Page 1 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
1.0 INDENTIFICATION
1.1 EPA 335.4 – Determination of Total Cyanide by Semi-Automated Colorimetry
2.0 APLICATION MATRIX
2.1 This method covers the determination of cyanide in drinking water.
3.0 DETECTION LIMIT
3.1 Refer to Laboratory Method Quality Performance Data File
4.0 SCOPE AND APPLICATION
4.1 This method covers the determination of cyanide in drinking water.
4.2 The applicable concentration range is 0.005 to 0.500 mg CN- / L.
5.0 SUMMARY OF TEST METHOD
5.1 Total cyanide is determined by distilling the sample and measuring cyanide
generated using a technique for cyanide ion detection (e.g., colorimetry).
5.2 In micro distillation, the hydrocyanic acid is released from cyanide complexes by
means of a manual reflux-distillation operation and absorbed in a scrubber
containing sodium hydroxide solution. The cyanogen ion in the absorbing solution
is converted to cyanogen chloride by reactions with chloramine-T, which
subsequently reacts with pyridine and barbituric acid to give a red-colored complex.
6.0 DEFINITION
6.1 Calibration Blank (CB) — A volume of reagent water fortified with the same matrix
as the calibration standards but without the analyte.
6.2 Calibration Standard (CAL) — A solution prepared from the primary dilution
standard solution or stock standard solutions. The calibration standard solution
solutions are used to calibrate the instrument response with respect to analyte
concentration.
6.3 Initial Demonstration of Capability – Include a reagent blank and four laboratory
fortified blank (LFB) is used to demonstrate proficiency in performing method and
obtaining acceptable results for each analyte.
6.4 Instrument Performance Check Solution (ICP) — A solution of one or more
methods analytes, or other test substances used to evaluate the performance of the
instrument system with respect to a defined set of criteria.
6.5 Laboratory Fortified Blank (LFB) — An aliquot of reagent water or other blank
matrices to which known quantities of the method analytes are added in the
laboratory. The LFB is analyzed exactly like a sample, and its purpose is to
determine whether the methodology is in control, and whether the laboratory is
capable of making accurate and precise measurements.
6.6 Laboratory Reagent Blank (LRB) — An aliquot of reagent water or other blank
matrices that are treated exactly as a sample including exposure to all glassware,
equipment, solvents, reagents, that are used with other samples. The LRB is used
to determine if method analytes or other interferences are present in the laboratory
environment, the reagents, or the apparatus.
Page 2 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
6.7 Matrix spike (MS) and matrix spike duplicate (MSD) - Aliquots of an
environmental sample to which known quantities of the analytes are added in the
laboratory. The MS and MSD are prepared and/or analyzed exactly like a field
sample. Their purpose is to quantify any additional bias and imprecision caused by
the sample matrix. The background concentrations of the analytes in the sample
matrix must be determine in a separate aliquot and the measured values in the MS
and MSD corrected for background concentrations.
6.8 Quality Control Sample (QCS) — A solution of method analytes of known
concentrations that is used to fortify an aliquot of LRB or sample matrix. The QCS
is obtained from a source external to the laboratory and different from the source
of calibration standards. It is used to check laboratory performance with externally
prepared test materials.
6.9 Linear Calibration Range (LCR) — The concentration range over which the
instrument response is linear.
6.10 Method Blank - An analyte-free sample to which all reagents are added in the same
volumes or proportions as used in sample processing. The method blank must be
carried through the complete sample preparation and analytical procedure. The
method blank is used to assess contamination resulting from the analytical process.
6.11 Method Detection Limit – The minimum concentration of an analyte that can be
identified measured and reported with 99% confidence that analyte concentration
is greater than zero.
6.12 Minimum Reporting Limit (MRL) – Is the lowest standard concentration that could
be analyzed obtaining a response using the established analyte’s calibration curve
and could be statistically treated.
6.13 Ongoing Precision and Recovery Standard (OPR) – The ongoing precision and
recovery standard test is used to ensure the laboratory meets the performance
criteria during the period that the samples are analyzed. The OPR solution is an
aliquot of method blank to witch known quantities of the methods analytes are add
in the laboratory. The OPR is analyzed in the same manner as samples.
7.0 INTERFERENCES
7.1 Aldehydes, nitrate-nitrite, oxidizing agents, such as chlorine, thiocyanate,
thiosulfate and sulfide. Some of these interferences are reduced or eliminated by
the distillation process.
7.2 Sulfides adversely affect the procedure by producing hydrogen sulfide during
distillation. If a drop of the sample on lead acetate test paper indicates the presence
of sulfide, treat 25 mL more of the sample volume required for the cyanide
determination with powdered cadmium carbonate. It forms a precipitate if the
sample contains sulfide. Repeat this operation until a drop of the treated sample
solution does not darken the lead acetate test paper. Filter the solution through a
dry filter paper into a dry beaker, and from the filtrate, measure the sample to be
used for analysis. Avoid a large excess of cadmium and a long contact time in order
to minimize a loss by complexation or occlusion of cyanide on the precipitated
material.
