Pro
du
ct Sp
ecification
sP R O C E S S W A T E R A N A L Y S I S
Markets:• DrinkingWater
• Wastewater
• FoodandBeverage
• Industrial
Thermo Scientific Orion Chlorine XP
Multiple parameter water quality analyzer
From brine and wastewater to clean drinking water, the Thermo Scientific™ Orion™ Chlorine XP™ analyzer provides the ultimate accuracy and stability for the measurement of routine water quality parameters. With minimal maintenance, low cost of ownership, and multiple parameter measurement in one unit, the Orion Chlorine XP analyzer provides one of the most comprehensive water quality monitoring solution on the market.
High Performance & Reliability• Provenaccuracyandrepeatabilityeven
inharshsampleconditions(seawater,coloredwater,oilwater)
• 0-10ppmmeasurementrange–inmostapplications
• Selfzerocalibrationbeforeeachreading,enables“0”reading
• Wettestedfor24hoursbeforeshipping• 2yearwarranty
Low Price & Cost of Ownership• F&TChlorine+pH,Temp(optional
ORP,Conductivity,andTurbidity)inoneanalyzer–savesthousandsofdollarsoverpurchasingseparateanalyzers
• Lowandcustomizablereagentconsumption(default~0.033ml/sample)– lastsupto2monthsat5minutescycletime,providingsignificantreagentsavingscomparedtoothermanufacturers
Low Maintenance• Lightsourceselfcalibrationadjustment• Hands-freeselfcleansingofthephotocell• Automaticeliminationofbubblinginthe
photocell• Maintenanceremindersandalarms(once
ayearonaverage)• Autobufferrecognitionat4,7,and10
Additional Features• Upto6,4-20mAisolatedcurrentoutputs• 6xrelays• RS485,Modbus®protocolsupport• OptionalEthernetprotocoltooperations
center
The Orion Chlorine XP analyzer measures free, total, and combined chlorine, providing accurate, objective, and fast readings of chlorine in water. It minimizes the need for periodic calibrations and is compatible with all disinfectant processes. It utilizes a DPD-based measurement technology that can be configured to use the least amount of reagents while maintaining utmost accuracy.
2 ApplicationsTheChlorineXPanalyzerisrecommendedasageneralpurposechlorinemeasurementanalyzeralongwithotherparameters-pH,temperature,ORP,andflowinwaterqualitymeasurementapplications.
Drinking Water Treatment -Mostdrinkingwaterplantsusechlorineasapreferredmethodofwaterdisinfectionanddistributionmonitoring.TheChlorineXPanalyzerprovidesrapidfeedbackforchlorineadditionsforwaterdisinfectionandaccuratecheckpointforchlorinelevelsduringwaterdistribution.AsanEPAapprovedDPDmethodwithuptotwomonthsofreagentusage,theChlorineXPanalyzerdeliverssignificantcostsavingscomparedtoconventionalonemonthusagesystems.
Wastewater Treatment-Mostwastewaterplantsusechlorineasapreferredmethodofdisinfectionbeforereleasingwaterintotheenvironment.TheChlorineXPanalyzeraccuratelymeasuresbothtotalandfreechlorinecontinuouslyinthesameunitascheckpointbeforethewaterisdischarged.Thiscombinationofmeasurementseliminatestheuseoftwodifferentinstruments,resultinginoperatingcostsavingsfortheplant.
Power Generation-Powerplantsmeasurefreeandtotalchlorineforcoolingwatereffluenttomeettheregulatorylimitsforchlorinedischarge.TheChlorineXPanalyzermeasuresaslowas10ppbfortotalchlorineinseawater,wellbelowtheregulatoryrequirementofminimumdetectionlimitsinallsamplesmatrices.
Food & Beverage Manufacturing-FoodandbeverageplantsneedtomeasurechlorinebeforewatergoesthroughtheROprocessbecausechlorineshortensthelifeoftheROmembranebychemicallyreactingtothemembrane.TheChlorineXPanalyzerprovideslowleveldetection(upto10ppb)andaquickresponsetime,extendinglifespanofROmembraneswhichtranslatestolowoperatingcostsfortheplant.
Dialysis Treatment-Dialysistreatmentcentersmonitorthechlorinelevelinkidneydialysismachinestomakesurethatitdoesn’tcomeincontactwithblood.Chlorine/chloraminesinwaterusedfordialysiscanresultinseriousadversepatientreactionsordeath.TheChlorineXPanalyzermeasureslessthan0.5ppm(max.limit)totalchlorineandprovidesquickresponsetime,resultinginROlifeextensionandimprovingcarbonfilterefficiency.
Theoperatingsystemisdesignedtobesimpleandintuitive.Onceinstalledandcalibrated,theChlorineXPanalyzerautomaticallyreleasestheproperquantityofchemicalsdependingonmeasurementfrequency.TheChlorineXPanalyzercomesstandardwithchlorinemeasurementandcanbeconfiguredtomeasureanycombinationoffreechlorine,totalchlorine,andbothfreeandtotalchlorine.
