TH Premium RIC LI 19Technical Data
S-Receiver ● 56 dB / 119 dB SPL (ear simulator)
● 45 dB / 108 dB SPL (2 ccm coupler)
M-Receiver ● 70 dB / 129 dB SPL (ear simulator)
● 60 dB / 119 dB SPL (2 ccm coupler)
P-Receiver ● 80 dB / 134 dB SPL (ear simulator)
● 70 dB / 124 dB SPL (2 ccm coupler)
HP-Receiver ● 82 dB / 138 dB SPL (ear simulator)
● 75 dB / 130 dB SPL (2 ccm coupler)
TH Premium RIC LI 19 | 2
TH Premium RIC LI 19 | Technical Data
Type S-Receiver M-Receiver
2 ccm coupler Ear simulator 2 ccm coupler Ear simulator
Output sound pressure level
at 1.6 kHz – 109 dB SPL – 123 dB SPLPeak 108 dB SPL 119 dB SPL 119 dB SPL 129 dB SPLHFA-OSPL 90 101 dB SPL – 113 dB SPL –Gain
Full on gain (FOG) at 1.6 kHz – 43 dB – 55 dBFull on gain (Peak) 45 dB 56 dB 60 dB 70 dBHFA-FOG 37 dB – 50 dB –Reference test gain 24 dB 34 dB 36 dB 48 dBFrequency, noise and directivity
Frequency range 100 - 10000 Hz 100 - 10000 Hz 100 - 9400 Hz 100 - 10000 HzEquivalent input noise 19 dB SPL 20 dB SPL 19 dB SPL 23 dB SPL
Total harmonic distortion at500 / 800 / 1600 / 3200 Hz 1 / 1 / 1 / 1 % 1 / 1 / 2 / - % 1 / 2 / 1 / 1 % 2 / 3 / 2 / - %
Tinnitus therapy broadband 65 dB SPL – 70 dB SPL –AI-DI 4.0 dB 4.0 dBInductive coil sensitivity
MASL (1 mA/m) at 1.6 kHz – – – –HFA MASL (1 mA/m) – – – –HFA SPLITS (left/right) – – – –RSETS (left/right) – – – –HFA SPLIV – – – –Battery
Battery voltage 1.25 V 1.25 VBattery current drain 1.2 mA 1.2 mA 1.4 mA 1.4 mABattery life (cell zinc air) – –Battery life (rechargeable) ~19 h ~19 hIRIL IEC 60118-13:2016 Ed. 4.0
700-960 MHz (rating) user user1400-2000 MHz (rating) user user2000-2700 MHz (rating) user userANSI C63.19-2011
800-950 MHz (rating) M4 M41600-2500 MHz (rating) M4 M4
TH Premium RIC LI 19 | 3
TH Premium RIC LI 19 | Technical Data
Type P-Receiver HP-Receiver
2 ccm coupler Ear simulator 2 ccm coupler Ear simulator
Output sound pressure level
at 1.6 kHz – 128 dB SPL – 137 dB SPLPeak 124 dB SPL 134 dB SPL 130 dB SPL 138 dB SPLHFA-OSPL 90 119 dB SPL – 123 dB SPL –Gain
Full on gain (FOG) at 1.6 kHz – 70 dB – 82 dBFull on gain (Peak) 70 dB 80 dB 75 dB 82 dBHFA-FOG 63 dB – 68 dB –Reference test gain 42 dB 53 dB 46 dB 62 dBFrequency, noise and directivity
Frequency range 100 - 7500 Hz 100 - 8100 Hz 100 - 7300 Hz 250 - 6100 HzEquivalent input noise 18 dB SPL 21 dB SPL 16 dB SPL 12 dB SPL
Total harmonic distortion at500 / 800 / 1600 / 3200 Hz 1 / 2 / 1 / 1 % 3 / 4 / 2 / - % 1 / 2 / 1 / 1 % 2 / 2 / 1 / - %
Tinnitus therapy broadband 75 dB SPL – 85 dB SPL –AI-DI 4.0 dB 4.0 dBInductive coil sensitivity
MASL (1 mA/m) at 1.6 kHz – – – –HFA MASL (1 mA/m) – – – –HFA SPLITS (left/right) – – – –RSETS (left/right) – – – –HFA SPLIV – – – –Battery
Battery voltage 1.25 V 1.25 VBattery current drain 1.3 mA 1.3 mA 1.3 mA 1.3 mABattery life (cell zinc air) – –Battery life (rechargeable) ~19 h ~19 hIRIL IEC 60118-13:2016 Ed. 4.0
700-960 MHz (rating) user user1400-2000 MHz (rating) user user2000-2700 MHz (rating) user userANSI C63.19-2011
800-950 MHz (rating) M4 M41600-2500 MHz (rating) M4 M4
TH Premium RIC LI 19 | 4
TH Premium RIC LI 19 | Fitting Range
S-Receiver M-Receiver
H
earin
g lo
ss
f
H
earin
g lo
ss
f
Open Click Domes+ Double Click Domes+ + Click Mold (no vent)
Open Click Domes+ Double Click Domes+ + Click Mold (no vent)
P-Receiver HP-Receiver
H
earin
g lo
ss
f
H
earin
g lo
ss
f
Double Click Domes+ Click Mold (no vent) Custom Shell (no vent)
TH Premium RIC LI 19 | 5
S-Receiver (Closed Click Dome) | Basic Data
2 ccm coupler Ear simulator
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Frequency response (LI = 60 dB)
f
L0
Basic acoustic response (LI = 60 dB)
f
TH Premium RIC LI 19 | 6
M-Receiver (Closed Click Dome) | Basic Data
2 ccm coupler Ear simulator
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Frequency response (LI = 60 dB)
f
L0
Basic acoustic response (LI = 60 dB)
f
TH Premium RIC LI 19 | 7
P-Receiver (Click mold) | Basic Data
2 ccm coupler Ear simulator
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Frequency response (LI = 60 dB)
f
L0
Basic acoustic response (LI = 60 dB)
f
TH Premium RIC LI 19 | 8
HP-Receiver (Custom Shell) | Basic Data
2 ccm coupler Ear simulator
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Output sound pressure level (LI = 90 dB)
Full on gain (LI = 50 dB)
f
L0
Frequency response (LI = 60 dB)
f
L0
Basic acoustic response (LI = 60 dB)
f
TH Premium RIC LI 19 | 9
