Taking Aim at Interference (Without Shooting Yourself in the Foot)
May 2020
3
IntroductionPopularity of LC-MS/MS-based methods for
clinical testing continues to increase
One of the major reasons:
superior analytical specificity
Despite that, these methods may still suffer from interference Affecting method accuracy and precision
Negatively impacting patient care
4
Aim of this presentation Sources of guidelines for interference testing and monitoring
What is analytical interference?
Where does it come from?
How do we define acceptable interference levels?
How do we test for interference in LC-MS/MS?
When do we test for interference?
Can we use internal standard to mitigate interference?
How do we monitor for interference in routine testing?
5
Guidelines for interference testing
CLSI – Clinical and Laboratory Standards Institute EP7-A2: Interference testing in clinical chemistry EP14-A2: Evaluation of matrix effects C62-A: Liquid chromatography-mass spectrometry methods
FDA – Food and Drug Administration Bioanalytical Method Validation, Guidance for Industry
SWGTOX – Scientific Working Group for Forensic Toxicology SWGTOX Doc 003: Standard Practices for Method Validation in Forensic
Toxicology
WADA – World Anti-Doping Agency WADA Technical Document – TD2015IDCR: Minimum criteria for LC-MS
confirmation of the identity of analytes for doping control purposes
European Medicines Agency Guidelines on bioanalytical method validation
6
What is analytical interference?
Interference =
the effect of a substance, identified or not, that causes the measured concentration of an analyte to differ from its true value.
Interferent or Interfering Substance =
the substance causing interference
Reference: CLSI document EP14-A2. Evaluation of matrix effects; Approved guideline – 2nd edition. Wayne (PA), 2005.
7
What is analytical interference?
Interference May appear as partially or
completely co-eluting peaks in the analyte/internal standard mass chromatograms
May be invisible to the naked eye –a matrix effect Interferent alters the efficiency of
analyte/internal standard ions reaching the MS detector
v-forvictory.blogspot.com
Time
Inte
nsity
8
Where does interference come from?
Interfering substances May come from many different sources
May be introduced at any time before or during the testing workflow
9
Compounds related to patient treatment medications parenteral nutrition plasma expanders
Metabolites produced in normal & pathological conditions
Substances ingested by patients alcohol drugs of abuse nutritional supplements food
paleoleap.comalamy.com
Examples of interfering substances
slideshare.netcapecodhealth.org
10
Substances added during sample preparation anticoagulants – heparin & EDTA salts preservatives stabilizers
Contamination during sample handling hand lotion, soap serum separators collection tube stoppers leachables from plastic consumables
Interference arising from the sample matrix hemolysis – hemoglobin and RBC contents icterus – bilirubin lipemia – lipoproteins
Examples of interfering substances
thermofisher.com
seadragonbotanicals.comhealthmanagement.org
11
How do we define acceptable interference levels?
Reference: CLSI document EP7-A2. Interference testing in clinical chemistry; Approved guideline – 2nd edition. Wayne (PA), 2005.
Acceptability criteria must be decided prior to conducting evaluation to ensure objectivity
Key questions: 1. How large a discrepancy is considered clinically significant?
2. What are the allowable analytical error limits? Accuracy requirements:
Have been proposed for some analytes (total allowable error) Can be established based on physiological variability (inter- & intrapersonal) Can be derived from clinical experience (consensus of clinical experts) Can be based on analytical variability (long-term imprecision)
12
Why do we test for interference?
Interference affects: Method accuracy Method precision Quality and validity of reported patient results
Assessing susceptibility to analytical interference– very important part of LC-MS/MS
method development and validation
13
How do we test for interference in LC-MS/MS?
