Measure turbidity to four decimal places. It enables you to monitor the quality of water which is good to drink or not. For managers at beverage waterworks and water purification plants. TC-Mi provides you with easy and accurate water quality management! Applications Easy installation Maintenance cost reduction High measurement accuracy Water intake management Scene 1 Detect membrane breakage Scene 2 Clean water management Scene 3 TC-Mi Ultra Low Turbidity Checker Ultra Low Turbidity Checker The compact design containing a deforming enables you to install easily. A LED light source eliminates the need for replacement. The condensation prevention function maintains a stable measure- ment. Manage raw water from rivers and wells Monitor the condition of the membrane unit and detect membrane breakage Monitor drinking-water quality
Transcript
Page 1: tc-mi - optex.co.jp · TC-Mi protocol (RS485) Signal output (analog 4 – 20 mA, resistance load 300 ohm or less) Self-checking output (open collector) 100 – 200 mL / min 0 to 40°C

Measure turbidity to four decimal places.It enables you to monitor the quality of water

which is good to drink or not.

For managers at beverage waterworks and water purification plants.TC-Mi provides you with easy and accurate water quality management!

Applications

Easy installation Maintenance costreduction

High measurementaccuracy

Water intakemanagement

Scene

1Detect membranebreakage

Scene

2Clean watermanagement

Scene

3

TC-MiUltra Low Turbidity Checker

Ultra Low Turbidity Checker

The compact design containing a deforming enables you to install easily.

A LED light source eliminates the need for replacement.

The condensation prevention function maintains a stable measure-ment.

Manage raw water from rivers and wells Monitor the condition of the membrane unit and detect membrane breakage

Monitor drinking-water quality

Page 2: tc-mi - optex.co.jp · TC-Mi protocol (RS485) Signal output (analog 4 – 20 mA, resistance load 300 ohm or less) Self-checking output (open collector) 100 – 200 mL / min 0 to 40°C

Copyright (C) 2018 OPTEX CO., LTD.

Specifications

・ Equipped with condensation prevention function・ Built-in deforming tank to remove air bubbles

Accurate and Stable measurement

・ High waterproof and dustproof structure to various installation sites

Water and dust resistance

・ No need to use or replace dehumidifier・ Employing a long-life light source

Low maintenance cost

76053-00-15747-1908

OPTEX CO., LTD.5-8-12 Ogoto Otsu Shiga 520-0101 JapanTel.+81-77-579-8680 Fax.+81-77-579-8199www.optex.co.jp/e/

Model No.Measuring method

Light source

Measuring range

Display

Display resolution

Source voltageConsumption current

Digital communication

Analog output

Measuring water flow rateMeasuring water temperature

Operating temperatureMain materials

DimensionWeight

Protective constructionOption

Note: Specifications and design are subject to change without prior notice.

Ultra Low Turbidity CheckerTC-Mi

Nephelometric 90° scattered lightLED

0.0001 – 100.00 (NTU/FNU)0.0010 – 20.000 (deg.: polystyrene standard solution)

LCD monitor0.0001 – 9.9999NTU: 0.0001NTU10.000 – 100.00NTU: 0.001NTU

24 VDC ±10%Normal time: 100 mA or less Startup time: 600 mA or less

MODBUS (RS485)TC-Mi protocol (RS485)

Signal output (analog 4 – 20 mA, resistance load 300 ohm or less)Self-checking output (open collector)

100 – 200 mL / min0 to 40°C (No freezing)

-20 to +50°C humidity 95%RH or lessPP-GF AES SUS316L

Approx. 157 (H) x 259 (W) x 147 (D) mmApprox. 2 kg

IP65TC-Mi-FL (Flowmeter)

mm(in)

Dimensions

Installation pitch 230(9.06)

259(10.2)

200(7.87)

Max

.140

(5.5

1)

111(

4.37

)17

(0.6

7)

35(1.38)16 (0.63)

Installation holeφ7(φ0.28)

147(5.79)

144(5.67)

116(4.57)Applicable cableφ6~8 (φ0.24~0.31)

Inst

alla

tion

pitc

h 65

(2.5

6)29

(1.1

4)


Recommended