The Blue Book The Blue Book P f Ji Ml Published, 1988, by predecessor of Prof Jim Malone Robert Boyle Prof of Medical Physics Trinity College, Dublin Published, 1988, by predecessor of Radiological Protection Institute of Ireland (RPII) IAEA, Vienna
Transcript
Page 1: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

The Blue BookThe Blue Book 

P f Ji M l

Published, 1988, by predecessor of

Prof Jim MaloneRobert Boyle Prof of Medical Physics

Trinity College, DublinPublished, 1988, by predecessor of Radiological Protection Institute of Ireland (RPII)

y g ,

IAEA, Vienna

Page 2: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Problems with Blue BookProblems with Blue Book

• Changes in Legislation,  • OBJECTIVESg gTechnology, Clinical Practice, Building Style, Building Materials • Preference for 

h i l l• Changes in Dose Limits and Constraints

comprehensive local solution

• Use of Upper Floors• Need illustrated floor plans

• Not to innovate, but to produce a reliableNeed illustrated floor plans

• Advice on Ceilings, • Higher levels of walls

to produce a reliable practical manual or code.g

• Practical Tips and Solutions

Page 3: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Issues 1: EquipmentIssues, 1: Equipment

Page 4: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Issues 2: New ProblemsIssues 2: New Problems

W d ll t lid• Ward walls not solid• Theatre workloads not consistent with modern practice.

• Other – recovery rooms, endo suites lithotripsyendo suites, lithotripsy, cardiac pacing

• Radionuclides in Theatres• PET shielding

Page 5: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Issues 3: Dose Constraints

Category of P l

Dose C t i t

Dose C t i tPersonnel Constraint

1998mSv/year

Constraint 2001

mSv/year

Exposed 5 0 1 0Exposed Worker

5.0 1.0

All others 0.05 - 1.0 0.3

Page 6: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Issues 4: Medical PhysicsIssues 4: Medical Physics

Medical Physics SupportMedical Physics Support 

• Deficits in numbers, leadership and d i i l / i iacademic involvement/connectivity

• Often poor, or inadequate training, and narrow or inadequate experience

• Risk of litigation and difficulties with public accountabilityaccountability

Page 7: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,
Page 8: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Design Code (2nd edition)Design Code (2nd edition)

• About 100 pages incl. Appendices

1. Legal and Administrative Framework

2 Radiation Protection Projectpp 2. Radiation Protection, Project Management and Building Projects

3 R di l R D i d3. Radiology Room Design and Layout

4. Nuclear Medicine5. Shielding Calculations6. Some Practical Considerations

www.rpii.ie RPII, 3 Clonskeagh Square, Dublin 14, Ireland

Page 9: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

1. Statutory Framework1. Statutory Framework

• The Radiological Protection Act, 1991 (Ionising Radiation) Order 2000 (SI No. 125 of 2000)

E C iti (M di l I i i R di ti• European Communities (Medical Ionising Radiation Protection) Regulations 2002, 2007 (SI. No. 478 of 2002 & SI. No. 303 of 2007)

• RPII Licensing System and Requirements

• Related EU Directives

Page 10: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

2. Radiation Protection, Project d ildi jManagement and Building Projects

• The Radiation Safety Committee

• The Radiation Protection Advisor

• Project Teams, New Building Design Cycle, Refitting Buildings

• Dose Limits and Dose Constraints

• Risk Assessments

• Site visits essential

Issue of New Build versus Conversion/Refit

Page 11: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Radiology Room Design and Layoutgy g y

• Radiology Room Types

• Fluoroscopy Rooms– General Fluoroscopyyp

• General Comments on Shielding

– Interventional Radiology and Cardiology

• Radiography Rooms– General

gy• CT• Shared Function

– Chest Room– Mammography

Rooms (A&E, Theatres, wards)

• Equipment in Trailers– DXA– Dental

• Equipment in Trailers

Page 12: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

3 Radiolog Room Design and La o t3. Radiology Room Design and Layout

General RadiologyTwo-corridor DesignLarge enough for trolleys table andLarge enough for trolleys, table and chest radiologyTypical room sizes givenGenerally 2mm Lead – assess on individual basisPrimary Beam absorberStaff entrance behind protective screenTypical screen lengths presented Chest stand positioned to minimise

tt t i t ti lscatter entering protective consoleChanging cubicles

Specific Requirements for each type of facilityExamples of Good Layouts

Page 13: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

3 Radiolog Room Design and La o t3. Radiology Room Design and Layout

Dental SurgeryNo shielding required if: < 20 exps/wk

CTSeparate staff area – Other staff present

and 2m between patient and all boundaries Need good view of door and patientScanner angled for access and visibility

