Design Flexibility with a Vast Range of Applications
The PSD/4 is a popular choice for customers looking for a high precision syringe pump at an economical price. From the semi-conductor to the medical device industries, this pump has proven to be a reliable addition to numerous continued projects with top automation manufacturers. The space-saving design uses 30 mm half-height syringes and performs all standard liquid handling functions, including dispensing, serial dispensing, and diluting.
PSD/4 Half-Height Syringe Pump Specification Sheet
Side
PSD/4 Product Drawings and Dimensions
Back Top
Front
1.400" (35.56 mm)
2 X Mounting Holes Threads: #6 – 32 UNC–2B Hole Depth: 0.40 ± 0.03
0.200" (5.08 mm)
3.170" (80.518 mm)
4.17" (105.92 mm)
5.000" (127 mm)
0.410" (10.414 mm)
1.350" (34.29 mm)
1.750" (44.45 mm)
0.603" (15.316 mm)
0.200" (5.08 mm)
0.183" (4.648 mm)
1.384" (35.154 mm)
4 X Mounting Holes (Top and Bottom) Threads: M3 X 0.5 – 6H Hole Depth: 0.25 minimum
4.20" (106.68 mm)
4.400" (111.76 mm)
0.300" (7.62 mm)
1.23" (31.24 mm)
0.200" (5.08 mm)0.603" (15.316 mm)
0.19" (4.83 mm)
4 X Mounting Holes (Both Sides) Threads: M3 X 0.5 – 6H Hole Depth: 0.25 minimum
2 X Mounting Holes Threads: M3 X 0.5 – 6H Hole Depth: 0.25 minimum
2 X Jack Screws Threads: #4 – 40 UNC-2B
Communication and Power Connection Type: Male DB15
4.40" (111.76 mm)
1.84"
0.87" (22.10 mm)
0.55" (13.97 mm)
0.30" (7.62 mm)
Fluid Logic: 8-5
Valve Description: Distribution Flow Path Eight ports, plus syringe port
Ceramic Part Number: 7993-01, 59943-01
Fluid Logic: 3-5
Valve Description: Distribution Flow Path Three ports, plus syringe port
Ceramic Part Number: 7991-01
Fluid Logic: 4-5
Valve Description: Distribution Flow Path Four ports, plus syringe port
Ceramic Part Number: 7992-01
Fluid Logic: Y-Block
Valve Description: Static non-switching Y-block
PTFE/CTFE Part Number: 9537-01
30 mm Syringes for PSD/4
Volume UHMW-PE P/N PTFE P/N Pressure Rating
12.5 µL N/A 5495-10
1000 psig (68.9 bar)
25 µL 8250-15 5495-15
50 µL 8250-20 5495-20
100 µL 8250-22 5495-22
125 µL 8250-25 5495-25
250 µL 8250-30 5495-30 500 psig (34.5 bar)
500 µL 8250-35 5495-35
200 psig (13.8 bar)
1 mL 8250-37 5495-38
1.25 mL 8250-40 5495-40
2.5 mL 8250-45** 5495-45**
5 mL 8250-50** 5495-50**
12.5 mL N/A 5495-55*
* Not compatible with 57252-01, 58889-01, 8778-01, 7992-01, and 7993-01.** Not compatible with 7992-01 and 7993-01. Use valve 59943-01 as a
substitute for 7993-01.
Valves for PSD/4
* Valve 57252-01 has a 0.059” (1.499 mm) fluid path diameter. All tubing connections are compatible with ¼" x 28 flanged or flangeless fittings. All valves have a pressure rating of 100 psi. Higher pressure ratings are available upon request. All valve housings are aluminum. Stainless steel, CTFE, and PEEK housings are available as a custom option.
Fluid Logic: Y-Valve
Valve Description: “Y” Flow Path Two ports, plus syringe port
Ceramic Part Number: 8778-01
PTFE/CTFE Part Number: 57252-01*
Fluid Logic: 4-2
Valve Description: 90° Flow Path Three ports, plus syringe port
Ceramic Part Number: 9234-01
PSD/4 Precision Syringe Pump
Description Part Number
PSD/4 Precision Syringe Pump
54848-01
PSD/4 Starter Kit Containing Software CD, Demo Communication Cable, Serial-to-USB Adapter, and Power Supply
64316-04
0.300" (7.62 mm)
Fluid Logic: 6-5
Valve Description: Distribution Flow Path Six ports, plus syringe port
Ceramic Part Number: 9998-01
Fluid Logic: 3-3
Valve Description: “T” Flow Path Two ports, plus syringe port
Ceramic Part Number: 8063-01
PTFE/CTFE Part Number: 58889-01
PSD/4 SpecificationsGeneral Specifications
Accuracy ± 1% at 100% stroke
Precision ≤ 0.05% at 100% stroke
Fluid Path Borosilicate glass, PTFE, PFA, CTFE, ETFE, UHMW-PE, or ceramic
Weight 2.6 lbs (1.2 kg)
RoHS Compliant Yes
Linear Force Capability 20 lbf. (9.07 kgf.)
