Thermo Scientific Matrix Electronic Pipette User Manual
Transcript
Page 1: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

Thermo Scientific Matrix Electronic Pipette

User Manual

Page 2: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

Manual N09238

Thermo Scientific Matrix Electronic Pipette

User Manual

Page 3: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

ii

CONTENTS

Congratulations! ............................................................................................. 1Before You Begin ............................................................................................. 1

A. INTRODUCTION ....................................................................................... 2

A.1ThermoScientificMatrixSingleChannelPipette ................................................3

A.2ThermoScientificMatrixMultichannelPipette ....................................................3

B. PIPETTE OVERVIEW ................................................................................. 4

B.1Components ........................................................................................................4

B.2GeneralDescription ............................................................................................5

B.3Keypad ................................................................................................................6

B.3.1OperationKeys ................................................................................................6

B.3.2ProgrammingKeys ...........................................................................................7

B.3.3SpecialFunctionKey .......................................................................................8

B.4Display ..............................................................................................................10

B.5PipetteSpecifications ........................................................................................11

B.6HazardsandPrecautions ..................................................................................12

C. PROGRAMMING AND OPERATION ....................................................... 13

C.1ProgrammingtheElectronicPipette .................................................................13

C.1.1OpenProgrammingModeor“ScratchPad”(P-0) .........................................13

C.1.2EnteringaPrograminOpenProgrammingMode ..........................................13

C.1.3StoringaProgram ..........................................................................................16

C.1.4ProgramRecall ..............................................................................................17

C.1.5AdjustingPipettingSpeed ..............................................................................17

C.1.6ReviewingtheProgram ..................................................................................17

C.2ProgrammingExamples ....................................................................................18

C.2.1SampleTransfers ...........................................................................................18

C.2.2IncrementalPipetting .....................................................................................19

C.2.3 Serial Dilutions ...............................................................................................19

C.2.4SimpleDilutions .............................................................................................20

C.2.5SupernatantCollection ...................................................................................20

C.3OperatingProcedures .......................................................................................21

Page 4: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

iii

D. CALIBRATION ......................................................................................... 23

D.1VolumeMeasurement .......................................................................................23

D.2PipetteCalibration .............................................................................................24

D.2.1FactoryCalibrationStep ................................................................................24

D.2.2“EASY”CalMode ...........................................................................................24

D.2.3“CALC”Mode .................................................................................................25

D.2.4CALVIEW .......................................................................................................27 D.2.5CALRESET ....................................................................................................27 D.3RecommendationsforAccuratePipetting .........................................................27

E. TROUBLESHOOTING .............................................................................. 29

F. MAINTENANCE ........................................................................................ 31

F.1SingleChannel Maintenance .............................................................................32

F.1.1SingleChannel Disassembly ..........................................................................32

F.1.2SingleChannelCleaningandAutoclaving ......................................................34

F.1.3PipetteReassembly ........................................................................................34

F.2Multichannel Maintenance .................................................................................35

F.2.1Multichannel Disassembly ..............................................................................35

F.2.2MultichannelCleaningandAutoclaving ..........................................................38

F.2.3MultichannelMaintenanceandReassembly ..................................................38

F.3Installing/ReplacingBatteries ............................................................................44

F.3.1Installing/ReplacingNiMHBatteries ...............................................................44

F.3.2Installing/ReplacingLithium-IonBattery(16Ch) ............................................48

F.4Charging/RechargingBatteries ..........................................................................50

F.4.1Charging/RechargingNiMHBatteries .............................................................50

F.4.2Charging/RechargingLithium-IonBattery(16Ch) ..........................................51

G. TECHNICAL SERVICE ............................................................................ 53

G.1InstrumentWarranty .........................................................................................53

G.2ExpertPipetteService ......................................................................................54

G.3OnSitePipetteServiceOptions .......................................................................55

G.4RecommendedEquipmentDecontaminationMethods ....................................56

G.5TipsandReservoirs ..........................................................................................58

Page 5: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available
Page 6: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

1

Congratulations!

Youhavepurchasedthestate-of-the-artinprecisionpipetting.TheThermoScientificMatrixElectronicPipetteisthefirstcordlesselectronicpipettethat“understands”themyriadofliquidhandlingprotocolstypicallyperformedintoday’slaboratory.

Ergonomicallydesignedandbalancedtoreducefatigue,theMatrixelectronicpipettefunctionssointuitively,youmayneedonlyaquicksurveyofthismanual to use it.

Theelectronicpipetteexecutesyourliquidpipettingtasksquickly,accurately,andwithlessriskofdevelopingrepetitivestraininjuries(RSI)typicallyassociatedwithmanualpipetteuse.Laboratoryresultswillbecomemorestandardized,whilelesstimeisexpendedontediousbenchprocedures.

Before You Begin

BeforeusingtheMatrixelectronicpipette,verifythecontentsofthepackage. Thepackageshouldcontainthefollowingitems:

1ElectronicPipette

1Operator’sManual

2RechargeableNiMHBatteriesor1Lithium-IonBattery(16 Channels)

1ACPowerTransformer

1WarrantyCard

1Accuracy/PrecisionDataSheet

Ifitemsaremissing,contactaThermoFisherScientificrepresentativeorcontactcustomerserviceat 800-345-0206.

Note: Batteries must be installed and recharged before using the Matrix electronic pipette Refer to the following sections in this manual:

Section F.3, Installing/Replacing Batteries Section F.4, Charging/Recharging Batteries

Page 7: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

2

A. INTRODUCTION

TheThermoScientificMatrixElectronicPipetteprovidestheconvenienceofamanualpipettor,withtheprogrammingcapabilitiesofanelectronicpipette.TheMatrixelectronicpipettefeaturesinclude:

•Expandedmemoryforstoringuptofivepipettingprograms.

•Eachprogramcanholdamaximumof40steps.

•“ScratchPad”programmingmodewhichallowsforfastandsimpleprogramchangesthatarenotstoredintomemory.

•Enhancedspeedcontrolofferssloweraspiratinganddispensingofviscousliquids.Ideal for gel loading!

•“EASY”CALwhichsimplifiesin-labcalibrationverificationusing distilledwater.

TheMatrixelectronicpipetteislightweightandcordless,utilizingrechargeablebatteries,thatcanbeusedduringrecharge(with the exception of single channel pipettes).Thepipettealsohasanautomaticshut-offfeaturetoextendbatterylife.

Theelectronicpipetteusesintegratedcircuittechnologytostoreprogramminginformation.Itwillholdupto40stepsinaprogram,withupto5programssavedinmemory,and1scratchpad(See section C.1.1 Open Programming Mode or “Scratch Pad”).Theprograminformationisretainedinmemoryevenifthebatteriesarefullydischargedorremoved.

Theelectronicpipette usesa“paceddispense”featurethatperformsrepeatpipettingstepswhilethetriggerisdepressed.

Page 8: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

3

Introduction

A.1 Thermo Scientific Matrix Single Channel Pipettes

Thesinglechannelelectronicpipetteareavailableinanumberofmodelconfigurations(see Figures A-1)from12.5µlto1250µl,ranginginvolumefrom0.5µlto1250µl.Eachpipetteiscolor-codedforeasyvolumeidentification(see section B.5, Pipette Specifications).Twomodels,250µland1250µl,areofferedwithashorterejectorsleevereducingtheoveralllengthcomparedtothestandardpipette.Idealforuseinconfinedspacessuchasbiosafetycabinet

A.2 Thermo Scientific Matrix Multichannel Electronic Pipettes

Themultichannelelectronicpipettesareavailableinanumberofconfigurationsfrom8to16channnels.AnexpandableEXPpipetteisalsoavailablein6-channelandcustomconfigurations.Modelsrangeinvolumefrom0.5µlto1250µl.Eachpipetteiscolor-codedforeasyvolumeidentification(see section B.5, Pipete Specifications).TheThermoScientificMatrixEqualizer®pipettehasequaltipspacingbetween4.5mmand14.15mm(depending on model).

6-ChannelExpandable

8-Channel 12-Channel 16-Channel384Model

1250µl 250µl 125µl 30µl 12.5µl1250µlShorty

250µlShorty

Figure A.1: Single Channel Models

Figure A.2: Multichannel Models

Page 9: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

4

Pipette Overview

B. PIPETTE OVERVIEW

B.1 Components

Figure B.1b: Multichannel Components

Figure B.1a: Single Channel Components

Page 10: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

5

Pipette Overview

B.2 General Description

Keypad.Thekeypadcontainstheoperationkeysthatareusedinprogrammingthepipettingsequenceandvolumes. (See Section B.3, Keypad).

Display.ThedisplayisanLCDscreenthatshowsthepipettingoperation(fill, dispense, mix, or purge)andfluidvolumeforeachstepinthepipettingprogram.(See Section B.4, Display).

Battery Case.Thebatterycasecontainstherechargeablebatteries.NickelMetalHydridebatteriesfullychargein14-16hours,whileLithium-Ionbatteriesfullychargein2hours.Thepipettecanperformafullday’sworkloadwithnoneedtorecharge.Batterylifeisextendedwithautomaticshut-offafter 10minutesofinactivity.(See Section F.3, Installing/Replacing Batteries).

Pipetting Trigger.Thepipettingtriggerinitiatesaprogram.Asinglepipettingstepisexecutedbyaquickpressandreleaseofthetrigger.Repeatpipettingisexecutedbypressingandholdingthetrigger.

Note: At the end of any dispense, the pipette tips should be removed from the fluid before releasing the trigger, to prevent aspiration of fluid.

Ejector Trigger.Theejectortriggerispressedtoejectthetips fromthepipette.

Tip Fittings.Thetipfittingsformanairtightsealwiththetipstoensurepipettingaccuracy.

Note: Electronic pipettes are equipped with chemically-resistant (PEEK) tip fittings (and ejector sleeves on single channel models) that allow pipetting of acidic solutions without damage to the tip fittings.

Color-coded Ring.Theringaroundthepipettehandleiscolorcodedtoidentifythepipettingvolumeofthehandle(See Section B.5, Pipette Specifications).

Transformer Connector.Thetransformerconnectorreceivestheplugfromthepowertransformerforrechargingthebatteries.Thepipettecanbeusedwhilerecharging(with the exception of single channel pipettes).

Page 11: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

6

Pipette Overview

B.3 Keypad

Thekeypadcontainstheoperationkeys,programmingkeys,andthespecialfunctionkey.

Figure B.3: Electronic Pipette Keypad

B.3.1 Operation Keys

[On/Off] TheOn/Offkeyturnsthepipetteonoroff.Whenthepipetteisturnedon,itemitsabeeptone,flashes“ZERO”onthedisplay,thenautomaticallyresetsthepistonstozeroposition.Afterzeroing,thedisplayshowsthefirststepoftheprogramintheRUNmode.

