TCMB 02 Conditioning Module
Contents
SECTION ONE
Page(s)User Instructions 3Preliminary notes 4Main electrical specifications 5Dimensions 6Types of connections 7Installation and power supply 8Electrical connections (conditioner and transducer) 9 - 12Transducer Extension cables 13 - 14Check of pre-calibration 15 - 16
SECTION TWO Pages(s)
Setup instructions 17 PCB layout 18Conditioner setup 19 - 21Conditioner calibration - Zero 22 - 23Conditioner calibration - Gain 23Conditioner calibration - Zero Offset 23Setup examples 24 - 27
Specifications in this document may be changed without notice
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 2 of 28
Section one
User Instructions
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 3 of 28
Preliminary notes
The TCMB 02 Series conditioning module are compact and flexible signal conditioner based on a standard TCM conditioner and boxed into plastic housing with options for different types of input connections, output signals and supply voltages.
The TCMB 02 conditioners are assembled and wired for use with Half Bridge inductive transducers and can also be supplied on request pre-adjusted (*) for use with specific types of 3TLX Series gauging transducers.
(*) Please note that a fine adjustment to be carried out by the customer may be required at the time of installation (see detalis on the following pages).
Environmental protection
The TCMB 02 conditioners (all versions) must be used in a dry environment and are not suitable for use in harsh environments with presence of high humidity / moisture. They must also be protected against contact with liquids and dust, especially if conductive or corrosive.
Temperature range from 0 °C to +50 °C (operating and storage).
Warm up
The TCMB 02 conditioners are precision modules and need approximately 15 minute warm up time to ensure stable measurement.
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 4 of 28
Main electrical specifications
Power Supply Min Typical Max Units
Optional power supplies
A) 5 5.2 5.5 B) 11.8 12 12.5
23.8 24.2 24.7 ± 14.5 ± 15 ± 18
Volt @ 120 mA Volt @ 65 mA Volt @ 25 mA Volt @ ± 25 mA
Selectable Options
kHz ±10%
2.2 V rms - Typical all frequencies 5.0 V rms - Typical all frequencies
C) D)
Transducer
Primary
Frequency
Amplitude (1) Amplitude (2)
Conditioning Min Typical Max Units
-11 +11 V dc Span
Zero Offset -100% 0 +100%
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 5 of 28
2 3 5 10 13
6
Dimensions in mm (drawings not to scale)
106
4,0
13
45
45
23
DIN rail mount option
Top view
Side view
66
45
Front view
Side view
3
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 6 of 28
7
Available types of electrical connections (drawings not to scale)
Transducer input socket on front side end Supply and Output - 1 m cable
-TCMB 02-1S - no DIN rail mount
TCMB 02-1A - DIN rail mount option
Supply and Output - 1 m cable
Supply and Output - 1 m cable
Transducer input Panel mount socket - with max. 10 m cable
Transducer input Flying socket - with max. 10 m cable
TCMB 02-2S - no DIN rail mount
TCMB 02-2A - DIN rail mount option
TCMB 02-3S - no DIN rail mount
TCMB 02-3A - DIN rail mount option
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 7 of 28
8
Installation and power supply
Make sure that the TCMB 02 is disconnected from any supply voltage during installation. Apply power only after all electrical connections are made between :
- TCMB 02 module and transducer- output from TCMB 02 and instrument or digital display.
For the supply of the TCMB 02 modules a regulated linear power supply is required (do not use switching-mode units).
Max. ripple 10 mV p.p. - @ 24 V output.
For a correct operation of the TCMB 02 module use a dedicated power supply that does not supply any other electronic equipment. When supplying the TCMB 02 make sure the correct value of output voltage from the power supply and the correct polarity of connection.
All TCMB 02 modules (all versions and styles) :
• are reverse polarity protected (power supply)• are not overvoltage protected (power supply)• are not short circuit protected (output signal)
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 8 of 28
9
Electrical connections - Power supply and Output signal
Colour code of Supply and Output cable
Description Cable colours
GND +VE0 V
= power supply return = power supply = output signal return
V OUT = output signal
0 V+24 V or +12 V or +5 V0 V±5 V or ±10 V or 0-10 V f.s.
blue red
green yellow
Supply and Output cable - 1 metre long (longer cables on request)
DBH2
L 2
L 1
DC
-DC
CO
NVE
RTE
R
xx15
S
0 / V
+VE
V OUT
S2
GND
TP3
FVout
0 V
+24 V
0 V
±5 V ( ±10 V) or 0-10 V f.s.