TH Premium RIC LI 19 | Features and AccessoriesAudiology
Own Voice Processing (OVP) 1)
3D Classifier
Signal processing (channels) / Gain/MPO (handles) 48 / 20Hearing programs 6Sound Clarity
HD SpatialExtended dynamic rangeExtended bandwidthEchoShieldHD Music (presets) 3eWindScreen binaural 1)
eWindScreenNoise Management
Speech and noise management (steps) 7SoundSmoothing® (steps) 3Directional speech enhancement (steps) 3
Feedback cancellationSpeech Quality
DirectionalityNarrow Directionality 1)
Spatial SpeechFocus 1)
SpeechFocusTwinPhone1)
Frequency compressionDirect Streaming
Made for iPhone®
Adaptive Streaming Volume 2)
Tinnitus
Notched Noise TherapyTinnitus therapyFitting
Smart Optimizer and Data LoggingAcclimatization managerPerformance GuideInsitugramLearning (classes) 6TeleCare
TeleCare 3.0
1) req. bilateral fitting2) streaming only
available highest feature performance — not available
TH Premium RIC LI 19 | 10
TH Premium RIC LI 19 | Features and AccessoriesStyle specific features
Ingress Protection Rating IP68Charging contacts —Battery Size —Battery door on/off function —Nanocoated housinge2e wireless® 3.0User controls coupling via e2eWireless programmingInstrument configurations
Flat cover —Rotary volume control —Push buttonRocker switch —Color conversion kitBattery door – integrated telecoil —Battery door – child lock —Small earhook —Programming accessories
ConnexxAir, ConnexxLink —Noahlink™ wirelessProgramming adapter / cable —Accessories
miniPocket®
TH CROS RIC 19StreamLine TVStreamLine MicInductive Charger mandatoryApps
myControl™ ApptouchControl™ App
available optional — not available
TH Premium RIC LI 19 | 12
Abbreviations and Standards
Abbreviations
The following abbreviations are used in this datasheet:
Output Sound Pressure LevelOSPLHigh Frequency AverageHFAFull On GainFOGMagneto Acoustical Sensitivity LevelMASLCoupler SPL for an Inductive Telephone SimulatorSPLITSRelative Equivalent Telephone SensitivityRSETSSPL In a Vertical magnetic fieldSPLIVArticulation Index - Directivity IndexAI-DIInput Related Interference LevelIRILReference Test FrequencyRTF
Standards and additional information
u All measurements with the 2 ccm coupler were performed according to ANSI S3.22-2014 and IEC 60118-0:2015 if applicable.u All measurements with an ear simulator were performed according to IEC 118-0/A1:1994 and to DIN 45605
(frequency range) if applicable.u Curves and figures representing FOG are measured with 20 dB reduction and 70 dB SPL input level.u Figures representing Equivalent Input Noise incorporate a moderate expansion.u Inductive coil sensitivity values, inductive response curves and T ratings apply for instruments with telecoil battery door only.u Tinnitus therapy measurement conditions: all tinnitus single frequency sliders in max position, master volume slider in
default position (0 dB) and local volume control in default position.u The current consumption is measured in reference test setting (RTS) according to the applicable standards. Due to the
settling behaviour of hearing instruments supporting RF (radio frequency), the battery current is measured 3 minutes after turning on (note: no pairing).
u The battery life is based on first fit settings using 60% of the fitting range and an ISTS (International Speech Test Signal) input signal at 65 dB SPL (note: pairing established). The actual battery life is determined by battery quality, hearing loss, sound environment, usage and activated feature set.
u The following acoustic connections / ear pieces were used: - S-Receiver Unit and M-Receiver Unit: Closed Click Dome - P-Receiver Unit: Click Mold - HP-Receiver Unit: Custom Shell
Note for power cell
u Operating times may vary due to hearing loss, use of binaural features and accessories, age of power cell as well as the sound enviroment.
“Made for iPod”, “Made for iPhone”, and “Made for iPad” mean that an electronic accessory has been designed to connect specifically to iPod, iPhone, or iPad, respectively, and has been certified by the developer to meet Apple performance standards. Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards. Please note that the use of this accessory with iPod, iPhone, or iPad may affect wireless performance.
WarningChoking hazard posed by small parts.u This instrument is not intended for the fitting of infants, children under 3 years and persons of mental incapacity.
