SYSTEM OPTIMIZATION MANUAL Page 1 www.phenomenex.com Get the Most out of your Kinetex ® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance benefits on any HPLC system. The key to achieving ultra-high performance using Kinetex core-shell columns on an existing legacy HPLC may include optimizing: 1) The HPLC Method pg. 2 2) HPLC System Volume pg. 6 3) Detector Settings pg. 13 All sections may be important for you to achieve maximum benefits from the core-shell media.
Transcript
Page 1: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

SYSTEM OPTIMIZATION MANUAL

Page 1www.phenomenex.com

Get the Most out of your Kinetex® Core-Shell Column

Kinetex core-shell media can generate UHPLC-type performance benefits on any HPLC system.

The key to achieving ultra-high performance using Kinetex core-shell columns on an existing legacy HPLC may include optimizing:

1) The HPLC Method pg. 2

2) HPLC System Volume pg. 6

3) Detector Settings pg. 13

All sections may be important for you to achieve maximum benefits from the core-shell media.

Page 2: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 2www.phenomenex.com

1. The HPLC Method

Why This is Important The following method-related technique improvements will improve performance for all column separations. When using Kinetex® core-shell columns, the following suggestions will dramatically improve performance.

HPLC methods should first work towards focusing the sample on the head of the column, •and then minimizing sample dispersion through the column.

Incorporating simple injector programming in the method can minimize extra-column volume to •improve peak shape and efficiency, especially for fast LC separations.

Trouble: All Peaks are BroadCheck: Sample Volume 1. Errors in sample loading may negate the performance gain the Kinetex core-shell particle can achieve. If you inject the sample in the same organic composition as the mobile phase under isocratic conditions, the initial bandwidth on column will be directly proportional to the injection volume. As such, a large volume sample will load as a broad band on the column (low efficiency).

SOLUTION

Check: Sample Diluent 2. If the organic strength of the diluent is greater than the HPLC mobile phase, then the sample will load as a diffuse band on the head of the HPLC column. This may occur in either isocratic or gradient methods and can negate much of the performance advantages that the Kinetex column delivers.

Gradient applications where the sample is injected in stronger diluents result in compound surfing effects, which will cause •peaks to broaden and split.

Isocratic methods where the diluent is stronger than the mobile phase will result in excessive sample dispersion. •

SOLUTION

Pre-concentrate your sample. To reduce the efficiency loss of large volume injections, it is possible to pre-concentrate the sample on the head of the column by using a weaker diluent than the mobile phase.

Inject a weak solvent. It is best to use a diluent either weaker or equal to the organic strength of the mobile phase so that the sample focuses on the head of the column, resulting in sharper peaks (i.e., higher efficiency).

Page 3: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 3www.phenomenex.com

Trouble: All Peaks are Broad (continued)Check: Injector Program 3. Extra-column volume before the column will cause dispersion of the sample on the way to the column; the injector loop is a major source of extra-column volume.

SOLUTION

Table 1:

Figure 1: Dwell Volume Reduction Using an Injector Program

* Mainpass peak is an injector artifact due to switching the injector valve back to include the injector loop in the flow (to flush injector in preparation for the next injection).

Steps Commands Comments1 DRAW Draw volume of sample (injection volume) from vial

2 INJECT Introduce sample into flow path

3 WAIT 0.06 min (calculated wait time) Flush sample loop after injection (wait time = 6x (injection volume + 5 μL) / flow rate)

4 VALVE bypass Direct flow from pump to column, bypassing injection valve to exclude delay volume (~200 – 500 μL from auto-injector path)

5 WAIT 1.5 min The period of VALVE bypass time (Wait time = Run time - 1 min)

6 VALVE mainpass Switch valve from bypass to injection position

1. The HPLC Method

Bypass the loop. It is possible to bypass the loop with an injector program during the analysis method (after the sample has left the injector loop). Table 1 and Figure 1 show an example where the injection loop is bypassed after injection to reduce system dwell volume.

