Academiejaar 2015 – 2016
'THE ULTRASOUND ‘POP-UP SIGN’ RELIABLY
PREDICTS CORRECT NASOGASTRIC TUBE
POSITION IN INTENSIVE CARE UNIT PATIENTS
Pieter VERSLYPE
Promotor: Prof. dr. P. Wouters
Co-promotor: Dr. W. Stockman
Masterproef voorgedragen in de master in de specialistische geneeskunde
Anesthesie en Reanimatie
Academiejaar 2015 – 2016
'THE ULTRASOUND ‘POP-UP SIGN’ RELIABLY
PREDICTS CORRECT NASOGASTRIC TUBE
POSITION IN INTENSIVE CARE UNIT PATIENTS
Pieter VERSLYPE
Promotor: Prof. dr. P. Wouters
Co-promotor: Dr. W. Stockman
Masterproef voorgedragen in de master in de specialistische geneeskunde
Anesthesie en Reanimatie
Abstract (English)
The ultrasound pop-up sign reliably predicts correct nasogastric tube position in ICU patients
Introduction:
Complications due to a malpositioned nasogastric tube (NGT) may have disastrous
consequences leading to prolonged length of stay or ultimately death. Therefore the correct
position must be confirmed before usage of the NGT. Although several methods have been
investigated, chest radiograph remains the golden standard to confirm correct position.
However, a chest X-ray in an ICU patient has disadvantages including radiation exposure,
cost and labor intensity.
Objective:
The aim of this study is to investigate the diagnostic accuracy of the ultrasound pop-up sign to
confirm correct position of a nasogastric tube in ICU patients compared with chest X-ray.
Material and methods:
In this single center prospective study, adult patients admitted to our ICU were included if a
NGT was placed. After placement the correct position was tested by ultrasound. The patient
remained in the dorsal decubitus position with the head of the bed elevated 30°. The antrum
was visualized by ultrasound and then NGT was rapidly insufflated with 60 ml of air. If a
sudden pop-up of air was identified in the stomach the sign was considered positive. A chest
X-ray was subsequently taken to confirm position.
Results:
Eighty patients were included. The mean patient age was 66 ± 14.6 years. The mean Body
Mass Index was 24.7 ± 5.5. Sixty-four patients (80%) were invasively ventilated. In 63/80
patients (79%) the pop-up sign could be identified. In all these 63 patients correct position of
the NGT was confirmed by chest X-ray (positive predictive value 100%). In 17 patients the
pop-up sign could not be identified. Of these the nasogastric tube was correctly placed in the
stomach in 14 patients (88%), while two (12%) were midesophageal and one postpyloric
(6%). The sensitivity of the method is thus 63/77 (=0.82).
Conclusion:
The ultrasound pop-up sign is a very reliable predictor of correct nasogastric tube position and
is present in 82% of correctly positioned nasogastric tubes. We believe a confirmatory chest
X-ray can be omitted after visualization of the pop-up sign.
Abstract (Nederlands)
Het ultrasound pop-up sign is een betrouwbare predictor van correcte nasogastrische tube
positie in intensieve zorg patienten.
Inleiding:
Complicaties ten gevolge van een verkeerd gepositioneerde nasogastrische sonde (NGS) kan
zeer ernstige gevolgen hebben leidend tot een verlengde hospitalisatie duur of overlijden. De
positie van de NGS dient geverifiëerd te worden voor het gebruik ervan. Verschillende
methodes zijn hiervoor reeds onderzocht, maar een thoraxfoto blijft de gouden standaard. Een
thoraxfoto op intensieve zorgen heeft wel nadelen zoals straling, kostprijs en werkintensiteit.
Doelstelling:
Het doel van deze studie is om de diagnostische performantie van het ultrasound pop-up sign
op intensieve zorgen te bepalen in vergelijking met de RX-Thorax voor de correcte positie
van de NGS.
Methodologie
In deze single-center prospectieve studie werden volwassen patienten geïncludeerd die
werden opgenomen op intensieve zorgen als er een NGS werd geplaatst. Na de plaatsing werd
de positie bepaald met echografie: de patient was gepositioneerd in ruglig met 30° elevatie
van het bed. Het antrum werd met echografie in beeld gebracht en daarna werd snel 60 ml
lucht geinsuffleerd in de NSG. Als er een plotse pop-up van lucht zichtbaar werd, dan werd
het teken als positief beschouwd. Nadien werd een thoraxfoto genomen ter controle.