Power supply
Output signal
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 9 of 28
10
Power supply
The current versions of TCMB 02 conditioners differ from earlier versions they now have both Power supply return (GND) and Output return (0/V) electrically connected together through a solder bridge on link L2 on PCB.
If this link causes any operational defect to the equipment connected to the TCMB 02, please remove the bridge across the two pads.
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 10 of 28
11
Electrical connections - Transducer connection
2
3
4
1
2
3
4
1
5
DBH4
P
3w3w f1
v2f2f3
4w 4w
4w
3w
DBH3
TP1TP2
10k2k P
P
Energization
Output
Energization
Screen
PCB - unpopulated side
1
3
5
PCB - populated side
Not connected
IC 1 AD698AP
DBH1
Half Bridge
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 11 of 28
12
Electrical connections - Transducer plug connections and colour code
Grey
Black
( 3 )
( 5 )
Screen ( 2 )( 1 )
Red
3TLX10…3TLX50 Standard PUR cable
Screen ( 2 ) ( 1 )
Green
( 3 ) White
( 5 ) Brown
3TLX10…3TLX50 Halogen free cable
Cable screen of 3TLX10…3TLX50 Series transducers is connected to the transducer body and to pin 2 of plug.
1 5
3 4 2
5 pin round plug - solder side view
Blue
( 3 ) Yellow
( 5 )
Screen ( * )( 1 )
Red
3TLX100 - Standard PUR cable
(*) Cable screen of 3TLX100 transducers is connected to the transducer body and to case of plug.
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 12 of 28
13
Electrical connections - Transducer extension cables
Conditioners TCMB 02-1S and TCMB 02-1A can be supplied on request with one of the T100/7xx extension screened cables for remote connection of transducers (max. length 10 metres). The electrical connections of the T100/7xx extension screened cables are available only for the 3TLX10…3TLX50 Series transducers
Extension cables are wired for INCREASING output signal.
T100/781 T100/782 T100/787 T100/790
= length 1 metre = length 2.5 metres = length 7.5 metres = length 10 metres
max. 10 m
Transducer side connection 5 pin flying
socket
Extension cable - T100/7xx
TCMB 02 side connection 5 pin flying
plug
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 13 of 28
14
Recommendation for the extension of the transducer cable
In general, the distance between a 3TLX Series transducer and its conditioning unit shouldn’t exceed the standard 2 metre length of its cable. If an extension of the transducer cable is required, the overall length cannot exceed 12 metres : 2 metres for the transducer cable + max 10 metres for the extension cable. This extension must be made with an original cable, available on request in different lengths up to 10 metres, The use of the original manufacturer’s extension cable cannot guarantee the best performance and is essential to reduce as much as possible to the actual required length necessary. Avoid running cable parallel to power lines. Position the cable away from electrical machines or noise sources that may interfere with the 12 kHz transducer signal transmitted along the transducer cable. Do not place the transducer cable inside pipes or sleeves containing other electrical conductors. As the transducer performance may be affected by an to excessive length of its cable it is essential to test the application for a correct evaluation.
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 14 of 28
15
Check of correct calibration of conditioners delivered with an associated transducer (full measuring system)
Although each TCMB 02 conditioner may be delivered on request configured and pre-adjusted for use with a specific transducer of the 3TLX Series models, it is recommended to check its calibration (accuracy of measurement) after installation. If an “on site” adjustment is required, this can be carried out by the customer referring to the setup procedure described in the setup instructions (section two of this manual). Each conditioner is pre-adjusted in association with a specific transducer (model and serial number), but if a fine adjustment is required, this will compensate any error that might have been introduced by placement of the transducer at the time of its installation and/or to compensate any possibile variation in the system calibration due to operating conditions that could not be simulated in our laboratory during the pre-adjustment operations :
for instance :
- environmental conditions (temperature, etc.)- presence of electrical or electromagnetic noise or interferences- conditioner supply (if not correctly regulated and filtered )- interaction with equipments connected to the conditioner output
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 15 of 28
16
TCMB 02 conditioner delivered as a spare part (without transducer)
In this case the conditioner will be delivered on request configured and pre-adjusted in association with a sample 3TLX Series transducer from stock. The sample transducer used for the conditioner setup will be of the same model of that which will be used by the customer. Also in this case it is recommended to check the correct calibration of the conditioner after its installation.
