1 / 17
Thermage Skin Therapy System (20MHz radiofrequency)
Model #: CRF 007
User manual
2 / 17
Contents Chapter One: About Thermage System…………..……….………………………………3
1. Treatment theory ........................................................................................................ 3
2. Applications ................................................................................................................. 3
3. Features ....................................................................................................................... 3
4. Contraindication .......................................................................................................... 4
5. Specifications .............................................................................................................. 4
Chapter Two: Operation Security Instruction .............................................................. 5
1. Security introduction .................................................................................................. 5
2. Safety of LVD and EMC ............................................................................................ 6
Chapter Three: Installation ............................................................................................... 7
1. Components of the device ........................................................................................ 7
2. Installation and Operation instruction .................................................................... 11
3. Operation instruction notice .................................................................................... 12
Chapter Four: Operation instructions ......................................................................... 13
1. greeting scene: ......................................................................................................... 13
2. Instruction of switches on platform ........................................................................ 14
Chapter Five: Clinic Operation ...................................................................................... 15
1. Clinic operating parameters .................................................................................... 15
2. Attentions ................................................................................................................... 15
3. Course of treatment ................................................................................................. 16
Chapter Six: Maintenance ............................................................................................... 17
1. Normal troubleshooting ........................................................................................... 17
2. Daily maintenance .................................................................................................... 17
3 / 17
1. Treatment theory Capacitance Thermage system is the high-tech beauty equipment, highly safe, lasting effect and non-invasive treatment. The collagen supporting skin tissue, especially the deep collagen and elastin gradually reduce and lost because of sun oxidation and air pollution etc. As the passage of time , elastic fibers and collagen fibers as supporting skin elasticity and tension is gradually reduced, the skin gradually mouth slack, the end of eye droop, the facial contours is not obvious. It shows that the characteristics of skin aging. Capacitance Thermage body sculpting is of intelligent epidermal cooling system. Take advantage of super Thermage producing heating, transfers to deep skin tissue and stimulates the movement of positive and negative ions, which heat collagen tissue under dermal and accelerate blood circulation under dermal and subcutaneous tissue. The fibrous tissue shrink instantly, stimulate long-term collagen freshmen and reconstruction, and can improve facial contours. The super Thermage can directly penetrate 7mm approximate under skin surface, effectively stimulate skin collagen tissue, to promote collagen continuous fresh and achieve natural skin tightening and enhance.
2. Applications ◆ Facial skin tightening, avoiding aging ◆ Wrinkle removal around eyes, nose, forehead, mouth, neck, hand etc. ◆ Skin tightening of sagging skin ◆ Breast enhance ◆ Body contouring ◆ Hand care
3. Features a. The most advanced equipment for wrinkle removal, skin tightening and body contouring
worldwide now
Chapter One: About Thermage System
4 / 17
b. Color touch LCD screen (8.4”), noble and generous, user friendly software, easy operation c. Adopt American “plug and play” handle connectors, and Germany original handle hose d. Capacitance thermage system, never happen electric shock caused by the improper
operation of traditional RF e. Instant skin shrink: the collagen instantly shrink when heat up to 45-600C, therefore the first
treatment will show you instant skin lifting and tightening f. Lasting effect: one course of treatment can last 2-3 years effect
4. Contraindication a. Pregnancy b. Allergic skin c. Serious disease, such as heart disease or unhealthy people d. People who has high expectations for beauty e. People who has metal in body f. People who feed babies g. Nipple and the around area
5. Specifications
POWER SUPPLY OUTPUT 2000W
Radio Frequency 20MHz
Working Mode Pulse
Shots Number No limited
Handpiece Number 1, with 3 probes
LCD Display 8.4” color touch LCD screen
Language English
Dimension 45*50*122cm
Gross weight 28kg
Voltage 220V/110V
5 / 17
1. Security introduction Capacitance Thermage system (20MHz) is a high-tech medical equipment with safety and reliability, mainly for removing surplus fat, body contouring, skin tightening, wrinkle removal and skin rejuvenation. For using the machine safely, operators and assistants must know the security notice in this chapter. Most important, it’s the safest design for both treatments all are as following: 1. Automatically inspection system after turning on the machine 2. Independent electro circuit, automatically stop working when electric leakage 3. The red emergency stop switch facilitates your urgency 4. Key switch to prevent any person free to open the device
【Warning】
The improper operation may causes injury. It will generate high pressure during operation; the operator should be vigilant and operate in accordance with this manual to take the necessary protective measures. Patients: The Safety of patients mainly relies on well-trained physicians as well as the layout
of the appropriate treatment room. Cultural quality of life of patients is also very important; they should understand the principles of treatment. Patients should be protected effectively when treatment.