Resultaten:
Tachtig patienten waren geïncludeerd. De mean leeftijd was 66 ± 14.6 years en de mean Body
Mass Index 24.7 ± 5.5. Vierenzestig patiënten waren invasief geventileerd (80%). In 63/80
(79%) van de patiënten kon het pop-up sign opgemerkt worden. Een controle RX-thorax
bevestigde de correcte positie van de NGS bij alle patiënten met een positief pop-up sign
(Positive Predictive value = 100%) In 17 patiënten was het pop-up sign niet visualiseerbaar.
De positie van de maagsonde was van deze patienten correct geplaatst in 88 %. Twee NGS
waren midoesophagaal geplaatst en één postpyloor. De sensitiviteit is 82%
Conclusie:
Het ultrasound pop-up sign is een betrouwbare predictor van correcte nasogastrische tube
positie en is aanwezig in 82% van de correct gepositioneerde tubes. We zijn van mening dat
een controle RX-Thorax achterwege kan gelaten worden na een positief pop-up sign.
Introduction
In Intensive care medicine, the nasogastric tube has multiple purposes and is frequently
placed in the Intensive care unit in patients of all ages. A nasogastric tube is a flexible single
or double lumen tube that is passed proximally from the nose distally into the stomach or
small bowel. The gastric tube can also be placed orally in certain indications (e.g. basilar skull
fracture, …). (7) The most common indications for the placement of nasogastric tubes are
gastrointestinal decompression (e.g. ileus or bowel obstruction, or after gastric distension due
to bag-valve-mask ventilation or assisted ventilation), administration of medication, gastric
lavage or enteral nutrition. (2, 8)
The placement of a nasogastric tube is frequently considered a harmless procedure and is
placed on a daily basis and the complications may range from simple to disastrous
consequences leading to prolonged length of stay or ultimately death. Nasally inserted tubes
may cause nasal alar ulceration, epistaxis or sinusitis. (9,10) Nasal placement of a feeding
tube can obstruct breathing in patients of all ages although it is a greater problem in newborns
who are obligate nasal breathers due to partial nasal obstruction, an increase in airway
resistance and increased work of breathing. (9) The presence of a nasogastric tube impairs the
normal function of the lower esophageal sphincter making the patient more susceptible to
reflux of gastric contents which may lead to esophagitis, esophageal stricture, gastrointestinal
bleeding or pulmonary aspiration. (8) The most serious complications are related to
malpositioned tubes or perforation which can cause pneumothorax, pneumomediastinum,
subcutaneous emphysema, pulmonary hemorrhage, pulmonary abcess, aspiration by proxy,
mediastinitis. Nasogastric tube misplacement appears between 0.5-11% and intracranial or
intravascular placement has also been described in rare cases. (2, 3, 7, 9, 23, 28-31)
The standard of care requires verification of the placement of the gastric tube prior to its use
in order to minimize complications. (11) Several methods have been thoroughly investigated
(e.g. pH-metry, visual inspection of aspiration of gastric content, auscultation of air,
ultrasound of the stomach for direct visualization of the tip, capnography, … ) but
radiographic verification is considered the golden standard to confirm the correct position of
the nasogastric tube and is preferred by many. (11-17) However, radiological confirmation in
an ICU patient has several disadvantages including radiation exposure, cost, labor intensity
and is only a single point in time. (17-20)
Ultrasound use has gained its place in the Intensive care unit as a daily monitoring tool. Over
the past few years a multitude of protocols for standard evaluation of the Intensive Care
patient have been published such as the BLUE protocol (Bedside Lung Ultrasound in
Emergency), FALLS protocol (Fluid Administration Limited by Lung Sonography), FATE
protocol (Focused Assessed of Transthoracic Echocardiography), EFAST (Extended Focused
Assessment with Sonography for Trauma). (32-37) Ultrasound assessment of gastric content
and volume is also used in anaesthesia practice to help determine the aspiration risk. The
antrum of the stomach is the most tractable gastric region for ultrasound evaluation and its
evaluation accurately reflects the content of the entire organ. (5)
Several studies have already reported the use of ultrasound for the assessment of the correct
position of the nasogastric tube in the intensive care unit and the emergency department,
however to our knowledge no studies have been performed with to evaluate the diagnostic
accuracy of the ultrasound pop-up sign in the intensive care department. (2-4, 40, 41)
The objective of this study is to investigate the diagnostic accuracy of the ultrasound “pop-up
sign” to confirm correct position of a nasogastric tube in intensive care unit patients compared
with chest radiograph.