7.3 High results may be obtained for samples that contain nitrate and/or nitrite. During
the distillation, nitrate and nitrite will form nitrous acid water that will react with
Page 3 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
some organic compounds to form oximes. The oximes will decompose under test
conditions to generate HCN. The interference of nitrate and nitrite is eliminated by
pretreatment with sulfamic acid.
7.4 Oxidizing agents, such as chlorine, decompose most of the cyanides. Test a drop
sample with potassium iodide-starch paper (KI-starch paper) moistened in acetate
buffer pH 4.0 at time of collection; a blue color indicates the need for treatment.
Add ascorbic acid, a few crystals at a time, until a drop of sample produces no color
on the indicator paper; then add an additional 0.06 g of ascorbic acid for each liter
of sample volume. Sodium arsenite has also been employed to remove oxidizing
agents. Add a small amount of sodium arsenite solution, until a drop of sample
produces no color on the indicator paper.
7.5 Method interferences may be caused by contaminants in the reagent water,
reagents, glassware, and other sample processing apparatus that bias analyte
response.
8.0 SAFETY
8.1 Each chemical should be regarded as a potential health hazard and exposure should
be as low as reasonably achievable.
8.2 A reference file of all reagent’s Safety Data Sheets (SDS) should be available to all
personnel involved in the chemical analysis.
8.3 The following chemicals have the potential to be highly toxic or hazardous:
8.3.1 Potassium Cyanide – Very hazardous in cases of skin and eyes contact
(irritant), of ingestion and inhalation. Severe over exposure can produce
lung damage, choking, unconsciousness or death.
Recommended personal protection:
8.3.1.1 Safety glasses or goggles.
8.3.1.2 Use an approved/certified respirator or manage under certified
fume hood.
8.3.1.3 Chemical resistant gloves.
8.3.1.4 Protective clothing
8.3.2 Hydrochloric acid - Very hazardous in case of skin contact (Corrosive).
Recommended personal protection:
8.3.2.1 Face Shield
8.3.2.2 Respirator with vapor protection filter (If not managed under
certified fume hood)
8.3.2.3 Strong Acids and Chemicals resistant gloves
8.3.2.4 Protective clothing
8.3.3 Pyridine - OSHA regulated air exposure (<250 ppm exposure)
Recommended personal protection:
8.3.3.1 Full face respirator with Organic vapor filter (If not managed
under certified fume hood)
8.3.3.2 Protective gloves
8.3.3.3 Protective clothing
8.3.4 Sulfuric Acid - Extremely corrosive, causes severe burns with contact and
if inhaled, mist can cause irritation
Recommended personal protection:
Page 4 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
8.3.4.1 Face shield and goggles
8.3.4.2 Protective clothing
8.3.4.3 Chemical resistant gloves (Strong acid resistant gloves)
8.3.4.4 Respirator with vapor/fumes/dust protection filters for
environments exposure up to 10mg/m3 (If not managed under
certified fume hood.
9.0 EQUIPMENT
9.1 Analytical Balance, capable of accurately weighing to the nearest 0.0001 g
9.2 Glassware Class A
9.3 Distillation apparatus – Macro distillation (Section 23.5) or Micro distillation
System (Section 23.3 & 23.4)
9.4 Heating block
9.5 Automated continuous flow analyzer- LACHAT Quickchem 8500
SN:070400000435
9.5.1 Sampling device (Sampler)
9.5.2 Multichannel pump
9.5.3 Reaction unit or manifold
9.5.4 Colorimetric detector
9.5.5 Data recording device
9.6 Buret
9.7 Filter Paper
9.8 Lead Acetate Test Paper
9.9 Potassium Iodide –Starch Paper
9.10 pH Meter
10.0 REAGENTS & STANDARD
10.1 Reagents
10.1.1 Cadmium Carbonate Powder, CdCO3(CARN: 513-78-0)
10.1.2 Reagent Water: Deionized Water
10.1.3 Sodium Arsenite Solution (5.0 g/L) (CASRN-7784-46-5) (Commercially
available)
10.1.4 Lead Carbonate Powdered, PbCO3
10.1.5 Distillation Reagents
10.1.5.1 Boiling Chips Environmental Express # C5249 or Equivalent
10.1.5.2 Magnesium Chloride (CASRN-7786-30-3) (MgCl2•6H2O)
10.1.5.3 Sodium Hydroxide (CASRN-1310-73-2), NaOH
10.1.5.4 Sulfamic Acid (CASRN-5329-14-6)
10.1.5.5 Sulfuric Acid (CASRN-7664-93-9) (H2SO4)
10.1.6 Standardization Reagents
10.1.6.1 Acetone
10.1.6.2 P-Dimethylaminobenzalrodanine
10.1.6.3 Silver Nitrate (CASRN-7761-88-8), AgNO3
10.1.7 Instrument Reagents
10.1.7.1 Ascorbic acid: Crystal (CASRN-50-81-7)
10.1.7.2 Barbituric Acid (CASRN-67-52-7)
Page 5 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
10.1.7.3 Chloramine-T (CASRN-127-65-1)
10.1.7.4 Hydrochloric Acid (CASRN-7647-01-0) (HCl)
10.1.7.5 Pyridine (CASRN-110-86-1)
10.1.7.6 Sodium dihydrogenphosphate (CASRN-10049-21-5)
(NaH2PO4•H2O)