Note: If the pipette is being programmed for the first time, the first step will be the fill operation, with the default fill volume. (See Section C.1, Programming the pipette, for the default settings).

[Prog] TheProgramkeyontheMatrixelectronicpipetteenablesyoutostoreuptofiveseparateprogramsinmemoryforfuturerecall.Eachprogramcanholdupto40steps.Programsarestoredinreservedprogramlocationnumbers1–5.Program“0”isusedasa“scratchpad”wheretemporaryprogramscanbewritten,run,andoverwritteneasily.Program“0”isconsideredtobean“open”program.(See Section C.1, Programming the Pipette).

Page 12: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

7

Pipette Overview

[Purge] ThePurgekeyentersapurgepromptanytimeduringapipettingoperationtoaborttheprogram.BypressingthePurgekey,thenpressingthetrigger,youcandispenseanyremainingliquidfromthepipettetipsandreturntothebeginningoftheprogram.

ThePurgekeyisalsousedtoenterthepurgestepintoaprogram.

Note: If there is insufficient liquid for a final dispense step, the pipette emits a beep tone and automatically displays the purge prompt.

Note: The RUN mode performs the operation that is displayed, such as fill, dispense, mix, or purge. When “Run” appears in the display, you can activate the RUN mode by pressing the trigger. At the end of a pipetting sequence, the pipette emits a triple beep tone and returns to the beginning of the program. The pipette remains in RUN mode until you program a new pipetting sequence.

B.3.2 Programming Keys

[Fill]

[Disp]

[Mix]

TheFill,Dispense,andMixkeysprogramthepipettingoperations.Bypressingthefillormixkey,youcanentertheprogrammingmodeoftheelectronicpipette.The“Run”promptdisappearsfromthedisplayandtheappropriateoperationappearswithaflashing“Vol”prompt.Fill,dispense,ormixvolumescanbeprogrammedbypressingtheVolumekey.Also,thenumberofmixcyclescanbeprogrammedbypressingtheMixkeytwice.

Automatic blowout:Attheendofafinaldispensestep,thepipetteautomaticallyexpelsanyremainingliquidfromthepipettetipswithashortburstofair(blowout).Thepistonsthenreversedirectionandreturntozeroposition.

Page 13: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

8

Pipette Overview

Note:Ifyoupressandholdinthetriggerduringthefinaldispensestep,thepipetteperformstheautomaticblowoutandholdsitspositionuntilyoureleasethetrigger.Thisdelayallowsyousufficienttimetoremovethetipsfromtheliquidtopreventaspirationofliquidasthepistonsreturntozeroposition.

TheVolumekeycontrolsthequantityofliquidtobeaspiratedordispensed.Thevolumecanbeenteredwhenthe“Vol”promptappears.

PipettingvolumecanbeincrementedquicklybycontinuouslydepressingtheVolumekey.Thelongeryouholdthekey,thegreatertheincrementalspeed.

TheVolumekeycanalsobeusedtochangepipettingspeedswithinastoredprogramorduringtheexecutionofaprogram.(See Section C.1.5, Adjusting Pipetting Speed).

[Enter] TheEnterkeystorestheprograminmemory.Aftereachstepisprogrammed,theEnterkeyispressedoncetoacceptthestep.Afterthefinalstepisprogrammed,theEnterkeyispressedtwicetodenotetheendoftheprogram.Thepipetteverifiestheendoftheprogramwithatriplebeeptone.

B.3.3 Special Function Key

TheSpecialFunctionkeyisthesmallholebelowtheVolumekey.Itcanbeactivatedwithapipettetip.ByrepeatedlypressingtheSpecialFunctionkey,youcandisplayfourfunctions:PipettingSpeed(SPd),BeepTone(TONE),PacedDispenseSpeed(PACE),andCalibration(CAL).ToexittheSpecialFunctionkeyatanytime,presstheEnterkey.

Vol

t

s

Page 14: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

9

Pipette Overview

SPd PipettingspeedcanbeadjustedbypressingtheVolumekeyinthespecialfunctionmode.Fivespeedsettingsareavailable:

SpeedSetting Display

Slow S

MediumSlow S z

Medium S z z

MediumFast S z z z

Fast S z z z F

TONE

Thebeeptonesignalsthecompletionofoneormoresteps,orsignalsawarning:

• Single beep tone:endofpipettingstep.

• Double beep tone:operationorprogrammingkeyisnotactiveincurrentmode;insufficientdispensevolume,purgerequired;orillegalprogrammingstepattempted(see Section E, Troubleshooting).

• Triple beep tone:endofpipettingsequence;endofprogrammingsequence.

Therearethreeoptionstochoosefrominsettingthebeeptone.ON1soundsalltonesattheappropriatetimeslistedabove.ON2usesonlydoubleandtriplebeeptones.OFFturnsoffalltonesexceptforthetonetosignifyanillegalprogrammingstep.PresstheSpecialFunctionkeytwicetodisplay“TONE”anditscurrentsetting.Tochangethesetting,presstheVolumekey.IfthebeeptoneissetforON1orON2,thebeeptonesymbol“ ”appearsatthebottomofthedisplay.

Note:Forpaceddispensing,thebeeptoneshouldbesetON1,toprepareyouforeachdispensestep.

Page 15: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

10

Pipette Overview

PACE Thepacecontrolsthespeedofrepeatpipetting(paced dispensing),FAST,SLOWorOFF.PresstheSpecialFunctionkeythreetimestodisplay“PACE”anditscurrentsetting.Tochangethesetting,presstheVolumekey.

PacedDispense.Whenusingthepaceddispensefeature,thepipettedispensestheprogrammedvolumeatthespecifiedpace(fast or slow).Immediatelyaftereachdispensestep,thepipetteemitsasinglebeeptone(if beep tone is on)toprepareyouforthenextdispensestep(see Section C.5, Operating Procedures).

CAL Thecalibrationofthepipetteisfactorysetfordistilledwateratroomtemperature.Ifyouarepipettingfluidswithsignificantlydifferentspecificgravitiesortemperatures,usethissettingtorecalibratethepipette.(See Section D, Calibration).

Thecurrentspeedsettingappearsatthebottomofthedisplay.Thesesettingscanbeselectedandchangedatanytimebeforeaprogramstepisexecuted.

B.4 Display

ThepipettedisplayisadynamicLCDscreen.Itdisplaysthecurrentoperation,volumeprompt(if entering a program),pipettingvolume,beeptonesymbol(if the beep tone is on),pipettingspeed,andprogramnumber.

Figure B.4: Pipette Display

Page 16: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

11

Pipette Overview

B.5 Pipette Specifications

ItemNo.

Volume ChannelVolume Range

Color Ring

Accuracy* (±)

Precision** (+)

1029 12.50µl Single Ch 0.50-12.50µl Red 1.0%/0.1ul 0.4%/0.06ul

1020 30.0µl Single Ch 1.0-30.0iµl Purple 1.0%/0.15µl 0.3%/0.05µl

1021 125.0µl Single Ch 2.0-125.0µl Yellow 0.6%/0.3µl 0.2%/0.1µl

1022 250µl Single Ch 5-250µl Blue 0.6%/0.5µl 0.15%/0.15µl

1024 1250µl Single Ch 15-1250µl Green 0.5%/3.0µl 0.13%/0.6µl

1122 250µl Shorty 5-250µl Blue 0.6%/0.5µl 0.15%/0.15µl

1124 1250µl Shorty 15-1250µl Green 0.5%/3.0µl 0.13%/0.6µl

2621 125.0µl 6-ChEXP 2.0-125.0µl Yellow 2.0%/1.0µl 1.0%/0.6µl

2622 250µl 6-ChEXP 5.0-250.0µl Blue 2.0%/1.5µl 0.7%/1.0µl

2624 1250µl 6-ChEXP 15-1250µl Green 1.5%/6.0µl 0.6%/3.0µl

2009 12.50µl 8-Ch 0.50-12.50µl Red 2.5%/0.15µl 2.0%/0.15µl

2029 12.50µl 8-ChEXP 0.50-12.50µl Red 2.5%/0.15µl 2.0%/0.15µl

2001 125.0µl 8-Ch 2.0-125.0µl Yellow 2.0%/1.0µl 1.0%/0.6µl

2021 125.0µl 8-ChEXP 2.0-125.0µl Yellow 2.0%/1.0µl 1.0%/0.6µl

2002 250.0µl 8-Ch 5.0-250.0µl Blue 2.0%/1.5µl 0.7%/1.0µl

2022 250.0µl 8-ChEXP 5-250µl Blue 2.0%/1.5µl 0.7%/1.0µl

2004 1250µl 8-Ch 15-1250µl Green 1.5%/6.0µl 0.6%/3.0µl

2024 1250µl 8-ChEXP 15-1250µl Green 1.5%/6.0µl 0.6%/3.0µl

2019 12.50µl 12-Ch 0.50-12.50µl Red 2.5%/0.15µl 2.0%/0.15µl

2229 12.50µl 12ChEXP 0.50-12.50µl Red 2.5%/0.15µl 2.0%/0.15µl

2011 125.0µl 12-Ch 2.0-125.0µl Yellow 2.0%/1.0µl 1.0%/0.6µl

2012 250.0µl 12-Ch 5.0-250.0µl Blue 2.0%/1.5µl 0.7%/1.0µl

2014 850µl 12-Ch 15-850µl Orange 1.5%/4.0µl 0.6%/2.5µl

2069 12.50µl 16-Ch 0.50-12.50µl Red 2.5%/0.15µl 2.0%/0.15µl

2060 30.0µl 16-Ch 1.0-30.0µl Purple 2.0%/0.30µl 1.7%/0.30µl

2061 125.0µl 16-Ch 2.0-125.0µl Yellow 2.0%/1.0µl 1.0%/0.6µl

Table B.5 Pipette Specifications

Note:ForAccuracyandPrecision,valuesshownareexpressedasapercent(%)deviationormicroliter(µl)value.Whenappliedtodesiredvolume,thegreaterofthetwovalueswillalwaysapply.

*Bothvaluesrepresentthedeviationfromthemean.** Percentagevaluesareexpressedasthecoefficientofvariation.Microlitervaluesare expressedasthestandarddeviation.

Page 17: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

12

Pipette Overview

B.6 Hazards and Precautions

TherearenoknownhazardsassociatedwiththeThermoScientificMatrixElectronicPipettewhenitisoperatedinaccordancewiththeinstructionsinthismanual.However,youshouldbeawareofsituationsthatcanresultinseriousinjury.

