Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis Scott Peterman 1 , David Sarracino 1 , Shen Luan1, Xiaolei Xie 1 , Edward Goucher 1 , Amol Prakash 2 , and Bradley Hart 1 1 Thermo Fisher Scientific BRIMS, Cambridge, MA; 2 Optys Technologies, Boston, MA Poster Note 64730 ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 μL aliquot was diluted 100-fold with 90:10:0.1 solvent of H 2 O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo Scientific TM Q Exactive TM Focus mass spectrometer equipped with a Thermo Scientific TM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 μL of the extracted hemoglobin samples directly onto an SephacrylS100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1) Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events. 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min) 2.72 2.90 2.65 2.97 2.59 3.09 2.53 2.47 2.41 2.34 Directed to Waste Directed to Mass Spectrometer 3.21 The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination. 797.11 650 700 750 800 850 900 950 1000 1050 1100 m/z 841.34 z=18 z=19 757.30 z=20 890.77 z=17 721.24 z=21 946.38 z=16 882.47 z=18 934.32 z=17 688.50 z=22 836.08 z=19 794.32 z=20 1009.33 z=15 992.71 z=16 658.61 z=23 1058.69 z=15 1081.43 z=14 690.85 z=23 662.10 z=24 719.15 z=22 2A 450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z 759.91 z=4 946.32 z=16 890.71 z=17 654.36 z=1 802.42 z=3 682.37 z=2 832.46 z=8 738.91 z=2 1134.32 z=14 994.51 z=8 714.06 z=3 451.27 z=1 512.77 z=2 553.31 z=3 609.09 z=4 967.87 z=8 1190.33 z=3 1086.58 z=3 2B Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 μscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.
Transcript
Page 1: Thermo Fisher Scienti c BRIMS, Cambridge, MA; β High ... · Development of a High-Throughput Hemoglobinopathies Work ow Using 30 High Resolution Accurate Mass Analysis Scott Peterman

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1 1Thermo Fisher Scientific BRIMS, Cambridge, MA; 2Optys Technologies, Boston, MA

Po

ster No

te 64

730

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Page 2: Thermo Fisher Scienti c BRIMS, Cambridge, MA; β High ... · Development of a High-Throughput Hemoglobinopathies Work ow Using 30 High Resolution Accurate Mass Analysis Scott Peterman

2 Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100 R

elat

ive

Abu

ndan

ce

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Page 3: Thermo Fisher Scienti c BRIMS, Cambridge, MA; β High ... · Development of a High-Throughput Hemoglobinopathies Work ow Using 30 High Resolution Accurate Mass Analysis Scott Peterman

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

Scott Peterman1, David Sarracino1, Shen Luan1, Xiaolei Xie1, Edward Goucher1, Amol Prakash2, and Bradley Hart1, 1Thermo Fisher Scientific BRIMS, Cambridge, MA, 2Optys Technologies, Boston, MA

MATERIALS AND METHODS – cont. RESULTS Successful hemoglobinopathies routines must confidently determine normal hemoglobin chains or variants. Therefore sample prep and data acquisiton must facilitate integrated and reliable search strategies. The knowledgebase compiled from the analysis of normal plasma (Figs. 2-4) are used as the first data set to screen for presence/absence of variants, and then facilitate characterization. To simulate the increased complexities of variant hemoglobin samples, normal human plasma was spiked with known amounts of bovine hemoglobin, doubling the number of chains. While the mass shifts between human and bovine hemoglobin chains are not as close as some reported human variants, the increased complexity does stress the independent data acquisition routines for both precursor charge state determination and fragmentation analysis. Thus the determination of sequence specific fragments from each chain become key in characterizing hemoglobin content. Finally, the routine must be reproducible and robust to handle the sample load facing most institutions. Therefore a larger number of replicates were prepared and analyzed to determine reproducibility. A set of 25 samples per spiking mixture was created in two 96-well plates and processed in a bulk manner. The resulting sample analysis was acquired in an automated manner and evaluation of response factors and carry-over was measured by randomizing the autosampler injection sequence. Of critical importance was the ability to reproducibly detect all hemoglobin chains, maintain mass accuracy for the majority of charge states per hemoglobin chain, determine the chain ratios, and measured the relative abundance of human to bovine ratios.