Power Requirements
Supply Voltage 24 VDC
Average Current 850 mA maximum
Syringe and Syringe Drive
Syringe Volumes 12.5 µL to 12.5 mL
Syringe MaterialsGlass barrel, PTFE or PFA insert, stainless steel or aluminum plunger with PTFE coating, PTFE or UHMW-PE plunger tip
Resolution Selectable 3,000 steps (standard) / 24,000 steps (high)
Syringe Drive Mechanism Stepper motor driven lead screw and optical encoder
Stroke Length 30 mm
Syringe Speeds 1 second to 50 minutes per stroke
Valve and Valve Drive
Valve Drive Speed 250 ms per 120° rotation
Valve Drive Mechanism Stepper motor with optical encoder feedback
Valve Fittings ¼" – 28
Valve Materials CTFE, PTFE, ETFE, or ceramic
Nominal Fluid Path Diameter All valves: 0.060” (1.524 mm) unless otherwise noted
Communication
Type RS-232, RS-485, or CAN
Protocols Terminal and standard
Baud Rate 9600 or 38,400 (RS-232 or RS-485); 100,000 or 125,000 (CAN)
Data Bits 8
Parity None
Stop Bit 1, Half duplex
Daisy Chain Length Up to 16 individual pumps
Programmable CapabilitiesRamps, cutoff velocity, backlash compensation, syringe speeds, loops, on-the-fly speed changes, terminate moves and delays, error detection, valve rotation selection, enhanced “h” Factor capabilities including valve rotation clockwise and counter-clockwise, 15 location method storage
Environmental Operating and Storage Range
Operating Temperature 59 °F (15 °C) to 104 °F (40 °C)
Operating Humidity 20% to 95% relative humidity, non-condensing
Storage Temperature -4 °F (-20 °C) to 149 °F (65 °C)
Storage Humidity 20% to 95% relative humidity, non-condensing
For more information or to place your order, call Hamilton at 1.800.648.5950 or visit www.hamiltoncompany.com.
©2017 Hamilton Company. All rights reserved. All other trademarks are owned and/or registered by Hamilton Company in the U.S. and/or other countries.
Lit. No. L60034 Rev. F — 05/2017
Hamilton Americas & Pacific RimHamilton Company Inc.4970 Energy WayReno, Nevada 89502 USATel: +1-775-858-3000Fax: [email protected]
Hamilton Europe, Asia & AfricaHamilton Bonaduz A.G.Via Crusch 8CH-7402 Bonaduz, SwitzerlandTel: +41-58-610-10-10Fax: [email protected]
To find a representative in your area, please visit hamiltoncompany.com/contacts.
Web: www.hamiltoncompany.comUSA: 800-648-5950Europe: +41-58-610-10-10
A Compact, Full-Height Pump, for Precision Dispensing of Small to Large Volumes
The PSD/6 is the newest full-height syringe pump to join the Hamilton Precision Syringe Drive (PSD) family. The PSD/6 can be fitted with a variety of valve and syringe combinations and performs all standard liquid handling functions, including dispensing, serial dispensing, and diluting. These full-height pumps are easy to integrate, self-contained syringe drive modules. They can be used as single units or in a daisy chain of up to 16 pumps.
The PSD/6 is a reliable turnkey system to meet the requirements of the most demanding applications. Ensure that your development projects stay on time and on budget with proven technology that is engineered to exceed expectations.
