Page 1: The next step on Magnetic nanoHeatingresearch...2015 Catalog Instruments for Magnetic nanoHeating The next step on Magnetic nanoHeatingresearch DM100 Series System Controller Main

2015 Catalog

Instruments for Magnetic nanoHeating

The next step onMagnetic nanoHeating research

Page 2: The next step on Magnetic nanoHeatingresearch...2015 Catalog Instruments for Magnetic nanoHeating The next step on Magnetic nanoHeatingresearch DM100 Series System Controller Main

DM100 Series System Controller

Main features

Fully automated software operation

Compatible with all DM100 applicator

USB and Serial port connetivity

Touch screen monitor

Maximum productivity, ergonomyand control power for all kinds of

magnetic heating experiments

DMC1 is the brain and the power of all DM100 Series setups.

nB's unique embedded software,MANIAC, helps you program and controls

your experiments, records the measured data, and processes your experimental results into truly useful

and ready to publish reports and graphic presentations.

A reliable, powerful, fully automated and easy way of moving forward on

your magnetic hyperthermia research plans

Local User Interfase15’ tactile color monitor

Optional mouse and keyboardTwo lines LCD status display

SoftwareControl and automation software

for embedded systemData Analysis software for

embedded system.SAR and ILP calculation

Experiment sequence programmingConnectivity

USB Storage portSerial port

Instruments forMagnetic nanoHeating

DMC1 System controller

Instruments for Magnetic nanoHeating

nB nanoScale Biomagnetics introduces the DM100 Series, the only integral, immediate and reliable solution for laboratory tests on Magnetic HyperThermia and Induction nanoHeating.

One sole equipment to apply, measure and analyze the results. One model for every application. The DM100 Series gives the answer to all your needs in MHT/InH instrumentation either with its series models or custom made applications. DM100 Series is also the only integral, immediate and reliable solution to the instrumentation needs of MHT through all its stages: from magnetic characterization to clinical application.

The design conceptualization of the DM100 series emerges from the real need of the Magnetic Hyperthermia lab of the Institute of Nanosciences of Aragón (Spain). The speci�c needs involved in the process of applying a magnetic �eld, measuring and analyzing the results were thoroughly examined and approached one by one, leading to an integrated and �nal solution that guarantees the highest standards in Magnetic Hyperthermia research. The results achieved with DM100 Series are now endorsed by published results on characterization of nanostructured materials, experiments with cell culture, parasithology, among other �elds.

Come and take a close look to DM100 devices and accessories and �nd the best setup for you.

Calle Panamá 2, Local 1Zaragoza 50012

Spain

t. +34 976 741 [email protected]

IndexDMC1 System controller 3

DM1 applicator for calorimetry 4

DM2 applicator for multiple uses 6

DM3 applicator for In vivo test 8

10DM100 Accessories Guide

MaNIaC DM100 Embedded Software 11

MAGNO Magnetic Colloid for InH 11

DM100 Series Configuration Guide 12

Electrical specifications

- Input voltage: 100-240 V AC, 50-60

Hz

- Maximum input power: 2600 W

- Maximum input current: 16 A

Mechanical specifications

- Dimmensions: 505 x 489 x 230

mm3

- Weight: 24 kg

Environmental and safety

specifications

- See DM1

System requirements

- Electrical supply:

o 100-240 VAC – 50-60 Hz, 2600W

Software featuresMANIAC v1.0 is the software

environment which allows the user to run and schedule induction heating tests for research in magnetic hyperthermia in DM100 devices, as well as monitoring and analyzing the results of such tests.

Main functions:

- Test: induction heating test implementation on DM100 series equipments

- Experiments: tests automation on DM100 series equipments

- Selection of magnetic field or temperature as control variable in tests and experiments

- Definition of set point as constant or in function of time in experiments

- Real-time monitoring of equipment sensors.

- Equipment sensors readings log.