Material and methods
In this single centre prospective study, patients over the age of 18 with no previous history of
surgery to the stomach or esophagus, admitted to our mixed surgical and medical ICU were
included if a nasogastric tube was required for either enteral feeding and/or medication
administration, abdominal decompression or gastric lavage. The study was performed
between January 2012 until November 2014.
The length of the tube was estimated by using the distance from the tip of the nose to the tip
of the ear to the xiphisternum plus 10 centimeter. (see image 1) (24, 25, 38, 39) The
nasogastric tube was then lubricated with a water soluble lubricant to reduce friction. The
right nostril was usually preferred except when resistance was felt, then the contralateral
nostril was used.
.
In mechanically ventilated patients the nasogastric tube was either placed in the esophagus by
a trained anesthesiologist under direct vision with a laryngoscope and Magill forceps and
advanced in the esophagus until the premeasured length of the nasogastric tube was reached
or the head was maintained neutrally and by using two fingers inside the patients mouth the
tip of the nasogastric tube was guided between the two fingers to the esophagus. If resistance
was felt or coiling occurred in the mouth a laryngoscope was used. (8, 26)
When the patient was awake, the patient was seated with the head tilted towards the chest.
The tube was then introduced into one of the nostrils horizontally over the nasal floor. The
patient was given a small cup of water with a straw and was to drink while the nasogastric
tube was advanced. When the nasogastric tube reached the posterior nasopharynx and the
patient felt like gagging, then swallowing was encouraged to help guide the tube to the
stomach. If the patient became short of breath or had to cough then the nasogastric tube was
withdrawn and a new attempt was performed after a minute. (8, 27) Once the insertion of the
nasogastric tube was finished it was secured to the patient’s nose with adhesive tape.
After placement of the nasogastric tube the correct position was tested initially as follows.
The patient was placed in the dorsal decubitus position with the head of the bed elevated 30°.
The antrum of the stomach was visualized by ultrasound with a low-frequency (2-5 MHz)
curvilinear abdominal probe (Philips CX50 – probe C5-1). The probe was placed horizontally
in the epigastric area and oriented towards the left upper abdominal quadrant. The left lobe of
the liver was used as an internal landmark for optimal visualization of the antrum of the
stomach. The antrum appears flat with juxtaposed anterior and posterior walls in a fasted
stomach, with increasing volume, the antrum becomes round and distended with thin walls. In
the unfasted patient, the air/solid mixture creates multiple ring-down artifacts on the anterior
gastric wall which obscure the posterior wall due to scattering. (2, 5). (see image 2A)
When an optimal image was acquired, a nurse or second investigator was asked to rapidly
insufflate the nasogastric tube with 60 ml of air. The insufflated air causes a typical
appearance due to the scattering of the ultrasound waves at the interface of soft tissue and air.
The reverberations of the ultrasound waves between the transducer and the air increase the
echogenicity and cause posterior artifactual reverberation echoes with characteristic comet-
tail echo appearance. (43) If a sudden pop-up of air, also known as dynamic fogging was
identified in the stomach in a maximum of four attempts, then the sign was considered
positive. (see image 2B)
The investigator was asked to only identify the pop-up sign as positive if he was absolutely
certain of visualization of the sign.
After the ultrasound investigation a chest radiograph was taken to confirm the position of the
nasogastric tube and to determine the precise location of the tip of the nasogastric tube in the
upper gastrointestinal tract. (see image 3) (42) Chest radiographs were interpreted by a
radiologist who did not perform the ultrasound visualization.
Statistical methods:
Observed data were entered into Microsoft Excel Workbook 2010. All analyses were
conducted using Statistical Package for the Social Sciences (SPSS) for Windows (IBM SPSS
Statistics version 23. Inc., Chicago, IL, USA). Continuous variables were reported in terms of
the mean and standard deviation, and categorical variables were reported in terms of
frequency and percentage. Student’s t-test was used for continuous variables and Fisher’s
exact test for categorical variables. A P-value under 0.05 was considered to be of statistical
significance for all tests. Indicators of the performance of a diagnostic test, sensitivity,
specificity, positive predictive value, and negative predictive value were determined.