PSD/6 Full-Height Syringe PumpSpecification Sheet
Front
Back
Side
PSD/6 Product Drawings and Dimensions
Bottom
.243" (6.17 mm)
.500"(12.7 mm)
1.264"(32.11 mm)
1.750"(44.45 mm)
1.000"(25.4 mm)
4 X Mounting HolesThreads: M3x0.5-6H
Hole Depth: 0.20” minimum
5.000"(127 mm)
8.990"(228.35 mm)
8.155"(207.14 mm)
.55" (13.9 mm)
1.937" (49.2 mm)
1.98" (50.3 mm)
2 X Jack Screws Threads: #4-40 UNC-2B
.184” (46.8 mm)
Communication and Power Connection Type: Male DB15
4.400" (111.76 mm)
.87" (22.2 mm)
4 X Mounting Holes Threads: M3x0.5-6H Hole Depth: 0.25” minimum
.30" (7.6 mm)
4 X Mounting Holes Threads: M3x0.5-6H Hole Depth: 0.20” minimum.200" (5.08 mm)
.603" (15.32 mm)
.183" (4.65 mm)
1.384” (35.15 mm)
4 X Mounting HolesThreads: M3x0.5-6HHole Depth: 0.20” minimum
8.390"(213.11 mm)
.300"(7.62 mm)
.603"(15.32 mm)
.20"(5.0 mm)
1.23"(31.3 mm)
4.19" (106.4 mm)
.200" (5.08 mm)
PSD/6 Precision Syringe Pump
Description Part Number
PSD/6 Precision Syringes Pump 63133-01
PSD Starter Kit Containing Software CD, Demo Communication Cable, Serial-to-USB Adapter, and Power Supply
64316-04
Demo Communication Cable 57025-01
Power Supply 4833-01
Syringes for PSD/6
Volume60 mm Long-Life Syringes P/N
Standard PTFE Tip Syringes P/N
Pressure Rating
25 µL N/A 80222
1000 psig (68.9 bar)50 µL 8300-15 80922
100 µL 8300-20 81022
250 µL 8300-25 81122500 psig (34.5 bar)
500 µL 8300-30 81222
1 mL 8300-35 81320
200 psig (13.8 bar)2.5 mL 8300-40 81420
5 mL 8300-45 81520
10 mL 8300-50 81620*
25 mL N/A 82521**100 psig (6.9 bar)
50 mL N/A 85021**
* Directly compatible with valve 7992-01, 7993-01, 9998-01, and 59943-01. Use adapter (P/N 209651) for compatibility with all remaining valves listed below.
** Directly compatible with valve 9998-01 and 59943-01. Use adapter (P/N 209651) for compatibility with all remaining valves except 7992-01 and 7993-01.
Fluid Logic: 8-5
Valve Description: Distribution Flow Path Eight ports, plus syringe port
Ceramic Part Number: 7993-01, 59943-01
Fluid Logic: 3-5
Valve Description: Distribution Flow Path Three ports, plus syringe port
Ceramic Part Number: 7991-01
Fluid Logic: 4-5
Valve Description: Distribution Flow Path Four ports, plus syringe port
Ceramic Part Number: 7992-01
Fluid Logic: Y-Block
Valve Description: Static non-switching Y-block
PTFE/CTFE Part Number: 9537-01
Valves for PSD/6
* Valve 57252-01 has a 0.059” (1.499 mm) fluid path diameter. All tubing connections are compatible with ¼" x 28 flanged or flangeless fittings. All valves have a pressure rating of 100 psi. Higher pressure ratings are available upon request. All valve housings are aluminum. Stainless steel, CTFE, and PEEK housings are available as a custom option.
