0 CMC Strategy Forum Japan 2017 Tokyo The development case of Combination product in Japan “Pre-filled Syringe / Auto-Injector for SC Injection” Seiji Ueno, Ph.D. Deputy Department Manager Production Engineering dev. (Formulation Technology) Chugai Pharmaceutical Co., Ltd. Dec. 5 th , 2017
Transcript
Page 1: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

0CMC Strategy Forum Japan 2017 Tokyo

The development case of Combination product in Japan

“Pre-filled Syringe / Auto-Injector for SC Injection”

Seiji Ueno, Ph.D.Deputy Department ManagerProduction Engineering dev.(Formulation Technology)Chugai Pharmaceutical Co., Ltd.

Dec. 5th, 2017

Page 2: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

1CMC Strategy Forum Japan 2017 Tokyo

Outline

Environmental Analysis of Combination Product Development

Development History in case of Chugai Product Development for SC Injection of ACTEMRA®

Target Product Profile (TPP) Key point of Success Challenge to Problems

• Formulation• Process• Container & Device• Regulation

Look back to development• Point to note for Development of “Combination products that

correspond to drugs”

Page 3: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

2CMC Strategy Forum Japan 2017 Tokyo

Environmental Analysis of Combination Product Development

AI: Autoinjector

30% ⇒ 48.4%

1

23

4

5 6US

Device

Sales trend of PFS, AI, and pen

24% ⇒ 30%

JP

Device

1

2

3

4

5

6

1

2

3

45

6

World Trend of Parenteral Injection

Self-injection CPs such as AI and pen grow consistently. Specifically, the US market shows drastic increase in sales.

Traditionally, Japanese pharmas were reluctant to develop CPs, but these days steadily increases. Device:Combination products as AI,Pen,Syringe

1

2

3

45 6

AI, pen, syringe

0

10,000

20,000

30,000

40,000

50,000

60,000

70,000

80,000

90,000

2006 2008 2010 2012 2014 2016

Annu

al sa

les o

f AI/

PEN

(mio

USD

)

Series1

Series2

Series3 2006 2016

2006 2016

Source : Based on IMS MIDAS 2006-2016Copy Right @ QuintilesIMS2017Reprinted with permission

Page 4: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

3CMC Strategy Forum Japan 2017 Tokyo

Environmental Analysis of Combination Product DevelopmentIn case of Japanese Trend of Parenteral Injection

Some of the major drivers for the growth of combination products in the Japanese market are an increase in syringes and decrease in ampoules.

Although shares of Pens and AIs slightly increased over the last 10 years, the shares initiated to rise recently. Specifically, AIs boosted its share in 2014.

A combination-product development race, which has been commonly observed in the competitive market such as Rheumatoid Arthritis in the USA and Eus since 2000, just gets started in Japan,

Source : Based on IMS MIDAS 2006-2016Copy Right @ QuintilesIMS2017Reprinted with permission

(Calculated form NDB open data (receipt data) of Ministry of Health, Labor and welfare,)

Page 5: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

4CMC Strategy Forum Japan 2017 Tokyo

Development History of Combination Product

2005, Launch First domestic antibody product

Indication : Castleman's disease

<Vial product for IV>antibody conc. : 20 mg/mL

2008, Additional indicationIndication : Rheumatoid

Arthritis (RA)

2013, Additional dosage form Indication : RA

2015, PFS barrel materialchange (glass⇒COP)

Indication : RA

<PFS product for SC><AI product for SC> antibody conc. : 180 mg/mL

In case of Chugai Product “ACTEMRA® “

Page 6: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

5CMC Strategy Forum Japan 2017 Tokyo

Development for SC Injection of ACTEMRA®

Indication Rheumatoid Arthritis (RA)

Dosage form PFS, AIAdministration route SC (max. 1mL)Antibody conc. min. 180mg/mL Administration frequency Once in two weeksQuality (Spec. and stability)

Same or better quality of IV product which already in market

< Target Product Profile (TPP) >

PFS/AI are desirable for QOL improvement of RA patient.

RA is chronic disease. RA patient has physically disabled hands. IV product with long treatment time (long infusion time) become a big burden to

RA outpatient.

Target Product Profile (TPP)

Page 7: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

6CMC Strategy Forum Japan 2017 Tokyo

Development for SC Injection of ACTEMRA®

Regulation

Formulation

ProcessContainer

&Device

Stable formulation was found in high concentrationof drug solution(180mg/mL).

Stable process was found in high concentration / highviscosity of drug solution(180mg/mL).

PFS & AI design was discussed with container and device suppliers, and quality improvement was achieved.

(Before amendment of the Act) License of “device marketingauthorization holder” and ”device manufacturer” was gotten by ourselves.

