Test Equipment Solutions Datasheet Test Equipment Solutions Ltd specialise in the second user sale, rental and distribution of quality test & measurement (T&M) equipment. We stock all major equipment types such as spectrum analyzers, signal generators, oscilloscopes, power meters, logic analysers etc from all the major suppliers such as Agilent, Tektronix, Anritsu and Rohde & Schwarz. As well as the headline benefit of cost saving, second user offers shorter lead times, higher reliability and multivendor solutions. Rental, of course, is ideal for shorter term needs and offers fast delivery, flexibility, try-before-you-buy, zero capital expenditure, lower risk and off balance sheet accounting. Both second user and rental improve the key business measure of Return On Capital Employed. All products supplied by Test Equipment Solutions include: Email: [email protected] Web: www.TestEquipmentHQ.com We are focused at the professional end of the marketplace, primarily working with customers for whom high performance, quality and service are key, whilst realising the cost savings that second user equipment offers. As such, we fully test & refurbish equipment in our in-house, traceable Lab. Items are supplied with manuals, accessories and typically a full no-quibble 1 year warranty. Our staff have extensive backgrounds in T&M, totalling over 150 years of combined experience, which enables us to deliver industry-leading service and support. We endeavour to be customer focused in every way right down to the detail, such as offering free delivery on sales, presenting flexible technical + commercial solutions and supplying a loan unit during warranty repair, if available. We are based at Aldermaston in the UK from where we supply test equipment worldwide. Our facility incorporates Sales, Support, Admin, Logistics and our own in-house Lab. - No-quibble parts & labour warranty (we provide transport for UK mainland addresses). - Free loan equipment during warranty repair, if available. - Full electrical, mechanical and safety refurbishment in our 40GHz in-house Lab. - Certificate of Conformance (calibration available on request). - Manuals and accessories required for normal operation. - Free insured delivery to your UK mainland address (sales). - Support from our team of seasoned Test & Measurement engineers. - ISO9001 quality assurance. T: 01183 800 800 F: 01183 800 804 Test Equipment Solutions Ltd Unit 3 Zodiac House Calleva Park Aldermaston Berkshire RG7 8HN
Transcript
Page 1: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

Test Equipment Solutions Datasheet

Test Equipment Solutions Ltd specialise in the second user sale, rental and distribution of quality test & measurement (T&M) equipment. We stock all major equipment types such as spectrum analyzers, signal generators, oscilloscopes, power meters, logic analysers etc from all the major suppliers such as Agilent, Tektronix, Anritsu and Rohde & Schwarz.

As well as the headline benefit of cost saving, second user offers shorter lead times, higher reliability and multivendor solutions. Rental, of course, is ideal for shorter term needs and offers fast delivery, flexibility, try-before-you-buy, zero capital expenditure, lower risk and off balance sheet accounting. Both second user and rental improve the key business measure of Return On Capital Employed.

All products supplied by Test Equipment Solutions include:

Email: [email protected]: www.TestEquipmentHQ.com

We are focused at the professional end of the marketplace, primarily working with customers for whom high performance, quality and service are key, whilst realising the cost savings that second user equipment offers. As such, we fully test & refurbish equipment in our in-house, traceable Lab. Items are supplied with manuals, accessories and typically a full no-quibble 1 year warranty. Our staff have extensive backgrounds in T&M, totalling over 150 years of combined experience, which enables us to deliver industry-leading service and support. We endeavour to be customer focused in every way right down to the detail, such as offering free delivery on sales, presenting flexible technical + commercial solutions and supplying a loan unit during warranty repair, if available.

We are based at Aldermaston in the UK from where we supply test equipment worldwide. Our facility incorporates Sales, Support, Admin, Logistics and our own in-house Lab.

- No-quibble parts & labour warranty (we provide transport for UK mainland addresses).- Free loan equipment during warranty repair, if available.- Full electrical, mechanical and safety refurbishment in our 40GHz in-house Lab.- Certificate of Conformance (calibration available on request).- Manuals and accessories required for normal operation.- Free insured delivery to your UK mainland address (sales).- Support from our team of seasoned Test & Measurement engineers.- ISO9001 quality assurance.