3Principle of OperationsTheDPD(N,N-diethyl-p-phenylenediamine)methodforresidualchlorinewasfirstintroducedbyDr.A.T.Palinin1957.Overtheyearsithasbecomethestandardmethodfordeterminingfreeandtotalchlorineinwaterandwastewater.WhenDPDreactswithsmallamountsofchlorineatanearneutralpH,acoloreddyeistheprincipaloxidationproduct.TheDPDdyecolorismeasuredphotometricallyatwavelengthsrangingfrom490to555nanometers(nm),whichresultsinaccuratereadingofbothfreeandtotalchlorineconcentrationsinwater.
Freechlorine(hypochlorousacid+hypochloriteions)reactswiththeFreeChlorineIndicator(DPD1)causingacolorchangeinthesamplefromcleartored.ThebufferisusedtoensurereactionataconsistentpH.Themorefreechlorinethatispresent,thedarkertheredcolorthatforms.ThecolorintensityisconvertedtoppmemployingtheBeer-Lambertlaw.
Totalchlorineisoftenusedformonitoringcombinedchlorine(chloramines)levels.Combinedchlorineisthedifferencebetweenthetotalchlorineandfreechlorine.Whenfreechlorineandtotalchlorineneedtobemeasuredinsequence,theTotalChlorineIndicator(DPD3)isaddedtothesamplealreadycontainingtheFreeChlorineIndicator(DPD1)andbuffer.ThecombinedchlorinereactswiththeTotalChlorineIndicatorcausinganincreaseintheredcolor.Thetotalchlorineisdeterminedfromthecolorchangeandthecombinedchlorineiscalculatedfromthedifferencebetweenthetotalandfreechlorine.Whenonlytotalchlorinetestisneeded,theTotalChlorineIndicator(DPD4)isaddedtothesamplecausingcolorchangeandthelevelsoftotalchlorineisdetermined.
Colorimeter water outlet
Colorimeter drain
Rotating flow switch
Reagent injection
Reagent pumps
Reagentsiphon
caps
Reagentbottles
set
Control unit
Control panel
Water intake filter
I/O module
Colorimeter module
pH, Redox, temperature module
Drain cap
Samplingcell
View window
Solenoid valve Analyzing unit
Pro
du
ct Sp
ecification
s
thermoscientific.com/processwater
©2013ThermoFisherScientificInc.Allrightsreserved.ModbusisaregisteredtrademarkofSchneiderElectric.AllothertrademarksarethepropertyofThermoFisherScientificInc.&itssubsidiaries.
Cat. No. Description
CXP71 FreeChlorineonlywith2X4/20mAoutputs
CXP72 TotalChlorineonlywith2X4/20mAoutputs
CXP73 Free & TotalChlorinecombined,2x4/20mAoutputs
CXP71PH FreeChlorine+ pH + Temp.2x4/20mAoutputs
CXP72PH TotalChlorine+pH + Temp.2x4/20mAoutputs
CXP73PH Free & TotalChlorinecombined + pH + Temp2x4/20mAoutputs
CXPRGDPD1F 5SetsofDPD1reagentsforFreeChlorine
CXPRGDPD4T 5SetsofDPD4reagentsforTotalChlorine
CXPRGDPD3T 5setsofDPD3reagentsforTotalChlorine(needtobeusedwithDPD1forCXP73andCXP73PH)
Product Ordering Information
Performance
Accuracy +/-3%
Repeatability 0.5%
MinimumDetectionLimit 10ppb
Zero-Point-Adjustment Selfzerobeforeeachreading
CycleTime 2to10minutes
FlowMonitoring Rotaryflowswitch(seeAdditionalRequirementssectionforinlet/outletpressure)
Parameters FC,TC,F&TC,ORP,pH,Temp.
MeasuringRange 0to10ppm(Cl);0-14(pH)
Reliability
Warranty 2years
Validation Contactfactory
USEPAAcceptedMethod Yes
CSACertified Yes
CECertified Yes
Operations Requirements
Maintenance 1-2monthsforreagentreplacementandfiltercleaning
Calibration Every6months(pHandORPonly)
ReagentUsage DPDupto2monthsat5min.cycletime
PowerConsumption Approx.60VA
PowerSupply 100-115VAC,50/60Hz,1.0Aor200-230VAC50/60Hz,0.5A
Weight 24lbs.(11kg)
Dimensions 67cmx33cmx14cm,26’’x13’’x5.5’’(WxHxD)
Product Specifications
Options/features
Alarms Optional,seemanual
RemoteMonitoring Notyetavailable
Enclosure IP-65ratedenclosures(NEMA4equivalent)
LocalI/O 2standard4to20mAoutputs4optional4to20mAoutputs
Memory 256K
Lines 1000
EventLogger Yes
TotalRelayOnTime Yes
DisplayType 5.5”graphicmonochromatic;characterLCDwithbackgroundlightalarmsandstatus
Password Operatorandtechnician
Servicing/Maintenance Self-cleaningphotocell(minimumservicerequirement)
Additional Requirements
SampleandDrainConnection
Pressurizedsampleinletandgravitydrain
SampleTemperature 32oFto212oF(0oCto100oC)AmbientTemperature:15oFto131oF;(-10oCto55oC)
InletPressure 5-15psi
SampleConditioning Notrequired
C.C. Lynch & Associates300 Davis Avenue, Pass Christian, Mississippi 39571Regional Offices: Austin, Baton Rouge, Dallas,Houston, Lafayette, MobileToll Free: 800.333.2252Fax 228.452.2563www.cclynch.com I [email protected]