Note: Do not perform troubleshooting procedures on the internal components unless instructed byThermo Fisher Scientific Technical Service personnel.

WARNING!EnsurethatthepowertransformerisconnectedtoapowerreceptaclethatprovidesvoltageandcurrentspecifiedbyThermoFisherScientific.Useofanincompatiblepowerreceptaclecancauseshockandfirehazard.

CAUTION! UseonlythepowertransformersuppliedbyThermoFisherScientific.Useofanincompatiblepowertransformercandamagethepipette.

CAUTION!Alwaysturnoffthepowerandunplugthepowertransformerbeforecleaningtheexterior.Fluidseepagecandamageinternalcomponents.

CAUTION!Donotimmersetheunitinliquid.Fluidseepagecandamageinternalcomponents.

CAUTION!Donotautoclavetheentireunit.Extremeheatcandamagethedisplayandotherelectroniccomponents.

Note: For autoclaving specific portions of the pipette, refer to the Autoclaving Procedures, or call Thermo Fisher Technical Service for instructions.

CAUTION!Donotcleanthekeypadwithbleachsolutionorothersolvents.Causticcleaningsolutionscandamagethekeypad.

CAUTION!Avoidexcessivechargingofthebatterieswhenthepipetteisnotinuse.Excessivechargingshortensbatterylifeandmaydamagethebatteries.

Page 18: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

13

Programming and Operation

C. PROGRAMMING AND OPERATION

C.1 Programming the Electronic Pipette

ProgrammingtheMatrixelectronicpipetteisfastandeasy.Pipettingstepsandvolumeareenteredinthesequencethattheywillbeperformed,thentheprogramissavedtomemory.Thisprogramwillstayinmemory(evenifthepipetteisturnedoff,ortheelectronicpipettebatteryrunslow)untilyouchangeit.Fiveprogramsandonescratchpad(Program0),eachcapableofholding40steps,canbeenteredandstoredintheelectronicpipette.

C.1.1 Open Programming Mode or “Scratch Pad” (Program 0)

TheMatrixelectronicpipetteprovidesampleprogramstorageforfive,40-stepprograms.Inaddition,itoffersopenprogrammingmode,alsoknownasthe“scratchpad.”Thismodeisidentifiedby“P-0,”whichappearsintheprogramsectionofthedisplay(refer to the pipette display on page 10).Thisfeatureenablesquickchangestopipettingprotocolswithoutaffectingpreviouslystoredprograms.

Tousethe“scratchpad”mode:Fromthestartofanyprogram,simplybeginenteringanewprotocol(selecting a pipetting operation as described below).Thisnewprotocolcannowbeused,andwillremainactiveuntilastoredprogramisretrieved.“P-0”appearsintheprogramIDsectionofthedisplaytoindicatethatthe(O)penprogrammingmode,or“scratchpad”mode,isactive.

C.1.2 Entering a Program in Open Programming Mode

Toenterapipettingprogram:

1. Turn on pipette. Pressthegreen[On/Off]keytoturnonthepipette.Thepipettewillbeeponceandflash“ZERO”onthedisplaywhileresettingthepistonstozeroposition.Thefirststepofthecurrentprogramisdisplayed.

Note: If the pipette is being programmed for the first time, the first step is the fill operation, with the default fill volume displayed (see the following table).

Page 19: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

14

Programming and Operation

VolumeFill

VolumeDispenseVolume

Mix Volume

12.50 10.00 10.00 5.00

30.0 25.0 25.0 10.00

125.0 100.0 100.0 50.0

250.0 250.0 50.0 50.0

850 850 100 100

1250 1250 100 100

Table C.1.2: Default Programming Volumes (in µl)

2. Select pipetting operation. Entertheprogrammingmodebypressingthe[Fill]or[Mix]key. For example, to begin programming with a filling operation, press the [Fill] key. The display shows “Fill” with a flashing “Vol” prompt:

Note: The displayed volume defaults to the last volume programmed for the selected operation, or to the default volume for the pipette, if it is being programmed

for the first time.

Fill.The[Fill]keyprogramsthepipettortoaspirateavolumeofliquid.Asinglefillstepcanbeprogrammedforfluidtransfers,oracombinationoffillstepscanbeprogrammedfordilutionsorsupernatantcollection.

Dispense.The[Disp]keyprogramsthepipettetodispenseavolumeofliquid.Thedispensevolumecaneitherbetheentirefillvolumeorasmallerincrementforrepeatpipetting.Thedispensevolumeneedstobeprogrammedonlyonceforrepeatpipetting.Thepipettewilldispensethevolumeeachtimethetriggerispressed(orrepeatedlyduringapaceddispense)untilthefillvolumeisreached,oruntilinsufficientvolumeremains.

Figure C.1.2a: Flashing Volume Display

Page 20: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

15

Programming and Operation

Mix.The[Mix]keyprogramsthepipettetoperformacombinationoffillsanddispenses.Youcanprogramamixvolumeormixcycle(singlefillanddispense).Youcanpress[Mix]oncetoprogramamixvolume,ortwicetoprogramamixcycle.Mixcyclesfrom2to9areavailableonthepipette. Thedefaultsettingisthreemixcycles.

Someprogrammingconsiderationsforthemixoperation:

•Iftheprogrambeginsamixcyclefromzeroposition,itwillcompletethecyclewithanautomaticblowout,thendisplay“ZERO.”Pressthetriggertoreturnthepistonstozeroposition.

•Iftheprogrambeginsamixcycleabovezeroposition(volumeisgreaterthanzero),thefluidisdispensed,andthemixcyclebegins.Themixcyclewillendatthesamevolumeitstartedwith.Also,themixoperationwillremaininthedisplaysoitcanberepeated(usefulforserialdilutions),unlessfollowedbyanotherprogramstep.Toendthemixoperation,pressthe[Purge]key,thenpressthetrigger.

Purge.The[Purge]keyprogramsthepipettetopurgethefluid.Thisstepcanbeenteredasaprogramstep,oritcanbepressedanytimewhilepipettingtoterminatetheprogramanddispensetheremainingliquid.

Note: If the program ends at a volume greater than zero, and the last step is not a mix cycle, the “Purge” prompt automatically appears. Also, the pipette emits a double beep tone to warn you that some fluid remains. Press the trigger to dispense the remaining fluid and return to the beginning of the program.

Examples of pipetting programs with different mix steps are shown in Section C.2, Programming Examples.

3. Enter volume.Enterthedesiredvolumefor[Fill],[Disp],or[Mix]bypressingtheupsordowntsectionofthe[Vol]key.Thelongeryouholdthekey,thegreatertheincrementalspeed.

Note: The [Purge] key does not use a volume; therefore, if a purge step has been entered, continue to the next step.

Page 21: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

16

Programming and Operation

4. Confirm programming step.Press[Enter]toconfirmtheprogrammingstep.Thedisplayflashesalloftheoperationkeystopromptthenextprogrammingstep:

5. Select next pipetting operation.Repeatsteps2to4untilallprogrammingstepshavebeenentered.

6. Save program and exit. Whenthedisplayflashesalloperationkeys,press[Enter]tosavetheprogramandexittheprogrammingmode.Thepipetteemitsatriplebeeptonetosignaltheendoftheprogrammingsequence.Thefirststepintheprogramisdisplayed.“Run”isdisplayedinthelowerleftcornertoindicatethatthenewprogramisreadytoberun.GotoSectionC.1.6forReviewingtheProgram,orSectionC.3forOperatingProcedures.

C.1.3 Storing a Program

Tocreateandstoreprogramsinthepipette:

1.Selecttheprogramnumberbypressingthe[Prog]key,thenpressthe[Vol]keyuntilthedesiredprogramnumber(P1-P5)isdisplayed.

2.Pressandholdthe[Prog]key,whilesimultaneouslypressingtheSpecialFunctionkeytoentertheprogrammingmode.Theprompt“Prog”willappearonthedisplay.

3.Beginenteringyournewprogram(refer to Section C.1.2, Entering a Program in Open Programming Mode, for instructions).

NOTE: When programming mode is active, the Speed setting and Beep Tone symbols will blink.

Uponpressingthe[Enter]keytosaveandexittheprogrammode,thenewprogramwillbestoredundertheselectedprogramnumber.

Figure C.1.2b: Flashing Display

Page 22: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

17

Programming and Operation

Theprogramnumberwillbedisplayedfortwoseconds.(The first step of the newly written program is displayed).Thisprogramisreadytoberun.

C.1.4 Program Recall

Torecallastoredprogram:

1.Pressthe[Prog]key,thenthe[Vol]keytoselectthedesiredprogramnumber.

2.Pressthe[Enter]keytoactivatethedisplayedprogram.

C.1.5 Adjusting Pipetting Speed

TheMatrixelectronicpipetteoffersuniquespeedcontrolprogrammingforFill,Dispense,Mix,orPurge.Speedswithinanyofthefivestoredprogramscanbeindependentlysetandstoredasanintegralpartofthepipettingprotocol.Pipettingspeediseasilychangedatthestartofeachpipettingfunction.

Tochangepipettingspeedduringtheexecutionofaprogram:

Pressthe[Vol]keytoselectthespeedforeachpipettingstep(Fill, Dispense, or Mix only).Ifpurgeisprogrammedasastep,itsspeedcanalsobeprogrammed.

Note:Purgemustbechangedintheprogram.

Thisnewlysetspeedwillremainactivewitheachfunctionevenifthepipetteisturnedoff.Speedsettingswithinanyofthefivestoredprogramswillalsoremaininmemorywiththeprogramfunctionsuntilchangedbytheoperator.

C.1.6 Reviewing the Program

Afterprogrammingthepipette,reviewtheprogrammingsteps:

1. Set pipette for RUN mode.EnsurethepipetteisintheRUNmode(“RUN”appearsinthelowerleftcornerofthedisplay).Ifthedisplayisflashingallpipettingoperations,itisstillintheprogrammingmode;press[Enter]tosavetheprogramandexit.

2. Begin program.Runtheprogramwithoutusingpipettetipsoraspiratingfluid.Pressthetriggerforeachstepintheprogram.

Page 23: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

18

Programming and Operation

3. Observe display. Ateachstepoftheprogram,observethedisplay.Ifnecessary,re-entertheprogrammingsequence.

Somehelpfulhintsforreviewingyourprogram:

•Iftheprogramiscomplexandcontainsseveralsteps,writethestepsonpaperbeforeenteringtheprogram.Checkoffeachstepasyoureviewtheprogram.

