Page 1: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Thank you for viewing this presentation.

We would like to remind you that this

material is the property of the author.

It is provided to you by the ERS for your

personal use only, as submitted by the

author.

2016 by the author

Page 2: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

OSCILLOMETRY (FOT and IOS) definition and application

PG11 Assessing respiratory system resistance in children and adults

Prof. Waldemar TOMALAK

Dept. Physiopathology of Respiratory System,

Institute for TBC & Lung Dis. Rabka Branch, Poland

Page 3: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

I declare no conflict of interest related to thispresentation

Page 4: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

The forced oscillation technique• Introduced in 1956 by A. DuBois and coworkers:

„Oscillatory mechanics of the lung and chest in man” J. Appl. Physiol. 1956; 8:587-594.

• The forced oscillation technique uses external pressuresignals (generated by loudspeaker) to evoke flowresponse of the respiratory system. Pressure and resulting flow are then analysed yielding to parametersof oscillatory mechanics (resistances and reactances orparameters of the model of respiratory system).

Page 5: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

The forced oscillation technique – twoapproaches

• „classic” (FOT) – with the excitation wavebeing the sum of several sinusoidal waves

• Impulse oscillometry (IOS) – using pressurepulses

Page 6: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

The forced oscillattion technique – the principle

Z – respiratory impedance; R – resistance; X - reactance

Page 7: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Multifrequency oscillations

………….

Typical frequency range: 2..4 – 32 Hz

Page 8: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

„classic” FOT realisations

Pulmosfor (SEFAM)

Ros (Sensormedics)

Cosmed

Page 9: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

The measurements• Oscillatory measurement are distorted by the so called

upper airways shunt (influence of the impedance of the cheeks and mouth floor). Therefore, the subjecthas to support cheeks with palms during the measurements.

• Interesting modification was put into practice with PULMOSFOR (SEFAM, F) where the pressure wave was applied around the head rather than at the mooth (the head generator technique) to minimize this.

Page 10: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Head generator of PULMOSFOR

Page 11: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Analysis of data

Data can be interpreted as values resistance or reactance at given frequency (e.g. R10, X10) or using a simple model of respiratory system with frequency dependent resistance Rrs, compliance od respiratory system Crs and respiratory system inertanceIrs. Resonant frequency Fn is a frequency where Xrs equals 0

Page 12: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Correlation of FOT indices with classicmeasurements

W. Tomalak et al. Pneumonologia i Alergologia Polska, 1995

Page 13: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Reference values for FOT

Many publications; especially for children. For different age ranges. Some examples:

• Solymar L et al. Pediatric Pulmonology 1985; 1; 134-140.

• Helinckx J. Eur. Respir J. 1998; 12:438-443

• Mazurek H et al. Pediatric Pulmonology 2000; 30: 47-55.

Page 14: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Reference values for FOTAdults: • E. Oostven et al. Respiratory impedance in healthy

subjects: baseline values and bronchodilator response.ERJ 2013, 42:1513-1523.

• Age range: 18-80; frequency range: 4-26 Hz; number of subjects: 368.

• Suggestion: „new reference equations based on different setups are recommended to replace those established with a single device.”

Page 15: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory
Page 16: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory
Page 17: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

IOS

Present on the market since 1994 – launched by Jaeger company (Vogel J, Smidt U. Impulse oscillometry. Frankfurt: PMI Verlagsgruppe; 1994.)

The realisation of measurements using IOS (impulse oscillometry system) usespressure pulses instead of sum of several sinusoidal waves.

The analysis of data is the same as in classic FOT realisations .

Page 18: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory
Page 19: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

IOS – the measuring head

loudspeaker

pneumotach

V’ Pm

Page 20: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

An example of IOS measurement

Page 21: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

IOS

• The data are analysed as resistances and reactances at different frequencies (5, 10, 15, 20, 25, 35 Hz)

• The results may be also analysed with built-in 7-element model (not frequently used)

Page 22: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Correlation of IOS indices with classicmeasurements

Correlation of R5 with FEV1: r=-0.66 (the same study)

W. TOMALAK et al. Pediatric Pulmonology 2006.

Page 23: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Reference values for children (IOS)

B. Nowowiejska, W. Tomalak et al. Pediatric Pulmonology 2008

Page 24: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Reference values for adults (IOS)

Page 25: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

ERMonograph 31, 2005

Page 26: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Reliability of FOT (IOS) data

• As natural breathing may interfere with excitationpressure waves (especially in lower frequencies) a coherence function values (calculated from power spectra of flow and pressure signals) arecalculated for each frequency of interest. The value of coherence is within 0 and 1; the higherthe value the ‚better’ is the measurement.

