medentis medical GmbH · 53507 Dernau · Gartenstraße 12 · Germany · Fon: +49 02643 902000-0 · Fax: +49 02643 902000-20 www.medentis.de Surgical manual 5.0 ICX templant® optimized surgical protocol SILVERSPONSOR of the DGI BRONZESPONSOR of the EAO Years 2 0 Y e a r s G u a r a n t e e d S p a r e P a r t A v a i l a b i l i t y O s s e o i n t e g r e a t i o n W a r r a n t y
Transcript

medentis medical GmbH · 53507 Dernau · Gartenstraße 12 · Germany · Fon: +49 02643 902000-0 · Fax: +49 02643 902000-20

w w w . m e d e n t i s . d e

Surgical manual 5.0ICX templant®

optimized surgical protocol

SilverSponSor of the DGi

Bronzesponsor of the eAoYears

20 Years

Guaranteed Spare Part A

vaila

bilit

yOsse

ointegreation

W a r r a n t y

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e2

Explanation of the color schemes and the color codingof the ICX templant® drills

ICX-templant drills for implant diameter 4,1 belong to the GREEN color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 green marker for very soft bones D4

2 green markers for soft/medium hard bones D3/D2

3 green markers for hard bones, D1

ICX-templant drills for implant diameter 4,8 belong to the bluE color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 blue marker for very soft bones D4

2 blue markers for soft/medium hard bones D3/D2

3 blue markers for hard bones, D1

Ø 4.1mm

Ø 4.8mm

ICX-templant drills for implant diameter 3,75 belong to the RED color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 red marker for very soft bones D4

2 red markers for soft/medium hard bones D3/D2

3 red markers for hard bones, D1

Ø 3.75mm

ICX-plus templant drills for implant diameter 3,45 belong to the yEllow color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 yellow marker for very soft bones D4

2 yellow markers for soft/medium hard bones D3/D2

3 yellow markers for hard bones, D1

Ø 3.45mm

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e 3

Explanation of the color schemes and the color codingof the ICX templant® drills

Explanation of the color schemes and the color codingof the ICX templant® drills

ICX-templant drills for implant diameter 4,1 belong to the GREEN color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 green marker for very soft bones D4

2 green markers for soft/medium hard bones D3/D2

3 green markers for hard bones, D1

ICX-templant drills for implant diameter 4,8 belong to the bluE color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 blue marker for very soft bones D4

2 blue markers for soft/medium hard bones D3/D2

3 blue markers for hard bones, D1

Ø 4.8mm

Color coding sample:Green = 3,45 implant diameter1 green marker = drill for very soft bone D4

Color coding sample:Blue = 3,45 implant diameter1 blue marker = drill for very soft bone D4

Ø 4.1mm

ICX-templant drills for implant diameter 3,75 belong to the RED color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 red marker for very soft bones D4

2 red markers for soft/medium hard bones D3/D2

3 red markers for hard bones, D1

Color coding sample:Red = 3,45 implant diameter1 red marker = drill for very soft bone D4

Ø 3.75mm

ICX-plus templant drills for implant diameter 3,45 belong to the yEllow color scheme.

Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).

1 yellow marker for very soft bones D4

2 yellow markers for soft/medium hard bones D3/D2

3 yellow markers for hard bones, D1

Color coding sample:Yellow = 3,45 implant diameter1 yellow marker = drill for very soft bone D4

Ø 3.45mm

Ø 2,9Surgical procedure for ICX-mini templant® Ø 2.9mm

Service-Fon: +49 02643 902000-06

Ø 2.9mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

drills for very soft

to medium hard bones,

D2/D3/D4

drills for

hard bones,

D1

10.0mm

12.5mm

8.0mm6.5mm

Ø 2,9Surgical procedure for ICX-mini templant® Ø 2.9mm D2/D3/D4 drilling sequence for ICX-mini templant® Ø 2.9mm

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e 7

Ø 2.9mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Drilling sequence for very soft to medium hard bones.