Results
Eighty patients were included in this study, of which 47 were male (58.8%). The mean age
was 66.3 years (Standard deviation (=SD) 14.6 years). The height and weight of the patient
were used to calculate the Body Mass Index. The mean BMI was 24.7 (SD 5.5). The
placement of the nasogastric tube was indicated for either abdominal decompression or
medication administration and/or enteral feeding in respectively 62.5% and 37.5% of the
cases. There were no indications for nasogastric tube placement for gastric lavage during the
study.
The nasogastric tube was inserted in awake patients in 16/80 cases (20%). In mechanically
ventilated patients, the nasogastric tube was inserted blind in 30 of 64 cases (46.9%) and with
the aid of the laryngoscope and Magill forceps in 34 of 64 cases (53.1%). (Table 1) The p-
value of the mean BMI and mean age between the group with a positive ultrasound pop-up
sign and a negative ultrasound pop-up sign when using a student’s t-test was >0.05 and was
not considered statistically significant. Fisher’s exact test showed no statistically significant
difference between a positive pop-up sign for either the indication for the nasogastric tube or
the method of insertion.
Table 1: Patient characteristics
Sexe Male 47/80 (58.8%)
Age (years) 66,3 ± 14,6 *
Body Mass Index1 24,7 ± 5,5*
Indication for nasogastric tube
Abdominal decompression 50/80 (62,5%)
Medication administration/enteral feeding 30/80 (37,5%)
Gastric lavage 0/80 (0%)
Method of nasogastric tube insertion
Awake patient 16/80 (20%)
Laryngoscope and Magill forceps 34/80 (42,5%)
Blind insertion 30/80 (37,5%)
*mean ± standard deviation 1: kg/m²
Three different types of nasogastric tubes were inserted, with each varying external diameters.
Thirty small bore enteral feeding tubes were inserted with an external diameter of 8 F
(French). The ultrasound pop-up sign was positive in 18/30 cases. Six 14 F Salem-sump tubes
were inserted with a positive pop-up sign in 83.3% of the cases. Forty Levin tubes were
inserted with four different diameters. One 10 F Levin tube was inserted with a positive pop-
up sign. Eight 12 F Levin tubes were inserted with a positive pop-up sign in 7/8 (87.5%) of
the cases. The majority of the inserted Levin tubes had a diameter of 14 F and had a positive
pop-up sign in 30/33 (90.9%). There were two 16 F Levin tubes inserted with both had a
positive pop-up. (Table 2) A Fisher’s exact test was performed and showed there was a
significant difference (p=0.031) between the external diameter of the nasogastric tube and the
visualization of the ultrasound pop-up sign.
The ultrasound pop-up sign was present in 63/80 cases (78.8%). The pop-up sign could be
identified after a single insufflation of air in 24/63 (38.1 %) patients, after two insufflations in
15/63 (24%) patients, after three insufflations in 9/63 (14%) and after four insufflations in
5/63 (8%) patients. In the remaining 10 patients (16%), the data of the number of attempts are
missing. (Table 3)
Table 2. Type of nasogastric tube
Ultrasound pop-up sign
Diameter Positive Negative
Small bore nasogastric tube 8 F1 18 (60%) 12 (40%)
Salem-sump tube 14 F 5 (83.3%) 1 (16.7)
Levin tube
Black 10 F 1 (100%) 0 (0%)
White 12 F 7 (87.5%) 1 (12.5%)
Green 14 F 30 (90.9%) 3 (9.1%)
Orange 16 F 2 (100%) 0 (0%) 1 F= External Diameter (mm) * 3
The nasogastric tube position was confirmed by chest radiograph in all patients with no cases
of serious complications. In all 63 patients with a positive pop-up sign, the correct position of
the nasogastric tube was confirmed by chest radiograph. On chest X-ray, the nasogastric tube
was identified in the corpus in 49/63 (77.8%), in the cardia in 4/63 (6.3%), in the antrum in
3/63 (4.8%) and in the fundus in 7/63 (11.1%). In 17 patients the pop-up sign could not be
identified. Of these the nasogastric tube was correctly placed in the stomach in 14 patients
(82.3%), while two (11.8%) were midesophageal and one was located postpyloric 1/17
