The Impact of Sample Containers on Large Particle Count for CMP Slurries Brian Kim, PhD Fujimi Corporation R&D 2017 CMPUG Spring Meeting April 13 th , 2017, Portland OR 1
Transcript
Page 1: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

The Impact of Sample Containers on Large

Particle Count for CMP Slurries

Brian Kim, PhD

Fujimi Corporation R&D

2017 CMPUG Spring Meeting

April 13th, 2017, Portland OR

1

Page 2: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Overview

• Problem statement

• LPC sources and measurement

• Sample bottles and its evaluation

• Identification of the sources of particles

• Summary

2

Page 3: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Problem Statement

Problem: A larger variation of LPC was observed for slurry A containing

surfactant when sample bottles w/ cap liner was used

3

Page 4: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

LPC and LPC Sources

• Large Particle Count (LPC) in CMP slurries

• Particles >0.5 um

• LPC is typically measured to predict defects

• Possible sources of LPC:

• In the slurry: large particles, agglomerates, micelles

• Extraneous sources: contamination

• In this study LPC contribution from sample bottles was identified and

reduced

4

Page 5: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Variations in LPC Measurement

• Sources of variations in LPC measurement

• Taking samples

• Handling samples

• Preparing samples

• Stability of LPC measurement tool

• Previous LPC studies at Fujimi:

• Settling and re-dispersion of large particles

5

Page 6: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

LPC as a Function of Depth in Sample Bottle

• Settling of large particles and its re-dispersion of particle was demonstrated

• Sample preparation right before LPC measurement is important to

obtain an accurate measurement

0

2

4

6

8

10

12

14

0 1000 2000 3000 4000

Dep

th in

sam

ple

bo

ttle

(cm

)

Slurry A: LPC >0.56 m

0 day

22 day, before mixing

22 day, after mixing

5466

0.5 L bottle

Set up for LPC measurement

To Particle Counter

14 cm

450 mL

Set up for LPC measurement

To Particle Counter

“Settling of Colloidal Silica Particles in CMP Slurry: Monitoring, Effect, and Handling” J. Lin, W.S. Rader, CSTIC 2017, March 12-13, 2017, Shanghai, China.

6

Page 7: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Bottles Used

• Bottles tested

• Bottle w/o cap liner (Bottle A)

• Bottle w/ cap liner (Bottle B)

Bottle ABottle B

HDPE

PP

HDPE: high density polyethylenePP: polypropylene

7

Page 8: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Slurry for This Study

• Slurry A

• HVM slurry: colloidal particle / surfactant

• Slurry at pH 10

• A typical LPC measurement variation of slurry A including variations

by LPC measurement operator and date

8

Particle size (um) Particle counts

>0.56 +/- 343

>0.99 +/- 50

Page 9: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Sample Handling

Operator Bottle handling method

Operator 1 Bottles tipped over

Operator 2 Bottles kept upright from the production line to

QC for LPC measurement

9

• Two operators collected samples directly from a production line and

delivered to Quality Control (QC)

Page 10: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Sample Preparation for LPC Measurement 10

• Sample preparation before LPC measurement

• Slurry samples were shaken on an orbital shaker for 3 min right

before sampling

• Sampling from bottle during the measurement

• Instrument for LPC measurement: Laser-based custom system

2 cm

Particle counter

Sample tubing

Page 11: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

LPC of Slurry A from Bottles w/ or w/o Cap Liner

• Higher LPC was observed from slurry A in bottle B w/ cap liner

Bottle A: w/o cap linerBottle B: w/ cap liner

11

Page 12: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Effect of Bottle Handling

• LPC for bottle B depended on how the sample bottle was handled

• Operator 1 tipped the bottle over – slurry contacted bottle cap

Bottle B: w/ cap liner

12

Page 13: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Bottle B after Removal of Cap Liner

• Cap liner is demonstrated to be a significant source of LPC

• Low LPC was observed from bottle B after the removal of cap liner

Bottle B: w/ cap linerBottle B: w/o cap liner

13

Page 14: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

SEM Images for Surface of Cap Liner

• Large particles up to 15 um were observed on a cap liner surface by

SEM

Cap liner

14

Page 15: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Cleaning Particles from Cap Liner

• Removal of particle from cap liner

• Typical rinsing with DIW did not help to remove particles on the cap

liners

Before cleaning After cleaning

15

Page 16: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Removal of Particle by Slurry A

• Particles removed by slurry A

• Bottle B (w/ cap liner) tipped over with slurry A in the bottle

• Large particles disappeared from cap liner

Before contacting slurry A

15

After contacting slurry A

Page 17: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Particle from Cap Liner of Bottle B on LPC

• By tipping over the bottle B particles on the cap liner could be pulled

into the slurry causing higher LPC

16

Bottle tipped over

SlurryBottle B with cap liner

Slurry contact

Cap liner

Particle

Particle migrated into the slurry

Particle

Page 18: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Bottle A vs Bottle B

• By using bottle A, LPC variation by sample collectors could be minimized

Bottle A: w/o cap linerBottle B: w/ cap liner

18

Particle size (um) Particle counts

Bottle A Bottle B

>0.56 +/- 58 +/- 343

>0.99 +/- 17 +/- 50

Measurement variation of slurry A

Page 19: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from

Summary

• LPC measurement in CMP slurry can be impacted by particles from cap

liner of sample bottle

• High LPC in slurry A was attributed to particles on a cap liner from

bottle B

• LPC from bottle B varied by how samples were collected

• Uncertain variation by operators could be eliminated by using bottle A

(w/o cap liner)

• To ensure that LPC is a valuable metric for monitoring slurry product

quality, extraneous modes of LPC generation from sampling process need

to be identified and eliminated

Bottle A: w/o cap linerBottle B: w/ cap liner

19

Page 20: The Impact of Sample Containers on Large Particle Count ... · Sample Handling Operator Bottle handling method Operator 1 Bottles tipped over Operator 2 Bottles kept upright from