T: 01183 800 800 F: 01183 800 804

Test Equipment Solutions LtdUnit 3 Zodiac HouseCalleva ParkAldermastonBerkshireRG7 8HN

Page 2: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

The 5700A/5720AHigh Performance

Multifunction CalibratorsTaking Accuracy to a

Whole New Level

Page 3: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

The 5700A/5720A CalibratorsSet the Highest Standards

for Accuracy and Performance

Growing global trade is causingmanufacturers to pay closeattention to controlling, improv-ing, and measuring quality inall processes from designthrough manufacturing. Whilethis is good for product quality,it also increases your calibrationworkload. Fluke offers a wide rangeof calibration solutions to helpyou handle that workload.Choices that fit your budgetand your business.

High performance plus achoice of confidence levelsSince 1988, the Fluke 5700Ahas set the standard for multi-

function calibrator performance.The Fluke 5700A Series II and5720A set even higher stan-dards, calibrating your highperformance workload up to 81/2

digit system DMMs. Bothmodels now state specificationsto the standard Fluke 99%confidence level as well as the95% confidence level to supporteasy measurement comparisonsaccording to internationalquality standards includingISO 9000 and ISO Guide 25.All specifications are absoluteand include the uncertainty ofthe calibration standards used.No additional analysis isrequired. Relative specificationsare also available.

Dramatically improvedperformance specificationsThe new high performance5720A offers an unprecedentedaccuracy level that delivers thelowest uncertainties available. Itcalibrates more of your precisionmeter workload, more easily andcost effectively than ever before.

Intuitive easy-to-use designThe 5700A Series II and the5720A are remarkably easy touse. All readouts appear in plainlanguage rather than in crypticerror codes. Just enter values onthe numeric keypad and pressthe ENTER key.

Page 4: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

Both the 5700A Series II andthe 5720A feature RS-232 andIEEE-488 interfaces, and cancontrol Fluke 5725A, 5220A,and 5205A/5215A Amplifiers, sothey are compatible with yourexisting equipment. Both alsofeature Artifact Calibrationwhich helps keep support costslow while providing confidencethat your instrument is perform-ing as you expect it to. The rugged 5700A/5720ACalibrators operate over a widetemperature range so they’re agood choice for on-site work aswell as in the lab.

A critical factor in specified calibratorperformance is the difference between theactual output value and the nominal outputvalue. The confidence interval is a statisticalexpression of the likelihood that any outputof any instrument will deviate beyond thisspecified difference or uncertainty. At Fluke we state calibrator specificationswith better than 99% confidence tominimize the risk for the user. However,because international metrology standardsrequire using a 95% confidence interval inall measurements, the 5700A Series II and5720A now offer both 99% and 95%confidence level specifications. This makesit easier to make valid comparisons ofmeasurements, and permits you to accept aslightly higher statistical risk in return forlower instrument uncertainty. In addition,both absolute and relative specificationsare provided.

Understanding confidence intervals

-4 -3 -2 -1 0 +1 +2 +3 +4Output Uncertainty (PPM)

Nominal Value

Uncertainty Specificationat 95% Confidence Level

Uncertainty Specificationat 99% Confidence Level

Page 5: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

Increase Confidenceand Efficiency While

Reducing Ownership Costs

The 5700A/5720A Calibratorsfeature Artifact Calibration. Onlythree artifact standards—a 10Vdc reference and 1Ω and 10 kΩresistance references—arerequired to calibrate all rangesand functions to full specifica-tions. Front panel instructionsprompt the operator to makeconnections and inputs eachstep of the way. The calibratorcontrols the process whichtakes only about an hour,compared to several hoursusing traditional methods.

In addition to saving timeand equipment costs, ArtifactCalibration extends the needfor external verification to twoyears or longer as required byyour standards and procedures.And, because the 5700A SeriesII and 5720A can tolerateoperating temperatures be-tween 15°C and 35°C they canbe calibrated where they’reused, rather than having to berecalled to the standardslaboratory for calibration.