Please only use drills with 2 markers.

Important Note: drills for very soft to medium hard bones D2/D3/D4 are labeled with 2 white markers

ICX -mini templant®Ø2.9mm

ICX guide drill ICX D2/D3/D4-drill,

2 markers, white

D1-drilling sequence for ICX-mini templant® Ø 2.9mm

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e8

Drilling sequencefor hard bones.

Please only use drills with 3 markers.

Important Note: drills for hard bones D1, are labeled with 3 white markers

Explanation of the color schemes and the color coding of the ICX-mini templant®– Ø 2.9mm drills

Ø 2.9mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

ICX -mini templant®Ø2.9mm

ICX guide drill

ICX D1-drill, 3 markers, white

w w w . m e d e n t i s . d e 9

Explanation of the color schemes and the color coding of the ICX-mini templant®– Ø 2.9mm drills

ICX-mini templant drills for implant diameter 2,9 belong to the white color scheme.

Single ICX mini templant drills have different white color codes subject to the bone quality (very soft, soft, hard).

2 white markers for soft/medium hard bones D3/D2

3 white markers for hard bones, D1

Color coding sample:White = 2,9 implant diameter2 white markers = drill for soft to medium-hard bones D4/D3/D2

Ø 2.9mm

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e8

Important note for the prosthetic provision of all ICX-plus implants.

both the open and the closed impression posts fit on the ICX and ICX-plus implants. only the universal lab Analogs differ.

Especially for a very low mucosa height, it is mandatory that the technician is aware, which implant is an ICX-plus implant. we recommend invariably informing the technician whether it‘s ICX or ICX-plus implants.(this is not mandatory. The prosthetic devices fit both on the ICX as well as on the ICX-plus implants.)

Especially for ICX-plus, we have developed the straight titanium abutment with 0 mm gingiva height.It only fits on the ICX-plus implants - all other prosthetic components fit both on ICX and on ICX-plus implants.

Important Note + Important Note!

Specificity for all ICX-plus implants.

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e 9

Important note for the prosthetic provision of all ICX-plus implants.

For ICX-plus implant Ø 3.45 mm, we recommend the ICX-plus universal lab Analog. Art-no.: C-006-010002.

Art-no.: C-006-010001 we do not recommend forICX-plus implant Ø 3.45mm!

Important Note + Important Note!

Specificity for all ICX-plus implants.

Ø 3,45

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e

Surgical procedure for ICX-plus templant® Ø 3.45mm

12

Drill for

very soft bone,

D4

Drill for soft/

medium-hard bones,

D3/D2

Drill for

hard bones,

D1

ICX stop drill,

mandatory use

10.0mm

12.5mm

15.0mm

8.0mm

6.5mm

Ø 3.45mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 3,45Drilling sequence for very soft bone

Please only use drills with 1 color marker.

Important Note: drills for very soft bones D4,are labeled with 1 yellow marker

w w w . m e d e n t i s . d e

Surgical procedure for ICX-plus templant® Ø 3.45mm D4 drilling sequence for ICX-plus templant® Ø 3.45mm

13

transgingival = collar does not sink subgingival = collar sinks

ICX Stop-drill, yellow

ICX Stop-drill, yellow

ICX-plusØ3.45mm

ICX-plusØ3.45mm

ICX D4-drill, 1 marker,yellow

ICX guide drill

ICX D4-drill, 2 markers, white

Note:If the collar is sunk, the cavity needs to be 1.7mm deeper.

Standard Optionally

ICX stop drill,

mandatory use

Ø 3,75mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 3.45mm

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e

D2/D3 drilling sequence for ICX-plus templant® Ø 3.45mm

14

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 yellow markers

transgingival = collar does not sink subgingival = collar sinks

ICX Stop-drill, yellow

ICX Stop-drill, yellow

ICX-plusØ3.45mm

ICX-plusØ3.45mm

ICX D2/D3-drills, 2 markers, white

ICX guide drill

ICX D2/D3-drills, 2 markers, yellow

Note:If the collar is sunk, the cavity needs to be 1.7mm deeper.