WarningInstrument has an output sound pressure level of 132 dB SPL or more.Risk of impairing the residual hearing of the user.u Take special care when fitting this instrument.
The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases and are subject to change without prior notice. The required features should therefore be specified in each individual case at the time of conclusion of the respective contract.
www.truhearing.com
Order No. 03802-99T1-7600, SI/18938-19© 09.2018, TruHearing Inc.All rights reserved
Manufactured forTruHearing Inc.12936 S. Frontrunner BlvdDraper, UT 84020United States
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 S Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240777 10/18 1.0 SI/18916-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
108 dB101 dB
Full-on gainPeakHF - averageReference test gain
45 dB37 dB24 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz10000 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%1%1%1%
Equivalent input noise 19 dB
Battery current drain 1.2 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 S Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240777 10/18 1.0 SI/18916-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
108 dB101 dB
Full-on gainPeakHF - averageReference test gain
45 dB37 dB24 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz10000 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%1%1%1%
Equivalent input noise 19 dB
Battery current drain 1.2 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 S Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240777 10/18 1.0 SI/18916-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
108 dB101 dB
Full-on gainPeakHF - averageReference test gain
45 dB37 dB24 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz10000 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%1%1%1%
Equivalent input noise 19 dB
Battery current drain 1.2 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 S Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240777 10/18 1.0 SI/18916-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
108 dB101 dB
Full-on gainPeakHF - averageReference test gain
45 dB37 dB24 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz10000 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%1%1%1%
Equivalent input noise 19 dB
Battery current drain 1.2 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 M Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240778 10/18 1.0 SI/18917-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
119 dB113 dB
Full-on gainPeakHF - averageReference test gain
60 dB50 dB36 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz9400 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 19 dB
Battery current drain 1.4 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 M Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240778 10/18 1.0 SI/18917-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
119 dB113 dB
Full-on gainPeakHF - averageReference test gain
60 dB50 dB36 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz9400 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 19 dB
Battery current drain 1.4 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 M Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240778 10/18 1.0 SI/18917-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
119 dB113 dB
Full-on gainPeakHF - averageReference test gain
60 dB50 dB36 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz9400 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 19 dB
Battery current drain 1.4 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 M Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240778 10/18 1.0 SI/18917-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with closed click dome.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
119 dB113 dB
Full-on gainPeakHF - averageReference test gain
60 dB50 dB36 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz9400 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 19 dB
Battery current drain 1.4 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 P Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240779 10/18 1.0 SI/18918-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with click mold.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
124 dB119 dB
Full-on gainPeakHF - averageReference test gain