Page 4: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 4www.phenomenex.com

Trouble: The First Peak is BroadCheck: Polar Compound Focusing

In reversed phase HPLC, early eluting compounds typically display the most peak distortion. Under isocratic conditions, the peak volume for early eluters is comparable to the extra-column volume, while under gradient conditions there is little or no refocusing of weakly retained compounds at the column inlet.

SOLUTION

Figure 2: Injector Program For Sample Focusing

1. The HPLC Method

Focus your sample. Inject a volume of water, approximately 3x the sample injection volume, after the sample using an injector program. This will focus the sample on the head of the column and increase retention, resulting in narrower, more symmetrical peaks. The improvement in resolution is larger with small diameter columns since these provide the smallest peak volumes. An example is shown in Figures 2 and 3.

Strong Solvent Sample

Solvent Dilutes in Injector Loop

Diluted Injection Solvent Concentrates On Column

Results in more focused analyte peak

with higher efficiency

Page 5: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 5www.phenomenex.com

Trouble: The First Peak is Broad (continued)

Figure 3: Using Sample Focusing to Improve Early Eluting Peak Shape

HPLC Method Optimization: Pros and Cons

HPLC Method Optimization: Pros and Cons Benefits

Reduced run time•

Improved peak efficiency •

Improved resolution of closely eluting peaks•

Improved peak shape of early eluting analytes•

Drawbacks

Increased equilibration time (injector program)•

Limited sample injection volumes•

Possible baseline disruptions in chromatogram (injector program)•

1. The HPLC Method

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min0.0

2000.0

4000.0

6000.0

8000.0

1.0e4

1.2e4

1.4e4

1.6e4

1.8e4

2.0e4

2.2e4

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.50.0

2000.0

4000.0

6000.0

8000.0

1.0e4

1.2e4

1.4e4

1.6e4

1.8e4

2.0e4

2.2e4

min

19168

Ap

p ID

191

68

RT = 0.88, 0.98 minS/N = 20,54Rs = 1.1

RT = 1.06, 1.15 minS/N = 27,68Rs = 1.4

Without sample focusing

With sample focusing

Page 6: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 6www.phenomenex.com

Why This is Important The HPLC systems that make up most of the instrument base today are rugged, high performance capable systems optimized by the manufacturer for use with very high volume, fully porous 5 μm media in standard analytical (250 x 4.6 mm) column dimensions. Such a design may significantly mask the separation power of Kinetex® core-shell columns.

To realize the maximum performance of the Kinetex core-shell, the link between extra-column system volume and performance should be considered.

The fraction of time that an analyte spends inside the HPLC column during a run is productive: •analytes are being separated with high efficiency.

The fraction of time that an analyte spends outside of the HPLC column during a run is unproductive: •analytes diffuse within connecting tubing, lowering efficiency.

Lower relative extra-column volume will result in higher overall system performance. • o Sample dispersion occurs before the column o Peak dispersion occurs after the column

Question: Where should I look to minimize sample dispersion?

1. Check: Volume Before the Column

The part of an HPLC system before the column is typically the most overlooked area to improve overall performance since its effect on overall column performance is less than post-column volume changes. There are many opportunities to reduce pre-column volume and improve system performance.

Needle Seat.• On many HPLC systems, the injector needle and needle seat are part of the injection loop and represent significant amounts of dead volume.

o For both integrated needle and needle seat systems, the standard needle seat can be easily replaced with a low volume needle seat. The standard needle seat capillary on the Agilent® 1100 (0.17 x 100 mm; green) has a volume of 2.5 μL.

SOLUTION

2. HPLC System Volume

?

Use lower volume needle seat. A lower volume needle seat capillary (0.12 x 100 mm; red) reduces the volume by almost one-half to 1.3 μL. Figure 1 shows an example after an Agilent 1100 has had the needle seat capillary replaced.

