UNCLASSIFIED AD NUMBER AD494667 NEW LIMITATION CHANGE TO Approved for public release, distribution unlimited FROM Distribution authorized to U.S. Gov't. agencies and their contractors; Critical Technology; OCT 1945. Other requests shall be referred to Commanding Officer, U.S. Army Aberdeen Research and Development Center, Aberdeen Proving Ground, MD. AUTHORITY AMSRL-CS-IO-SC [380] ltr dtd 1 Nov 2001 THIS PAGE IS UNCLASSIFIED
Transcript
Page 1: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

UNCLASSIFIED

AD NUMBER

AD494667

NEW LIMITATION CHANGE

TOApproved for public release, distributionunlimited

FROMDistribution authorized to U.S. Gov't.agencies and their contractors; CriticalTechnology; OCT 1945. Other requests shallbe referred to Commanding Officer, U.S.Army Aberdeen Research and DevelopmentCenter, Aberdeen Proving Ground, MD.

AUTHORITY

AMSRL-CS-IO-SC [380] ltr dtd 1 Nov 2001

THIS PAGE IS UNCLASSIFIED

Page 2: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

-JAD

MEMORANDUM REPORT NO. 397

AN IVPROVED TOURMALINE AIR BLAST GAGE

by

Thomas D. Carr D D CMary A. Bakinowskl '

OCT2 519L68 fj

October 1945 B

This document is subject to special export controls and each transmittalto foreign governments or foreign nationals may be made only with priorapproval of Commanding Officer, U.S. Army Aberdeen Researcn and DevelopmentCenter, Aberdeen Proving Ground, Maryland.

U.S. ARMY ABERDEEN RESEARCH AND DEVELOPMENT CENTERBALLISTIC RESEARCH LABORATORYABERDEEN PROVING GROUND, MARYLAND

Page 3: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

BALLISTIC RESEARCH LABORATORY

MNJORA LU REPORT NO. 397

Ordnance Research and Develop-

ment Center Project No. 5336

Carr/Bakino-ski/rwAberdeen Proving Ground, Md.17 October 1945

AN DIPROVED TOU0M'1LII E AIR BLAST GAGE

Abstract

In order to meet the requirements of a program formeasuring blast pressures of large caliber bombs atthe Aberdeen Proving Ground a piezoelectric tournalinecrystal blast gage has been developed and methods ofproducing it in large quantities have been worked out.Successive calibrations taken on individual gages shova standard deviation from the mean rarely exceeding 1%;and an output that shows no significan departure from

linearity in the pressure range from three to sixteenpounds per square inch. Details of construction and

calibration are given.

Page 4: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

AN IMPROVED TOURMALINE AIR BLAST GAGE

I. INTRODUCTION

A. The Piezoelectric Blast Gage. Piezoelectric gages areused extensively in the measurement of the time variation of the airblast pressure from bombs and guns. Changes in the pressure appliedto a gage produce proportional amounts of electric charge which chargesup the capacity of the gage and connecting cable. The resultingvoltage is amplified and recorded photographically as a function oftime by means of a cathode ray oscillograph. If the oscillographdeflection with gage and cable attached is recorded for a known chargeand the calibration constant for the gage (i.e. charge per unit pres-sure) is known from previous measurement, the gage record can readilybe calibrated in units of pressure.

B. Types of Piezoelectric Crystal. The crystals most widelyused in piezoelectric gages are quartz, tourmaline, Rochelle salt, andammonium dihydrogen phosphate. Of these crystals quartz and tourmalinehave proven most suitable for quantitative blast measurement. Quartzgages have the disadvantage that they must be constructed with a pistonor diaphragm for directing the force to particular surfaces of the cry-stals, and other faces must be protected from the pressure changes.Unlike quartz, tourmaline is sensitive to hydrostatic pressure. Thusthe mechanical design of a gage can be much simpler if tourmaline isused instead of quartz because no force-directing structure is necessary.A disadvantage of tourmaline is that it exhibits a pronounced pyro-electric sensitivity, whereas quartz does not. Despite the pyroelec-tric effect, however, the piezoelectric constant of tourmaline is in-dependent of temperature. If the crystals in tourmaline gages arecoated with a layer of thermal insulation, the pyroelectric sensi-tivity is not objectionable in the measurement of blast. In the de-velopment of a satisfactory gage at Ballistic Research Laboratory,Aberdeen ?r,;iing Ground, Md., it was found that the difficulties oftourmaline were more easily overcome than those of quartz.