10.1.8 Stock Standard Reagents
10.1.8.1 Potassium Cyanide (CASRN-151-50-8) (KCN)
10.1.8.2 Potassium Hydroxide (CASRN-1310-58-3) (KOH)
10.2 Distillation Solutions
10.2.1 Sodium Hydroxide Solution 1.00 N
10.2.1.1 Dissolve 40g of NaOH in reagent water
10.2.1.2 Mix with magnetic stirrer until dissolve.
10.2.1.3 Cool solution
10.2.1.4 Dilute solution to 1L using volumetric flask
10.2.2 Sodium Hydroxide Solution 1.25 N
10.2.2.1 Dissolve 50g of NaOH in reagent water
10.2.2.2 Mix with magnetic stirrer until dissolve.
10.2.2.3 Cool solution
10.2.2.4 Dilute solution to 1L using volumetric flask
10.2.3 Sodium Hydroxide Solution 0.25N
10.2.3.1 Dilute 200 mL of 1.25N Sodium Hydroxide Solution (Section
10.2.2) to 1L with reagent water.
10.2.4 Sulfuric Acid, 18N (1+1)
10.2.4.1 Place a 1L glass beaker in an ice bath with magnetic stirrer.
10.2.4.2 Add 500 mL reagent water.
10.2.4.3 Add slowly 500 mL of concentrated Sulfuric Acid.
10.2.4.4 Cool solution
10.2.4.5 Dilute to 1L with reagent water.
10.2.5 Magnesium Chloride
10.2.5.1 Dissolve 510g of Magnesium Chloride with reagent water in 1L
volumetric flask
10.2.5.2 Mix with magnetic stirrer until dissolve.
10.2.5.3 Dilute to the mark with reagent water, invert to mix
10.2.6 Sulfamic Acid
10.2.6.1 Dissolve 20g of reagent grade sulfamic acid in reagent water.
10.2.6.2 Mix with magnetic stirrer until dissolve.
10.2.6.3 Dilute to 1L volumetric flask with reagent water.
10.3 Instrument Solutions
10.3.1 Pyridine Burbuturic Acid Reagent
10.3.1.1 Weight 15g of Barbituric Acid in a 1L volumetric flask.
10.3.1.2 Wash the side of the beakers with about 100mL of reagent water.
10.3.1.3 Mix with magnetic stirrer.
10.3.1.4 Add 75mL of pyridine.
10.3.1.5 Add 15 mL of HCl and mix.
10.3.1.6 Dilute to 900mL with reagent water and mix until the barbituric
acid had dissolved.
Page 6 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
10.3.1.7 Dilute to the mark with reagent water.
10.3.2 Sodium Dihydrogenphosphate Buffer
10.3.2.1 Dissolve 138g NaH2PO4·H2O with reagent water in 1000mL
volumetric flask
10.3.2.2 Mix with magnetic stirrer.
10.3.2.3 Dilute to the mark with reagent water.
10.3.3 Chloramine-T
10.3.3.1 Dissolve 2.0 g chloramine-T with reagent water in 500 mL
volumetric flask.
10.3.3.2 Mix with magnetic stirrer until dissolve.
10.3.3.3 Dilute to mark with reagent water.
10.4 Standardization Reagents
10.4.1 Indicator Solution
10.4.1.1 Dissolve 20mg p-dimethylaminobenzalrhodaine in 100mL
acetone.
10.4.2 Standard Silver Nitrate Solution, 0.0192 N
10.4.2.1 Prepare by crushing approximately 5g AgNO3 crystals and
drying to constant weight at 40ºC.
10.4.2.2 Weight out 3.2647g of dried AgNO3.
10.4.2.3 Dissolve with reagent water in 1000mL volumetric flask.
10.4.2.4 Dilute to mark with reagent water.
10.4.3 Sodium Hydroxide Dilution Solution
10.4.3.1 Dissolve 1.6g NaOH in 1000mL volumetric flask.
10.4.3.2 Mix with magnetic stirrer.
10.4.3.3 Dilute to mark with reagent water.
10.5 Stock & Working Standard Solutions
10.5.1 Cyanide Stock Standard Solution 1000mg/L
10.5.1.1 Dissolve 2.51g Potassium Cyanide (KCN) and 2.0g Potassium
Hydroxide (KOH) with reagent water in 1000mL volumetric
flask.
10.5.1.2 Mix with magnetic stirrer
10.5.1.3 Dilute to the mark with reagent water.
10.5.2 Cyanide Intermediate Standard Solution 100mg/L
10.5.2.1 Transfer 10.0mL of Cyanide Stock Standard Solution 1000mg/L
(Section 10.5.1) to 100mL volumetric flask.
10.5.2.2 Dilute to volume with Reagent Water and mix well.
10.5.3 Cyanide Working Standard Solution 10.0mg/L
10.5.3.1 Transfer 20.0mL of Cyanide Stock Standard Solution 100mg/L
(Section 10.5.2) to 200mL volumetric flask.
10.5.3.2 Dilute to volume with Reagent Water and mix well.
10.5.3.3
10.6 Calibration Curve
Note: Is not imperative that all standards to be distilled in the same manner as the
sample. At least two standards (a high and low) and a blank should be distilled and
compared to similar value on the standard curve to insure that the distillation
technique is reliable. Before distillation, standards should contain 4mL NaOH
Page 7 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
0.25N (Section 10.2.2) per 50.0mL. Theses distilled standard must agree within
±10% of the undistilled standards.