Cal Check monitors perfor-mance between calibrationsFor extra confidence that5700A/5720A Calibrators staywithin specifications betweencalibrations, the built-in auto-mated Cal Check functionchecks each range and functionagainst internal standards tosee that all functions andranges are within specification.These Cal Check results can alsobe downloaded to a computervia the IEEE-488 or RS-232port to develop control chartsthat predict the calibrator’slong-term performance. Or youcan print the results directly.

Page 6: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

Artifact Calibration transfers theassigned values of an externalartifact to a large array of multi-dimensional parameters within aninstrument. The instrument takesover the manual metrology functionsof establishing ratios and makingcomparisons, as well as controllingthe measuring process. As a result, Artifact Calibrationrequires just three standards totransfer external traceable values

Save time and support costs with Artifact Calibration

into the instrument. At that point, thecalibrator measures, verifies, andadjusts itself, reducing an eight-hourjob to about one hour. Thousands of5700A calibrators in service aroundthe world prove Artifact Calibrationdelivers fast, easy, and inexpensivecalibration along with the confidencethat your instrument is performing asexpected between calibrations.

732BdV Ref.

742A1Ω Ref.

742A10 kΩ Ref.

InternaldV Ref.

Internal1Ω Ref.

Internal10 kΩ Ref.

InternalPrecisionDivider

InternalAC/DC

Xfer Std.

InternalNull

Detector

Page 7: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

A simple, calculator-style keyboardmakes it easy to quickly enter values.

Plain language display states results insentences rather than in cryptic error codes.

Press the SPEC key to displaythe total uncertainty for thepresent output, calculated at a99% or 95% confidence level.

Quickly select between OPERATE and STANDBYmodes by pressing a single button. STANDBYmode disconnects output and sense terminalsallowing the calibrator to share a common outputcable with other calibration equipment.

Low-loss Type-N connector supports awideband voltage option that provides flat,low-noise alternating voltage output from

Auxiliary current output for meterswith separate current terminals.

Interfaces for RS-232, IEEE-488,and the Fluke 5725A, 5220A, and5205A/5215A Amplifiers.

Resistances in x1.0and x1.9 decades.

The 5720A pushes accuracystandards for multifunctioncalibrators to a new level, yetis designed to be remarkablyeasy to use.

Works with what you haveBoth the 5700A Series II and5720A are designed to protectyour investment in existingFluke instruments and proce-dures. For example, you can

connect a Fluke 5220ATransconductance Amplifier or5205A/5215A Voltage Amplifi-ers directly to dedicated portson either calibrator. In auto-mated systems, you can config-ure both the 5700A Series IIand 5720A to emulate thepopular 5100B Calibrator,minimizing the need to rewrite

your procedures. In addition,the 5700A Series II and 5720Acan emulate the Fluke 5200AAC Voltage Calibrator, and theirremote programming interfaceremains fully compatible withprevious versions of the 5700A.

Designed to go the distanceThe 5700A/5720A Calibratorsare designed for long, reliable,

External sense minimizes lead resistance errors.

Complete CalibrationSolutions that Set NewStandards for Usability

10 Hz to 30 MHz to handle RF voltmeters.

Page 8: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

To verify the reading simply adjustthe output knob and the error isdisplayed directly in ppm or %.

OFFSET and SCALE keys make it easyto compensate for zero offset and scaleerrors and permit direct display oflinearity errors at any scale level.

Press the x10 and ÷10 buttons toinstantly increase or decreaselevels in even decade steps.

Presetting output limits protects theoperator and the unit under test.

Use function keys work with the 3-linemessage display to select and store I/O portparameters including 5200A emulation, andto select external phase lock, phase referenceoutput, and stored procedures for internalverification or zero, internal referencing,external traceability, and diagnostics.

Display saver automatically blanksthe dot matrix display after aspecified interval of idle time.

Phase-in and phase-out portsallow you to lock two calibratorstogether to simulate power.

trouble-free operation. Statisti-cal process control techniquesassure that quality is continu-ously maintained and improvedfrom component test throughfinal assembly. Every instrumentis subjected to 2g of randomvibration to prevent subtleimperfections from causingfailures later.