C. The INDRC - Stanolind Tourmaline Gage. A tourmaline gagefor air blast measurement was developed by Division 2, NDRC and(yIsmanufactured in improved form by the Stanolind Gas and Oil Co. 'Itconsists of a stack of four to eight tourmaline discs soldered to-gether, with electrodes between the crystals. A process was developedfor soldering directly to the crystal surface. The stack is surroundedfor a plastic heat insulating material and a flexible electrostaticshield. A lead of relatively non-microphonic shielded cable is con-nected with the crystal electrodes which are in parallel electrically.

(1) See 0R.D Report Ro. 76h9 "Development of BIE.st Gages & RecordingEquipment" - T.I.B. No. 754 9a.

-1- __

Page 5: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

The Stanolind gages were used by Ballistic Research Laboratory intesting bombs for about a year and very good results were obtainedat first. However, the gages were fragile and when used constantlywould deteriorate rapidly. Furthermore, there were often gages forwhich the calibration varied. These gages were almost alw foundto have loose electrodes, due to defects in the soldering."' Finally,a sufficient number of gages to satisfy the needs of Ballistic Re-search Laboratory could not be obtained from Stanolind. Due to thesedifficulties, it was decided to attempt the development at BRL of arugged and relatively simple tourmaline gage.

D. The BRL Tourmaline Gage. After a long period of de-velopment, a satisfactoiy tourmaline gage was produced at BRL. Itssensitivity is approximately 75pg C/psi. The gage combines rugged-ness with constancy of calibration, and can easily be constructed inquantity. A large number of these gages have been in almost con-tinous use for the past six months in testing bombs. With the ex-ception of those hit by fragments, veryfew have developed any fault.Blast records obtained with them are of good quality and show noapparent gage frequencies.

II. GAGE CONSTRUCTION

A. General Description. Figure 1. is a photograph of theBRL tpurmaline gage at several stEEges of construction. It containsfour M2 tourmaline plates with five silier foil electrodes, platesand foils alternating. The plates and foils in the stack are cementedtogether with a thermosetting cement, monomeric butyl methacrylate(MBM). The stack is cemented to a steel block. A brass ring of thesame outside diameter as the block and slightly higher than the stackis bolted to the same face of the block to which the stack is cemented.Thus the stack of crystals is fastened to the bottom of a cup-likemetal housing.

The crystals are so arranged that the sides which arepositive -,ith compression are in the directions indicated in Figure2a. Electrical connection is made to the foils by means of project-ing tabs. The positive tabs project from the stack in one directionand the negative in another. The plates are circular except on oneside, where the boundary is a chord of the circle. This flattenedside (see Figure 2b) allows space for making the tab connections. The

negative tabs vhich include the top and bottom ones of the entirestack are fastened to the steel block.

The cable is British Telconax. It is especially made formeasuring blast with piezoelectric gages, and is practically free ofthe signal resulting from mechanical agitation which is encounteredin other types. A length of about three feet is attached to the gage.

(2) -t is uderstood that recent improvements have been.ade wvhich ayhave eliminated these difficulties.

(3) Another model has eight plates, sensitivity 320 W C/psi.

-2-

Page 6: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

It enters the gage housing through a brass sleeve in the side. The

shield at the end is unbraided for a shoi't distance and anchored under

a notch in the ring and soldered to the outside of the ring, as shown

in Figure 3. The central conductor is soldered to the positive tabs.