The Zar v1.0 application of the software MANIAC analyzes the heating curves of a magnetic colloid type sample which are obtained when carrying out an induction heating test with a DM100 series device.

Main functions:- Visualization of the heating curves - Numerical adjustment of the

heating curves - Calculation of the sample's SPA

and ILP indexes

Page 3: The next step on Magnetic nanoHeatingresearch...2015 Catalog Instruments for Magnetic nanoHeating The next step on Magnetic nanoHeatingresearch DM100 Series System Controller Main

DM100 Series Applicator for calorimetry

Magnetic hyperthermia tests

SAR measurement

Magnetic colloid characterization

Magnetic nanoparticle heating

Precision and reproducibility in magnetic colloid characterization.

DM1 is the only device on the market specially designed for an

accurate determination of the physical parameters that describe the heating effects of alternating

magnetic fields on magnetic nanosystems.

A reliable, powerful, fully automated and easy way of moving forward on

your magnetic hyperthermia research plans

Main features

SensorsOptic fiber temperature probeWater pressure sensorControl methodB field user selectable curveSample temperature user selectable curveCombined B-T user selectable curve

RF SectionUp to 8 user selectable frequencies from 10KHz to 850KHzUp to 350 Oe (27825 A/m) field.Water cooling connection (chiller not included)Sample SectionOpen Dewar flask with vacuum socket 2ml disposable vial sample holder

Low frequency DM1 applicators

f [kHz] Bmin [G] Bmax [G]

610 48 300

300 48 300

270 48 300

230 48 300

215 48 300

185 48 300

175 48 300

130 48 300

125 48 300

120 48 300

115 48 300

110 48 300

105 48 300

Frequency and field corrections < 2%

Full frequency range DM1 applicators

f [kHz] Bmin [G] Bmax [G]

810 48 250

725 48 250

590 48 300

540 48 300

480 48 300

460 48 300

400 48 300

325 48 300

300 48 300

295 48 300

285 48 300

275 48 300

270 48 300

265 48 300

250 48 300 Frequency correction < 2%, field

correction < 3%

Standard operation modesA maximum of 8 modes can be activated on a single device. The tables cannot be merged into one single device.Intensity magnitudes provided in peak Gauss value. For peak to peak value, multipy by 2.

Custom made frequency and higher intensity modes on demand

Volume conf. Axial field attenuation Radial field boost

0.5ml <1.5% <1.5%

1ml <5% <1.5%

1.5ml <11% <1.5%

Field distribution on the sample volumeTypical values. Each device is individually callibrated and documented.

Electrical specifications

- Operating connection I:

o Voltage range: 0-360 V DC

o Maximum power: 1500 W

o Maximum current: 15A

- Operating connection II:

o Voltage: 115 V, 50 Hz

o Maximum power: 130 W

o Maximum current: 2.5 A

Mechanical specifications

- Dimmensions: 415 x 712 x 285 mm3

- Weight: 29 kg

Environmental and safety specifications

- Temperature ranges:

o Normal operation: 5-40 ºC

o At maximum power: 5-30 ºC

o Storage: -20 – 70ºC

- Humidity: 20-80%

- Electromagnetic Compatibility: EMC

2004/108/EC according UNE-EN

61326-1:2006 Class A

- Safety: Low Voltage Directive (LVD)

2006/95/EC according

UNE-EN-61010-1:2011

System requirements

- Electrical supply: connect the applicator-

controller connecting hose before starting

- Liquid refrigeration system

o Minimum cooling power: 1600W

o Typical flow: 5-7 l/min

o Typical input pressure: 2.9 bar

o Two flexible tubes of 10mm

internal diameter to connect to

the provided series 24 hydraulic

connectors .