Interference testing falls into two categories:
1) Direct testing of the effect of specific substances on analyte concentration
2) Evaluation of unidentified interferents arising from sample matrix and anything added to it
14
1. Testing for specific interference
clker.com123RF Stock Photos Online Science Mall
+
LC-MS/MS analysis
Patient specimens w/ analyte of interest Sample pool
Interferent 1Interferent 2
Interferent 3
Solvent
Analyze both test and control samples in the same manner as patient specimens with adequate replication within one analytical run
Test samples Control sample
15
1. Testing for specific interference Evaluate interference as a relative bias of the target analyte
concentration in test vs control sample:
𝐁𝐁𝐁𝐁𝐁𝐁𝐁𝐁(%) = 𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀 −[𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐜𝐜𝐜𝐜𝐀𝐀𝐀𝐀𝐜𝐜𝐜𝐜𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀][𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐜𝐜𝐜𝐜𝐀𝐀𝐀𝐀𝐜𝐜𝐜𝐜𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀]
×100%
Analyte – Test at 2 (or more) concentrations
Potential interferent – Spike at the highest concentration expected in patient specimens When clinically significant interference present =>
evaluate further at lower concentrations
Reference:CLSI EP7-A2: Interference testing in clinical chemistry – guidelines for test concentrations for many analytes and potential interferents
16
1. Testing for specific interferenceExample calculations table
Measured Concentration %Bias from Control Mean
Potential Interferent % Spike Volume
Intf. Conc. (µmol/L) Repl #1 Repl #2 Repl #3 Mean Repl #1 Repl #2 Repl #3 Mean
Control_Methanol Mean 5 52.1 52.5 52.2 52.3 -0.3% 0.5% -0.1%Caffeine 5 257 52.5 52.2 52.2 52.3 0.4% -0.1% -0.1% 0.1%Theobromine 5 278 52.1 51.9 52.1 52.0 -0.2% -0.7% -0.3% -0.4%Acetaminophen 5 331 52.2 51.6 48.3 50.7 0.0% -1.2% -7.7% -3.0%Aspirin (as salicylate) 5 362 52.3 51.6 52.1 52.0 0.0% -1.2% -0.4% -0.5%Diphenhydramine 5 196 52.0 52.0 51.8 51.9 -0.6% -0.6% -1.0% -0.7%Control_0.1 M NaOH 10 45.6 45.9 46.0 45.9 -0.5% 0.2% 0.4%Uric acid 10 1487 47.6 47.2 47.7 47.5 3.9% 3.0% 4.0% 3.6%Icterus - bilirubin 10 248 46.6 46.7 46.5 46.6 1.6% 1.7% 1.3% 1.6%Control_Normal Saline 10 47.4 47.1 47.4 47.3 0.3% -0.5% 0.2%Hemolysis - hemoglobin 10 166 41.6 41.4 41.5 41.5 -12.1% -12.4% -12.4% -12.3%Lipemia - triglycerides 10 N/A 48.6 48.3 48.0 48.3 2.7% 2.1% 1.5% 2.1%Ibuprofen 10 485 47.2 47.2 47.2 47.2 -0.3% -0.2% -0.3% -0.2%
Note: Think about the “mechanism” of the interference.Adding the interferent solution may not always give you a clear pictureof how the interferent acts in a biological system.