The material used for heat insulation is General

Electric Cable Joint Compound No. 227. It is a high melting point

wax similar to pitch in appearance. The wax is melted and pouredover the stack until its surface is level with the top of the ring,

It clings very tightly to the crystal and metal surfaces. The

thickness of the ring is such that the layer of wax covering thestack is about 116 inch. The wax is relatively plastic incomparison with the crystal. Thus it can transmit hydrostatic pressureto the crystal. Evidence supporting this was obtained by a simple

test. A gage which was calibrated with several different thicknesses

of wax always gave the same calibration constant.

A disc of copper foil is attached to the top of thering, in contact with the wax, to complete the electrostatic

shielding of the stack.

The leakage resistance of the MBM cement is veryhigh. Thus it does not matter if cement adheres to the edges ofthe crystals, bridging the gap between successive foils. A testwas made to determine whether or not cement between the crystalscontributes to the piezoelectric output. A dummy gage was madeof glass plates instead of crystal and was calibrated. It gaveno charge, indicating that the cement is inactive. It is believed

that in this type of gage soldering crystals together has no ad.-vantage over cementing.

Detailed construction information is presented in

the appendix.

III. CALIBRJiTION RESULTS

A. Method of Calibration. Most of the calibrating hasbeen done oscillogiaphically. From one to four gages are calibra-ted at a time. They are placed in a small pressure tank with acellophane window. The air pressure in the tank is raised, and itsvalue read with a Bourdon gage. The pressure is then released verysuddenly by puncturing the cellophane window. A multi-trace os-cillograph record taken at the time of release gives the deflectionfor each gage. The charge liberated by each gage is determined from

the ratio of this deflection to that produced by a standard charge. (4)The calibration constant is the ratio of the initial tank pressure(excess above atmospheric) to the liberated charge. Pressure oscilla-tions persist within the tank until about five to eight millisecondsafter release. The deflection is measured at a point on the recordten milliseconds after release, since within ten milliseconds alloscill.ation is completely damped.

(T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to

that descril'ed in OSRD Report No. 4619.

-3-

Page 7: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

B. Consistency of Calibrations. Nine successive cali-bration records are taken for each gage constructed, all at approxi-mately 20 psi. The standard deviation of individual measurementsfrom the mean for the nine corresponding values of calibrationconstant is almost "always less than 1%. Definite trends in aseries of values never appear, although with the Stanolind gagesthis often occurred. Mhen a gage which has been used for severalmonths is recalibrated, its constant almost always agrees with theoriginally determined value to within 2%.

C. Linearity. No evidence of nonlinearity has ever beenfound. Table I gives mean values of a calibration constant measuredat several defferent pressures, with the standard deviation of themean for each value. There are no significant differences.

TABLE I

Gage 44

Pressure Sensitivitypsi. p C/psi

3 74.o + 1.5

6 74.7 + 0.3

10 74.8 + 0.2

16 74.7 + 0,07

D. Teitq)erature Dependence. Gages were calibrated at twowidely separated temperatures. The observed differences in calibra-tion constant are not significant. The data is given in Table II.

TABLE II

Gage Sensitivity Sensitivityat 25 0 C. at 40 C.vp C/psi P11 C/psi

T42 78.8 + 0.3 77.7 + 0.3

T68 63.8 + 0. 4 65.4 + 0.5

Thomas D. Carr, Physicist

Ka-y A. Bakinowski, T/4Laboratory Technician

-4-

Page 8: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

APPENJDIX I

Details of Gage Const11ruction

Page 9: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

APPENDIX I

Details of Gage Construction

1. Preparation of Crystal Plates

Cutting: Tournialine crystal plates were cut from the follow-ing specifications: 1) Crystal stock to be supplied by the Government.Plates to be cut perpendicular to crystal axis for maximum piezoelectricresponse to compression 0.060 inch plus or minus 0.02 inch thick andground flat plus or minus 0.0001 inch, uniform in thickness, naturalsize and shape. Finish fine ground.