- If vacuum in the dewar flask is desired, an

appropriate connection to the applicator

connector DN16 ISO-KF is required

Standard accessories

- Two 24 series hydraulic connectors for the

cooling system

- Temperature measuring system:

o FPI-HR-2-F2-SCAI-V dual channel

high resolution fiber optic signal

conditioner

o FOT-L-NS-705 fiber optic

temperature contact sensor with

protected tip

Technical specifications:

o Measuring range: -10ºC to 120 ºC

o Precision: ±0.2 ºC

- Probe-column sample holder set:

o Temperature sensor

o Sample holder column

o Kit of sample/probe position

calibration parts

Instruments forMagnetic nanoHeating

The next step on Magnetic nanoHeating research

DM1 applicator

www.nbnanoscale.com

Page 4: The next step on Magnetic nanoHeatingresearch...2015 Catalog Instruments for Magnetic nanoHeating The next step on Magnetic nanoHeatingresearch DM100 Series System Controller Main

Instruments forMagnetic nanoHeating

The next step on Magnetic nanoHeating research

DM2 applicator

DM100 Series Applicator for in vitro tests and multiple uses

Main features

Magnetic hyperthermia in vitro tests

Magnetic colloid characterization

Combined experiments

Custom designed accesories

Precision and reproducibility in magnetic heating experiments.

DM2 is the most versatile tool you can imagine for studying

the heating effects of alternating magnetic fields on magnetic

nanosystems on in vitro, in vivo or new material tests

A reliable, powerful, fully automated and easy way of moving forward on

your magnetic hyperthermia research plans

RF SectionUp 4 user selectable frequencies from 100KHz to 500KHz Up to 200 Gauss (15.9 kA/m) field.Water cooling connection (chiller not included)Sample SectionMultiple optional sample holders60 mm/80 mm diameter versatile sample chamber in vitro, in vivo and calorimetry accessories.Sensors1 or 3 fiber optic temperature sensorWater pressure sensorControl methodB field user selectable curveSample temperature user selectable curveCombined B-T user selectable curve

Compatible assessories (see accesories datasheet on page 10)

CAT: Controlled atmosphere accessory for in vitro testsCAL: Vacuum dewar flask calorimeter with sensor probeHBM: Heartbeat monitor system for small animalsIR1: Infrared thermal image systemDRM: Drug release monitor system for in vitro tests

Custom designed sample holders, probes and accessories.

CAT

CAL

IR1

Field distribution on the sample volumeTypical values. Each device is individually callibrated and documented.

Volume configuration

Axial field attenuation

Minimum radial field attenuation

Maximum radial field attenuation

H = 40mm <9% <8% <18%

Low frequency DM2

applicators

f [kHz] Bmin [G] Bmax [G]

230 36 200

120 36 200

100 36 200

85 36 200

Frequency correction < 2%,

Field correction < 4%

High frequency DM2

applicators

Frequency correction < 2%,

Field correction < 4%

f [kHz] Bmin [G] Bmax [G]

405 36 172

315 36 172

285 36 180

250 36 180

HBM

DM100 Series Applicator for in vitro tests and multiple uses

www.nbnanoscale.com

Electrical specifications

- Operating connection I:

o Voltage range: 0-360 V DC

o Maximum power: 1500 W

o Maximum current: 15A

- Operating connection II:

o Voltage: 115 V, 50 Hz

o Maximum power: 130 W

o Maximum current: 2.5 A

Mechanical specifications

- Dimmensions: 439 mm x 712 mm x

259 mm

- Weight: 29 kg

Environmental and safety specifications

- Temperature ranges:

o Normal operation: 5-40 ºC

o At maximum power: 5-30

ºC

o Storage: -20 – 70ºC

- Humidity: 20-80%

- EMC 2004/108/EC according

UNE-EN 61326-1:2006 Class A

- Low Voltage Directive (LVD)

2006/95/EC according

UNE-EN-61010-1:2011

System requirements

- Electrical supply: connect the

applicator-controller connecting

hose before starting

- Liquid refrigeration system

o Minimum cooling power:

1600W

o Typical flow: 5-7 l/min

o Typical input pressure: 2.9

bar

o Two flexible tubes of

10mm internal diameter

to connect to the

provided series 24

hydraulic connectors .