Both the 5700A Series II andthe 5720A feature a rugged,modular design that minimizesfailures and makes it easy tofind and repair any faults thatmay occur. Internal diagnosticsexercise both digital and analogfunctions and can isolateproblems to the board level, sorepair often requires nothingmore than replacing a plug-inmodule. And their rugged

chassis are built to go to theworkload with minimal risk ofdamage. DIN connectors sealout potential contaminantswhile keeping circuit boardsfirmly connected. You can also count on Fluke’sworldwide network of serviceand calibration centers whichoffer quality work and fastturnaround at competitive prices.

Page 9: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

5720A MultifunctionCalibratorThe 5720A calibrator raises theperformance of the worldstandard 5700A to the nextlevel. The 5720A covers themost demanding workload,covering DMMs up to 81/2 digitsand delivering the highest levelof performance of any calibratoron the market. It states specifi-cations to both 99% and 95%confidence levels and includespowerful internal features, suchas Artifact Calibration and CalCheck, to simplify support anddramatically reduce your costof ownership.

5700A Series II MultifunctionCalibratorIn service since 1988, the

5700A has undergone continu-ous improvements to becomethe 5700A Series II, one of themost tested and reliable highprecision calibrators Fluke hasever produced. Considered thecalibration standard worldwide,the 5700A Series II delivershigh value as well as accuracy,covering 51/2 to 71/2

digit DMMs.

Plus it offers the same ease ofuse, low cost of ownership,rugged design, simplifiedsupport, and confidence build-ing features as the 5720A.

5500A Multi-ProductCalibratorThe 5500A covers an unprec-edented range of dc and low-frequency electrical calibrationworkload, including digital andanalog multimeters, thermom-eters, handheld wattmeters,

current clamps, oscilloscopes,process calibrators, powerharmonics analyzers, andmuch more at an extremelyaffordable price.

5790A AC MeasurementStandardThe 5790A is a completeautomated ac measurementstandard designed for the mostdemanding calibration applica-tions. It provides a 24 ppmtotal uncertainty and covers awide 700 µV to 1000V voltagerange and a 10 Hz to 1 MHzfrequency range with theoption to extend it to 30 MHz.

732B Direct VoltageReference StandardThe 732B is a small, rugged,solid-state, direct voltage

The Fluke Calibration FamilySets the Standard

for Value and Performance

Page 10: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

reference standard. Designed forreliable convenient transfers, itfeatures very predictable perfor-mance and long battery life.

742A Resistance StandardsThese high-accuracy workingstandards for precision on-siteresistance calibration feature arugged design and can be usedover a wide temperature range.

792A AC/DC TransferStandardThe 792A offers extraordinarytransfer accuracy with totaluncertainties to ±10 ppm trace-able through Fluke. Its robustinput protection and fast settlingtimes enable you to makemeasurements covering voltagesfrom 2 mV to 1000V, and fre-

quencies from 10 Hz to1 MHz—in just 30 seconds.

5725A AmplifierThe 5725A Amplifier increasesmaximum direct and alternatingcurrent to 11A for calibrating thehigh current ranges of popularlow-cost, handheld DMMs. Italso extends the calibrator’s

alternating Volt-Hertz productto 1100V at 30 kHz and 750Vat 100 kHz to cover the calibra-tion requirements of high-accuracy bench and systemmeters. It is compatible withthe 5700A, 5720A, and 5500A.

Functions Ranges Output 5720A 5700ABest traceable Best traceableuncertainty uncertainly(95% 180 Days) (95% 180 Days)

DC Voltage 0 - ±1100V 10V ±3.25 ppm ±6.4 ppmAC Voltage 220 µV - 1100V 1V ±55 ppm ±87 ppmResistance 0 - 100 MΩ 10 kΩ ±9 ppm ±12 ppmDC Current 0 - ±2.2A 10 mA ±37 ppm ±65 ppmAC Current 9 µA - 2.2A 100 mA ±140 ppm ±190 ppm

The 5700A/5720A Multifunction Calibrators at a glance

Page 11: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

Today’s quality standardsare imposing more and morestringent requirements fordocumentation and reporting.Fluke calibration softwareprovides an easy, affordablesolution.

MET/CAL is a powerfulenvironment for creating,editing, testing, and document-ing calibration procedures,and for performing automatedcalibrations. It supportsa wide variety of standardsand includes more than300 procedures.