17
1. Testing for specific interference
Advantages
Ability to define acceptable/unacceptable sample collection conditions
Ability to set limitations for sample abnormalities
Ability to provide guidelines for patient preparation(medications, supplements, and foods to avoid prior to sample collection)
To obtain a valid test resultTo reduce the need for repeat specimen collection and analysis
Disadvantages
Laboriousness of testing a large number of substances
No practical interference study can identify all potential interferents
18
2. Testing for unidentified interference
LC-MS/MS allows testing for interference that cannot be anticipated or identified beforehand (matrix effects)
Interference arising from sample matrix can cause: Signal enhancementSignal suppression
Evaluation of matrix effects:A. Qualitative post-column infusion studyB. Quantitative matrix effects study
19
A. Qualitative post-column infusion study Analyte solution infused into the LC column effluent
While analyzing blank sample matrix
When blank sample matrix not available/not representative of patient specimens IS may be infused while analyzing a patient specimen
HPLC
Syringe Pump
Mass Spectrometer
LC column Tee
Sample
Analyte solution(IS solution)
Patient specimen
Void volume
Inte
nsity
(cps
)
Time (min)
Infusion trace
20
A. Qualitative post-column infusion study
Signal suppression
Ion suppression (enhancement) evaluated as the presence of negative (positive) peaks in a steady signal trace of the infused analyte or IS
Advantage
Allows for visualization of the position and width of matrix effects regions
Useful in optimizing separation conditions
Calibrator
Void volume
Inte
nsity
(cps
)
No signal suppression
Time (min)
Infusion trace
21
B. Quantitative matrix effects study
LC-MS/MS analysis
Extracted samples: Various blank matrices Patient specimens
with no/low analyte concentrations
Analyte Analyte Analyte Analyte Analyte
Analyze both test and control samples
in the same manner as patient specimens with adequate replication within one analytical run
#1 #2 #3 #n…Test samples
+Solvent
Control sample
22
B. Quantitative matrix effects study
Analyte signal in test sample expressed as % of control:
𝐌𝐌𝐁𝐁𝐀𝐀𝐜𝐜𝐁𝐁𝐌𝐌 𝐀𝐀𝐞𝐞𝐞𝐞𝐀𝐀𝐜𝐜𝐀𝐀(%) = 𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀[𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐜𝐜𝐜𝐜𝐀𝐀𝐀𝐀𝐜𝐜𝐜𝐜𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀]
×100%
Matrix effect: <100% indicates signal suppression >100% indicates signal enhancement
Alternatively, as %Bias from control:
𝐌𝐌𝐁𝐁𝐀𝐀𝐜𝐜𝐁𝐁𝐌𝐌 𝐀𝐀𝐞𝐞𝐞𝐞𝐀𝐀𝐜𝐜𝐀𝐀(%) = 𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀 −[𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐜𝐜𝐜𝐜𝐀𝐀𝐀𝐀𝐜𝐜𝐜𝐜𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀][𝐀𝐀𝐀𝐀𝐁𝐁𝐀𝐀𝐀𝐀𝐀𝐀𝐀𝐀 𝐁𝐁𝐀𝐀 𝐜𝐜𝐜𝐜𝐀𝐀𝐀𝐀𝐜𝐜𝐜𝐜𝐀𝐀 𝐁𝐁𝐁𝐁𝐬𝐬𝐬𝐬𝐀𝐀𝐀𝐀]
×100%
Matrix effect: <0% indicates signal suppression >0% indicates signal enhancement
23
B. Quantitative matrix effects study
The extent of a matrix effect can be calculated as: Non-normalized (using peak areas) or Normalized to internal standard (using concentrations)
Both matrix effect values provide valuable information: Non-normalized – actual magnitude of ion suppression/enhancement Normalized values – how well the IS compensates for the matrix effect
These experiments should be performed: At two analyte concentrations expected in the patient population With several native matrix sources, such as different patient specimens or
different vendor sources
Useful especially when: Qualitative post-column infusion experiment cannot be performed
(e.g. LC flow rate high => too high a backpressure for syringe pump) Testing matrices used for calibrator or QC preparation