(5.9%). (Table 4)
Table 4: Location of tip of nasogastric tube on chest radiograph
Ultrasound pop-up sign
Tip location on chest radiograph Positive Negative
Corpus 49/63 (77.8%) 7/17 (41.1%)
Cardia 4/63 (6.3%) 2/17 (11.8%)
Antrum 3/63 (4.8%) 0/17 (0%)
Fundus 7/63 (11.1%) 5/17 (29.4%)
Midesophageal 0/63 (0%) 2/17 (11.8%)
Postpyloric 0/63 (0%) 1/17 (5.9%)
Table 3: Insufflation of air until appearance of pop-up sign
Positive ultrasound pop-up sign 63/80 (78,8%)
Attemps of insufflation
One insufflation 24/63 (38,1%)
Two insufflations 15/63 (23,8%)
Three insufflations 9/63 (14,3%)
Four insufflations 5/63 (7,9%)
Unknown 10/63 (15,9%)
The ultrasound pop-up sign has a sensitivity of 81.8% with 95% confidence interval [71.4%-
89.7%] and a specificity of 100%, 95% CI [29.2%-100%]. The positive predictive value is
100%, 95% CI [94.3%-100%] and a negative predictive value of 17.6%, 95% CI [3.8%-
43.4%]. The Youden’s index is 0.81. (Table 5)
Table 5: Diagnostic accuracy of ultrasound pop-up sign for confirming correct position
of nasogastric tube
Confirmation chest radiograph of nasogastric tube
Correct Position Incorrect Position Total
Pop-up sign
Positive 63 0 63
Negative 14 3 17
Total 77 3 80
Discussion
The frequency of malposition varies between 0.5-11%. The incidence of tracheopulmonary
complications are approximately 2% and 0.3% died from the complications. Therefore it is
necessary to verify the correct position of the nasogastric tube. (3) Although many
verification methods have been described the most common used techniques are monitoring
for signs of respiratory distress, observing the aspirate of the nasogastric tube and the ‘whoosh
test’(auscultation over the epigastrium following the rapid insufflation of air). (48-49)
The ‘whoosh’ test however has proven to be unreliable and widely used despite
recommendations to the contrary. PH measures are used in combination with visual inspection
of the aspirate, a pH lower than 5.5 indicates gastric placement however the use of H2-
blockers and proton pump inhibitors in the intensive care unit confounds the usefulness of
assessing the pH. Enteral feeding, certain medications and sterile water (pH 5.5-7) can also
alter the pH. It is often difficult to obtain an aspirate and small bore tubes may collapse when
negative pressure is applied. (50)
Other methods exist to confirm the correct position such as pepsin and trypsin concentrations
in the aspirate, capnography, bilirubin testing of the aspirate or with an electromagnetic
device for special nasogastric tubes.
Chest radiograph remains the golden standard by which all tests are compared. Although the
radiation exposure is small for the a single chest radiograph, repeated exposures may result in
high cumulative doses which can be associated with cancer. Ultrasound examination requires
little patient manipulation, whereas a chest radiograph can be more time consuming and labor
extensive. We believe that ultrasound may become another verification method with a
widespread accessibility in the intensive care unit. (18, 19, 21, 50)
The use of ultrasound has become a standard of care in the intensive care unit. The objective
of this study was to evaluate the ultrasound pop-up sign compared with the golden standard to
determine the correct position of the nasogastric tube in the intensive care unit. The sensitivity
was 81.8% and specificity 100% confirm the diagnostic performance of the ultrasound pop-up
sign. The diagnostic odds ratio cannot be calculated since the positive likelihood ratio is
infinite (sensitivity/1-specificity). To approximate the diagnostic odds ratio, 0.5 was added in
each cell in table 5. The diagnostic odds ratio is 242 and indicates a high discriminatory test
performance which is independent of prevalence. (51)
Our study shows that the pop-up sign becomes less visible with a smaller diameter. This could
be due to the inability to rapidly insufflate the air through a smaller diameter.
