive remedy for any breach or violation of thiswarranty is as follows: The Baker Company, Inc., will F.O.B.Sanford, Maine, furnish without charge repairs to or replace-ment of the parts or equipment which proved defective inmaterial or workmanship. No claim may be made for any incidental or consequential damages. THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHERWARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANYIMPLIED WARRANTY OF MERCHANTABILITY OR FITNESSFOR A PARTICULAR PURPOSE UNLESS OTHERWISEAGREED IN WRITING SIGNED BY THE BAKER COMPANY.(THE BAKER COMPANY SHALL NOT BE RESPONSIBLE FORANY IMPROPER USE, INSTALLATION, SERVICE OR TESTING OF THE GOODS.)

OW7188.18 3/04

P.O. Drawer E, Sanford, Maine 04073 (207) 324-8773 1-800-992-2537 FAX (207) 324-3869 www.bakerco.com©2004 The Baker Company

Purchase SpecificationSterilGARD® III Advance˚Class II, Type A2 BiologicalSafety Cabinet

S t e r i l G A R D ® I I I A d v a n c e °


Recommended