24
B. Quantitative matrix effects studyControl (solvent) spiked
with analyteTest sample (Plasma 1) spiked
with analyteTest sample (Plasma 1)
Baseline
Analyte Repl #1
Repl #2
Repl #3 Mean CV Repl
#1Repl #2
Repl#3 Mean CV Repl
#1Repl #2
Repl #3 Mean CV
Analyte 1 101 103 107 104 3% 113 114 112 113 1% 13 14 13 13 3%Analyte 2 98 99 96 98 2% 337 315 312 321 4% 213 225 233 224 5%Analyte 3 103 111 106 107 4% 103 102 104 103 1% 1 1 1 1 14%Analyte 4 103 107 104 104 2% 357 332 329 339 5% 256 254 254 255 0%Analyte 5 100 106 112 106 6% 98 102 101 101 2% 0 0 0 0 ####Analyte 6 94 93 99 95 3% 98 95 97 97 2% 8 8 9 8 6%Analyte 7 96 98 106 100 6% 212 214 187 204 7% 113 110 110 111 2%Analyte 8 105 103 104 104 1% 271 257 251 260 4% 139 146 146 144 3%Analyte 9 99 101 99 100 1% 280 287 276 281 2% 191 193 205 196 4%Analyte 10 102 100 104 102 2% 211 204 198 204 3% 110 115 115 113 2%Analyte 11 94 96 94 95 1% 185 182 176 181 3% 93 96 93 94 2%Analyte 12 97 100 100 99 2% 152 146 152 150 2% 53 55 56 55 3%Analyte 13 97 96 95 96 1% 190 191 183 188 2% 101 101 106 103 3%Analyte 14 97 102 101 100 3% 95 96 99 97 2% 7 7 7 7 0%Analyte 15 113 112 111 112 1% 506 483 489 493 2% 390 422 422 411 4%
Precision: %CV ≤15%10% ≤15% 16% >15%, ≤20%21% >20%
25
B. Quantitative matrix effects study
AN Spike
Target Conc
Control (Solvent) Plasma 1 Plasma 2 Plasma 3 Plasma 4 Plasma 5 Plasma 6
Mean Conc
Mean %ME
Mean Conc Mean Rec
Mean %ME
Mean Conc Mean Rec
Mean %ME
Mean Conc Mean Rec
Mean %ME
Mean Conc Mean Rec
Mean %ME
Mean Conc Mean Rec
Mean %ME
Mean Conc Mean Rec
Mean %MEAnalyte Spkd Bsln Spkd Bsln Spkd Bsln Spkd Bsln Spkd Bsln Spkd Bsln
Analyte 1 100 104 4% 113 13 100 -4% 103 8 95 -9% 103 7 97 -7% 105 7 99 -5% 108 8 100 -4% 104 6 99 -5%Analyte 2 100 98 -2% 321 224 97 -1% 422 320 101 3% 413 300 113 15% 551 414 137 39% 344 227 117 19% 374 246 129 31%Analyte 3 100 107 7% 103 1 102 -4% 100 0 100 -6% 96 0 96 -9% 101 0 100 -6% 104 0 104 -2% 103 0 103 -4%Analyte 4 100 104 4% 339 255 85 -19% 415 340 76 -27% 302 222 79 -24% 636 513 123 18% 360 272 88 -16% 463 386 78 -26%Analyte 5 100 106 6% 101 0 101 -5% 95 0 95 -10% 96 0 96 -9% 97 0 97 -8% 101 0 101 -4% 103 0 103 -2%Analyte 6 100 95 -5% 97 8 88 -7% 96 11 85 -11% 92 10 82 -14% 99 10 89 -7% 110 19 91 -4% 108 16 92 -4%Analyte 7 100 100 0% 204 111 93 -7% 179 90 88 -12% 200 105 96 -4% 235 128 106 7% 198 96 102 2% 180 88 92 -8%Analyte 8 100 104 4% 260 144 116 12% 254 148 106 2% 226 110 116 12% 280 160 121 16% 310 201 109 5% 266 141 125 21%Analyte 9 100 100 0% 281 196 85 -15% 287 201 86 -13% 191 106 85 -14% 212 120 92 -8% 303 219 84 -16% 273 184 89 -11%Analyte 10 100 102 2% 204 113 91 -11% 243 154 88 -13% 194 108 87 -15% 152 56 96 -6% 211 114 97 -5% 207 113 94 -8%Analyte 11 100 95 -5% 181 94 87 -9% 168 87 81 -14% 125 43 82 -14% 171 78 93 -2% 157 63 94 -1% 147 56 91 -4%Analyte 12 100 99 -1% 150 55 95 -5% 106 16 91 -9% 101 8 93 -6% 127 30 98 -2% 118 20 98 -1% 110 13 97 -2%Analyte 13 100 96 -4% 188 103 85 -11% 188 111 77 -20% 140 49 91 -6% 146 56 90 -7% 212 119 93 -4% 188 98 90 -6%Analyte 14 100 100 0% 97 7 90 -10% 90 3 88 -12% 90 1 89 -11% 90 2 88 -12% 95 5 90 -10% 92 2 90 -10%Analyte 15 100 112 12% 493 411 81 -27% 697 620 77 -31% 701 625 76 -32% 702 586 117 4% 647 574 73 -34% 625 541 83 -26%
20% Within ± 20% of mean analyte concentration in solvent - acceptable level of matrix effects (based on assay accuracy & precision)-25% < -20% of mean analyte concentration in solvent - signal suppression25% > 20% of mean analyte concentration in solvent - signal enhancement
Legend:
26
When do we test for interference?