Standard Optionally

Ø 3.45mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

w w w . m e d e n t i s . d e

D2/D3 drilling sequence for ICX-plus templant® Ø 3.45mm D1 drilling sequence for ICX-plus templant® Ø 3.45mm

15

Drilling sequence for hard bones Please only use drills with 3 color markers.

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 yellow markers

Important Note: drills for hard bones D1, are labeled with 3 yellow markers

transgingival = collar does not sink subgingival = collar sinks

ICX Stop-drill, yellow

ICX Stop-drill, yellow

ICX-plusØ3.45mm

ICX-plusØ3.45mm

ICX D1-drill, 3 markers, yellow

ICX guide drill

ICX D1-drill, 3 markers, white

Note:If the collar is sunk, the cavity needs to be 1.7mm deeper.

Standard Optionally

Ø 3,75mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 3.45mm

Ø 3,75Drilling sequence for very soft bone

Please only use drills with 1 color marker.

Surgical procedure for ICX–templant® Ø 3.75mm

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e16

Drill for

very soft bone,

D4

Drill for soft/

medium-hard bones,

D3/D2

Drill for

hard bones,

D1

ICX stop drill,

mandatory use

12.5mm

15.0mm

10.0mm

8.0mm 6.5mm

Ø 3.75mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 3,75Drilling sequence for very soft bone

Please only use drills with 1 color marker.

Important Note: drills for very soft bones D4,are labeled with 1 red marker

D4 drilling sequence for ICX–templant® Ø 3.75mm

w w w . m e d e n t i s . d e 17

ICX implantØ3.75

ICX Stop-drill, red

ICX D4-drill, 1 marker,red

ICX guide drill

ICX D4-drill, 2 markers, white

Ø 3.75mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Service-Fon: +49 02643 902000-0

D2/D3 drilling sequence for ICX–templant® Ø 3.75mm

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 red markers

Important Note: drills for hard bones D1,are labeled with 3 red markers

18

ICX implantØ3.75

ICX Stop-drill, red

ICX D2/D3-drills, 2 markers, red

ICX guide drill

ICX D2/D3-drills, 2 markers, white

Ø 3.75mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

19Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e

D2/D3 drilling sequence for ICX–templant® Ø 3.75mm D1 drilling sequence for ICX–templant® Ø 3.75mm

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Drilling sequence for hard bones Please only use drills with 3 color markers.

Important Note: drills for hard bones D1,are labeled with 3 red markers

ICX implantØ3.75

ICX Stop-drill, red

ICX D1-drill, 3 markers, red

ICX guide drill

ICX D1-drill, 3 markers, white

Ø 3.75mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 4,1

20 Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e

Surgical procedure for ICX–templant® Ø 4.1mm D4 drilling sequence for ICX–templant® Ø 4.1mm

Drilling sequence for very soft bone Please only use drills with 1 color marker.

Drill for

very soft bone,

D4

Drill for soft/

medium-hard bones,

D3/D2

Drill for

hard bones,

D1

ICX stop drill,

mandatory use

12.5mm

15.0mm

10.0mm

8.0mm 6.5mm

Ø 4.1mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 4,1

w w w . m e d e n t i s . d e

D4 drilling sequence for ICX–templant® Ø 4.1mm

Drilling sequence for very soft bone Please only use drills with 1 color marker.

