This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump Youngseok Park 1,2 , Yunmok Son 2 , Hocheol Shin 2 , Dohyun Kim 2 , and Yongdae Kim 2 1 NAVER Labs 2 System Security Laboratory, KAIST 10th USENIX Workshop on Offensive Technologies (WOOT '16) Aug.09.2016
Transcript
Page 1: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

This Ain’t Your Dose:Sensor Spoofing Attack on Medical

Infusion Pump

Youngseok Park1,2, Yunmok Son2, Hocheol Shin2, Dohyun Kim2, and Yongdae Kim2

1 NAVER Labs2 System Security Laboratory, KAIST

10th USENIX Workshop on Offensive Technologies (WOOT '16)

Aug.09.2016

Page 2: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Sensor

v Sensing changes in physical property and converting to electric signal

v Gyroscope, Accelerometer, Radar, Sonar, Infrared sensor, etc.

2

Page 3: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Sensing and Actuation System

3

Real World

Processor

Sensor Actuator

Sensing Actuation

System

ADC

ADC: Analog-to-Digital Converter

Converting Processing

GyroscopeRadarFlightcontrol

Crashavoidance

Page 4: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Sensing and Actuation System

4

Real World

Processor

Sensor Actuator

Sensing Actuation

System

ADC

ADC: Analog-to-Digital Converter

Converting Processing

GyroscopeRadarFlightcontrol

Crashavoidance

No Authentication

Vulnerable to sensor spoofing attack

Spoofing!

Page 5: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Sensor Spoofing Attack

v Manipulating sensors with a malicious signal

v Previous works- Attacking Circuit using EMI: Injecting EMI into a wire of a defibrillator (S&P’13)

- Canceling and injecting Active Sensor Signal: magnetic signal on ABS sensor (CHES’13)

- Generating Resonance (DoS): Injecting sound noise into a gyroscope of a drone (SEC’15)

5

EMI: Electromagnetic InterferenceABS: Anti-lock Braking System

Page 6: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

6

This Work: Manipulating Sensing Valuesby Saturating Receiver

Page 7: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Target: Medical Infusion Pump

v Controlling infused volume of medicine to patients

v Sometimes using a drop sensor for accuracy

7 Infusion Pump (body)

Display

Controlpanel

Actuator(PeristalticFingers)

IV Tube

To human’s body

From drop sensorMedicine

IR receiver

IR emitter

To infusion pumpbody

Drop sensor

Drop

IV TubeDrip

chamber

Output

~

Page 8: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Infusion Pump Operation

8

Light

Page 9: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Sensor Saturation

v New type of sensor spoofing attack using saturation- Sensors have typical operating region- Output is saturated when exceeding a saturation point- Blinding sensors

9

In case of the infusion pump

Page 10: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Medical Infusion Pump

v Two infusion pumps with drop sensors

10

Infusion pump Drop sensor

JSB-1200(Pump1)

BYS-820(Pump2)

Page 11: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Hardware Analysis

v Pump1 (JSB-1200)

11

Peristalticfingers

Tube

Infusion pump

LED

Drop sensor

IR emitter

IR receiver

IR Filter

Page 12: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Hardware Analysis

v Measuring signal with oscilloscope- Connector = 4 pins: VCC, GND, LED, and IN (signal)

12

Connector(Device side)

Four pins(Sensor side)

Normal drop

Page 13: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Simple Test (Saturation, w/o filter)

13

Page 14: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Simple Test (Saturation, w/o filter)

14

Page 15: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Hardware Analysis

v Mainboard (2 MCUs)

15

W78E516D(MCU2)

AT89S52(MCU1)

Internal structure

SPI Port

Drop sensor port

Page 16: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Hardware Analysis

v Sensor output is inserted to MCU1 after ADC- 8-bit ADC (0 to 255)- Digital signal indicates voltage level of the drop sensor

16

Output of ADC

8-bit ADC

IN(sensor output)

MCU1

Page 17: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Firmware Extraction

v Extracting firmware of MCU1 via SPI port- Reading Flash memory using USBISP and AVR Studio- Data section -> 8051 assembly -> IDA Pro