Key point of Success

Page 8: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

7CMC Strategy Forum Japan 2017 Tokyo

Development for SC Injection of ACTEMRA®

1. Formulation Aggregation with high conc. antibody (180 mg/mL)

Large-scale stabilizer screening was performed, and stable formulation (same or better quality of IV product which already in market) was found.

2. Process High viscosity & gelation with high conc. antibody (180 mg/mL)

Appropriate condition was found in ultrafiltration (UF) process.

3. Container & Device AI design (Quest for safety & certainty) Improvement of AI injection time Change of PFS Barrel Material (glass⇒COP)

4. Regulation AI product (before & after change of the Act)

Challenge to Problems

Page 9: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

8CMC Strategy Forum Japan 2017 Tokyo

Challenge to ProblemsFormulation Development (Overall Image)

(F1) Buffer optimization (Species, Conc, pH)

(F3) Surfactant optimization (PX188, PS20, PS80)

(F4) Stabilizer optimization (Arg, His, etc.)(F4-1) Optimization main stabilizer(F4-2) Optimization of subsidiary stabilizer

(F5) Stability of candidate formulations

(F2) Stabilizer screening(F2-1): Main stabilizer selection(F2-1): Subsidiary stabilizer selection

Candidate stabilizer (Arg, His, others)

Surfactant (species, conc.)

Candidate formulations

Buffer (species, conc., pH)

Challenge to Problems

Page 10: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

9CMC Strategy Forum Japan 2017 Tokyo

• < Stabilizer screening using design of experiments (DoE) >

⊿Di mer

-0. 14

-0. 12

-0. 1

-0. 08

-0. 06

-0. 04

-0. 02

0

0. 02

0. 04

0. 06

0. 08

Arg

Buf

fer

Arg

+ Bu

ffer

Met

Arg

+Me

t

Buf

fer

+Met

Ser

+Tr

p

Ser

Arg

+Se

r

Buf

fer

+Ser

Met

+Tr

p

Met

+Se

r

Buf

fer

+Trp

Arg

+Tr

p

Trp

His

Arg

+Hi

s

Lys

Arg

+Ly

s

Gly

Arg

+Gl

y

Pro

Arg

+Pr

o

Phe

Arg

+Ph

e

Coef

fiti

ent

Coef f i ci ent

*; P<0. 05, **; P<0. 01

*

*

**

**

****

*

**

**

*

Level of statistical significance: P<0.01

⊿ DimerHisbuffer

100mMArg 10mM

Met10mM

Trp 100mMHis

100mMPro

Candidate stabilizer is selected.Main: Arg

Sub:Met, Trp

Challenge to ProblemsFormulation Development (Example of Study Result (1))Challenge to Problems

Page 11: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

10CMC Strategy Forum Japan 2017 Tokyo

< Stabilizer optimization>

Arg

Met

Arg+Met(10,30,50mM)

Arg+Met(30mM)+Trp(10mM)

i.v.

Target Quality Same or better stability of IV product Isotonicity (Osmotic pressure ratio: 1)

Challenge to ProblemsFormulation Development (Example of Study Result (2))

(Target)

Only Arg: more than 300 mM (Osmotic pressure ratio: more than 2) Combination: Arg 100mM + Met 30 mM (Osmotic pressure ratio: 1)

Challenge to Problems

Page 12: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

11CMC Strategy Forum Japan 2017 Tokyo

Syringe

PFS product AI product

Challenge to ProblemsContainer & Device Development

< Composition of PFS/AI product for SC Injection of ACTEMRA®>

partsAccessary

Challenge to Problems

Page 13: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

12CMC Strategy Forum Japan 2017 Tokyo

< Quest for safety and certainty > Needle-stick accident-protected feature

– Injection needle is covered by safety cover when remove the cap.– Safety cover is locked when AI is removed from site of administration.

Malfunction-injection-protected feature– Injection button will be stand by when press safety cover to site of

administration. Needle is automatically stick and discharge drug solution when injection button is pressed.

Challenge to ProblemsContainer & Device Development (AI design)

For self-administration of RA patient with disabled hands, Better grip on body and cap Easy cap removal

WindowWindow Injection button

Safety coverStopper

Cap

Before use

After use

Challenge to Problems

Page 14: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

13CMC Strategy Forum Japan 2017 Tokyo

< Instruction for use>

Challenge to ProblemsContainer & Device Development (AI design)

(1) Ready for using auto-injectorRemove the cap from auto-injector. Grip the body of injector firmly and lightly turn the body and pull on the side-way while touch the anti-slip aria .Caution!: Remove the cap just before use.It is important to inject immediately after removing cap. Drug solution will be solidified when the cap is left opened.

(2) Ready for PositionHold injection site and lift your skin to ready for position auto injector.Caution!: Do not hold auto injector opposite side.

(3) PositionPush auto injector firmly against your skin at 90 degree angleand hold the position until the purple safety cover disappear.Caution!: Do not touch or press the injection button while push auto injector against your skin. You will no be able to press the button to start your injection until push firmly against your skin at 90 degree angle.