Fluid Logic: Y-Valve
Valve Description: “Y” Flow Path Two ports, plus syringe port
Ceramic Part Number: 8778-01
PTFE/CTFE Part Number: 57252-01*
Fluid Logic: 4-2
Valve Description: 90° Flow Path Three ports, plus syringe port
Ceramic Part Number: 9234-01
Fluid Logic: 6-5
Valve Description: Distribution Flow Path Six ports, plus syringe port
Ceramic Part Number: 9998-01
Fluid Logic: 3-3
Valve Description: “T” Flow Path Two ports, plus syringe port
Ceramic Part Number: 8063-01
PTFE/CTFE Part Number: 58889-01
PSD/6 SpecificationsGeneral Specifications
Accuracy ± 1% at 100% stroke
Precision ≤ 0.05% at 100% stroke
Fluid Path Borosilicate glass, PTFE, PFA, CTFE, UHMW-PE, or ceramic
Weight 3.65 lbs (1.65 kg)
RoHS Compliant Yes
Linear Force Capability 20 lbf (9.07 kgf)
Power Requirements
Supply Voltage 24 VDC
Average Current 850 mA maximum
Syringe and Syringe Drive
Syringe Volumes 25 µL to 50 mL
Syringe MaterialsGlass barrel, PTFE, PCTFE, or PFA insert, stainless steel or aluminum plunger with PTFE coating, PTFE or UHMW-PE plunger tip
Resolution Selectable 6,000 steps (standard) / 48,000 steps (high)
Syringe Drive Mechanism Stepper Motor driven lead screw and optical encoder
Stroke Length 60 mm
Syringe Speeds 2 seconds to 100 minutes per stroke
Valve and Valve Drive
Valve Drive Speed 250 ms per 120° rotation
Valve Drive Mechanism Stepper motor with optical encoder feedback
Valve Fittings ¼" - 28
Valve Materials CTFE, PTFE, ETFE, or ceramic
Nominal Fluid Path Diameter All valves: 0.060” (1.524 mm) unless otherwise noted
Communication
Type RS-232, RS-485, or CAN
Protocols Terminal or Standard
Baud Rate 9,600 or 38,400 (RS-232 or RS-485); 100,000 or 125,000 (CAN)
Data Bits 8
Parity None
Stop Bit 1, Half Duplex
Daisy Chain Length Up to 16 individual pumps
Programmable Capabilities
Ramps, cutoff velocity, backlash compensation, syringe speeds, loops, on-the-fly speed changes, terminate moves and delays, error detection, valve rotation selection, enhanced “h” Factor capabilities including valve rotation clockwise and counter-clockwise, 15 location method storage
Environmental Operating and Storage Range
Operating Temperature 59 °F (15 °C) to 104 °F (40°C)
Operating Humidity 20% to 95% relative humidity, non-condensing
Storage Temperatures -4 °F (-20 °C) to 149 °F (65 °C)
Storage Humidity 20% to 95% relative humidity, non-condensing
For more information or to place your order, call Hamilton at 1.800.648.5950 or visit www.hamiltoncompany.com.
©2017 Hamilton Company. All rights reserved. All other trademarks are owned and/or registered by Hamilton Company in the U.S. and/or other countries.
Lit. No. L60038 Rev. E — 07/2017
Hamilton Americas & Pacific RimHamilton Company Inc.4970 Energy WayReno, Nevada 89502 USATel: +1-775-858-3000Fax: [email protected]
To find a representative in your area, please visit hamiltoncompany.com/contacts.
Web: www.hamiltoncompany.comUSA: 800-648-5950Europe: +41-58-610-10-10
Hamilton Europe, Asia & AfricaHamilton Bonaduz A.G.Via Crusch 8CH-7402 Bonaduz, SwitzerlandTel: +41-58-610-10-10Fax: [email protected]
Hamilton Modular Valve Positioners are self-contained, bi-directional valve actuators used for fluid selection and redirection. There are two models to choose from, a standard torque valve positioner and a high torque valve positioner. A built-in DC stepper motor ensures accurate valve port alignment and location for optimum fluid delivery. This module can be fitted with a choice of valves to meet even the most demanding application.
MVP/4 Valve PositionerSpecification Sheet
MVP/4 Valve Positioner Product Drawings and Dimensions
4X 6-32 UNC–2BMOUNTING HOLES
.875”(22.23 mm)
2X 1.616”(41.05 mm)
2X 1.157”(29.39 mm)
2X .430”(10.92 mm)
2X .430”(10.92 mm)
3.410”(86.61 mm)
1.425”(36.20 mm)
2X M3 X 0.5–6HMOUNTING HOLES
BOTH SIDES
5.240”(133.10 mm)
1.750”(44.45 mm)
.500”(12.70 mm)
.250”(6.35 mm)
.650”(16.51 mm)
1.380”(35.05 mm)
94747-01 94747-02