17

USBISP

AVR Studio 4 Intel HEX format

Data sectionAT89S52(MCU1)

SPI Port

Page 18: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Firmware Analysis

v Finding sensor output in Timer interrupt function

18

Put 8-bit sensor output to RAM

Page 19: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Firmware Analysis

19

Page 20: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Drop Detection Algorithm

20

Sensing drop when voltage decreases by 𝟎.𝟑𝟐𝑽

Send command (0x11) through serial port,connected to MCU2

Page 21: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Pump1 Structure

1. Drop sensor output enters into AT89S52 (MCU1)

2. MCU1 sends data to W78E516D (MCU2) via serial comm.

3. MCU2 actuates peripherals with this data- Pins of MCU2 are directly connected to motor, display and alarm

21

Page 22: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Vulnerability

v Drop sensor- Saturated with an external source- Cannot sense drops in saturation

v Drop detection algorithm- Counting drops based on a relative change in voltage- Making a voltage drop to sensor output

22

Saturation

Fake drop

Page 23: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Experimental Setting

23

Measuringcylinder

IR Laser(905nm, 30mW) Drop sensor

Arduino

Infusion pump

Page 24: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Experiment

v Performed on both infusion pumps (Pump1, Pump2)

v Saturation (failed in Pump2)- Sensor is saturated when injecting IR laser to receiver

- Drop sensor cannot sense real drops -> Over-infusion

v Fake drops- Sensor is deceived by fake drops with external IR

- Pump perceives that there are drops already -> Under-infusion

v Both cases cause an alarm

24

Page 25: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Spoofing Pattern

v Over-infusion- Alarm: “No drop is detected”- Inject some period and compensate insufficient drops

v Under-infusion- Alarm: “Too many drops are detected”- Find properly interval of fake drops experimentally

v Example (60mL/h setting)- 1 drop per 3 seconds

25

Normal operation

Continuous saturation

Over-infusion

Saturation time (13s)

Real drop interval (3s) drop fake drop

Alarm

Under-infusion

Fake drop interval

2s

Page 26: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Demo (Over-infusion)

26

Page 27: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Demo (Under-infusion)

27

Page 28: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Spoofing Pattern

v Over-infusion- Alarm: “No drop is detected”- Inject some period and compensate insufficient drops

v Under-infusion- Alarm: “Too many drops are detected”- Find properly interval of fake drops experimentally

28

Normal operation

Continuous saturation

Over-infusion

Saturation time

Real drop interval drop fake drop

Alarm

Under-infusion

Fake drop interval

2s

Page 29: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Results

v Controlling infused volume is possible- By adjusting saturation time or fake drops- Measured in 10 minutes and 5 times each (No alarm rings over 30 minutes)- Over-infusion fails on Pump2

29

Page 30: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Discussion

v Attack distance- Related to power of source- Possible in the range of 12m with 30mW IR laser

v Mitigation- Authentication between emitter and receiver

• PyCRA (CCS ‘15)

• Generate random zero signal in an emitter

- Voltage level detection• Checking boundary of legitimate signal

- Physical isolation

30

Saturation(by spoofing)

Sensor output

Real drops(without spoofing)

Boundary check

Detect!

Concept of PyCRA

Voltage level detection

Page 31: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Discussion

v Attack distance- Related to power of source- Possible in the range of 12m with 30mW IR laser

v Mitigation- Authentication between emitter and receiver

• PyCRA (CCS ‘15)

• Generate random zero signal in an emitter

- Voltage level detection• Checking boundary of legitimate signal

- Physical isolation

31

Page 32: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious

Conclusion

v Presenting a new type of sensor spoofing attack- Deceiving a sensor by saturation

v Analysis on medical infusion pumps- Finding vulnerability in drop detection algorithm

v Controlling infused fluid from 65% to 330%

v Note- Infusion pump was not communicating at all. - IR lay is invisible to human eyes.- FDA approved US devices?

v Sensor security- Most sensors are exposed to receive signal- Must be considered for safety

32

Page 33: This Ain’t Your Dose: Sensor Spoofing Attack on Medical ... · This Ain’t Your Dose: Sensor Spoofing Attack on Medical Infusion Pump ... v Manipulating sensors with a malicious