Technology Challenges for Active Cardiac Implantable Devices
Alain Ripart – Senior VP and CSO
Heart rhythm disorders and associated therapy devices
Bradycardia
Atrial Arrhythmias
Ventricular arrhythmias
Congestive Heart Failure
Pacemaker
Pacemaker
ICD
CRT-D2
Smart, Powerful, Small
CRT-D
PMMarket value
2 844 2 890
211 190
2 7614 372
1 814
3 120
2007 2012
M€ 7,629
M€ 10,573
Heart Failure
Tachycardia
Bradycardia
11,5
9,6
0,2
CAGR, 2004-09 (%)
CRT-D
ICD
Total 6,7%
PM
CRT-P
3
Active Cardiac Implantable Devices
RFcommunication
Low powerCPU
Analog sensing & pacing
RAM ROM
MEMSSensors
High voltageDefibrillation
Wireless RemoteMonitoring
Programming System
Titaniumpackage
Hermeticpackage 4
Requirements for an implantable sensor
Physiologic but simple
Miniature
Integration
Low current consumption
Reliable Hermetically sealed
Biocompatible
Leadless / Wireless
5
1D Accelerometer
Accéléromètre
Batterie
What's inside the package?
Feedthrough with ground wire and capacitor filters
Hybrid circuit encompasses only components automatically managed by "Pick & Place" machine
Flex circuit provide high density interconnection and cancels complex & risky feedthrough wire bending
7
What’s inside? Hybrid circuit
8
Shock circuits
9
1 sec.1 sec.
20 J, T/N.20 J, T/N.DDD, 90 bpmDDD, 90 bpmV: 4.8 V, 0.37 msV: 4.8 V, 0.37 ms
LAE DALAE DA
Défibrillation
10
What’s inside? Ovatio defibrillator
11
Defibrillator downsizing
0
20
40
60
80
100
120
140
1993 1994 1995 1996 1997 1998 1999 2000
VVIDDD
cc
2001
010
2030
40
5060
7080
78 c
c
49 c
c
39 c
c
29 c
c
Mechanics & interconnexionConnectorElectronicsCapacitorsrsBattery
2002 2003
12
PoP and WDoD package
Innovations from : IPEDIAIntegration of passive components in silicon
3D
Adapted from NXP
PN532
3DCS
P5CN072
PN532
3DCS
P5CN072
3DCS3DCS
A technical challenge in CRM : circuit & package miniaturization
15
Hybrid Platform
1 2 3
4 5 6
8 7 9
10
Detailed views of DIAMOND stack process
Hybrid
Heart failure
2005: up to 14 million Europeans currently suffer from heart failure.
2020: increasing to 30 million.
Over 3.6 million new cases of heart failure are reported each year in Europe.
First cause of cardiovascular mortality in Europe.
Heart Failure is the most common cause of hospital admission in people over 65.
17
HF heart status change over time• CRT settings must be repeatedly tailored
to the individual patient
Self-adjustment of CRT parameters • Cardiac resynchronization is permanently
tailored to the patient• Replaces time-consuming echocardiographic assessment required with
conventional CRT devices require
Monitoring of the patient’s status
Optimal Management of Heart Failure
18
Heart failure sensor
Distal End, with accelerometer inside
Pacing Tip
Proximal end; connector fits into pacemaker receptacle
Lead body (length=60cm)
Lead Connections
19
Capteur SonR
20
What does the signal look like?
PEA I
PEA II
0 1 2
0
100
200
-0.5
0
0.5
EA (g)
PHONO
BP (mmHg)
ECG
t (s)
Isovolumic Contraction Phase Isovolumic Relaxation Phase
21
Clinical Efficacy of CRT continuous optimization with SonR versus standard clinical practice.