Page 7: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 7www.phenomenex.com

Figure 1: Chromatography Improvement by System Optimization

Kinetex® 100 x 2.1 mm Un-optimized HPLC System

Fixed Volume Loop• . A fixed volume loop is used as part of the injector and oversized sample loops can contribute greatly to extra-column volume of a system.

SOLUTION

Solvent Heat Exchangers• . In the case of heat exchangers, tubing volume is being added to change the temperature of the mobile phase to more closely match the temperature of the column.

SOLUTION

Column Switching Valves• . Column switching valves are a great method development tool that adds significant volume to the system.

SOLUTION

Fittings and Unions• . These can add significant dead volume if fittings are improperly connected. Older tubing unions can also add dead volume to a system (Figure 2).

SOLUTION

2. HPLC System Volume

Replace with lower volume loop. In all cases, the injection loop volume can be reduced to volumes more appropriate for a particular application. In some cases where very large diluent volumes are being used, a trap column can be investigated to minimize injection loop volumes.

Temperature control of solvent reservoir. If mobile phase absolutely must be heated (or chilled), then one should consider controlling the temperature of the solvent reservoir instead and avoid using heat exchangers.

Bypass the switching valve. Alternatively, a lower volume column switching valve can be used.

Zero volume finger-tight fittings. Tubing unions should be zero dead volume (ZDV) unions. Finger-tight fittings ensure a better connection with parts from multiple vendors.

0.0 1.0 min

0

100

200

300

400

mAU

19614

1 2

3

4

5

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 min

0

100

200

300

400

mAU

19615

1

2

3

4

5

Kinetex 100 x 2.1 mm Optimized HPLC System

Ap

p ID

196

15

Ap

p ID

196

14

Page 8: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 8www.phenomenex.com

Figure 2: PEEK® Sure-Lok™ Nuts and Zero Dead Volume Union

Tubing• . Most legacy HPLC systems are plumbed throughout the system with 0.010 in. ID tubing. The 0.010 in. ID tubing has a solvent volume of 1.3 µL per inch (0.51 μL per cm), adding a significant amount of dead volume to a HPLC system.

SOLUTION

SOLUTION

Table 1: Estimating Capillary Tubing Volume

Figure 3: Large ID Tubing Increases System Volume and Sample Dispersion

2. HPLC System Volume

Use 0.005 in. tubing. 0.005 in. tubing has 0.3 µL per inch (0.13 μL per cm) internal volume. This will provide an approximate extra-column volume reduction of 1 µL per inch (or 0.38 μL per cm) and should be plumbed from the solvent mixing “T” all the way to the HPLC column inlet. See Table 1 and Figure 3.

ID (inches) ID (mm) µL/inch µL/cm0.010 0.254 1.287 0.51

0.007 0.178 0.633 0.25

0.005 0.127 0.322 0.13

0.010 in. ID tubing

0.005 in. ID tubing

Peak becomes diluted and dispersed

Peak integrity is maintained

Proper Tubing Cuts. Rough cut tubing can potentially increase dead volume if such fittings compromise a smooth coupling. Precut tubing of fixed lengths can be pur-chased from numerous sources.

Page 9: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 9www.phenomenex.com

2. HPLC System Volume

Pre-Column System Optimization: Pros and ConsBenefits

Reduced run time•

Increased efficiency•

Better resolution of closely eluting peaks•

Drawbacks

Higher system backpressure•

Potential method reproducibility concerns (column heater and static mixer)•

Limited sample injection volumes•

Question: Where should I look to minimize peak dispersion?

1. Check: Volume After the Column

Minimizing post-column volumes significantly decreases peak dispersion and has the greatest effect on the apparent performance of a column. The areas of post-column volume where system optimization can influence column performance will depend on the type of detection used.

UV Detection.• The UV flow cell is an important place where a minor change like using a lower volume flow cell can make a big difference in column performance. Larger flow cells tend to disperse narrow peaks and reduce the apparent efficiency of the column; this can also result in loss of resolution for closely eluting peaks.