10.6.1 Calibration Curve Standard 0.300 mg/L
10.6.1.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 3.0mL to 100 mL volumetric flask.
10.6.1.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.6.1.3 Dilute to volume with Reagent Water.
10.6.2 Calibration Curve Standard 0.100 mg/L
10.6.2.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 1.0mL to 100 mL volumetric flask.
10.6.2.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.6.2.3 Dilute to volume with Reagent Water.
10.6.3 Calibration Curve Standard 0.05 mg/L
10.6.3.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 0.50mL to 100 mL volumetric flask.
10.6.3.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.6.3.3 Dilute to volume with Reagent Water.
10.6.4 Calibration Curve Standard 0.030 mg/L
10.6.4.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 0.30mL to 100 mL volumetric flask.
10.6.4.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.6.4.3 Dilute to volume with Reagent Water
10.6.5 Calibration Curve Standard 0.01 mg/L
10.6.5.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 0.10mL to 100 mL volumetric flask.
10.6.5.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.6.5.3 Dilute to volume with Reagent Water
10.6.6 Calibration Curve Standard 0.005 mg/L
10.6.6.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 0.05mL to 100 mL volumetric flask.
10.6.6.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.6.6.3 Dilute to volume with Reagent Water
10.7 Standards & Quality Control Solutions Preparation
10.7.1 Calibration Blank
10.7.1.1 Add reagent water in 100mL volumetric flask
10.7.1.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.7.1.3 Dilute to volume with Reagent Water.
10.7.2 Quality Control Sample (QCS) 0.05mg/L
10.7.2.1 From Cyanide Working Standard Solution 10.0mg/L using a
different source preparation transfer 0.50 mL to 100 mL
volumetric flask.
10.7.2.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.7.2.3 Dilute to volume with Reagent Water.
10.7.3 Instrument Performance Check (IPC) 0.03mg/L
10.7.3.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 0.30mL to 100 mL volumetric flask.
Page 8 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
10.7.3.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.7.3.3 Dilute to volume with Reagent Water
10.7.4 Laboratory Reagent Blank (LRB) & Laboratory Reagent Blank
Duplicate
10.7.4.1 Add reagent water in 100mL volumetric flask
10.7.4.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.7.4.3 Dilute to volume with Reagent Water.
10.7.5 Laboratory Fortified Blank (LFB) & Laboratory Fortified Blank
Duplicate 0.03mg/L
10.7.5.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 0.30mL to 100 mL volumetric flask.
10.7.5.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.7.5.3 Dilute to volume with Reagent Water
10.7.6 On-going Precision & Recovery 0.05mg/L
10.7.6.1 From Cyanide Working Standard Solution 10.0mg/L (Section
10.5.3) transfer 0.50mL to 100 mL volumetric flask.
10.7.6.2 Add 20mL NaOH 1.25N (Section 10.2.2)
10.7.6.3 Dilute to volume with Reagent Water
10.7.7 Sample spikes (Laboratory fortified matrix):
10.7.7.1 Spikes concentration should be the same as used in the
laboratory fortified blank. Perform sample spikes a minimum of
10% of the total of samples to be analyzed by this method.
Spikes are to be prepared in duplicate.
11.0 SAMPLE COLLECTION, PRESERVATION SHIPMENT AND STORAGE
11.1 Use plastic or glass bottles.
11.2 Preserve with ascorbic acid and sodium hydroxide (pH>12) and cool to ≤4 ºC at
the time of collection.
11.3 Samples should be analyzed as soon as possible after collection sample should be
checked for possible interferences. If storage is required, preserved samples are
maintained at ≤4 ºC and may be held for up to 14 days.
12.0 QUALITY CONTROL
12.1 The laboratory must have records of the implementation of proper quality controls
to ensure the validity of the results that the client receives after this technique is
applied. This must contain evidence of the following being performed and verified
periodically (if applicable):
12.1.1 Linear Calibration Range (LCR)
12.1.1.1 Must be determined initially and verified every six month. The
initial demonstration of linearity must use sufficient standards to
insure that the resulting curve is linear.
12.1.1.2 The verification of linearity must use a minimum of a blank and
three standards.
12.1.1.3 Is determined by successive analyses of higher concentration
standards until the results are less than 90% of the target value
of the standard.
Page 9 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
12.1.2 Initial Demonstration of Capability (IDC)
12.1.2.1 Must be determined when every new analyst performs this
analysis to demonstrate proficiency with the analytical method.
The analyst must recalculate IDC when a change in the method,
analyst or instrument is made, after prolonged shutdowns, mayor
instrumental repairs and whenever a change could affect the
precision, accuracy or sensitivity.
12.1.2.2 Run a reagent blank and four laboratory fortified blank at the
concentration between 10 times the MDL and the midpoint of
the calibration curve. Then calculate the percent of recovery of
each standard.
12.1.2.3 The percent recovery should within 90-110% of it true value,
and the calculated percent relative standard deviation should be
at or below 15%.
