QUAL-QA-TEMP-0220_rev15 Part Number: RAN429G Date: 22/9/14 ECR: 5567 ABX Pentra 60C+ / Pentra ES60 Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 Tn TN
Transcript
Page 1: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

QUAL

-QA-

TEM

P-02

20_r

ev15

Part Number: RAN429G

Date: 22/9/14ECR: 5567

ABX Pentra 60C+ / Pentra ES60Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2

Tn

TN

Page 2: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G

Objective Software version update for P60C+/PES60 with or without takasago valves

Intervention

informationIntervention priority For information - As needed - Mandatory on next visit - Immediate and mandatory

Tracking number TN-RAN429G Intervention estimated duration: 30 min

User process impact YES NO Potential patient result impact YES NO

User training needed YES NO User documentation update YES NO

Applied in factory

on instrument serial #P60C+: PCP13129, 13152PES60: PES12976, 12980, 12985, 12990, 12994, 13038

since instrument serial #P60C+: PCP13156PES60: PES13075

Note history Note index Reason for change

C to D Release of version V2.8.1 (ECR4284)

D to E Release of version V3.0.0 (ECR5180)

E to F Warning regarding the installation on Windows 7 for the Chinese language pack (ECR5437)

F to G V3.0.0 to V3.0.1 (ECR5567)

Software

version

Note index Previous New Mini. requirements

A to B

V2.5.0 PES60V2.4.1 P60C+

V2.7.0EPROM V2.3.0P60C+ version > v2.2.9

B to C canceled canceled

C to D V2.7.0 V2.8.1 EPROM V2.3.3 + Takasago Valves

D to E V2.8.1 V3.0.0 EPROM V3.0.0

E to F V3.0.0 V3.0.0 EPROM V3.0.0

F to G V3.0.0 V3.0.1 EPROM V3.0.0

New parts Parts designation Previous New Reason for change

KIT V3.0.1 P60C+/ES60

XEC130BS XEC130CS Software release

Motherboard P60 OT/C+/ES TAKA

- XAA722AS Software release

Motherboard P60 OT/C+/ES NON TAKA

XAA472DS XAA472ES Software release

New

adjustments

Adjustments Previous New Reason for change

2/8

Page 3: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G

New

documents

Documentation Previous New Reason for change

Documentation CD-ROM

PES60 : RAX071EP60C+ : RAX059F

PES60 : RAX071FP60C+ : RAX059G

V3.0.1 software release

Additionnal

requirements

Designation Part number Link

Part details kit or Assy DesignationComponent p/n

Component designation Qtyavailable for order

XEC130CS KIT V3.0.1 P60C+/ES60 Yes

EPROM V3.0.0-1 1 No

EPROM V3.0.0-2 VTK 1 No

NAJ128BP60 RANGE WORKSTATION V3.0.1 CD-ROM

1 No

RAN429GV3.0.1 PES60/P60C+ technical note

1 No

RAX071F Documentation CD-ROM PES60 1 Yes

RAX059G Documentation CD-ROM P60C+ 1 Yes

3/8

Page 4: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G

1. Modifications and evolutionsThis new software version operates under Windows XP and Windows 7. It no longer operates under Windows NT.

1.1. Russian interface

The ABX Pentra 60C+ and the Pentra ES60 interface is now available in Russian.

1.2. Correction of the PCT Unit

The unit for PCT has been changed for SI and mmol/L. It has been changed from “10-2 /L” to “10-2 L/L”.

1.3. Optical Gain Tolerance Change

Following recent findings, the optical gain frame has been increased. It still allows a minimum tension of 5.5 V, but itcan now allow a maximum tension of 7.5 V.Refer to RAN589A for more information on this adjustment.

2. EPROM installation■ Switch the instrument off and remove the left-hand side panel.

■ Using an antistatic mat, replace the instrument EPROM (refer to the diagram on the following page).Replace the EPROM by the correct one, depending on whether the instrument is equipped with Takasago valves ornot. The EPROM V3.0.0-1 must be used for instruments not equipped with Takasago valves, and the EPROM V3.0.0-2 VTK must be used for instruments equipped with Takasago valves.

■ Re-install the left-hand side panel.

Without Takasago valves With Takasago valves

4/8

Page 5: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G

Without Takasago valves With Takasago valvesV3.0.0-1 V3.0.0-2 - VTK

U43 EVEN U42 ODD

5/8

Page 6: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G

3. V3.0.1 Software installation

■ Create a new worklist before installing the new version in order to have an archive file.

■ Switch the instrument and the workstation off.

■ Switch the instrument and the workstation on and log in as the administrator.

■ Insert the CD-ROM NAJ128B in the computer CD-ROM drive.

■ Start the Windows Explorer application and click the CD-ROM drive device letter.

■ Run "setup.exe" from the CD-ROM root. The following window is displayed:

■ Click "Yes" to continue the installation. The following window is displayed:

■ Click "Next".

■ Select the correct instrument from the scrolling list.

Version V3.0.1 must be installed over version 2.2.9 (or higher) except for the computers Fujitsu Siemens E620 and E5615, which operate under Windows XP. For Fujitsu Siemens E620 and E5615, version 3.0.1 can be installed directly or over version 2.2.9 (or higher).Refer to the latest version of RAN056 to check the compatibility between the software and the computer.

The installation program installs all the files needed to carry out the implementation of the software in the case of a first installation (case of the installation on the Fujitsu Siemens E620 and E5615). In other cases it updates the software but keeps the customer settings and data.

The installation of software version V3.0.1 on Windows 7 is not compatible and not functional for the Chinese language pack.EPSON 6100, EPSON EPL-59000 and Lexmark E210 printers are not compatible with Windows 7.

6/8

Page 7: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G

■ Click "Next".A status bar is displayed during the copy:

◆ The instrument proceeds to the software version installation. At the end of the installation the following window is displayed:

◆ Click "Finish" to restart the computer.◆ Open a session with the "ABX" login.◆ Check the instrument and user setup.◆ Control the correct operation of the instrument.

7/8

Page 8: Tn TN - its-information.horiba-abx.comits-information.horiba-abx.com/_basedoc/hematology/abx_pentra_60_c... · Tn TN. Kit Version 3.0.1 ... Using an antistatic mat, replace the instrument

Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G

8/8