【MAIN ALARM】
For safe using, should carefully read the alarm notice: 1. Except the authorized mechanic, any others have no right to examine and repair this
structure; especially inner structure inside of machine that including power supply control block, cooling system, and adjustment of probes etc.
2. Make sure whether the voltage of the system is accordant with your resident voltage or not.(220V/110V AC)
3. Maintenance must be done after turning off the machine and drawing out the electrical source line. Or it’s most possible to hurt people and damage the equipment.
4. keep the treatment head/probe clean, avoid any dirty thing absorbed to internal machine 5. Don’t turn on machine in case of water leakage anywhere. Or must cut down power supply
immediately.
Chapter Two: Operation Security Instruction
6 / 17
6. Take using of the ground patch for each treatment, and keep it close touch skin without any space during the whole treatment. (note: we adopt the medical rubber material outside the ground patch, which can offer a more comfortable treatment and no need to use the gel at all.)
2. Safety of LVD and EMC 1. Keep close both panels and cover board, it may cause risk when opening the panel at
random 2. High pressure inside equipment. It’s possible to deposit electricity in some parts even after
switching off. So it’s not allowed to disconnect equipment beyond the factory’s authorization. 3. In maintenance, please turn off machine and disconnect power supply firstly; and then do
observe in person. 4. Care in moving. 5. Ground line connect 【SYSTEM SAFETY SETTINGS】 1. There are some safety settings inside equipment, so please know their places and using
methods 2. Key: it’s used to turn on or turn off machine, only provide by our company 3. Emergency stop switch: the red switch is used to switch off power under emergency. Press
downwards to stop power supply immediately; turn right to recover power supply.
【AUTOMATICALLY INSPECTION】
On turning on the machine, it will be inspecting automatically. About 1 minute running automatically, then it enters into the next step.
【LEVEL OF EQUIPMENT】
1. Electric shock protect: level 1,BF level
2. Corrosion resistance liquid: common
3. Operating way:continuous / time setting
7 / 17
Please open the packaging box to check up whether machine in good condition and the
accessories are complete or not, if machine damaged or the accessories have lost in
transportation, please contact the courier and the factory both at the first time. Thanks. We will
do best to help you!
1. Components of the device
Picture one Picture two
External package for main
machine
Chapter Three: Installation
8 / 17
How to open the package: a. Life up the spanner in picture one, marks in red line b. 900 switch of the spanner and then able to open the package Specification of spare parts:
Spare part name Photo Amount
1 Handpiece
1
2 Treating probe 3
3 Grand patch
1
4 Water input 1
5 Foot pedal 1
9 / 17
6 Power cable 1
7 Keys 2
Treatment probes x 3pcs (sequence from left to right) Smallest probe: for the arrow area treatment, such as eyes surrounding
Middle probe: for facial treatment
Big probe: for body treatment
10 / 17
About mainframe
Handpiece and treatment probes
Handpiece holder
Emergency stop switch
keys
controller
Connector of power line
Power control switch
Power control switch
Water inlet
Handpiece socket
Water drain
Water overflow
Connector of grand patch
11 / 17
2. Installation and Operation instruction Placement of device: place the machine on level ground, make sure the treatment room is clean and it is about 22~28℃ and humidity more than 20%. Input water:
A. water source: distilled water or pure water ; B. screw off the caps for “water inlet” and “water overflow” at the same time plug water pipe
& funnel to “water inlet” spill from “water spilt” automatically means enough water input.
Step 1: Switch off both caps for water inlet and Water overflow at the same time Step 2: connect tools for water input till water spilt from water overflow automatically.
Attention: Do not tighten the caps for water inlet and water overflow till good water circulations (means probe turns to cold). Install of handpiece:
Handpiece plug
Press the white button
downwards firstly Plug to machine side, “click” sound means good connection
as well as the white button bounce at the same time
Handpiece socket
12 / 17
Afterwards connect the power line and foot pedal in turn. After all above installation, please insert keys and turn right to turn on machine. The water circulation system works at the same time, the treatment probe will turn to cold 3 minutes later if good water circulations.