Bedside sonography for confirming the tip of nasogastric tubes can rapidly taught and
although we only studied the ultrasound pop-up sign, ultrasound can also be used as an
adjuvant method during the insertion of the nasogastric tube to confirm the location of the
nasogastric tube in the cervical esophagus or it can be used to directly visualize the location of
the tip in the stomach. The larger the nasogastric tube is, the easier the direct visualization. (2-
4, 21). Weighted-tip nasogastric tubes are easily identified and appear as an hyperechogenic
line. (21) Ultrasound verification of the correct position of the nasogastric tube by direct
visualization has also been performed in children. Other studies have also used the ultrasound
pop-up sign but they only used it as an adjuvant technique when the tip was not visible. (2-4,
41, 46-47) Kim et al (4) described one case in which dynamic fogging was false positive. The
nasogastric tube was located midesophageal and dynamic fogging could be seen in the
stomach. However they used a solution of 40 ml normal saline and 10 ml of air, whereas in
our study only air was used.
Limitations
The time required to perform the ultrasound examination was not measured and could not be
compared with chest radiograph. A recent study by Brun however showed that the insufflation
of air and echographic verification can be performed in one minute. (3)
There is no report on whether the patient was fasted or ‘unfasted’. It seems likely that the
ultrasound pop-up sign could not be visualized in certain cases due to gas interposition which
remains a major limitation for ultrasound. (2) The duration of the study was almost three
years and this is due to the fact that only a single examiner included the patients and
performed the ultrasound examination. It is possible that a selection bias occurred.
Conclusion
Bedside ultrasound examination is becoming standard of care in the intensive care unit and
sonography for confirming the correct location of the nasogastric tube can be easily and
rapidly taught. The combination of ultrasound techniques to confirm the correct position of
the nasogastric tube supplement one another and we believe the ultrasound pop-up sound is an
essential part of this examination. The ultrasound pop-up sign is a very reliable predictor of
correct nasogastric tube position and is present in 82% of correctly positioned nasogastric
tubes. We believe a confirmatory chest X-ray can be omitted after visualization of the pop-up
sign and should be reserved for when the examination is indecisive.
References
(1)Sequin P, Le Bouquin V, Aquillon D, Maurice A, Laviolle B, et al.Testing nasogastric tube
placement: evaluation of three different methods in intensive care unit. ANN FR ANESTH
REANIM. 2005 Jun;24(6):594-9
(2)Chenaitia H, Brun P, Querellou E, Leyral J, Bessereau J, Aimé C, Bouaziz R, et al.
Ultrasound to confirm gastric tube placement in prehospital management. Resuscitation 2012
Apr;83(4):447-51
(3)Brun P, Chenaitia H, Lablanche C, Pradel A, Deniel C, Bessereau J, et al. 2-point
ultrasonography to confirm correct position of the gastric tube in prehospital setting. MIL
MED 2014 Sep;179(9):959-63
(4)Hyung Min Kim, Byung Hak So, Won Jung Jeong, Se Min Choi and Kyu Nam Park The
effectiveness of ultrasonography in verifying the placement of a nasogastric tube in patients
with low consciousness at an emergency center. SCANDINAVIAN JOURNAL OF
TRAUMA, RESUSCITATION AND EMERGENCY MEDICINE 2012,20:38 Opgehaald op