ABSTRACT A high-throughput hemoglobinopathies workflow is presented which is designed to perform research to automated data acquisition and processing for normal human alpha and beta chains, as well as profiling for abnormal hemoglobin chains. The presentation demonstrates a single method to profile samples regardless of complexity and the data acquisition scheme facilitates sequence detection for small mass deviations from normal hemoglobin chains. A large set of samples were used to evaluate the workflow, consisting of normal human hemoglobin (from dilute DBS) as well as aliquots spiked with known amounts of bovine hemoglobin. Spiking ranges were ca. 9-fold for both conditions (human:bovine and bovine:human). Results show excellent reproducibility with %CVs less than 6% across the 25 samples per spiking level and %RDS for measured AUC ratios were less than 10% for the expected SNP ranges and 20% for the extreme spiking ranges. INTRODUCTION Hemoglobin profiling incorporates detecting, characterizing, and determining relative quantitation of all hemoglobin chains arising from normal protein sequence translation as well as sequence variants consisting of SNPs, truncations, and omissions. In addition, PTMs associated with glycation must be detected and quantified. Currently there are over 1000 reported hemoglobin forms prohibiting effective screening methods incorporating targeted/biased detection methods. Recently, mass spectral detection based on HRAM MS and AIF has emerged as a suitable method to perform hemoglobinopathies due to the advances in resolution and reproducible product ion detection over large dynamic ranges. MATERIALS AND METHODS Hemoglobin samples were created directly from a stock of whole blood from a healthy donor. A 10 µL aliquot was diluted 100-fold with 90:10:0.1 solvent of H2O/MeOH/Acetic Acid and loaded into separate wells of a cation exchange plate. A bovine hemoglobin (purchased from Sigma Aldrich) stock solution and used to spike into selected sample wells at known concentrations to increase the complexity. The sample set was comprised of 20-25 replicates of neat human plasma, 9:1 human/bovine, 7:3, 5:5, 3:7, 1:9, and neat bovine hemoglobin. The plate was washed to remove salts and small molecules, and then desired molecular weight range was displaced and collected using an ammonium formate solution. All experiments were performed on a Thermo ScientificTM Q ExactiveTM Focus mass spectrometer equipped with a Thermo ScientificTM Vanquish TM UHPLC autosampler and pump, with the peak tubing running through a 6-port divert value. The autosample loaded 20 µL of the extracted hemoglobin samples directly onto an Sephacryl™ S100 SEC column (GE Healthcare Life Sciences) for concentration washing for 1.5 minutes, then the divert valve was switched to the mass spectrometer for data acquisition for 3.5 minutes prior to diverting the LC flow back to waste to clean and condition the column for the next injection. (Figure 1)

CONCLUSIONS The presentation shows an integrated workflow for high-throughput hemoglobinopathies research using the Q Exactive Focus mass spectrometer. The results demonstrate an effective method to combine robust sample preparation, delivery into the mass spectrometer, qualitative and quantitative data acquisition, and routine data processing. The workflow enables 384 samples to be analyzed per day with automated data processing on a single autosampler/LC stack. Specific attributes relative to different workflow aspects include: • Sample preparation

• Method can support DBS or plates • Plate methods require ca. 8 minutes for clean up

• Introduction into the mass spectrometer • Utilization of 6-port valve to load on column, and wash salts prior to diverting flow to ion

source • Coupled with sample preparation steps, increased ruggedness of ion source

• Data acquisition • Alternating HRAM MS and AIF enabled many data points to be acquired across the 1.5

minute elution profile • AIF increased fragment ion detection from a wide range of precursor charge states • Use of 10 microscans significantly increased spectral quality for automated sequencing