Standard accessories

- Two 24 series hydraulic connectors

for the cooling system

- Temperature measuring system:

o Dual or quad channel high

resolution fiber optic signal

conditioner

o 1 or 3 fiber optic temperature

sensor with protected of

unprotected tip

Technical specifications:

o Measuring range: -10ºC to 120

ºC

o Precision: ±0.2 ºC

- Probe-holder sample set:

o Temperature sensor

o Sample holder

Compatible accessories

- DMR: Drug release monitor for in

vitro tests

- CAT: Controlled atmosphere and

temperature accessory for in vitro

tests

- CAL: Vacuum dewar flask

calorimeter with sensor probe

- HBM: Heartbeat monitor for small

animals

- IR1: Infrared thermal image system

- Custom designed sample holders,

probes and accessories

Standard operation modesFrequencies and intensities on available devices may differ. Please, ask nB before ordering.

Exceptionally, values on final devices may present up to 5% factory tolerance in some high energy modes. Typical tolerance is 2%.

Intensity magnitudes provided in peak Gauss value. For peak to peak value, multipy by 2.Custom models can host up to 8 modes with 40mm diameter sample chamber

Custom frequency modes on demand

Page 5: The next step on Magnetic nanoHeatingresearch...2015 Catalog Instruments for Magnetic nanoHeating The next step on Magnetic nanoHeatingresearch DM100 Series System Controller Main

Instruments forMagnetic nanoHeating

The next step on Magnetic nanoHeating research

DM100 Series Applicator for in vivo hyperthermia tests

Magnetic hyperthermia in vivo tests

Local and systemic temperature sensing

Local or whole body versions

Heartbeat monitoring

Thermal imaging

Maximum precision and ergonomyfor in vivo developing and testing

of magnetic hyperthermia and drug realease therapies

DM3 is the only device on the market specially designed for an

accurate application of alterating magnetic field, local sensing

and experiment monitoring in mice and rats on laboratory hyperthermia procedures

A reliable, powerful, fully automated and easy way of moving forward on

your magnetic hyperthermia research plans

Main featuresRF Section4 user selectable frequencies from 80KHz to 450KHz (According to user requirements)Up to 450/300 Gauss/Oersted (35.7/27.8 kA/m) field.Water cooling connection (chiller not included)Sample SectionAnimal bed for mice or rats Surface and whole body application versionsSensors1 or 3 local temperature sensorWater pressure sensorControl methodB field user selectable curveLocal temperature user selectable control optionCombined B-T user selectable curveInfrared imaging control loopCompatible assessoriesHBM: Heartbeat monitor system for small animalsIR1: Infrared thermal image systemDRM: Drug release monitor system for in vitro testsMicro CAT: Controlled atmophere for tests in small petri platesCustom designed sample holders, probes and accessories.

DM3 applicator

Standard operation modesFrequencies and intensities on available devices may differ. Please, ask nB before ordering.

Exceptionally, values on final devices may present up to 10% factory tolerance in some high energy modes. Typical tolerance is 2%.

Intensity magnitudes provided in peak Gauss value. For peak to peak value, multipy by 2.A maximum of 8 modes can be activated on a single device

Custom frequency modes on demand

DM3 - GWhole Body applicator

DM3 - JSurface applicator

www.nbnanoscale.com

Electrical specifications

- See DM1

Mechanical specifications

- Dimmensions: 439 mm x 712 mm x 259 mm

- Weight: 29 kg

Environmental and safety specifications

- See DM1

System requirements

- See DM1

Standard accessories

- Two 24 series hydraulic connectors for the cooling

system

- Temperature measuring system:

o Dual channel high resolution fiber optic signal

conditioner

o 1 or 3 fiber optic temperature sensor with protected or

unprotected tip

Technical specifications:

o Measuring range: -10ºC to 120 ºC

o Precision: ±0.2 ºC

Animal bed:

o Adjustable positioning

Compatible accessories

- HBM: Heartbeat monitor for small animals

- IR1: Infrared thermal image system

- DRM: Drug release monitor for in vitro tests

- Micro CAT: Controlled atmosphere for tests in small

petri plates

Custom designed sample holders, probes and accessories

available

DM3 - HWhole Body

open view applicator NEW!