Interference testing often performed as part of method validation
Waiting until method validation to perform these experiments can result in unwanted surprises
To ensure a developed LC-MS/MS method is robust and provides high quality data, test for interferences: as part of the method development process by performing the experiments outlined above
27
Post-column infusion study – very useful for: Designing LC conditions to maneuver analytes
out of suppression zones (especially in the case of dilute-and-shoot methods prone to matrix effects)
Assessing extract cleanliness when determining which sample preparation method or which conditions may best mitigate matrix effects
When do we test for interference?
28
A couple of notes:1) Interference testing and the adjustment of
method parameters may need to be an iterative process
2) Use as many patient specimens as practical to ensure that they capture the biological variability of interference
When do we test for interference?
29
Investigate interference EARLY on, and with a large variety of specimens!
mafiatoday.com
Don’t shoot yourself in the foot!
30
Interference mitigation by internal standard
Many ways to reduce interference, but no method is completely immune to it
Use of stable isotope-labeled internal standards (SIL-IS) to mitigate interference (matrix effects) common
When analyte elutes in a suppression region, compensating with a SIL-IS often deemed adequate
31
Interference mitigation by internal standard
Problems:1.Separation effect (kinetic isotope effect):
= IS and analyte not co-eluting => Differential suppression => Assay accuracy compromised
2.Severe suppression by matrix: = Coeluting IS not compensating for matrix effects
=> Analyte and IS S/N ratio drastically reduced => Assay performance compromised, especially near LLOQ
32
Problem #1: Separation effect Deuterated analogs mostly affected esp. when deuteriums in positions
impacting chromatographic retention
Differences in physicochemical properties between H and D
=> differences in interactionwith mobile & stationary phases
=> differences in retention times
Degree of separation Molecule size Number of D labels Position of D labels LC conditions
IS Analyte
Time (min)
Analyte and IS suppressed by matrix to different degrees
Quantitation errors Over-quantitating:
if IS more suppressed Under-quantitating:
if analyte more suppressed
The larger the separation, the lower the IS ability to compensate for matrix effects
Separation effect results in:Differential suppression by sample matrix
33
D&S
Infusion trace
34
Example 1: Oxycodone/oxycodone-d6 matrix effect
Scenario: Opioid panel Dilute-and-shoot Differential
suppression between analyte and IS
Solution: Selective sample
cleanup
SPE
2.0 3.0 4.0
1.0e6
2.0e6
3.0e6
4.0e6
Inte
nsity
(cps
)
1.0
NOC
OC
M
OM
OC-d6
0.0 Time (min)
Infusion trace
0.0 1.0 2.0 3.0 4.0Time (min)
1.0e6
2.0e6
3.0e6
4.0e6 NOCOC-d6
M
OM
Dilute-and-shoot
OC
Inte
nsity
(cps
)Infusion trace
Example 2: Quantitation w/ HIAA-d5 vs HIAA-13C6Post-column infusion experiment: samples with < ±10% conc. deviation
35
[HIAA] = 8.97 mg/LDeviation = 2%HIAA
HIAA-13C6
Time (min)
HIAA-d5
Infusion trace
[HIAA] = 2.27 mg/LDeviation = 0%
HIAA
HIAA-13C6
Time (min)
HIAA-d5
Infusion trace
36
[HIAA] = 1.66 mg/LDeviation = 27%
HIAA
HIAA-13C6
Time (min)
HIAA-d5[HIAA] = 4.82 mg/LDeviation = -11%