Replacement of a 3TLX Series transducer
In most cases the replacement of a faulty 3TLX Series transducer with a new one of the same type does not require recalibration of its conditioner. This is because during factory calibration, the sensitivity of each transducer of the same type is attenuated to a specific value, within tight tolerance, in order to standardise the output of all transducers and reduce the measurement error introduced by the replacement of the transducer. This error is generally accepted by industry, and unless the application requires very accurate measurements, fine recalibration of the conditioner is not necessary. It is recommended to check the correct calibration of the conditioner after replacement of the faulty transducer.
As a general rule, if after replacement of a faulty transducer fine recalibration of the conditioner is required, this is to compensate any possible measurement error introduced by :
- non-accurate positioning of the new transducer during installation onits fixture
- possible non-simmetry between positive and negative stroke of thenew transducer
- different sensitivity between faulty transducer and replacementtransducer
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 16 of 28
17 Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 17 of 28
Section two
TCM - Setup Instructions
18
PCB - Printed circuit board layout (drawings not to scale)
DBH4
DBH1 DBH3
GAIN
P 2
OFFSET
P 1
DBH2
DC
-DC
CO
NVE
RTE
R
xx15
S
L 2
L 1
IC 1 AD698AP
P
TP1TP2
10k2k P
P
0 / V
+VE
V OUT
S2
GND
TP3
FVout
4w 4w
2
3
4
1
4w
3wP
+
4w 4w 3w 3w f1 f2 f3 v2
Coarse Gain (span) Fine Gain (span)
_
Half Bridge Load, 2k or 10k TP1 available for shorting the LVDT output to determine TCM zero TP2 available for shorting the Half Bridge to determine TCM zero Note : Always position link in park position P (as shown) after setting TCM zero
3 kHz frequency 5 kHz frequency (used mainly for LVDT) 10 kHz frequency (used mainly for Half Bridge) With link in park position P (as shown) default is 13 kHz Primary 5 V rms energising
Coarse Zero Offset Fine Zero Offset
Handbag position for 3 wire Half Bridge transducers
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 18 of 28
19
Conditioner Setup
Before calibration, the TCM conditioner must be configured in accordance to the type of transducer to be connected (Half Bridge or LVDT) and the requested output signal.
This is carried out using the following 4 handbag links :
Handbag link DBH1 - Coarse Gain selection
Handbag link DBH2 - Coarse Offset selection
Handbag link DBH3 Link to simulate transducer at its centre stroke (yellow handbag link) Demodulator load selection (green handbag link)
Handbag link DBH4 Transducer type selection : Half Bridge or LVDT (2 black handbag links) Transducer energizing voltage and frequency selection (1 or 2 blue handbag links)
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 19 of 28
20
Conditioner Setup
Make sure the TCM is disconnected from any supply voltage. Following the manufacturer’s transducer information, wire the transducer according to the type of device, by either soldering to the appropriate pads or plug in to the optional in-line socket.
1. Transducer type, Energizing frequency and Enerigizing voltage
Position the handbag links on DBH4 to suit the type of transducer being used (Half Bridge or LVDT ), making sure the correct frequency and supply voltage are selected. The supply frequency of 5 kHz is mainly used for LVDT transducers. The supply frequencies of 10 or 13 kHz are mainly used for Half Bridge transducers. Select also the appropriate supply voltage of the transducer (2.2 V or 5 V).
DBH4
4w 4w3w 3w f1 f2
P
HALF BRIDGE transducers - position handbag links on 3w
LVDT transducers - position handbag links on 4w
f3 v2
3 kHz frequency - position handbag link on f1 5 kHz frequency - position handbag link on f2 10 kHz frequency - position handbag link on f3 13 kHz frequency - position handbag link on park position P
2 kHz frequency - position handbag links on f1 + f2
2.2 V voltage - position handbag link on park position P 5 V voltage - position handbag link on v2
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 20 of 28
21
2. Calibration load
Select the correct calibration load, as shown below (DBH3 handbag link). 2 kOhm or 10 kOhm is mainly used for Half Bridge transducers (check the correct value of calibration load in the transducer data sheet). For LVDT transducers position the green handbag link in park position P and check that the transducer linearity is not affected.
DBH3
Half Bridge transducers - select 2k or 10 kOhm Load (gree n handbag link)
LVDT transducers - Green handbag link in park position P ( this may affect the LVDT, resulting in non-linear output)
TP1TP2
2k10k
P
TP1TP2
2k10k
P
TP1TP2
2k10k
P
PPP
2 kOhm 10 kOhm LVDT
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 21 of 28
22
3. Conditioner calibration
For a correct calibration of the conditioner, the TCM must be first set for electrical zero. Fit the Coarse Offset handbag link DBH2 to the centre (park) position P.