• Data processing • The Pinnacle software leverages known protein sequences to expedite data

characterization and facilitates screening for truncations, PTMs, and sequence variants • Precursor and product ion isotopic profiling was used to increase confidence in assigning

hemoglobin sequence to each chain

For research use only. Not for use in diagnostic procedures. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Development of a High-Throughput Hemoglobinopathies Workflow Using High Resolution Accurate Mass Analysis

797.11

650 700 750 800 850 900 950 1000 1050 1100 m/z

841.34 z=18 z=19 757.30

z=20 890.77 z=17

721.24 z=21 946.38

z=16 882.47 z=18 934.32

z=17 688.50 z=22 836.08

z=19 794.32

z=20 1009.33

z=15 992.71

z=16 658.61 z=23

1058.69 z=15 1081.43

z=14 690.85 z=23 662.10

z=24

719.15 z=22

2A

Figure 1. Representative data acquisition strategy for the routine hemoglobin profiling experiment showing the chromatographic profile for direct sample loading, washing, and data analysis using alternating HRAM MS (black lines) and AIF (red lines) scan events.

1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.5 3.7 Time (min)

2.72 2.90 2.65 2.97 2.59

3.09 2.53

2.47 2.41

2.34 Directed to Waste

Directed to Mass Spectrometer

3.21

450 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

759.91 z=4

946.32 z=16

890.71 z=17 654.36

z=1 802.42 z=3

682.37 z=2

832.46 z=8 738.91

z=2

1134.32 z=14 994.51

z=8 714.06 z=3 451.27 z=1

512.77 z=2

553.31 z=3

609.09 z=4 967.87

z=8 1190.33

z=3 1086.58 z=3

2B

β 2-147

α 2-146

α chain 2-142

β chain 2-147

Figure 2. Example data for neat human whole blood analysis following sample preparation and LC analysis where 2A shows full scan HRAM MS data and 2B shows the AIF spectrum. Each spectrum is one scan consisting of 10 µscans. The peaks labeled in blue font are from the alpha chain and red font represents charge states attributed to the beta chain.

Figure 3. Automated protein sequence determination in the Pinnacle software based on default Uniprot sequences and partial cleavage assumptions. The insets shows the matched isotopic profiles used for sequence determination of the partial sequence for the +16 charge state for the alpha chain and +17 charge state for the beta chain. Matching algorithms compare empirical isotopic and charge state m/z values with theoretical values for all possible sequences (assuming partial and PTMs) to identify the highest scoring match.

Figure 4. Automated AIF data analysis for the refined alpha and beta chain sequences. The matched product ions are marked in red triangles based on 10 ppm mass tolerance, charge state determination, and isotopic distribution for the proposed sequence (Fig. 3). Peptide diagrams for each chain are displayed in the inset with b- and y-type coverage maps indicated.