� DM3-H DM3-G f [kHz] Bmáx[G] f [kHz] Bmáx[G] f [kHz] Bmáx[G]

461 300 455 190 449 300

261 400 273 250 278 400

227 400 234 260 237 400

151 400 148 280 154 400

143 400 141 285 141 400

126 400 125 290 135 400

99 400 106 295 100 400

87 400 92 300 92 400

81 400 86 300 80 400

DM3-J

Page 6: The next step on Magnetic nanoHeatingresearch...2015 Catalog Instruments for Magnetic nanoHeating The next step on Magnetic nanoHeatingresearch DM100 Series System Controller Main

The next step on Magnetic nanoHeating research

DM2 Controlled Atmosphere andTemperature cell culture holder

CAT

DM100 Thermal Image SystemIR1

DM2 Accessory for Calorimetryand SAR measurement

CAL

Hot point detectionHistogramVisible video monitoringImage backfed field control loopImage data analysisRaw data available for external analysisOptional local probe callibrationUseful for in vitro, ex vivo, materials and in vivo tests.

Allows you to make real time surface thermal images and movies in DM2 and DM3 experiments.

Fully integrated with MANIAC embedded software, featuring advanced analysis functions.

Enables IR data-controlled experiments

Main features

Turns your DM2 in a high quality calorimeter for SAR determination and thermal characerization of liquid samples.

Includes ZAR software option, for instant data fitting, SAR/SPA and ILP calculations.

2ml vial sample holderFiber optic high accuracy thermometerOpen dewar flask with vacuum pumpconnection socket (pump not included)ZAR data analysis software package

Main features

Enables atmosphere and temperature controlin cell culture experiments in DM2

Micro CAT available for DM3 for small petri plates

The only blood pressure monitoring systemfor small animals fully compatible

with high intensity AC fields

Advanced integration with MANIAC embedded software

Enables heart rate-controlled experiments

Hydraulic sensing via internal catheter(minor surgery needed)Distant electric transductionReal time blood pressure graphReal time heart rate measureIntegrated dataloggingCompatible with IR1

Main features

Designed for

DM2 DM3

Designed for

DM2 DM3

Designed for DM3

Designed for DM2

Compatible withDM2 HBM

DM100 HeartBeat Monitor system for small animals

Main internal chamber with input andoutput socket for gas mix flowSecondary external chamber with input and output sockets for thermalization waterCompatible with up to 30 mm petri dishesEnables multiple dish experimentsAdjustable high accuracy optic thermometerCustom design for IR imaging with IR1 accessory

Main features

DM100 Series Accessories guide

MaNIaCDM100 Series Embedded Software

Fully automatic test execution

Ease of use and high productivity

Complex experiment sequencing

Integration and automation of accessories

Quick calculations and data processing

Data import/export

Expandability

Performing individual testsMANIAC allows you to carry out inductive heating tests in a matter of seconds. Using “test” mode, an automatic fi el application can be started and dcontrolled just by indicating the field values you wish to attain.

Programming complex testsOne important characteristic of MANIAC is that it is able to carry out complex tests made up of fi elds of different characteristics, with pre-set intervals of time and even to control them through different parameters. A versatile tool for trying out new ideas.

Data registrationMANIAC keeps you informed in real time about the data registered during the experiment. Both the fi eld characteristics as well as variations measured in the sample are shown on screen, while they are fi led away together with all the other test data. MANIAC is compatible with USB storage devices, which allows you to manage all the files in standard format.