MET/TRACK is a powerfulmetrology property manage-ment package designed to helpyou document and report on allaspects of your measurementassets, including inventory,calibration history, results,location and repair.

Its powerful client-serverdesign makes it ideal forapplications from single-userinstallations up to enterprise-wide networks. Plus its indus-try-standard SQL databaseoffers easy data access from avariety of applications andoperating systems. CrystalReports™ Professional isincluded with every serverpackage to provide virtuallyunlimited reporting capabilities.

Calibration service, repair,and supportFluke provides extensivecalibration support and serviceto maximize the value of yourcalibration investment. Ourworldwide network of Calibra-tion Centers has traceabilityto national standards. Serviceand support are just a telephonecall or fax away. We offer fast,quality repair and calibrationservices including a moduleexchange program, comprehen-sive training, and full supportin setting up your lab. Andwe provide a wide range oftechnical information andtraining, even a Software User'sSupport Group.

Fluke Calibration: the powerand flexibility you need tomeasure upWhen you examine ourcalibrators, standards, software,service, and support, you'll seewhy Fluke Calibration productsare the worldwide leaders.Fluke can help you meet justabout any electrical calibrationchallenge, including meetingISO 9000 or other safety,nuclear regulatory or environ-mental quality standards.We also offer a wide varietyof other instruments to helpmeet your specific applicationrequirements. Just contact yourlocal Fluke representative forexpert advice.

Automate Your Calibration andDocumentation Management

with MET/CAL® and MET/TRACK®

for Microsoft Windows®

Ordering InformationModel5720A Calibrator5700A Series II Calibrator5500A Calibrator

Options5700A-03 Wideband AC Voltage

(compatible with both the 5700Aand the 5720A)

5500A-SC 300 MHz Oscilloscope Calibrationoption for the 5500A

Accessories5725A Amplifier5220A Transconductance Amplifier5790A AC Measurement Standard734A DC Reference Standard732B DC Standard742A Resistance Standards792A AC/DC Transfer StandardMET/CAL - 4 Calibration SoftwareMET/TRACK - 4 Metrology Property Management

SoftwareMET/TRACK - 4+ 5-user VersionMET/TRACK - ND4 Workstation Node

For more informationFor more information on the5700A Series II and 5720ACalibrators, other Fluke calibra-tors, or MET/TRACK or MET/CALsoftware call your local Flukerepresentative.

For more information on confi-dence intervals and the devel-opment of calibrator specifica-tions, see: -Calibration: Philosophy andPractice -Application note B0254,Understanding and ComparingInstrument Specifications.

For more information on ArtifactCalibration ask for: -Artifact Calibration: Theoryand Application – outlines thedesign, verification and testingbehind the concept. -An Evaluation of ArtifactCalibration by respected me-trologist Les Huntley – looks athundreds of 5700As at theirtwo-year verification to assessthe effectiveness of the concept.

Page 12: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,
Page 13: Test Equipment Solutions DatasheetSince 1988, the Fluke 5700A has set the standard for multi-function calibrator performance. The Fluke 5700A Series II and 5720A set even higher stan-dards,

Fluke CorporationPO Box 9090Everett, WA USA 98206-9090

Fluke Europe B.V.PO Box 1186, 5602 BDEindhoven, The Netherlands

For more information call:In the U.S.A. (800) 443-5853 orFax (425) 356-5116In Europe (31 40) 2 678 200 orFax (31 40) 2 678 222In Canada (905) 890-7600 orFax (905) 890-6866

From other countries (425) 356-5500or Fax (425) 356-5116Web access: http://www.fluke.com

© 1996 Fluke Corporation. All rights reserved.Crystal Reports for Windows is a trademark ofCrystal Computer Services, Inc. MET/CAL andMET/TRACK are registered trademarks of FlukeCorporation. Microsoft and Windows areregistered trademarks of Microsoft Corporation.Printed in U.S.A. G0378UEN Rev B

CERTIFIED TO MEET ISO9001

QUALIT

Y MANAGEMENT SYSTEM

ISO 9001

Fluke. Keeping your worldup and running.