12.1.3 Method Detection Limit (MDL) should be determined every six month by
an experience analyst, when a new analyte begins to work with the method
or whenever there is a significant change in the background instrument
response. Refer to procedure ALQAP DW 009- Determination of Method
Detection Limit and Initial Demonstration of Performance to determine the
method detection limit value.
12.2 Laboratory Performance
12.2.1 Laboratory Reagent Blank (LRB) must be analyzed at least once with
duplicate with each batch of samples. Result must not exceed the laboratory
establish MDL value.
12.2.2 Laboratory Fortified Blank (LFB)
12.2.2.1 Must be analyze at least once with duplicate with each batch of
samples.
12.2.2.2 Laboratory will establish control limit of data generated from
laboratory fortified blanks using as minimum 20 to 30 analyses
results when data becomes available.
12.2.2.3 Until laboratory establish acceptable control limit, calculate
accuracy as percent of recovery and it results must be within 90-
110%.
12.2.3 Instrument Performance Check Solution (IPC is a mid-range check
standard) must be analyze following daily calibration, after every tenth
sample and at the end of the sample run. Analysis of the IPC solution and
calibration immediately following calibration and the subsequent analyses
of the IPC solution must be verify that the instrument is within 90 – 110%
of its true value.
12.2.4 Laboratory Fortified Sample Matrix (LFM) must be performed in duplicate
on 10% of samples to be analyzed with every batch of sample. The analyte
concentration must be high enough to be detected above the original sample.
The added analyte concentration should be the same as that used in the
laboratory fortified blank.
12.2.5 Minimum Reporting Limit (MRL) must be verify in every batch of sample
by analyzing fortified blank at the minimum reporting limit. The reporting
Page 10 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
limit must be below the minimum contamination level. Percent of
recovery must be within 90-110%.
12.2.6 Quality Control Sample (QCS) is obtained from a source external to the
laboratory and different from the source of calibration standards. Calculate
accuracy as percent of recovery, results must be within 90-110%.
12.2.7 For applicability and acceptance criteria refer to Section 23.2.
13.0 CALIBRATION & STANDARDIZATION
13.1 Prepare a series of six standards (Section 10.6), covering the desired range, and a
blank.
13.2 Pump all reagents and let the system equilibrate until a stable baseline is achieved.
13.3 For additional parameters refer to Section 23.2.
13.4 Place appropriate standards in the sampler in order of decreasing concentration
(Section 23.1).
13.5 Calibrate the instrument with the calibration curve standards and a calibration blank
(Section 23.1)
13.6 Run the standard preparation check or working laboratory fortified blank (Section
23.1).
13.7 Run analysis assuring suitability parameters in Section 23.2 are met.
13.8 Prepare standard curve by plotting instrument response against concentration
values.
13.9 A calibration curve may be fitted to the response data using computer based
regression curve fitting techniques.
13.10 Calculate the concentration of each calibration point. The percent of recovery of
each standard should be within 90 to 110% of it true value. If the difference between
the measure values of the calibration solution exceed ±10% the true value
concentration, the analysis should be terminated and the instrument recalibrated.
14.0 ROCEDURE
14.1 Standardization of Cyanide Stock Standard Solution 1000 mg/L (Section
10.4.2)
14.1.1 Transfer 20 mL of Cyanide Stock Standard Solution 1000 mg/L (Section
10.5.1) to a beaker.
14.1.2 Dilute to 100 mL with Sodium Hydroxide dilution solution (Section 10.4.3).
14.1.3 Add 0.5 mL Indicator Solution (Section 10.4.1)
14.1.4 Titrate with standard AgNO3 (10.4.2) titrate to the first change in color from
canary yellow to a salmon hue.
14.1.5 Record volume of titrant used.
14.1.6 Titrate a blank, NaOH dilution solution.
14.1.7 Record volume of titrate used.
14.1.8 Calculate mg CN-/L of Cyanide Stock Standard Solution (Section 15.5)
14.2 Microdistillation Procedure
14.2.1 Transfer 50mL of sample or standard into the MicroBlock Reaction Flask
(Section 23.4).
14.2.2 Set distillation system. Refer to the schematic in Section for the appropriate
placement positions and flexible tubing connections to cool the Cold Finger
Page 11 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
with running tap water. All air/vacuum valve are in OFF position.(Section
23.3)
14.2.3 Add a pinch of boiling chips. For solid sample weigh 1.0g or less to the
nearest 0.01g, dilute to 50.0mL and add boiling chip.
14.2.4 Pipette 50.0mL of 0.25N NaOH into the absorption Flask (Section 23.4).
14.2.5 Position the Cold Finger and Absorber Impingers.
14.2.6 Turn tap water ON after assuring all tubing connections are firmly placed.
A water flow rate of 50 to 60 gallons per hour is required.
14.2.7 Turn ON the vacuum and slowly adjust each valve to provide an air flow to
maintain a bubble rate of about several bubbles per second for each position
as viewed in the absorption vessel.
14.2.8 Add 5mL of sulfamic acid reagent solution (Section 10.2.5) to eliminate
Nitrite & Nitrate interference.
14.2.9 Slowly add 5mL sulfuric acid 18N (Section 10.2.3) through the reagent
inlet. Allow the air flow to mix the contents of the flask for several minutes.