•Iftheprogramisintendedforpaceddispensing,reviewtheprogramagain,withthetriggerdepresseduntilthelaststep.Checkthedispensepacetoensurethedispensestepsareacomfortablespeed.

C.2 Programming Examples

Examplesofdifferentpipettingsequencesaredescribedbelow,andareidealforuseastrainingexercises.Pressthekeysshowninbrackets[],thenpressthe[Vol]keyuntilthevolumeinparentheses()isdisplayed.Attheendoftheprogram,atriplebeeptone issounded.(see section D.3 Recommendations for Accurate Pipetting for suggestions for better precision and accuracy).

C.2.1 Sample Transfers

Example:Fillanddispense100µl.

Option1 Option2

[Fill]

Vol

t

s

(100)[Enter]or

[Fill]

Vol

t

s

(100)[Enter][Enter]

[Disp]

Vol

t

s

(100)[Enter][Enter]

Thesecondoptionisaone-stepprogram.Becausetheprogramendsatavolumegreaterthanzero,andthelaststepisnotamix,a“Purge”promptautomaticallyappearsattheendofthefilltodispensetheliquidandreturntothebeginningoftheprogram.

Page 24: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

19

Programming and Operation

C.2.2 Incremental Pipetting

Example:Fillanddispense250µl.

[Fill]

Vol

t

s

(250)[Enter]

[Disp]

Vol

t

s

(10)[Enter]

[Disp]

Vol

t

s

(45)[Enter][Enter]

Inthisprogram,onlyonedispensestepneedstobeprogrammed.Ifanyprogramendsinadispensestep,thepipetterepeatsthedispensestepuntilthevolumeiszero,orislessthantheprogrammeddispensevolume.Iftheremainingvolumeislessthantheprogrammeddispensevolume,thepipettedisplaysa“Purge”prompttodiscardtheremainder.Thisdiscardedvolumecontainsthecumulativeerrorofpreviouspipettings(seesectionD.3RecommendationsforAccuratePipettingforfurtherexplanation).

C.2.3 Serial Dilutions

Example: Transfer100µltothefirstcolumnofaplatepreviouslyfilledwithreagent.Mix200µlfor2cycles.Transfer100µltothesecondcolumnandrepeatthemixstep.Followthisprocedurefortherestoftheplate.

[Fill]

Vol

t

s

(100)[Enter]

[Mix]

Vol

t

s

(200)[Mix]

Vol

t

s

(2)[Enter][Enter][Enter]

That’stheentireprogram!Themixstepalwaysreturnstothevolumeatwhichitstarted(100 µl);therefore,thepipettecontinuesthemixstepindefinitelyuntilyoupressthe[Purge]keytoendtheprogram.Inthisexample,themixcycleisprogrammedfor2cycles(default is 3 cycles).

Note: The 10 µl Primer Dispense reduces the percentage of error in incremental dispensing (see section D.3 Recommendations for Accurate Pipetting)

Page 25: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

20

Programming and Operation

C.2.4 Simple Dilutions

Example:Prepare1:10dilutionwitha10µlsamplevolume.

[Fill]Vol

t

s

(100)[Enter]

[Fill]Vol

t

s

(20)[Enter]

[Fill]Vol

t

s

(10)[Enter]

[Purge][Enter][Enter]

Thefirstfillvolumeisforthediluent(100 µl),followedbyanairgap(20 µl),andfinallybythesamplevolume(10 µl).Thepurgestepwilldispensethetotalvolume.

C.2.5 Supernatant Collection

Example: Aspirate100µlfrom8rowsofaMicrotiter®plateanddiscardthe

totalvolume.

[Fill]Vol

t

s

(100)[Enter]

[Fill][Enter]

[Fill][Enter]

[Fill][Enter]

[Fill][Enter]

[Fill][Enter]

[Fill][Enter]

[Fill][Enter]

[Purge][Enter][Enter]

Thisprogramisperformedwiththe850µlhandle.All100µlfillvolumesarecollectedinthesametips.Whencompleted,thetotalvolumeisdiscardedwiththepurgestep.

Page 26: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

21

Programming and Operation

C.3 Operating Procedures

Afterprogrammingthepipette,thedisplaywillshowthefirstprogramstep,withthe“Run”promptdisplayedinthelowerleftcorner:

Thefollowingstepsdescribeabasicprogramforpipettingreagentintoamicrotiterplate.Thepipetteisprogrammedasfollows:

[Fill]Vol

t

s

(250)[Enter]

Note: The 10 µl Primer Dispense reduces the percentage of error in incremental dispensing (see section D.3 Recommendations for Accurate Pipetting)

[Disp]Vol

t

s

(10)[Enter]

[Disp]Vol

t

s

(30)[Enter][Enter]

1. Attach pipette tips.Whenreadytobeginpipetting,attachthetipstothepipette.

Note: When applying tips, avoid ‘banging’ or applying excessive force to the tip(s), as this stretches the tip seal and will affect volume. Use Matrix tips which have been specifically designed to match the tip fittings on Matrix pipette.

2. Aspirate liquid.Immersethepipettetipsintotheliquidtobeaspirated,thenpressthetriggertoaspirate250µl.Submergethetipsapproximately1/4”belowthesurfaceoftheliquid.

3. Dispense primer volume.Discardthe10µlprimervolumebacktothesourcevesselorintoawastereceptacle.

Figure C.3: Run Prompt

Page 27: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

22

Programming and Operation

4. Dispense liquid.Positionthetipsoverthefirstcolumnofwellsinthemicrotiterplate,andpressthetriggertodispense30µl.Atthecompletionofthedispensestep,thepipetteemitsasinglebeeptone(if the beep tone is set to ON1).

5. Repeat dispense.Repeatthepreviousstepfortherestofthecolumnsinthemicrotiterplate.Atthecompletionoftheeighthdispense,thepipetteemitsadoublebeeptone(if the beep tone is on, otherwise it emits a single beep tone)anddisplaysthe“Purge”prompt.Thereisinsufficientvolumeforanother30µldispense,andthepipettesignalsforapurgestep.

-or-

Paced dispense.Steps4and5canalsobeperformedasapaceddispense.Insteadofpressingthetriggerforeachdispensestep,press andholdthetriggertoautomaticallydispenseeachvolume.Immediatelyaftereachdispensestep,thepipetteemitsasinglebeeptone(if the beep tone is on),promptingyoutoprepareforthenextdispensestep.Tostopapaceddispense,releasethetriggerbeforetheendofthelastdispensestep.

Note: For paced dispensing of small volumes, set the pace to SLOW. Ensure that you have fully released the trigger before the end of the last dispense step to prevent an additional, unwanted dispense step.

6. Discard remaining liquid. Positionthepipettetipsoverawastecontainerandpressthetriggertopurgetheremainingliquid.Thepipetteemitsatriplebeeptonetosignaltheendoftheprogram,thenreturnstothefirststepoftheprogram.

7. Eject tips.Presstheejectortriggertoejecttheusedtips.

Page 28: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

23

Calibration

D. CALIBRATION

Eachelectronicpipetteisfactory-calibratedfordistilledwaterat20°C.Foreasycalibrationusingdistilledwater,thepipettefeaturesan“EASY”CALmode(seeSectionD.2.2,“EASY”CALmode).Whenpipettingliquidsofsignificantlydifferentspecificgravitiesortemperatures,however,theCalibration(CALC)featuremaybeusedtoachievegreateraccuracy(seeSectionD.2.3,“CALC”mode).

Toprepareforcalibration,ensurethepipetteandtipsareequilibrated.Whencalibratingamultichannelpipette,selectonlyonechanneltopipettewith.Usingananalyticalbalance,firstobtaintheactualvalueoftheprogrammedvolumeatfactorycalibration.Theprogrammedvolume(Vp)usedistypicallythecalibrationpointforthatpipette(RefertoTableD.2.2)

D.1 Volume Measurement

Tomeasuretheactualvalueoftheprogrammedvolume:

1. Filltheprogrammedvolume(Vp),thendispensetheentirevolumeintoacontainerontheanalyticalbalance.

2.Readtheweightontheanalyticalbalance.Repeatthisproceduretentimes.

Note: Use a clean and dry pipette tip with each pipetting cycle. When filling the liquid, immerse the pipette tip approximately 1/4 inch (approx. 0.6 cm) below the liquid surface. When dispensing the liquid, touch the side of the container to ensure a complete dispense.

3.Determinetheaverageweightoftheprogrammedvolumeandconvertittovolume.Toconverttovolume,correcttheweightforspecificgravityandtemperature.

Ifweighingwater,useoneofthecorrectionfactorsbelow:

Temperature CorrectionFactor

20°C-22.5°C 1.003

23°C-25°C 1.004

Theresultingvolumeisthecorrectedactualvolume.Ifthepipetteiscorrectlycalibrated,theprogrammedvolumeshouldequaltheactualvolume(within the specifications of the pipette).

Page 29: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

24

Calibration

D.2 Calibration

Electronicpipetteuserscancalibratetheirpipetteswith“EASY”CALmodefordistilledwaterorwith“CALC”modetocalibrateotherliquidsandfortemperaturesotherthanroomtemperature.

D.2.1 Factory Calibration Step

Thefactorycalibrationstepisneededwhencalibratingyourelectronicpipette.Thisislocatedonthecalibrationcertificatecalled“PipetteCalibrationReport”enclosedinthedocumentionforthispipette.(See comments on page 1 of Pipette Calibration Report).Viewthepipette’scurrentcalibrationstepbyusingtheCALVIEWmodelistedinsectionD.2.4.

Note: When resetting the calibration step to a default value, as listed in section D.2.5 CALRESET, the step is set to an average value for all pipettes of its class. This value may differ from the factory-optimized calibration step documented on the “Pipette Calibration Report” . Please use the step listed there for best results. If you cannot locate your documentation that lists the factory calibration step, contact technical service at 800-345-0206 and have your pipette’s serial

number ready.

D.2.2 “EASY” Cal Mode

Electronicpipetteuserscanquicklycalibratetheirpipettebyusingtheunique“EASY”CALfeature.“EASY”CALhasbeendevelopedforfast,easy,in-labverificationand/orcalibrationusingdistilled water at room temperature (20-22.5°C).

Volume CalibrationPoint(µl) FactoryCalibrationSteps

12.50µl 10.00µl Locatedonthecalibrationcertificate“PipetteCalibrationReport”sheetincludedinthepipettedocumentation.