Heart Rhythm 2010 ;7(5s):AB27_4
.L.Padeletti et al
The future: exercise adaptive & monitoring of resynchronization
Neuronal Co-Processor
Micro-controller
Hemodynamic sensors
RA lead
RV lead
LV lead CRT / CRT-D Device
PulseGenerator,
Sense Amps
Real time AV and VV intervals optimization during exercise
23
Monitoring of implanted patients for detectionof
Cardiac Decompensation
(early detection of pulmonary oedema condition)
24
Ventilation measurement
ela
Injection (I)
Measure (dV)
Bonnet JL, Ritter P, Pioger G et al. Measurement of minute ventilation with different DDDR pacemaker electrode configurationsPacing And Clinical Electrophysiol 1998; 21 ; 1 [Pt I] : 4-10
ΔZ=ΔV/I and VE=k.ΔZ
25
Clinical case: True positive
Alarm by the RestRule first thenthe Activity rule
Patient hospitalized for CHF at the end of the 3-month FU phase D88).An alarm was delivered 1 month before at D58.
ALARMS
Rese
arch
softw
are
26
CRT system with multielectrodeslead
AMICA ASICs
Electrodes
Lead
Multiple distributed & communicating hemodynamicsensors for diagnosis and treatment of HF
27
Biocompatible packaging with ASIC substrate and sensor
0.3mm
2 mm1mm
0.8 mm
0.9 mm
0.9 mm
chip
MEMSAccelerometer
28
Evolution of implanted devices
S
S
SA
A
A
S S
FROM
A device with one, simple sensor on one lead
A device with- multiple, complex sensors- multiple, complex actuators- wired and wireless
TO
S S
S S
A
A
29
30
RESEARCH in FP7: Leadless pacemakerIC Device 1 (Technology 1)
MetallisationSystem 1
IC Device 2 (Technology 2) with TSV
MetallisationSystem 2
IC Device 2 (Technology 2) with TSV
MetallisationSystem 2Metallisation
MEMS/NEMS Device possibly with TSV
Cap -Chip (Wafer)
MEMS/NEMS Device possibly with TSV
Cap -Chip (Wafer)
MEMS/NEMS Device possibly with TSV
Cap -Chip (Wafer)
e-BRAINS
25mm
Diameter: 6 mmLength: 26 mm
Leadless pacing system
Heart beat
Energy scavenger
Leadless pacemaker
Communication Synchronization
31
32
RESEARCH in FP7: Leadless pacemakerIC Device 1 (Technology 1)
MetallisationSystem 1
IC Device 2 (Technology 2) with TSV
MetallisationSystem 2
IC Device 2 (Technology 2) with TSV
MetallisationSystem 2Metallisation
MEMS/NEMS Device possibly with TSV
Cap -Chip (Wafer)
MEMS/NEMS Device possibly with TSV
Cap -Chip (Wafer)
MEMS/NEMS Device possibly with TSV
Cap -Chip (Wafer)
e-BRAINS
24mm*15mm*5mm
8 cc
1CC
Remote Follow-upof implanted
Cardiac active devices
Telemedecine… was born
a long time ago !
33
non emergencymonitoring of device
and patient
Homemonitor
CRMpatient
IMD
Clinician access
home monitor inventoryHome Monitors and ICD
deliveries
Helpdesk
Orange hosting centre Orange application service
Sorin remote monitoring :a global solution
34
35
Hearing aids
Cardiac implants
Cochlear implants
Insulin pumps
2.4GHzmW‐level
<4x4x1 mm3
WBAN
RF+IF+LF MEMS
RF & DSP SoC
RADIO IC65nm CMOS
RF & LF MEMS
WiserBAN microsystem
Heterogenous SiP
Miniature antenna
Smart miniature low-power wirelessmicrosystem for Body Area Networks
Smart Systems for CRM What do we need ?
Ultra miniature micro-systems including
High density sub micron technology chips.
Large capacity static RAM
Sensors : accélérometers, pressure, impédance to measure activity,workloadposition,contractility,volume,minute-ventilation ,left side pressures etc…
RF transmission : antennas, Baw filters ,RF Mems for switches or sensors.
HV switches and storage capacitors.
Low leakage Hi energy density battery ,efficient energy scavengers
Micro-encapsulation, micro-connectors
Low current consumption
Reliability and Biocompatibility: Blood tightness.
36
THANK YOU FOR YOUR ATTENTION
37
Biocompatible packaging with glass substrate• ENCAPSULATION STEPS