14.2.10 Add 2.0mL of magnesium chloride (Section 10.2.4) reagent and wash
down with a stream of water through the reagent inlet.
14.2.11 If excess foaming occurs, add an additional 2.0mL magnesium chloride.
14.2.12 Allow a few minutes to mix.
14.2.13 Turn the heat ON and set the temperature of the heating block to 150˚C.
14.2.14 Wait until the system reach the mentioned temperature, then turn off the
heater block in an hour.
14.2.15 Let the boiling flask cool.
14.2.16 Turn air/vacuum valves to the OFF position.
14.2.17 Disconnect the Distillation Head and Absorption Impinger by separating
the Slip Connector.
14.2.18 Repeat steps 14.1.18 to for each sample position.
14.2.19 Allow the tap water to continue circulating through the Cold Fingers which
will allow the heater block more rapidly stabilize in preparation for the
next run.
14.2.20 The distilled cyanide in the 0.25N NaOH is now ready to be analyses.
14.3 Macro Distillation Procedure 14.3.1 Add 500 mL sample, containing not more than 10mg CN-/L (diluted if
necessary with reagent water) to the boiling flask.
14.3.2 Add 10 mL Sodium Hydroxide 1N (10.2.1) to gas scrubber and dilute with
reagent water to obtain an adequate liquid depth in the absorber. Do not use
more than 225 mL total volume of absorber solution.
14.3.3 When sulfide generation from the distilling flask is anticipated add 50 or
more mg powdered lead carbonate (PbCO3) to the absorber solution to
precipitate sulfide.
14.3.4 Connect the train, consisting of boiling flask air inlet, flask, condenser, gas
washer, suction flask trap, and aspirator. Adjust suction so that least 1 air
bubble/s enters the boiling flask. This air rate will carry HCN gas from flask
to absorber and usually will prevent a reverse flow to HCN through the air
inlet. If this air rate does not prevent sample backup in the delivery tube,
Page 12 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
increase air flow rate to 2 air bubbles. Observe air purge rate in the absorber
where the liquid level should be raised not more than 6.5 to 10 mm.
Maintain air flow throughout the reaction.
14.3.5 Add 2 g sulfamic acid through the air inlet tube and wash down with reagent
water.
14.3.6 Add 50 mL Sulfuric Acid 1+1 (10.2.4) through the air inlet tube. Rinse tube
with reagent water and let air mix flask content for 3 min.
14.3.7 Add 20 mL Magnesium Chloride reagent through air inlet and wash down
with stream of water. Precipitate that may form redisolves on heating.
14.3.8 Heat with rapid boiling, but not flood condenser inlet or permit vapor to rise
more than halfway into condenser. Adequate refluxing is indicated by a
reflux rate of 40 to 50 drops/min from the condenser lip.
14.3.9 Reflux for at least 1 hour. Discontinue heating but continue air flow for
15min.
14.3.10Cool and quantitatively transfer absorption solution to 250 mL volumetric
flask. Rinse absorber and its connecting tubing sparingly with reagent water
and add to flask.
14.3.11Dilute to volume with reagent water and mix thoroughly.
Note: When following the macro distillation procedure, the cyanide ion
present in the original sample is concentrated by a factor of two from
distillation procedure. The initial sample volume is 500mL; however, the
cyanide which has distilled over into 0.25M NaOH absorbing solution is
only diluted to a 250mL final volume. Because the nondistilled standards
are not carried through the distillation procedure, they are not
concentrated by factor or two. To compensate for this factor, the standard
concentration are doubled.
14.4 Colorimeter Analysis by Automated Continuous Flow Analyzer 14.4.1 Place the Cyanide manifold (Section 23.6 & Section 23.7) on channel 1 of
the instrument. If this channel isn’t working properly the channel 2 can be
used but make sure to turn on the channel in the software. Verify that the
run sequence is the same one as Section 23.1 and verify the condition of the
instrument.
14.4.2 Turn on the LACHAT (the switch is placed in the right side of the
equipment)
14.4.3 Go to the icon Omnion
14.4.4 Open Alchem methods search Cyanide year month
open a run worksheet.
14.4.5 As soon the worksheet appear a window will be shown asking “Do you want
to change the set points of the relevant heater?” Select YES
14.4.6 Line/Pump Startup
14.4.6.1 Go to configuration instrument pumps
normal function close
14.4.6.2 Pull line to ends of cartridge to ensure proper tension on the line.
Clamp the cartridge to the pump. Clamp pump cartridge down
Page 13 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
only when pump on and rolling. Snap cartridge in one end and
then the other.
14.4.6.3 Set tension on line by clicking the tabs on top of the cartridge
back on click.
14.4.6.4 Keep pump speed on 35.
14.4.6.5 Tubes can pop off the cartridge. If that occurs slow the pump
down and place tube in fluid to be drawn through.
14.4.6.6 PVC (clear) and Duraprene (tan color) pump tube can be used
can be used for this method
14.4.6.7 Place all tube ends in reagent water and run the reagent water
through all the line for at least 5 minutes to check for leaks and
flush potential contaminants.
14.4.7 The method requires a heating unit. Make sure the heater is turn at 60°C and
it has reached at that temperature.