2X M3 X 0.5–6HMOUNTING HOLES
2x .260”(6.60 mm)
1.384” (35.15 mm)
.183”(4.65 mm)
4.400”(111.76 mm)
2X M3 X 0.5–6HMOUNTING HOLES
0.300”(7.62 mm)
2X .875”(22.23 mm)
Suggested Mounting Holes for Hamilton MVP/4 Module
Valves for the MVP/4 Valve Positioner
Valve Information
Valve Model Valve Description
HV PTFE, Small Body, MVP Valve
HVXM PTFE, Large Body, MVP Valve
HVC Ceramic, Small Body, MVP Valve
HVCX Ceramic, Large Body, MVP Valve
Fluid Logic: Y
Valve Description: Distribution Flow Path Two ports
Ceramic HVC P/N: 97464-01
PTFE HV P/N: 97465-01
Fluid Logic: 3-2T
Valve Description: 90° Flow Path Three ports
Ceramic HVC P/N: 97467-01
PTFE HV P/N: 97466-01
Fluid Logic: 3-5
Valve Description: Distribution Flow Path Three ports
Ceramic HVC P/N: 97468-01
Fluid Logic: 4-2
Valve Description: 90° Flow Path Four ports
Ceramic HVC P/N: 97469-01
Fluid Logic: 4-3T
Valve Description: “T” Flow Path Four ports
Ceramic HVCX P/N: 6600338-01
Fluid Logic: 4-4
Valve Description: Loop Flow Path Four ports
PTFE HVXM P/N: 98642-01
Fluid Logic: 4-5
Valve Description: Distribution Flow Path Four ports
Ceramic HVCX P/N: 97470-01
Fluid Logic: 6-5
Valve Description: Distribution Flow Path Six ports
Ceramic HVCX P/N: *97471-01
Fluid Logic: 6-6
Valve Description: Loop Flow Path Six ports
PTFE HVXM P/N: *98643-01
Fluid Logic: 8-5
Valve Description: Distribution Flow Path Eight ports
Ceramic HVCX P/N: *97472-01
Fluid Logic: 8-7
Valve Description: Loop Flow Path Eight ports
PTFE HVXM P/N:*6600295-01
Ordering Information
P/N Description
94747-01 MVP/4 Valve Positioner
94747-02 MVP/4 High Torque Valve Positioner
64316-04 Starter Demo Kit*
*Kit includes: software CD, demo communication cable, serial-to-USB adapter, and power supply
*Hamilton recommends using the High Torque MVP/4 (P/N 94747-02) to drive these valves if using a salty solution or lock up is a concern.
MVP/4 Valve Positioner Specifications
General Specifications
Fluid Path PTFE, CTFE, or ceramic
Weight1.75 lbs (0.79 kg) (P/N 94747-01) 2.25 lbs (1.02 kg) (P/N 94747-02)
Dimensions Height: 3.41 inches (86.6 mm), Width: 1.75 inches (44.5 mm), Depth: 5.49 inches (139.4 mm)
RoHS Compliant Yes
Power Requirements
Supply Voltage 24 VDC
Average Current 1.7 A maximum
Valve and Valve Drive
Valve Drive Speed 250 ms per 120° rotation
Valve Drive Stepper motor with optical encoder
Valve Torque Capability25 in-oz (0.18 N-m) (P/N 94747-01) 75 in-oz (0.53 N-m) (P/N 94747-02)
Rotational Speed (RPM) 52.5–115.2
Valve Fittings ¼”–28
Valve Materials PTFE, CTFE, or ceramic
Nominal Fluid Path Diameter All valves: 0.060 inches (1.524 mm) unless otherwise noted
Communication
Type RS-232, RS-485, or CAN
Protocols Terminal or Standard
Baud Rate9,600 or 38,400 (RS-232, RS-485) 100,000 or 125,000 (CAN)
Data Bits 8
Parity None
Stop Bit 1, Half duplex
Daisy Chain Length Up to 16 individual MVP units
Programmable CapabilitiesLoops, terminate moves, and delays, error detection, valve rotation selection, enhanced “h” Factor capabilities including valve rotation clockwise and counter-clockwise, 15 location method storage
Environmental Operating and Storage Range
Operating Temperature 59°F (15°C)–104°F (40°C)
Operating Humidity 20–95% relative humidity, non-condensing
Storage Temperature -4°F (-20°C)–149°F (65°C)
Storage Humidity 20–95% relative humidity, non-condensing
©2018 Hamilton Company. All rights reserved. All other trademarks are owned and/or registered by Hamilton Company in the U.S. and/or other countries.
Lit. No. L60039 Rev C — 10/2018
Hamilton Americas & Pacific RimHamilton Company Inc.4970 Energy WayReno, Nevada 89502 USATel: +1-775-858-3000Fax: [email protected]
Hamilton Europe, Asia & AfricaHamilton Central Europe S.R.L.str. Hamilton no. 2-4307210 Giarmata, RomaniaTel: +40-356-635-055Fax: [email protected]
To find a representative in your area, please visit hamiltoncompany.com/contacts.
Web: www.hamiltoncompany.comUSA: 800-648-5950Europe: +40-356-635-055