03 januari 2015 van http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3477076/
(5)Van de Putte P, Perlas A. Ultrasound assessment of gastric content and volume. BR J
ANAESTH. 2014 Jul;113(1):12-22
(6)Tamhne S, Tuthill D, Evans A. Should ultrasound be routinely used to confirm correct
positioning of nasogastric tubes in neonates? ARCH DIS CHILD FETAL NEONATAL ED
2006;91:F388-390
(7)Baskaya M. Inadvertent intracranial placement of a nasogastric tube in patients with head
injuries. SURG NEUROL 1999;52:426-7
(8)Hodin A, Bordeianou L. Nasogastric and nasoenteric tubes. Online 2015. Opgehaald op 20
april 2016 van www.Uptodate.com/contents/nasogastric-and-nasoenteric-tubes/
(9)Metheny N, Meert K, Clouse R. Complications related to feeding tube placement. CURR
OPIN GASTROENTEROL. 2007 Mar;23(2):178-82
(10) van Zanten A, Dixon J, Nipshagen M, et al. Hospital-acquired sinusitis is a common
cause of fever of unknown origin in orotracheally intubated critically ill patients. CRIT
CARE 2005;9:R583–590
(11)Proehl J, Heaton K, Naccarato M, Melanie A. Crowly et al. Emergency nursing resource:
Gastric Tube placement verification J EMERG NURS 2011;37:357-62
(12)Baskin W. Acute complications associated with bedside placement of feeding tubes
NUTR CLIN PRAC 2006;21:40-55
(13)Arbogast D.Enteral feedings with comfort and safety. CLINCIAL JOURNAL OF
ONCOLOGY NURSING 2002;6:275-283
(14)Bourgault A, Halm M. Feeding tube placement in adults; Safe verification method for
blindly inserted tubes. AMERICAN JOURNAL OF CRITICAL CARE 2009;18:73-76
(15)Westhus N. Methods to test feeding tube placement in children. AMERICAN JOURNAL
OF MATERNAL CHILD NURSING 2004;29:283-291
(16)Wilkes-Holmes C. Safe placement of nasogastric tubes in children. PEDIATRIC
NURSING 2006;18:14-17
(17)M. Gordon, Society of Pediatric nurses. Best Evidence: Nasogastric Tube placement
Verification. J PEDIATR NURS 2011 Aug ;26(4):373-376
(18)Mettler F, Huda W, Yoshizumi T, Mahesh M. Effective doses in radiology and diagnostic
nuclear medicine: a catalog RADIOLOGY 2008 Jul;248(1):254-63
(19)Berrington de González A, Darby S. Risk of cancer from diagnostic X-rays: estimates for
the UK and 14 other countries. LANCET 2004 Jan 31;363(9406):345-51
(20)http://www.inami.fgov.be/nl/themas/kost-terugbetaling/door-ziekenfonds/individuele-
verzorging/honoraires/Paginas/arts-deel04.aspx#.Vxzzf3r-DUA. Online opgehaald op 15 april
2016
(21)Vigneau C, Baudel JL, Guidet B, Offenstadt G, Maury E. Sonography as an alternative to
radiography for nasogastric feeding tube location. INTENSIVE CARE MED. 2005
Nov;31(11):1570-2
(22)Nikandros M, Skampas N, Theodorakopoulou M, Ioannidou S, Theotokas M et al.
Sonography as a tool to confirm the position of the nasogastric tube in ICU patients. CRIT
CARE 2006;10(Suppl 1):P216
(23)Bankier A, Wiesmayr M, Henk C.et al. Radiographic detection of intrabronchial
malpositions of nasogastric tubes and subsequent complications in intensive care unit patients.
INTENSIVE CARE MED 1997;23:406-410
(24)Hanson R. Predictive criteria for length of nasogastric tube insertion for tube feeding. J
PARENTER ENTERAL NUTR 1979;3:160
(25)Beckstrand J, Cirgin E, McDaniel A. Predicting internal distance to the stomach for
positioning nasogatric and orogastric feeding tubes in children. J ADV NURS 2007;59:274-
289
(26)Mandal M, Dolai S, Ghosh S, Mistri P, Roy R, et al. Comparison of four techniques of
nasogastric tube insertion in anaesthetised, intubated patients: A randomized controlled trial.