14.4.8 Check that the wash bath line for the auto samples valves is in reagent water.
14.4.9 Load rack with the run sequence (Section 23.1)
14.4.10Switch reagent tube lines from the reagent water to their proper reagent
bottles and allow the system to equilibrate until stable baseline is achieved.
14.4.11 Select PREVIEW and wait for a stable baseline.
14.4.12 Select the STOP function.
14.4.13 Begin the analysis by selecting the START function.
Note: If sample concentration are greater than the high standard, the
distilled sample should be diluted with 0.25M Sodium Hydroxide (NaOH)
diluent. Do not dilute samples or standard with reagent water.
15.0 DATA ANALYSIS AND CALCULATION
15.1 Prepare a calibration curve by the peak area of distilled standard against
concentration values. The data system will then prepare a calibration curve by plot
15.2 Compute sample concentration by comparing sample response with the standard
curve. Result obtained by the instrument is in mg/L. Multiply answer by appropriate
dilution factor. Result can be reported in mg/L and µg/L.
15.3 Report only those values that fall between the lowest and the highest calibration
standards
15.4 Calculations: (For manual demonstration only. QuickChem Software will perform
these calculations automatically)
15.4.1 Spike % Recovery
%R = Cs – C x 100
S
where: R = Percent Recovery
Cs = Average result of analysis of spiked sample and duplicate
C = Average result of the sample without spike
s = Concentration of spike added to sample
15.4.2 Relative % Difference (RPD) Between Spikes:
Page 14 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
RPD = |(Spike Result – Spike Duplicate Result) | x 100
(Spike Result + Spike Duplicate Result)/2
15.5 Calculation of mg CN-/L of Cyanide Standard Stock Solution
Mg/L CN-/L = (A-B) x 1000 X ____ 250______
mL original sample mL portion used
A = mL AgNO3 for sample; B = mL AgNO3 for blank
15.6 For manual integrations (if needed to be performed) refer to SOP ST 001
16.0 METHOD PERFORMANCE
16.1 Refer to laboratory Method Quality Performance Data file.
17.0 POLLUTION PREVENTION
17.1 Pollution Prevention should be the management first option.
17.2 When wastes cannot he feasibly reduced at the source recycling is the next best
option.
18.0 DATA ASSESMENT AND ACEPTANCE CRISTERIA FOR QUALITY CONTROL
MEASURES
18.1 See Procedure ALQAP DW 006-Statistical Control Procedure, Section 23.2 –Table
2- Acceptances Criteria
19.0 CORRECTIVE ACTIONS FOR OUT-OF-CONTROL DATA
19.1 See Procedure ALQAP DW 008- Corrective and Preventive Actions.
20.0 CONTIGENCIES FOR HANDLING OUT –OF-CONTROL OR UNACCEPTABLE
DATA
20.1 See Procedures ALQAP DW 006 – Statistical Control Procedure, ALGAP DW 002
– Procedure of Non-Conforming Results, ALQAP DW 008 – Corrective and
Preventive Action Procedure
21.0 WASTE MANAGEMENT
21.1 Excess reagent, samples and method process wastes should be characterized and
disposed in an acceptable manner.
21.2 See Procedures ALGAP DW 004 – Laboratory Waste Disposal, ALGAP DW 012
– Sample Storage and Disposition.
22.0 REFERENCES
22.1 Lachat Quick Chem 8500 Autoanalyzer Operations Manual
22.2 Quikchem Method 10-204-00-1-A
22.3 USEPA Method 335.4
22.4 Standard Method for the Examination of Water and Wastewater 4500-CN- (22nd
Edition)
Page 15 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
22.5 Environmental Express MicroBlock Distillation System Operation and Instruction
Manual
22.6 ISO 17025 Quality Standard
22.7 Puerto Rico Dept. of Health Regulation for Drinking Water – Manual for the -
Certification of Laboratories Analyzing Drinking Water PRDOH-016-8-90
23.0 TABLES, DIAGRAMS, FLOWCHARTS
23.1 Table 1 - Run Sequence
Sample or Standard Concentration (mg/L)
Calibration Blank 0.00
Calibration Standard Solution A (Lowest Concentration) 0.300
Calibration Standard Solution B 0.100
Calibration Standard Solution C 0.0500
Calibration Standard Solution D 0.0300
Calibration Standard Solution E 0.0100
Calibration Standard Solution F (Highest Concentration) 0.0500
Calibration Blank 0.00
Quality Control Sample 0.0500
Initial Instrument Performance Check 0.0300
Initial Calibration Blank <MDL
Instrument Blank <MDL
Reporting Limit Verification (MRLV) 0.0500
Laboratory Reagent Blank <MDL
Laboratory Reagent Blank Duplicate <MDL
Laboratory Fortified Blank 0.0300
Laboratory Fortified Blank Duplicate 0.0300
On-going Precision & Recovery 0.0500
Method Blank <MDL
Continuance Instrument Performance Check 0.0300
Continuance Calibration Blank <MDL
Sample 1 Unknown
Sample 2 Unknown
Sample 3 Unknown
Sample 4 Unknown
Sample 5 Unknown
Sample 6 Unknown
Sample 7 Unknown
Sample 8 Unknown
Sample 9 Unknown
Page 16 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
Sample or Standard Concentration (mg/L)
Continuance Instrument Performance Check 0.0300
Continuance Calibration Blank <MDL
Sample 10 Unknown
Sample 10 Duplicate Unknown
Sample 10 Spike Unknown + Known Standard
Concentration
Sample 10 Spike Duplicate Unknown + Known Standard
Concentration
Calibration Verification Standard 0.0300
Calibration Verification Blank <MDL
Note: After quality control sample for every 10 sample an instrument performance check
and calibration blank should be performed
23.2 Table 2 – Acceptance Criteria
Parameter Frequency Acceptance Criteria
Method Blank or Laboratory
Reagent Blank (LRB)
1 Method Blank per set of
20 samples or less Must be below MDL value
Laboratory Fortified Blank
(LFB)
Performed 1 with Duplicate
with each batch of sample Recovery % 90.0-110.0%
Standard Calibration Curve
Performed with every
analysis for sample
calculations
Correlation Coefficient
> 0.995. Percent of Recovery of each
standard must be within 90-110% of
it true value.