SOLUTION Install a lower volume flow cell. Moving to a lower volume flow cell (≤3 µL) can produce sharper peak shapes and better resolution for most closely eluting peaks (Figure 4).

?

Page 10: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 10www.phenomenex.com

2. HPLC System Volume

Figure 4: Peak Dispersion Due to Flow Cell Incompatibilities

MS Interface• . Extra volume in the electrospray interface can reduce apparent performance similar to what is seen with a UV detector flow cell. You must also consider that when an MS detector is being used, the distance between the column outlet and the MS detector can be quite large.

SOLUTION Use a lower volume ESI interface. Using capillary lines in the interface can reduce volume as can using a sheath-flow for low volume separations. Figure 5 shows an example of the differences between the API 3000™ and API 4000™ interfaces; the API 4000™ shows a four-fold improvement in peak width due in part to a lower volume ESI interface.

Standard Flow Cell

A narrow concentrated peak will diffuse in a large flow cell resulting in lower efficiency

A low volume flow cell minimizes peak dispersion through the flow cell

Low Volume Flow Cell

Page 11: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 11www.phenomenex.com

Figure 5: ESI Interface Volume Influences Peak Dispersion

Meperidine peak width comparison using two different MS sources

Fittings and Unions• . These can add significant dead volume if fittings are improperly cut and connected. Older tubing unions can also add dead volume to a system.

SOLUTION

Tubing• . Most legacy HPLC systems are plumbed throughout the system with 0.010 in. ID tubing. The 0.010 in. ID tubing has a solvent volume of 1.3 µL per inch (0.51 μL per cm), adding a significant amount of dead volume to a HPLC system.

SOLUTION

SOLUTION

Table 2: Estimating Capillary Tubing Volume

2. HPLC System Volume

Proper Tubing Cuts. Rough cut tubing can potentially increase dead volume if such fittings com-promise a smooth coupling. Precut tubing of fixed lengths can be purchased from numerous sources.

Zero volume finger-tight fittings. Tubing unions should be zero dead volume (ZDV) unions. Finger-tight fittings ensure a better connection with parts from multiple vendors.

Use 0.005 in. tubing. 0.005 in. tubing has 0.3 µL per inch (0.13 μL per cm) internal volume. This will provide an approximate extra-column volume reduction of 1 µL per inch (or 0.38 μL per cm) and should be plumbed from the HPLC column outlet to the inlet of the UV flow cell or electrospray interface (Table 2 and Figure 6).

ID (inches) ID (mm) µL/inch µL/cm0.010 0.254 1.287 0.51

0.007 0.178 0.633 0.25

0.005 0.127 0.322 0.13

Ap

p ID

1885

0A

pp

ID19

177

PW @ base = 5.4 s

PW @ base = 1.2 s

0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5 min0.0

5.0e5

1.0e6

1.5e6

2.0e6

2.5e6

3.0e6

3.5e6

4.0e6

4.5e6

5.0e6

5.5e6

18850

Inte

nsity

, cps

1.080.855 0.860 0.865 0.870 0.875 0.880 0.885 0.890 0.895 min

0.01.0e52.0e53.0e54.0e55.0e56.0e57.0e58.0e59.0e51.0e61.1e61.2e61.3e61.4e61.5e61.6e61.7e61.8e61.9e62.0e62.1e62.2e62.3e62.4e62.5e6

Inte

nsity

, cp

s

0.87

1.97 1.98 1.99 2.00 2.01 2.02 2.03 2.04 2.05 2.06 2.07 2.08 2.09 2.10 2.11 min0.0

5.0e4

1.0e5

1.5e5

2.0e5

2.5e5

3.0e5

3.5e5

4.0e5

4.5e5

5.0e5

5.5e55.7e5

Inte

nsity

, cp

s

2.01

API 4000™ QTRAP®

0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 min0.0

1.0e52.0e53.0e54.0e55.0e56.0e57.0e58.0e59.0e51.0e61.1e61.2e61.3e61.4e61.5e61.6e61.7e61.8e61.9e62.0e62.1e62.2e62.3e62.4e6