Important Note: drills for very soft bones D4, are labeled with 1 green marker

21

ICX implantØ4.1

ICX D4-drill, 1 marker,green

ICX D4-drill, 1 marker,red

ICX guide drill

ICX D4-drill, 2 markers, white

ICX Stop-drill, green

ICX stop drill,

mandatory use

Ø 4.1mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Service-Fon: +49 02643 902000-0

D2/D3 drilling sequence for ICX–templant® Ø 4.1mm

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 green markers

Important Note: drills for hard bones D1,are labeled with 3 green markers

22

ICX implantØ4.1

ICX Stop-drill, green

ICX D2/D3-drills, 2 markers, green

ICX D2/D3-drills, 2 markers, red

ICX guide drill

ICX D2/D3-drills, 2 markers, white

Ø 4.1mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e

D2/D3 drilling sequence for ICX–templant® Ø 4.1mm D1 drilling sequence for ICX–templant® Ø 4.1mm

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Drilling sequence for hard bones Please only use drills with 3 color markers.

Important Note: drills for hard bones D1,are labeled with 3 green markers

23

ICX ImplantatØ4.1

ICX Stop-drill, green

ICX D1-drill, 3 markers, green

ICX D1-drill, 3 markers, red

ICX guide drill

ICX D1-drill, 3 markers, white

Ø 4.1mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 4,8

Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e

Surgical procedure for ICX–templant® Ø 4.8mm D4 drilling sequence for ICX–templant® Ø 4.8mm

24

Drill for

very soft bone,

D4

Drill for soft/

medium-hard bones,

D3/D2

Drill for

hard bones,

D1

ICX stop drill,

mandatory use

12.5mm

15.0mm

10.0mm

8.0mm 6.5mm

Ø 4.8mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Ø 4,8

w w w . m e d e n t i s . d e

D4 drilling sequence for ICX–templant® Ø 4.8mm

Drilling sequence for very soft bone Please only use drills with 1 color marker.

Important Note: drills for very soft bones D4,are labeled with 1 blue marker

25

ICX D4-drill, 1 marker,blue

ICX D4-drill, 1 marker,red

ICX implantØ4.8

ICX guide drill

ICX D4-drill, 2 markers, white

ICX Stop-drill, blue

ICX stop drill,

mandatory use

Ø 4.8mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Service-Fon: +49 02643 902000-026

D2/D3 drilling sequence for ICX–templant® Ø 4.8mm

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 blue markers

Important Note: drills for hard bones D1, are labeled with 3 blue markers

ICX D2/D3-drills, 2 markers, blue

ICX implantØ4.8

ICX Stop-drill, blue

ICX guide drill

ICX D2/D3-drills, 2 markers, white

ICX D2/D3-drills, 2 markers, red

Ø 4.8mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

27Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e

D2/D3 drilling sequence for ICX–templant® Ø 4.8mm D1 drilling sequence for ICX–templant® Ø 4.8mm

Drilling sequence for soft to medium hard bones

Please only use drills with 2 color markers.

Drilling sequence for hard bones Please only use drills with 3 color markers.

Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 blue markers

Important Note: drills for hard bones D1, are labeled with 3 blue markers

ICX D1-drill, 3 markers, blue

ICX implantØ4.8

ICX guide drill

ICX D1-drill, 3 markers, white

ICX Stop-drill, blue

ICX D1-drill, 3 markers, red

Ø 4.8mm from 25 rpm

ICX templant-Implant rpm

from 400 rpmICX templant-Drill

rpm

Service-Fon: +49 02643 902000-0 w w w . t e m p l a n t . d e

Handling/Insertion of the ICX templant®

Compatibly to elbow and hand adapter

4 5

1 2 3

10

6

7 8 9

Sterile ICX-packageExtraction of the ICX templant®with the ICX torque wrench driver

ICX torque wrenchDriver short and medium

ICX torque wrenchInserted in the dental motor

ICX torque wrenchDriver inserted in hand-driver

Finally insertedcover cap

Insertion of the ICX templant®

Extraction of the ICX templant®with the ICX torque wrench driver

Extraction of the ICX templant®with the ICX torque wrench driver

Insertion of the cover cap


Recommended