30µl 25.0µl

125.0µl 100µl

250.0µl 200µl

850µl 700µl

1250µl 1000µl

Table D.2.2: Calibration Specifications

Page 30: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

25

Calibration

1. Weighouttenaliquotsofdistilledwateronananalyticalbalanceusingonepipettingchannel.Theprogrammeddispensevolumemustcorrespondtothecalibrationpointforthatpipette(for example, 200 µl for a 250 µl pipette).RefertoTableD.2.2,CalibrationSpecifications,forpropercalibrationpoints.

Note: Ensure that the distilled water used is at room temperature (20-22.5 °C). An internal correction factor of 1.003 is used by the pipette to convert weight (gm) to volume (ml).

2. Calculateanaverageforthetenreadingsofdistilledwaterobtainedfromthebalance.Iftheresultingaverageweightiswithinyouraccuracyspecifications,nofurthercalibrationisrequired.Iftheaverageweightdoesnotmeetexpectedperformance,enterthe“EASY”CALmodetoquicklyre-calibratethepipette.

3. Enterthe“EASY”CALmodebypressingtheSpecialFunctionkey(using a pipette tip)until“EASY”flashes,thenpressthe[Enter]keytodisplaythecalibrationpoint.

4. Usethe[Vol]keytoadjustthedisplayedcalibrationpointuntilitmatchestheaverageweightobtainedinstep2.Storethisnumberbypressingthe[Enter]key.Thenewcalibrationsettingisconfirmedbythe“CAL”and“SET”messagesonthedisplay.

5. VerifythisnewcalibrationsettingbyrepeatingSteps1and2.

D.2.3 “CALC” Mode

Tocalibratethepipette:

1.Weighouttenaliquotsofdistilledwateronananalyticalbalance,usingonepipettingchannel.Theprogrammeddispensevolumemustcorrespondtothecalibrationpointforthatpipette(for example, 200 µl for a 250 µl pipette).RefertoTableD.2.2,CalibrationSpecifications,forpropercalibrationpoints.

Note: Ensure that the distilled water used is at room temperature (20-22.5 °C).

Page 31: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

26

Calibration

2.Calculatedisplayvolumeofthepipette:

Vp=ProgrammedVolume

Va=CorrectedActualVolume(correctedfortemperature)

Cp=CalibrationPoint(each pipette handle has a unique calibration point, see table D.2.2: Calibration Specifications.)

Dv=DisplayVolume

3.EntertheCALmodeonthepipettebypressingtheSpecialFunctionkey(use a pipette tip to press the key)until“EASY”appearsonthedisplay.

4.Pressthe[Vol]keytodisplay“CALC”(pressing the [Vol] key again will return you to the “EASY” mode).

5.Pressthe[Enter]keytodisplaythecalibrationpoint.

6.Pressthe[Vol]keytoentertheDisplayVolume(Dv)calculatedabovein step2.

7.Pressthe[Enter]keytosetthecalibrationpoint.Thedisplayshows“CAL”then“SET”toconfirmthenewcalibrationpoint,andreturnstothecurrentpipettingprogram.

Example:Calibratea1250µl8-channelpipette.

Watertemperature=24°C ProgrammedVolume(Vp)=1000µl AverageofActualVolume=990µl CorrectedActualVolume(Va)=990x1.004=994µl CalibrationPoint(Cp)=1000

Dv=xVp

VaCp

DV =xCp

Vp

Va

DV = x10001000 µl

994 µl

DV = 1006

Page 32: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

27

Calibration

D.2.4 CALVIEW

CALVIEWdisplaysthepipettingstroke(numberofmotorsteps)usedtoreachthecalibrationpoint.Thisvalueallowsyoutoconfirmthecalibrationduringpipettingandaftercalibration.ToactivateCALVIEW,entertheCALmodeusingtheSpecialFunctionkey.Whenthecalibrationpointisdisplayed,pressthe[Fill]key.Thenumberofstepsisdisplayed.PressanykeytoexitCALVIEW.

D.2.5 CALRESET

CALRESETallowsyoutoreturntoadefaultcalibrationstep.Toresetthecalibration,entertheCALmodeusingtheSpecialFunctionkey.Whenthecalibrationpointisdisplayed,pressthe[Purge]key.Thedisplayflashes“FACT,”“CAL,”then“SET”andthepipetteresetstothedefaultstep.Thisdefaultsettingisanaveragevalueforallpipettesofitsclass.This value may differ from the factory-optimized calibration step documented on the “Pipette Calibration Report” certificate enclosed in this pipette’s documentation.

D.3 Recommendations for Accurate Pipetting

MatrixElectronicpipettesarefactorycalibratedtodeliveraccuratevolumeswithdistilledwaterat21°Cand50%humidity.Byusingthecalibrationproceduresdescribedabove,youcanpipetteliquidsofdifferentspecificgravitywithahighdegreeofaccuracyandreproducibility.Foroptimalpipettingresults,refertothefollowingrecommendations:

• Pipette at a constant temperature. Allowliquidsandequipmenttoequilibratetoambienttemperature.Volumeaspirationwithairdisplacementpipettesvarieswithairpressure,relativehumidityandtheliquidsvaporpressure,allofwhicharetemperaturedependent.

• Attach tips with slow, even pressure.Whenapplyingtips,avoid‘banging’orapplyingexcessiveforcetothetip(s),asthisstretchesthetipsealandwillaffectvolume.UseMatrixtips,whichhavebeenspecificallydesignedtomatchthetipfittingsonMatrixpipettes.

Page 33: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

28

Calibration

• Calibrate the pipette for the liquid.Iftheliquidhasaspecificgravityandvaporpressureequivalenttodistilledwater,calibrateusing“EASY”CALmode.Otherwise,use“CALC”mode.

• Pre-wet the pipette tip.Aspirateanddispenseafewtimesbeforeaspiratingtheliquidtoincreasehumiditywithinthetip,reducingsampleevaporationthatresultsinlowerdispensevolumesininitialfewpipettings.Theneedtopre-wetincreaseswhenworkingwithliquidswithhighvaporpressureand/orhighsurfacetension.

• Immerse tips fully.Tipsshouldbeimmersedapproximately1/4inch(approx. 0.6 cm)belowtheliquidsmeniscusandawayfromthecontainersurfacestoavoidaspirationofairandexcessdropletsclingingtotheoutsideofthetip.

• Discard a primer volume.Whenthemotorchangesfromaspirationtodispense,thepercentageoferrorisgreaterinthefirstdispensethaninsubsequentdispenses.

• Pause consistently.Ittakesamomentfortheliquidinthetiptofinishmoving.Pausewiththetipintheliquidforaboutonesecondafteraspiratingthesampletoallowtheliquidtofinishmoviingintothepipettetipafterthepipettestops.Otherwise,thevolumewillbelowerthanitshould.

• Aspirate slowly, dispense quickly.Asageneralrule,pipetteliquids,especiallyviscousliquids,ataslowerspeedtoensureaccuratepipetting.Dispenseaqueoussolutionsatfasterspeedstoexpelliquidmore accurately.

Note: Viscous liquids should be dispensed slowly.

• Tip touch.Duringthedispensingofaliquid,touchthetipagainstthesideofthereceivingcontainertopreventanyliquidfromremainingonthetip.

• Discard final dispense.Whenaccuratedispensingofseveralincrementsofliquidareneeded,discardthelastincrement.Thecumulativepercentageoferrorfrompreviousincrementscarriestothelastvolume.(i.e. The pipette aspirates 100 µl and dispenses 20 µl five times, and if each of the first four dispenses were off by +1.0 µl, then the last dispense is missing 4 µl before you even dispense it).

Page 34: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

29

Troubleshooting

E. TROUBLESHOOTING

Possibleproceduralandsystemproblems,aswellasrecommendedsolutions.Ifyoucannotresolvetheproblem,calltechnicalserviceat800-345-0206.

No power to the pipette.

ProbableCause RecommendedSolution

Batteries are incorrectly installed.

Rechecktheorientationofthebatteries;theyshouldmatchthediagraminthebatterycase

Batterypackcablesarenotproperlyconnected.

Ensuretheconnectionstothebatterypackaresecurelypluggedintotheconnectionfromthepipettebody.

BatterySpacerisnotinunit or is not in correctly (Multichannels only with NiMH).

EnsurethatthebatteryspacerisinplaceforallmodelsthatuseNickelMetalHydride(see section F.3.1).

Double beep tone occurs.

ProbableCause RecommendedSolution

Anillegalprogrammingstepisbeingentered,suchasprogrammingtwofillsthatexceedthemaximumfillvolume.

Rechecktheprogrammingsequence.Ensureyouarenotattemptingtofillabovethemaximumfillvolumeofthepipette,orattemptingtodispensemorethanthetotalfillvolume.

“ERR_1” is displayed. (Signifies one of the following conditions)

ProbableCause RecommendedSolution

Liquidisaspiratedintothepipette.

Disassembleandcleanthepipette.(See Section F.1 and/or F.2, for pipette cleaning and autoclaving instructions).

Motorisdefectiveortheflexcircuitassemblyisdamaged.

Contacttechnicalserviceat:800-345-0206

Page 35: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

30

Troubleshooting

The motor sounds rough and aspiration is very slow.

ProbableCause RecommendedSolution

Thepipettehasbeensettotheslowestspeed.

Verifyspeedsetting,theslowestspeedsetting normally causes the motor to run slowlyandlouderthanusual.

Motorhasnotbeenusedforanextendedperiod.

Cycletheunit(fill and dispense)fivetoseventimestore-lubethepiston.

After reassembly, the pipettor leaks or does not fill liquid.

ProbableCause RecommendedSolution

O-ringwasnotproperlyseated.DisassemblethepipetteandreseattheO-ring.(See Section F.1 and/or F.2 for pipette disassembly/reassembly).

O-ringwasdamagedduringreassembly.

ReplacetheO-ring.(See Section E.2.2, Pipette Cleaning and Autoclaving (Relubrication step) and Section E.2.3, Pipette Reassembly).

Cylinderwasnotscrewedtightlyintothepipettebody.(single channel only)

Securelyscrewthecylinderintothepipettebody,finger-tightened.

Tubingwasdisconnectedfromtipfittingorcylinder.

Disassemblethepipetteandreattachthetube.Iftubefitspoorly,itmayhavestretchedandneedstobereplacedortrimmedofstretchedsection.

Pipette does not eject pipette tip properly.

ProbableCause RecommendedSolution

Duringreassemblyofthecolletandejectorsleeve,themetalejectortriggertabwasnotcorrectlypositionedbehindtheplastictaboftheejectorsleeve.(Single channel only)

Reassemblethecolletandejectorsleevewiththecorrectorientationoftheejectortriggertabtotheejectorsleevetab.(See Section F.1.3, Pipette Reassembly).