3. Operation instruction notice Thermage system can be used in hospital or beauty salon. attentions in stallation is very important: ◆ Open package, place machine on flat ground. We suggest that place machine in a fix place. ◆ Check up machine and accessories, make sure no damage, no loss ◆ Make sure good condition of power supply, including the ground wire ◆ Good condition of all accessories ◆ Take care water away from the power socket ◆ Function test
13 / 17
1. greeting scene: Turn on machine and the first scene is greeting scene
1. Alarm of water temperature:
2. Alarm of water flow:
Water flow alarm Water temperature alarm
Treatment menus
System settings
clock
Introduction of alarm:
Alarm of water temperature: when temperature is higher than
system setting‐‐‐‐550C, the indicator will show as this and alarm
sound following at the same time. In response, the machine will
stop all working immediately
Alarm of water flow: no water or little water or no water
circulation inside, the indicator will show as this and alarm sound
following at the same time. In response, the machine will stop all
working immediately
Chapter Four: Operation instructions
14 / 17
2. Instruction of switches on platform
BUTTON FUNCTION OPERATION
1 Power adjust energy from 5-300W/cm2 touch “△” to increase energy; Touch “▽” to reduce energy
2 Time setup of working time from 5-90minutes
touch “△” to increase; Touch “▽” to reduce
3
adjust probe temperature cyclical change by touch continously
4 save parameter settings touch “save” and do setting directly. It does not affected by turn on/off machine , means show your settings in next turnon machine.
5 Ready let machine under working condition touch “ready” only
6 Standby stop working condition touch “standby” only
7 Return return to last platform touch only
1,2,3 present 3 different groups of parameters setting.
counte
Cooling level adjustment: cyclical change by finger touch
Time settings Energy adjustment
Switch of “save”
15 / 17
1. Clinic operating parameters
TREATMENT
AREA POWER TIME
COOLING LEVEL
1 smallest probe
Eye surrounding
60-80W/cm2 adjustable according to actual condition of patients
30 minutes approximate
5 level
2 middle probe
face 60-100W/cm2 adjustable
according to actual condition of patients)
40 minutes approximate
5 level
3 biggest probe
body
120-180W/cm2 adjustable according to
actual condition of patients)
40 minutes approximate (mainly
depend on the treatment area)
5 level
2. Attentions a. It’s necessary and important to use the ground patch during the whole treatment. Please
Chapter Five: Clinic Operation
16 / 17
keep it close touch skin and avoid some special parts of body, such as heart. (it’s not allowed to keep people’s heart, treatment probe and ground patch in one line, avoid any potential risk to people’s heart. It’s no need to spread cooling gel on ground patch, keep it touch skin directly but close please)
b. Keep probe vertical & close touch skin during the whole treatment c. If move probe away from skin, please release foot pedal firstly (generally step pedal
continuously in treatment). If stop treatment, please release foot pedal firstly also. d. Before start the treatment, please adjust parameters and put probe on treatment area
vertically and closely; and then touch “ready’ and continuously step pedal. e. After treatment, please keep probes clean.
3. Course of treatment
TIME TREATMENT
TREATMENT TIMES NEEDED / 3
MONTHS
INTERVAL BETWEEN TWO TREATMENTS
Body contouring, wrinkle removal, skin tightening, skin rejuvenation
The first month twice per week
The second month one time per week
The third month one treatment every two weeks
17 / 17
1. Normal troubleshooting
SOLUTION No power in the
machine
A. Check the emergency button, keep it bounced up in working; or power off to machine
B. Check up the power supply control switch at the rear panel. Press the side of “I” downwards, means power on
C. check the power line broken or not, and whether there is short circuit
D. check the fuse good or not E. check the external power supply system
Machine stop working immediately
A. high pressure interfere, check up whether high-power electrical around or not
B. the software out of control, should replace the microchip or replace the controller
Nothing feeling of energy
A. check up whether treatment probe close touch skin or not B. check up whether counter works normally or not C. check up whether good connection of probe
2. Daily maintenance a. Keep machine away from sun exposure b. Cover treatment probes and mainframe after treatments, avoiding dust cover. It’s very
important c. Carefully hand the treatment probes (important also) d. Change water every 3 months (pure water or distilled water only) e. Place machine in dry and dust-free room (very important) f. Non-professional operator is not allowed to operate this machine
Chapter Six: Maintenance