The cutting to gage size is done in the laboratory as follows:

A "biscuit cutter" is constructed of 1-3/4" I.D.copper tubing and W 2" wall thickness for usein a drill press. Four notches are filed ingrinding edge of the biscuit cutter with abruptleading side and sloping trailing side.

Silicon Carbide Abrasive Grain is used as cutting agent. Mix2/5 of size 220 with 3/5 of size 100; add water until of loose pasteconsistency.

Use melted wax to attach crystal to a steel plate for cutting.Cool wax thoroughly before cutting. Place plate on drill press tableand fix biscuit cutter in drill chuck. Adjust plate position untilbiscuit cutter projects slightly beyond the edge of the crystal platein order to cut out a piece of crystal of approximate dimensions asshown in Figure 2.

Clamp plate to drill press table. Attech weights and springto handle of drill press in such a way that con;.ant pressure will beapplied in cutting. Start drill press, using medium speed. Applycompound to crystal, bring cutter into contact .ith crystal. As cut-ting progresses, add compound and water when necessary. Cutting pro-cess takes from 3 to 5 minutes.

To remove cut crystal from plate, heat back of plate untilwax is melted. Remove with spatula. Do not force. Clean excesswax off plate and repeat the process for next crystal.

Cleaning Crystals. Rermove excess wax and compound with razorblade. Clean with carbon tetrachloride and cotton. Clean thoroughly.

(1) 7nis was done for VH1L by Reeves Sound Laboratory, New York.(2) I.:ethod is described In Strong, "Procedures in E]xperimental Physics"

-6-

Page 10: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

Test-.g for Polarity. A cathode ray oscilloscope is used to

determine crystal polarity. Lay crystal on metal block which is con-nected to oscilloscope ground. On crystal, place electrode connectedby shielded cable to input. Lay a piece of thin sheet bakelite onelectrode and tap with pencil eraser. Deflection is upward or downward,depending on whether top side of crystal is positive or negative.Mark negative side with pencil.

2. Preparation of M. B. M.

Partially polymerized Monomeric Butyl Methacrylate (3) is used

as a cement for securing stacks of crystals and foils together and tothe base of the gage.

Precaution! M.B.M. must be kept in a dry ice refrigeratorwhen not in use. Dissolve 3 or 4 grains of benzoyl peroxide catalystin a 25 c.c. beaker of M.B.M.

Lay a metal sheet over gas burner and use a moderate gas

flame for cooking. Lay a piece of paper toweling over metal sheet.

Pour mixture into glass cooking vessel and put over flame.

Stir constantly with thermometer, watching temperature rise.

Material can be heated to about 1500 C. initially, and thencooled to 1000 C. for partial polymerization.

Once the polymerization reaction has started, it is exothermicand must be watched. To keep under control it is necessary to lowerthe flame just as the temperature reaches 1500 C. In the case of per-sistent rise in temperature, turn out flame or :emove from stove at1500 C.

Keep stirring until temperature comes dov:n to 1000 C. Pourinto glass container. Preparation is ready to uzc. 25 c.c. of thispreparation is enough for approxiimately 25 gagez of 4 crystals each.

M.B.M is good for about one week when l.ft standing.

3. Assembly of Gage

Preparation of Gage Base

The metal parts of the gage are nade accordinC to Figure 3.Clean Eare ba-e of oil and frit. Screw tinned -oiderinG lug to basefor negative term'[hal.

(3) Furrchascd from, Rolm-n and .ass, Pila13elphia, Pa.

I-

-. 7-

Page 11: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

I

Preparation of Crystals and Foils

Electrodes for crystals are made of 0.001" silver foil.Select 4 crystal plates and cut foil to match each plate. Leavetab projecting from straight side. Cut tab at such an angle thatit will fit over its proper terminal. (See Figure 2 b) Clean foilswith carbon tetrachloride.