Analysis of ResultsEach applicator has an option for additional software with tools for analyzing experiments, which can speed up work fl ow in the lab. MANIAC is a fl exible and powerful environment that adapts to each experiment and makes the researcher’s daily life easier.

Depending on the available options, MANIAC can calculate a sample’s SAR, make graphic presentations and statistical charts, and is open to tailor-made designs and future modifications in accordance with the researcher’s needs.

Accessories managementMaNIaC is prepared to integrate and control every active DM100 accessory, like IR1 or HBM. All data and automatic functions of DM100 accessories can be viewed and processed within MaNIaC environment.

Magnetic colloid for Magnetic nanoHeating

Ma nogMAGNO is the �rst commercially available magnetic colloid specially designed for Inductive nanoHeating applications.

Optimizing the heating properties of MNPs is cumbersome and takes long time and dedicated manpower. Magno offers the solution for those researchers that need accurate data, repetibility and reliability to test new ideas such as physical models, or any heating applications.

MAGNO is a water-based magnetic colloid consisting of magnetite (Fe3O4) nanoparticles with average sizes of 25nm functionalized with a biocompatible polymer. The magnetic nanoparticles display an impressive SPA value (typically >210W/g, for f = 580KHz, H0 =23.877 KA/m).

MAGNO has been manufactured and synthesized under controlled

conditions using the DM100 technology. The precise control of the manufacturing conditions gives MAGNO a highly reproducible heating behavior and SPA values under different conditions, and an outstanding stability over the time.

MAGNO is the ideal raw material for a �rst approximation to inductive nano heating research as well as for a great variety of advanced applications, only limited by your imagination. Covered by Polyethylenimine (PEI), MAGNO nanoparticles constitutes a powerful tool not only for basic research on power absorption but also for different applications in most �elds of biomedicine.

MAGNO has been developed by nB nanoScale Biomagnetics under three basic principles for inductive heating research: repetibility, traceability and reliability.

www.nbnanoscale.com

A fully integrated system for the quantificationof a drug release experiment on liquid

and solid samples

Advanced integration with MANIAC software

Enables delivery rate-controlled experiments

Integrated spectrophometerPeristaltic pumpDynamic magnetic separationHolders for liquid and solid samples

Main featuresDesigned for

DM2 Compatible with

DM1 DM3 DRMDM100 Drug Release

Monitor system

NEW!Coming fall

2015

Page 7: The next step on Magnetic nanoHeatingresearch...2015 Catalog Instruments for Magnetic nanoHeating The next step on Magnetic nanoHeatingresearch DM100 Series System Controller Main

DM2 Accessory for Calorimetryand SAR measurement

CAL

HBMDM100 HeartBeat Monitor system for small animals

DM2 Controlled Atmosphere andTemperature cell culture holder

CAT

DM100 Thermal Image SystemIR1

100v-240v AC

Water chiller

DM100 Series Applicator for in vitro tests and multiple uses

DM100 Series Applicator for in vivo hyperthermia tests

DM100 Series Applicator for calorimetry

Rotary or turbomolecular vacuum pump(recommended for calorimetry)

Basic Setup

Mandatory third party gear Recommendedthird party gear

DM100 Series is a set of instruments, accesories and software tools that can be combined to form different configurations covering every kind of Magnetic Hyperthermia and Induction nanoHeating

experimental setup. Each DM100 configuration is a complete workstation that allows you to automatically run complex tests, register data and analyze your results.

DM100 Series is designed under a concept of modularity. One controller can be used to drive multiple applicators -not simultaneously. Here is a quick guide to how your DM100 system must to

be configured.

Accessories

Instruments for Magnetic nanoHeating Reseach

Calle Panamá 2, Local 1Zaragoza 50012

Spain

t. +34 976 741 [email protected]

MaNIaCDM100 Series Embedded Software

DRMDM100 Drug Release Monitor system


Recommended