2.30

19177

API 3000™

Page 12: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 12www.phenomenex.com

Figure 6: Large ID Tubing Post-Column Increases Peak Dispersion and Reduces Peak Efficiency

Post-Column System Optimization: Pros and Cons

Benefits

Increased peak efficiency •

Improved resolution of closely eluting peaks•

Drawbacks

Potential reduction in sensitivity (flow cell)•

2. HPLC System Volume

0.010 in. ID tubing

0.005 in. ID tubing

Peak becomes diluted and dispersed

Peak integrity is maintained

Page 13: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 13www.phenomenex.com

3. Detector Settings

Why This is Important

Most HPLC detectors have their data acquisition optimized for high volume peak widths typically seen with 5 μm fully porous media.

The ultra high performance chromatography of Kinetex• ® core-shell columns produce narrow analyte peaks that are highly concentrated, elute quickly and are low volume (Figure 1). The detector sampling rate may require adjustment.

On a variety of different systems this can be found as sampling rate, peak width, or detector time constant.•

Figure 1: Kinetex fast, low volume peaks

Ketone mixtureRs (10,11) = 3.11 Rs (10,11) = 3.12

Fully Porous 5 µm C18

•250x4.6mm

•0.7mL/min

•Agilent® 1100

Kinetex 2.6 µm C18

•50x4.6mm

•3.4mL/min

•Agilent1100

60 min 2.5 min

Average Peak Width: 0.37 min (22 seconds)

Average Peak Volume: 300 µL

Average Peak Width: 0.043 min (2.5 seconds)

Average Peak Volume: 146 µL

0

18280

20 40 60 min

1 23

46

89

11

1312

107

5

10

18276

2 min

1

23

4

5

6

7

8

9

10

11

1213

Ap

p ID

1828

0

Ap

p ID

1827

6

Page 14: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 14www.phenomenex.com

3. Detector Settings

Trouble: Broad and Low Intensity PeaksCheck: UV Detector Scan Rate 1. Many chromatographers using fully porous 10, 5, and even 3 μm analytical columns are not in the habit of changing the detector scan rate.

SOLUTION

Figure 2: Detector Setting Influence on Peak Efficiency and Peak Shape

Check: UV Detector Time Constant 2. A slow detector time constant is normally set to filter out high frequency noise. Unfortunately, this noise filtering can also filter out the sharp peaks and high resolution that Kinetex® core-shell columns deliver.

SOLUTION

Collect 20 points across the peak. For proper integration of a peak, one ideally needs at least 20 data points. Increasing the sampling rate to 20 Hz results in an accurate determination of 1 sec-ond wide peaks. An example of the limitations of low sampling rate is shown in Figure 2.

Increase Detector time constant. Figure 3 and Table 1 show the efficiency improvement when the detector time constant is optimized on an Agilent® 1100 using ChemStation. Figure 4 and Figure 5 show where the settings for the detector sampling rate can be modified in the software for the Agilent® and Waters® instruments, respectively.

Page 15: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 15www.phenomenex.com

3. Detector Settings

Trouble: Broad and Low Intensity Peaks (continued)2. Check: UV Detector Time Constant (continued)

Figure 3: Effect on Peak Efficiency by Changing Time Constant

Table 1: Lists of Time Constants

Time Constant (min) Efficiency (plates/column)0.0025 8073

0.005 8027

0.01 7955

0.03 7598

0.05 6612

0.1* 3978

0.2 1459

*The system default for Agilent® HPLC instruments

Page 16: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 16www.phenomenex.com

3. Detector Settings

Trouble: Broad and Low Intensity Peaks (continued)2. Check: UV Detector Time Constant (Continued)

Figure 4: Agilent® System Detector Rate Settings

Figure 5: Waters® Empower Acquisition Rate Settings

Page 17: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 17www.phenomenex.com

Trouble: Broad and Low Intensity Peaks (continued)

Changing UV Detection Rate: Pros and Cons

Benefits

Increased efficiency of all analyte peaks •

Improved resolution of closely eluting peaks•

More accurate quantitation of peak height and area •

Drawbacks

Increased size of chromatography data file•

Increased baseline noise •

Poor quantitation of very low level peaks•

3. Check: MS Sampling Rate The total number of sample points recorded per second (pts/s). Sampling rate is dictated by the dwell time, settling time, and pause between mass ranges.