Page 36: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

31

Maintenance

F. MAINTENANCE

TheThermoFisherMatrixfamilyofelectronicpipetteshasbeenspecificallydesignedforserviceability,enablingtheusertoserviceinstrumentswithrelativeease.Undercertaincircumstanceswhereon-siteserviceisnotfeasible,theinstrumentmayneedtobereturnedtoThermoFisherScientificforrepair(seePipetteService,SectionsG.2andG3).

Electronicpipettesrequireverylittlemaintenance.Theexteriorofthepipettecanbecleanedperiodicallywithasoftclothmoistenedwithmethylalcohol.Periodiccleaningofthetipfittingmaybenecessaryusingtheproceduresformaintenancelistedbelow.

Incaseswherepipetteshavebeenexposedtobiohazardousmaterials,autoclavinginternalcomponentsisrequired.ThermoFisherScientificprovidesthefollowingPreventiveMaintenance/AutoclaveKit:

Preventive Maintenance/Autoclave Kit

•Replacementcrosstubes •O-ringlubricant

•Woodenapplicator •O-rings

Item # Description Item # Description

9809 12.5µl-6,8,&12channel 9804 850µl-12channel

9801 125µl-6,8,&12channel 9805 1250µl-6&8channel

9802 250µl-6,8,&12channel

CAUTION! DO NOT AUTOCLAVEentireunitorcleanthekeypadwithbleachsolutionorothersolvents.Extremeheatcandamagethedisplayandotherelectroniccomponents.Causticcleaningsolutionscandamagethekeypad.

Note: For autoclaving specific portions of the pipette refer to the autoclaving steps in Section F.1.2 for single channel pipettes, or section F.2.2 for multichannel pipettes; or call technical service for instructions.

Note: Avoid prolonged exposure to ultraviolet light. The pipette housing may become discolored or damaged.

Page 37: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

32

Maintenance

Figure F.1.1b: Pipette Tip Disassembly

Figure F.1.1a: Detaching the Collet

F.1 Single Channel Electronic Pipette Maintenance

F.1.1 Single Channel Electronic Pipette Disassembly

Withproperuseofthepipette,thedisassemblyandcleaningofthepipettetipisnotnecessary.However,iffluidisinadvertentlydrawnintothetipfitting,thenthelowerendofthepipetteshouldbedisassembledfor thorough cleaning.

Todisassemblethelowerendofthepipette:

1.Detachthecolletbytwistingitcounter-clockwise(see Figure F.1.1a).

2.Removethecolletandejectorsleevebyslidingbothpiecesdownoverthecylinder(see Figure F.1.1b).Forreassemblylater,notethepositionofthelargespringinthecollet.

3. Unscrewcylindercounter-clockwiseandremoveitfromthepipettebody.

Note: Check for the presence of the O-ring on the piston. If it is not on the piston, look inside the cylinder and remove the O-ring.

4.RemoveO-ring,O-ringsupport,and1/2”diameterspringforcleaningbyslidingthemoffofthethepiston(refertoFigureF.1.1c).

IMPORTANT! Leave the piston on the motor screw, DO NOT try to disassemble!

Note: Check for the presence of the O-ring on the piston. If it is not on the piston, look inside the cylinder and remove the O-ring.

Page 38: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

33

Maintenance

Figure F.1.1c: Pipette Tip Assembly

Page 39: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

34

Maintenance

F.1.2 Single Channel Cleaning and Autoclaving

General Cleaning: Carefullywipedownalldisassembledplasticandmetalpartswithmethylalcoholor70%isopropylalcohol.

Decontamination/Autoclaving: Decontaminatealldisassembledpartsbywipingthemwithableachsolution(5%sodiumhypochlorite)orbyautoclavingthepartsat121°Cfor20minutes.

Re-lubrication: LightlycoatpistonwithO-ringlubricantandreplaceorre-installO-ringonthepiston,asdescribedinreassemblyprocedure.

Note: Do not clean the O-ring. Replace the O-ring after decontamination or autoclaving (O-ring and O-ring lubrication is available from Thermo Fisher Scientific).

F.1.3 Pipette Reassembly

Aftercleaningordecontaminatingthedisassembledparts,thepipettecanbereassembled(refer to Figure F.1.1c for parts identification and reassembly order).

1.Slidethe1/2”diameterspring,O-ringsupport,andO-ringoverthepistonandmotorscrew.(see Figure F.1.3a).

2.SlidethecylinderovertheO-ringandpistonassemblyandscrewintothepipettebodysothatitisfingertight.Donotover-tighten.

3.Assemblecollet,1.0”diameter ejectorspring,andejectorsleeve.

Note: Ensure the tab on the ejector sleeve is aligned with the notch in the collet.

Figure F.1.3a: O-ring and piston assembly

Page 40: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

35

Maintenance

4.Slidethecollet,withspringandejectorsleeve,overthecylinder.Positionthemetaltaboftheejectortriggerbehindthetabontheejectorsleeve (see Figure F.1.3b).

5. Attach the collet onto the pipettebodybytwistingleft slightly,thenright,toinsertthe tabsofthepipettebodyinto thelocksonthecollet (see Figure F.1.3b).

6.Testtheejectorfunctionto ensurethattheejectorsleeve movesdownwhentheejector triggerispressed.

F.2 Multichannel Electronic Pipette Maintenance

F.2.1 Multichannel Electronic Pipette Disassembly

1.Removethecolorringbygentlyprying looseoneoftheedgesnearthe handletrigger.

2.UseaPhillipsscrewdriverto removeallthreehousingscrews.

3.Carefullyremovetheplasticcover housingandobservethelayoutof theinternalparts.

Figure F.2.1a: Removing Cover Housing

Figure F.1.3b: Detaching the Collet

Page 41: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

36

Maintenance

4.Removeplasticejectortriggerwithspringandslidethealuminumejectorcomboffofthetipfittings(see Figure F.2.1b).On16channelmodels,reversethisprocessandremovecombfirst,thentrigger.

5.Removethepiston/cylinderassemblybypullingthemotorscrewdownfromthemotorshaftcouplingandoutofthemainhousing.Removethemanifoldassemblybyfirmlyliftingitoutofthemainhousingsupportribs(seeFigureF.2.1c).

Figure F.2.1c: Removing Piston/Cylinder and Manifold Assemblies

Figure F.2.1b: Removing Ejector Trigger and Ejector Comb

Page 42: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

37

Maintenance

Item Description Item Description

1. Keypad 15. Tipfitting

2. Mainboard 16. Consolecover

3. Flexcircuit 17. Consoledisplay

4. Batterybox 18. Batteries

5. Driverboard 19. Mainhousing

6. Motor 20. Triggerhingepin

7. Color ring 21. Handletrigger

8. Motorcoupling 22. Handletriggerspring

9. Motorscrew 23. Ejectortrigger

10. Pistonbackplate 24. Piston

11. Photosensor 25. O-ring

12. Cylinder 26. Crosstube

13. Coverhousing 27. Manifoldplate

14. Housingscrew

Figure F.2.1d: Pipette Components

Page 43: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

38

Maintenance

F.2.2 Multichannel Cleaning and Autoclaving

1.Removethepistonassemblyfromthecylinder.

2.Removeanddiscardthecrosstubesintoabiohazardouscontainer.

3.RemovetheO-rings(only if necessary, see section F.1 and/or F.2 for O-ring Replacement)anddiscard themintoabiohazardous container.

4.Placeallcomponents(see Figure F.2.2)intotheautoclaveandautoclaveat121°Cfor20minutes.

F.2.3 Multichannel Maintenance and Reassembly

Oncecomponentsareautoclavedandcompletelydry,usethefollowingprocedurestorelubricateandreassembletheunit:

O-ring Replacement

TheO-ringsmayrequirereplacementifliquidshavebeenoveraspirated.Inthiscase,theO-ringsshouldberemovedpriortoautoclaving.SpareO-ringpacksareavailablethroughMatrixforallpipettevolumes.

1. Remove old O-rings.

• 1250 µl, 850 µl, 250 µl Pipettes:O-ringscanberemovedfromthepistonsbysqueezingonesideoftheO-ringandliftingitoutoftheO-ringgrooveonthepiston.Onceremoved,theO-ringsshouldbediscardedasbiohazardouswaste.

• 125 µl, 30 µl, 12.5 µl Pipettes: O-ringsfromthesepipettesarelocatedinthecylinder.ToremovetheO-rings,unscrewthefourscrewsthatholdthecylindercaptothecylinder(see Figure F.2.3a).Onceremoved,theO-ringsshouldbediscardedasbiohazardouswaste.

Figure F.2.2: Components to be Autoclaved

Page 44: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

39

Maintenance

2. Install new O-rings. After theoldO-ringshavebeenremovedandallcomponentshavebeenautoclaved,thenewO-ringsmaybeinstalled.BeforeinstallingtheO-rings,applyathincoatoflubricanttoeachO-ring,usingtheO-ringlubricantsuppliedinthePreventiveMaintenance/AutoclaveKit.

Re-lubrication• 1250 µl, 850 µl, 250 µl Pipettes:UsethewoodenapplicatorandO-ringlubricantsuppliedinthePreventiveMaintenance/AutoclaveKittoapplyathinlayeroflubricationapproximately0.5inches(approx. 1.3 cm)downintoeachpistonholethatrequiresit.

Spreadthelubricantevenlyalongtheinnerwallofeachcylinderbyusingacircularmotionwiththewoodenapplicator (see Figure F.2.3b).

•125µl,30µl,12.5µlPipettes:Spreadthelubricantdirectlyontothepistons,approximately0.5inches(approx. 1.3 cm)fromthebottomofeachpiston(see Figure F.2.3c)

Figure F.2.3c: Re-lubricating 125 - 12.5 µl Pipettes

Figure F.2.3a: Removing O-rings

Figure F.2.3b: Re-lubricating 1250 - 250 µl Pipettes

Page 45: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

40

Maintenance

Reassembly

1.Reinsertthepistonassemblyintothelubricatedcylinder.Ensurethatthemotorscrewengagesintotheholeinthecenterofthecylinder(see Figure F.2.3d).

NOTE: To correctly orient the piston and cylinder assembly, ensure that the lot number of the cylinder is facing up and the painted section of the piston backplate is facing down. Also, care should be taken to keep the O-rings from being pinched when inserting the pistons into the cylinder.

2. Cross Tube Replacement Refertothefollowingfigureandtableforproperconnectionofcrosstubesfromthetipfittingstothecylinderpositions.