HIAAHIAA-13C6
Time (min)
HIAA-d5
HIAA over-quantitated using HIAA-d5
HIAA under-quantitated using HIAA-d5
Example 2: Quantitation w/ HIAA-d5 vs HIAA-13C6Post-column infusion experiment: samples with > ±10% deviation
37
Example 2: Quantitation w/ HIAA-d5 vs HIAA-13C6
HIAA-d5 HIAA
ΔtR = 0.03 min
HIAA-13C6 HIAA
ΔtR = 0.00 min
Scenario: Urinary HIAA method with
HIAA-d5 IS Dilute-and-shoot Analyte and IS not coeluting Differential suppression
between analyte and IS Quadratic calibration curve fit
Solution:Replace HIAA-d5 with coeluting IS:
HIAA-13C6(Alternate isotope label)
Alternate isotope labels: 13C,15N,…
Advantages Do not suffer from label
instability
Closer co-elution with analyte
Better compensation for matrix effects
Drawbacks
More difficult to make
More expensive
Not always commercially available
38
39
Problem #2: Coeluting IS not compensating for matrix effects
Scenario: Urinary Vanillylmandelic acid (VMA)
and Homovanillic acid (HVA) Dilute-and-shoot
4 min LC method
Co-eluting IS (VMA-d3) Interferences:
~3% VMA samples
( vs. < 0.03% HVA samples) Suppression zone partially coeluting
with analyte in validated method
Inte
nsity
VMA
Time (min)
40
Problem #2: Coeluting IS not compensating for matrix effects
Normal scenario:No to very little signal suppression by matrix components
Time (min)
Inte
nsity
VMA
Time (min)In
tens
ity
VMA
Severe matrix effect:Signal suppression severely decreases S/N for analyte and IS=> Assay performance compromised, especially near LLOQ
41
Solution: Chromatographic separation
Primary LC method for regular testing
Alternate LC method for reinjection of samples with VMA interference
Injection-to-injection time remained 4 min
Inte
nsity
VMA
Time (min)
Primary method
Time (min)
Inte
nsity
VMAAlternate method
Problem #2: Coeluting IS not compensating for matrix effects
42
Interference mitigation by internal standard???
Only when…
Using IS exactly coeluting with analytei.e. one with stable isotope labels that do not
exhibit separation effect (13C, 15N)better compensate for matrix effects
AND
Signal suppression is NOT severe
43
You validated a robust method, thoroughly tested for interference, and handed it off to the clinical lab.
Are you done with interference?
Even the best method development strategies rarely able to prevent interference completely
Monitor for interference in routine testing to avoid reporting compromised results
44
How do we monitor for interference?
Most relevant data quality metrics: Ion ratios
Interferents isobaric with analyte/IS(appear on analyte/IS transitions)
Absolute Internal Standard areas Interferents causing matrix effects
(signal suppression/enhancement)
Retention times Near-eluting isobaric interferent integrated instead of analyte/IS
Deviations can signal the presence of interference in either the analyte or internal standard mass chromatograms
45
𝑰𝑰𝑰𝑰𝑰𝑰 𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝑰𝑰 =𝑸𝑸𝑸𝑸𝒓𝒓𝑰𝑰𝒓𝒓𝒓𝒓𝑸𝑸𝒓𝒓𝒊𝒊𝒓𝒓 𝑴𝑴𝒓𝒓𝑴𝑴𝑴𝑴 𝑻𝑻𝒓𝒓𝒓𝒓𝑰𝑰𝑴𝑴𝒓𝒓𝒓𝒓𝒓𝒓𝑰𝑰𝑰𝑰 𝑷𝑷𝒊𝒊𝒓𝒓𝑷𝑷 𝒓𝒓𝒓𝒓𝒊𝒊𝒓𝒓𝑸𝑸𝑸𝑸𝒓𝒓𝑸𝑸𝒓𝒓𝑸𝑸𝒓𝒓𝒊𝒊𝒓𝒓 𝑴𝑴𝒓𝒓𝑴𝑴𝑴𝑴 𝑻𝑻𝒓𝒓𝒓𝒓𝑰𝑰𝑴𝑴𝒓𝒓𝒓𝒓𝒓𝒓𝑰𝑰𝑰𝑰 𝑷𝑷𝒊𝒊𝒓𝒓𝑷𝑷 𝒓𝒓𝒓𝒓𝒊𝒊𝒓𝒓
Quantifier
Qualifier
Ion ratios –Monitoring for isobaric interferents in chromatograms
46
Ideally calculated for both analyte and internal standard
Individual specimen ion ratios compared to mean ion ratio
Mean ion ratio – calculated from ion ratios of calibration standards / quality controls