Dwell time: the scan time required for one sample point. An MRM will be monitored for the period specified by the dwell time •giving one sample point.

Pause between mass ranges: the length (millisecond) of pause between scanning separate mass ranges.•

Settling time: the amount of time (millisecond) allowed for the system to pause at a m/z setting before data acquisition begins. • o Table 2 shows an example calculation for the sampling rate used in an application (5 MRM transitions monitored) (Phenomenex HPLC 2010 Poster)

Table 2: Relationship Between Dwell Time and Sampling Rate

3. Detector Settings

Dwell time for each MRM transition (ms)

Settling Time (ms) Pause between mass ranges (ms)

Total sampling time for 5 transitions (ms)

Sampling Rate (pts/s)

10 0.1 3 (10x5) + (0.1) + (3x4) = 62.1 ms/pt

16

Page 18: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 18www.phenomenex.com

Trouble: Broad and Low Intensity Peaks (continued)

Fixed Mass Methods (MRM, SIM, etc.)

Fixed mass methods optimization is often straightforward for maximizing performance of Kinetex• ® core-shell columns.

SOLUTION

For methods with 5 or more analytes being monitored, it is often difficult for the mass spectrometer to analyze all the analytes in •parallel with a high sampling rate.

SOLUTION

Signal Sensitivity versus Analyte Quantitation•

SOLUTION

Table 3: Relationship Between Dwell Time and Sampling Rate for an API 3000™ and API 4000™ MS

3. Detector Settings

Optimize the dwell time to obtain 20 or greater points per second for methods with only a few analytes (Table 3).

Time programmed acquisition method. Such methods can maximize the data acquisition rate for the limited time during analyte elution and make high performance possible for methods with large numbers of analytes.

Sometimes one must balance between maximizing analyte sensitivity with increasing dwell time, and improving peak shape and quantitation with higher sampling rate. An example of this balance between sensitivity and resolution is shown in Figure 6.

Dwell Time (ms) Sampling Rate (SR = pts/s)

Peak Width (PW) in sec (5th Peak @ Base)

# Points across the peak (PW x SR)

10 16 11.6 s 186

17 10 11.8 s 118

50 4 19.8 s 79

100 2 22.7 s 45

AP

I 3

00

0

Dwell Time (ms) Sampling Rate (SR = pts/s)

Peak Width (PW) in sec (5th Peak @ Base)

# Points across the peak (PW x SR)

10 16 7.6 s 122

17 10 8.8 s 88

50 4 15.1 s 60

100 2 19.1 s 38

AP

I 4

00

0

Page 19: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 19www.phenomenex.com

Trouble: Broad and Low Intensity Peaks (continued)

Figure 6: Balancing Between Sensitivity and Resolution by Adjusting MS Dwell Time

3. Detector Settings

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 min0.0

5000.0

1.0e4

1.5e4

2.0e4

0.84

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.00.0

5000.0

1.0e4

1.5e4

0.84

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.00.0

5000.0

1.0e4

1.5e4

0.86

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.00.0

5000.0

1.0e4

1.5e4

0.85

19169

min

min

min

S/N 54S = 1.6e4N = 300

S/N 88S = 1.6e4N = 180

S/N 207S = 1.2e4N = 60

S/N 388S = 1.2e4N = 30

2.2e4 cps DT = 10 msRS @ 50 % 4,5 = 2.47PWbase 5 = 11.6 s

DT = 17 msRS @ 50 % 4,5 = 2.35PWbase 5 = 11.8 s

DT = 50 msRS @ 50 % 4,5 = 2.17PWbase 5 = 19.8 s

DT = 100 msRS @ 50 % 4,5 = 2.10PWbase 5 = 22.7 s

Ap

p ID

1916

9

Decreased dwell time reduces signal-to-noise ratio

and total signal

As sampling rate increases by lower dwell times, peak

resolution increases

Page 20: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

Page 20www.phenomenex.com

3. Detector Settings

Trouble: Broad and Low Intensity Peaks (continued)