600 650 700 750 800 850 900 950 1000 1050 1100 1150 m/z

10 20 30 40 50 60 70 80 90

100

Rel

ativ

e Ab

unda

nce

841.34 z=18 797.06

z=19 757.25

z=20 890.71 z=17

837.28 z=18

721.29 z=21 793.27

z=19 886.41

z=17 946.31

z=16 753.61 z=20

882.41 z=18 717.81

z=21

688.55 z=22

934.32 z=17

1009.40 z=15 992.59

z=16 685.19

z=22 1081.43 z=14 1058.76

z=15 658.61 z=23 1134.31

z=14 655.48

z=23 1164.54 z=13

Human α 2-142 Human β 2-147 Bovine α 2-142

881 883 885 887 889 891 893 m/z

890.71 z=17 886.41

z=17 882.47 z=18

887.41 z=18

Bovine α 2-142

Human α 2-142

Human β 2-147

Bovine β 1-145

Human α 2-142 Human β 2-147 Bovine α 2-142

Bovine β 1-147

881 882 883 884 885 886 887 888 889 890 891 892 m/z 0

50

100 0

50

100 0

50

100

Rel

ativ

e Ab

unda

nce

0

50

100 0

50

100 890.77

z=17 882.47 z=18

886.53 z=17

887.41 z=18

890.77 z=17 882.41

z=18 886.41 z=17 887.41

z=18

890.71 z=17 886.41

z=17 882.47 z=18 887.36

z=18

886.41 z=17 890.71 z=17

887.30 z=18

882.41 z=18

886.42 z=17 887.36 z=18 890.77

z=17 882.41 z=18

9:1

7:3

5:5

1:9

3:7

The data acquisition sequence consisted of alternating high resolution/accurate mass (HRAM) MS and all ion fragmentation (AIF) events for the 3.5 minutes of data acquisition. Data acquisition parameters were consistent for both MS and AIF scan events including AGC = 5e6, max ion fill times of 100 msec, 10 microscans, and a resolution setting of 70,000 (FWHM) at m/z 200. Figure 2 shows representative spectra from the HRAM MS (2A) and AIF spectra (2B) from the human hemoglobin. All data processing was performed in the Pinnacle software (Optys Technologies). Automated qualitative and quantitative data analysis utilized a FASTA file of the reported alpha and beta human and bovine hemoglobin sequences as downloaded from the UniProt database. (www.uniprot.org) Search parameters used partial sequence considerations as well as dynamic modifications for oxidation, deoxy, and hexose. A mass tolerance of 10 ppm was used for precursor and product ion isotopic determination.

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

0 5 10 15 20 25

AUC

Val

ues

Injection Number

Human α

Human β

Bovine β

Bovine α 3:7

5.4

4.1

6.4

5.2

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

2.5E+10

3.0E+10

3.5E+10

4.0E+10

4.5E+10

0 5 10 15 20 25

AU

C V

alue

s

Injection Number

Human α

Human β

Bovine α

7:3

4.7

3.9

5.7

Bovine β 4.4

7.43 (17.4%)

2.31 (1.1%)’

1.00 (0.2%) 0.43

(1.4%)

0.13 (20.6%)

Figure 5. HRAM MS analysis of a hemoglobin mixture from human and bovine. The spiked bovine levels were ca. 1:1 with estimated human amounts. The charge states are labeled by the color coded key. Due to the narrow mass difference, not all detected charge states are labeled. The inset shows a narrow mass range to demonstrate the precursor distribution identification between the different alpha and beta chains.

Figure 6. Comparative detection levels for the human and bovine hemoglobins from the different spiking levels used in the study. The displayed mass range covers the +17 and +18 precursor charge states for the alpha and beta chains, respectively. The spiked ratios are labeled for each spectrum (human:bovine).

Figure 7. Comparative AIF spectra covering different spiking levels of human:bovine. The labeled fragment ions are matched based on sequences identified in Figure 4 and 5. Nomenclature used to label fragment ions indicate species, chain, fragment type, and position. The predominant fragments from human hemoglobin are labeled in the 9:1 spectrum and the predominant bovine hemoglobin fragments are labeled in the 1:9 spectrum.

0 20 40 60 80

100

Rel

ativ

e A

bund

ance

759.914 z=4

802.426 z=3 767.445

z=? 749.844 z=2 682.367

z=2 832.466 z=8 862.212

z=12 738.910 z=2 795.813

z=10 843.467 z=8 714.057

z=3 690.038 z=3 868.232

z=4 811.783 z=3 781.419

z=5 732.724

z=3 850.994 z=9 822.377

z=1 725.407 z=2 693.393

z=?