Acceptance limits set during method development
based on clinical requirements for the assay
±20 or 30% common
Ion ratios –Monitoring for isobaric interferents in chromatograms
47
No interference Interference
Quantifier
Qualifier
Quantifier
Qualifier
Quantifier peak area 1.02E+05Qualifier peak area 6.43E+04Ion ratio 1.58IR±20% range 1.26-1.89
Ion ratios –Monitoring for isobaric interferents in chromatograms
Quantifier peak area 1.18E+05Qualifier peak area 5.57E+04Ion ratio 2.11IR±20% range 1.26-1.89
48
Absolute Internal Standard areas –Monitoring for matrix effects (mainly signal suppression)
Mean or median value calculated For the entire run, or
For calibration standards set
Acceptance limits set during method development Based on clinical requirements for the assay, and
Overall method performance
Individual specimen peak area values compared to mean/median value
49
Absolute Internal Standard areas –Monitoring for signal suppression
0
50,000
100,000
150,000
200,000
250,000
300,000
0 50 100 150
IS P
eak
Area
(cou
nts)
Sample #
Internal Standard 1
Well-controlled assay
200% of peak area median
50% of peak area median
peak area median
50
Absolute IS areas –Monitoring for signal suppression
Not so well-controlled assay
0
50,000
100,000
150,000
200,000
0 50 100 150
Peak
Are
a (c
ount
s)
Sample #
Internal Standard 2
200% of peak area median
50% of peak area median
peak area median
Typical of dilute&shoot methods – large variations due to signal suppression
Can indicate samples extraction issues –variable recovery
51
Retention times (absolute or relative to IS) –Monitoring for near-eluting isobaric interferent integrated in place of analyte/IS
Analyte
IS
Interference
2.59 min
2.53 min
2.75 min Interference
IS2.53 min
2.75 min
52
Retention times (absolute or relative to IS) –Monitoring for near-eluting isobaric interferent integrated in place of analyte/IS
Set up acceptance limits in MS software for retention time (RT) differences between analytes and their internal standards
Absolute RT differences
Percentage of target RT for IS
Acceptance limits based on method performance evaluated during development and validation
Analyte RTs beyond the acceptance limits are flagged by the software
53
Metanephrine interference
Original Qual
Original Quant
No interference
Interference on Quant
Interference on Qual
Scenario: Urinary metanephrines Deuterated (d3) IS SPE sample cleanup 5 min LC method
Interferences on both Quant and Qual MRM in original method
54
Metanephrine interference
Original Qual
Original Quant
Alternate Qual
No interference Interference on Quant Interference on Qual
Use: Alternate Qual Original Qual as
new Quant MRM
Use: Alternate Qual Original Quant
Solution: Alternate quantitation method using a 3rd MRM transition
Use: Original Qual Original Quant
55
OC+IntfOC-d6
Specimen with Oxycodone and coeluting interference
316.15 241.2316.15 256.2
Coeluting Oxycodone interference
Scenario: Opioid panel
Unresolved Oxycodone interference in original opioid method
SPE sample prep
5 min run time
Added all viable MRMs
316.15 212.3316.15 181.3316.15 168.3316.15 128.2316.15 115.2
56
316.15 168.3316.15 115.2316.15 241.2316.15 256.2316.15 181.3316.15 212.3316.15 128.2
Scenario: All MRMs give signal
for both analyte and interference
But at different intensities!!!