Figure F.2.3d: Reassemble Piston/Cylinder Assembly

Table F.2.3a: Cylinders

Page 46: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

41

Maintenance

TipFittingPositions

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

CylinderPositions

16 Channel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

12 Channel 1 2 3 4 5 6 7 8 9 10 11 12

8 Channel 1 3 4 6 7 9 10 12

6 Channel 1 2 3 10 11 12

5 Channel 1 2 3 10 11

8 Channel 1 2 3 4 5 6 7 8

6 Channel 1 3 4 6 7 8

5 Channel 1 3 4 5 7

Note: If the Cross Tube orientation between the cylinder and Tip Fittings is incorrect, performance of the pipette will be impaired.

3.Carefullyinserttheflatendofthemotorscrewintotheholeofthemotorcoupling.Thiscanbeaccomplishedbyholdingthepiston/cylinderassemblyataslightanglewhileinsertingthemotorscrewintotheholeofthemotorcoupling.Slowlyspinthemotorscrewuntiltheflatendofthemotorscrewisfullyengagedintothemotorcouplinghole(see Figure F.2.3e).Thecylindercannowbeseatedbetweentheplasticribsinthemain housing.

NOTE: The lot number on the face of the cylinder should be facing toward you and the painted section of the piston backplate should be facing the photosensor. The painted section is used by the photosensor to track the location of the pistons.

Table F.2.3b: Cylinder/Cross Tube Positions

12ChCylinder

8ChCylinder

16 C

h

Page 47: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

42

Maintenance

4.Afterinstallingthepiston/cylinderassembly,installthemanifoldassemblyintothemainhousing.Tokeepthetubingfreefromejectorcontact,rotatethemanifoldplateandtubingonecompleteturncounterclockwise(see Figure F.2.3f).Insertthemanifoldassemblyintothemainhousingsupportribs.

5.Slidetheejectorcomboverthetipfittings withtheslotsonthebackofthecomb exposed(see Figure F.2.3g).

Figure F.2.3e: Installing Piston/Cylinder Assembly

Figure F.2.3f: Rotating Manifold Assembly

Figure F.2.3g: Attaching Ejector Comb

Page 48: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

43

Maintenance

6.Holdtheejectorcombinplacewhileinsertingtheejectortrigger(see Figure F.2.3h).Placetheendoftheejectorspringagainsttheplasticstopandcompressthespringwhilepositioningtheejectortriggeroverthealuminumejectorcomb(the ribs on the ejector trigger should mate with the slots on the aluminum ejector comb).For16-channelmodels,installthetriggerpriortoconnectingtheejectorcomb.

NOTE: Be careful not to pinch the cross tubes when inserting the ejector trigger.

Testtheejectortriggertoensurethattheejectorcombmovesdownwhenthetriggerispressed(hold the ejector comb firmly in place while testing the trigger).

7.Placethecoverhousingontothemainhousingandfastenthe threescrews.

8.Re-attachthecolorring.

9. Testpipetteforsmoothoperation.(Reassembleandrepeatsteps 3-7ifneccessary.)

The pipette is now ready for use.

Figure F.2.3h: Inserting Ejector Trigger

Page 49: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

44

Maintenance

F.3 Installing/Replacing Batteries

TheMatrixelectronicpipettecomeswithrechargeableNickelMetalHydride(NiMH)batteriesorwithaLithiumIonbatteryfor16channelmodels.Undernormaluse,thebatteriescanberechargedapproximately400-500times.Whentheycannolongerholdtheircharge,thebatteriesmustbereplaced.Replacementbatteriescanbepurchasedthroughcustomerservice.ThepipettemayalsobereturnedtoThermoFisherScientificforbatteryreplacement.

Replacement batteries:

6,8,12Multichannels 2NiMH ItemNo.6092

Single Channel 1NiMHPack ItemNo.2092

16 Channel only 1Lithium-IonBattery ItemNo.2094

Note: When replacing batteries, ensure the pipette is turned off before removing the battery case.

F.3.1 Installing/Replacing Nickel Metal Hydride Batteries

F.3.1A Installing/Replacing Single Channel Electronic Pipette Batteries

1. Remove screws. Useasmall,flatheadscrewdrivertoremovethetwoscrewsholdingthetopofthepipette(containing the display and keypad)tothepipettehousing.Thescrewsarelocatedjustabovethepipettingtrigger(see Figure F.3.1Aa).

2. Lift pipette top. Carefully lift the topofthepipettetoexposethebatterypack.

Figure F.3.1Aa: Removing Pipette Cover

Page 50: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

45

Maintenance

3. Disconnect wire connectors.Pullapartthetwowireconnectorsoverthebatterypack(see Figure E.3.1Ab).

4. Remove battery pack.Pullgentlyonthebatterywireconnector(withoutthepins)todislodgethebatterypack,thenliftthebatterypackoutofthemainhousing(see Figure F.3.1Ac).

5. Insert new battery pack. Insertthenewbatterypackintothecavityofthemainhousing.Ensurethatthewireisloopedunderthebatterypackandtheconnectorisaboveit(refer to Figure F.3.1Ac).

NOTE: The battery pack fits very snugly in the main housing; therefore, ensure that the pack is oriented as shown in Figure F.3.1Ac. The connectors for the battery pack and the pipette

should be on opposite sides.

6. Reconnect wire connectors. Reconnectthewireconnectorsabovethebatterypack.Gentlypushthewiresofthebatterypackdownuntilthewireconnectorslayflatoverthetopofthebatterypack.

Figure F.3.1Ac: Routing battery wires

Figure F.3.1Ab: Disconnect Wires

Figure F.3.1Ad: Reconnect Wires

Page 51: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

46

Maintenance

7. Reassemble pipette top. Attachpipettetoptothemainhousing.Ensureflexcableisbehindthetopofthepipettetriggerandthetrigger contacts the limit switchproperly(see Figure F.3.1Ae).Fittheedgesofthepipettetopsnuglywithintheedgesalongthetopofthepipettehousing.Ensurethatthecablestothepowerjackareinsidethemainhousingwellbelowtheedgestoavoidpinchingthecables.Holdassemblytogetherfirmlywhileresecuringthetwoscrews.

8. Charge battery. Charge the newbatterybeforeusingthepipette (see Section F.4, Recharging Batteries).

F.3.1B Installing/Replacing Multichannel Electronic Pipette Batteries

TheMatrixElectronicpipetteisshippedwithNickelMetalHydridebatteriespackagedseparately.Installthebatteriesandrechargethemfor14-16hours.Usethefollowingproceduretoinstallthebatteries,thencontinuetoSectionF.4,RechargingBatteries.

Figure F.3.1Ba: Removing Battery Case

Figure F.3.1Ae: Assembling Pipette Top

Page 52: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

47

Maintenance

1. Remove battery case. Positionyourthumboverthethumbgrips(see Figure F.3.1Ba)andbraceyourfingersagainstthefrontofthekeypad.Applyfirmpressuretothepanelwiththethumbgripstoreleasethelatch.

2. Remove batteries.Pressapipettetipagainsteitherbatterythroughtheslotinthebottomofthebatterycasetoremovethebatteries(see Figure F.3.1Bb).

3. Install new batteries. Insert thespacerintothebottomofthebatterybox(as shown in Figure F.3.1Bc).Insertnewbatteriesbyplacingthepositiveendintothebatterycasefirst,thenthenegativeendinplaceoverthespacer(an exploded view of the assembly is shown in Figure F.3.1Bc, or the diagram in the battery case).Ensurethatthebatteriesarecorrectlyoriented.

Figure F.3.1Bc: Installing NiMH Batteries

Figure F.3.1Bb: Removing the Batteries

Page 53: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

48

Maintenance

F.3.2 Installing/Replacing Lithium-Ion Battery (16 Channel pipette)

Yourelectronic16channelpipetteisshippedwithaLithium-onbattery.Beforeusingthepipette,chargeitfor2hours.Usethefollowingprocedurewheninstallinganewbatterypack.Otherwise,continuetoSectionF.4,RechargingBatteries.

1. Remove battery case.Positionyourthumboverthethumbgrips(see Figures F.3.2a,b)andbraceyourfingersagainstthefrontofthekeypad.Applypressuretothepanelwiththethumbgripstoreleaselatch.Carefullyslidebatterycaseoutofthepipette,allowingsufficientslackforattachedwiresofthebattery.

2. Remove Console Cover.Unclipthetwosnapsonconsolecoverassembly,thenunhooktheendofconsolecoverassembly(see Figures F.3.2c,d).

Figure F.3.2a: Battery Case (Closed)

Figure F.3.2b: Battery Case (Open)

Figure F.3.2c: Unclip Snaps Figure F.3.2d: Unhook Cover

Page 54: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

49

Maintenance

3. Carefully Remove Console Cover.Pressandliftoffconsolecoverassembly(see Figure F.3.2e).BECAREFULnottoteartheattachedflexcable! (Refer to Figure F.3.2f).

4. Unclasp Battery Connector. Liftonesideofbatteryconnectorwithasmallflatheadscrewdriver.Graspallthreewiresandpulloutconnector(see Figures F.3.2f,g).

CAUTION: Be careful not to tear the Flex Cable

Figure F.3.2f: Lift Battery Connector

Figure F.3.2g: Unclasp Battery Connector

Figure F.3.2e: Remove Console Cover

Page 55: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

50

Maintenance

5. Connect New Battery. Attachthewiresofthenewbatteryinplaceoftheoldbattery.

6. Reassemble Console Cover.Attachthepipettetoptothemainhousing.Fittheedgesofthepipettetopsnuglywithintheedgesalongthetopofthepipettehousing.

F.4 Charging/Recharging Batteries

F.4.1 Charging/Recharging Nickel Metal Hydride (NiMH) Batteries

TheelectronicpipetteusestwoNickelMetalHydride(NiMH)rechargeablebatteries,locatedinabatterycaseunderthekeypad.Thebatteriesarechargedinthepipettebyusingthepowertransformerincludedinthepackage,withtheexceptionofsinglechannelpipetteswhichdonothavethisability.Youcanoperatethepipettewhileitisrecharging,oryoucanoperateitsolelywiththepowertransformer(withoutbatteries).Thebatteriescanberechargedapproximately400-500times.Toextendbatterylife,thepipettewillturnoffautomaticallyifitisnotusedwithintenminutes.

1. Attach transformer.Attachthecableendofthetransformerintothepipette.Plugthetransformerintotheelectricaloutlet.

2. Recharge batteries.Rechargebatteriesfor14-16hourstoensureafullcharge.Afterthebatterieshavebeenrecharged,unplugthepipettefromthe transformer.

Note: Batteries hold their maximum charge if they are fully discharged before recharging again. However, overnight recharging is acceptable if the pipette is in normal use during the day.