705.399 z=3

7:3

0 20 40 60 80

100 767.162

z=4 690.038

z=3 759.915 z=4

843.468 z=8

749.845 z=2 802.426

z=3 732.725 z=3 832.467

z=8 739.411 z=2

682.368 z=2 862.130

z=12 770.421 z=3

787.214 z=9 818.951

z=4 714.059 z=3

868.736 z=4 811.783

z=3 850.995 z=9 693.394

z=2

5:5

680 690 700 710 720 730 740 750 760 770 780 790 800 810 820 830 840 850 860 870 m/z

0 20 40 60 80

100 767.162

z=4 690.038

z=3 843.467 z=8 732.724

z=3 749.844

z=2 835.793 z=8 787.213

z=9 770.420 z=3 868.233

z=4 818.699 z=4 759.915

z=4 683.340 z=2 802.424

z=6 850.994 z=9 774.432

z=5 714.059 z=?

693.393 z=2

3:7

0 20 40 60 80

100 759.914

z=4

802.425 z=3 682.367

z=2 832.464 z=8 738.909

z=2 767.444 z=1

862.129 z=12 749.844

z=2 795.812 z=10 714.057

z=3 841.394 z=18 811.782

z=3 781.419 z=5 822.377

z=2 860.216

z=9 868.231 z=4 725.406

z=2 699.888 z=4

9:1 Hβb13

Bβy19

Hβy20

Hβb22 Hβy96

Hαy28

Hαy6 Hβb14

Bβy20

Bαy28

Hβy14 Hβy36 Hβy23 Hαy61

Bβb31

Bαb65 Bαb68 Bαb64

Bβb43 Bβb33

Hαb75

Rel

ativ

e A

bund

ance

Figure 9. Summary report for quantitative hemoglobin spiking analysis. Histogram shows average AUC values and error bars for each chain per spiking ratio. In addition, the AUC ratios and %RSDs is reported between human and bovine. The ratios were corrected based on the estimated 1:1 ratios.

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

05

101520253035404550

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range (Da)

9:01

7:03

5:05

3:07

1:09

0

10

20

30

40

50

60

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

0102030405060708090

450-600 600-800 800-1000 1000-2000

Num

ber o

f Fra

gmen

t Ion

s

Product Ion m/z Range

9:01

7:03

5:05

3:07

1:09

Human α Chain

Bovine α Chain Bovine β Chain

Human β Chain

Figure 8. Evaluation of the workflow reproducibility based on the analysis of 25 replicate samples. The coefficient of variance was determined from measured AUC values for each hemoglobin chain per sample. Two representative plots are shown for the hemoglobin mixtures of 7:3 (Fig. 8A) and 3:7 (Fig. 8B) human/bovine. Reported hemoglobin chain AUC values were based on summed charge states and isotopes, as presented in Fig. 3. The %CVs are labeled by each chain. The trend shows excellent agreement for the species response, as the AUC trends track across the 25 samples. In addition, the AUC ratios between chains from the same species are consistent.

8A

8B

Figure 10. AIF product ion detection per hemoglobin chain across the different spiking ratios. The histograms are broken down into product ion m//z range. The key at the right is conserved for all histograms. Fragment ion confirmation was based on mass accuray, correct charge state and product ion isotopic distribution overlap (determined based on dot product correlation coefficients) for b- and y-type ions. The Pinnacle software also determines uniqueness of fragment ion association across the four hemoglobin chains.

PN64730-EN 0616S

Africa +43 1 333 50 34 0Australia +61 3 9757 4300Austria +43 810 282 206Belgium +32 53 73 42 41Brazil +55 11 2730 3006Canada +1 800 530 8447China 800 810 5118 (free call domestic)

400 650 5118

Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0Finland +358 10 3292 200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591

Japan +81 6 6885 1213Korea +82 2 3420 8600Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00

Russia/CIS +43 1 333 50 34 0Singapore +65 6289 1190Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00Taiwan +886 2 8751 6655UK/Ireland +44 1442 233555USA +1 800 532 4752

www.thermofisher.com©2016 Thermo Fisher Scientific Inc. All rights reserved. Sephacryl is a trademark of GE Healthcare Lifesciences. UniProt is a trademark of the European Molecular Biology Laboratory. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001Certified.


Recommended