Steps in Cementing Stack and Baking

Mark gage base by setting brass ring in its proper positionand going around the inside of it with a scribe. This mark servesas a guide for placing the stack. When stack is centered, there shouldbe a space of about 1/8" between edge of stack and the mark.

A small paint brush is best for applying the cement (M.B.14).

Paint negative foil and place directly on the gage base. Ad-just it so that tab will. lie over negative terminal and that negativeside of the crystal plate will fit over it.

Immediately paint negative side of the first crystal andplace it over the silver foil. Apply pressure with the fingers, be-ginning in the center of plate and moving outword to the edges toforce out the air bubbles.

Top of crystal is positive. Paint top and lay positivematching silver foil over it with tab projecting from side oppositeto the negative one. Use pressure as before. It is preferable thatpressure is applied over cheesecloth or a rietal block in order to keepfinger prints and perspiration off parts. Lay the painted positiveside of another plate on the positive silver foil.

Plates are always laid with adjacent sides of polarity asshown in Figure 2 a.

Top surface Is now negative. Continue process until allare stacked. Stack is completed with a negative crystal surface andfoil on top.

When stack is completed, place brass ring in position andrnake it secure with screws. Fill space around oith a material thatwill stand heat, to keep stack frcm sliding during the final poly-merization. Folded sheets of cellophane has been found to be usefulas a filler.

Cover stuck 'ith a sheet of cellophane and set a weighton it. -'.eight should fit into the inside diorzet,.r of ring. Dis-tribu;t.e veights eve-nly.

Let afr dry for an hour. Then place rg-'e vith 'eir.hts inan ol-en c'f a ce.:ta,. t:.:F,'r.t.,e of 6&O C. ( Y5O° v. ).

For a stack of hi -ry-tals of 1-3/1i" di=s.tier bake for 2

hours.

-8.

Page 12: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

Bake a longer or shorter time depending upon the area andheight of stack.

Remove from oven, remove ring and weights. Take care inremoving cellophane which may be stuck to top silver foil.

Cut excess M.B.M. from around stack with razor blade.

Clean with carbon tetrachloride all excess cement that can-not be cut.

Solder negative tabs to terminal lug.

Test leakage resistance from positive to negative tabs withvoltohmist. Should be infinite (> 109Rf])

4. Connecting Cable and Filling of Gage

The cable (British type) is inserted through the brass tubeinside of gage base. Use 3 feet of cable for each gage.

The following method is best for threading the cable throughthe tube:

Strip about 2- 1/2" of the rubber cover exposing shielding.Slip shielding back and bare inside lead for ]". Slide shield back inplace and use it as a guide for pulling cable through. Pull untilrubber covering of cable is all the way in the brass tube.

Shielded part of cable is now on gage base. Unweave shieldingthreads and bring out between ring and base to be soldered to outsideof ring for contact.

Strip and cut inside conductor to suitable length and solder

to positive silver foil.

Test terminals for infinite resistance.

File a little hollow in brass ring, enough to permit un-braided shielding to pass through. Screw ring in position and soldershielding to ring.

Prepare a copper foil diaphragm .002" in thickness and of thesame diameter as gage base. Cut a tab for contact to one side of theoutside of ring. Tin contact points.

Assembled gage is ready for filling, providing resistance isinfinite.

"e-

Page 13: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

If voltohmist shows low resistance, examine silver foilsaround edges to make sure there is no contact. If low resistance per-sists it may be due to humidity. Heating or placing gage in jar towhich Calcium Chloride (CaCL2) grains have been added will help.

Wax (G.E. Cable Joint Compound No. 227) is used to fill thegage.

Melt wax over gas burner. Heat gage slightly so that hotwax will flow better for filling.

Use toothpick. Start at one point and work around and around

layers until gage is filled a little above level of ring.