70 dB63 dB42 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7500 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 18 dB
Battery current drain 1.3 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 P Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240779 10/18 1.0 SI/18918-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with click mold.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
124 dB119 dB
Full-on gainPeakHF - averageReference test gain
70 dB63 dB42 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7500 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 18 dB
Battery current drain 1.3 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 P Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240779 10/18 1.0 SI/18918-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with click mold.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
124 dB119 dB
Full-on gainPeakHF - averageReference test gain
70 dB63 dB42 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7500 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 18 dB
Battery current drain 1.3 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
TH Premium RIC LI 19 P Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240779 10/18 1.0 SI/18918-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with click mold.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
124 dB119 dB
Full-on gainPeakHF - averageReference test gain
70 dB63 dB42 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7500 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 18 dB
Battery current drain 1.3 mA
Battery life (typical) Li-Ion power cell up to 19 hrs.
TH Premium RIC LI 19 HP Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240780 10/18 1.0 SI/18919-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with custom shell.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
130 dB123 dB
Full-on gainPeakHF - averageReference test gain
75 dB68 dB46 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7300 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 16 dBBattery current drain 1.3 mABattery life (typical) Li-Ion power cell up to 19 hrs.
TH Premium RIC LI 19 HP Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240780 10/18 1.0 SI/18919-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with custom shell.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
130 dB123 dB
Full-on gainPeakHF - averageReference test gain
75 dB68 dB46 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7300 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 16 dBBattery current drain 1.3 mABattery life (typical) Li-Ion power cell up to 19 hrs.
TH Premium RIC LI 19 HP Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240780 10/18 1.0 SI/18919-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with custom shell.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
130 dB123 dB
Full-on gainPeakHF - averageReference test gain
75 dB68 dB46 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7300 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 16 dBBattery current drain 1.3 mABattery life (typical) Li-Ion power cell up to 19 hrs.
TH Premium RIC LI 19 HP Receiver Receiver-in-Canal Hearing Instrument
Hearing Instruments made in Singapore.10240780 10/18 1.0 SI/18919-18
All data specified were determined under test conditions which comply with the Specifications of Hearing Aid Characteristics ANSI S3.22 — 2014.Hearing aid test settings according to the test mode, selectable from the CONNEXX® fitting menu, configures the instrument for full-on gain, no compression and all adaptive signal analysis and processing turned off. SPLITS (Sound Pressure Level for
Inductive Telecoil Simulator) Battery life stated is measured at 65 dB input
and reference test gain. Actual battery life depends on the output level. All tests performed with custom shell.
Output Sound Pressure LevelANSI S3.22-2014
Frequency ResponseANSI S3.22-2014
OSPLL
I= 90dB
FOGL
I= 50dB
RTG L
I= 60dB
Standard ANSI S3.22 - 2014 2 ccm coupler
Output Peak OSPL 90HF - average OSPL 90
130 dB123 dB
Full-on gainPeakHF - averageReference test gain
75 dB68 dB46 dB
Frequency range
Low frequency limitHigh frequency limit
100 Hz7300 Hz
Total harmonic distortion
500 Hz800 Hz1600 Hz3200 Hz
1%2%1%1%
Equivalent input noise 16 dBBattery current drain 1.3 mABattery life (typical) Li-Ion power cell up to 19 hrs.