Full scan MS or MS/MS methods

Depending on the mass spectrometer being used, scan rates are based on several factors: dwell time •(time acquiring a specific m/z), mass resolution (mass difference between neighboring m/z), and scan window (mass range of m/z from low to high mass).

SOLUTION

Reducing Dwell Time and Increasing MS Sampling Rate: Pros and Cons

Benefits

Improved resolution of closely eluting peaks•

Improved quantitation of analytes•

Drawbacks

Reduced MS sensitivity•

Reduced signal-to-noise ratio•

Narrower MS scan range•

Reduced isotopic resolution•

Depending on the analytes being scanned, all three parameters can be modified to increase the sampling rate. Dwell time can be reduced if sensitivity is not a factor in the method, mass resolu-tion can be decreased if isotopic resolution is not necessary, and the scan range can be narrowed if some information is known about the analytes being analyzed.

Page 21: SYSTEM OPTIMIZATION MANUAL - Microsoft...SYSTEM OPTIMIZATION MANUAL Page 1 Get the Most out of your Kinetex® Core-Shell Column Kinetex core-shell media can generate UHPLC-type performance

SYSTEM OPTIMIZATION MANUAL

Page 21www.phenomenex.com

t: f:

Australia02-9428-6444 02-9428-6445 [email protected]

t: f:

Austria 01-319-1301 01-319-1300 [email protected]

t: f:

Belgium+31 (0)30-2418700 +31 (0)30-2383749 [email protected]

t: f:

Canada(800) 543-3681 (310) 328-7768 [email protected]

t: f:

Denmark4824 8048 4810 6265 [email protected]

t: f:

Finland(09)4789 0063 +45 4810 6265 [email protected]

t: f:

France01 30 09 21 10 01 30 09 21 11 [email protected]

t: f:

Germany 06021-58830-0 06021-58830-11 [email protected]

t: f:

Ireland 01 247 5405 +44 1625-501796 [email protected]

t: f:

Italy 051 6327511 051 6327555 [email protected]

t: f:

Luxembourg+31 (0)30-2418700 +31 (0)30-2383749 [email protected]

t: f:

Mexico(55) 5018 3791 (310) 328-7768 [email protected]

t: f:

Netherlands030-2418700 030-2383749 [email protected]

t: f:

New Zealand09-4780951 09-4780952 [email protected]

t: f:

Norway 810 02 005 +45 4810 6265 [email protected]

t: f:

Puerto Rico (800) 541-HPLC (310) 328-7768 [email protected]

t: f:

United Kingdom 01625-501367 01625-501796 [email protected]

t: f:

All other countries: Corporate Office USA (310) 212-0555 (310) 328-7768 [email protected]

www.phenomenex.comPhenomenex products are available worldwide. For the distributor in your country, contact Phenomenex USA, International Department at [email protected]

Terms and Conditions Subject to Phenomenex Standard Terms and Conditions which may be viewed at http://www.phenomenex.com/TermsAndConditions.

Trademarks Kinetex is a registered trademark of Phenomenex in the U.S., EU, and other jurisdictions. Sure-Lok is a trademark of Pheonomenex, Inc. PEEK is a registered trademark of ICI Americas, Inc. Agilent is a registered trademark of Agilent Technologies. QTRAP is a registed trademark, and API 3000 and API 4000 are trademarks of Applied Biosystems/MDS SCIEX, a joint venture between Applera Corporation and MDS Inc.