MSCT: 3-4mm Pb

Page 14: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

DXA and CTDXA and CT 

Page 15: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Design C ite iaDesign Criteria

orC

orrid

o

Corridor

CorridorCorridor

Page 16: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

9/3/2009

Page 17: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Mobile X‐Ray Equipment

Page 18: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Equipment not in RoomsEquipment not in Rooms

Page 19: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Equipment not in RoomsEquipment not in Rooms

Page 20: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Equipment not even in BuildingEquipment not even in Building

• Re‐shielding often • TrailerRe shielding often means complete refit

• Hospital can’t do 

Trailer

pwithout equipment for 6 months

• Trailer arrives ‐‐‐‐‐

Page 21: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

5 Shielding Calculations5. Shielding Calculations

X‐Ray• Review of two widely used 

Shielding MethodologiesShielding Methodologies– BIR, 2000– NCRP, 2004

• Variables– Distance from Barrier– Workload

• BIR: Workload is based  (ESD) and (DAP)

• NCRP: Workload based on “beam– Occupancy • NCRP: Workload based on  beam‐on” time. in mA min per week

Issue of New Build versus Conversion/Refit

Page 22: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,
Page 23: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

WorkloadWorkload

Page 24: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

WorkloadWorkload

• Either DAP (BIR) or mA‐min/week (NCRP)Either DAP (BIR) or mA min/week (NCRP)

• Try and base figures on real audit/projection

i i l bli h d d i l d• Historical or published data can mislead

• If no other option, use published data

• Transparency and accountability to the publicTransparency and accountability to the public

• Defendable legally (reasonable patient, not reasonable doctor)

Page 25: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Occupancy

Page 26: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

OccupancyOccupancy

• Occupancy of adjoining areas to be assessedOccupancy of adjoining areas to be assessed

• Try and get real information

C id h id f id• Consider rooms on other side of corridor

• Extremes: Office, 100%;   Unattended car‐park, 2.5 to 5 %

• Reservation about NCRP door value in new builds, and Remember:

• Transparency and accountability to the publicTransparency and accountability to the public

• Defendable legally (reasonable patient, not reasonable doctor)

Page 27: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

General Radiographic Room Ceiling (BIR method)

Page 28: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

General Radiographic Room Ceiling (NCRP method)

General Rm Window scatter only at 10 mGeneral Rm Window, scatter only, at 10 m

Page 29: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

6. Practical Considerations

• Building Materials– Lead sheet and lead 

products

• Walls• Floors and Ceilings• Doorsproducts

– Concrete and concrete BlocksBarium Plaster

• Doors• Windows• Staff Areas– Barium Plaster

– Brick– Gypsum Wallboard

• Joints, Services, Openings and Perforations

• Assessment of Shielding– Lead Glass– Lead Acrylic

Assessment of Shielding• Nuclear Medicine

ALWAYS NEED TO VISIT AND SEE IMPLEMENTATION

Page 30: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

9/3/2009

Page 31: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Lead CodesLead Codes

Code No. 3 4 5 6 7 8

Nominal Thickness (mm)Nominal Thickness (mm)1.32 1.80 2.24 2.65 3.15 3.55

Weight (kgm‐2) 14 6 19 5 24 4 29 3 34 2 39 1Weight (kgm ) 14.6 19.5 24.4 29.3 34.2 39.1

Cost (Relative)

• When installed as part of a new build, lead is not very dear relative to other costs

Page 32: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Some Data 2:Some Data 2: 

Page 33: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Some Data Figure C 6Some Data Figure C 6

• Also f room; Some Data Table C2Also f room; Some Data Table C2 and C3,4,or Fig C6 and page 101

Page 34: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,
Page 35: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

Issues

• Advice for imaging facilities located on upper floors

• Advice for shielding of windows on upper ppfloors

• Transparent,

bl• Accountable 

• Defendable

Page 36: The Blue Book - Shielding Code of Practice in … Blue...The Blue Book PfJi Ml Published, 1988,by predecessor of Prof m Malone Robert Boyle Prof of Medical Physics Trinity College,

CONCLUSIONCONCLUSION