INDIAN J ANAESTH 2014;58:714-8
(27)Margolis G. Paramedic, Airway management. American Academy of Orthopaedic
Surgeons.2004
(28)Hendry P, Akyurekli Y, McIntyre R, Quarrington A, Keon W. Bronchopleural
complications of nasogastric feeding tubes. CRIT CARE MED 1986;14(10):892-4
(29)Hand R, Kempster M, Levy J, Rogol P, Spirn P. Inadvertent transbronchial insertion of
narrow-bore feeding tubes into the pleural space. JAMA 1984;251(18):2396-7
(30) Metheny N, Aud M, Ignatavicius D. Detection of improperly positioned feeding tubes. J
HEALTHC RISK MANAG 1998;18(3):37-48
(31) Dobranowski J, Fitzgerald J, Baxter F, Woods D. Incorrect positioning of nasogastric
feeding tubes and the development of pneumothorax. CANADIAN ASSOCIATION OF
RADIOLOGISTS JOURNAL 1992;43(1):35-9
(32)Hernandez C, Shuler K, Hannan H, Sonyika C, Likourezos A, Marshall J. C.A.U.S.E.:
Cardiac arrest ultra-sound exam: a better approach to managing patients in primary non-
arrhythmogenic cardiac arrest. RESUSCITATION. 2008 Feb;76(2):198-206
(33)Lichtenstein D, Malbrain M. Critical care ultrasound in cardiac arrest. Technological
requirements for performing the SESAME-protocol - a holistic approach. ANAESTHESIOL
INTENSIVE THER. 2015;47(5):471-81
(34)Lichtenstein D, Mezière G. Relevance of Lung Ultrasound in the Diagnosis of Acute
Respiratory Failure :The BLUE Protocol CHEST 2008;134;117-125
(35)Lichtenstein D. BLUE-protocol and FALLS-protocol: two applications of lung ultrasound
in the critically ill. Chest 2015 Jun;147(6):1659-70
(36)Jensen M., Sloth E. , Larsen M, Schmidt M. Transthoracic echocardiography for
cardiopulmonary monitoring in intensive care. EUROPEAN JOURNAL OF
ANAESTHESIOLOGY 2004;21:700-707
(37)Kirkpatrick A, Sirois M, Laupland K, Liu D, Rowan K, et al. Hand-held thoracic
sonography for detecting post-traumatic pneumothoraces: the Extended Focused Assessment
with Sonography for Trauma (EFAST). J TRAUMA 2004;57:288-95
(38)NHS: Patient Safety Alert NPSA/2011/PSA002: Reducing the harm caused by misplaced
nasogastric feeding tubes in adults, children and infants March 2011 . Online 2011.
Opgehaald op 20 april 2016 van http://www.nrls.npsa.nhs.uk/alerts/?entryid45=129640
(39) Taylor S, Allan K, McWilliam H, Toher D Nasogastric tube depth: the 'NEX'
guideline is incorrect. BR J NURS. 2014 Jun 26-Jul 9;23(12):641-4
(40) Gok F, Kilicaslan A, Yosunkaya A. Ultrasound-guided nasogastric feeding tube
placement in critical care patients. NUTR CLIN PRACT. 2015 Apr;30(2):257-60
(41) Atalay Y, Aydin R, Ertugrul O, Gul S, Polat A, Paksu M. Does Bedside Sonography
Effectively Identify Nasogastric Tube Placements in Pediatric Critical Care Patients? Nutr
Clin Pract. 2016 Mar 30
(42)Gray H. Gray's Anatomy: The Anatomical Basis of Clinical Practice; Fortieth edition
2008
(43)Ashraf H, Fikri A Sonographic diagnosis of intraperitoneal free air. J EMERG TRAUMA
SHOCK. 2011 Oct-Dec;4(4):511–513
(44)Larson M. Descriptive Statistics and Graphical Displays CIRCULATION. 2006;114:76-
81
(46)Brun P, Chenaitia H, Bessereau J, Leyral J, Barberis D et al. Contrôle échographique de
la position de la sonde nasogastrique en préhospitalier. ANNALES FRANCAISES
D’ANESTHESIE ET DE REANIMATION 2012;31:416-420
(47)Greenberg M, Bejar R, Asser S. Confirmation of transpyloric feeding tube placement by
ultrasonography. J PEDIATR. 1993 Mar;122(3):413-5
(48)Bourgault A, Heath J, Hooper V, Sole M, Nesmith E. Methods Used by Critical Care
Nurses to Verify Feeding Tube Placement in Clinical Practice CRITICAL CARE NURSE.
2015;35[1]:e1-e7 Opgehaald op 16 april 2016 van http://ccn.aacnjournals.org/content/
35/1/e1.abstract?sid=11152374-e3b2-4474-a2a0-c81683741590
(49) Metheny A, Stewart B, Mills A. Blind insertion of feeding tubes in intensive care units: a
national survey AM J CRIT CARE 2012;21:352-360
(50)Irving S, Lyman B, Northington L, Bartlett J, Kemper C et al. Nasogastric tube placement
and verification in children: review of the current literature. NUTR CLIN PRACT
2014;29:267-276
(51)Glas A, Lijmer J., Prins M, Bonsel G, Bossuyt P. The diagnostic odds ratio: a single
indicator of test performance. Journal of Clinical Epidemiology 2003;56 (11): 1129–1135