Page 56: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

51

Maintenance

CAUTION! Avoidexcessivechargingofthebatterieswhenthepipetteisnotinuse.Excessivechargetothebatterieswillshortenthebatterylifeandmaydamagethebatteries.

F.4.2 Charging/Recharging Lithium-Ion Battery (16 Channel Electronic Pipette)

The16channelelectronicpipetteusesaLithium-Ionrechargeablebatterylocatedinabatterycaseunderthekeypad.Thebatteryischargedwhileitisinthepipettebyusingthepowertransformerincludedinthepackage.Thepipettorcanbeoperatedwhileitisrecharging.Thebatterycanberechargedapproximately400-500times.Toextendbatterylife,thepipettewillturnoffautomaticallyifitisnotusedwithintenminutes.

Rapid Charge - Thebatteryshouldbechargedfor1.5to2hourstoachieveafullcharge.Refertothefollowingtableforpartialchargetimes:

Amount of Charge Time

80% 1.25 hour

95% 1.5 hours

100% 2hoursmax

Thebatterypackforthe16channelElectronicpipettefeaturesanindicatorlighttoindicatethestateofchargeofthebatterywhilepluggedintothepowersupply.Withtheunitpluggedin,theindicatoroperatesasfollows:

• Green -Indicatesthatthepowersupplyisconnected.• Orange -Indicatescharging(charge current into the battery).• Flashing Orange to Green -Whenthebatteryisfullydrained,theindicatorwillremainasolidorangeforawhileasthebatteryrequiresconstant charge current.

Page 57: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

52

Maintenance

Astheunitcontinuestocharge,currentissuppliedinpulses.Thisisindicatedbytheflashingoftheindicator.Asthebatterynearsfullcharge,theflashingslowsdownbecomingnearlyaconstantgreen.Whenthebatteryisfullycharged,theindicatorwillbealmostaconstantgreen,flashingonlyaboutonceevery3-5secondstoensurethebatterystaysfullycharged.

Note: The typical life expectancy of the battery is approximately 400-500 full recharges. To prolong this and attain the maximum usage from the battery, it is recommended the battery be recharged whenever it is not being used instead of waiting until the battery is completely discharged. Completely draining the battery will not shorten the life expectancy. However, recharging before fully discharged will prolong the battery life.

CAUTION! Avoidexcessivelychargingthebatterieswhenthepipetteisnotinuse(excluding16channelpipettes).Excessivechargetothebatterieswillshortenthebatterylifeandmaydamagethebatteries.

CAUTION! Thebatterypackandpowersupplyforthe16channelelectronicpipetteisnotinterchangeablewiththepowersupplyandbatterypackofotherelectronicpipettemodels.Thepowerjack/connectorandbatteryconnectoraredifferenttopreventusingthembymistakewithothermodels.

Page 58: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

53

G. TECHNICAL SERVICE

G.1 Instrument Warranty

ThisinstrumentiswarrantedtotheoriginalpurchaserbyThermoFisherScientifictobefreeofdefectsinmaterialsorworkmanshipforaperiodof12monthsfromthedateofpurchase,providingthatishasbeenoperatedaccordingtotheinstructions,notabusedormisused,thattheserialnumberhasnotbeenremoved,andthattheinstrumenthasnotbeendisassembled,otherthanforproceduresfornormalmaintenance.Nootherwarrantyisexpressedorimplied,includingthewarrantyofmerchantability.

Shouldaunitneedtobereturned,thepurchasermustpayinsuranceandshippingcharges.ThermoFisherScientificwillassumethecostofreturningtherepairedunittothepurchaser.Thisproductspackagingiscapableofwithstandingnormalshippinghazards.

ContactThermoScientificTechnicalServiceforrecommendedequipmentdecontaminationmethods(see also section G.4),ortoobtainrepairsorreplacementswithinthetermoftheabovewarrantyat800-345-0206 fax: (603) 577-7682.

ForEuropeandAsialocations,pleasevisit: www.thermo.com/matrix

Page 59: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

54

Technical Service

Trusted ServiceCertified technicians calibrate and repair Thermo Scientific Matrix pipettes as well as most other pipette brands and models. With more than 20 years experience manufacturing and servicing pipettes, we offer a broad range of services to fit every need: basic calibrations, preventive maintenance and repair services.

Streamlined ProcessSpecially designed pre-addressed pipette shipping boxes provide the ultimate in convenience with a single documentation form to select service options and certify proper decontamination.

Free shipping via FedEx ensures pipettes arrive at our facility quickly, reducing the time away from your lab. Contact your sales representative or call 800-345-0206 to have these pre-paid pipette shipping boxes delivered right to your door.

Accuracy & PrecisionEvery pipette is verified on a NIST-traceable gravimetric balance at two volumes.

If you require “as received” performance data or ISO-compliant testing protocols, we will gladly provide custom calibration services upon request.

G.2 Expert Pipette Service

Page 60: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

55

Technical Service

G.3 On-site Pipette Service Options

Services include:

• Physical pipette inspection

• Leak test

• Replacement of seal and O-ring, as needed

• Gravimetric performance validation using NIST-traceable balance

• Performance validation of five aliquots at two volumes

• Calibration adjustment, as needed

• Choice of validation sheet or CD containing results of performance validation

ValidationIf a pipette fails validation on-site, we will offer to repair it at our facility and, if necessary, deal with the original manufacturer to ensure any major repairs are completed properly.

Service and SupportOur support specialists work in partnership with you to keep your pipettes performing at their peak. We understand how vital these instruments are to your success and take pride in setting the industry standard through service excellence.

Page 61: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

56

Technical Service

G.4 Recommended Equipment Decontamination Methods

ThefollowingtableliststheproperdecontaminationproceduresforreturningMatrixElectronicPipettestoThermoFisherScientificforservice.Contactcustomerserviceat800-345-0206foracurrentequipmentdecontaminationverificationform.

Product/Component

Heat at 65 °C/10 Hr.

70%/30% IPA/Water

10% Bleach

Ethylene Oxide

Autoclave 121 °C/15 min.

Multichannel Pipettes

PipetteExterior YES Wipe Wipe YES NO

Piston/CylinderAssembly YES YES NO YES YES

TipManifold YES YES NO YES YES

Battery NO Wipe NO NO NO

PowerSupply NO Wipe NO NO NO

PipetteExterior NO Wipe Wipe NO NO

Single Channel Electronic Pipettes

Pipettew/BatteryRemoved Yes Wipe Wipe YES NO

Lowercylinder YES YES YES YES YES

Piston(donotremove) YES YES NO YES NO

PowerSupply NO Wipe NO NO NO

WARNING:Donotexposebatteriestotemperaturesabove60°C.

Page 62: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

57

Technical Service

SafetyTestingofourProducts

ThermoScientificMatrixElectronicPipetteshavepassedtherequiredsafetytestsforelectrical

productsfortheapprovalsandmarksmentioned.

Tra

nsfo

rmer

Item

#

6099,6096,1096,

1099,9068

6098,1098/1097,

9068

6098,1098/1097,

9068

Standard

EN61010-1:1993

S

afet

y re

gula

tions

EN55022/E

N55011-1

Emissions

EN50082-1:1992

Im

mun

ity

EN60825

DENAN

Directive

73/23/EEC

LowVoltage

89/336/EEC

Electromagnetic

compatibility

LawNo.234,Electical

ApplianceandMaterial

SafetyLaw,Japanese

MinistryofEconomy,

TradeandIndustry.

Approvalmark

Europe

Page 63: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

58

Technical Service

G.5 Tips and Reservoirs

Thetipsandotherconsumablesyouuseplayarolethat’sequallyasimportantasyourinstruments.InnovationatThermoFisherScientificdoesn’tstopwithpipettes.

PipetTips Pipettes BulkNonSterile Sterile

Sterile Filtered ECOTIPs

30µlTallTip 12.5µl/30µl 7600 7631 7632 7635*

250µlTip 125µl/250µl 7250 7151 7152 7275** 7156

250µlTallTip 125µl/250µl 7280 7281 7282 7285**

300µlTip 125µl/250µl 7320 7321 7322 7325***

300µlTallTip 125µl/250µl 7080 7081 7082 7085***

1250µlTip 1250µl 8050 8041 8042 8046 8045

1250µlTallTip 1250µl 8250 8241 8242 8246 8245

PipetTips Pipettes Non-Sterile

Sterile Sterile Filtered

12.5µl384Tip 12.5µl,384/16Channel 7421 7422 7245

30µl384Tip 30µl,384/16Channel 7431 7432 7435

125µl384Tip 125µl,384/16Channel 7441 7442 7445

*12.5µlfunctionalvolume**200µlfunctionalvolume***250µlfunctionalvolume

Page 64: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

59

Index

A

Autoclaving 34,38Automaticblowout 7

B

Batteries,Installation(Lithium-Ion) 48Batteries,Charging(Lithium-Ion) 51Batteries,Charging(NiMH) 50Batteries,Installation(NiMH) 44Battery Case 5

C

CAL 9“CALC”Mode 25Calibration(Procedures) 23-27Calibration(“CALC”Mode) 25Calibration(“EASY”Cal) 24Calibration(Specifications) 24CALRESET 27CALVIEW 27ColorRing 5

D

DecontaminationForm 57DecontaminationMethods 56DefaultProgrammingVolumes 14DispKey 7Dispense 14Display 5,10

E

EjectorTrigger 5“EASY”CalMode 24EnterKey

F 8

FactoryCalibrationStep 24Fill 14FillKey 7HazardsandPrecautions 12

K

Keypad 5,6

L

Lubrication 34,40

M

Maintenance(Multichannel) 35-43Maintenance(Singlechannel) 32-35MaintenanceProcedures 31-52Mix 15MixKey 7

O

On/OffKey 6OpenProgrammingMode 13-16OperationKeys 6-7

P

PACE 9PacedDispense 9,22PipettingTrigger 5ProgKey 6Program(Examples) 18-20Program(Recall) 17Program(Storing) 16ProgramKeys 7-8Purge 15PurgeKey 7

R

RUNmode 7

S

ScratchPad(Program0) 13SPd 9SpecialFunctionKey 8-10SpeedAdjustment 17

T

Tansformer Connector 5TechnicalService(InFactory) 55TechnicalService(OnSite) 54TipFittings 5TONE 9Troubleshooting 29

V

Vol Key 8VolumeRange(s) 11

W

Warranty 53

INDEX

Page 65: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available
Page 66: Thermo Scientific Matrix Electronic Pipette · 2018-11-29 · 3 Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available

Thermo Fisher Scientific, Inc.81 Wyman StreetP.O. Box 9046Waltham, Massachusetts 02454-9046United States