Watch for air bubbles while filling. Eliminate all.

Remove gage from heat and let cool a little before placingcopper diaphragm.

Lay diaphragm over gage and press out excess pitch withblock of wood over paper toweling. Apply even pressure and wipe offexcess. Repeat until all excess has been forced out and diaphragm islevel with ring. Allow to cool and test for infinite resistance.

5. Finishing Touches

Clean outside of gage with carbon tetrachloride or benzine.

Solder contact of diaphragm to ring.

To make air tight and prevent leaks apply adhesive (M 3) (4)

at cable entrance into tube. Force some of it into tube and paintaround cable. Allow to harden overnight. Two or three thin appli-cations are better than one thick one.

Place label made from copper sheeting stamped with numberof gauge about one foot avay from brass tubing on cable.

Gage is finished and ready for calibrating.

() Purchased from Mirnesota Mining 8; Mfg. Co.

-10-

Page 14: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

APPENDIX II

Photographs and Drawings

Fig. 1. APG No. A32901 - Gages at several stages of

construct ion.

Fig. 2 - Drawing BG 211 - Diagram of Gage

Fig. 3 - Drawing BG 210 - Tourmaline Blast Gage

-1l- I5

Page 15: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

CRY sT.4z t, :

tA 2 )'.T/2L PZ,9Tt

Rt..

Page 16: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

jt

///

Ia

a

Page 17: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

7-W- A- '.-. 1-

AA

Page 18: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

Ogle .41- ,,%/

*NRM FRVN .4.',X*VA4

H-4 4etC' RESEARCH AN DMOM. U

/5_ f- I " .. -III' \

'5, r APP Mm

' - IO

Page 19: TO - Defense Technical Information Center · (T~~TH-e metf "Z applying the stanca-d.-c-nrgo imil-ar to that descril'ed in OSRD Report No. 4619.-3-B. Consistency of Calibrations. Nine

AMSRL-CS-IO-SC (380) 1 $f7l' 2

MEMORANDUM FOR Chief, Technical Library, ATTN: Ms. L. LeTendre

SUBJECT: Distribution Statements for Ballistic ResearchLaboratory Reports

1. References:

a. Ballistic Research Laboratory Memorandum Report No. 397,"An Improved Tourmaline Air Blast Gage", by T. D. Carr and M. A.Bakinowski, October 1945, AD number 494667, UNCLASSIFIED,enclosed.

b. Ballistic Research Laboratories Memorandum Report No.1778, "Detonation Pressure Measurements in TNT and OCTOL", by R.Jameson and A. Hawkins, August 1966, AD number 802251,UNCLASSIFIED, enclosed.

c. Ballistic Research Laboratory Memorandum Report No. ARBRL-MR-03115, "Blast Computations over a Hemicylindrical AircraftShelter", by J. Wortman, July 1981, AD number B058960,UNCLASSIFIED, enclosed.

d. Ballistic Research Laboratory Memorandum Report No. ARBRL-MR-03125, "Combinatorial Geometry Computer Models of Sitting andStanding Crew Personnel" by L. R. 1 ruse and C; H. Lee, August1981, AD number B060185, UNCLASSIFIED, enclosed.

e. Ballistic Research Laboratories Report No. 734, "Responseof Air Blast Gauges of Various Shapes as a Function of PressureLevel", by S. T. Marks, August 1950, AD number 801219,UNCLASSIFIED, enclosed.

f. Ballistic Research Laboratories Report No. 775, "Responseof Air Blast Gauges of Various Shapes to One-Pound SphericalPentolite Charges as a Function of Pressure Level", by S. T.Marks, September 1951. AD number 801726, UNCLASSIFIED,. enclosed.

2. Subject area experts have reviewed the referenced reports andhave determined that they do not contain any information thatrequires limited distribution. Document release authorities haveapproved the reports for public release. This office will notifythe Defense Technical Information Center about the change in thedistributicn statements.