Page 27: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Reliability of FOT (IOS) data

• It is assumed that for classic measurementsthe value of coherence should be 0.95 to have the coefficient of variation of impedancedata <10% (Landser et al., 1976.)

• For IOS coherence >0.6 at 5 Hz is anacceptable threshold (Smith et al., 2005.)

Page 28: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Difference between IOS and FOT (n=79 subjects)f = 5 Hz f = 10 Hz f = 20 Hz

RIO

S[

hPa

-1ls

]

(RPULM + RIOS)/2 [ hPa-1ls ]

RIO

S-

RP

ULM

[ h

Pa-1

ls ]

RPULM [ hPa-1ls ]

J. Radliński, W. Tomalak et al. Comparison of measuring system used to evaluate input respiratory impedance by forced oscillation technique – pilot studyEur. Respir. J 2002; 20; suppl 38; 32

Page 29: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Difference between FOT and IOS

f

5

10

20

0.82 (***)

0.81 (***)

0.74 (***)

Correlation

0.56 ± 0.91

0.37 ± 0.65

0.22 ± 0.55

0.24 (*)

0.24 (*)

0.16 (NS)

0.16

0.17

0.12

differenciesmean±sd R slope

BA graphs

R

The analysis showed high correlation between PULMOSFOR and IOS resistances; but also systematic differences at lower frequencies, which can be attributed to differencesin pressure signal energy.

J. Radliński, W. Tomalak et al. Comparison of measuring system used to evaluate input respiratory impedance by forced oscillation technique – pilot studyEur. Respir. J 2002; 20; suppl 38; 32

Page 30: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Oscillometric measurements in obstructive diseases

Page 31: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Impedance data in obstructive diseases.

DA Kaminsky. Resp Care 2012; 57:85-96

Page 32: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

IOS (FOT) allows measurements of inspiratory and expiratory resistances

Page 33: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Main advantages of FOT and IOS

• Requires only passive cooperation (natural breathing)

• Fast, easy to perform and reproducible

• Useful especially in children (from 2 yrs of age.), olderadults and patients unable to perform spirometry

• Useful in bronchodilator tests and bronchial challenges

• Standardisation documents

• Allows analysis of respiratory resistance in real time

Page 34: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Performance of IOS/FOT in youngestchildren

• IOS: Own experience: among 133 children aged3-4 years, 6 were unable to perform the measurements; 27 refused to make examination. Overall cooperation succes rate: 61%

• Spirometry: (also own experience) age group 3-4 yrs. Cooperation yielding to interpretable results: 3.6%

Nowowiejska B et al. Pediatric Pulmonology 2008.Tomalak W et al. Pneumonologia i Alergologia Polska 2008

Page 35: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

IOS (FOT) in assessing small airwayproperties

Page 36: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Bickel S et al. Chest 2014; 146:841-847

AX and the difference R5-R20 areregarded as small airway status indices

Page 37: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

R5-R20/R5 15% 33% 39% 44%

Page 38: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

277 persons aged 65-97 yrs; 197 with obstruction(FEV1/FVC<LLN)

(R5-R20)/R5 in patientswithout obstruction:

33%

(R5-R20)/R5 in patients with obstruction (mainly COPD)

50.7%

Page 39: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

Application of IOS (FOT)• COPD;• Asthma• Bronchopulmonary displasia• OSA• Central airways obstruction• Adult interstitial lung diseases• Occupational and environmental exposure• Response to bronchodilators;• Bronchoprovocation testing

Page 40: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

DA Kaminsky. Resp Care 2012; 57:85-96

Page 41: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

To conclude… FOT and/or IOS measurements:• Are fast and easy to perform (for the patient, as well as

for the technician)• Need only passive cooperation (tidal breathing)• Are reproducible (although variability is greater than in

spirometry, is comparable to other techniques of respiratory resistance measurements)

• Can be used in bronchodilator and bronchoprovocationtests

Page 42: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

To conclude…

• Can be used in patients from 2..3 years of age up to 100 or more (?)

• There exist several reference values sets (much morefor children than for adults) for both approaches

• There exist recommendation for performing measurements (ERJ 2003; ERM 2005)

• Offers new possibilities for studying respiratory physiology and pathology

Page 43: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory

LIMITATIONS

• Much less popular (than spirometry for exampleaccording to Cosmed representative >50 RESMONs sold (excluding US); on the other hand, according to Jaeger representatives, >9500 IOS devices were sold throughout the world)

• Difficult theoretical background

• Different models used; not completely verified

Page 44: Thank you for viewing this presentation. We would like to remind … · 2016. 8. 11. · The forced oscillation technique •Introduced in 1956 by A. DuBois and coworkers: „Oscillatory