Calibration Verification
Standard (IPC)
Performed with every
analysis for calibration
preparation verification
Recovery % 90.0-110.0%
Instrument Performance
Check Solution (IPC)
Mid-Range Check Standard
analyses with calibration
blank, immediately
following daily calibration
and after every tenth
sample.
Recovery % 90.0-110.0%
Laboratory Fortified Sample
Matrix (LFM) OR Sample
Spiking
Performed in duplicate on
10% of samples to be
analyzed with every
analysis.
Relative Percent Difference Between
Spikes &
Percent of Recovery of Spike - Refer
to Laboratory Control Limit. Relative
Percent Difference Between Sample
Must be ≤ 20%
Page 17 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
23.3 Figure 1 – MicroBlock Equipment Wiring Connection
Parameter Frequency Acceptance Criteria
Method Detection Level
(MDL)
Initial Determination,
verified every six months,
new analyst performing the
method and/or when a
significant change in
instrument response is
detected
Average of the seven aliquots cannot
exceed 10 time the MDL value.
Linear Calibration Range
(Dynamic Range)
Initial Determination for
method and verified every
six months
Linearity Correlation Coefficient >
0.995
Standards Recovery 90.0-110.0%
Quality Control Standard
(QCS)
(External Source)
Initial Determination for
method and verified every
six months
Recovery 90.0-110.0%
Page 18 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
23.4 Figure 2 – MicroBlock Glassware
Page 19 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
23.5 Figure 3 – Macro Distillation System
Page 20 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
23.6 Figure 4 – LACHAT Cyanide Manifold Setup
Carrier: 0.25M Sodium Hydroxide Solution (Section 10.2.2)
Manifold Tubing: 0.80 mm (0.032 in) i.d. This is 5.2µL/cm
Sample Loop: 50 cm x 0.8mm
Interference Filter: 570nm
Apparatus: An injection valve, a 10mm path length flow cell, and a
colorimeteric detector module are required. The
shows 650cm of tubing wrapped around the heater block at
the specified temperature.
4.5: 70cm of tubing on a 4.5cm coil supoort.
Note 1: 200cm backpressure loop, 0.52mm
Page 21 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
23.7 Figure 5 – Valve Diagrams
23.8 Appendice 1- Cyanide LACHAT Data System Parameters
Sample throughput: 90 sample/h, 40 s/sample
Pump Speed: 35
Cycle Period: 40
Analyte Data:
Concentration units: mg CN-/L
Chemistry: Direct
Calibration fit Type: 1st Order Polynomial
Weighting Method: 1/X
Force through zero: No
Page 22 of 22 SABANETAS INDUSTRIAL PARK · BLD. M-1380 · PONCE, PR 00731
P.O. BOX 359 · MERCEDITA, PR 00715-0359 TELS. (787) 835-4242 · (787) 848-6050 · FAX (787) 848-6299 · (787) 835-2004
www.altolenterprises.com
TITLE: Determination of Total Cyanide by Semi- Automated Colorimetry
REVISION: C
PROCEDURE: SOP DW 002 EFFECTIVE DATE: September 19, 2016
24.0 CHANGE HISTORY
10/04/13 REV. A First Issue
09/04/14 REV. B Correction of all quality manual procedure id used as
reference.
09/19/16 REV. C 1. Add Section 6 (Definition) the definition Initial
Demonstration of Capability, and Minimum Reporting Limit
2. Improved detail related to Initial Demonstration of
Capability in section 12.1.2.
3. Improved detail to each section of Laboratory Performance
(Section 12.2). Also added method requirement for
Minimum Reporting Limit (Section 12.2.5) and Quality
Control Sample (Section 12.2.6).
4. Added to section 13.0 (Calibration & Standardization)
acceptable criteria for calibration curve standard (Section
13.10)
5. Added to Table 1 – Run Sequence (Section 23.1) Reporting
Limit Verification standard.
6. Added to Table 2 – Acceptance Criteria (Section 23.2)
Instrument Performance Check Solution (IPC) and Quality
Control Standard requirement.
7. Added and detailed procedure for the uses of Macro
Distillation System (Section 14.3).