Short the output from the transducer fitting the yellow handbag link of DBH3 to TP2 position for Half Bridge transducers (3w), or to TP1 position for LVDT transducers (4w). This simulates the transducer at electrical zero positions.
DBH2
Coarse Zero
PHan dbag link in park position P
LVDT transducer - handbag link in TP1 position Half Bridge transducer - handbag link in TP2 position
DBH3
4w - LVDT3w - 1/2 Ponte
TP1TP2
P
P
Half Bridge
TP1TP2
P4w - LVDT3w - 1/2 Ponte
P
LVDT
TP1TP2
P
P
Pos. P
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 22 of 28
23
Connect the DC supply voltage and measuring instrument, checking the correct polarity, and adjust the Fine Zero (Offset) control P1 to achieve zero DC volts. Remove the TP yellow handbag link and place in a park (P) position.
Gain (Span)
Set the transducer to its mechanical (electrical) zero position, then with the core adjusted to its required stroke, fit the Coarse Gain handbag link DBH1 to a position that produces the nearest required result e.g. 10.00 V then adjust the Fine Gain control P2 for final setting.
Zero Offset (if needed)
With the transducer at its desired stroke (i.e. -5 mm) for zero volts, fit the Coarse Offset handbag link DBH2 to a position closest to zero volts adjusting the Fine Offset control P1 for final setting. Position the transducer to the other end of its stroke (i.e. +5 mm) and adjust Gain controls for required voltage e.g. +10.00 V on the measuring instrument. Due to the possibility of slight interaction, please repeat the above procedure.
DBH1
Coarse gain
increase decrease + _
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 23 of 28
24
SAMPLES
1) Sample Setup for Half Bridge using no Offset and a ±10 V Increasing Output
DBH4
DBH1 DBH3
GAIN
P 2
OFFSET
P 1
DBH2
DC
-DC
CO
NVE
RTE
R
xx15
S
L 2
L 1
IC 1 AD698AP
P
3w3w f1
v2f2f3
TP1TP2
10k2k P
P
0 / V
+VE
V OUT
S2
GND
TP3
FVout
4w 4w
2
3
4
1
54w
3wP
Higher Lower
To increase or decrease the Gain (span), move this handbag left or right for coarse adjustment before final adjustment of the Gain pre-set
-10 V 0 V +10 V
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 24 of 28
25
SAMPLES
2) Sample Setup for Half Bridge using 100% Offset and a 0-10 V Increasing Output
DBH4
DBH1 DBH3
GAIN
P 2
OFFSET
P 1
DBH2
DC
-DC
CO
NVE
RTE
R
xx15
S
L 2
L 1
IC 1 AD698AP
P
3w3w f1
v2f2f3
TP1TP2
10k2k P
P
0 / V
+VE
V OUT
S2
GND
TP3
FVout
4w 4w
2
3
4
1
54w
3wP
Higher Lower
To increase or decrease the Gain (span), move this handbag left or right for coarse adjustment before final adjustment of the Gain pre-set
0 V +10 V
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 25 of 28
26
SAMPLES
3) Sample Setup for LVDT using no Offset and a ±10 V Increasing Output
DBH4
DBH1 DBH3
GAIN
P 2
OFFSET
P 1
DBH2
DC
-DC
CO
NVE
RTE
R
xx15
S
L 2
L 1
IC 1 AD698AP
P
4w4w f1
v2f2f3
TP1TP2
10k2k P
P
0 / V
+VE
V OUT
S2
GND
TP3
FVout
4w 4w
2
3
4
1
54w
3wP
Higher Lower
To increase or decrease the Gain (span), move this handbag left or right for coarse adjustment before final adjustment of the Gain pre-set
-10 V 0 V +10 V
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 26 of 28
27
SAMPLES
4) Sample Setup for LVDT using 100% Offset and a 0-10 V Increasing Output
DBH4
DBH1 DBH3
GAIN
P 2
OFFSET
P 1
DBH2
DC
-DC
CO
NVE
RTE
R
xx15
S
L 2
L 1
IC 1 AD698AP
P
4w4w f1
v2f2f3
TP1TP2
10k2k P
P
0 / V
+VE
V OUT
S2
GND
TP3
FVout
4w 4w
2
3
4
1
54w
3wP
Higher Lower
To increase or decrease the Gain (span), move this handbag left or right for coarse adjustment before final adjustment of the Gain pre-set
0 V +10 V
Copyright © 2019 Innovative Measurement Technology Ltd. | All Rights Reserved Page 27 of 28