Challenge to Problems

Page 15: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

14CMC Strategy Forum Japan 2017 Tokyo

Challenge to ProblemsContainer & Device Development (AI design)

(4) Press buttonYour will hear a 1st “click” as you press the button. Automatically beginning to be delivered.Caution!: Keep holding auto injector firmly against your skin and make sure it stick to your skin.

(5) Count 15 secondRemove your finger from the injection button after press. You will hear 2nd “click”. The 2nd “click” means the lock on the safety cover and is not complete injection. Count 15 second until complete injection. Your will see the purple indicator in the view window lower down and the purple indicator will fill the viewing window.

(6) Lifting needleAfter lifting auto injector from your skin, release your finger from injection site.Caution!: Used auto injector should be placed into the case and bring to your health care institution.

< Instruction for use (continued) >

Challenge to Problems

Page 16: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

15CMC Strategy Forum Japan 2017 Tokyo

Challenge to ProblemsContainer & Device Development (Improvement of AI injection time)

RWN TWN

RWN: Regular wall needle, TWN: Thin wall needle

0.0

20.0

40.0

60.0

80.0

100.0

120.0

1 2

AI In

ject

ion

time

(%)

To improve AI injection time,overbore needle

Improvement of certainty for administration

Viscosity in higher due to high concentration of antibody

Antibody conc. (mg/mL)

AIInjection

time

Needle

20min

15min

Challenge to Problems

Page 17: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

16CMC Strategy Forum Japan 2017 Tokyo

Challenge to ProblemsContainer & Device Development (Change of PFS Barrel Material)

<Excellent feature of COP> Durable (high intensity) Lower dimensional tolerance (highly molding accuracy) Needle can be fixed on COP resin by itself (without glue)

PFS barrel was changed from glass to COPfor safer usage.

Challenge to Problems

Page 18: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

17CMC Strategy Forum Japan 2017 Tokyo

Challenge to ProblemsRegulation

< Handling of combination product with amendment of the Act>[Application for approval of combination product] (2014/10/24 drug food review division 1024/2)

Definition of combination product was clearly stated. Malfunction report is required on mechanical part of sub formation

by pharmaceutical license holder.

(Quoted from the 49th General Assembly Materials of the Pharmaceutical Association of the Ministry of Health, Labor and Welfare / Pharmaceuticals Evaluation Committee PMS Subcommittee)

Combination products that correspond to drugs:Pharmaceutical license holder is responsible for design control of mechanical part of medical device.

Challenge to Problems

Page 19: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

18CMC Strategy Forum Japan 2017 Tokyo

< Drug > Chugai

< Device > Chugai

License for drug marketing authorization holder

License for drug manufacturer

GMP inspection for drug manufacturer

License for device marketing authorization holder

License for device manufacturer

QMS inspection for device manufacturer

PMDA

Device application(Brand name: ACTO-Pen)

Drug application(Brand name: ACTEMRA auto injector)

Report ofadverse drugreaction

(AI Design developer)(AI parts manufacturer)

Supplier Aprocurement

(unclear !!)Report ofdevice malfunction

use

Regulation (AI product : before amendment of the Act, ~2014)

Challenge to Problems

Page 20: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

19CMC Strategy Forum Japan 2017 Tokyo

< Combination Product > Chugai

< Device > Chugai

License for drug marketing authorization holder

License for drug manufacturer

GMP inspection for drug manufacturer

License for device marketing authorization holder

License for device manufacturer

QMS inspection for device manufacturer

PMDA

Device application(Brand name: ACTO-Pen)

Drug application(Brand name: ACTEMRA auto injector)

Report ofadverse drugreaction

Report ofdevice malfunction

appropriate procurement controlunder GMP

(AI Design developer)(AI parts manufacturer)

Supplier A

Regulation (AI product : after amendment of the Act, 2015~)

Challenge to Problems

Page 21: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

20CMC Strategy Forum Japan 2017 Tokyo

Look back to developmentPoint to note for Development of “Combination products that correspond to drugs”

Regulation

Formulation

ProcessContainer

&Device

Very important !!Relationship with PFS & device supplierto achieve patient-oriented design.

Pharmaceutical license holder is responsible for design control of mechanical part of medical device.This means we need to catch up regulatory requirement of PFS & device in order to get approval.

< Pharmaceutical license holder point of view >

Very important !!Relationship with PFS & device supplier

Very important !!Relationship with PFS & device supplierfor maintenance and improvement of product quality after launch.

Look back to development

Page 22: The development case of Combination product in Japan “Pre ... · “Pre-filled Syringe / Auto -Injector for SC Injection ... Self-injection CPs such as AI and pen grow consistently.

21CMC Strategy Forum Japan 2017 Tokyo

Thank you very much !!