Coeluting Oxycodone interference
CalibratorOC
OC-d6316.15 241.2316.15 256.2316.15 212.3316.15 181.3316.15 168.3316.15 128.2316.15 115.2
Specimen with OC interference only
OC intf
OC-d6
57
OC intf
OCOC-d6 OC-d6
OCOC intf
Coeluting Oxycodone interference
Solution: Chromatographic separation to resolve analyte from interferent Change in LC gradient profile
58
References:1. CLSI. Evaluation of matrix effects; Approved guideline – second edition. CLSI document EP14-A2.Wayne (PA): CLSI; 2005.2. CLSI. Interference testing in clinical chemistry; Approved guideline – second edition. CLSI document EP7-A2.Wayne (PA): CLSI;
2005.3. CLSI. Liquid chromatography-mass spectrometry methods; Approved guideline. CLSI document C62-A. Wayne (PA): CLSI; 2014.4. FDA. Guidance for Industry – Bioanalytical Method Validation, 2001. http://www.fda.gov/cder/guidance/index.htm.5. SWGTOX. Scientific Working Group for Forensic Toxicology Standard Practices for Method Validation in Forensic Toxicology,
Doc 003, 2013. https://academic.oup.com/jat/article/37/7/452/765476.6. World Anti-Doping Agency (WADA) Technical Document – TD2015IDCR. https://www.wada-
ama.org/sites/default/files/resources/files/td2015idcr_-_eng.pdf.7. European Medicines Agency. Guideline on bioanalytical method validation, 2012.
http://www.ema.europa.eu/docs/en_GB/document_library/Scientific_guideline/2011/08/WC500109686.pdf.8. Clark ZD, Balloch S, Calton L, Mason D. Interference Testing and Mitigation in LC-MS/MS Assays. Clinical Laboratory News
2017;43(8):22-5.9. Clark ZD, Cutler JM, Pavlov IY, et al. Simple dilute-and-shoot method for urinary vanillylmandelic acid and homovanillic acid by
liquid chromatography tandem mass spectrometry. Clin Chim Acta 2017;468:201–8.10. Clark ZD, Cutler JM, Frank EL. Practical LC-MS/MS method for 5-hydroxyindoleacetic acid in urine. J Appl Lab Med 2017;1:387–
99.11. Clark ZD, Strathmann FG, McMillin GA. Diluting and shooting yourself in the foot: Complications with sample-to-sample variations
in signal suppression. MSACL 2013 podium presentation.12. Clark ZD, Frank EF. Urinary metanephrines by liquid chromatography tandem mass spectrometry: Using multiple quantification
methods to minimize interferences in a high throughput method. J Chrom B 2011;879(31):3673-80.13. Matuszewski BK, Constanzer ML, Chavez-Eng CM. Strategies for the assessment of matrix effect in quantitative bioanalytical
methods based on HPLC-MS/MS. Anal Chem 2003;75:3019–30.14. Annesley TM. Ion suppression in mass spectrometry. Clin Chem 2003;49:1041–4.15. Bonfiglio R, King RC, Olah TV, et al. The effects of sample preparation methods on the variability of the electrospray ionization
response for model drug compounds. Rapid Commun Mass Spectrom 1999;13:1175–85.16. King R, Bonfiglio R, Fernandez-Metzler C, et al. Mechanistic investigation of ionization suppression in electrospray ionization.
J Am Soc Mass Spectrom 2000;11:942–50.17. Lynch KL. LC-MS/MS quality assurance in production: The real work begins after validation. Clinical Laboratory News
2017;43(5):28–9.18. Zabell APR, Stone J, Julian RK. Using big data for LC-MS/MS quality analysis. Clinical Laboratory News 2017;43(5):30–1.
AcknowledgementsMSACL organizers and committee members
Donald MasonLisa CaltonSteven Balloch
Dr. Marzia PasqualiDr. Frederick StrathmannDr. Elizabeth FrankDr. Gwen McMillin
Staff of ARUP’s Mass Spectrometry II, Clinical Toxicology, Analytic Biochemistry, and Biochemical Genetics Labs
ARUP Institute for Clinical and Experimental Pathology®
60